Citation: Romero, M.; Herrero-Climent, M.; Ríos-Carrasco, B.; Brizuela, A.; Romero, M.M.; Gil, J. Investigation of the Influence of Roughness and Dental Implant Design on Primary Stability via Analysis of Insertion Torque and Implant Stability Quotient: An In Vitro Study. J. Clin. Med. 2023,12, 4190. https://doi.org/10.3390/ jcm12134190 Academic Editor: Daniele De Santis Received: 17 May 2023 Revised: 11 June 2023 Accepted: 16 June 2023 Published: 21 June 2023 Copyright: © 2023 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https:// creativecommons.org/licenses/by/ 4.0/). Journal of Clinical Medicine Article Investigation of the Influence of Roughness and Dental Implant Design on Primary Stability via Analysis of Insertion Torque and Implant Stability Quotient: An In Vitro Study Marta Romero 1, Mariano Herrero-Climent 1,2 , Blanca Ríos-Carrasco 1, Aritza Brizuela 3, Manuel María Romero 1and Javier Gil 4,* 1Department of Periodontology, School of Dentistry, Universidad de Seville, C/Avicena S/N, 41009 Seville, Spain; romer[email protected] (M.R.);
[email protected] (M.H.-C.); [email protected] (B.R.-C.); mmromer[email protected] (M.M.R.) 2Porto Dental Institute, Periodontology Department, Symmetrya Prothesis, Av. de Montevideu 810, 4150-518 Porto, Portugal 3Densia Reserach Group, Facultad de Ciencias de la Salud, Universidad Europea Miguel de Cervantes, C/del Padre Julio Chevalier 2, 47012 Valladolid, Spain; [email protected] 4Bioengineering Institute of Technology, Faculty of Medicine and Health Sciences, Universidad International de Cataluña, C/Josep Trueta s/n, Sant Cugat del Vallés, 08195 Barcelona, Spain *Correspondence: xavier[email protected] Abstract: In the placement of dental implants, the primary fixation between the dental implant and the bone is of great importance and corresponds to compressive mechanical fixation that aims to prevent micromovement of the implant. The aim of this research was to determine the role of roughness and the type of dental implant (tissue-level or bone-level) in implant stability, measured using resonance frequency analysis (RFA) and insertion torque (IT). We analyzed 234 titanium dental implants, placed in fresh calf ribs, at the half-tissue level and half-bone level. The implant surface was subjected to grit-blasting treatments with alumina particles of 120, 300, and 600 µ m at a projection pressure of 2.5 bar, resulting in three types of roughness. Roughness was determined via optical interferometry. The wettability of the surfaces was also determined. Implant stability was measured using a high-precision torquemeter to obtain IT, and RFA was used to determine the implant stability quotient (ISQ). The results show that rough surfaces with Sa values of 0.5 to 4 µ m do not affect the primary stability. However, the type of implant is important; bone-level implants obtained the highest primary stability values. A good correlation between the primary stability values obtained via IT and ISQ was demonstrated. New in vivo studies are necessary to know whether these results can be maintained in the long term. Keywords: dental implant; titanium; primary stability; insertion torque; implant stability quotient 1. Introduction Several papers since the 1980s have confirmed the importance of the implant surface in implant osseointegration. Improved implant surface properties result in faster and stronger bone formation, which may confer better stability during the healing process. In this sense, a positive correlation has been found between surface roughness and bone-to-implant contact and pushout strength [ 1 ]. The final goal of implant surface modification is definitely to create a more osteophytic surface that attracts bone-forming cells [2]. This is the reason why researchers have tried to improve implant properties using methods such as roughening the implant surface using different techniques, such as subtractive and additive methods, including physical (turning and blasting), chemical (acid etching and alkalis), electrochemical (electropolishing and anodizing), deposition (plasma-spraying, sol–gel), and biochemical (proteins) methods [ 3 , 4 ]. Many of these studies have been performed with different dental implant macro-designs, lengths, diameters, thread depths, J. Clin. Med. 2023,12, 4190. https://doi.org/10.3390/jcm12134190 https://www.mdpi.com/journal/jcm
J. Clin. Med. 2023,12, 4190 2 of 14 thread spacing, micro-threads, and other geometrical aspects of dental implants [5–10] . Conical or cylindrical implant shapes and the use of platform switching have also been studied, and in all cases, the authors determined the mechanical load transfer to the bone [ 11 , 12 ]. Different studies have reported a positive correlation between roughness and osseointegration, using machined surfaces as controls [ 13 – 17 ]. The roughness produces changes in the wettability, surface energy, zeta potential, and other physico-chemical properties that improve bone growth. [18–22]. Therefore, there is sufficient evidence of how implant surface modifications have achieved a stronger bone response, which may explain the high implant survival, and equally, many treatment indications have become more predictable, as in the case of the immediate placement and immediate loading of implants. Moreover, surface-modified implants also lead to the preservation of marginal bone, without the clinically significant superiority of any particular surface or design [23]. On the other hand, primary stability has been defined as the biometric stability measured immediately after implant insertion; it is determined by bone–implant contact, and is the result of the mechanical engagement of an implant with the surrounding bone. It is considered a prognostic marker for long-term implant success. Primary stability will be greater the smaller the micromovements between the surrounding bone and implant. This allows for proper and faster healing, and correct osseointegration [24]. Implant stability at the moment of implant insertion and the absence of micromovements are also critical to determining whether it is possible to immediately load the implant. Good fixation will prevent the formation of connective tissue between implant and bone. It is well known that certain micromovements cause the destruction of new cells and blood vessels. It has been suggested that micromotion between the implant and the surrounding bone must not exceed a threshold value of 150 µ m for successful osseointegration. Any movement at the micrometer range above that threshold can induce stress and strain, hinder the recruitment of new cells, and negatively influence healing bone and bone remodeling, leading to the formation of fibrous tissue [24–27]. Different methods have been utilized to measure the stability of implants, among them, resonance frequency analysis (RFA), with the Osstell device (Osstell AB, Goteborg, Sweden), commonly used today. This method is based on measurement of the oscillation frequency of the implant in the bone, induced by a magnetic pulsing stimulus, which is transformed into ISQ values (implant stability quotient) in a range between 1 and 100 [ 28 – 30 ]. In vitro studies have shown that the ISQ increases with the stiffness of the bone–implant interface. RFA has proven to be a useful tool for assessing the evolution of implant osseointegration because it allows for clinical measurement of the quality of the bone–implant interface immediately after insertion (primary stability), as well as biological anchorage or secondary stability that occurs in the following weeks [ 29 ]. Insertion torque (IT) analysis using a torquemeter is another method that provides information about bone density at the implant placement moment, and to some extent, about implant stability. High values of IT indicate a reduced micromovement in the initial healing period, before osseointegration is achieved in the cortical bone [29,30]. Sennerby and Meredith [ 31 ] reported that implant design and surface structure could have an impact on implant stability during initial healing. In a study in dogs, Rompen et al. [ 32 ] showed that surface-modified implants maintained implant stability, whilst machined surfaces experienced a decrease in stability during the early healing period. In another study, a machined surface was compared with an oxidized one using an immediate loading protocol, and the authors reported a greater decrease in stability for the machined implants during the first 3 months post-loading [ 33 ]. Many researchers have found that implant surface, rough or smooth, has no influence on implant stability [ 34 – 37 ], although some papers have reported different implant stability with different implant surfaces [ 38 ]. Although there are several studies in the literature that compare different surfaces of commercialized implants and their influence on osseointegration and primary stability,
J. Clin. Med. 2023,12, 4190 3 of 14 we didn’t find any papers in which, on the same commercialized implant, the surface roughness was modified to assess its effect on implant stability. On the other hand, in relation to implant macro-design, we can distinguish between implants placed at the bone level—at the same level or slightly below the bone crest (bonelevel implants)—and those with the active surface of the implant located within the bone, and the coronal portion—a polished titanium ring—on the bone crest (transmucosal); on this supracrestal part is established the soft tissue attachment for tissue-level implants. These transmucosal implants have been shown to represent a valid option in daily practice, particularly when applied in immediate-loading full-arch rehabilitation [ 39 , 40 ]. In this sense, we have not found any primary stability studies comparing different levels of roughness on two different implant designs, tissues, and bone-level dental implants. The aim of this research was to determine the role of surface roughness and of the type of dental implant (tissue-level or bone-level) on the implant stability, measured via RFA and IT. In addition, the correlation of RFA with IT was analyzed. The null hypothesis was a follows: The surface topography of dental implants, as well as the presence, or lack thereof, of a polished ring at the neck (tissue- or bone-level dental implant), have no influence on their primary stability after insertion into the bone. This study presents an original study of the primary stability of dental implants using two types of dental implant and, at the same time, with three levels of roughness. The study of different levels of roughness with the same design of dental implants and the comparison between the two types of implant (tissue- and bone-level) are what give this research originality. 2. Methods 2.1. Dental Implants Two hundred and thirty-four implants were employed from Klockner Implant System (SOADCO, Les Escaldes, Escaldes-Engordany, Andorra). Two different implant designs were used: Essential ® Cone (internal connection and double-threaded with an atraumatic apex, progressive core, and machined collar (a tissue-level implant)), and VEGA ® (internal connection, double-threaded with an atraumatic apex and progressive core designed for placement at the crestal level). The dental implants were made with commercially pure grade 3 titanium. The implants used are shown in Figure 1. J. Clin. Med. 2023, 12, x FOR PEER REVIEW 3 of 15 Although there are several studies in the literature that compare different surfaces of commercialized implants and their influence on osseointegration and primary stability, we didn’t find any papers in which, on the same commercialized implant, the surface roughness was modified to assess its effect on implant stability. On the other hand, in relation to implant macro-design, we can distinguish between implants placed at the bone level—at the same level or slightly below the bone crest (bonelevel implants)—and those with the active surface of the implant located within the bone, and the coronal portion—a polished titanium ring—on the bone crest (transmucosal); on this supracrestal part is established the soft tissue attachment for tissue-level implants. These transmucosal implants have been shown to represent a valid option in daily practice, particularly when applied in immediate-loading full-arch rehabilitation [39,40]. In this sense, we have not found any primary stability studies comparing different levels of roughness on two different implant designs, tissues, and bone-level dental implants. The aim of this research was to determine the role of surface roughness and of the type of dental implant (tissue-level or bone-level) on the implant stability, measured via RFA and IT. In addition, the correlation of RFA with IT was analyzed. The null hypothesis was a follows: The surface topography of dental implants, as well as the presence, or lack thereof, of a polished ring at the neck (tissue- or bone-level dental implant), have no influence on their primary stability after insertion into the bone. This study presents an original study of the primary stability of dental implants using two types of dental implant and, at the same time, with three levels of roughness. The study of different levels of roughness with the same design of dental implants and the comparison between the two types of implant (tissue- and bone-level) are what give this research originality. 2. Methods 2.1. Dental Implants Two hundred and thirty-four implants were employed from Klockner Implant System (SOADCO, Les Escaldes, Escaldes-Engordany, Andorra). Two different implant designs were used: Essential Cone (internal connection and double-threaded with an atraumatic apex, progressive core, and machined collar (a tissue-level implant)), and VEGA (internal connection, double-threaded with an atraumatic apex and progressive core designed for placement at the crestal level). The dental implants were made with commercially pure grade 3 titanium. The implants used are shown in Figure 1. Figure 1. (A) Essential dental implant: tissue-level. (B) VEGA dental implant: bone-level. Figure 1. (A) Essential dental implant: tissue-level. (B) VEGA dental implant: bone-level.
J. Clin. Med. 2023,12, 4190 4 of 14 2.2. Surface Treatments Dental implants were prepared with three different surface treatments and divided into six groups (three different surface treatments × two types of dental implant—tissue- and bone-level) with thirty-nine implants for each group. A total of 234 dental implants. The treatment was realized using alumina particles (120, 300, and 600 µ m size) with 0.25 MPa blasting pressure until roughness saturation was achieved. The distance of projection from the gun to the surface was 0.80 m. The different abrasive sizes produced different roughnesses on the surfaces of the dental implants. The sterilization and packing of all implants were carried out at SOADCO, S.L. (Klockner Dental Implants, Escaldes Engordany, Andorra). The consensus report of implant surfaces and design (working group 4) established three types of surface roughness for oral implants [ 34 ], specified by Albrektsson and Wenneberg [35]: •Minimally rough surfaces: Sa values 0.5–1 µm. •Moderately rough surfaces: Sa values 1–2 µm. •Rough surfaces: Sa values > 2 µm. The implants were distributed among the following groups: Group A: 39 Essential®Cone minimally rough surface implants. Group B: 39 VEGA®minimally rough surface implants. Group C: 39 Essential®Cone rough surface implants. Group D: 39 VEGA®rough surface implants. Group E: 39 Essential®Cone moderately rough surface implants. Group F: 39 VEGA®moderately rough surface implants. 2.3. Roughness White light interferometer microscopy (Wyko NT1100, Veeco, Plainview, NY, USA) was used. Data analysis was performed using Wyko Vision 232 TM software (Veeco, Plainview, NY, USA). A Gaussian filter was used to separate waviness and form from the roughness of the surface. Measurements were taken of three different surfaces of each type of surface treatment to characterize the amplitude and spacing roughness parameters, Sa and Pc, respectively. Sa and Pc were calculated by averaging the values of all individual profiles that were evenly distributed along the surface analyzed. A scanning electron microscope (SEM) (JSM 6400, Jeol, Tokyo, Japan) was used to qualitatively analyze the surface topography of the implants. 2.4. Wettability Contact angle analysis was performed with ultrapure distilled water and formamide, and the corresponding data were analyzed using SCA20 (Dataphysics, Filderstadt, Germany). Contact angle measurements were taken using the sessile drop method. Drops were generated using a micrometric syringe and were deposited over discs. 2.5. Implant Stability Analysis The implants were placed in fresh calf ribs, simulating bone quality type II according to the expertise of the experienced clinician, who performed the insertion of the implants and compared the bone radiographs with the images of the Lekholm and Zarb classification by bone density [ 41 ]. All implants used were 10 mm in length and 4 mm in diameter. Figure 2 shows the implantation in bovine bone.
J. Clin. Med. 2023,12, 4190 5 of 14 J. Clin. Med. 2023, 12, x FOR PEER REVIEW 5 of 15 Figure 2. Surgeon placing the implants in fresh bovine bone. The bone site preparations were performed according to the manufacturer’s protocol by an experienced clinician. The protocol began with the use of a sequence of drills, starting with a lanceolated marking bur, continuing with different pilot drills of growing diameter (1.2, 2.8, and 3.5 mm, and finishing with a profile bur. To insert the implant, a lowspeed (5 rpm) surgical motor with handpiece was used. After that, a hand torque wrench was used to submerge the implants up the right level. All implants were 4 × 10mm. The osteotomy for each implant was performed under abundant saline solution irrigation with a minimum distance of 4 mm between osteotomies. The rough/smooth interface was placed at the bone crest level for the tissue-level implants (groups A, C, and E), while bone-level implants (groups B, D, and F) were placed at the crestal level. Implant insertion was performed manually, and IT was registered using a calibrated Analog Dynamometer—BTG90CN (Tohnichi, Tokyo, Japan). Figure 3 shows the torquemeter used for the determination of insertion torque. Figure 3. Insertion torque analysis for the determination of primary stability. Figure 2. Surgeon placing the implants in fresh bovine bone. The bone site preparations were performed according to the manufacturer’s protocol by an experienced clinician. The protocol began with the use of a sequence of drills, starting with a lanceolated marking bur, continuing with different pilot drills of growing diameter (1.2, 2.8, and 3.5 mm, and finishing with a profile bur. To insert the implant, a low-speed (5 rpm) surgical motor with handpiece was used. After that, a hand torque wrench was used to submerge the implants up the right level. All implants were 4 × 10mm. The osteotomy for each implant was performed under abundant saline solution irrigation with a minimum distance of 4 mm between osteotomies. The rough/smooth interface was placed at the bone crest level for the tissue-level implants (groups A, C, and E), while bone-level implants (groups B, D, and F) were placed at the crestal level. Implant insertion was performed manually, and IT was registered using a calibrated Analog Dynamometer—BTG90CN (Tohnichi, Tokyo, Japan). Figure 3shows the torquemeter used for the determination of insertion torque. J. Clin. Med. 2023, 12, x FOR PEER REVIEW 5 of 15 Figure 2. Surgeon placing the implants in fresh bovine bone. The bone site preparations were performed according to the manufacturer’s protocol by an experienced clinician. The protocol began with the use of a sequence of drills, starting with a lanceolated marking bur, continuing with different pilot drills of growing diameter (1.2, 2.8, and 3.5 mm, and finishing with a profile bur. To insert the implant, a lowspeed (5 rpm) surgical motor with handpiece was used. After that, a hand torque wrench was used to submerge the implants up the right level. All implants were 4 × 10mm. The osteotomy for each implant was performed under abundant saline solution irrigation with a minimum distance of 4 mm between osteotomies. The rough/smooth interface was placed at the bone crest level for the tissue-level implants (groups A, C, and E), while bone-level implants (groups B, D, and F) were placed at the crestal level. Implant insertion was performed manually, and IT was registered using a calibrated Analog Dynamometer—BTG90CN (Tohnichi, Tokyo, Japan). Figure 3 shows the torquemeter used for the determination of insertion torque. Figure 3. Insertion torque analysis for the determination of primary stability. Figure 3. Insertion torque analysis for the determination of primary stability.
J. Clin. Med. 2023,12, 4190 6 of 14 Once the implants were in place, primary stability was measured by means of resonance frequency analysis (Osstell ISQ ® ), following the manufacturer’s instruction and using the appropriate transducers (SmartPeg) for each implant design (Figure 4). Resonance frequency analysis (RFA) is used to determine primary stability. It is a recent technique but is highly sensitive and non-destructive. This technique determines the stability as a function of the rigidity between the dental implant–bone tissue system by means of a magnetic peg that is connected to the implant. The values range from 1 to 100, and these are the implant stability quotient (ISQ). J. Clin. Med. 2023, 12, x FOR PEER REVIEW 6 of 15 Once the implants were in place, primary stability was measured by means of resonance frequency analysis (Osstell ISQ®), following the manufacturer’s instruction and using the appropriate transducers (SmartPeg) for each implant design (Figure 4). Resonance frequency analysis (RFA) is used to determine primary stability. It is a recent technique but is highly sensitive and non-destructive. This technique determines the stability as a function of the rigidity between the dental implant–bone tissue system by means of a magnetic peg that is connected to the implant. The values range from 1 to 100, and these are the implant stability quotient (ISQ). The transducers were hand-screwed, at approximately 6–8 N/cm of torque. For each SmartPeg, two measurements were taken at a perpendicular level. Between measurements, the transducer was unscrewed and screwed again. A new SmartPeg was used every 5 implants. Figure 4. Resonance frequency analysis to determine implant stability quotient. 2.6. Statistical Analysis Variables are described in terms of their mean, median, and standard deviation values. Variable normality was checked with the help of the Shapiro–Wilk test. Due to the lack of normality of the variables, two different non-parametric tests were applied. In the case of two categories, the Mann–Whitney–Wilcoxon test was employed [42,43]. In the Figure 4. Resonance frequency analysis to determine implant stability quotient. The transducers were hand-screwed, at approximately 6–8 N/cm of torque. For each SmartPeg, two measurements were taken at a perpendicular level. Between measurements, the transducer was unscrewed and screwed again. A new SmartPeg was used every 5 implants. 2.6. Statistical Analysis Variables are described in terms of their mean, median, and standard deviation values. Variable normality was checked with the help of the Shapiro–Wilk test. Due to the lack of normality of the variables, two different non-parametric tests were applied. In the case of
J. Clin. Med. 2023,12, 4190 7 of 14 two categories, the Mann–Whitney–Wilcoxon test was employed [ 42 , 43 ]. In the case of more than two categories, the Kruskal–Wallis test was applied, and in order to find differences among groups, the post hoc Dunn’s test with Holm’s correction was employed [44–46]. 3. Results Figure 5shows images of the three types of roughness obtained via grit-blasting with different sizes of abrasive alumina particles. As is well known, the projection of larger particles at the same pressure and projection distance generates a higher roughness value. The roughness values obtained for the different surfaces and implants can be seen in Table 1. As expected, the roughness between the two types of implants does not show statistically significant differences. J. Clin. Med. 2023, 12, x FOR PEER REVIEW 7 of 15 case of more than two categories, the Kruskal–Wallis test was applied, and in order to find differences among groups, the post hoc Dunn’s test with Holm’s correction was employed [44–46]. 3. Results Figure 5 shows images of the three types of roughness obtained via grit-blasting with different sizes of abrasive alumina particles. As is well known, the projection of larger particles at the same pressure and projection distance generates a higher roughness value. The roughness values obtained for the different surfaces and implants can be seen in Table 1. As expected, the roughness between the two types of implants does not show statistically significant differences. Figure 5. The different topographies studied. (A) Minimally rough surfaces: Sa values 0.5–1 μm. (B) Moderately rough surfaces: Sa values 1–2 μm. (C) Rough surfaces: Sa values > 2 μm. Table 1. Values of the different parameters of roughness for the different treatments studied in both types of dental implant. Samples Sa (µm) Sz (µm) S Area Index Average Desv.est. Average Desv.est. Average Desv.est. A 0.55 0.01 3.47 1.53 1.04 0.01 B 0.54 0.07 16.74 1.11 1.03 0.01 C 3.85 0.18 20.96 1.65 1.74 0.05 D 2.76 0.21 19.65 2.00 1.13 0.07 E 1.60 0.22 13.42 2.54 1.66 0.10 F 1.67 0.19 19.72 3.02 1.88 0.09 The contact angles were determined and are shown in Table 2. For the determination of these values, the area index values were corrected. The results are shown in Table 2. Table 2. Wettability of the different surfaces. Samples A B C D E F Mean 90.88 92.34 76.70 78.45 80.92 82. 32 Standard Dev. 5.90 6.00 3.09 2.90 1.85 1.97 The mean IT of the whole sample is 23.87 ± 11.56 N/cm and the mean ISQ value of the whole sample is 67.12 ± 8.75. The descriptive analysis of the study sample’s rugosity Figure 5. The different topographies studied. ( A ) Minimally rough surfaces: Sa values 0.5–1 µ m. (B) Moderately rough surfaces: Sa values 1–2 µm. (C) Rough surfaces: Sa values > 2 µm. Table 1. Values of the different parameters of roughness for the different treatments studied in both types of dental implant. Samples Sa (µm) Sz (µm) S Area Index Average Desv.est. Average Desv.est. Average Desv.est. A 0.55 0.01 3.47 1.53 1.04 0.01 B 0.54 0.07 16.74 1.11 1.03 0.01 C 3.85 0.18 20.96 1.65 1.74 0.05 D 2.76 0.21 19.65 2.00 1.13 0.07 E 1.60 0.22 13.42 2.54 1.66 0.10 F 1.67 0.19 19.72 3.02 1.88 0.09 The contact angles were determined and are shown in Table 2. For the determination of these values, the area index values were corrected. The results are shown in Table 2. Table 2. Wettability of the different surfaces. Samples A B C D E F Mean 90.88 92.34 76.70 78.45 80.92 82.32 Standard Dev. 5.90 6.00 3.09 2.90 1.85 1.97 The mean IT of the whole sample is 23.87 ± 11.56 N/cm and the mean ISQ value of the whole sample is 67.12 ± 8.75. The descriptive analysis of the study sample’s rugosity is shown in Table 3. The results obtained show statistically significant differences in the
J. Clin. Med. 2023,12, 4190 8 of 14 medians among groups for both the IT (H = 96.82.df = 5), with a p-value < 2.2 × 10 −16 , and ISQ variables (H = 94.92. df = 5. p-value = 2.2 × 10 −16 ). The Pearson correlation value of the ISQ and IT variablea is 0.059 (p-value = 0.371), which can be considered a low value. All the ISQ values are shown in Supplementary File (Table S1) and the IT values (Table S1) for different roughness and dental implants. Table 3. Descriptive statistics of IT and ISQ variables by type of dental implant surface. Implants Sa IT ISQ Mean SD Mean SD Median Mean SD Median A 0.55 0.01 25.85 10.44 24 64.13 5.40 65.50 B 0.54 0.07 29.41 11.84 28 70.77 8.59 74.00 C 3.85 0.18 28.44 11.40 26 63.08 8.12 64.00 D 2.76 0.21 14.69 4.35 15 73.04 4.43 74.00 E 1.60 0.22 29.00 11.01 27 59.68 9.51 62.00 F 1.67 0.19 15.28 7.40 14 72.03 5.73 72.50 Figure 6shows a box-plot of the IT for each type of implant, and Figure 7shows a box-plot of the RFA analysis. Table 4shows the p-values of the Dunn post hoc test applied to the IT variable with the Holm correction. According to the p-values obtained, the different dental implant types can be divided in two categories: Essential minimally rough surface, Essential moderately rough surface, and Essential rough surface versus Vega minimally rough surface, VEGA moderately rough surface, and VEGA rough surface. J. Clin. Med. 2023, 12, x FOR PEER REVIEW 8 of 15 is shown in Table 3. The results obtained show statistically significant differences in the medians among groups for both the IT (H = 96.82.df = 5), with a p-value < 2.2·× 10−16, and ISQ variables (H = 94.92. df = 5. p-value = 2.2·× 10−16). The Pearson correlation value of the ISQ and IT variablea is 0.059 (p-value = 0.371), which can be considered a low value. All the ISQ values are shown in supplementary file (Table S1) and the IT values (Table S1) for different roughness and dental implants. Table 3. Descriptive statistics of IT and ISQ variables by type of dental implant surface. Implants Sa IT ISQ Mean SD Mean SD Median Mean SD Median A 0.55 0.01 25.85 10.44 24 64.13 5.40 65.50 B 0.54 0.07 29.41 11.84 28 70.77 8.59 74.00 C 3.85 0.18 28.44 11.40 26 63.08 8.12 64.00 D 2.76 0.21 14.69 4.35 15 73.04 4.43 74.00 E 1.60 0.22 29.00 11.01 27 59.68 9.51 62.00 F 1.67 0.19 15.28 7.40 14 72.03 5.73 72.50 Figure 6 shows a box-plot of the IT for each type of implant, and Figure 7 shows a box-plot of the RFA analysis Figure 6. Boxplot of IT by dental implant type. The median values are included. Figure 6. Boxplot of IT by dental implant type. The median values are included.
J. Clin. Med. 2023,12, 4190 9 of 14 J. Clin. Med. 2023, 12, x FOR PEER REVIEW 9 of 15 Figure 7. Boxplot of ISQ by dental implant type. It can clearly be observed that the ISQ values are, in general, higher for VEGA dental implants. Table 4 shows the p-values of the Dunn post hoc test applied to the IT variable with the Holm correction. According to the p-values obtained, the different dental implant types can be divided in two categories: Essential minimally rough surface, Essential moderately rough surface, and Essential rough surface versus Vega minimally rough surface, VEGA moderately rough surface, and VEGA rough surface. Table 4. p-values of the Dunn post hoc test applied to variable IT with the Holm correction. Essential Moderately Rough Surface Essential Rough Surface VEGA Minimally Rough Surface VEGA Moderately Rough Surface VEGA Rough Surface Essential minimally rough surface 1 1 1 5.23 ×10 −6 2.79 ×10 −6 Essential moderately rough surface 1 0.9106 2.18 ×10−9 9.83 ×10−10 Essential rough surface 1 4.44 ×10 −8 2.05 ×10 −8 VEGA minimally rough surface 4.17 ×10−9 1.93 ×10−9 VEGA moderately rough surface 1 The equivalent results for the ISQ variable are presented in Table 5. In this case, the results are similar, also showing differences between VEGA and Essential, but no intragroup differences in either VEGA or Essential. Figure 7. Boxplot of ISQ by dental implant type. It can clearly be observed that the ISQ values are, in general, higher for VEGA dental implants. Table 4. p-values of the Dunn post hoc test applied to variable IT with the Holm correction. Essential Moderately Rough Surface Essential Rough Surface VEGA Minimally Rough Surface VEGA Moderately Rough Surface VEGA Rough Surface Essential minimally rough surface 1 1 1 5.23 ×10−62.79 ×10−6 Essential moderately rough surface 1 0.9106 2.18 ×10−99.83 ×10−10 Essential rough surface 1 4.44 ×10−82.05 ×10−8 VEGA minimally rough surface 4.17 ×10−91.93 ×10−9 VEGA moderately rough surface 1 The equivalent results for the ISQ variable are presented in Table 5. In this case, the results are similar, also showing differences between VEGA and Essential, but no intra-group differences in either VEGA or Essential. Table 5. p-values of the Dunn post hoc test applied to variable ISQ with the Holm correction. Essential Moderately Rough Surface Essential Rough Surface VEGA Minimally Rough Surface VEGA Moderately Rough Surface VEGA Rough Surface Essential minimally rough surface 0.0462 0.0875 1.94 ×10−42.17 ×10−55.12 ×10−7 Essential moderately rough surface 0.0536 1.68 ×10−81.26 ×10−97.16 ×10−12 Essential rough surface 6.64 ×10−51.10 ×10−52.28 ×10−7 VEGA minimally rough surface 1 0.0978 VEGA moderately rough surface 1 In order to find out whether there were statistically significant differences among the median values by type of implant (Essential versus VEGA) in the IT and ISQ variables,