scieee Science in your language
[en] (orig)
Laboratory based GIXRF and GEXRF spectrometers
for multilayer structure investigations
Veronika Szwedowski-Rammert, *
a
Jonas Baumann,
a
Christopher Schlesiger,
a
Ulrich Waldschl¨
ager,
b
Armin Gross,
b
Birgit Kanngießer
a
and Ioanna Mantouvalou
a
This work reports laboratory angle resolved measurements with the goal of establishing laboratory
techniques to obtain a more complete idea of the intralayer composition of multilayer samples. While X-
ray reectometry is a widely available technique for the characterization of multilayer samples, angle
resolved XRF measurements (grazing emission/incidence X-ray uorescence) are usually performed at
synchrotron radiation facilities. With the development of ecient laboratory spectrometers and
evaluation algorithms for angle resolved measurements, these methods become suited for routine
measurements and screening. We present two laboratory spectrometers which make quantitative non-
destructive elemental depth proling feasible. For reasons of comparison a validation sample, a nickel
carbon multilayer sample, is measured with both setups and additional information on krypton
contamination and its distribution is retrieved. Additionally, the rst application for the characterization of
multilayer structures with sub-nanometer layer thicknesses is shown.
Introduction
In the non-destructive characterization of multilayer structures,
the X-ray reectometry (XRR) method is widely used in indus-
trial and scientic applications. In XRR measurements, X-ray
radiation is directed onto a sample at varying incidence
angles and the reected radiation is detected with a photo-
diode. The measurements can be performed at synchrotron
radiation facilities
1,2
as well as with benchtop instruments.
3,4
Due to the broad applications of XRR, several codes for the
quantication of the measurements exist, with IMD
5,6
being the
most popular. Though, as has been shown in ref. 7 and 8,
relying solely on XRR measurements for the characterization of
complex, stratied materials is insucient. The number of
degrees of freedom in the quantication rises rapidly with the
complexity of the sample, calling for complementary measure-
ments. A possible way of obtaining element specic in-depth
information from the sample non-destructively is the use of
angle resolved X-ray uorescence (XRF) methods.
9
As in XRR, X-
ray radiation is used, though not the reected radiation but the
element specicuorescence radiation emitted from the
sample is detected. This provides additional information about
the composition of the investigated sample since the elemental
distribution is measured. Contamination as well as intermixing
regions can be characterized. Several studies have already
combined XRR with grazing incidence XRF analysis.
1012
Grazing angles refer to the angular range from below the
angle of total reection up to a few multiples of the value of the
angle of total reection. Depending on the energy of the inci-
dent beam and the investigated material, it is the angular range
where interference eects from the samples are observable.
Angular resolution of XRF measurements can be achieved in
a twofold manner: by varying the shallow incidence angle of the
primary X-ray radiation on the sample (grazing incidence X-ray
uorescence: GIXRF) or by measuring the uorescence radia-
tion from the sample at distinct shallow emission angles
(grazing emission X-ray uorescence: GEXRF). Both methods
yield equivalent information about the sample as was shown in
ref. 13.
A benecial phenomenon for analysis occurring in both
angle resolved measurements when the interfaces of the layers
are smooth is an enhancement of excitation due to the forma-
tion of X-ray standing waves (XSW) within the sample. In the
case of GIXRF the incident X-ray beam can interfere with the
beam reected at layer boundaries
14
and therefore constructive
and destructive interference nodes are formed. Essential for the
standing wave eld is a monochromatic primary beam with low
divergence. Contrarily, in GEXRF measurements the quality of
the primary beam is not crucial for interference to occur. In that
case, the probability density functions of the possible paths of
the uorescence radiation released by atoms within a layer can
interfere. This eect is known as the Kossel eect
15
and can be
observed under the grazing emission conditions in all
a
Technische Universit¨
at Berlin, Hardenbergstr. 36, 10623 Berlin, Germany. E-mail:
jonas.baumann@physik.tu-berlin.de
b
Bruker Nano GmbH, Am Studio 2D, 12489 Berlin, Germany
Cite this: J. Anal. At. Spectrom.,2019,
34,922
Received 5th December 2018
Accepted 4th February 2019
DOI: 10.1039/c8ja00427g
rsc.li/jaas
922 |J. Anal. At. Spectrom.,2019,34,922929 This journal is © The Royal Society of Chemistry 2019
JAAS
PAPER
Open Access Article. Published on 11 February 2019. Downloaded on 6/17/2019 10:33:20 AM.
This article is licensed under a
Creative Commons Attribution 3.0 Unported Licence.
View Article Online
View Journal
| View Issue
investigations with beams that excite uorescence radiation
from atoms of the sample (protons,
16
electrons,
15
and
photons
17,18
).
The result of both angle resolved techniques is an angular
prole where the intensity of a given uorescence line of
a specic element is shown for each measured angle. Since each
angle corresponds to a dierent penetration or information
depth for a given sample, a depth-resolved elemental prole can
be calculated. Additional conclusions about the sample may be
drawn from interference structures in the GI- or GEXRF prole
in the case of multilayer samples. The position of the structures
depends on the thickness of the layers of the sample. Rough-
ness and interdiusion between the layers decrease the ampli-
tude of the structures and a density variation shis the position
of the angle of total reection.
In the following, measurements of a validation sample with
a laboratory GEXRF and a commercial GIXRF setup and the
respective quantication routines are presented. The goal is to
ensure a reliable characterization of these laboratory setups,
which is crucial for the quantication of unknown samples
beyond synchrotron radiation measurements. The measure-
ments show the capabilities of these laboratory spectrometers,
their strengths, challenges and benets.
Laboratory setups for angle resolved XRF
Angle resolved XRF methods are dominantly applied at
synchrotron radiation facilities. The high ux of the X-ray beam
and a small focus size are benecial for fast and precise
measurements of X-ray uorescence. However, a rapid devel-
opment of laboratory equipment in recent years accompanied
by the improvement of ecient evaluation algorithms enables
angle resolved measurements in the laboratory with compa-
rable angular resolution and statistics.
19
This has the potential
to expand the availability of these methods for a broader range
of users. Especially in the development of new, more ecient
multilayer structures with high reectivity complementary
methods to XRR are benecial for the characterization of
interlayer diusion, contaminations and roughness.
Grazing incidence X-ray uorescence
In GIXRF measurements the incidence angle of the radiation on
the sample is varied in a range depending on the beam's energy
and the investigated sample's material(s). Most commonly,
a GIXRF scan starts below the angle of total reection and
surpasses it.
20
To benet from the interference eect enhance-
ment resulting from the formation of X-ray standing waves,
GIXRF measurements are usually performed with mono-
chromatized radiation either from an X-ray tube
11
or at
synchrotron radiation facilities.
7
The sample is tilted with
respect to the beam at dened angles and the uorescence
radiation is detected above the sample with an energy dispersive
detector, see Fig. 1. In most setups the detector is a silicon dri
detector, SDD. The achievable angular resolution is limited by
the divergence of the X-ray beam used and inuences the
intensity of the interference patterns. It is thus crucial to char-
acterize the divergence of a GIXRF setup to distinguish between
the eect it has on the prole and the damping resulting from
roughness or intermixing of the interfaces of the layers. With
the NANOHUNTER II from Rigaku and the BRUKER S4 T-STAR,
benchtop GIXRF setups have become commercially available.
Grazing emission X-ray uorescence
GEXRF measurements are typically performed with the primary
X-ray radiation incident on the sample at angles much larger
than the angle of total external reection. Two possible setups
with distinct congurations are established to detect the uo-
rescence radiation from the sample at grazing emission angles:
the conventional and the so-called scan-free approach. In the
former the radiation is detected with an SDD and the angular
resolution is achieved e.g. by placing a slit in front of the
detector.
13
The detector or sample is moved to measure the
angular prole. The angular resolution that can be achieved is
usually limited by the slit system. Additionally, the sensitivity of
such a setup is low due to the small solid angle of detection.
These limitations prompted the development of the scan-free
GEXRF (SF-GEXRF) setup,
21
where the slit-detection unit is
replaced with a 2D area detector. This facilitates the recording
of several angles in one measurement without the necessity to
move parts of the setup. This version of a GEXRF setup is shown
in Fig. 2. A crucial condition to perform energy dispersive SF-
GEXRF experiments is that all photons counted by the
detector in each frame are well separated from the adjacent
photons. If this is satised, the energy and position of the
photon detected by the CCD can be reconstructed. Thus, several
frames are recorded, with the number of frames depending on
the required statistics. To full the single photon condition
either the exposure time of each frame or (if possible) the ux of
the source can be regulated.
A signicant benet of GEXRF compared to GIXRF follows
from the investigation of samples at normal incidence in the
former. This enables the scanning of laterally inhomogeneous
samples, since the footprint on the sample is smaller than that
for grazing incidence conditions as the beam is not broadened
Fig. 1 Schematic representation of the GIXRF measurement geom-
etry. The incident beam of the primary radiation on the sample is varied
by either changing the position of the X-ray source or tilting the
sample relative to the beam. The uorescence radiation from the
sample is detected in a short distance above the surface with an SDD
detector.
This journal is © The Royal Society of Chemistry 2019 J. Anal. At. Spectrom.,2019,34,922929 | 923
Paper JAAS
Open Access Article. Published on 11 February 2019. Downloaded on 6/17/2019 10:33:20 AM.
This article is licensed under a
Creative Commons Attribution 3.0 Unported Licence.
View Article Online
by the shallow incidence angle. This was used to map impurities
in silicon wafers.
22
While the SF-GEXRF setup was rst
successfully introduced at a synchrotron radiation facility
21
it
could be transferred to the laboratory. Measurements were
performed in the soX-ray range
23
as well as with poly-
chromatic radiation in the hard X-ray range.
24
Materials and methods
For the validation of the laboratory angle resolved XRF tech-
niques a nickelcarbon (NiC) multilayer manufactured by AXO
DRESDEN is chosen. The carbon/nickel bi-layer structure is
repeated 15 times on a silicon substrate. From XRR measure-
ments performed by the manufacturer and evaluated with the
IMD soware package the bi-layer thickness is calculated to be
5.55 nm, with 2.42 nm for nickel and 3.13 nm for carbon. This
sample is well suited for validation purposes since a sharp
interference structure forms due to its smooth interfaces on the
X-ray wavelength length scale and neglectable interdiusion.
Laboratory grazing incidence X-ray uorescence
The GIXRF scans are performed with a commercial BRUKER S4
T-STAR. This setup was developed for total reection XRF
analysis of trace elements and a new feature allows the execu-
tion of angular scans. The molybdenum Ka-line (17.4 keV) is
used as the excitation radiation, which is monochromatized
and focused on the sample with parabolic graded multilayer
optics. The beam shape is calculated by measuring the FWHM
of the beam with a CCD camera at dierent distances from the
monochromator. It is determined to be Gaussian in the x-
direction with a FWHM of 100 10 mm and a rectangular
function in the y-direction with a width of 5.62 0.01 mm in the
focus. From the same measurement, beam divergence in the x-
direction is calculated to be FWHM ¼0.25 0.06 mrad. From
that the angular resolution of the GIXRF measurements can be
calculated to be FWHM ¼0.0140.003since beam diver-
gence is the limiting factor. The sample is placed on a robotic
arm and the uorescence radiation detected with an SSD
detector. The robotic arm positions the sample beneath the
SDD and tilts the sample relative to the X-ray beam about the y-
axis to perform a GIXRF scan. The SDD detector is also tilted
with the sample. Due to the shallow incidence angles the foot-
print on the sample changes during the measurement. In this
case this concerns the x-direction of the beam, thus only the
divergence in this axis is considered in calculations. The GIXRF
scan of the NiC multilayer sample is executed in an angular
range from 0up to 0.6with a detector real time of 10 seconds
and a step size of 0.001. Considering motor movements this
results in an overall measurement time of 125 minutes. In view
of the step size exceeding the limitations imposed by beam
divergence, measurements with a larger step size would be
appropriate. The divergence calculations were performed aer
the measurements, and thus this discrepancy. Including these
considerations an improvement of measurement time by
a factor of 10 is not only achievable in this case but also rational.
For each angular position a distinct spectrum is recorded and
Fig. 3 shows the sum spectrum of the whole measurement
depicted in blue. The energy resolution is 150 eV@Ni Ka.
Laboratory grazing emission X-ray uorescence
The laboratory scan-free GEXRF spectrometer for hard X-rays is
set up at the Berlin Laboratory for innovative X-ray technologies
(BLiX). An X-ray tube with a rhodium target operated at 50 kV
and 600 mA is equipped with a polycapillary lens to focus the
Fig. 2 Schematic representation of the SF-GEXRF measurement
geometry. The primary radiation is incident on the sample at a xed
angle, usually 90. The uorescence radiation from the sample is
detected with a position and energy sensitive detector. The geometry
of the setup remains xed throughout the measurement in which
several CCD frames are recorded. For each uorescence line a CCD
frame solely containing the events with the respective energy is
created in the evaluation. Here, the CCD frame for a Ni Kauores-
cence line is shown. The GEXRF prole is calculated from the CCD
frame by summing the events along isoangular lines as indicated by the
Ni Kaprole on the right.
Fig. 3 Sum spectra of the GI- and GEXRF measurements. The GIXRF
sum spectrum (blue) is generated from 600 individual spectra recor-
ded during a scan in the range 00.6with a step size of 0.001. For
the GEXRF sum spectrum (red) 2200 CCD frames were evaluated. The
elemental labels are indicated in red if the element can only be found
in the GEXRF sum spectrum, blue if solely the GIXRF spectrum shows
it, and black when the element is present in both sum spectra.
924 |J. Anal. At. Spectrom.,2019,34,922929 This journal is © The Royal Society of Chemistry 2019
JAAS Paper
Open Access Article. Published on 11 February 2019. Downloaded on 6/17/2019 10:33:20 AM.
This article is licensed under a
Creative Commons Attribution 3.0 Unported Licence.
View Article Online
radiation to the pivotal point of a vacuum chamber.
32
A back
illuminated deep-depletion CCD (BI Princeton PI-SX:1300) with
1340 1300 pixels, each with the dimensions of 20 mm20
mm, is used to detect the uorescence radiation from the
sample. The deep-depleted sensor of the CCD is a crucial
improvement of the setup compared to the rst proof-of-
principle measurements published in ref. 24, since hard X-
rays of up to 20 keV are detected. Throughout the measure-
ment the CCDs temperature was set to 45 C and was moni-
tored to ensure an unchanged dark current. To achieve an
angular resolution of below 0.005the CCD is mounted 57 cm
away from the sample. This resolution is required to resolve ne
interference structures in the GEXRF prole of the multilayer
(corresponding to Kiessig fringes in XRR) and can be calculated
from trigonometric considerations when the pixel size of the
CCD is known. For this specic measurement the covered
angular range is 2.2(01.4shown in Fig. 4 top) and the
angular resolution is 0.005. For the SF-GEXRF measurement
a total number of 2200 CCD frames and additionally 18 dark
frames are recorded. The exposure time is 3 seconds and a read-
out speed of 1 MHz is used. This results in an overall
measurement time of 160 minutes. The GEXRF frames are dark
frame corrected, which means that from each pixel a median
value calculated from the dark frames for the respective pixel is
subtracted. Furthermore, from the dark frames the standard
deviation of each pixel is calculated and used as a threshold to
evaluate the CCD frames. Thus, there is a necessity to ensure
a constant temperature throughout the measurement. The
clustering PEE algorithm
25
is applied and on average 600 events
are registered in one GEXRF frame. The sum spectrum of the
NiC multilayer is calculated and shown in Fig. 3 depicted in red.
The energy resolution is 306 eV@Ni Ka.
Comparison of the sum spectra of GI- and GEXRF
measurements
Both sum spectra show nickel, silicon and scattering peaks
from the primary radiation as expected. In the sum spectrum of
the laboratory SF-GEXRF measurement there are multiple lines
which do not originate from the sample (aluminum, chromium,
etc.) and which are missing in the GIXRF spectrum. The pres-
ence of these lines results from the primary radiation scattered
from the sample exciting uorescence radiation from the walls
of the experimental chamber. The small detectorsample
distance in the GIXRF setup prevents such artefacts in the
respective sum spectrum. On the other hand, the GIXRF sum
spectrum contains a prominent krypton peak that cannot be
assigned to artifacts originating from the setup. Krypton is used
as the sputtering gas in the magnetron sputtering process of
manufacturing the sample, and hence it is present in the
sample. In the SF-GEXRF measurement a CCD is used as
a detector and the quantum eciency of such detectors dras-
tically decreases for energies above 10 keV.
24
Thus, the statistics
of the SF-GEXRF measurement are insucient for the detection
of the krypton peak at 12.6 keV. Moreover, the polycapillary lens
in the GEXRF setup has a neglectable transmission above 20
keV, additionally limiting the excitation of krypton. The GIXRF
measurements are performed in air (which explains the pres-
ence of argon) and a beryllium window is in front of the
detector. This all together prevents the detection of low Z
elements and the L-line of nickel in the GIXRF experiment. The
fact that carbon could not be measured in the in-vacuum SF-
GEXRF experiment is a result of using hard X-rays as the exci-
tation radiation and additionally the prominent noise peak for
low X-ray energies. The lines not assigned to any element in
both sum spectra are either escape or pile-up peaks. Even
though the measurement time of the SF-GEXRF measurement
is longer, the GIXRF measurement has better statistics. This
results from the small solid angle of detection of the measure-
ment in the SF-GEXRF case. Improving this factor or more
ecient excitation of uorescence radiation would signicantly
improve the statistics of the SF-GEXRF measurement.
Angular GI- and GEXRF proles
From the measured spectra angular XRF proles for each
measurement and each relevant element are generated. In the
Fig. 4 Angle resolved elemental proles of the NiC multilayer. For the
SF-GEXRF measurements (top) only the Ni Kauorescence line prole
is shown. An angular resolution of 0.005enables the discrimination of
ne interference structures not visible in the respective GIXRF prole
(bottom). The GIXRF proles of the Ni Ka(green) and Si K (red) uo-
rescence lines are shown. The prole of silicon is multiplied by a factor
5 to display it in one plot with nickel.
This journal is © The Royal Society of Chemistry 2019 J. Anal. At. Spectrom.,2019,34,922929 | 925
Paper JAAS
Open Access Article. Published on 11 February 2019. Downloaded on 6/17/2019 10:33:20 AM.
This article is licensed under a
Creative Commons Attribution 3.0 Unported Licence.
View Article Online
case of the GIXRF measurement each spectrum recorded at
each respective angle is deconvolved with a code provided by the
manufacturer. The calculated intensity of the individual
elemental lines is plotted over the angular position of the cor-
responding measurement.
Contrarily, in the SF-GEXRF measurements only the Ni Ka
uorescence line prole is shown. This results from the absorp-
tion of the silicon uorescence line due to the extended paths
through the sample at low emission angles. The intensity of the
detected silicon signal starts rising at around 1.7.Noadditional
information can be extracted from the measured silicon prole
range in the SF-GEXRF case, consequently it is not shown or
discussed further. To generate the SF-GEXRF prole a region-of-
interest (ROI) around the Ni Kauorescence line peak is chosen
(7.077.69 keV) and the 2200 CCD frames aer dark frame
correction are searched for events within this ROI. A new CCD
frame is generated and at the x,y-position of each Ni Kauo-
rescence photon detection position in the recorded CCD frames
the value of the corresponding pixel in the new frame is increased
by one. As a result, a CCD image with the interaction positions of
all Ni Kauorescence photons is generated. This image is shown
in Fig. 2 on the CCD. To calculate a GEXRF prole from the CCD
image the pixels are summed along isoangular lines and each
pixel intensity is normalized to its solid angle of detection
resulting in a GEXRF prole as depicted next to the ROI image in
Fig. 2. The geometrical information necessary to perform these
calculations is gathered beforehand in a calibration routine.
24
These geometrical parameters are the distance between the
sample and the detector and the tilt of the CCD since the CCDs
position cannot be adjusted once mounted on the chamber.
Additionally, with these parameters the angular range covered in
one scan and the theoretical angular resolution of the measure-
ment can be calculated.
The resulting GIXRF proles of silicon and nickel are shown
in the bottom of Fig. 4, where the intensity of silicon is multi-
plied by a factor 5 to display the proles clearly in one plot. The
rapid increase of intensity above 0.1occurs when the angle of
total reection is surpassed and the signal from the depth of the
sample is detected. The prominent interference structure with
a minimum at 0.38and a maximum at 0.42in the nickel
prole corresponds to angles where the intensity of the
standing wave eld in the nickel layers is the largest and
smallest, respectively. The top of that gure shows the SF-
GEXRF prole of the Ni Kauorescence line with the equiva-
lent interference structure and additionally a substructure of
ne oscillations. These correspond to the Kiessig fringes visible
in XRR measurements between two reectivity maxima.
26
Only
those maxima in the angular prole correspond to minima in
the XRR prole and vice versa. What additionally can be seen in
this prole is that for higher angles angle resolved proles
become approximately constant. Compared to the SF-GEXRF, in
the GIXRF prole no ne interference structure can be detected.
This is mainly caused by the lower angular resolution due to the
beam divergence of this setup. This decreases the amplitude of
the structure and is a crucial factor in GIXRF setups.
To measure the interference fringes a combination of su-
cient temporal and spatial coherence needs to be considered.
The temporal coherence is mainly inuenced by the natural line
width which is a comparable factor for both angle resolved
setups. In the GIXRF setup, this concerns the characteristic
emission lines from the X-ray tube. The applied multilayer
optics is adapted for the respective energy, and the brems-
strahlung background within the bandwidth of the multilayer is
negligible. In GEXRF the temporal coherence is given by the
natural line width of the uorescence radiation if the detectors
energy resolution is sucient to isolate the line. Contrarily,
spatial coherence is inuenced by angular resolution. This
means divergence of the incidence beam in GIXRF and the
geometry of the setup (pixel size and detectorsample distance)
for GEXRF. Thus, these factors are entirely setup dependent and
there always is a compromise between setup eciency and
angular resolution.
Comparison with simulations
The quantication of angle resolved measurements can be
performed by varying the parameters of a simulated sample to
achieve an agreement with the measured proles. This requires
well-known and well-characterized instrumental parameters of
the setup.
20
In the demonstrated case a well-characterized
sample, the NiC multilayer, is used to validate the established
experimental parameters of the introduced setups. The data
from the XRR measurements of the NiC multilayer are applied
to perform simulations of the angular proles with an in-house
code based on the solution of the Sherman equation:
27
the
xrfLibrary. However, the simple Sherman approach is insu-
cient if an X-ray standing wave eld is formed in the sample. In
that case reection and refraction need to be additionally
considered. In the xrfLibrary the algorithm introduced by de
Boer
9,14
is used in the case of GIXRF and of Urbach and de
Bokx
28
for GEXRF. The code facilitates the application of any X-
ray excitation radiation, the use of optics and both scanning
geometries. The angular scans are implemented by varying the
excitation angle (GIXRF) or the detection angle (GEXRF). In the
latter the angular calibration is used in the calculation of the
isoangular lines and the solid angle of detection.
For GIXRF the geometry of the setup is additionally included
in the calculations of the solid angle of detection. Furthermore,
beam divergence must be considered in the simulations since it
causes damping of the amplitude of the interference structures.
The solid angle is calculated by means of a Monte-Carlo
approach. The geometry of the setup is implemented in the
code as obstructions for the X-ray trajectories. 10
6
X-rays are
generated within the point of origin, which is the footprint of
the beam on the sample considering its shape: Gaussian in one
and rectangular in the perpendicular direction. From there the
X-rays proceed in arbitrary directions within a half sphere solid
angle. Whenever the X-ray paths collide with the collimator the
trajectory is disregarded from further calculations. The solid
angle of detection is hence the ratio of photons hitting the
detector without collision to the number of emitted photons
times the emission solid angle (2p). A respective solid angle is
calculated for each angular position of the sample by smearing
the area of origin with respect to the X-ray beam. Smearing of
926 |J. Anal. At. Spectrom.,2019,34,922929 This journal is © The Royal Society of Chemistry 2019
JAAS Paper
Open Access Article. Published on 11 February 2019. Downloaded on 6/17/2019 10:33:20 AM.
This article is licensed under a
Creative Commons Attribution 3.0 Unported Licence.
View Article Online
the footprint at shallow incidence angles is thus included in the
calculation.
Fig. 5 top shows the comparison of the simulated and
measured GIXRF proles of nickel and silicon. The data are
normalized since the eciency of the focusing multilayer and the
detecting system is unknown. This is justiable since the sample
is dened by the shapes of the proles and not the absolute
intensities. In the bottom of the same gure the GEXRF prole of
the Ni Kauorescence line is shown and the region around the
prominent ne interference structures is enlarged. In contrast to
GIXRF, the geometrical parameters of the setup are not included
in the simulation, rather they are used to generate proles from
CCD frames. Since no investigations on the detector's eciency
were performed and a consistent investigation into the poly-
capillary lensesabsolute transmission is currently executed, the
GEXRF proles are also normalized.
The presented simulated and measured data overlap for
both setups and in the GIXRF case for both uorescence lines.
The validation sample is well dened and thus these results
allow the characterization of these laboratory setups. This
enables the application of recursive algorithms for quantitative
results when unknown samples are investigated.
As already mentioned, krypton was found in the GIXRF sum
spectra. Interestingly, the analysis of the krypton Kauores-
cence line prole shows that this sputtering gas is preferably
deposited in the carbon layers of the multilayer. This can be
seen in Fig. 6 where the measured GIXRF krypton uorescence
line prole (dark blue) and the simulated proles of carbon
(light blue), krypton (orange) and nickel (green) are shown. The
shapes of the carbon and measured krypton proles follow
a similar trend, indicating that krypton uorescence is excited
when the standing wave eld is formed within the carbon
layers. In the simulation of the krypton prole it was assumed
that 1 at% of the carbon layer is krypton. Except for the region
around 0.1this assumption generates an overlap of the simu-
lated and measured krypton proles. The mismatch in the low
angle region indicates that krypton is preferably concentrated
deeper inside the rst carbon layer. This qualitative analysis of
the results indicates how the system can be used to determine
in-depth contaminations in samples with nanometer layer
thickness. A dedicated sample could be used to characterize the
angle dependent lower limits of detection of the nal setup.
Discussion and outlook
For both presented angle resolved laboratory XRF measure-
ments the measured and simulated proles agree. Since the
investigated sample is well-known and well-characterized by
XRR measurements, this proves the applicability of these
laboratory spectrometers. The benet of the SF-GEXRF method
is the independence of the interference structure from the
quality of the primary radiation. This facilitates the resolution
Fig. 5 Comparison of normalized simulated and measured angle
resolved XRF proles of the NiC multilayer. For GIXRF (top) the proles
of the uorescence lines of Ni Ka(green) and Si K (red) are evaluated,
and for SF-GEXRF (bottom) only those of Ni Ka. A magnied image of
the angular region from 0.3up to 0.6of the SF-GEXRF prole is
shown to emphasize the overlap of the ne interference structures of
the prole.
Fig. 6 The measured GIXRF prole of the krypton Kauorescence line
(dark blue) and the simulated GIXRF proles of carbon (light blue),
nickel (green) and krypton (orange). In the simulation the krypton is
assumed to accumulate in the carbon layers and make up 1 at% of the
respective layers.
This journal is © The Royal Society of Chemistry 2019 J. Anal. At. Spectrom.,2019,34,922929 | 927
Paper JAAS
Open Access Article. Published on 11 February 2019. Downloaded on 6/17/2019 10:33:20 AM.
This article is licensed under a
Creative Commons Attribution 3.0 Unported Licence.
View Article Online
of ne interference structures in the NiC multilayersNi Ka
uorescence line prole which are not present in the GIXRF
measurement due to beam divergence. On the other hand, the
angular resolution of the SF-GEXRF strongly depends on the
pixel size of the CCD and its distance from the sample. Due to
the better excitation and detection eciency of krypton in the
GIXRF measurement additional information such as the pref-
erable deposition of krypton in carbon layers can be extracted.
In the discussed case the sample has sharp interfaces and
a period thickness in the nm range. For samples with sub-nm
layer thickness and worse quality of the layer boundaries the
measurement time for the SF-GEXRF case increases drastically
to achieve good statistics. This is demonstrated in Fig. 7 with
the comparison of a GIXRF (top) and SF-GEXRF (bottom) prole
of the Cr Kauorescence line of a chromiumscandium
multilayer with a bi-layer thickness of 1.7 nm (repeated 100
times) and strong intermixing between the respective layers.
The measurements are performed under the same conditions as
the measurements of the NiC multilayer. Though, to resolve the
interference structure of the chromium Kauorescence line
from this sample for the SF-GEXRF prole 10 000 frames with
an exposure time of 3 seconds are recorded. This yields an
overall measurement time of 950 minutes compared to the 15
minutes needed to record the GIXRF scan. This dierence
results mainly from the small solid angle of detection of the SF-
GEXRF setup. Additionally, the excitation at normal incidence is
less ecient than at grazing incidence angles.
One approach to tackle the challenge of long measurement
times of the SF-GEXRF setup is the use of a dierent position
sensitive detector, a single photon detector as in the studies of
Kayser
21
and Baumann.
23
These specialized detectors facilitate
fast read-out speeds while maintaining very low noise.
29,30
For
laboratory angle resolved setups, the high price and the
complexity of operation of these detectors are serious draw-
backs. Especially considering that with the specic GEXRF
geometry introduced in this work within an exposure time of 3
seconds around 2000 photons are detected by the CCD for the
samples presented. Thus, a faster read-out would reduce the
measurement time by around 20% and a further reduction of
measurement time could only be achieved by a higher ux on
the sample. This is possible by using rotating anode X-ray tubes
or metal-jet anodes,
31
which have the same drawbacks as the
photon counting detectors and additionally the former
produces a large footprint on the sample. An increase of ux in
the SF-GEXRF measurements could also be achieved if the
sample was investigated at shallower incidence angles. From
geometrical considerations it follows that for an incidence
angle of 6an increase of detected uorescence intensity by
a factor 10 is expected. Though, the inuence on the solid angle
of detection of this modication needs to be included in the
calculations as it can have an impact on the measured prole.
Another option is to decrease the distance between the sample
and the detector, which also decreases the angular resolution if
the pixel size remains unchanged. Another perspective is the
use of cmos detectors, since they enable fast read-out and small
pixel size. So far, the depletion zone thickness has been insuf-
cient for this kind of investigation but there is a rapid devel-
opment in this area.
The main drawback of GIXRF is the lower angular resolution
which can only be improved by lowering the divergence of the X-
ray beam. This is possible by adding collimators to the setup or
crystal optics. The former would result in a lowered ux on the
sample, but due to the small sampledetector distance this
parameter is not critical in the GIXRF spectrometer used.
Furthermore, the necessity to scan the sample for GIXRF
measurements increases the complexity of the setup. A
substantial benet of GIXRF is the possibility to combine the
GIXRF setup with XRR measurements. This has been shown in
the literature
11
and provides two complementary measurements
at once. Additionally, a combined analysis lowers the uncer-
tainties of the results as thoroughly discussed e.g. by D. Ingerle
et al.
33
Conclusions
The demonstrated experiments prove that angle resolved tech-
niques are reliable enough to serve as complementary
Fig. 7 GI- (top) and SF-GEXRF (bottom) of a chromiumscandium
multilayer with a period thickness of 1.7 nm. The GIXRF scan is per-
formed with an angular resolution of 0.001and each point is
measured for 5 seconds. In the SF-GEXRF measurement 10 000 CCD
frames are recorded with an exposure time of 3 seconds.
928 |J. Anal. At. Spectrom.,2019,34,922929 This journal is © The Royal Society of Chemistry 2019
JAAS Paper
Open Access Article. Published on 11 February 2019. Downloaded on 6/17/2019 10:33:20 AM.
This article is licensed under a
Creative Commons Attribution 3.0 Unported Licence.
View Article Online
measurements to XRR in the laboratory. Especially for samples
with strong intermixing the angle resolved methods benet
from the element specic in-depth information. Opposed to
XRR, not only contaminations can be identied but also their
distribution in the sample can be calculated. Since both
methods yield theoretically equivalent information about the
sample the choice of which method, GI- or SF-GEXRF, to use
depends on the sample and available hardware.
Conicts of interest
There are no conicts to declare.
Acknowledgements
The authors would like to thank Helga for her reliability, Bruker
Nano GmbH for the permission to perform the GIXRF
measurements and the constant support, AXO DRESDEN
GmbH for manufacturing the validation sample, Philippe Jon-
nard and Meiyi Wu from Universit´
e Pierre et Marie Curie in
Paris for providing the Cr/Sc sample, Daniel Gr¨
otzsch for con-
structing substantial components of both setups, and Michael
Haschke for inspiring this comparative work. J. B. would like to
thank the Helmut Fischer Foundation for nancial support.
Notes and references
1 W. H. Briscoe, F. Speranza, P. Li, O. Konovalov,
L. Bouchenoire, J. Van Stam, J. Klein, R. M. J. Jacobs and
R. K. Thomas, SoMatter,2012,8,50555068.
2 L. Zhou, L. Fox, M. Włodek, L. Islas, A. Slastanova, E. Robles,
O. Bikondoa, R. Harniman, N. Fox, M. Cattelan and
W. H. Briscoe, Carbon, 2018, 136, 255261.
3 D. N. Gurgew, D. M. Broadway, M. Gubarev, B. D. Ramsey
and D. A. Gregory, Rev. Sci. Instrum., 2016, 87, 104501.
4 Y. A. Babanov, Y. A. Salamatov, D. A. Ponomarev,
L. I. Naumova, V. V. Proglyado, M. A. Milyaev and
V. V. Ustinov, Phys. Met. Metallogr., 2015, 116, 11161126.
5 D. L. Windt, Comput. Phys., 1998, 12, 360370.
6 D. G. Stearns, J. Appl. Phys., 1989, 65, 491506.
7P. H
¨
onicke, B. Detlefs, M. M¨
uller, E. Darlatt, E. Nolot,
H. Grampeix and B. Beckho,Phys. Status Solidi, 2015, 212,
523528.
8 A. Haase, S. Bajt, P. H¨
onicke, V. Soltwisch and F. Scholze,
Appl. Crystallogr., 2016, 49, 21612171.
9 D. K. G. de Boer, Phys. Rev. B, 1991, 44, 498511.
10 J. Spear, H. Murakami and S. Terada, AIP Conf. Proc., 2003,
683, 646650.
11 D. Ingerle, M. Schiebl, C. Streli and P. Wobrauschek, Rev. Sci.
Instrum., 2014, 85,16.
12 W. W. Van Den Hoogenhof and D. K. G. De Boer,
Spectrochim. Acta, Part B, 1993, 48, 277284.
13 R. S. Becker, J. A. Golovchenko and J. R. Patel, Phys. Rev. Lett.,
1983, 50, 153156.
14 L. G. Parratt and C. F. Hempstead, Phys. Rev., 1954, 94, 1593
1600.
15 W. Kossel, V. Loeck and H. Voges, Z. Phys., 1935, 94, 139144.
16 M. Wu, K. Le Guen, V. Ilakovac, I. Vickridge, D. Schmaus,
E. Briand, C. Burcklen, E. Meltchakov, F. Delmotte and
P. Jonnard, Nucl. Instrum. Methods Phys. Res., Sect. B, 2016,
386,3943.
17 E. Langer, M. Haschke and S. D¨
abritz, Microchim. Acta, 2008,
161,3943.
18 G. Bortel, G. Faigel, M. Tegze and A. Chumakov, J.
Synchrotron Radiat., 2015, 23, 214218.
19 S. Terada, H. Furukawa, H. Murakami and K. Nishihagi, Adv.
X-Ray Anal., 2000, 43, 504509.
20 B. Beckho,J. Anal. At. Spectrom., 2008, 23, 845853.
21 Y. Kayser, J. Szlachetko and J. S`
a, Rev. Sci. Instrum., 2013, 84,
123102.
22 J. Szlachetko, D. Bana´
s, A. Kubala-Kuku´
s, M. Pajek, W. Cao,
J.-C. Dousse, J. Hoszowska, Y. Kayser, M. Szlachetko,
M. Kavˇ
ciˇ
c, M. Salome and J. Susini, J. Appl. Phys., 2009,
105, 086101.
23 J. Baumann, C. Herzog, M. Spanier, D. Gr¨
otzsch, L. L¨
uhl,
K. Witte, A. Jonas, S. G¨
unther, F. F¨
orste, R. Hartmann,
M. Huth, D. Kalok, D. Steigenh¨
ofer, M. Kr¨
amer, T. Holz,
R. Dietsch, L. Str¨
uder, B. Kanngießer and I. Mantouvalou,
Anal. Chem., 2017, 89, 19651971.
24 V. Szwedowski, J. Baumann, I. Mantouvalou, L. Bauer,
W. Malzer and B. Kanngießer, Phys. Status Solidi C, 2017,
14, 1700158.
25 J. Baumann, R. Gnewkow, S. Staeck, V. Szwedowski-
Rammert, C. Schlesiger, I. Mantouvalou and
B. Kanngießer, J. Anal. At. Spectrom., 2018, 33, 20432052.
26 H. Kiessig, Ann. Phys., 1931, 10, 770788.
27 J. Sherman, Spectrochim. Acta, 1955, 7, 283306.
28 H. P. Urbach and P. K. De Bokx, Phys. Rev. B, 1996, 53, 3752
3763.
29 L. Str¨
uder, U. Briel, K. Dennerl, R. Hartmann, E. Kendziorra,
N. Meidinger, E. Pfeermann, C. Reppin, B. Aschenbach,
W. Bornemann, H. Br¨
auninger, W. Burkert, M. Elender,
M. Freyberg, F. Haberl, G. Hartner, F. Heuschmann,
H. Hippmann, E. Kastelic, S. Kemmer, G. Kettenring,
W. Kink, N. Krause, S. M¨
ulle, A. Oppitz, W. Pietsch,
M. Popp, P. Predehl, A. Read, K. H. Stephan, D. St¨
otter,
J. Tr¨
umper, P. Holl, J. Kemmer, H. Soltau, R. St¨
otter,
U. Weber, U. Weichert, C. von Zanthier, D. Carathanassis,
G. Lutz, R. H. Richter, P. Solc, H. B¨
ottcher, M. Kuster,
R. Staubert, A. Abbey, A. Holland, M. Turner, M. Balasini,
G. F. Bignami, N. La Palombara, G. Villa, W. Buttle,
F. Gianini, R. Lain´
e, D. Lumb and P. Dhez, Rev. Mex.
Astron. Astros., Ser. Conf., 2001, 365, L18L29.
30 B. Henrich, A. Bergamaschi, C. Broennimann, R. Dinapoli,
E. F. Eikenberry, I. Johnson, M. Kobas, P. Kra,
A. Mozzanica and B. Schmitt, Nucl. Instrum. Methods Phys.
Res., Sect. A, 2009, 607, 247249.
31 O. E. Hemberg, M. Otendal and H. M. Hertz, Opt. Eng., 2004,
43(7), 1682.
32 M. Spanier, C. Herzog, D. Gr¨
otzsch, F. Kramer, I. Mantouvalou,
J. Lubeck, J. Weser, C. Streeck, W. Malzer, B. Beckhoand
B. Kanngießer, Rev. Sci. Instrum., 2016, 87, 035108.
33 D. Ingerle, G. Pepponi, F. Meirer, P. Wobrauschek and
C. Streli, Spectrochim. Acta, Part B, 2016, 118,2028.
This journal is © The Royal Society of Chemistry 2019 J. Anal. At. Spectrom.,2019,34,922929 | 929
Paper JAAS
Open Access Article. Published on 11 February 2019. Downloaded on 6/17/2019 10:33:20 AM.
This article is licensed under a
Creative Commons Attribution 3.0 Unported Licence.
View Article Online