Trident: A dual oxygenation and fluorescence imaging platform for real-time and quantitative surgical guidance
Full text
Trident: A dual oxygenation and fluorescence imaging platform for real-time and quantitative surgical guidance Silvère Ségaud 1 , Luca Baratelli 1 , Eric Felli 2 , Elisa Bannone 2 , 3 , Lorenzo Cinelli 2 , 4 , María Rita Rodríguez-Luna 1 , 2 , Nariaki Okamoto 1 , 2 , Deborah S Keller 5 , Michel de Mathelin 1 , Sylvain Lecler 1 , 6 , Michele Diana 1 , 2 and Sylvain Gioux 1 , 7 * 1 University of Strasbourg, ICube Laboratory, Illkirch, France, 2 Research Institute Against Digestive Cancer (IRCAD), Strasbourg, Alsace, France, 3 Department of General and Pancreatic Surgery—The Pancreas Institute, University of Verona, Verona, Italy, 4 Department of Gastrointestinal Surgery, San Raffaele Hospital IRCCS, Milan, Italy, 5 University of CA-Davis Medical Center, Department of Surgery, Sacramento, CA, United States, 6 INSA Strasbourg, Strasbourg, Alsace, France, 7 Intuitive Surgical, Aubonne, Switzerland Despite recent technological progress in surgical guidance, current intraoperative assessment of tissue that should be removed (e.g., cancer) or avoided (e.g., nerves) is still performed subjectively. Optical imaging is a non-contact, non-invasive modality that has the potential to provide feedback regarding the condition of living tissues by imaging either an exogenously administered contrast agent or endogenous constituents such as hemoglobin, water, and lipids. As such, optical imaging is an attractive modality to provide physiologically and structurally relevant information for decision-making in real-time during surgery. The Trident imaging platform has been designed for real-time surgical guidance using state-of-the-art opticalimaging.Thisplatformiscapable of dual exogenous and endogenous imaging owing to a unique filter and source combination, allowing to take advantage of both imaging modalities. This platform makes use of a real-time and quantitative imaging method working in the spatial frequency domain, called Single Snapshot imaging of Optical Properties (SSOP). The Trident imaging platform is designed to comply with all relevant standards for clinical use. In this manuscript, we first introduce the rationale for developing the Trident imaging platform. We then describe fluorescence and endogenous imaging modalities where we present the details of the design, assess the performance of the platform on the bench. Finally, we perform the validation of the platform during an in vivo preclinical experiment. Altogether, this work lays the foundation for translating state-of-theart optical imaging technology to the clinic. KEYWORDS optical imaging, oxygenation imaging, fluorescence imaging, surgical guidance/ navigation, clinical translation OPEN ACCESS EDITED BY Kristen M. Meiburger, Politecnico di Torino, Italy REVIEWED BY Mengyang Liu, Medical University of Vienna, Austria Dimitris Gorpas, Helmholtz Association of German Research Centres (HZ), Germany *CORRESPONDENCE Sylvain Gioux, [email protected] SPECIALTY SECTION This article was submitted to Biophotonics, a section of the journal Frontiers in Photonics RECEIVED 31 August 2022 ACCEPTED 18 October 2022 PUBLISHED 09 November 2022 CITATION Ségaud S, Baratelli L, Felli E, Bannone E, Cinelli L, Rodríguez-Luna MR, Okamoto N, Keller DS, de Mathelin M, Lecler S, Diana M and Gioux S (2022), Trident: A dual oxygenation and fluorescence imaging platform for realtime and quantitative surgical guidance. Front. Photonics 3:1032776. doi: 10.3389/fphot.2022.1032776 COPYRIGHT © 2022 Ségaud, Baratelli, Felli, Bannone, Cinelli, Rodríguez-Luna, Okamoto, Keller, de Mathelin, Lecler, Diana and Gioux. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms. Frontiers in Photonics frontiersin.org01 TYPE Original Research PUBLISHED 09 November 2022 DOI 10.3389/fphot.2022.1032776
1 Introduction The capabilities of near-infrared (NIR) optical imaging greatly expanded over the past few years, providing feedback for tissue status assessment during surgery, and particularly blood perfusion assessment. In the context of this work, several intraoperative optical imaging technologies allowing non-contact, large field-of-view macroscopic (i.e., greater than 100 cm 2 ) imaging of the surgical field have been developed all the way from concept to clinical trials testing. These technologies differ from microscopic imaging technologies having smaller fields of view and requiring the instrument to be in contact or very close to the specimen. In this context, the development of fluorescence imaging for surgical guidance paved the way with the growing number of available contrast agents and the development of new macroscopic imaging technologies (Gioux et al., 2010;Gibbs, 2012;Vahrmeijer et al., 2013). However, the lack of quantitative capabilities leads to subjective assessment of fluorescence images, introducing variability in the surgical outcome (Pogue et al., 2018a;Pogue et al., 2018b;Mieog et al., 2022). Alternatively, the ability of multispectral and hyperspectral endogenous imaging to extract functional parameters for clinical applications has been reported for surgical guidance (Lu and Fei, 2014;Shapey et al., 2019; Clancy et al., 2020). Yet, it suffers from the lack of quantitative extraction of tissue constituents’concentrations and cannot provide real-time feedback intraoperatively. Spatial Frequency Domain Imaging (SFDI) is a relatively recent optical imaging method that has the capability to perform quantitative imaging of hemoglobin concentrations, and in turn of tissue oxygenation over large fields of view (>100 cm 2 ) (Dognitz and Wagnieres, 1998;Cuccia et al., 2009;Gioux et al., 2019). Recent developments in acquisition and processing led to a real-time implementation of SFDI–namely Single Snapshot imaging of Optical Properties (SSOP) (Vervandier and Gioux, 2013;van de Giessen et al., 2015). The most advanced form of SSOP makes use of deep learning-based processing, leading to high-quality images while preserving the real-time quantitative capabilities of the method (Aguenounon et al., 2020;Smith et al., 2022). To overcome the lack of tools capable of quantitative endogenous imaging for tissue assessment in clinical settings, we propose a novel imaging platform, called Trident, integrating the latest SSOP developments for performing real-time oxygenation imaging as well as fluorescence imaging. Combining these state-of-the-art imaging modalities in a single imaging platform enables the comparison of their respective performances as well as the potential for their combination. In this manuscript, following a background section on optical imaging for surgery, the Trident imaging platform is described in detail, as well as the fluorescence and oxygenation imaging methods used. Next, bench characterization experiments are presented, showing the performances of the platform for optical properties imaging and fluorescence detection. Finally, in vivo validation demonstrates the ability of the imaging platform to dynamically detect ischemia in the small bowel by measuring tissue oxygenation and fluorescence in real-time. 2 Background 2.1 Quantitative optical imaging for surgery Several implementations of either fluorescence imaging or oxygenation wide-field (>100 cm 2 )imagingexistinthe literature (DSouza et al., 2016;Shapey et al., 2019;Clancy et al., 2020). Depending on the imaging methodology used, these implementations greatly vary in their capability to quantify the measured signal. Without trying to provide an extensive review, it is worth mentioning that the most common form of optical imaging for either fluorescence or oxygenation consists of using temporally and spatially constant illuminations (Bigio and Fantini, 2016). These methods are inherently limited in their capacity to quantify signalssincetheyareunabletoseparatescatteringfrom absorption, creating a crosstalk between the two sources of contrast. By contrast, methods using modulations in either time or space allow for the separation of scattering and absorption, thereby allowing a more accurate quantification of the signals. In this work, we use a method called Spatial Frequency Domain Imaging (SFDI). A recent review has been published in the literature, explaining in detail the principles, acquisition and processing methods as well as applications (Gioux et al., 2019). In a nutshell, SFDI relies on the projection of a spatially modulated illumination pattern (also called structured illumination, typically a sinusoidally modulated pattern) and the quantification of the amplitude dampening of the reflectance pattern as a function of spatial frequency. A light propagation model is then used (typically Monte Carlo) to relate the dampening of the amplitude of the reflectance pattern to the optical properties of the medium (absorption coefficient and reduced scattering coefficient). Our group developed a real-time implementation of SFDI, called Single Snapshot of Optical Properties (SSOP) capable of capturing absorption and reduced scattering coefficients as well as profile information (for 3D profile correction) of a sample with only one image (Vervandier and Gioux, 2013;van de Giessen et al., 2015;Angelo et al., 2017). The latest implementation makes use of GPU, fast lookup tables and deep learning to allow accurate and high image-quality imaging with profile correction in real-time (Angelo et al., 2016;Panigrahi and Gioux, 2018;Schmidt et al., 2019;Aguenounon et al., 2020). Resources regarding the instrumental and processing aspects of SFDI and SSOP, Frontiers in Photonics frontiersin.org02 Ségaud et al. 10.3389/fphot.2022.1032776
including material references and processing code, are freely available in OpenSFDI (Applegate et al., 2020). 2.2 Fluorescence imaging The most common form of fluorescence imaging for imageguided surgery relies on the injection of an exogenous contrast agent to highlight a particular structure (e.g., lymph node, vessels, tumor) (Frangioni, 2003;Ntziachristos, 2006;Gioux et al., 2010; Gibbs, 2012;Vahrmeijer et al., 2013;Mieog et al., 2022). A dedicated imaging system that matches the optical properties of the contrast agent (namely absorption and emission spectra) is then used to highlight the presence of the contrast agent in realtime during the surgery. Current implementations used in the clinic (many being commercial products) are qualitative in nature, meaning the fluorescence signal measured by the imaging system is not directly related to the local amount of contrast agent (i.e., its concentration). The fluorescence signal is dependent on multiple external and internal factors such as the illumination homogeneity, the distance between the imaging system and the surgical field, the local tissue optical properties (absorption and scattering properties), the variation of the fluorescence properties due to the dye environment, or the quantum yield of the molecule. Recent effort has tried to address these challenges (Pogue et al., 2018a;Pogue et al., 2018b;Mieog et al., 2022). Some current commercial systems allow to correct for the distance between the imaging system and the surgical field, and some even correct for the illumination inhomogeneities. Research prototypes have been designed to further correct not only distance and field homogeneity, but also optical properties, even in real-time. Finally, the temporal behavior of the fluorescence signal is being investigated as a source of contrast, notably for the imaging of blood perfusion (Matsui et al., 2009;Diana et al., 2015;Meijer et al., 2021). This method referred to as fluorescence dynamics is less dependent on several external and internal factors. In the implementation described in this work, we chose to enable three modes of fluorescence imaging in the same prototype. The first mode consists of illuminating the field with a temporally constant (called continuous wave, CW) near infrared illumination. CW fluorescence imaging is the most common implementation both commercially and in research (Frangioni, 2003;Gioux et al., 2010). The second mode consists of capturing both the fluorescence image and the excitation illumination. One then divides the fluorescence image by the excitation image to automatically correct for distance, illumination inhomogeneity and partially for optical properties (Ntziachristos et al., 2005;Themelis et al., 2009). The third mode consists of using SFDI or SSOP to extract the optical properties of the surgical field at both the excitation and emission wavelength and use this information to correct for distance, sample profile, illumination inhomogeneity and optical properties. The method used to obtain the so-called quantitative fluorescence images have been previously described in the literature (Sibai et al., 2019;Valdes et al., 2019). We use our own method called qF-SSOP (Valdes et al., 2017). Note that for fluorescence imaging, the excitation wavelength and filter used to detect the signal are dependent on the contrast agent imaged. In our case, we intend to image indocyanine green (ICG), and from prior work, we use an excitation at 760 nm and collection through filters starting at 780 nm. More details about our filtration strategy are described in section 2.4.2. 2.3 Oxygenation imaging Oxygenation imaging has gained increasing interest for surgical applications (DSouza et al., 2016;Kohler et al., 2019; Shapey et al., 2019;Clancy et al., 2020;Felli et al., 2020;Felli et al., 2021). Oxygen saturation is computed as the ratio of the concentration of oxy-hemoglobin to the concentration of total hemoglobin (oxy-hemoglobin + deoxy-hemoglobin). Note that oxygen saturation here refers to a mix of arterial and venous blood, as opposed to pulse oximetry that only measures the amount of oxygen saturation in arterial blood. Most methods used for oxygenation imaging rely on reflectance imaging at several wavelengths, also called multispectral imaging, or hyperspectral imaging in cases where a large number of wavelengths are used (e.g., >10 wavelengths). These methods rely on CW illumination and typically fit the measured reflectance spectrum to a theoretical spectrum and extract an oxygenation value. In the case of oxygenation imaging, because it is a ratio (oxy-hemoglobin to total hemoglobin), the influence of scattering and distance between the sample and the imaging system can be managed. Several authors demonstrate the use of these CW multispectral and hyperspectral methods in surgery. Another approach for oxygenation imaging consists of properly separating scattering and absorption on the measured signal by using a quantitative optical imaging method, such as SFDI (Gioux et al., 2011;Ponticorvo et al., 2013). In this case, the absorption coefficient at a minimum of two wavelengths is then used in conjunction with Beer’s law to directly quantify the concentration of oxy-hemoglobin and deoxy-hemoglobin. These values are then used to compute the oxygen saturation of the sample. In our case, prior work in SFDI determined that optimal wavelengths for oxygenation imaging were close to 665 nm and 860 nm (Mazhar et al., 2010). Our group and others have translated similar technology to preclinical and clinical experiments (Gioux et al., 2011; Nadeau et al., 2013;Ponticorvo et al., 2013;Ghijsen et al., 2018;Schmidt et al., 2019;Weinkauf et al., 2019;Chen and Durr, 2020;Ren et al., 2020;Zhao et al., 2021;Lyu et al., 2022). However, none of the prior work includes a real-time methodology and combines the measurement of oxygenation Frontiers in Photonics frontiersin.org03 Ségaud et al. 10.3389/fphot.2022.1032776
with the measurement of fluorescence. The Trident imaging platform enables such feature to understand potential clinical use of either or both technologies. 3 Materials and methods 3.1 System design This section describes the design of the imaging platform for each subsystem. An overview of the complete system is given in Figure 1A. This medical cart-based platform features a three channel-imaging head, fiber-coupled white light sources and laser sources, and a workstation. 3.1.1 Cart and arm The platform is built upon a custom medical-grade cart (Symbio cart, ITD, Neubiberg, Germany). The wheeled base is equipped with an isolation transformer to regulate power supply to the device and counterweights to safely balance the assembly (compliant with IEC60601). Two articulated arms bear display screens both for the surgeons and the operator. A third articulated arm (Elite 5220, ICW, Medford, OR, United States) is fixed to the central column to accommodate the imaging head. Its large reach enables a user to operate the system while keeping distance from the sterile field. Cables can be guided from the devices through the central column and cable retainer on the articulated arm. Power is supplied to the imaging head using a 24 V AC/DC converter (AKM90PS24, XP Power, Singapore). All three articulated arms conveniently fold for storage. The wide castors and large handle provide good mobility when deploying or rolling out the platform. 3.1.2 Imaging head and filters design The imaging head is depicted in Figures 1B,C. The optomechanical design features three independent imaging channels: one visible light channel (COLOR) and two NIR channels (NIR1 and NIR2). These channels are co-registered using tilting mirrors, producing a field of view of 12 cm × 15 cm. The detection optics enable fine focusing and individual aperture control. An RGB camera (GO-5000C-USB, JAI Ltd., Kanagawa, Japan) and two monochrome cameras (GO-5000M-USB, JAI Ltd., Kanagawa, Japan; Edge 4.2, Excelitas PCO GmbH, Kelheim, FIGURE 1 Trident imaging platform: (A) Picture of the entire imaging platform, including the medical-grade tart, the articulated arm, the imaging head, the light sources (white light and lasers), the computer and the displays; (B) picture of the imaging head at the end of the articulated arm including the cameras, the Dptomechanical coupling system, the projector and the optical lenses; (C) detailed schematics of the imaging head including the cameras, the optomechanical coupling system, the optical lenses, the dichroic mirrors, the light sources, the illumination ring, the projector and the light paths; (D) details of the filters design of the imaging system: note the separation between the color, NIR1 and NIR2 cameras in combination with the light sources allowing the imaging of the surgical field in color, oxygenation or fluorescence. Frontiers in Photonics frontiersin.org04 Ségaud et al. 10.3389/fphot.2022.1032776
Germany) are used for the detection in the COLOR, NIR1 and NIR2 channels, respectively. Custom filter sets (Chroma Inc., Bellows Falls, VT, United States) provide wavebands selection for optical properties and fluorescence imaging with minimized crosstalk between channels. Two filter cubes (Axio cubes, Zeiss, Oberkochen, Germany) are composed of two emission filters and a dichroic mirror. These filters can be swapped easily with minimal re-alignment depending on the desired imaging configuration. Figure 1D depicts the filter design in combination with the sources for the presented configuration of oxygenation imaging and 800 nm fluorescence imaging. A first filter cube allows for the isolation of the COLOR channel (400–650 nm). A second filter cube separates the NIR1 and NIR2 channels. The NIR1 channel in the presented configuration allows for the detection of either the 665 nm for oxygenation imaging or 760 nm for fluorescence imaging. The NIR2 channel in the presented configuration allows for the detection of either the emission fluorescence signal from ICG above 780 nm or 860 nm for oxygenation imaging. Note that the dichroic filters optical density is around 2 and the channels emission filters optical densities are around 6–7. This configuration allows appropriate separation of the visible anatomical information in the COLOR channel, concurrently with either oxygenation imaging or fluorescence in NIR1 and NIR2 channels. 3.1.3 White light source Light emitting diodes (LEDs) based lamps (LO-35, Fiberoptics Technologies Inc., Pomfret, CT, United States) with custom 650 nm short-pass filters and fiber bundles provide a bright illumination of the surgical field. Though the LEDs emit low amounts of infrared light, the additional filter ensures an optimal visible illumination while preserving the NIR channels from background noise. The two fiber bundles are split in two outputs each to produce a multi-angle illumination from the imaging head over a large area over 50 cm diameter at a 45 cm working distance. An illumination ring with custom lensing allows to illuminate the field in the most homogenous manner. 3.1.4 Laser source Custom laser sources were designed to equip the platform. A first source is dedicated to the projector illumination to collect reflectance data for optical properties imaging and a second source is dedicated to fluorescence imaging. The two sources are referred to as reflectance source and fluorescence source, respectively to their purpose. Figures 2A,B presents one complete source closed and open (top-view) respectively. A single source houses six independent modules that can be loaded with different laser diodes (LDX Optronics, Maryville, TN, United States) to achieve multispectral imaging, or to match the excitation of fluorophores of interest. Identical laser diodes can also be combined to provide higher illumination power. For real-time oxygenation imaging, a combination of four 665 nm (2 W each) and two 860 nm (5 W each) laser diodes equip the reflectance source. In order to excite 800 nm fluorophores–especially indocyanine green (ICG)—six 760 nm (6 W each) laser diodes are mounted in the fluorescence source. Each diode is thermally regulated by a thermoelectric cooler (TEC) (TEC-15.4-3.9-33.4-72-40-CH4,7, Arctic TEC Technologies, Dortmund, Germany) and a thermistor placed on the laser diode chassis (TH10K, Thorlabs, Newton, NJ, United States). Individual TEC drivers (PTC5K-CH, Wavelength electronics, Bozeman, MT, United States) enable temperature regulation and monitoring of the diodes. Similarly, laser diode drivers (LD10CHA, wavelength electronics, Bozeman, MT, United States) control the laser output power, and monitoring of the integrated photodiode feedback. A custom optomechanical assembly featuring optical filters (Chroma Inc., Bellows Falls, VT, United States) and graded-index lenses (Newport, Irvine, CA, United States) performs the laser output coupling into a dedicated fiber bundle (CeramOptec, Bonn, Germany). In the case of the reflectance source, the modules are coupled into a 6-legged fiber bundle combined in a single output for coupling in the projector. In the fluorescence source, the 6 legs of the bundle contain multiple fibers which are randomized and redistributed into 6 legs that are plugged in an illumination ring. In addition, each module is interfaced with USB analog input/ouput modules (USB-202, Measurement Computing, Norton, MA, United States) and custom electronics to control the modules and record feedback signals from a computer. The interfacing of a single module is depicted in Figure 2C. As shown in Figure 2D, in order to dissipate the great amounts of heat generated by the laser diode cooling and different drivers, all the heating modules components are mounted on a tunnel heatsink equipped will push-pull fans. This design produces a compact assembly and while limiting noise due to the limited number of fans and their large diameter. 3.1.5 Illumination Two independent illuminations can be used with the Trident imaging platform. The first illumination is a single illumination ring containing four lensed outputs for the white light (from the LED white light source), and six SMA outputs for the CW fluorescence illumination (from the fluorescence laser diode source). The pattern and lensing have been optimized to the best of our effort to provide the most homogeneous illumination possible. A projector based on a digital micromirror device equipped with a NIR compatible projection optics (V7000 HiSpeed and STAR-CORE, Vialux GmbH, Chemnitz, Germany) and a custom fiber coupling adapter enables the projection of patterns at a 45 cm working distance for optical properties imaging using SFDI. Linear polarizers (RCV8N2EC, Moxtek, Orem, UT, United States) are cross-polarized in front of the projection and detection paths to reject specular reflections. One 5 V DC/DC converter (PYB30-Q24-S5Frontiers in Photonics frontiersin.org05 Ségaud et al. 10.3389/fphot.2022.1032776
U, CUI Inc., Tualatin, OR, United States) provides power supply to the digital micromirror device (DMD) board. 3.1.6 Workstation A high-performance workstation has been assembled to achieve the imaging device control, data processing, data storage management and visualization. A CameraLink HS frame grabber (ME5 Marathon AF2, Silicon Software, Mannheim, Germany) and an additional USB3.0 controller ensure the communication with the imaging head. Powerful CPU (Intel Core i5-10600K, Intel, Santa Clara, CA, United States) and GPU (Quadro RTX 4000, NVIDIA, Santa Clara, CA, United States) and 64 GB of high-speed RAM can perform intensive data processing and visualization on multiple displays. Finally, 5 TB of fast SSD storage are installed for data stream management up to 1 GB/s. 3.2 System testing 3.2.1 Optical performances A USAF1951 resolution chart (RES-2, Newport, Irvine, CA, United States) was placed under the imaging head at 45 cm working distance and imaged with each channel: COLOR, NIR1 and NIR2. Optical properties extraction was characterized by imaging an array of 12 tissue-mimicking phantoms of 6 × 6 × 2.5 cm. These cured silicone-base Polydimethylsiloxane (PDMS) samples were fabricated using varying amounts of alcohol-soluble nigrosine and Titanium dioxide (TiO 2 ) to tune the absorption and scattering of the material, respectively, spanning a large range of optical properties to be measured: µ a ranging from 0.013 to 0.08 mm −1 and µ s ’ranging from 0.75 to 1.8 mm −1 at 665 nm. These samples were imaged at 665 nm and 860 nm under a 0.2 mm −1 spatial frequency illumination and processed using the deep learning-based implementation of SSOP. The measurements were compared to those obtained with a reference benchtop imaging system using 7 phase-SFDI. That reference SFDI system was validated in a multi-laboratory performance assessment of diffuse optics instruments (BitMap exercise) (Lanka et al., 2022) and used to obtain the reference optical property values for this work. 3.2.2 Illumination Peak fluence rates were measured at 45 cm working distance in the center of the field of view using a digital handheld power meter (PM100D, Thorlabs, Newton, NJ, United States) at 665 nm and 860 nm while projecting a continuous wave pattern, and at 760 nm by fiber illumination. FIGURE 2 Laser light source: (A) Picture of the light source closed; (B) picture of the light source opened (top view); (C) detailed schematics of a single laser diode unit including the laser diode (LD) and its driver, the thermoelectric cooler (TEC) and its driver, the thermistor, the optomechanical module and the output fiber; (D) details of the cooling design of the source: note that the single laser diode units (D1 to D6) are assembled onto a unique heatsink with forced airflow (in and out) for more efficient cooling while maintaining a compact design. Frontiers in Photonics frontiersin.org06 Ségaud et al. 10.3389/fphot.2022.1032776
3.2.3 Sensitivity and noise The sensitivity of the NIR2 channel for fluorescence detection was assessed by imaging eight different dilutions of ICG diluted in dimethyl sulfoxide (DMSO) at concentrations ranging from 25 nM up to 800 nM. The vials were placed at 45 cm from the imaging head and imaged with varying exposure settings successively onto two different background materials: a large tissue-mimicking phantom with µ a = 0.012 mm −1 and µ s ’= 1.08 mm −1 at 665 nm and a sheet of white paper. Average fluorescence signals were extracted and compared to the noise floor level for sensitivity assessment. Sensitivity was defined as the concentration producing an intensity twice as high as the noise floor level. In order to quantify the background noise in the fluorescence images, the contributions from dark noise and both white light and excitation light leakage were measured. Images of the previously mentioned background materials from the NIR2 channel were acquired with varying exposure settings–exposure time, sensor binning–with successively all lights switched off, and white light and excitation laser light switched on. 3.3 Preclinical experiments 3.3.1 Bowel ischemia model One large swine was involved in this non-survival study, which received full approval from the local Ethical Committee on Animal Experimentation (ICOMETH No. 038.2019.01.121) and by the French Ministry of Superior Education and Research (MESR) under the following reference: APAFIS #208192019052411591088 v3. After performing a midline laparotomy, a 20 cm small bowel loop was exposed. A central region was delineated, and corresponding arcade branches were isolated. A clamp was used to create an occlusion on these vessels, causing ischemia on a limited portion of the bowel while the rest of the loop remained perfused. The occlusion was maintained during 14 min until release. 3.3.2 Fluorescence imaging A dose of 4 mg of ICG powder diluted in 1.6 ml of 5% glucose solution (Infracyanine, SERB, Paris, France) was injected intravenously 2 min before occlusion release. Color images from the COLOR channel and fluorescence images from the NIR2 channel were acquired starting from the injection timepoint and for 2 min and 30 s. The imaging working distance was maintained at 45 cm. Fluorescence excitation was produced using 760 nm laser light. 3.3.3 Oxygenation imaging A sinusoidal pattern with spatial frequency of 0.2 mm −1 was projected at both 665 nm and 860 nm wavelengths using the DMD projector. Reflectance images from the NIR1 and NIR2 channels were acquired, as well as color images from the COLOR channel for anatomical reference. Images were recorded starting 2 min before occlusion to establish the oxygen saturation baseline, and during 12 min after occlusion. After performing fluorescence imaging, oxygenation imaging resumed to record for 3 min during the end of the reperfusion phase. 4 Results 4.1 System testing 4.1.1 Optical performances At 45 cm working distance and performing 2 × 2 binning on the camera sensor (for increased sensitivity), the optical system produces images of a 12 × 15 cm field-of-view, with a resolution of 1024 × 1280 pixels. The measured lateral resolution with fully open apertures for channels COLOR and NIR2 was 5.04 lp/mm and 5.66 lp/mm for the NIR1 channel. A set of 12 tissue-mimicking phantoms was imaged first with a benchtop imaging system using 7 phase-SFDI as reference. The images produced by the clinical platform were processed using deep learning-based SSOP. High image quality and good accuracy with respect to optical properties extraction were achieved. Figure 3 presents the extracted optical properties maps and error estimation in comparison to standard SFDI. Errors in µ a and µ s ’measurements were respectively below 2.7% and 2.1% for both 665 nm and 860 nm oxygenation imaging wavelengths. 4.1.2 Illumination A continuous-wave illumination was produced using the DMD projector at both 665 nm and 860 nm oxygenation imaging wavelengths. The projector has a small Gaussian-like variation of intensity across the projected field (55% maximum). This inhomogeneity does not affect the performance of the system as the signal-to-noise ratio allows proper processing of the data at all locations of the image. At 45 cm working distance, a power meter probe was placed in the center of the field-of-view. Fluence rates of 0.93 mW/cm 2 at 665 nm and 1.51 mW/cm 2 at 860 nm were measured and classified as laser Class 1 as per IEC60825. The fluorescence excitation field at 760 nm showed a peak fluence rate of 31.51 mW/cm 2 and classified as laser Class 3 R as per IEC60825. 4.1.3 Sensitivity and noise A set of 8 ICG dilutions with concentrations ranging from 25 nM to 800 nM was imaged in the same previously stated conditions. Figure 4A shows the images from the COLOR and NIR2 channels as well as an overlay. Figure 4B shows the evolution of intensity in the NIR2 images for each dilution at various exposure times with fixed 2 × 2 sensor binning as well as Frontiers in Photonics frontiersin.org07 Ségaud et al. 10.3389/fphot.2022.1032776
using a silicone phantom as background. The intensity increases linearly with the exposure time until reaching saturation. From the entire dilution series, the sensitivity was estimated by measuring the concentration corresponding to an intensity 2 times higher than the noise floor level, yielding a signal-tobackground ratio of 2. At 40 ms exposure time and 2 × 2 binning factor, the sensitivity was estimated at 18.8 nM. Background noise in the NIR2 channel was measured using successively a tissue-mimicking phantom and a sheet of white paper placed at 45 cm working distance. Figure 4C showstheevolutionofdarknoise and leakage with exposure settings and background materials. Dark noise was first measured with exposure times ranging from 5 ms to 160 ms and sensor binning factors of 1 × 1, 2 × 2, and 4 × 4. Noise levels representing 0.1%, 0.6%, and 2.4% of the dynamic range of the NIR2 camera were measured respectively with 1 × 1, 2 × 2, and 4 × 4 binning factors. The same set of images was acquired with fully open aperture, and both white light and 760 nm laser excitation light switched on. Noise levels were on average 17% and 46% higher than the dark noise using a phantom background or a paper background, corresponding mostly to excitation light leakage into the NIR2 channel. 4.2 Preclinical experiments Figures 5A,B shows the results of the preclinical experiments in COLOR, fluorescence and oxygenation. A comparison of fluorescence and oxygenation imaging in given at three timepoints: during the baseline phase, after 12 min of occlusion and 1 min after reperfusion. 4.2.1 Fluorescence imaging At 12 min into the occlusion phase, a dose of ICG was intravenously injected. Images from the NIR2 and COLOR channels were recorded at 12 frames per second with 50 ms exposure time and 2 × 2 sensor binning while illuminating the surgical field with 760 nm laser light. Fluorescence intensity was monitored in a 120 × 180 pixels region of interest (ROI) located in the ischemic area and in a 50 × 50 pixels ROI located in a perfused area. The fluorescence signal reached its peak in the perfused area 35 s after injection before decreasing and stabilizing at 58% of the peak intensity 1 min later. After 14 min of occlusion, the clamp was released. The fluorescence intensity immediately increases in the ischemic area, as the perfusion resumes. When stabilizing, the intensity reaches 72% of the intensity in the control area. FIGURE 3 Bench results: endogenous imaging: (A) Tissue mimicking phantoms having different absorption and reduced scattering properties imaged with the Trident imaging platform: optical properties are imaged on all phantoms using the NIR1 (665 nm) and NIR2 (860 nm) cameras; (B) plots comparing the results obtained using a referencing SFDI acquisition method on a bench system compared with the results obtained with the SSOP acquisition method on the Trident platform: note the good agreement in absorption and reduced scattering (errors lower than 2.7% and 2.1%, respectively). Frontiers in Photonics frontiersin.org08 Ségaud et al. 10.3389/fphot.2022.1032776
4.2.2 Oxygenation imaging Oxygenation imaging was achieved at 12 frames per second using structured illumination at 0.2 mm −1 spatial frequency at 665 nm and 860 nm. Figure 5B shows the oxygen saturation trends for the two ROIs over time. During the baseline phase, both ROIs show stable StO 2 rates of 77.9% and 74.6% respectively in the control and ischemic areas. While the perfused area remains 4% of its baseline value of StO 2 , oxygenation in the ischemic area decreases until reaching 63.8% after 12 min of occlusion. From 2 min to 30 s after the occlusion release, the StO 2 rate decreases back its baseline state, reaching 75% after 5 min and 30 s of reperfusion. 4.2.3 Video A video is included in the supplementary material. It presents the merged of the oxygenation and fluorescence results over the course of the experiment. In vivo oxygenation and fluorescence imaging. An occlusion of a portion of a bowel loop was performed for 14 min. Oxygenation was monitored prior to clamping and during the 12 first minutes of occlusion. Fluorescence images were recorded from the ICG intravenous injection until 30 s after release. Oxygenation imaging resumed for the 3 min of the reperfusion phase. The absence of fluorescence signal in the ischemic area correlates with the low StO2 rates observed. After reperfusion, this area shows fluorescence signal levels and StO2 rates closetothoseoftheperfusedareas. Oxygenation rates are steady in the control area whereas ischemia could be detected and quantified in the occluded area. 5 Discussion We presented the design and testing of a clinicallycompatible imaging platform for real-time quantitative optical imaging during surgery. The platform is designed to image either fluorescence in three modes (CW, semi-quantitative and quantitative) or oxygenation. The platform imaging modality can be changed on the fly during the surgery allowing to compare both fluorescence and oxygenation modalities. The bench characterization of the platform demonstrates its performances for optical properties imaging and fluorescence imaging. First, the comparison between tissue-mimicking phantom optical properties obtained from standard profile corrected SFDI and state-of-the-art deep learning-based profile corrected SSOP shows good agreement for both absorption and reduced scattering, at 665 nm and 860 nm, as expected from previous published validation of SSOP. Average error below 2.7% in absorption and 2.1% in reduced scattering were observed. The considered optical properties were matching FIGURE 4 Bench results: fluorescence imaging: (A) Results obtained with the Trident imaging platform from a dilution series of indocyanine green (ICG) diluted in dimethyl sulfoxide (DMSO) at concentrations ranging from 25 nM to 800 nM in color (left), NIR2 fluorescence (middle) and a merged image of the color with the fluorescence (right); (B) plot showing the dilution series and a phantom background imaged at various exposure times from 5 to 80 ms; (C) plot showing the dark noise and leakage with different binning, exposure settings and background materials. Frontiers in Photonics frontiersin.org09 Ségaud et al. 10.3389/fphot.2022.1032776