Full text
Evaluation of impurity densities from charge exchange recombination spectroscopy measurements at ASDEX Upgrade R M McDermott 1 , R Dux 1 , T Pütterich 1 , B Geiger 1,2 , A Kappatou 1 , A Lebschy 1,3 , C Bruhn 1,3 , M Cavedon 1 , A Frank 4 , N den Harder 1 , E Viezzer 5 and the ASDEX Upgrade Team 1 Max-Planck-Institut für Plasmaphysik, Boltzmannstr. 2, D-85748 Garching, Germany 2 Max-Planck-Institut für Plasmaphysik, Wendelsteinstr. 1, D-17491 Greifswald, Germany 3 Physik-Department E28, Technische Universität München, Garching, Germany 4 Physik-Department, Ludwig-Maximilians-Universität, München, Germany 5 Department of Atomic, Molecular and Nuclear Physics, University of Seville, Spain E-mail: [email protected] Received 24 April 2018, revised 20 June 2018 Accepted for publication 10 July 2018 Published 26 July 2018 Abstract At ASDEX upgrade (AUG)a new framework for the evaluation of impurity densities based on measurements from charge exchange recombination spectroscopy (CXRS)diagnostics has been developed. The charge exchange impurity concentration analysis code, or CHICA, can perform these calculations for all of the beam-based CXRS diagnostics at AUG and is equipped with the atomic data for all of the regularly measured charge exchange spectral lines (He, Li, B, C, N, O, and Ne). CHICA includes four different methods for the evaluation of the neutral density populations, which feature different implementations and contain varying levels of sophistication. These methods have been thoroughly benchmarked against one another, enabling the important processes for the evaluation of neutral densities to be identified as well as the neutral populations that are most critical to the accurate interpretation of the measured CXRS intensities. For the AUG neutral beams, charge exchange with the ground state of the first energy component is typically the dominant contribution to the measured CXRS intensities, but emission from reactions with the n=2 beam halo population can contribute up to 35% to the total signal and must be included in the analysis. Neglect of this population leads to incorrect magnitudes and incorrect profile shapes of the calculated impurity density profiles. The edge lines of sight (LOS)of the core CXRS diagnostics at AUG intersect the edge pedestal inside of the neutral beam volume. Therefore, the impurity density is not constant along the LOS, complicating the interpretation of the measured intensities. Within CHICA a forward model for the edge impurity densities has been implemented, enabling the reconstruction of accurate edge profiles. Keywords: charge exchange recombination spectroscopy, impurity density, neutral beam attenuation, impurity transport, tokamak (Some figures may appear in colour only in the online journal) 1. Introduction Charge exchange recombination spectroscopy (CXRS)[1,2] is used in many fusion devices to diagnose the impurity ion Plasma Physics and Controlled Fusion Plasma Phys. Control. Fusion 60 (2018)095007 (21pp)https://doi.org/10.1088/1361-6587/aad256 Original content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s)and the title of the work, journal citation and DOI. 0741-3335/18/095007+21$33.00 © 2018 IOP Publishing Ltd Printed in the UK1
temperatures, rotations and densities. The charge exchange reaction is given by hHA HA HA ,1 ZZ Z11 *n++ + + ++ -+ + - () () () where H is, typically, a neutral hydrogen or deuterium atom and A is an ion of elementary charge Z. After the transfer of the electron, the impurity is left in an excited state and subsequently decays to a lower energy level and emits characteristic radiation: the measured CXRS spectrum. In fusion plasmas, the natural population of neutral particles is very low due to the high temperatures and densities. Thus, to make active CXRS measurements, neutral atoms are introduced into the plasma. This is typically done either by means of neutral beam injection (NBI)or neutral gas puffs. The injection of neutrals provides the added benefit of localizing the CXRS measurements to the intersection of the diagnostic line of sight (LOS)and the volume enclosing the injected neutral population. While the impurity ion temperature and rotation along the LOS can be derived directly from the measured spectra via the Doppler broadening and shift of the active spectral line, the impurity density evaluation is more complicated as the measured intensity is not only proportional to the impurity density, but also to the densities of the various energy components of the neutral beam and their excited state populations [3,4]. For neutral beam based charge exchange, there are typically at least three energy populations in the beam: the full, half, and third energy components. These correspond to H + , H 2 +and H 3 +ions in the beam source that are all accelerated to the same energy by high voltage grids. During the neutralization process the H 2 +and H 3 +ions break up, resulting in H neutrals with one half and one third of the full acceleration energy. Since the charge exchange cross-sections are dependent on the collision velocity, all three of these populations contribute differently to the total CXRS emission and must be accounted for separately. In addition, there is a fourth population that must be taken into account, namely the beam halo [2,5–8]. The halo is a cloud of thermal neutral particles that forms in and around the beam volume due to charge exchange reactions between beam neutrals and main ions in the plasma. These new thermal neutrals can then undergo charge exchange reactions with impurity ions leading to additional emission. The importance of the halo to CXRS measurements has been known since the 1980s [2,9]and is discussed in several publications including [2,5,7]. Regardless, it is not routinely included in present day CXRS impurity density calculations. This is possibly due to the additional complexity of modeling these components and possibly due to the high beam energies used in present day fusion devices. As will be shown in section 2.3, the production of halo neutrals decreases strongly with increasing beam energy. Together with the very small charge exchange crosssections at thermal energies, this justifies the neglect of the ground-state halo population and perhaps contributed to the general neglect of halo neutrals in CXRS calculations. However, the charge exchange cross-sections for different excited state populations differ significantly and must be considered separately to correctly determine the total charge exchange emission. In the case of the halo, the n=2[4] population can not a priori be neglected. This was also pointed out by Isler et al [2]and first assessed quantitatively for AUG data in [7]. The measured CXRS intensity is given by equation (2), Lhnsn s vss 4 d, 2 Z nj Znj njZ j CX, LOS 0, , ,, , eff ò åå ln p s = ´á ñ l () () () () () () where hνis the energy of the photon, n Z is the impurity density, n 0,n,j are the neutral density populations, the sums on nand jare over the principle quantum number and the neutral energy components, respectively. v njZ j,, , eff s á ñ l() is the effective CX emission rate with σthe charge exchange crosssection and v j the collision velocity. Here, the angle brackets, <K>, indicate integration over the appropriate velocity distributions: thermal distributions in the case of the halo and thermal and beam velocity in the case of beamimpact charge exchange. λrefers to the specific emission line and sto the position along the diagnostic LOS. Note that the effective emission rates and the neutral densities in the plasma are also dependent on local plasma parameters such as the electron density, electron and ion temperatures, and the effective charge, all of which can vary along a LOS. Hence, these quantities have been kept inside the LOS integral, as has the impurity density itself. It is often assumed that these quantities do not vary along the intersection of the LOS and the neutral population as was done, for example, in [8]. With these assumptions, the impurity density can be removed from the LOS integral and it is straightforward to solve directly for n Z nh L vnss 4 d.3 Z z njnjZ j nj CX, ,, , eff LOS 0, , ò p n l s =ååáñ l () () () With these assumptions the integral nssd nj LOS 0, , ò() contains all of the 3D geometric information of the sightline and neutral beam geometry. This is the conventional way of solving for the impurity density. However, it can be necessary to maintain the LOS dependences of the impurity density and the charge exchange cross-sections, particularly if the LOS samples many flux surfaces inside the neutral volume or if it intersects the neutral volume in a region with steep gradients, for example, the edge pedestal. This will be discussed further in section 5. At ASDEX Upgrade (AUG)a new framework for the evaluation of impurity densities from CXRS measurements has been developed. The charge exchange impurity concentration analysis code, or CHICA, is capable of calculating impurity densities using measurements from all of the CXRS diagnostics in the AUG CXRS suite [10,11]and is equipped with the atomic data for all of the regularly measured CX spectral lines including He, Li, B, C, N, O, and Ne. The atomic data used for the calculations will be discussed further in section 3. Presently, CHICA only evaluates beam based CXRS measurements. CHICA supports four different methods for calculating the neutral beam densities in the plasma: FIDASIM [12], two versions of a ‘pencil’code dubbed 2 Plasma Phys. Control. Fusion 60 (2018)095007 R M McDermott et al
‘FAST’and ‘COLRAD,’and indirect measurements of the neutral populations from beam emission spectroscopy (BES) [7,13,14]. The details of these methods and their comparative pros and cons are presented in section 2. In section 3the effective emission rates used for the CXRS impurity density evaluation are presented and discussed. In addition, the contributions from the different energy components and excited states to the total measured CXRS intensity are examined. In section 4these codes are applied to AUG experimental data and the resultant impurity density profiles obtained via the different methods are compared. Section 5discusses the evaluation of CXRS impurity densities when they vary along the LOS of the diagnostic and, therefore, can not be removed from the LOS integral. Finally, this work is summarized in section 6. 2. Neutral density evaluation 2.1. Neutral beam geometries A top-down view and poloidal cross-section of the AUG tokamak are shown in figure 1. Here one can see the trajectories of the AUG neutral beam sources as well as the LOS of the BES and CXRS diagnostics. The LOS of the core toroidal CXRS and BES systems are embedded in the same optical heads, providing measurements along nearly parallel paths through the neutral beams, and are marked simply as ‘Core tor.’in figure 1. There are two separate core toroidal CXRS/ BES systems, one on each of the neutral beam boxes. For clarity, only a subset of the LOS from each system are shown. On NBI box I, there are 91 LOS imaging the plasma from the magnetic axis to the pedestal top on the low field side (LFS), separated into three vertically stacked rows of 30, 31, and 30 fibers. The vertical separation is roughly 1.5 cm. These LOS are divided between three charge exchange diagnostics, typically measuring He, B, and N, a fast ion D-alpha (FIDA) system, and a BES diagnostic. On NBI box II, there are 70 LOS imaging the plasma from the pedestal top on the LFS to the pedestal top on the high field side (HFS). These LOS are divided between two CXRS diagnostics, typically measuring either B or N, and a second BES system. As can be seen from figure 1the AUG NBI boxes [15] are separated toroidally by 180°. Each box is comprised of four neutral beam sources capable of injecting 2.5 MW of power in D operation for a total of 20 MW. NBI Box 1 uses arc plasmas sources and has a maximum injection energy of 60 keV. NBI Box 2 uses radio frequency sources and has a maximum injection energy of 93 keV in D. The extraction voltages of these beams are not fixed. On a discharge-to- discharge basis, the energy of NBI Box 1 can vary between 30 and 60 keV (0.5–2.5 MW)and the energy of box 2 from 50 to 93 keV (0.72–2.5 MW). When reducing the voltage of the beams, also the beam current is reduced to keep the beams operating at the optimum points of their respective perveance curves. As such, at reduced voltage the CXRS signal suffers from both a reduction in the CX cross-sections and a reduction in the overall number of neutral particles injected, resulting very quickly in a significant reduction in the CXRS signal. As a result, standard CXRS analysis at AUG focuses on discharges with full-voltage beams, although CHICA is capable of handing reduced voltage beams, provided sufficient CXRS signal is present. CHICA is also capable of Figure 1. Toroidal and poloidal cross-sections of the ASDEX Upgrade tokamak showing the trajectories of all of the neutral beams from NBI box 1 and NBI box 2 and the LOS of the beam-based CXRS and BES diagnostics. 3 Plasma Phys. Control. Fusion 60 (2018)095007 R M McDermott et al
handling the impurity density analysis when the beams are operated in hydrogen. The extracted beam power is very well known, as are the losses in the acceleration grids and in the neutralization chamber. In addition, the fraction of particles in the different energy components is well diagnosed. The biggest uncertainty on the power reaching the plasma comes from the losses in the duct, which are not measured and are assumed to result in a 15% loss. While this assumption results in injected powers that agree very well with BES measurements of the beam energy components in the plasma, the error bars on the BES data are also of order 10% and the BES data is not analyzed routinely. There are pressure gauges in the duct itself, which show a relatively large variation in pressure (factor of two), which indicates that the flat 15% assumption might not be appropriate for all plasma discharges. As such, NBI losses due to variable duct pressure could well contribute to the 10%–20% variations observed between the measured CXRS intensities on NBI box 1 and NBI box 2. This could be clarified through the analysis of the BES data for a large number of discharges. Such an analysis task, however, is beyond the scope of this work. Not only are the injection energies and particle distributions of the two boxes different, but the beam geometries are as well. In addition, the geometries of sources 6 and 7 (NBI box 2)are variable and, for a given plasma discharge, can be in any position between the extrema of the range. The CXRS and BES diagnostics on NBI Box 1, while focused on source 3, also collect significant emission from reactions with sources 1, 2, and 4, see figure 1. The diagnostics on NBI Box 2[11]are focused on source 8, but also see emission from sources 5 and 7 at the LFS plasma edge. Thus, all of these beam sources need to be sufficiently well characterized to reliably produce impurity densities from the full suite of CXRS diagnostics. NBI source 6 is situated significantly below the LOS of the diagnostics, see figure 1, and does not contribute to the CXRS signal. It too, however, has been characterized and included in the codes. The geometries of the NBI sources (trajectories and divergences)have been determined by a combination of BES measurements using both toroidal and poloidal LOS, thermal images of the beam impact positions on the inner wall, and direct measurements of the source locations. The impact positions of the beams on the inner wall of the machine are known to an accuracy of ∼1 cm, which translates into an uncertainty on the horizontal and vertical angles of the beam of ∼0.05°. The full width half maximum (FWHM)of the beams are on the order of 20 cm and the radial and vertical dimensions of the LOS integration volumes are approximately 1 cm, making a 1 cm uncertainty in the overall positions of the beam centers acceptable. The uncertainties on the beam widths are more difficult to characterize. Perhaps the best measure we have of how well the widths are characterized are from CXRS impurity density profiles on different beam boxes and using different beam configurations, particularly using beams that are centered well below the diagnostic LOS, such that only the wings of the neutral distributions contribute to the signal. Comparisons of impurity density profiles from NBI 2 and 3 (box 1)or NBI 5 and 8 (box 2), for example, produce the same impurity density profiles within the error bars of the measurements. As previously mentioned, CHICA supports four different methods for calculating the neutral beam densities in the plasma, three different beam attenuation codes as well as indirect measurements via BES. All three beam attenuation codes (FIDASIM, FAST, and COLRAD)use identical neutral beam geometries, although the implementation between FIDASIM and the pencil codes is different. FIDASIM uses set beam divergences and foci of the particle distributions accelerated by the grids to describe the downstream beam geometry. FAST and COLRAD use the product of two Gaussians to describe the fall-off of the neutral densities perpendicular to the beam direction. These Gaussians are characterized by vertical and horizontal widths, a u,n,j and a v,n,j , that increase as a function of distance along the beam path to account for the beam divergence. These widths were obtained via fits to the same beam model that is included in FIDASIM. The beam widths and divergences are very similar for the different energy components and excited states in the beam, but not identical. Therefore, the differences as a function of beam energy, j, and excited state, n, are explicitly maintained. The fit function is given in equations (4)–(6) Guvl alal u al v al ,, 1 exp exp , 4 nj unj vnj unj vnj , ,, ,, 2 ,, 2 2 ,, 2 = ´-- ⎡ ⎣ ⎢⎤ ⎦ ⎥⎡ ⎣ ⎢⎤ ⎦ ⎥ () () () () () () ala a la l,5 unj unj unj unj,, 0,,, 1,,, 2,,, 2 =+´+´() ( ) ala a la l.6 vnj vnj vnj vnj,, 0,,, 1,,, 2,,, 2 =+´+´() ( ) Here, uand vare orthogonal directions, perpendicular to the direction of the neutral beam injection, l. This function is with respect to a fixed starting point at the edge of the plasma outside of which the geometry is not valid. For all three codes, the neutral densities are calculated on the same three dimensional spatial grid. Care has been taken to ensure that the grid for each beam box encompasses the entire beam-LOS intersection volume and is dense enough to describe the beam geometry well. For the analysis of the core impurity densities the grids extend typically 80 cm in the direction perpendicular to the beams and have a 1.5 cm resolution. A smaller, finer spatial grid is used for the edge diagnostics. Figure 2provides a comparison of the beam geometries implemented within the pencil codes (left)and FIDASIM (right)for NBI 3 (top)and NBI 8 (bottom). Here, the neutral densities in the ground state, first energy component have been used for the comparison. The beams are shown in a topdown view and the densities were summed in the vertical direction. Note that the scale differs between the NBI 3 and the NBI 8 plots. There is almost a factor of two difference between the densities of this population (n=1, first energy) at the plasma edge. The x- and y-axes correspond to the outer boundaries of the grids used for these calculations and the 4 Plasma Phys. Control. Fusion 60 (2018)095007 R M McDermott et al
directions of xand yare the same as in figure 1. The LOS of the core toroidal systems are also shown as red lines terminating in filled circles at the center of the beams. The grids are sufficient to encompass the entire intersection volume of the LOS with the neutral distribution, also in the zdirection (not shown). One can see that the geometries agree very well between the two methods. This is also the case for the other six neutral beams at AUG. The slight differences between the two methods (left and right columns)are attributed to the fact that FIDASIM takes into account plasma rotation as well as the shape of the magnetic flux surfaces. This leads to slightly asymmetric attenuation between the top and bottom and left- and right-hand sides of the beam. In addition, one can see larger discrepancies between the two codes in the plasma scrape off layer, i.e. outside of the solid red line. This is because the geometry used in the pencil models (FAST/ COLRAD)is not applicable in this region. The valid comparison region starts just outside of the last closed flux surface (LCFS)(solid red line)on the low-field side of the machine. It is important to remember that the integral along the LOS as written in equation (2)gives the effective measurement position of the diagnostic. This is often approximated as the intersection of the LOS with the center of the beam. This is appropriate for LOS that cross only a single NBI source, are perpendicular to the beam injection, and do not cross-regions with steep gradients, such as the edge pedestal. In practice at AUG, these conditions are only very rarely fulfilled. Therefore, the default measurement positions are calculated as the center of mass of the distribution given by the integrand in equation (2) [11]. This is done using the equilibrium and plasma profiles from a standard H-mode discharge and for all possible combinations of the NBI sources. Even so, when running CHICA the calculation is re-done with the actual plasma equilibrium and kinetic profiles. This can result in shifts of the effective measurement positions by up to 1 cm in the core and even more at the edge, depending on the position of the LCFS compared to the value used in the default evaluation. 2.2. Neutral beam attenuation methods The beam attenuation codes implemented within CHICA vary significantly in terms of their capabilities, the physics Figure 2. Comparison of neutral beam geometries as implemented within the pencil codes (left)and FIDASIM (right)for NBI 3 (top)and NBI 8 (bottom). Here the neutral densities in the ground state, first energy component have been used for the comparison. These were summed in the vertical direction. The color bar indicates the absolute magnitude of the neutral densities. The red lines and circles depict a subset of the LOS of the CXRS diagnostics. 5 Plasma Phys. Control. Fusion 60 (2018)095007 R M McDermott et al
included, the beam model used, and the atomic data. To provide a reference for the discussions that follow, an overview of the main features of each code is provided in table 1. FIDASIM [12,16]is a Monte Carlo (MC)code capable of simulating the entire D α spectrum. As part of these calculations, FIDASIM calculates the neutral densities of all beam energy components including the halo in the first four excited states. Within CHICA, this information can be extracted and used for the determination of the CXRS impurity density profiles. The results of the FIDASIM simulations have been compared to the measured beam emission spectra, including halo, and reproduce the experimental spectra well [17]. Due to the MC approach, the analysis of the neutral density via this method requires relatively long computation times; the simulation of an entire discharge can take several days. Using FIDASIM for the neutral density calculation also has the disadvantage that the noise level on the resultant neutral densities is inversely proportional to the number of Monte Carlo particles used. This means that to get a smooth neutral density profiles on the HFS requires larger number of particles and longer computation times. However, FIDASIM is the only method within CHICA that correctly handles the plasma edge, taking regularly into account the plasma curvature and also neutral particle losses out of the plasma. In addition, FIDASIM is the only model that fully treats the movement of halo neutrals. The FAST code originated as a very simple, hands-on, attenuation model for the assessment of the CXRS impurity densities. Through comparisons of FAST with COLRAD and FIDASIM, it was possible to identify errors in all three codes as well as the physics that was important for the evaluation of the impurity densities. During this process, the physics included in FAST was evolved until it became possible to run COLRAD on similarly fast time scales within the CHICA framework and all of the remaining differences between the codes could be attributed to specific pieces of physics missing in FAST. At this point, COLRAD became the standard analysis tool and there was no need for further evolution of the FAST code. However, it remains as a reduced model within CHICA that is still useful as a benchmark and for the understanding of the behavior of its more complicated counter-parts. The FAST method calculates the neutral densities of the jth energy component in the ground state (equation (8)) and first excited state (equation (9)) along the center of each neutral beam according to equations (7)through (9). Al Al dl S ln l t1,7 jj Bj j,e =- - D() ( )[ () () ] ( ) nln AlEl01 , 8 nj nj j j0, 1, 0, 1, =- == () ( ) ()( ()) ( ) nln AlEl0. 9 nj nj j j0, 2, 0, 2, = == () ( ) () () ( ) Here, A l j()is the beam attenuation factor with A 01 j=() and lbeing the position along the beam path. S B,j are the total beam stopping rate coefficients in m s 31- ,n e is the electron density and Δt j =v j /dl, where v j is the beam velocity per energy component and dl is the distance of one iteration step along the beam path. The beam stopping cross-sections are taken from ADAS [18]using an implementation [19]that takes into account the temperature, density, beam energy, and individual impurity density profiles. In equations (8)and (9) E j is the fraction of particles in the first excited state from a simplified model considering only ion impact excitation. The normalization factor, n 0,n=1,j , is the starting neutral density in the ground state at the center of the beam outside of the plasma, i.e. before attenuation. The FAST code also includes a simplified halo estimate in which the source of the halo, CX reactions between beam neutrals and plasma main ions (righthand side of equation (10)), is balanced against ionization losses nln lSl vn n l.10 n j jj n j e0,1,Haloe DD, 0, 1, D ås= = = () () () () ( ) Here, n 0,n=1,Halo is the halo density in the ground state, S e is the ionization loss term, σ DD is the cross-section for deuterium charge exchange with beam neutrals, j vis the velocity of the neutral atoms, n 0,n=1 is the neutral density in the ground state, the subscript lrefers to the position along the NBI trajectory, jto the neutral beam energy component, and n e and Table 1. Overview of the three beam attenuation codes implemented within CHICA. The entry ‘MC’is short for ‘Monte Carlo’and ‘full col. rad.’indicates a full collisional radiative model treatment including electron, ion, and impurity ion impact excitation and charge exchange. Codes: FAST COLRAD FIDA Type: Local at beam center Local at beam center MC, 3D geometry NBI Geom.: Gaussian (equation (4)) Gaussian (equation (4)) Fixed foci and divergences for ‘launched’MC particles Halo Geom.: Gaussian (equation (4)) Ionization versus CX fall-off lengths MC treatment Comp. time: ∼5s/time point ∼5s/time point ∼1–5 min/time point Curvature: Yes (not standard)Yes (not standard)Yes Rotation: No No Yes n-states n=1, 2 n=1–10 n=1–10 Beam excit. : Ion-impact equations (7)–(9)full col. rad. full col. rad. Halo source: σ DD v NBI , not Maxwell averaged v DD NBI sáñ v DD NBI sáñ Halo excit.: elec. impact full col. rad. full col. rad. Z eff profile: Array of impurities Single impurity Single impurity 6 Plasma Phys. Control. Fusion 60 (2018)095007 R M McDermott et al
n D are the electron and main ion densities, respectively. For the ionization, only electron impact ionization as a function of electron temperature and density along the beam is taken into account. The atomic data used is taken from the ADAS ADF11 scd96_h.dat file. For the main ion charge exchange cross-sections only the energy dependence has been taken into account as the variations with density and temperature in the parameter space of interest are comparatively small. A proper analysis would integrate the product of these crosssections and beam velocities over a Maxwellian velocity distribution. Within the FAST code this has not been done. However, this is properly treated in both FIDASIM and COLRAD. As will be shown in section 3, the emission from charge exchange reactions between impurity ions and ground state halo atoms to the total CXRS signal is negligible. However, the contribution from the n=2 halo atoms is important. Within the FAST model, the fraction of halo atoms in the first excited state is estimated as nln l l n lAPEC , 11 nne 0,1,Halo 0,2,Halo 21 = == () () () () ( ) where n 0,n=1,Halo is the ground state halo population, n 0,n=2,Halo is the population in the n=2, PEC is the photon emissivity coefficient as a function of electron temperature and density along the beam taken from the ADAS ADF15 pec96#h_pju#h0.dat file, and A 21 is the hydrogen Einstein 21 rate coefficient. Despite its simplicity, the neutral densities from this model compare well against those from the more complete physics models included in FIDASIM and COLRAD and deviations from this model highlight where additional physics effects become important. All of the calculations described above are performed for the parameters at the center of the neutral beam and are then extended onto the three dimensional spatial grid using the function described in equation (4). The vertical and horizontal widths of the halo populations depend strongly on local plasma parameters. To describe this population via a Gaussian decay, widths as a function of plasma parameters must first be defined. For this purpose, FIDASIM was run for a matrix of ion and electron temperatures and electron densities. The resultant halo populations were fit to the same Gaussian function given in equation (4)and these fitted widths are used as a ‘look-up’table when running FAST. It should be noted, however, that the single Gaussian fall-off length is often a poor description of the halo distribution and FAST, therefore, significantly underestimates the spatial extent of the halo populations. The impact of this geometry error on the calculated CXRS densities depends strongly on local plasma parameters and is somewhat mitigated by the LOS integral through the neutral distribution. This is discussed further in sections 2.3 and 4. A simplified toroidal curvature model is implemented for the FAST code. However, the impact on the neutral densities when using it is very small. As this module increases the computation time of the FAST code it is not routinely included in the calculations. The effect of both toroidal and poloidal curvature on the neutral density distribution have also been tested with COLRAD and found to be negligible. This is consistent with the good agreement in edge neutral densities between the pencil codes and FIDASIM, as will be shown in section 2.3. The COLRAD model, short for the collisional radiative model [20], combines the simplified beam model of the FAST code discussed above, avoiding the need for a time consuming Monte Carlo approach, and the same complete collisional radiative model used within FIDASIM. COLRAD solves the collisional radiative model for the beam neutrals of each energy component and the halo neutrals. It calculates the densities in the states with principal quantum number n=1–10 by solving the corresponding system of 10 coupled rate equations. The collisional rate coefficients in COLRAD are due to excitation, de-excitation and ionization by electron, hydrogen and impurity ion impact as well as charge exchange reactions with main ions (HorD)and impurities. The evolution along the beam trajectory is calculated on a spatial grid with 1cm step width, where all beam neutrals start in n=1at the outermost grid point. The rates in the model vary in time as the beam penetrates regions with different n e ,T e ,T i and Z eff . The losses due to charge exchange of all energy components of the beam with the hydrogen background are the source for the n-levels of the halo neutrals. The population of the halo neutrals in all n-states is then calculated from the equilibrium solution of the collisional radiative model, where the corresponding rates are evaluated with the local plasma parameters at the source location on the beam axis. Similar to FAST, within COLRAD all calculations are performed for the plasma parameters at the center of the beam and are then extended onto the 3D grid using equation (4). The shape and extent of the halo population, however, is treated more carefully. Within COLRAD characteristic falloff lengths for ionization and charge exchange reactions of the halo population are calculated for each position along the beam path. The relative strengths of these two processes is used to estimate the profile of the halo neutral cloud perpendicular to the beam trajectory. Therefore, within COLRAD the halo population is not characterized by equation (4). This ingredient was found to be critical to reproducing the radial profile of the halo as measured by BES and calculated by FIDASIM. Not taking into account the spreading of the halo cloud leads to significant differences in the shape of the calculated impurity density profiles, see section 4. The final possibility to determine the neutral density populations in the plasma is direct BES measurements along lines of sight (LOS)parallel to those used for CXRS [7,13,14]. The D α spectra provide measurements of the total emission from the beam neutrals in the n=3 for all energy components including the halo. By using effective photon emission coefficients, this can be translated into n=3 neutral particle densities, which in conjunction with a collisional radiative model, enables the populations in other excited states (n=1, 2)to be determined. The end result is the LOS integrated neutral densities as a function of beam energy and excited state. While this direct measurement of the neutral densities is the ideal method to determine the neutral densities in the 7 Plasma Phys. Control. Fusion 60 (2018)095007 R M McDermott et al
plasma, there are several limitations to using this technique. First, the NBI sources in the NBI beam boxes are not completely separated vertically. Therefore, when multiple NBI sources are used, the BES LOS collect emission from all of them, blurring the spectra and making it difficult to unambiguously identify the emissions from the different beams and energy components. This can lead to significant errors in the deduced neutral densities. This problem is worse for NBI box I than for NBI box II. Second, if the LOS are too perpendicular to the beam, as is ideal for CXRS measurements, then there is insufficient Doppler-shift separation between the energy components for them to be clearly identified. This is the case for the edge LFS LOS of both BES systems and can also be a general problem for the third energy component. Fortunately, the third energy component is relatively unimportant for the analysis of the impurity density profiles. Third, the evaluation of the BES data itself is a complicated and time consuming process. Therefore, the BES is mainly used as a check on the attenuation codes and to evaluate the impurity densities in situations, such as beam blips (short injections of NBI power of 20 ms or less), in which the beam shapes are not well described by equation (4)and the number of neutral particles in each energy component is not well characterized. Similar analysis is needed, for example, at DIII-D [21], where the behavior of the beam in the start-up phase is also found to evolve, and not be well described by the steady-state parameters. 2.3. Comparison of different methods The beam attenuation models included within CHICA have been thoroughly benchmarked against one another to understand the limitations of a given code and to identify the dominant physics mechanisms. To do this, CHICA was run with all three methods using a matrix of flat kinetic profiles (n e ,T e ,T i )as inputs. The electron density was varied between 1 and 9×10 19 m −3 , and the electron temperature and the ion temperature between 250 eV and 5 keV. In all cases a linear pedestal assumption was used starting at ρ f =0.94 (ρ f is the normalized toroidal flux coordinate)and enforcing T e =T i =100 eV and n e =1×10 18 m −3 at the LCFS. This is necessary within the codes to ensure reasonable temperature and density values are extrapolated onto grid points outside of the LCFS. For the calculations the equilibrium from AUG discharge 33910 was used. In figure 3the resultant attenuations of all three beam energy components in the ground state for both NBI 3 and NBI 8 are shown. These curves are for the case T e =T i =500 eV and n e =5×10 19 m −3 , which roughly corresponds to pedestal top parameters at AUG. Here, the neutral density distributions have been integrated along the LOS of the CXRS systems viewing the respective beams and are plotted against the major radius of the intersection position at the center of the beam. The agreement between the codes is extremely good, making it difficult to distinguish the three curves per energy component. Differences are visible for the third energy component, but are small enough to be of no importance for the evaluation of the impurity densities, as will be shown in section 3. The attenuation of these components calculated by all three codes agree well across the entire matrix of densities and temperatures considered. In addition to the ground state neutral densities, each code is capable of calculating the n=2 neutral densities. FIDASIM and COLRAD can also provide higher n-state information. These, however, are not used in the analysis. Section 3will present a brief discussion of the effect of neglecting these populations on the total calculated CXRS impurity density. In figure 4the calculated neutral densities in the n=2 for all three beam energy components are shown for NBI 3 at mid-radius (R=1.9 m). In the left panel the n=2 densities are shown as a function of the electron density and on the right as a function of the electron temperature. Figure 3. Comparison of the ground state neutral densities of all beam energy components for NBI 3 and NBI 8 calculated via all three beam attenuation codes with CHICA. These attenuations correspond to a flat electron density profile of 5×10 19 m −3 and flat ion and electron temperature profiles of 500 eV. 8 Plasma Phys. Control. Fusion 60 (2018)095007 R M McDermott et al
The dependence on electron density in all three codes is very similar, although the increase in n=2 population calculated by FAST is not as steep as in the other two. This discrepancy decreases as the electron temperature is increased. The excitation and de-excitation model of the beam neutrals within FAST does not include a dependence on temperature. For excitation of beam neutrals into the n=2, FAST only considers ion impact excitation, which depends on the beam energy and electron density. Ionization out of the n=2 is not considered, only de-excitation back to the ground state given by the relevant Einstein coefficient, A 21 . The slight decrease in n=2 population at low T e reported by FAST on the righthand side of figure 4is in fact a carryover from the ground state densities, which also show this slight decrease with electron temperature for all three codes. The other codes show a larger n=2 population at low T e . This indicates that FAST is missing an atomic process that causes increased population of the n=2 at low electron temperature. The missing process is electron impact excitation, which is included in both FIDASIM and COLRAD. The population of the higher nstates included in COLRAD and FIDASIM, but not in FAST may also play a role. The last neutral density population of interest is the halo. Figure 5shows the ground state halo densities computed by the three codes integrated along a LOS near mid-radius as a function of electron density, electron temperature, and ion temperature. Here the halo densities are normalized by the sum of the ground state beam energy components (j=1–3), which represents the source of the halo particles. Note that the density of the halo population tends to be larger than the densities of the ground state beam energy components, which are the source of the halo. This is because of the long lifetime Figure 4. Neutral densities in the n=2 calculated by all three attenuation codes within CHICA for NBI 3 around mid-radius as a function of (left)electron density and (right)electron temperature. Figure 5. Ground state halo densities computed by the three beam attenuation codes as a function of electron density (left), electron temperature (middle), and ion temperature (right). These densities are normalized by the sum of the ground state densities of the first, second, and third beam energy components and have been integrated along a diagnostic LOS near mid-radius. 9 Plasma Phys. Control. Fusion 60 (2018)095007 R M McDermott et al
than 10% across the entire profile, while for NBI 8 this component is more important contributing over 20% near the plasma edge. The halo contribution on NBI 3 (COLRAD and FIDASIM)is only 10% at the plasma edge, but increases to 35% of the total signal in the plasma core. A similar trend is seen on NBI 8, but in this case the core n=2 halo contribution does not exceed 20% of the total signal. The corresponding boron impurity density profiles are shown in the top panel of figure 14. Here, for clarity, only the profiles from COLRAD and FIDASIM have been shown as well as the BES profile from NBI 8, which featured a different halo density profile and causes the impurity density profile to be more peaked. The absolute intensity calibration from the NBI 3 CXRS system was adjusted upwards by 20% to get the match to the NBI 8 data. This adjustment is not unreasonable considering that the absolute intensity calibration on each system has an uncertainty on the order of 10% and that the invessel optical components of the different systems are exposed to different degrees of coating and damage during plasma operation, increasing the uncertainty on the absolute and relative calibrations of the systems during an experimental campaign. In general, the uncertainty on the injected neutral beam power, as discussed in section 2.1, could also play a role in the observed intensity differences. In this case, however, the good agreement between the neutral beam attenuation codes and the BES systems, suggests that this is not the source of the mismatch. The agreement in profile shape resulting from the two systems and the different methods is very good. This can be seen in the bottom panel of figure 14 where the normalized logarithmic gradients ( R LRnn nBB B=- )are shown. Each boron density profile from COLRAD and FIDASIM in the 400 ms time window was fit with a spline and the corresponding R LnBdetermined. The profiles shown here are the mean of the R LnBprofiles and the error bars are the standard deviation. Therefore, these R LnBprofiles do not correspond directly to the averaged profiles shown in the top panel. The R LnBprofiles corresponding to the averaged profiles do fall well within the error bars shown. While small differences between the impurity density profiles calculated from the different CXRS systems (NBI 3 system versus NBI 8 system) and the different codes remain, it is difficult to pinpoint the source of these differences and highlights the challenge of determining the absolute value of R LnZto better than 0.5 with even very high quality CXRS data. This is an important consideration for low-Zimpurity transport experiments and comparisons to modelings. The red triangles correspond to the profile derived using the BES data on NBI 8. Inside of mid-radius, where the agreement in the halo profile is still good, the BES data overlaps with the other curves. Outside of mid-radius, where the halo-profile starts to diverge, there is a change to the impurity density profile shape outside of the error bars. However, as the halo becomes less important in this region (∼10%), the effect on the profile shape is relatively small. Figure 13. Ground state neutral densities for the first and second energy components of NBI 3 and 8 as well as the n=2 halo population multiplied by their respective effective emission rates from discharge #33856. 16 Plasma Phys. Control. Fusion 60 (2018)095007 R M McDermott et al
The top panel of figure 15 shows the boron density profiles calculated using neutral densities from FIDASIM for both NBI 3 and NBI 8 with (blue full symbols)and without (red hollow symbols)including the halo contributions. The blue profiles here are identical to the FIDASIM profiles shown in figure 14. Neglecting the halo results in impurity densities that are too high and an artificial peaking of the profile, as was first shown for AUG in [7]. The bottom panel of figure 15 shows the corresponding R/Ln B profiles. In this case, neglecting the halo causes an artificial increase of order ΔR/Ln B ∼0.6 across the entire profile for NBI 8 and 1.0<ΔR/Ln B <1.5 for NBI 3. At AUG the halo is observed to have a very large impact on the evaluation of impurity density profiles. However, this does not a priori mean that this impact will translate directly to the profiles evaluated at other devices where the halo is not included in the analysis. The magnitude of the halo depends strongly on the NBI energy and is larger for lower energy beams. Moreover, it increases with electron density and decreases with Z eff . At devices with higher energy beams, higher fractions of particles in the first energy component, lower electron densities, or higher Z eff , the contribution of the n=2 halo to the total CXRS signal may well be smaller. Dedicated calculations at each machine are required for a proper assessment. 5. Variations along CXRS LOS The simplest way to evaluate the LOS integrals in equation (2)is to assume that all plasma parameters, impurity density included, are constant along the LOS through the neutral beam volume and can, therefore, be taken out of the integral. In the presence of strong gradients or LOS geometries that are not near tangency to magnetic flux surfaces, however, this assumption is not necessarily valid. This is the case for the edge LOS of the core CXRS diagnostics on both NBI 3 and NBI 8. These LOS intersect the edge pedestal while still in the neutral beam volume, as can be seen in figure 2. These LOS collect significantly less emission than the innermost LOS of the edge diagnostics, which are Figure 14. (top)CXRS boron impurity density profiles from NBI 3 and NBI calculated using COLRAD and FIDASIM in discharge # 33856. The profile from BES on NBI 8 is also shown in red. (bottom)Normalized logarithmic gradients of the boron impurity density profiles shown in the top panel. Figure 15. (Top)Boron density profiles for discharge #33856 calculated from the CXRS intensity measurements on NBI 3 (circles) and NBI 8 (squares)including all neutral populations (blue closed symbols)and excluding the halo (red open symbols). The neutral densities were calculated with FIDASIM. These profiles correspond to averages over 400 ms intervals and the error bars show the standard deviation. (Bottom)Corresponding normalized boron density gradients. 17 Plasma Phys. Control. Fusion 60 (2018)095007 R M McDermott et al
optimized to be perpendicular to the beam and tangent to magnetic flux surfaces. This leads to an apparent discrepancy in the measured intensities between the core and edge CXRS diagnostics. This can be see in figure 16. In this discharge the plasma position was swept slowly across the LOS of the diagnostics by several cm allowing an in situ cross-calibration of the intensity calibrations of neighboring CXRS channels, as the intensity measured on each LOS must produce the same profile as the plasma is swept across it. In figure 16 the measured boron IV intensity profiles from the core NBI 3 and NBI 8 systems as well as the edge system on NBI 3 are shown for three different plasma positions, corresponding to three different time-points in the discharge. On the left-hand side the outermost radius of the plasma was 2.16 m. Here, the LOS from the edge diagnostic are inside of the pedestal top. In the middle panel, the outer radius is 2.14 m. Here, the edge diagnostic characterizes the boron pedestal very well and a clear decrease of the intensities on the core LOS inside of this pedestal location can be seen. On the right-hand side, the maximum plasma radius was 2.13 m, and this discrepancy is even more pronounced. The time windows shown in figure 16 were specifically chosen to exclude large type I ELMs, as these events add significant scatter to the data. As the plasma was moved away from the wall, the overall impurity content in the plasma decreased. This can be seen by comparing the intensities measured at a fixed position in the three panels of figure 16. The measured profile shape, however, remains unchanged. Therefore, for the comparison of the impurity density profiles below, a single, smooth, time dependent correction factor was applied to all of the data from all three diagnostics such that the average density measured between ρ ψ =0.87 and ρ ψ =0.93 remains constant at the value from the outermost point of the radial sweep (left panel in figure 16). In addition, an overall 10% correction factor was applied to the intensity data of the core system on NBI 8 (upward)and the edge system on NBI 3 (downward). The intensities measured on the core NBI 3 system have not been adjusted. These corrections were applied to get the match in the absolute value of the impurity densities shown in figure 18. Potential sources for these differences were already discussed in the previous section. To evaluate the impurity density profiles for the edge LOS of the core diagnostics, one must assume a functional form for the edge impurity density profile. In CHICA, the default option for this form is: n C C C if 0.9 1 10.1 if 1.0 1.1 if 1.0 1.1, 12 B ped.top ped.top ped.top r rr r rrr rr = < ´- -+<< ´- + < < y y y y yy ⎧ ⎨ ⎪ ⎪ ⎩ ⎪ ⎪ ⎛ ⎝ ⎜ ⎜ ⎞ ⎠ ⎟ ⎟ () () () where, ρ ped.top is the normalized poloidal flux at the pedestal top, which can be set by the user before the run based on measured electron density profiles or measured edge impurity intensity profiles, and Cis a constant that can be taken out of the integral. This is the standard form used. However, it is straightforward to implement other functional forms. For the evaluation of the impurity density from the LOS that intersect the edge pedestal (or other steep gradient region) it is insufficient to model the LOS as a single line-trajectory through the plasma. Rather, the full geometry of the lightcone, defined by the diameter of the lens in the optical head [11], the diameter of the focal spot, and the distance between them, is needed. The finite volume of the light-cone means that at each position along the LOS, light is collected from regions of both smaller and larger radii than the central position. In CHICA the light-cone of each LOS is split into four quadrants that are aligned with the major radius of the machine such that the geometric centers of the quadrants on the low- and high-radius sides of the cone have the maximum difference in major radius. The emission collected in each quadrant at each position dsalong the LOS is given equal weight. A sketch of the light-cone geometry used in CHICA Figure 16. CXRS intensities measured by the core NBI 3 and NBI 8 diagnostics as well as the edge NBI 3 system for three different plasma positions in discharge #31916. Here, a 10% upward correction factor was applied to the NBI 8 intensities and a 10% downward correction to the edge NBI 3 intensities. No correction was applied to the data from the NBI 3 core system. 18 Plasma Phys. Control. Fusion 60 (2018)095007 R M McDermott et al
is shown in figure 17. Division of the light-cone into even smaller pieces was also tested, but found to make almost no difference (<2%)to the LOS integrals compared to the four quadrant method. By including the light-cone geometry and the non-con- stant impurity density profile along the LOS, it is possible to interpret the measured intensity profiles. Figure 18 shows the boron density profiles corresponding to the intensity profiles shown in figure 16, but including data from the entire plasma position sweep. Time frames encompassing large ELMs have been excluded. On the left-hand side, the boron density was assumed to be constant along the LOS of the diagnostics and the light-cone geometry was not included. Here, one can see the three diagnostics give three different gradients for the edge impurity density profiles between ρ ψ =0.9 and 1.0. The LOS of the core system on NBI 3 are more strongly affected than those of the NBI 8 system because they are less tangent to flux surfaces and less perpendicular to the NBI. The effect on the edge system of NBI 3 is negligible because of the very small light cones (2–3 mm spot-sizes)and because the LOS are significantly more tangent to the magnetic flux surfaces and also more perpendicular to the NBI. For the evaluation of the CXRS densities from the edge CXRS diagnostics, therefore, these effects are not routinely included and are not included in the evaluation shown in figure 18. On the right-hand side of figure 18 the evaluation of the impurity densities taking these effects into account (ρ ped.top =0.96)is shown. Here, the densities measured along the edge LOS of the core diagnostics have been corrected upwards by the model resulting in flat impurity density profiles out to ρ ψ =0.95. Outside of this location, the core LOS do a much better job of reproducing the pedestal profile measured by the edge system. The agreement is not perfect as the effective measurement location is in the pedestal region and the functional form assumed is not a perfect model. In this region, the correction applied is extremely sensitive to the details of the assumed profile shape and further iteration and analysis would be required to fully correct the data from the core systems. In particular, although large ELMs have been removed from this dataset, small ELMs remain and cause a time dependent evolution of the edge impurity density profile shape, which would have to be included. More careful modeling of the light-cone geometry and the relative alignment of the pedestal profiles would likely also be needed. However, when the effective measurement position of the LOS is inside the top of the impurity density pedestal, the corrections introduced by the simple model reproduce the correct profile shape well. 6. Summary At AUG a new framework for the evaluation of impurity densities measured by CXRS diagnostics has been developed. CHICA can assess the impurity densities using data from all of the beam-based CXRS diagnostics at AUG and includes four different methods for the evaluation of the beam neutral density populations. These methods feature different implementation and contain varying levels of sophistication. While FIDASIM represents the most comprehensive physics package included in the code, the long computation times associated with the Monte Carlo calculations make running FIDASIM as a routine tool for impurity densities impractical. Overall, the reduced physics models of the FAST and COLRAD, COLRAD in particular, perform well against FIDASIM, eliminating the need for extensive runs with FIDASIM. These codes also fare well compared to indirect measurements of the neutral density populations from BES. While in theory, BES is the ideal method for determining the neutral density integrals, the practical implementation is often challenging. At AUG, due to very small angles between the diagnostic LOS and the neutral beams, the BES evaluation at the plasma edge is not possible. In addition, when multiple NBI sources are in use the interpretation of the BES spectra becomes complicated and greatly increases the uncertainty of the analysis. Lastly, the evaluation of the BES data itself is a complicated and time consuming process and is not routinely done. Therefore, the BES is mainly used as a check on the attenuation codes when such checks are possible and to evaluate the impurity densities in situations, such as beam blips, in which the beam shapes and particle fractions are not well characterized. CHICA is equipped with the atomic data for all of the regularly measured charge exchange spectral lines (He, Li, B, C, N, O, and Ne). This data has been taken for the most part from ADAS cross-section data and processed to produce a consistent set of effective emission rates for both beam impact and thermal charge exchange with D(H)in the n=1 and n=2. Possible contributions from n=3 populations are not included due to lack of atomic data. However, the AUG data leaves little room for n=3 contributions to play a significant role in determining the shape or magnitude of CXRS impurity density profiles. From the neutral density calculations and the available atomic data, it is clear that for AUG neutral beams and plasma parameters, there are three neutral populations that contribute significant emission to the total measured CXRS signal. For the 60 keV beams, the measured CXRS emission is dominated by the ground state of the first energy component and the n=2 beam halo population. The latter can contribute up to 35% to the total signal. For the 93 keV beams, the first Figure 17. Sketch of the geometry used within CHICA to define the light-cone of each LOS and calculate the total collected emission. 19 Plasma Phys. Control. Fusion 60 (2018)095007 R M McDermott et al
energy component is responsible for 50%–70% of the CXRS intensity, while the contribution of the n=2 halo is typically less than 20%, and the remainder comes from the ground state of the second beam energy component. For the halo calculations, it is important to not only determine the source and loss of halo neutrals, but also to determine the spread of the halo perpendicular to the beam volume. Correctly including this population in the calculation of CXRS impurity densities is critical, as neglecting it leads to incorrect overall magnitudes and an artificial peaking of the profiles for both high and low energy beams. The increase in measured R LnBvalues can be on the order of RL 0 .5 1.5 nB <D < , which will seriously impact any impurity transport studies that compare to predictions from either neoclassical or gyrokinetic theory. The edge LOS of the core CXRS diagnostics at AUG intersect the edge pedestal inside the neutral beam volume. Therefore, the impurity density is not constant along the LOS, complicating the interpretation of the measured CXRS intensities. Within CHICA a forward model for the edge impurity densities has been implemented that enables the reconstruction of the edge profiles out to the pedestal top and greatly improves the agreement between the core and edge CXRS diagnostics. Acknowledgments The authors would like to thank Luis Fernandez Menchero and the ADAS group for their assistance with the n=2 thermal charge exchange cross-sections and Matthew Reinke for his contributions to the initial development of the CHICA code. We thankfully acknowledge the financial support from the Helmholtz Association of German Research Centres through the Helmholtz Young Investigators Group program. In addition, this work has been carried out within the framework of the EUROfusion Consortium and has received funding from the Euratom research and training programme 2014–2018 under grant agreement No. 633053. The views and opinions expressed herein do not necessarily reflect those of the European Commission. ORCID iDs R M McDermott https://orcid.org/0000-0002-8958-8714 A Kappatou https://orcid.org/0000-0003-3341-1909 C Bruhn https://orcid.org/0000-0003-1196-326X E Viezzer https://orcid.org/0000-0001-6419-6848 References [1]Fonck R J et al 1984 Phys. Rev. A29 3288 [2]Isler R C et al 1994 Plasma Phys. Control. Fusion 36 171 [3]Anderson H et al 2000 Plasma Phys. Control. Fusion 42 781 [4]Hoekstra R et al 1998 Plasma Phys. Control. Fusion 40 1541 [5]Duval B P 1986 Spectroscopic studies of highly ionised atoms Doctoral Thesis Keble College, Oxford, Bodleian Library [6]Hogan J et al 1982 J. Nucl. Mater. 111 413–9 [7]Dux R et al 2012 39th EPS Conf. and 16th Int. Congress on Plasma Physics (Stockholm, Sweden)(http://ocs.ciemat.es/ EPSICPP2012PAP/pdf/P2.049.pdf) [8]von Hellermann M G et al 2005 Phys. Scr. T120 19 [9]Isler R C et al 1984 Nucl. Fusion 24 1559 [10]Viezzer E et al 2012 Rev. Sci. Instrum. 83 103501 [11]McDermott R M et al 2017 Rev. Sci. Instrum. 88 073508 [12]Heidbrink W et al 2010 Commun. Comput. Phys. 10 716–41 [13]Dux R et al 2011 38th EPS Conf. on Plasma Physics (Strasbourg, France, 2011)(http://ocs.ciemat.es/ EPS2011PAP/pdf/P1.056.pdf) [14]Mandl W et al 1993 Plasma Phys. Control. Fusion 35 1373–94 Figure 18. Edge boron density profiles evaluated during the plasma position sweep in discharge #31916. On the left the impurity densities are evaluated assuming constant impurity density along the LOS and without including the full light-cone geometry of the LOS. On the right, these two effects have been included in the analysis. An impurity density pedestal top location of ρ ψ =0.96 was used for these calculations. 20 Plasma Phys. Control. Fusion 60 (2018)095007 R M McDermott et al
[15]Streibl B et al 2003 Fusion Sci. Technol. 44 578 [16]Geiger B et al 2013 Rev. Sci. Instrum. 84 113502 [17]Geiger B 2013 Fast-ion transport studies using FIDA spectroscopy at the ASDEX upgrade tokamak Dissertation LMU München, Germany (http://edoc.mpg.de/get.epl? fid=101711&did=633269&ver=0) [18]Summers H P 2004 The ADAS User manual version 2.6 http://adas.ac.uk/ [19]Kraus M 2010 Heiz- und Stromprofile bei Neutralteilcheninjektion in Tokamakplasmen Max Planck Institute for Plasma Physics Internal Report 5/122 Max Planck Institute for Plasma Physics http://edoc.mpg.de/display.epl? mode=doc&id=478136&col=33&grp=1311#cb [20]Summers H P et al 2006 Plasma Phys. Control. Fusion 48 263 [21]Grierson B et al 2014 Rev. Sci. Instrum. 85 103502 [22]Igenbergs K et al 2012 J. Phys. B: At. Mol. Opt. Phys. 45 065203 [23]Guzman F et al 2010 J. Phys. B: At. Mol. Opt. Phys. 43 144007 [24]Kappatou A et al 2018 Plasma Phys. Control. Fusion 60 055006 21 Plasma Phys. Control. Fusion 60 (2018)095007 R M McDermott et al