Comparison of visual evoked potential variability in eyes affected by optic neuritis and fellow eyes
Abstract
The copyright was changed from CC-BY-ND to CC-BY. Check for a new version.
Full text
Vol.: (0123456789) Doc Ophthalmol https://doi.org/10.1007/s10633-025-10061-y ORIGINAL RESEARCH ARTICLE Comparison ofvisual evoked potential variability ineyes affected byoptic neuritis andfellow eyes MarieChutná · JanKremláček · MiroslavKuba · ZuzanaKubová · JanaSzanyi · FrantišekVít · JanaLangrová Received: 11 September 2024 / Accepted: 20 October 2025 © The Author(s) 2025 Abstract Purpose This study compared the variability of visual evoked potential (VEP) in response to stimulation of eyes affected by unilateral optic neuritis with that of fellow (non-affected) eyes. Methods Pattern-reversal VEP (PVEP) and motiononset VEP (MVEP) recordings from thirty-six subjects with unilateral optic neuritis at different intervals from disease onset were retrospectively evaluated, and differences in the following parameters were compared: signal‒to‒noise ratio (SRN), interquartile range of the response jitter (jitter IQR), and number of trials corresponding to the average response (corresponding N). Results In the PVEP recordings, the P1 peak times of the fellow eyes were significantly shorter than those of the affected eyes (Cohen’s d = -1.470, p < 0.001). P1 amplitudes were significantly greater in fellow eyes (d = 1.17, p < 0.001). Significant differences were found in the SNR (d = 0.782, p < 0.001), jitter IQR (d = -0.874, p < 0.001), and corresponding N (d = 0.700, p < 0.001). MVEP presented significantly shorter N2 peak times in fellow eyes than in affected eyes (d = 0.840, p < 0.01) and significantly greater amplitudes (d = 0.494, p = 0.002). There was a significant difference in the SNRs (d = 0.440, p = 0.01) and corresponding N values (d = 0.415, p = 0.01). There was no difference in the jitter IQR (d = 0.143, p = 0.230). Conclusions The increased variability in eyes affected by optic neuritis compared with fellow eyes (in particular, in pattern-reversal VEP, which predominantly represents the activity of the macular– papillary fibers of the optic nerves) may represent important pathophysiologic features and may add valuable information to diagnostics via VEP examinations. Keywords Signal‒to‒noise ratio· Averaging methods· Latency jitter· Trial–to–trial variability Introduction The visual evoked potential (VEP) provides objective functional assessment of the visual pathway, and since its first clinical usage [1], it has proven to be a useful tool for both clinical and experimental investigations of visual pathways [2]. Averaging is often used for VEP processing. This method has enabled tremendous progress in VEP M.Chutná(*)· M.Kuba· Z.Kubová· J.Szanyi· F.Vít· J.Langrová Department ofPathological Physiology, Faculty ofMedicine inHradec Kralove, Charles University, Simkova 870, 50003HradecKralove, CzechRepublic e-mail: [email protected] J.Kremláček Department ofMedical Biophysics, Faculty ofMedicine inHradec Kralove, Charles University, Simkova 870, 50003HradecKralove, CzechRepublic
Doc Ophthalmol Vol:. (1234567890) investigations and has made this method available for routine clinical diagnostics [3]. However, evoked responses of the central nervous system are rather variable even for a precisely defined stimulus, which is kept constant for all trials; thus, due to averaging, some information about the variability of VEP could be lost [4]. The degree of variability could be affected by several factors, e.g., the current conditions of the examined persons, their ability to maintain attention during the examination, by brain adaptation to the visual stimuli [5–7], and, as several studies have suggested, the level of variability in the CNS may be functionally relevant, and may even be linked to some pathological conditions [8–13]. Increased variability may also affect the final trace gained by the averaging of single trials and possibly interfere with test interpretation. For example, time fluctuation of single trials, a phenomenon called latency jitter, may cause an amplitude decrease in averaged recordings or can affect the peak shape, so assessing a VEP on the basis of only the nominal peak time may not be sufficient in some cases [9, 14–16]. Therefore, it would be useful to explore whether there is increased variability in diseases investigated by VEP, and whether the assessment of VEP response variability might serve as beneficial additional information to increase the diagnostic yield of the investigation in general. Assessment of variability is often based on single trial examinations, but our aim was to find simple and easily evaluable parameters that reflect the variability of VEP and to examine whether the variability is greater in a pathological condition. For this purpose, we evaluated variability in eyes affected by unilateral optic neuritis compared with fellow eyes. Optic neuritis (ON) is an inflammatory disorder of the optic nerve that causes unilateral acute impairment of vision. It can affect one or both eyes either simultaneously or successively [17]. There could be several underlying pathologies, but one of the most common is multiple sclerosis [18]. The diagnosis of ON was chosen because it represents one of the most common and thoroughly studied pathologies for VEP examination, and when it is unilateral, it enables variability of affected and healthy eyes to be compared within the same subject to the other eye, thus minimizing the interference from inter-subject variability which is quite high [15]. Two types of VEP stimuli were analyzed in this study: pattern-reversal VEP (PVEP), which represents the most frequently employed VEP, and motiononset VEP (MVEP). MVEP are routinely examined in our laboratory because a wider range of stimulus types can increase the diagnostic sensitivity of VEP examination [2, 19], and in some cases, especially if the magnocellular pathway is predominantly affected, pathological findings can be observed only in MVEP, whereas PVEP is normal [20–22]. Subjects andmethods Subjects The data represent retrospective VEP analysis of patients saved at the database of the Electrophysiological Laboratory in the Department of Pathological Physiology in Hradec Kralove from 2010 to 2024 (examined under strictly uniform conditions). Thirty-six subjects were chosen for the assessment, 13 males and 23 females, aged 22–50years, with a mean age of 33.9years (see Table1 for demographic data). There were more females in the sample, which is consistent with the incidence of optic neuritis being greater in females [23]. The inclusion criteria were as follows: age 18–65years, visual acuity assessment available, readable VEP in both eyes for at least one of the stimuli: pattern-reversal VEP (PVEP)/motion-onset VEP (MVEP), and clinical diagnosis of unilateral optic neuritis based on suggestive symptoms (confirmed by the clinical expert who made the referral for the VEP investigation) and supported by a paraclinical test, an abnormal VEP. The typical clinical course and symptoms, together with supporting evidence from one paraclinical test, are regarded as sufficient for the diagnosis of optic neuritis [24, 25]. To confirm an abnormal VEP, we assessed the peak time (latency) of the main peak P1/N2, its amplitude and the interocular difference. The assessment was based on comparisons with normative data for the adequate age group set in our laboratory. Optic nerve affection was defined as a prolonged nominal peak time and/or significant interocular nominal peak time/amplitude asymmetry or the absence of the P1/N2 wave. The exclusion criteria included a history of optic neuritis in the other eye, other ocular comorbidities and diseases potentially affecting vision (such as
Doc Ophthalmol Vol.: (0123456789) diabetes mellitus, amblyopia, and neuroborreliosis). Patients older than 65 years were not included because it cannot be ruled out that there may be an effect of even healthy aging on VEP variability [26]. Patients younger than 18 years were not included because the visual system is known to mature during childhood and adolescence, which is reflected especially by the response to motion stimulation [27]. Ethics All the patients clearly agreed with the testing and possible subsequent use of their data for research. This analysis was a part of a larger research project Table 1 Demographics and visual acuity at the time of testing VA – Visual acuity; FE – Fellow eye; AE – affected eye Patient no Age Gender VA FE VA AE logMAR VAFE logMAR VAAE 1 42 F 1.00 0.25 0.00 0.60 2 24 M 1.00 0.13 0.00 0.90 3 46 M 0.80 0.20 0.10 0.70 4 50 F 0.25 0.13 0.60 0.90 5 22 F 0.40 0.10 0.40 1.00 6 38 F 0.80 0.25 0.10 0.60 7 30 M 1.00 1.00 0.00 0.00 8 24 F 1.00 0.40 0.00 0.40 9 38 M 0.80 0.50 0.10 0.30 10 22 F 0.67 0.17 0.17 0.77 11 36 F 1.00 0.50 0.00 0.30 12 26 F 1.00 0.67 0.00 0.17 13 46 M 0.80 0.80 0.10 0.10 14 25 F 0.50 0.13 0.30 0.90 15 35 M 1.00 0.10 0.00 1.00 16 34 M 1.00 0.80 0.00 0.10 17 45 F 1.00 0.10 0.00 1.00 18 39 F 1.00 0.20 0.00 0.70 19 46 M 1.00 0.33 0.00 0.48 20 47 F 0.80 0.20 0.10 0.70 21 31 F 0.50 0.25 0.30 0.60 22 33 F 0.40 0.20 0.40 0.70 23 31 M 0.25 0.10 0.60 1.00 24 41 M 0.80 0.13 0.10 0.90 25 23 F 1.00 1.00 0.00 0.00 26 22 F 1.00 0.10 0.00 1.00 27 31 F 0.67 0.67 0.17 0.17 28 36 F 1.00 0.33 0.00 0.48 29 28 M 1.00 0.80 0.00 0.10 30 39 M 0.25 0.17 0.60 0.77 31 26 F 1.00 0.50 0.00 0.30 32 44 F 1.00 1.00 0.00 0.00 33 23 M 0.40 0.25 0.40 0.60 34 28 F 1.33 0.80 -0.12 0.10 35 32 F 1.00 1.00 0.00 0.00 36 36 F 0.80 0.40 0.10 0.40
Doc Ophthalmol Vol:. (1234567890) that adhered to the tenets of the Declaration of Helsinki and was approved by the Ethical Committee of the Medical Faculty in Hradec Kralove. Visual acuity Information on visual acuity was obtained via the Landolt C test, and the logMAR score was used for statistical analysis (see Table1). Visual stimulation Two types of stimulation were chosen for the analysis: pattern reversal VEP (PVEP) and motion onset VEP – centrifugal/centripetal motion (‘‘expansion/ contraction motion” – MVEP). There are some differences in our parameters from the current ISCEV standards for PVEP examination [28]. The reason for this is that our laboratory would like to maintain the continuity of our stimulus and recording parameters and published results for a long period of time, starting in approximately 1990. Pattern-reversal stimulation (PVEP) was elicited using a black and white checkerboard (contrast 96% according to Michelson) with check sizes of 40’ reversing at a frequency of 2/s. For the standard PVEP examination we routinely use three checker sizes (40´, 20´and 10´ or more in the case of low visual acuity). Smaller check sizes were not analyzed because in many patients, the responses to smaller checks were not well readable and therefore not suitable for this type of analysis. Motion onset VEP (MVEP) was elicited using the radial motion of low luminance contrast (10%) circles with randomly selected centrifugal or centripetal (‘‘expansion/contraction ”) directions [20]. The spatial frequency of the structure decreased, and the motion velocity increased from the center (fixation point) toward the periphery, respecting the size of the retinal receptive fields and the sensitivity to motion velocity across the retina [29]. The temporal frequency of this stimulus was 5Hz in all parts of the visual field. The timing was 200 ms of motion followed by a one second interstimulus interval, during which the stationary pattern was presented, to prevent adaptation to motion. All visual stimuli were generated via VSG 2.5; CRS Ltd., UK on a 21’’ Iyama monitor (Japan) with a vertical frequency of 105 Hz. The stimulus field subtended 37 × 28° at a viewing distance of 0.6 m, and the average luminance of 17cd/m2 was constant. We use rather low luminance, different from ISCEV guidelines [28], because it may lead to increased examination sensitivity in optic nerve pathologies without increasing falsely positive results [19, 30]. The electrophysiological examination was performed in a Faraday cage to decrease the electromagnetic noise. Patients were instructed to keep their gaze on the red cross in the center of the screen during recording. To minimize the influence of poor fixation, we used an infrared CCD camera to check the correct fixation of the stimulus field center, and according to an assessment of examinators, all the chosen patients fixated quite well. All recordings were performed monocularly with the non-tested eye covered with an eye patch. All patients were tested with their best spectacle correction if needed, and no additional short-distance correction (for 0.6m viewing distance) was provided. Ag–AgCl electrodes for the VEP recording were placed according to the International 10–10 system [31] using pseudounipolar derivations from the midline Oz, Pz, Cz and Fz. Two additional electrodes were placed 5cm to the left (OL) and right (OR) of the Oz position as originally suggested by Göpfert etal. [32], since at these locations, MVEP has a maximum amplitude (dependent on individual lateralization). The reference electrode was placed on the right ear lobe. EEG poststimulus epochs of 440ms duration were sampled at 500Hz. The signal was amplified 20,000 times (Contact Precision Instruments – PSYLAB, System 5, UK) and filtered to include frequencies ranging from 0.3 to 45 Hz. Forty single trials were recorded and averaged, because this number of trials in our recording settings provides a good signal‒to‒ noise ratio. We standardly use more (six) different types of stimulation for routine clinical VEP investigations (40´, 20´and 10´ sizes of PVEP and three types of MVEP) because we believe it increases the diagnostic yield of the investigation [2]. However, the examination takes longer, and patients’ fatigue and decreased attention might interfere with the results [33]. For these reasons, we record forty single responses, not the hundred responses recommended by ISCEV [28].
Doc Ophthalmol Vol.: (0123456789) Variability analysis To assess VEP variability, we set two parameters that reflect the variability of the recordings: 1. jitter IQR represents the interquartile range of the latency shifts (jitters) in milliseconds applied to individual responses (epochs) during the alignment process with a template waveform. Single responses were subsequently shifted to match an iteratively constructed average potential (template) using Woody’s approach [34]. This iterative process continued until the root mean square (RMS) of the difference between succeeding averages was less than 0.1, indicating convergence and optimal alignment. The fitting between each response and the template was based on the cross-correlation function computed over a 1–440 ms time window, encompassing the entire recorded epoch of the VEP response. The latency shifts applied to each epoch were recorded as jitters, reflecting the adjustments needed for optimal alignment with the template. We evaluated the variability in response timing across trials by calculating these jitters’ interquartile range (IQR). A smaller jitter IQR indicates more consistent neural response timing across trials, suggesting lower variability in neural processing. Conversely, a larger IQR signifies greater variability in the timing of VEP responses. 2. corresponding N represents the number of epochs that closely correspond with the template VEP within a ± 30 ms jitter window. By limiting the jitter window, we avoid falsely enhancing noise due to misalignment with background EEG activity. The ± 30 ms was selected to be shorter than a shift that erroneously aligned counter-phase alpha activity to the average. In addition, we analyzed the signal-to-noise ratio (SNR) because it is an important parameter that reflects the actual change in the response. If we assume constant noise levels, a decrease in the SNR may reflect both a diminished signal amplitude and increased time variability. The SNR calculation was based on the ratio between the root mean square (RMS) of the average evoked response and the reference noise calculated using the "plus-minus" principle [35]. The average waveform is computed from single trials to estimate the signal level. The plus-minus averaging technique is employed to assess the noise level. This involves inverting the polarity of half of the recorded trials before averaging. When these inverted and non-inverted averages are combined, the consistent VEP signals cancel each other out because of their opposite polarities, effectively isolating the random noise components. The RMS values of the signal and noise waveforms are then calculated. The SNR is then determined by dividing the RMS of the averaged VEP response by the RMS of the noise estimated through the plusminus method Analysis of the recordings Recordings from the Oz derivation were evaluated in all patients for PVEP stimulation. For MVEP, the optimal derivation, which represents the derivation with the largest amplitude of the main motionspecific peak N2 (with a normal peak time of app. 160–200 ms), [20] was chosen for each patient individually. The recordings were evaluated in MATLAB R2023a (MathWorks, USA). In each recording, the peak time and the interpeak amplitude of the main peak were determined, as were the signal‒to‒noise ratio (SNR), the jitter IQR, and the corresponding N (for definitions, see above). In three participants the PVEP was undetectable; in three other participants, the MVEP was undetectable; thus, thirty-three PVEP and thirty-three MVEP recordings were analyzed. Statistical analysis Statistical tests were performed and graphs were generated with Jamovi (version 2.3.17) [36]. Based on the Shapiro–Wilk W test, some of the data sets did not fulfill the criteria of normality; therefore, they were summarized with the median and ranges, and the Wilcoxon pair test was used for the assessment of the differences in variability between affected and fellow eyes. Cohen’s d was also calculated to illustrate the effect size corresponding to the parametric tests. SNR = RMS of Average Response RMS of Reference Noise
Doc Ophthalmol Vol:. (1234567890) Table 2 Summarized results of the Pattern reversal VEP (PVEP) parameters. The last column includes p-values for each paired test, which indicate the significance of the observed differences between AE and FE. Additionally, 95% confidence intervals (CI) for the estimated difference in means between the eyes are provided, showing the range within which the true difference is likely to fall with 95% confidence. Values (Q1, Q3) represent lower and upper quartiles. Statistically significant differences that were subjected to Holm-Bonferroni correction are marked in bold, appropriate alpha levels were as follow: 0.01, 0.0125, 0.0166, 0.025, and 0.05 FE – fellow eye; AF – affected eye; SNR – signal-to-noise ratio; CI – Confidence Interval FE Median (Q1, Q3) AE Median (Q1, Q3) FE – AE (95% CI) Wilcoxon paired test: p value P1 peak time [ms] 112.0 (106.0, 118.0) 132.0 (118.0, 138.0) (-inf, -14.0) p < 0.001 P1 amplitude [µV] 12.2 (8.5, 16.9) 7.6 (5.0, 12.3) (3.1, inf) p < 0.001 SNR [-] 9.6 (4.9, 20.1) 3.8 (2.3, 8.0) (4.06, inf) p < 0.001 Jitter IQR [ms] 7.5 (5.0, 10.0) 12.0 (9.0, 16.0) (-inf, -3.25) p < 0.001 Corresponding N [-] 38.0 (36.0, 40.0) 36.0 (33.0, 38.0) (2.0, inf) p < 0.001 Fig. 1 Pattern-reversal VEP (PVEP) (1a) and motiononset VEP (MVEP) (1b) comparisons of jitter IQR and Corresponding N values between fellow and affected eyes. Scatter plots illustrating the relationship between variability parameters values for each patient, comparing the fellow eye (FE, horizontal axis) to the affected eye (AE, vertical axis). Each dot represents an individual patient’s data point. The dotted diagonal line indicates the identity line (i.e., where fellow and affected eye values would be equal). Panel 1b: MVEP – Left Plot Scatter plot depicting corresponding N values. More points falling below the identity line suggest that the fellow eye often has higher corresponding N values than the affected eye (p = 0.01). Right Plot: Scatter plot depicting jitter IQR values of the fellow and affected eye. Here, the difference was not statistically significant (p = 0.23).
Doc Ophthalmol Vol.: (0123456789) The Spearman correlation coefficient was used to test the correlation between amplitude magnitude and variability across trials (reflected by the SNR, jitter IQR and corresponding N). Within subject (paired) comparisons were chosen to eliminate interindividual variability in VEP parameters. Significance was assessed at the alpha level 0.05. For multiple testing correction, we used the Holm‒ Bonferroni correction to adjust the alpha level appropriately [37]. Clustered ROC (Receiver Operating Characteristic) analysis was performed to illustrate the discriminative performance of the following parameters. ROC analysis was conducted in the R environment (ver. 4.3.1) [38] based on a nonparametric analysis of clustered ROC [39, 40] Fig. 2 Pattern-reversal VEP (PVEP) (2a) and motion-onset VEP (MVEP) (2b) comparisons of the nominal peak time, amplitudes and SNR between fellow and affected eyes. Scatter plots illustrating the relationship between nominal peak time, amplitudes and SNR values for each patient, comparing the fellow eye (FE, horizontal axis) to the affected eye (AE, vertical axis). Each dot represents an individual patient’s data point. The dotted diagonal line indicates the identity line (i.e., where fellow and affected eye values would be equal). Panel 1a: PVEP—Left Plot Scatter plot illustrating P1 peak times. A clear asymmetry around this line, with almost all points falling above it, shows that the affected eye has longer P1 peak time than the fellow eye (p < 0.001). Middle Plot: Scatter plot depicting P1 amplitudes. A clear asymmetry around the identity line, with almost all points falling below it, shows that the affected eye has lower P1 amplitude than the fellow eye (p < 0.001). Right Plot: Scatter plot illustrating SNR values with more points falling below the identity line suggesting that the fellow eye often has higher SNR values than the affected eye (p < 0.001). Panel 1b: MVEP – Left Plot Scatter plot depicting N2 peak times. A visible asymmetry around the identity line, with more points falling above it, shows that the affected eye has longer N2 peak time than the fellow eye (p < 0.001). Middle Plot: Scatter plot illustrating the relationship between N2 amplitudes with more points falling below the identity line demonstrating greater amplitude of fellow eye compared to the affected eye (p = 0.002). Right Plot: Scatter plot illustrating SNR values with more points falling below the identity line suggesting that the fellow eye often has higher SNR values than the affected eye (p = 0.01).
Doc Ophthalmol Vol:. (1234567890) Results The results for pattern-reversal VEP (PVEP) and motion-onset VEP (MVEP) are summarized in Tables2 and 3 and areplotted in Figs.1, 2, and 3 Panel 1a: PVEP—Left Plot Scatter plot depicting corresponding N values. A visible asymmetry around the identity line, with most points falling below it, suggests that the fellow eye often has higher corresponding N values than the affected eye (p < 0.001). Right Plot: Scatter plot depicting jitter IQR values. A visible asymmetry around the identity line, with most points falling above it, suggests that fellow eye often has lower jitter IQR values than the affected eye (p < 0.001), indicating more consistent temporal response patterns in the fellow eye. Pattern-reversal VEP (PVEP) Significant differences between fellow and affected eyes were found in the P1 peak time, P1 amplitude, signal–to–noise ratio (SNR), jitter IQR and corresponding N. The median P1 peak time was 112.0ms (Q1, Q3: 106.0 ms, 118 ms) for the fellow eyes and 132.0 (118.0, 138.0) for the affected eyes. The P1 peak times in the fellow eyes were significantly shorter than those in the affected eyes (d = -1.470, p < 0.001; 95% Confidence Interval: -inf, -14.0) as expected, because the peak time was used as a diagnostic marker and selection criterion. The P1 interpeak amplitudes of the fellow eyes, with a median of 12.2µV (Q1, Q3: 8.5µV, 16.9µV), were significantly greater (d = 1.170, p < 0.001, 95% CI: 3.1, inf) than those of the affected eyes, with a median of 7.6µV (5.0, 12.3µV). The median SNR for the fellow eyes was 9.6 (Q1, Q3: 4.9, 20.1), which was significantly higher than that for the affected eyes (d = 0.782, p = < 0.001(95% CI: 4.06, inf), where the median was 3.8 (Q1, Q3: 2.3, 8.0). The median jitter IQR was 7.5 ms (Q1, Q3: 5.0, 10.0 ms) for the fellow eyes and 12.0 ms (Q1, Q3: 9.0, 16.0ms) for the affected eyes, which represent a significant difference, d = -0.874, p < 0.001 (95% CI: -inf, -3.25). The values of the corresponding N were also significantly different, d = 0.700, p < 0.001 (95% CI: 2.0, inf). The median values for the fellow eyes were 38.0 (36.0, 40.0) and for the affected eyes 36.0 (33.0, 38.0). Fig. 3 Illustrative traces of response variability in a patient with unilateral optic neuritis (female, age 22). 1a) PVEP fellow eye – final trace, 1b) PVEP fellow eye – final trace divided into four subgroups of trials (quarters), 1c) PVEP affected eye – final trace, 1d) PVEP affected eye – final trace divided into four subgroups of trials, 2a) MVEP fellow eye – final trace, 2b) MVEP fellow eye – final trace divided into four subgroups of trials, 2c) MVEP affected eye – final trace, 2d) MVEP affected eye – final trace divided into four subgroups of trials
Doc Ophthalmol Vol.: (0123456789) Spearman correlation coefficients were calculated to confirm the assumption that increased variability may underlie a reduction in amplitude. The Spearman correlation coefficient between the amplitudes and SNRs was 0.558 (p < 0.001) for the affected eyes and 0.592 (p < 0.001) for the fellow eyes. The Spearman correlation coefficient between the amplitudes and corresponding N was 0.631 (p < 0.001) for the affected eyes and 0.351 (p = 0.022) for the fellow eyes. The Spearman correlation coefficient between the amplitudes and jitter IQRs was -0.514 (p = 0.001) for the affected eyes and -0.546 (p < 0.001). ROC analysis for illustration of the discrimination power of the assessed parameters revealed the highest area under the curve (AUC) of 86.5% (95% confidence interval: 79.44%-93.56%) for P1 peak time, for the SNR (74.75; 65.86–83.64%), for the jitter IQR (78.01; 66.24–86.77%), and for the corresponding N (71.58; 62.24–86.77%) the AUC was smaller, although not significantly. Motion-onset VEP (MVEP) Significant differences between fellow and affected eyes were found in the peak times of the main motiononset specific N2 peaks and their amplitudes and in the SNRs and corresponding N values of the MVEP recordings. No significant difference was found in the jitter IQR. The peak times of the main motion-onset specific N2 peaks were significantly shorter (d = -0.840, p < 0.001 95% CI: -inf, -9.00) in the fellow eyes, with a median of 158.0 ms (Q1, Q3: 152.0, 170.0 ms) than in the affected eyes, with a median of 174.0ms (166.0 ms, 178.0 ms). The amplitudes were greater (d = 0.494, p = 0.002). The amplitudes of N2 peak were significantly greater (d = 0.494, p = 0.002; 95% CI: 0.552, 0.780) in the fellow eyes, median 12.0µV (Q1, Q3: 9.3µV, 14.7µV), than in the affected eyes, with a median of 9.7µV (Q1, Q3: 8.4µV, 14.6µV). The median SNR for the fellow eyes was 11.3 (Q1, Q3: 7.5, 14.9), which was significantly higher than that for the affected eyes (d = 0.440, p = 0.010; 95% CI: 1.01, inf), where the median was 7.3 (Q1, Q3: 4.1, 15.3). No significant difference was found in the jitter IQR (d = -0.143, p = 0.230; 95% CI: -inf, 1.25), where the median was 13.0ms (Q1, Q3: 8.0, 18.0ms) for the fellow eyes and 15.5ms (Q1, Q3: 8.0, 20.0ms) for the affected eyes. The values of the corresponding N differed significantly, d = 0.415, p = 0.010; 95% CI: 0.5, inf). The median values for the fellow eyes were 36.0 (Q1, Q3: 33.0, 38.0) and for the affected eyes 35.0 (Q1, Q3: 33.0, 36.0). Spearman correlation coefficients were also calculated for the MVEP recordings to confirm the assumption that increased variability may underlie a reduction in amplitude: The Spearman correlation coefficient between the amplitudes and SNRs was 0.380 (p = 0.015) for the affected eyes and 0.364 (p = 0.019) for the fellow eyes. The Spearman correlation coefficient between the amplitudes and corresponding N was 0.568 (p < 0.001) for the affected eyes and 0.573 (p < 0.001) for the fellow eyes. The Table 3 Summarized results of the Motion-onset VEP (MVEP) parameters. The last column includes p-values for each paired test, which indicate the significance of the observed differences between AE and FE. Additionally, 95% confidence intervals (CI) for the estimated difference in means between the eyes are provided, showing the range within which the true difference is likely to fall with 95% confidence. Values (Q1, Q3) represent lower and upper quartiles. Statistically significant differences that were subjected to Holm-Bonferroni correction are marked in bold, appropriate alpha levels were as follow: 0.01, 0.0125, 0.0166, 0.025, and 0.05 FE – fellow eye; AF – affected eye; SNR – signal-to-noise ratio; CI – Confidence Interval FE Median (Q1, Q3) AE Median (Q1, Q3) FE – AE (95% CI) Wilcoxon paired test: p value N2 peak time [ms] 158.0 (152.0, 170.0) 174.0 (166.0, 178.0) (-inf, -9.00) p < 0.001 N2 amplitude [µV] 12.0 (9.3, 14.7) 9.7 (8.4, 14.6) (0.552, 0.780) p = 0.002 SNR [-] 11.3 (7.5, 14.9) 7.3 (4.1, 15.3) (1.01, inf) p = 0.01 Jitter IQR [ms] 13.0 (8.0, 18.0) 15.5 (8.0, 20.0) (-inf, 1.25) p = 0.23 Corresponding N [-] 36.0 (33.0, 38.0) 35.0 (33.0, 36.0) (0.5, inf) p = 0.01