research papers 722 http://dx.doi.org/10.1107/S1600576716003873 J. Appl. Cryst. (2016). 49, 722–735 Received 10 November 2015 Accepted 7 March 2016 Edited by K. Chapman, Argonne National Laboratory, USA Keywords: limit of quantification; spiking method; high-energy laboratory X-ray powder diffraction. Supporting information:this article has supporting information at journals.iucr.org/j Accuracy in Rietveld quantitative phase analysis: a comparative study of strictly monochromatic Mo and Cu radiations L. Leo ´n-Reina, a M. Garcı ´a-Mate ´, b,c G. A ´lvarez-Pinazo, b,c I. Santacruz, b O. Vallcorba, d A. G. De la Torre b and M. A. G. Aranda b,d * a Servicios Centrales de Apoyo a la Investigacio ´n, Universidad de Ma ´laga, 29071-Ma ´laga, Spain, b Departamento de Quı ´mica Inorga ´nica, Cristalografı ´a y Mineralogı ´a, Universidad de Ma ´laga, 29071-Ma ´laga, Spain, c X-ray Data Services S.L., Edificio GREEN RAY, Primera Planta, Avenida Louis Pasteur 47 (Ampliacio ´n Campus Teatinos), 29010-Ma ´laga, Spain, and d ALBA Synchrotron, Carrer de la Llum 2-26, Barcelona, E-08290 Cerdanyola, Spain. *Correspondence e-mail: [email protected] This study reports 78 Rietveld quantitative phase analyses using Cu K 1 , Mo K 1 and synchrotron radiations. Synchrotron powder diffraction has been used to validate the most challenging analyses. From the results for three series with increasing contents of an analyte (an inorganic crystalline phase, an organic crystalline phase and a glass), it is inferred that Rietveld analyses from highenergy Mo K 1 radiation have slightly better accuracies than those obtained from Cu K 1 radiation. This behaviour has been established from the results of the calibration graphics obtained through the spiking method and also from Kullback–Leibler distance statistic studies. This outcome is explained, in spite of the lower diffraction power for Mo radiation when compared to Cu radiation, as arising because of the larger volume tested with Mo and also because higher energy allows one to record patterns with fewer systematic errors. The limit of detection (LoD) and limit of quantification (LoQ) have also been established for the studied series. For similar recording times, the LoDs in Cu patterns, 0.2 wt%, are slightly lower than those derived from Mo patterns, 0.3 wt%. The LoQ for a well crystallized inorganic phase using laboratory powder diffraction was established to be close to 0.10 wt% in stable fits with good precision. However, the accuracy of these analyses was poor with relative errors near to 100%. Only contents higher than 1.0 wt% yielded analyses with relative errors lower than 20%. 1. Introduction Most industrial materials are multiphase systems and the accurate determination of their phase assemblage is key to understanding their performance. The Rietveld method is nowadays the most employed methodology to achieve quantitative phase analysis (QPA) of crystalline materials in general (Madsen et al., 2001; Scarlett et al., 2002) and cements in particular (Leo ´n-Reina et al., 2009; Stutzman, 2005). These inter-laboratory comparisons gave some valuable recommendations for performing accurate Rietveld QPA (RQPA). The factors affecting the accuracy and precision of RQPA results can be gathered into three main groups: (i) instrumental; (ii) sample preparation; and (iii) data analysis protocol(s). The latter is related to the fact that every quantitative X-ray diffraction method requires the scaling of observed diffraction intensities with a suitable analytical standard. The Rietveld method is considered as a standardless methodology as it uses the crystal structure descriptions of each crystalline component to calculate its powder pattern. Consequently, the correct ISSN 1600-5767
choice of the crystal structure description for each phase in a multiphase system is key (Madsen et al., 2001; Zevin & Kimmel, 1995). The influence of the instrument type on the RQPA has been previously evaluated (Madsen et al., 2001); those authors concluded that both neutron and synchrotron (short-wavelength) powder diffraction yielded the best results, where the obtained values were the closest to the true ones. This was attributed mainly to the higher irradiated volumes, and also to the minimization of the microabsorption effects. Employing high-energy (short-wavelength) radiation allows us (i) to minimize absorption and microabsorption effects and (ii) to measure a higher number of Bragg peaks (useful mainly for structural studies). In addition, the use of short-wavelength X-rays enables an increase in the specimen irradiated volume. Molybdenum radiation combined with a flat sample in transmission geometry gave an irradiated volume of 100 mm 3 , while for copper radiation (flat sample in reflection geometry) the irradiated volume was 2mm 3 (Cuesta et al., 2015). In spite of these advantages, the angular resolution may be compromised when using X-rays with short wavelengths owing to the squeezing of the patterns. Consequently, the optics path must include appropriate elements [monochromator(s), slits, collimators etc.] to overcome severe peak overlap problems. It must also be noted that Mo radiation has a major drawback when compared to Cu radiation. The 3 dependence of diffraction intensity favours Cu diffraction by a factor of 10.2. So, a detector receives 10 times as many diffracted X-ray photons with Cu as with Mo (this calculation neglects the different fraction of photons lost in the diffractometer optical paths). This drawback could be partially overcome in modern X-ray detectors by increasing the counting time in Mo patterns without reaching prohibitively long values. In addition, sample preparation for RQPA is very important as the reproducibility of peak intensity measurements is governed by particle statistics (Elton & Salt, 1996). It is generally accepted that the diffraction intensities have to be collected with an accuracy close to 1% to obtain patterns suitable for good RQPA results (Dinnebier & Billinge, 2008). Particle statistics can be improved by (i) using short wavelengths as mentioned above, (ii) spinning the sample continuously during data collection and (iii) milling the sample to reduce the particle size, although this last approach should be executed with caution to avoid peak broadening or amorphization (Buhrke et al., 1998). Finally, another important issue in QPA of mixtures is the limit of detection and the limit of quantification. In this context, the limit of detection can be defined as the minimal concentration of analyte that can be detected with acceptable reliability in a sample (Zevin & Kimmel, 1995). ‘Acceptable reliability’ is a very elusive criterion as it depends upon the type of problem to be tackled. Madsen et al. (2001) assessed the limits of detection at the 1 wt% level; limits of quantification were not explicitly mentioned in that paper. Obviously, the limit of detection can be reduced (improved) by increasing the intensity of the X-ray source, for example, using synchrotron radiation. The aim of this study is to test a simple hypothesis: highenergy Mo radiation, combined with high-resolution laboratory X-ray powder diffraction optics, could yield more accurate RQPA, for challenging samples, than well established Cu radiation procedure(s). In order to do so, three sets of mixtures with increasing amounts of a given phase (spiking method) have been prepared and the corresponding RQPA results have been evaluated with calibration curves (leastsquares fit). Since the amorphous content of the single phases was unknown, the independent study of these mixtures does not allow the accuracy of the methodology to be established. The three designed series had increasing complexity. Firstly, a series of crystalline inorganic phase mixtures with increasing amounts of an analyte, from 0.12 to 4.0 wt%, was studied. This series does not represent a great challenge but it should allow us to determine if the Mo K 1 methodology is as robust as the well established Cu K 1 methodology. Secondly, a series of crystalline organic phase mixtures with increasing amounts of an organic compound, from 0.12 to 4.0 wt%, was analysed. This series was selected because of the challenge of working with low-absorbing samples that can result in transparency problems in reflection and inhomogeneous loading in narrow capillaries for transmission studies in diffractometers with parallel optics. This type of mixture is obviously of high interest in the pharmaceutical industry. Finally, a third series with variable amorphous ground glass content, from 0 to 32 wt%, was also studied. This is the most challenging work, as for the internal standard approach, the amorphous content is obtained from the small overestimation of the amount of analysed standard with respect to the weighed value. Any error in the procedure propagates to give large deviations in the derived amorphous content. Furthermore, the effect of preferred orientation on RQPA has been considered by including calcite and gypsum in the inorganic mixtures and lactose in the organic ones. Amorphous content determination is important in a number of industries including, but not restricted to, cements, glasses, pharmaceuticals and alloys. 2. Materials and methods 2.1. Materials Table 1 shows details of the single phases used in this work: d-(+)-glucose (99%), d-()-fructose (99%) and -lactose monohydrate (99%) from Sigma; d-(+)-xylose (>99%) and calcite (>99%) from Sigma–Aldrich; quartz (99.56%) from ABCR; zincite (99.99%) from Aldrich; micronized gypsum marketed by BELITH SPRL (Belgium). Insoluble anhydrite (i-A) was synthesized by heating the micronized gypsum at 973 K for 1 h in a furnace. All the mixtures were prepared by grinding the weighed phases by hand in an agate mortar for 20 min to ensure homogeneity. 2.1.1. Crystalline inorganic mixtures. A constant matrix of calcite (C), gypsum (Gp) and quartz (Q) was prepared. Then, six samples with known increasing amounts of i-A were produced, labelled as CGpQ_xA, where xstands for the target research papers J. Appl. Cryst. (2016). 49, 722–735 L. Leo ´n-Reina et al. Accuracy in Rietveld quantitative phase analysis 723
insoluble anhydrite content: 0.00, 0.12, 0.25, 0.50, 1.0, 2.0 and 4.0 wt%. 2.1.2. Crystalline organic mixtures. A constant matrix of glucose (G), fructose (F) and lactose (L) was prepared. Then six samples with known increasing amounts of xylose (X) were produced, labelled as GFL_xX, where xstands for the target xylose content: 0.00, 0.12, 0.25, 0.50, 1.0, 2.0 and 4.0 wt%. 2.1.3. Variable amorphous content within an inorganic crystalline phase matrix. A constant matrix of calcite (C) and zincite (Z) was prepared. Then five samples with increasing contents of amorphous ground glass (Gl), obtained by grinding a very thin optical glass plate by hand in an agate mortar for 30 min, were produced. The elemental composition of the ground glass, determined by X-ray fluorescence, was given by Garcı ´a-Mate ´et al. (2014). The amorphous content was determined by adding 20 wt% of quartz (Q) as an internal standard. The mixtures were labelled as CZQ_xGl, where xstands for 0, 2, 4, 8, 16 and 32 wt% of ground glass. 2.2. Analytical techniques 2.2.1. Laboratory X-ray powder diffraction. All single phases and mixtures were studied with both Mo K 1 (transmission geometry, trm) and Cu K 1 (reflection geometry, rfl) strictly monochromatic radiations. RQPA was performed for all the patterns to obtain the phase assemblages. Mo K 1 powder patterns were collected in transmission geometry (/), in constant irradiated volume mode, on a D8 ADVANCE (Bruker AXS) diffractometer (188.5 mm radius) equipped with a Johansson Ge(111) primary monochromator, which gives strictly monochromatic Mo radiation (= 0.7093 A ˚). The X-ray tube worked at 50 kV and 50 mA. The optics configuration was a fixed divergence slit (2 mm) and a fixed diffracted beam anti-scatter slit (9 mm). The energydispersive linear detector LYNXEYE XE (500 mm), optimized for high-energy radiation, was used with the maximum opening angle. Under these conditions the samples were measured between 3 and 35(2) with a step size of 0.006and with a total measurement time of 3 h and 5 min. Cu K 1 powder patterns, for exactly the same samples, were recorded in reflection geometry (/2)onanX’PertMPD PRO (PANalytical BV) diffractometer (240 mm radius) using strictly monochromatic Cu K 1 radiation (= 1.54059 A ˚) obtained by a Ge(111) primary monochromator. The X-ray tube worked at 45 kV and 40 mA. The optics configuration was a fixed divergence slit (1/2), a fixed incident beam antiscatter slit (1), a fixed diffracted beam anti-scatter slit (1/2) and an X’Celerator RTMS (real-time multiple strip) detector, working in the scanning mode with the maximum active length. Under these conditions the samples were measured between 6.5 and 81.5(2) with a step size of 0.0167and with a total measurement time of 2 h and 36 min. For Mo K 1 transmission geometry, samples were placed into cylindrical holders between two Kapton foils (Cuesta et al., 2015). The absorption factor of each sample was experimentally measured by comparing the direct beam with and without the sample (Cuesta et al., 2015). The amount of sample loaded (which determines the height of the cylinder) in the holders was adjusted to obtain a total absorption t’1, which corresponds to an absorption factor of 2.7 or 63% of direct beam attenuation. For the organic samples this criterion was not followed as it would lead to very thick specimens. In this case, the maximum holder thickness was used (1.7 mm). For Cu K 1 reflection geometry, the flat samples were prepared by rear charge of the flat sample holder in order to minimize preferred orientation. In both diffractometers, all the samples were rotated at 10 r min 1 during data collection. The lowest analyte content samples, CGpQ_0.12A and GFL_0.12X, were measured three times using both radiations, Mo K 1 and Cu K 1 . Regrinding and reloading of the mixtures in the sample holder were carried out prior to every measurement. Table 1 also reports the X-ray linear absorption coefficients for all the phases as microabsorption is always a concern in Rietveld X-ray quantitative phase analyses. 2.2.2. Transmission synchrotron X-ray powder diffraction. The powder patterns of the lowest analyte content samples, CGpQ_0.12A and GFL_0.12X, were also measured using synchrotron radiation. Synchrotron X-ray powder diffraction (SXRPD) data were collected in Debye–Scherrer (transmission) mode using the diffractometer of the ALBA light source (Fauth et al., 2013). The wavelength, = 0.77439 (2) A ˚, was selected with a double-crystal Si(111) monochromator and determined using the Si640d NIST standard (a= 5.43123 A ˚). The diffractometer is equipped with a MYTHEN-II detector system. The samples were loaded in glass capillaries of 0.7 mm diameter and rotated during data collection to improve diffracting particle statistics. The data acquisition time was 20 min per pattern to attain a very good signal-to-noise ratio (S/N) over the angular range 1–35(2). Three patterns, taken research papers 724 L. Leo ´n-Reina et al. Accuracy in Rietveld quantitative phase analysis J. Appl. Cryst. (2016). 49, 722–735 Table 1 Cambridge Structural Database/Inorganic Crystal Structure Database (CSD/ICSD) reference codes for all phases used for Rietveld refinements in this work and linear absorption coefficients for all the used wavelengths. (cm 1 ) Phases CSD/ICSD refcode Cu K 1 = 1.5406 A ˚Mo K 1 = 0.7093 A ˚= 0.7744/ 0.4959 A ˚ Glucose 1 Glucsa10 12 1 1.3/– Fructose 2 Fructo11 12 1 1.4/ – -Lactose monohydrate 3 Lactos10 12 1 1.3/ – Xylose 4 Xylose 12 1 1.2/ – Gypsum 5 151692 141 16 22/ – Quartz 6 41414 92 10 11/ 2.9 s-Anhydrite 7 16382 219 24 31/ – i-Anhydrite 8 79527 219 24 31/ – Zincite 9 65120 285 244 –/89.1 Calcite 10 80869 194 22 27/ 7.3 SrSO 411 22322 299 187 40/ – References: (1) Brown & Levy (1979); (2) Kanters et al. (1977); (3) Fries et al. (1971); (4) Hordvik et al. (1971); (5) De la Torre et al. (2004); (6) Will et al. (1988); (7) Kirfel & Will (1980); (8) Bezou et al. (1995); (9) Albertsson et al. (1989); (10) Maslen et al. (1995); (11) Garske & Peacor (1965); CSD: http://www.ccdc.cam.ac.uk/solutions/csd-system/ components/csd/; ICSD: http://www2.fiz-karlsruhe.de/icsd_home.html.
at different positions along the capillaries, were collected for each sample. SXRPD data for the amorphous content series, CZQ_xGl, were also measured at ALBA. The experimental setup was the same as described just above but the working wavelength was = 0.49591 (2) A ˚. 2.2.3. Data analysis. The powder patterns for all the samples were analysed by the Rietveld method as implemented in the GSAS software package (Larson & Von Dreele, 2000) by using a pseudo-Voigt peak shape function (Thompson et al., 1987) with the asymmetry correction included (Finger et al., 1994) to allow RQPA. The refined overall parameters were phase scale factors, background coefficients (linear interpolation function), unit-cell parameters, zero-shift error, peak shape parameters and preferred orientation coefficient, when needed. The March–Dollase preferred orientation adjustment algorithm was employed (Dollase, 1986). The modelling direction must be given as input for the calculations. In this case, the directions for the different phases were taken from previous studies. Alternatively, this direction is extracted from the pattern from the analysis of the differences between observed and calculated intensities for non-overlapped diffraction peaks. The crystal structure descriptions used in this study are reported in Table 1. In order to provide a single numerical assessment of the performance of each analysis, a statistic based on the Kullback–Leibler distance (KLD) has been employed (Kullback, 1968). This approach was previously used to evaluate the accuracy of RQPA applied to standard mixtures (Madsen et al., 2001; Scarlett et al., 2002; Leo ´n-Reina et al., 2009). Both phase-related KLD distances and absolute values of the Kullback–Leibler distance (AKLD) have been calculated. Accurate analyses are mirrored in low values of AKLD. 2.2.4. Amorphous content determination. The overall amorphous content was determined from the internal standard methodology approach (De la Torre et al., 2001; Aranda et al., 2012). Quartz was used as internal standard. If the original sample contains an amorphous phase, the standard will be overestimated in the RQPA. From the (slight) overestimation of the standard, the amorphous content of the investigated sample is derived (De la Torre et al., 2001). 2.2.5. Scanning electron microscopy characterization.All single-phase samples were characterized in terms of particle size by scanning electron microscopy (SEM; JEOL JSM 840, Tokyo, Japan). The powders were gold sputtered prior to SEM observation for better imaging. In addition to the as-received gypsum powder, and for the sake of comparison, a second gypsum powder sample, viz. a gypsum single crystal (Ma ´laga, Spain) that had been ground in an agate mortar for 10 min, was also characterized by SEM. 3. Results and discussion 3.1. Crystalline single phases All the single phases were selected according to several parameters, such as purity, particle size of the powder, preferred orientation and relevance for selected applications. All the phases were previously studied with Mo K 1 in order to check the suitability of the used crystal structures (see Table 1). These preliminary studies were of special interest for organic phases as the CIFs obtained from the CSD did not contain the atomic displacement parameters. 1 For lactose and fructose, the atomic displacement parameters were obtained from the reported data in the original publication and introduced manually in the GSAS control file. However, for glucose and xylose phases, these values were not reported in the original publications. Consequently, three groups of isotropic atomic displacement parameters were refined for glucose and xylose: 0.01 A ˚ 2 as starting value for carbon, hydrogen and oxygen atoms. Table 2 reports the final atomic displacement parameters for glucose and xylose, as well as R F values before and after their optimization, showing the improvements of the fits. The values reported in Table 2 were obtained from the fits to the Mo K 1 patterns for the single phases. In the RQPA of the organic mixtures, the atomic displacement parameters were kept fixed to these values. Preferred orientation was modelled by the March–Dollase algorithm along the [001] axis for both glucose and lactose. Table S1, deposited as supporting information, includes final refined profile function parameters and preferred orientation parameters for all the Mo K 1 refinements for organic single phases. These values were used as starting data to perform the refinements of the organic mixtures. Final Rietveld plots of the four Mo K 1 patterns for the organic single phases are given as supporting information in Figs. S1–S4. Since microparticle sizes and the distribution of different phases may allow us to explain some sample-related effects, such as preferred orientation, microabsorption and ‘graininess’, all powders were characterized by SEM. Fig. 1 shows SEM micrographs for all the organic single phases. The inorganic phases were also analysed by SEM and Mo K 1 . Fig. 2 reports micrographs for all inorganic phases. Table S2 includes the refined profile function parameters obtained for the inorganic phases analysed with Mo K 1 radiation. Final Rietveld plots for quartz, calcite, insoluble anhydrite and zincite are given as supporting information in Figs. S5–S8. All inorganic samples were single phases except gypsum and insoluble anhydrite. The gypsum sample used in this work was selected because of its small and homogeneous particle size. Fig. S9 shows SEM micrographs of the gypsum research papers J. Appl. Cryst. (2016). 49, 722–735 L. Leo ´n-Reina et al. Accuracy in Rietveld quantitative phase analysis 725 Table 2 Refined atomic displacement parameters for glucose (G) and xylose (X) single phases from the Rietveld fit of the Mo K 1 pattern. Atomic displacement parameters (A ˚ 2 )U iso =0.01A ˚ 2 U iso (refined) Phase C H O R wp (%) R F (%) R wp (%) R F (%) Glucose 0.013 (1) 0.041 (4) 0.0241 (6) 6.0 3.4 5.6 2.8 Xylose 0.023 (1) 0.059 (7) 0.0234 (6) 7.3 5.5 6.9 4.9 1 This is important as unaware researchers/analysts downloading CIFs from the CSD lose the atomic displacement parameter values with the current software.
powder used in this study (Fig. S9a) and a ground gypsum single crystal (Fig. S9b). The latter shows an inhomogeneous particle size distribution and quite large particle sizes, and therefore it was not used. The selected gypsum powder shows a more homogeneous particle size distribution; it contained, as minor phases, 2.25 (4) wt% of soluble anhydrite (s-A) and 1.13 (4) wt% of SrSO 4 , which were considered in the analysis of the mixtures. Table S3 shows the full phase assemblage of the used gypsum powder from the Mo K 1 RQPA. Table S3 includes refined/used profile function parameters for all the phases. Fig. S9calso shows Rietveld plots for gypsum collected with Mo K 1 radiation. Both organic and inorganic phases were also measured by using Cu K 1 radiation in reflection mode. The profile parameters were adjusted and preferred orientation was modelled as in the Mo K 1 patterns. The transparency effect of light compounds was observed in the Cu K 1 patterns for organic samples, as expected (Buhrke et al., 1998). Fig. S10 shows raw Mo and Cu K 1 patterns for glucose, as an example, to highlight the transparency effect. The peaks in the Cu K 1 patterns show a strong left-peak asymmetry and some are split, making them relatively difficult to fit. 3.2. Limit of detection and quantification The limit of detection (LoD) and limit of quantification (LoQ) are two important quantities in any analytical method validation. They have not been widely investigated in powder diffraction but they have been thoroughly used and discussed in the context of analytical measurements of drugs and pharmaceutical compounds [see for instance the review by Shrivastava & Gupta (2011)]. LoD/LoQ are terms used to describe the smallest concentration of an analyte that can be reliably detected/measured by an analytical procedure. The ‘reliability’ criterion is flexible and may be defined by the regulatory agencies, mainly the case for active pharmaceutical ingredients. In powder diffraction studies, the LoD for an analyte within a heterogeneous sample can be defined as the minimum amount of the analyte yielding a powder pattern with its strongest (not overlapped) diffraction peak with an S/N larger than 3.0. For techniques such as Rietveld analysis where the full powder pattern is evaluated, this approach is not straightforward. In this context, the LoQ can be defined as the minimum content of an analyte that can be determined with a value at least three times larger than its associated standard deviation and determined to an acceptable reliability level. For RQPA, this type of approach is straightforward although the accuracy for the very low content phases may be quite poor. Finally, and although it may seem counterintuitive, the Rietveld method applied to overlapped powder diffraction patterns may lead to a lower limit of quantification (for the full pattern) than the measured limit of detection, which is (currently) based on single-peak studies. Fig. 3 shows Mo K 1 and Cu K 1 raw patterns for the inorganic series with increasing amounts of insoluble anhydrite (highlighted with solid squares). Fig. 4 shows the strongest diffraction peak for i-A in the mixtures containing 0.12 wt% of anhydrite, CGpQ_0.12A, and 0.25 wt% of anhydrite, CGpQ_0.25A, to evaluate the limits of detection in the conditions reported in x2. For CGpQ_0.12A, both laboratory powder patterns yielded peaks with S/N lower than 3.0 (see top panels in Fig. 4). For CGpQ_0.25A, its Cu K 1 pattern yielded a clear peak with S/N = 4.1, so it can be concluded that the LoD for insoluble anhydrite, with this radiation in this research papers 726 L. Leo ´n-Reina et al. Accuracy in Rietveld quantitative phase analysis J. Appl. Cryst. (2016). 49, 722–735 Figure 2 Scanning electron microscopy micrographs for the studied inorganic phases (1000). The inset of the zincite micrograph shows the powder at higher magnification (20 000). Figure 1 Scanning electron microscopy micrographs for the studied organic phases (1000).
mixture, is very close to 0.2 wt%. For Mo K 1 radiation, the CGpQ_0.25A and CGpQ_0.50A samples yielded patterns with peaks having S/N = 2.4 and 5.1. Therefore, it can be concluded that the LoD for i-A, with this radiation in this mixture, is close to 0.3 wt%. The LoQ for i-A in this matrix was also studied. We chose to investigate the sample with the lowest anhydrite content to check the influence of using the full powder pattern, although this phase could not be reliably detected by analysing its strongest peak. Three Mo K 1 patterns and three Cu K 1 patterns were collected for CGpQ_0.12A. The RQPA results for these analyses, allowing the variation of just the phase scale factor, are reported here. For the three Mo K 1 patterns, the analysis results for i-A were 0.28 (3), 0.26 (2) and 0.29 (2) wt%. So, the anhydrite content could be quantified, yielding 0.28 (2) wt%, but the accuracy of the obtained value is poor, since the expected value was 0.12 wt%. Similarly, the results for the analyses of the three Cu K 1 patterns were 0.22 (3), 0.25 (3) and 0.26 (3) wt%, the average value being 0.24 (2) wt%. Full RQPA results are given in Table S4. Therefore, i-A can be quantified in this mixture at the level of 0.12 wt% but with a relative error close to 100%. If the ‘acceptable reliability’ criterion in the analysis were taken into account then the LoQ value would be close to 1.0 wt% as the relative associated error would be lower than 20% (see Table 3). CGpQ_0.12A was also studied by SXRPD. Fig. 3(c) shows SXRPD patterns collected in three different positions of the capillary, these patterns being almost identical. The main diffraction peak of anhydrite was clearly observed for this sample (see Fig. 4 bottom left). The S/N for the strongest diffraction peak of anhydrite was 12.8, and so the limit of detection for i-A, with synchrotron radiation in this matrix, is well below 0.10 wt%. Moreover, Table S5 gives the RQPA results obtained from the three patterns, with very little deviation for all the phases. The average phase assemblage was 34.0 (1) wt% of calcite, 31.4 (4) wt% of gypsum, 33.6 (4) wt% of quartz and 0.20 (1) wt% of insoluble anhydrite. As expected, the accuracy in the SXRPD analysis was better than that attained using laboratory radiation. To quantify the accuracy of the analysis results shown in Tables S4 and S5, the KLD methodology has been applied. Tables S4 and S5 also report the AKLD values for each analysis as well as the KLD values for the i-A phase in the analysis. The AKLD values for the synchrotron, Mo and Cu radiation analyses are 0.024, 0.031 and 0.057, respectively, these numbers being the average of the results from the three independent analyses for each radiation. The synchrotron analysis is indeed better than the laboratory radiation analyses. Furthermore, the Mo K 1 radiation analysis is better than the Cu K 1 one. Fig. 5 shows Mo K 1 and Cu K 1 raw patterns of the organic mixtures with increasing amounts of analyte (in this case xylose). The strongest powder diffraction peak for xylose was not observed in the GFL_0.12X patterns (both Mo and Cu ones). The corresponding peak was observed in the GFL_0.25X patterns. So, the LoD can be established to be close to 0.25 wt%. The analysis results for xylose in GFL_0.25X (see Table 4) were 0.33 (4) and 0.57 (9) wt% for the Mo K 1 and Cu K 1 patterns, respectively. These values showed that the results for Mo K 1 were slightly more accurate. research papers J. Appl. Cryst. (2016). 49, 722–735 L. Leo ´n-Reina et al. Accuracy in Rietveld quantitative phase analysis 727 Figure 3 (a) Raw Mo K 1 powder patterns for the inorganic series composed of a constant matrix of calcite, gypsum and quartz and increasing amounts of insoluble anhydrite (peaks highlighted with a solid square). (b) Raw Cu K 1 powder patterns for the same inorganic series. (c) Raw SXRPD patterns for CGpQ_0.12A collected at three different positions of the capillary (red, black and blue traces, almost overlapping).
The LoQ for xylose was also studied. Three Mo K 1 patterns and three Cu K 1 patterns were collected for GFL_0.12X. The analyses of the three Mo patterns gave 0.17 (5), 0.27 (5) and 0.11 (5) wt% of xylose, and so the research papers 728 L. Leo ´n-Reina et al. Accuracy in Rietveld quantitative phase analysis J. Appl. Cryst. (2016). 49, 722–735 Figure 4 Selected region of the powder patterns showing the main diffraction peak of insoluble anhydrite for the low-content samples used to investigate the limit of detection. Top left: Cu K 1 pattern for CGpQ_0.12A. Intermediate left: Cu K 1 pattern for CGpQ_0.25A. Bottom left: SXRPD pattern for CGpQ_0.12A. Top right: Mo K 1 pattern for CGpQ_0.12A. Intermediate right: Mo K 1 pattern for CGpQ_0.25A. Bottom right: Mo K 1 pattern for CGpQ_0.50A. The main peak of anhydrite, sin()/= 0.143 A ˚ 1 , is located at 25.4, 11.6 and 12.72for Cu K 1 ,MoK 1 and synchrotron radiations, respectively. The peak at sin()/= 0.1445 A ˚ 1 is due to the soluble anhydrite coming from gypsum (constant content in all the samples). The very tiny peak at sin()/= 0.1457 A ˚ 1 , slightly visible only in the SXRPD pattern, arises from SrSO 4 coming also from gypsum.
average value was 0.18 (8) wt%. Similarly, the results for the analyses of the three Cu patterns were 0.35 (9), 0.28 (10) and 0.40 (8) wt%, the average value being 0.34 (6) wt%. Full RQPA results are included in Table S6. Therefore, the LoD for xylose in this mixture for the two radiations can be established to be close to 0.12 wt%. If one applied an ‘acceptable reliability’ criterion, the LoQ would be much higher, above 1 wt%. Finally, the output using Cu K 1 radiation is less accurate than that obtained from Mo K 1 data, although both values were overestimated. GFL_0.12X was also studied by SXRPD in a rotating capillary in transmission. Fig. 5(c) shows SXRPD patterns for GFL_0.12X collected at three different positions in the same capillary. The powder patterns showed different peak ratios. It is known that filling a capillary with (some) organic compounds is not easy owing to electrostatic charge effects. Furthermore, the phase ratio within the part of the capillary bathed by the X-rays must be the same as that of the sample under study, which cannot be ensured under these circumstances. The differences between the patterns in Fig. 5(c) can be explained by this effect, which results in variable RQPA for the powder patterns of this sample (reported in Table S7). The mean values and the standard deviations of the three analyses, for the three positions, were 35 (4) wt% of glucose, 30 (2) wt% of fructose and 35 (3) wt% of lactose. For samples displaying this behaviour, SXRPD based on glass capillaries is clearly not suitable for obtaining accurate RQPA results. Selfsupported sample preparation and other types of holders/ capillaries are currently under investigation for pharmaceutical compounds at ALBA, but the results of this ongoing optimization are out of the scope of the present paper. 3.3. Increasing inorganic crystalline phase content series Table 3 reports the RQPA results for the six inorganic mixtures with increasing amounts of i-A measured with Mo K 1 (in transmission) and Cu K 1 (in reflection). In general, the values obtained from both radiations are similar. Fig. 6 displays the Rietveld plots of the mixture with 4 wt% of i-A measured with the two radiations. It must be noted that the gypsum contained soluble anhydrite and SrSO 4 , which were taken into account to calculate the phase assemblage reported in Table 3. The AKLD values and the KLD values for the i-A phase are also reported in Table 3. The AKLD research papers J. Appl. Cryst. (2016). 49, 722–735 L. Leo ´n-Reina et al. Accuracy in Rietveld quantitative phase analysis 729 Table 4 RQPA for the crystalline organic mixtures measured with Cu K 1 and Mo K 1 radiations. Weighed amounts (%Wt) are shown for the sake of comparison (in bold). The AKLDs for each mixture and the KLD values for xylose are also included. GFL_0.0X GFL_0.25X GFL_0.50X GFL_1.0X GFL_2.0X GFL_4.0X Phases %Wt Mo trm Cu rfl %Wt Mo trm Cu rfl %Wt Mo trm Cu rfl %Wt Mo trm Cu rfl %Wt Mo trm Cu rfl %Wt Mo trm Cu rfl G33.4 33.8 (1) 33.5 (3) 33.3 33.6 (1) 33.1 (2) 33.2 32.3 (2) 33.5 (2) 33.0 34.7 (1) 33.6 (2) 32.7 32.2 (1) 31.5 (2) 32.0 32.8 (1) 33.6 (2) F33.5 31.7 (1) 32.7 (3) 33.4 32.3 (1) 34.3 (2) 33.3 32.1 (2) 33.4 (2) 33.1 32.6 (1) 33.7 (2) 32.8 31.7 (1) 34.4 (2) 32.2 30.7 (1) 32.5 (2) L33.1 34.5 (1) 33.7 (3) 33.0 33.7 (1) 32.0 (2) 33.0 35.0 (3) 32.5 (2) 32.8 31.6 (2) 31.4 (2) 32.5 34.3 (1) 32.0 (2) 31.8 32.9 (1) 30.5 (2) X–––0.27 0.33 (4) 0.57 (9) 0.55 0.53 (8) 0.61 (9) 1.1 1.10 (5) 1.3 (1) 2.0 1.76 (5) 2.1 (1) 3.9 3.70 (5) 3.4 (2) AKLD sum 0.0362 0.0150 0.0216 0.0231 0.0410 0.0096 0.0338 0.0280 0.0363 0.0339 0.0361 0.0372 (X) KLD – – 0.001 0.002 0.000 0.001 0.000 0.002 0.003 0.001 0.002 0.005 Table 3 Rietveld quantitative phase analyses for the crystalline inorganic mixtures measured with Cu K 1 and Mo K 1 radiations. Weighed amounts (%Wt) are shown for the sake of comparison (in bold). The AKLDs for each mixture and the KLD values for i-anhydrite are also included. CGpQ_0.0A CGpQ_0.25A CGpQ_0.50A CGpQ_1.0A CGpQ_2.0A CGpQ_4.0A Phases %Wt Mo trm Cu rfl %Wt Mo trm Cu rfl %Wt Mo trm Cu rfl %Wt Mo trm Cu rfl %Wt Mo trm Cu rfl %Wt Mo trm Cu rfl C32.9 32.6 (1) 30.4 (2) 32.8 32.0 (1) 33.6 (1) 32.7 33.2 (1) 32.8 (1) 32.5 32.8 (1) 32.6 (2) 32.2 31.3 (1) 31.4 (1) 31.6 31.2 (1) 31.8 (1) Gp 31.7 31.7 (1) 34.5 (1) 31.7 32.5 (1) 31.6 (1) 31.6 30.1 (1) 30.7 (1) 31.5 30.4 (1) 30.7 (1) 31.1 32.1 (1) 32.3 (1) 30.5 30.7 (1) 30.5 (1) Q34.2 34.6 (1) 33.7 (1) 34.1 33.9 (1) 33.0 (1) 34.0 34.6 (1) 34.2 (1) 33.8 34.1 (1) 33.8 (1) 33.5 33.5 (1) 32.6 (1) 32.8 32.8 (1) 32.0 (1) s-A 0.8 0.66 (3) 0.76 (5) 0.8 0.77 (4) 0.78 (5) 0.8 0.97 (3) 1.15 (5) 0.8 1.03 (4) 1.11 (5) 0.7 0.54 (3) 0.58 (5) 0.7 0.67 (3) 0.77 (4) SrSO 4 0.4 0.44 (4) 0.70 (6) 0.4 0.44 (4) 0.67 (5) 0.4 0.39 (4) 0.56 (5) 0.4 0.43 (4) 0.68 (5) 0.4 0.48 (4) 0.68 (6) 0.4 0.45 (4) 0.63 (5) i-A – – – 0.28 0.42 (3) 0.42 (4) 0.52 0.71 (3) 0.71 (4) 1.02 1.23 (3) 1.17 (5) 2.02 2.05 (4) 2.38 (9) 4.02 4.30 (8) 4.33 (9) AKLD sum 0.0089 0.0605 0.0198 0.0235 0.0295 0.0180 0.0214 0.0152 0.0218 0.0358 0.0095 0.0156 (i-A) KLD –0.001 –0.001 –0.002 –0.002 –0.002 –0.001 0.000 –0.003 –0.004 –0.003 Table 5 R F factors of the crystalline phases for the inorganic (CGpQ_4.0A) and the organic (GFL_4.0X) mixtures with 4.0 wt% of the minor phase. CGpQ_4.0A† GFL_4.0X‡ Mo K 1 Cu K 1 Mo K 1 Cu K 1 R F (C) (%) 3.8 2.8 R F (G) (%) 1.7 4.5 R F (Gp) (%) 2.5 2.9 R F (F) (%) 2.1 4.6 R F (Q) (%) 1.5 1.3 R F (L) (%) 1.7 4.2 R F (s-A) (%) 6.3 6.3 R F (X) (%) 2.2 5.6 R F (i-A) (%) 2.6 2.6 R F (SrSO 4 )(%) 5.2 3.2 †The R wp values for the Mo K 1 and Cu K 1 patterns were 6.8 and 8.2%, respectively. ‡ The R wp values for the Mo K 1 and Cu K 1 patterns were 5.1 and 13.2%, respectively.
values from Mo K 1 radiation for most of the samples are slightly smaller than the corresponding ones obtained from Cu K 1 radiation (see Table 3). Hence, we can conclude that the Mo K 1 analyses are slightly better than those derived from Cu K 1 . As discussed above, the investigated samples are exactly the same and their structural descriptions are also identical. Hence, the phase-dependent R F agreement factors give an indication of the quality of the data. As an example, Table 5 reports the R F values of all phases for the sample with 4 wt% of insoluble anhydrite. The R F values obtained for both patterns are good and quite similar, indicating that both data sets have reproducible peak diffraction intensities. Furthermore, calcite and gypsum presented preferred orientation, the axis being [104] and [010], respectively. This effect was modelled by using the March–Dollase algorithm (insets in Fig. 6). Preferred orientation causes the 00lreflections for gypsum to have higher intensities in the Cu K 1 patterns than those calculated from its crystal structure. On the other hand, these reflections in the Mo K 1 patterns have smaller intensities than those derived from the gypsum structure (see Fig. 6 top). As a consequence, the refined values for flat samples in reflection and transmission geometries were smaller and larger than 1.0, respectively (Cuesta et al., 2015). For the mixture with 4 wt% of i-A, as a representative example, the optimized coefficients were 0.815 (2) and research papers 730 L. Leo ´n-Reina et al. Accuracy in Rietveld quantitative phase analysis J. Appl. Cryst. (2016). 49, 722–735 Figure 6 Selected range of the Rietveld plots for CGpQ_4.0A: (a)MoK 1 and (b) Cu K 1 patterns. The insets highlight the effect of preferred orientation for gypsum and calcite. Figure 5 (a) Raw Mo K 1 powder patterns for the organic series composed of a constant matrix of glucose, fructose and lactose and increasing amounts of xylose (peaks highlighted with an asterisk). (b) Raw Cu K 1 powder patterns for the same organic series. (c) Raw SXRPD patterns for GFL_0.12X collected at three different positions of the capillary.