Verification of the analytical performance of thyroglobulin autoantibodies on ALINITY ci ® experience from the biochemistry laboratory of Mohammed VI University Hospital in Oujda

Douichi, Imane; Kidoun, Asmae; Bourchid, Mohamed Khalil; Rchid, Sara Moulay; Halas, Meryem El; Moujtahide, Dounia El; Sebbar, El Houcine; Choukri, Mohammed

Abstract

The objective of this study was to assess the analytical performance of thyroglobulin autoantibody determination using a two-step immunoassay based on microparticle chemiluminescence immunoassay (CMIA) technology. This evaluation was conducted in compliance with the Scope A criteria outlined in the guidelines for the verification and validation of medical biology methods. We investigated the assay’s repeatability and intermediate precision. The results were highly satisfactory across three concentration levels (low, medium, and high). For intermediate precision, the coefficients of variation (CV) were: CV1 = 2.56 %, CV2 = 3.60 % and CV3= 2.70 %, respectively. For repeatability, the coefficients of variation were CV1 = 3.13 %, CV2 = 2.90 and CV3= 2.40 %, respectively. These findings confirm the method’s performance and its alignment with the analytical objectives set by the manufacturer and professional societies. This ensures compliance with both regulatory and normative requirements.

Full text

 Corresponding author: Imane Douichi Copyright © 2025 Author(s) retain the copyright of this article. This article is published under the terms of the Creative Commons Attribution Liscense 4.0. Verification of the analytical performance of thyroglobulin autoantibodies on ALINITY ci ® experience from the biochemistry laboratory of Mohammed VI University Hospital in Oujda Imane Douichi 1, 2, *, Asmae Kidoun 1, 2, Mohamed Khalil Bourchid 1, 2, Sara Moulay Rchid 1, 2, Meryem El Halas 1, 2, Dounia El Moujtahide 1, 2, El Houcine Sebbar 1, 2 and Mohammed Choukri 1, 2 1 Mohammed First University, Faculty of Medicine and Pharmacy of Oujda, Morocco. 2 Biochemistry laboratory of Mohammed VI University Hospital, Oujda, Morocco. World Journal of Biology Pharmacy and Health Sciences, 2025, 21(01), 427-434 Publication history: Received on 02 December 2024; revised on 11 January 2025; accepted on 13 January 2025 Article DOI: https://doi.org/10.30574/wjbphs.2025.21.1.0016 Abstract The objective of this study was to assess the analytical performance of thyroglobulin autoantibody determination using a two-step immunoassay based on microparticle chemiluminescence immunoassay (CMIA) technology. This evaluation was conducted in compliance with the Scope A criteria outlined in the guidelines for the verification and validation of medical biology methods. We investigated the assay’s repeatability and intermediate precision. The results were highly satisfactory across three concentration levels (low, medium, and high). For intermediate precision, the coefficients of variation (CV) were: CV1 = 2.56 %, CV2 = 3.60 % and CV3= 2.70 %, respectively. For repeatability, the coefficients of variation were CV1 = 3.13 %, CV2 = 2.90 and CV3= 2.40 %, respectively. These findings confirm the method’s performance and its alignment with the analytical objectives set by the manufacturer and professional societies. This ensures compliance with both regulatory and normative requirements. Keywords: Thyroglobulin autoantibodies; Analytical performance; Repeatability; Reproducibility; Alinity CI analyzer; Immuno-chemiluminescence 1. Introduction The verification and validation of an analytical method involve assessing the performance of the analytical process, which includes parameters such as precision, accuracy, trueness, measurement range, sensitivity to interference, and detection limit when applicable. This process requires quantifying these performance characteristics using a standardized operating protocol [1], then judging them against criteria defined by learned societies (RICOS, FSCB) or the supplier. Moreover, the verification of an analytical performance is both a regulatory requirement outlined in The Moroccan Guide for the good performance of Medical Laboratory Analysis (GBEA) and a normative standard according to ISO 15189:2022 [2][3], The central laboratory of the Mohammed VI University Hospital in Oujda has instituted a quality strategy encompassing a method verification protocol, of which our study is an integral component. World Journal of Biology Pharmacy and Health Sciences, 2025, 21(01), 427-434 428 In this study, we carried out a method verification protocol for the thyroglobulin autoantibodies (TgAb), using Abbott's Alinity ci® automated system. The aim of our work is to carry out a study which forms an essential basis for an accreditation procedure and is part of the quality process to which our laboratory is strongly committed. 1.1. TgAb and dosing interests Thyroglobulin (TG) is a 660 kDa hyperglycosylated protein expressed in thyrocytes and secreted into the follicular lumen, where it accumulates. Dimeric TG undergoes iodination at specific tyrosine residues in a process influenced by dietary iodine intake. Iodinated TG is transported to the thyrocyte cytosol via pinocytosis and digested, releasing the thyroid hormones triiodothyronine (T3) and thyroxine (T4). TG serves as both a precursor for thyroid hormone biogenesis, which is regulated by thyroid-stimulating hormone (TSH) signaling, and as a carrier protein responsible for iodine storage within the follicular colloid [4] [5]. Thyroglobulin antibodies (TgAb) are predominantly of the IgG class, primarily IgG1, with some IgG2, minimal IgG3, and very low levels of IgG4. TgAb levels tend to fluctuate in parallel with TPOAb (thyroid peroxidase antibody) levels under various clinical conditions, such as during methimazole therapy, after 131I therapy, and during pregnancy or the postpartum period. TgAb are less prevalent than TPOAb in patients with Graves' disease but are present at similar rates in patients with Hashimoto’s thyroiditis and in family members of individuals with autoimmune thyroid disease. [6] TgAb is a crucial marker in the context of thyroid cancer, particularly for detecting residual disease or recurrence. Measuring TgAb levels provides valuable insights into the progression from subclinical thyroid disorders to overt hypothyroidism and enhances our understanding of the immune mechanisms underlying thyroid autoimmunity. [6] While TgAb levels often mirror those of TPOAb in clinical conditions, TgAb holds particular importance in monitoring thyroid cancer but is generally less sensitive than TPOAb for predicting autoimmune thyroid disease. [6] Figure 1 Tg structure and function [5] World Journal of Biology Pharmacy and Health Sciences, 2025, 21(01), 427-434 429 1.2. Principle of the assay method This assay is a two-step immunological test for the quantitative determination of IgG-type autoantibodies directed against thyroglobulin (anti-Tg) in human serum and plasma, utilizing chemiluminescent microparticle immunoassay (CMIA) technology. The sample, paramagnetic microparticles coated with thyroglobulin, and the assay diluent are combined and incubated. Anti-Tg antibodies present in the sample bind to the thyroglobulin-coated microparticles. After washing, an acridiniumlabeled anti-human IgG antibody conjugate is added to form a reaction mixture, which is then incubated. Following another wash cycle, pre-activation and activation solutions are added. The resulting chemiluminescent reaction is measured in relative light units (RLUs). There is a direct relationship between the amount of anti-Tg antibodies in the sample and the RLUs detected by the optical system. 2. Material and methods Our study was carried out in the biochemistry laboratory of the Mohammed VI University Hospital in Oujda over a period of 33 days. TgAb was tested for analytical performance on the Abbott Alinity ci automated system, following the reproducibility and repeatability protocol described in the COFRAC GTA 04 accreditation technical guide. In our study, we chose the multi-specific control as the sample to guarantee more reliable and reproducible results. Three sample levels (Low, Medium and High) were used; each tested 33 times to assess repeatability. Then, over a period of 31 days, we assessed reproducibility by running the control daily at the three levels: low, medium and high. The standard deviation (SD), mean and coefficient of variation (CV) are processed by BYG Informatics EVM statistical software during the evaluation process, and then compared with the standard values of the learned societies (RICOS), no CV reference values were estabilished by Frensh Society of Clinical Biology SFBC. The findings of this investigation are detailed in the sections that follow . 3. Results 3.1. Repeatability Results Repeatability is evaluated by repeatedly analyzing the same samples under uniform conditions by the same operator. This process considers all aspects of measurement, including reagents, calibration, instrumentation, and operator performance, conducted within the shortest possible timeframe. The repeatability test determines initial system performance and verifies the proper functioning of the instrument-reagent system for the analyte in question [7]. Variability is assessed using coefficient of variation (CV) values. As shown in Table 1, the results for the various TgAb assay verification criteria indicate satisfactory repeatability at three level low, medium and high with CV values of 3.13%, 2.90 % and 2.40%, respectively. These findings have been graphically presented using Levey-Jennings plots to further illustrate the obtained results (Figures 2,3 and 4). Table 1 Repeatability results for TgAb on the Alinity i® automated system by level with comparison to RICOS data Levels N Mean Standard deviation (SD) CV % CV% (RICOS) LOW 33 16.71 UI/ml 0.523 UI/ml 3.13 % 3.19 % MEDIUM 33 33.94 UI/ml 0.985 UI/ml 2.90 % 3.19 % HIGH 33 44.31 UI/ml 1.066 UI/ml 2.40 % 3.19 % World Journal of Biology Pharmacy and Health Sciences, 2025, 21(01), 427-434 430 Figure 2 Low Level of Repeatability of TgAb: Levey Jennings graph and the distribution around the mean Figure 3 Medium Level of Repeatability of TgAb: Levey Jennings graph and the distribution around the mean Figure 4 High Level of Repeatability of TgAb : Levey Jennings graph and the distribution around the mean World Journal of Biology Pharmacy and Health Sciences, 2025, 21(01), 427-434 431 3.2. Reproducibility results The reproducibility test involves analyzing the same sample under varying conditions to evaluate the impact of factors such as operators, time, reagent batches, and calibrations on the results. This test aims to define acceptance criteria, particularly for decision support systems. Variability is quantified using the Coefficient of Variation (CV). [7] For the low, medium, and high levels, the CV values are provided (CV1 = 4.25 %, CV2 = 5.12%, CV3 = 5.13 %) These results are illustrated on the Levey-Jennings graphs (Fig. 5,6 and 7fig). Table 1 Reproducibility results for TgAb on the Alinity i® automated system by level with comparison to RICOS data Levels N Mean Standard deviation (SD) CV % CV% (RICOS) LOW 31 30.56 UI/ml 0.781 UI/ml 2.56 % 4.25 % MEDIUM 31 90.30 UI/ml 3.253 UI/ml 3.60 % 4.25 % HIGH 31 199.44 UI/ml 5.391 UI/ml 2.70 % 4.25 % Figure 5 Low level of reproducibility of TgAb : Levey Jennings graph and distribution around the mean Figure 6 Medium level of reproducibility of TgAb : Levey Jennings graph and distribution around the mean World Journal of Biology Pharmacy and Health Sciences, 2025, 21(01), 427-434 432 Figure 7 High level of reproducibility of TgAb : Levey Jennings graph and distribution around the mean 4. Discussion Autoimmune thyroiditis was first identified by Hashimoto in 1912. [8]This form of autoimmunes thyroid disease, characterized by the presence of a goiter, is known as Hashimoto's thyroiditis. In 1956, Roitt et al. [9] were the first to demonstrate the existence of thyroglobulin autoantibodies (TgAb) in patients with this condition, using a precipitin reaction. Unlike TPOAb (thyroid peroxidase antibody), anti-Tg antibodies are not considered pathogenic but rather serve as markers of the disease [10]. These antibodies have been shown to be polyclonal and exhibit heterogeneity in their heavy chain subclasses. [11] [12] Thyroglobulin is a 670,000 Dalton glycoprotein composed of two identical subunits, making it the primary protein in the thyroid gland. It provides 40 of its 140 tyrosine residues for iodination, playing a crucial role in the biosynthesis of thyroxine (T4) and triiodothyronine (T3), and is therefore essential for iodine storage within the thyroid gland. [13] While anti-Tg antibodies commonly occur alongside anti-TPO antibodies in the majority of cases of Hashimoto's thyroiditis, primary myxedema, and Graves' disease [14] [15], they are the sole antibodies present in up to 1% of hypothyroidism cases [16]. These antibodies are associated with mild forms of hypothyroidism or hyperthyroidism and are frequently detected in individuals with other autoimmune conditions, such as rheumatoid arthritis, pernicious anemia, and type I diabetes [17] [18]. Anti-Tg antibodies are found in 30% to 60% of patients with thyroid carcinoma. In these cases, the measurement of thyroglobulin antigen must account for the potential presence of significant concentrations of anti-Tg antibodies, as their presence can affect the accuracy of antigen detection and quantification [19] [20]. Additionally, low levels of antiTg antibodies are identified in about 20% of asymptomatic individuals, particularly among the elderly and more frequently in women than men, though the clinical relevance of these autoantibodies remains uncertain [21] [22]. The Abbott Alinity ci is a multiparametric system capable of integrating clinical chemistry and immunoassay, enabling the measurement of a wide range of standard biochemical parameters as well as specific proteins. The CMIA (microparticle chemiluminescence immunoassay) method is already being utilized for the TgAb assay. As a result, validation is not necessary; instead, we only need to conduct verification according to a ‘‘scope A verification/validation” where the recognized methods, are pre-validated within their designated field of application, to ensure the accuracy and the reliability of our results [3]. This verification is essential, meeting both regulatory standards (as per the Moroccan Guide for the Proper Execution of Medical Laboratory Analyses GBEA and normative requirements (ISO 15189:2022). Setting predetermined analytical goals through this control ensures the production of precise and dependable results [23]. The reproducibility test is employed to assess the consistency of assay results when different variables are introduced. [23]Our study results affirmed the reliability of the TgAb assay for reproducibility assessment. The three levels low, medium and high yielded satisfactory outcomes. For each level, 31 values were analyzed, revealing means of m1 = 0.74 IU/ml and m2 = 25.19 IU/ml, along with coefficients of variation (CV) of CV1 = 2.56 %, World Journal of Biology Pharmacy and Health Sciences, 2025, 21(01), 427-434 433 CV2 =3.60% and CV3= 2.70 %. The low CV values signify that even when modifying various factors, the test consistently produces results close to the mean value. This reliability is crucial in medical testing, where consistency ensures the dependability of test results for clinical decisions. The fact that CV values align with established quality control limits indicates that the test adheres to industry standards for reproducibility, enhancing its suitability for precise diagnostic applications. The precision of the assay under regulated and ideal circumstances is the main emphasis of the repeatability test. This evaluation is crucial as it gauges the method's capability to produce consistent results when analyzing the same sample repeatedly. In examining the repeatability across two levels (low and high), 30 values were scrutinized for each level, revealing remarkably low coefficients of variation (CV): CV1 = 3.13 %, CV2 = 2.90 % and CV3 = 2.40%. These values indicate a small degree of variability, underscoring the high precision of the assay. The extremely low CV values highlight the assay's outcomes as highly stable and predictable when operating under controlled conditions. Such precision is of utmost importance in clinical testing, where even minor variations can carry significant implications for patient care. The Mohammed VI University Hospital's central laboratory in Oujda has implemented a quality strategy incorporating a method verification protocol. Conducting this type of investigation will enable the establishment of a credible accreditation process for the analyses conducted in our laboratory [24]. As a pivotal reference center in the Eastern region of Morocco, our laboratory serves not only the needs of referred or hospitalized patients but also contributes to assessing the overall health of the region's general population through various scientific studies [25 ]. 5. Conclusion The verification of the TgAb assay demonstrated robust analytical performance, with excellent precision and reproducibility across all tested levels. These findings establish the assay's reliability for routine clinical applications, particularly in diagnosing and monitoring of autoimmune thyroid disorders and differentiated thyroid cancer (DTC). By meeting ISO 15189 and other regulatory standards like RICOS standards, the Mohammed VI University Hospital’s central laboratory underscores its dedication to delivering accurate diagnostic results, enhancing both patient care and regional health outcomes. Compliance with ethical standards Disclosure of conflict of interest No conflict of interest to be disclosed. References [1] Vassault A, Hulin A, Chapuzet E, Arnaud J, Giroud C, members of the SFBC subgroup 2 on analytics. [Verification/validation of the performances of analytical methods]. Ann Biol Clin (Paris). Dec 2010;68 Spec No 1:247-94. [2] Masson E. EM-Consulte. [cited 2024 Jan 14]. Guide to good practice in medical biology testing (GBEA): its application in the public hospital sector. Available from: https://www.em-consulte.com/article/191898/guidede-bonne-execution-des-analyses-de-biologie- [3] International Organization for Standardization (ISO). ISO 15189:2022 – Medical laboratories – Requirements for quality and competence. [cited 2024 Jan 14]. Available from: https://www.iso.org/standard/76677. [4] Citterio, C. E., Targovnik, H. M., & Arvan, P. (2019). The role of thyroglobulin in thyroid hormonogenesis. Nature Reviews Endocrinology, 15(5), 323–338. World Journal of Biology Pharmacy and Health Sciences, 2025, 21(01), 427-434 434 [5] Di Jeso, B., & Arvan, P. (2016). Thyroglobulin from molecular and cellular biology to clinical endocrinology. Endocrine Reviews, 37(1), 2–36. [6] Spencer, C., & Fatemi, S. (2013). Thyroglobulin antibody (TgAb) methods – Strengths, pitfalls and clinical utility for monitoring TgAb-positive patients with differentiated thyroid cancer. Best Practice & Research Clinical Endocrinology & Metabolism, 27(5), 701–712. https://www.elsevier.com/locate/beem [7] Technical guide for accreditation, verification (scope A)/validation (scope B) of medical biology methods, Document SH GTA 04, Revision 02, COFRAC. [8] Hashimoto H. On the knowledge of lymphomatous changes in the thyroid gland (Struma lymphomatosa). Arch Klin Chir. 1912; 97:219-248. [9] Roitt IM, Doniach D, Campbell PN, et al. Auto-antibodies in Hashimoto’s Disease (Lymphadenoid Goitre). Lancet. 1956; 6947:820-821. [10] Tomer Y. Anti-thyroglobulin autoantibodies in autoimmune thyroid diseases: cross-reactive or pathogenic? Clin Immunol Immunopath. 1997; 82:3-11. [11] Laing P. Both K and λ light chain types are present in thyroid microsomal and thyroglobulin autoantibodies. Proc Univ Otago Me. Sch. 1983; 61:75-77. [12] McIntosh RS, Asghar MS, Weetman AP. The antibody response in human autoimmune thyroid disease. Clinical Science. 1997; 92:529541. [13] DeGroot LJ, Larsen PR, Hennemann G, editors. Thyroid hormone synthesis and secretion. In: The Thyroid and its Diseases 6th ed. New York: Churchill Livingstone; 1996:45-48. [14] Rosenbaum D, Davies TF. The clinical use of thyroid autoantibodies. The Endocrinologist. 1992;2(1):55-62. [15] Burek CL, Rose NR. Thyroglobulin autoantibodies. In: Peter JB and Shoenfeld Y, editors. Autoantibodies. Amsterdam: Elsevier Science B.V.;1996: 810-815. [16] Nordyke RA, Gilbert FI Jr, Miyamoto LA, et al. The superiority of antimicrosomal over antithyroglobulin antibodies for detecting Hashimoto’s thyroiditis. Arch Intern Med. 1993; 153:862-865. [17] Ruf J, Feldt-Rasmussen U, Hegeds L, et al. Bispecific thyroglobulin and thyroperoxidase autoantibodies in patients with various thyroid and autoimmune diseases. J Clin Endocrinol Metab. 1994;79(5):14041409. [18] Walker DJ, Griffiths M, Griffiths ID. Occurrence of autoimmune diseases and autoantibodies in multicase rheumatoid arthritis families. Ann Rheum Dis. 1986; 45:323-326. [19] Feldt-Rasmussen U, Rasmussen K. Serum thyroglobulin (Tg) in presence of thyroglobulin autoantibodies (TgAb). Clinical and methodological relevance of the interaction between Tg and TgAb in vitro and in vivo. J Endocrinol Invest. 1985; 8:571-576. [20] Schaadt B, Feldt-Rasmussen U, Rasmusson B, et al. Assessment of the influence of thyroglobulin (Tg) autoantibodies and other interfering factors on the use of serum Tg as tumor marker in differentiated thyroid carcinoma. Thyroid. 1995;5(3):165-170. [21] Ericsson U-B, Christensen SB, Thorell JI. A high prevalence of thyroglobulin autoantibodies in adults with and without thyroid disease as measured with a sensitive solid-phase immunosorbent radioassay. Clin Immunol Immunopathol 1985; 37:154-162. [22] KIDOUN, Asmae & DOUICHI, Imane & NACIRI, Mehdi & HALAS, Meryem & KRIMI, Kholoud & Moujtahide, Dounia & Sebbar, El-Houcine & Choukri, Mohammed. (2024). Verification of the analytical performance of the anti-cyclic citrullinated peptide assay on ALINITY ci ® experience from the biochemistry laboratory of Mohammed VI University Hospital in Oujda. World Journal of Biology Pharmacy and Health Sciences. 18. 001-008. 10.30574/wjbphs.2024.18.3.0320. [23] Beyyoudh, S., Mokhtari, I., Grari, O., Douzi, N., eddine El Khamlichi, I., houcine Sebbar, E., & Choukri, M. (2023). Verification of analytical performance of the aspartate aminotransferase assay on the Abbott Alinity ci®: Experience of the central laboratory Mohammed VI Oujda. World Journal of Biology Pharmacy and Health Sciences, 15(3), 037-042 [24] SEBBAR, El-Houcine, SAALAOUI, Ennouamane, CHOUKRI, Mohammed. Evaluation of dietary vitamin D intake in a population in eastern Morocco. Revue Pratiques en Nutrition, 2018, Vol 14 - N 55, p. 44-45 Doi : 10.1016/j.pranut.2018.05.012. [25] SEBBAR El-Houcine, SAM Hicham, SIDQI Zaina, et al. Calcium Intakes in the Diet of Eastern Morocco’s Population. Journal of Pharmacy and Nutrition Sciences, 2018, vol. 8, no 3, p. 91-96. DOI: https://doi.org/10.6000/19275951.2018.08.03.2.