International Journal of Environmental Research and Public Health Review Review of the Mechanical Behavior of Different Implant–Abutment Connections Ana Sofia Vinhas 1, Carlos Aroso 2, Filomena Salazar 1, Paula López-Jarana 1, JoséVicente Ríos-Santos 3,* and Mariano Herrero-Climent 4 1Department of Periodontology, Instituto Universitário de Ciências da Saúde, 4585-116 Gandra, Portugal; [email protected] (A.S.V.); [email protected] (F.S.); [email protected] (P.L.-J.) 2Department of Prosthodontics, Instituto Universitário de Ciências da Saúde, 4585-116 Gandra, Portugal; [email protected] 3Department of Periodontology, School of Dentistry, Universidad de Sevilla, C/Avicena S/N, 41009 Sevilla, Spain 4Porto Dental Institute, 4150-518 Porto, Portugal;
[email protected] *Correspondence: [email protected]; Tel.: +34-954-481-121; Fax: +34-954-481-157 Received: 14 October 2020; Accepted: 17 November 2020; Published: 23 November 2020 Abstract: Introduction: Different implant–abutment connections have been developed to reduce mechanical and biological failure. The most frequent complications are loss of preload, screw loosening, abutment or implant fracture, deformations at the different interfaces, and bacterial microleakage. Aim: To review the evidence indicating whether the implant–abutment connection type is significant regarding the following issues: (1) maintenance of the preload in static and dynamic in vitro studies; (2) assessment of possible deformations at the implant–abutment interfaces, after repeated application of the tightening torque; (3) evaluation of the sealing capability of different implant connections against microleakage. Materials and Methods: In June 2020, an electronic literature search was performed in Medline, EBSCO host, and PubMed databases. The search was focused on the ability of different implant connections to maintain preload, resist deformation after tightening and retightening, and prevent microleakage. The related titles and abstracts available in English were screened, and the articles that fulfilled the inclusion criteria were selected for full-text reading. Results: The literature search conducted for this review initially resulted in 68 articles, among which 19 articles and 1 systematic review fulfilled the criteria for inclusion. The studies were divided according to the three proposed objectives, with some studies falling into more than one category (maintenance of preload, surface abutment–implant deformation, and resistance to microleakage). Conclusions: Conical abutment appears to result in fewer mechanical complications, such as screw loosening or fractures, and higher torque preservation. After SEM evaluation, damage was observed in the threads of the abutment screws, before and after loading in internal and external connections. Internal hexagon implants and predominantly internal conical (Morse taper) implants showed less microleakage in dynamic loading conditions. We suggest further studies to guarantee excellence in methodological quality. Keywords: implant–abutmentconnection; preload; tighteningtorque; cyclicloading; misfit;microleakage 1. Introduction In recent years, geometries of implant connections have been developed with different mechanical, biological, and esthetic characteristics. Two basic geometries are available: internal and external connections. External connections usually have an external hexagon on the implant platform, whereas internal connections can be divided into internal hexagons, internal octagons, and Morse taper Int. J. Environ. Res. Public Health 2020,17, 8685; doi:10.3390/ijerph17228685 www.mdpi.com/journal/ijerph
Int. J. Environ. Res. Public Health 2020,17, 8685 2 of 20 connections [ 1 ]. The osseointegrated implant and the prosthetic abutment are joined by a screw and have, therefore, been named a screw joint [2]. The external hexagon was the first connection system adopted in modern implantology by Branemark [ 3 ], based on the existence of a hexagon (0.7 mm height); however, this connection has been extensively modified in terms of diameter, height, and insertion torque. This kind of connection presents some advantages. First, it is adequate for the two-step surgical procedure preferred by Branemark [ 3 ] because it alleviates the second stage and the connection phase with the healing abutment. Second, it simplifies the recording of the external connection in the impression and the prosthetic phase due to its adjustability and compatibility with different prosthetic solutions. However, it also presents a number of disadvantages, such as little contact length between the restoration and the hexagonal part of the implant head, some degree of rotation between the platform and the internal hexagon of the restoration, and high tension created in the screw connection. It has been speculated that under high occlusal loads, the external hexagon might allow for micromovements of the abutment, causing instability of the joint, which may result in abutment screw loosening or even fatigue fractures [ 4 – 6 ]. The literature has shown the loosening rate of this type of connection to be between 6% and 48% [ 7 ], thus presenting a mechanical difficulty for the maintenance of the preload (torque for the removal of the pillar must be 10% lower than that of installation) [8,9]. Internal connections have been introduced with a Morse cone of different degrees of inclination, depending on the commercial brand [ 10 ], to lower or eliminate the mechanical complications of the external connection and to reduce the stress transferred to the crestal bone [ 11 – 14 ]. In the internal hexagonal system, the hexagon and the screw pass into the implant body so the prosthetic component is more stable. The internal hexagon connection was developed as an evolution of the external hexagon, with the aim of increasing the load absorption under a lateral force. This reduces mechanical and biological complications, such as screw loosening, fracture, and marginal bone loss. The greater depth of the connection in the fixture body allows more homogeneous dissipation of the mechanical stress; the stress is spread on the implant wall and, consequently, to the bone surrounding the entire implant and not only at the crestal level. [ 15 ] A conical connection is a particular kind of internal connection in which the abutment is fixed to the implant using the mechanical properties of a machine taper. A male member of a conical shape fits into a female socket, which has a matching taper of equal angle. The connection works by locking the two components by mechanical friction between the wall of the abutment and the implant. Although mechanical friction has been demonstrated to be strong enough, implant companies have also implemented screw retention and antirotational systems. However, to date, no qualitative data exists comparing the mechanical behavior of external and internal connections [ 7 ]. The internal connection using a Morse cone creates a more accurate bond between the implant and the abutment, which reduces the movement of the interface and decreases the loosening of the screw (torque for the removal of the abutment must be 17% greater than that of installation) [ 8 , 9 ]. The Morse cone has an internal cone of 8◦or 11◦, which can protect against screw loosening [9,16]. As described previously, the different implant–abutment connection designs have very different characteristics, which can affect mechanical stability. The failure of an implant is related to two problems: biological and mechanical factors. Biological causes essentially relate to periimplantitis, which affects the soft and hard tissues around the implants, whereas mechanical causes involve prosthetic components, namely, overload of the prothesis–implant–pillar complex, implant fracture, abutment fracture, loosening of the screws, and fracture of the superstructure (metal/ceramic) [ 17 , 18 ]. Abutment screw stability can be affected by preload, the effect of settling, and screw geometry [2,19,20] . Preload is the force that is generated when the screw is tightened using a given torque [ 2 , 19 , 21 ]. Torque is defined as the movement produced by applying tangential force to the screw and is usually expressed in newton centimeters (Ncm). When applying the preload to a screw, the connected elements are kept in compression, and the screw receives small impacts because most of the load is absorbed by the components of the implant–abutment junction [ 22 ]. The initial preload on the screw is usually inserted by applying torque using a torque wrench. One of the major causes of screw loosening is the
Int. J. Environ. Res. Public Health 2020,17, 8685 3 of 20 “loss of preload”. Only 10% of the initial torque is transformed into preload, whereas the remaining 90% is used to overcome the friction between the surface irregularities [ 2 , 19 , 21 ]. In the tightening of the components of the connection, tension occurs with a consequent compression between the structures of the joint. Screw loosening is one of the most common mechanical complications of implant treatment, with an estimated annual rate of 2.1% [ 23 ]. Estimated rates are 10.4% and 20.8% over 5 and 10 years, respectively [ 24 ]. From a clinical perspective, the loosening of the screw is greater in external connections than in internal connections, with an incidence of loose screws of 38% in external hexagon systems [ 25 , 26 ]. However, the ratio of torque to preload is not linear and is affected by several factors: coefficient of friction, geometry, and properties of contact surface materials. The first is the most influential and depends on the hardness of the threads, the finish of the surfaces, the lubricant used, and the tightening speed [ 27 ]. A preload torque between 10 and 35 Ncm is recommended by different manufacturers, depending on the screw manufacturing material and the morphology of the abutment–implant connection [2]. Another important phenomenon experienced by the screw joint is the settling effect. This occurs because neither the interior torque nor the screw is perfectly fabricated and without irregularity; therefore, these rough areas are smoothed, causing a loss of 2–10% of the initial preload [ 2 , 19 ]. Torque loosening causes micromovements in the interface screw, abutment, and implant body, which cause both mechanical and biological problems. This misalignment of the fitting results in the colonization of bacteria at the interface and is a major challenge to the success of the implant. Microleakage may be defined as the clinically undetectable passage of bacteria, fluids, molecules, or ions between a cavity wall and the restorative materials. Microleakage depends directly on the marginal accuracy of the components (fit, tolerance, presence of microgaps) [ 28 ]. The problems associated with microgaps between the implant and the abutment are both biological and mechanical. The biological problem relates to the presence of bacteria that have been found in the apical portion of the abutment screw; in vivo , this could produce a bacterial reservoir that could interfere with the long-term health of the peri-implant tissues. The mechanical problem relates to the micromovements and possible loosening or fracture due to the fatigue of screw-retained abutments [ 29 ]. The external hexagon, in fact, is subject to micromovements under lateral load, and this may create a microgap at the abutment–fixture interface [ 30 ]. Different authors have investigated this condition, and this microgap can lead to microleakage and bacterial infiltration that may affect the long-term success of dental implants [31]. The stability and integrity of the abutment–implant connection, by means of a screw, are fallible from the moment the prosthetic elements are attached. This fallibility depends on the applied preload, wear of the components, and function. It is necessary to evaluate and quantify, with in vitro studies, the loss of torque before and after loading and the integrity of the system structures in the different connections. The current work aims to review the existing literature to evaluate, according to the type of connection, the maintenance of the preload, the assessment of possible deformations at the different interfaces after repeated application of the tightening torque, and the sealing capability of different implant connections against microleakage. 2. Materials and Methods 2.1. Search Strategy and Study Selection The following analysis was performed according to the guidelines and the principles of an integrative review. The review is focused on the guiding question, “is the implant–abutment connection design important in the mechanical behavior of dental implants?”. Dental literature in Medline, EBSCO Host, and PubMed databases was searched from January 2004 to June 2020. The literature search was limited to journals available in English. The keywords were free-text words and included a combination of the following: implant abutment connections; preload; tightening torque; cyclic loading; implant abutment deformation; misfit; microleakage. Manual and electronic searches were performed to select the relevant articles.
Int. J. Environ. Res. Public Health 2020,17, 8685 4 of 20 2.2. Inclusion and Exclusion Criteria The screening of the articles was conducted, as shown in the flow chart of Figure 1. In the present review, the following inclusion criteria were as follows: articles in English published between January 2004 and June 2020; in vitro studies and systematic reviews, with a clear aim of investigating the relationship of different implant connections to loss of preload, surface abutment–implant deformation, and resistance to microleakage. Case reports, human trials, and studies involving animals were excluded. Int. J. Environ. Res. Public Health 2020, 17, x 4 of 19 2.2. Inclusion and Exclusion Criteria The screening of the articles was conducted, as shown in the flow chart of Figure 1. In the present review, the following inclusion criteria were as follows: articles in English published between January 2004 and June 2020; in vitro studies and systematic reviews, with a clear aim of investigating the relationship of different implant connections to loss of preload, surface abutment–implant deformation, and resistance to microleakage. Case reports, human trials, and studies involving animals were excluded. Figure 1. Flow chart presenting the screening of articles related to the mechanical behavior of different implant connections to be included in the review. 2.3. Data Extraction All studies meeting the criteria were obtained, screened independently, and analyzed according to the stages of an integrative review process. The literature search initially resulted in a total of 68 articles, of which 27 were selected after an evaluation of their titles and abstracts. Full articles were analyzed, and 19 in-vitro studies and 1 systematic review were considered eligible for the review [Table 1]. Figure 1. Flow chart presenting the screening of articles related to the mechanical behavior of different implant connections to be included in the review. 2.3. Data Extraction All studies meeting the criteria were obtained, screened independently, and analyzed according to the stages of an integrative review process. The literature search initially resulted in a total of 68 articles, of which 27 were selected after an evaluation of their titles and abstracts. Full articles were analyzed, and 19 in-vitro studies and 1 systematic review were considered eligible for the review [Table 1].
Int. J. Environ. Res. Public Health 2020,17, 8685 5 of 20 Table 1. Summary of included articles. Author Year Type of Study Purpose Type of Connection Number of Implants Torque Used Intervention Evaluation Outcome of the Study Khraisat A et al. [32] 2004 In Vitro Effect of lateral cyclic loading with different load positions on abutment screw loosening EXTERNAL HEXAGON (Branemark) divided into three groups (n=50) 32 Ncm A-Cyclic load 50 N 1 ×106cycles applied centrally. B- cyclic load 50 N 1 ×106cycles applied eccentrically. C-control unloaded. Torque gauge Micrometer microscope (1 µm) SEM RTVs were preserved under eccentric lateral loading compared with centric loading. SEM analysis revealed mild burnishing and scuffing of the screw thread surfaces in all groups. Marked burnishing was observed at the hexagon corners on the compression sides for group B implants. Piermatti J et al. [21] 2006 In Vitro Examine effects of connection design upon screw stability EXTERNAL HEXAGON (10 BioLok; 10 Nobel Biocare) INTERNAL CONNECTION (10 Zimmer; 10 Astra Tech) (n=40) 32 Ncm 10 samples of each system were loaded to 200 N for 1 ×106cycles. Screws were tightened to manufacturers’ recommendations. Torque audits were done after 250,000, 500,000, 750,000, and 1,000,000 cycles. Torque meter The design of the joint (i.e., internal vs. external connections) was not a significant factor in loss of torque. What did appear important was the screw design. Bio-Lok abutment screws (external connection) maintained their tightness in the best of the four systems. Tsuge T et al. [33] 2009 In Vitro Evaluate the effect of eccentric cyclic loading on abutment screw loosening in internal and external hexagon implants with two screw materials INTERNAL HEXAGON EXTERNAL HEXAGON (n=64) 4 groups20 Ncm Load was applied one million times (1.0 × 106cycles] RTV measured before (initial preload) and after loading (post-loading). Changes in the superstructure and movement of the abutment were inspected using visual and tactile checks every 50,000 cycles. Damage to the surfaces of selected abutment screws was observed using SEM. Torque gauge Fatigue testing machine SEM In all the groups, postloading preload was significantly higher than the initial preload. Implant–abutment connection did not have an effect, but the abutment screw material did. Screws damage was probably due to screw tightening and was observed on the flank near the crest. However, no abnormal wear or damage due to micromovement or bending caused by cyclic loading was observed on the abutment screws in all the groups. Ricomini Filho AP et al. [34] 2010 In Vitro Evaluate the preload loss and bacterial penetration through the implant–abutment interface of conical and external hexagon connection systems subjected to thermal cycling and mechanical fatigue. EXTERNAL HEXAGON MORSE TAPER (n=6) 32 Ncm 15 Ncm The assemblies were subjected to a thermal cycling regimen (1000 cycles of 5 and 55 ◦C) and to mechanical fatigue (1.0 million cycles, 1.0 Hz, 120 N). The assemblies were immersed in tryptic soy +yeast extract broth containing S sanguinis and incubated at 37 ◦C and 10% CO2for 72 h. Detorque values were recorded. The bacterial penetration was assessed, and the abutments were observed by SEM. Electronic torque controller Mechanical fatigue machineSEM All screw abutment systems showed significantly higher detorque values when subjected to TM, and all conical systems presented bacterial penetration. The results show no relationship between the preload loss and bacterial penetration. SEM micrographs show no bacterial cells on the surface of the external hexagon abutment screw, confirming the microbiological assay.
Int. J. Environ. Res. Public Health 2020,17, 8685 6 of 20 Table 1. Cont. Author Year Type of Study Purpose Type of Connection Number of Implants Torque Used Intervention Evaluation Outcome of the Study Cashman PM et al. [35] 2011 In Vitro Compare the abutments fatigue resistance to simulated function, using Removal torque Values as an indication of residual preload at the implant–abutment interface. INTERNAL HEXAGON (Straumann) (n=40) Four groups: 10 abutments from each manufacturer were evaluated for RTV with and without fatigue loading 35 Ncm Baseline: torque was carried out twice, 10 min apart. RTV obtained 1 h later. Postfatigue: torque was carried out twice, 10 min apart. RTV postfatigue obtained 1 h later. 10 to 200 N, 15 Hz for 5 × 10 6 cyclesSEM was carried out at baseline and post fatigue cycling to visualize thread geometry and abutment–implant interface. Digital torque gauge Moving-magnet linear motor to load specimens SEM The effect of component manufacturer resulted in a significantly higher RTV in the control group, indicating greater residual preload. There was no significant decrease in postfatigue RTV for either manufacturer compared to baseline. Differences in surface finish and machining tolerances were visualized with SEM. Ferreira MB et al. [36] 2012 In Vitro Evaluate the torque maintenance of retention screws of tapered abutments and cylinders of Morse taper implants submitted to retightening and detorque measurements MORSE TAPER (n=12) 30 Ncm Detorque values were measured by an analogic torque gauge after 3 min of torque insertion. The detorque was measured 10 times for each retention screw of groups I and II, for a total of 120 detorque measurements in each group. Analogic torque gauge The abutment and cylinder screws exhibited torque loss after insertion, which indicates the need for retightening during the function of the implant-supported prostheses. MurmuraG et al. [37] 2013 In Vitro Evaluation of the preload distribution in screw-retained implant systems under cyclic load. INTERNAL HEXAGON (35 Xsign) INTERNAL OCTAGON (35 SSO) (n=70) 25 Ncm 35 Ncm Cyclic load between 20 and 200 N for 1 × 10 6 cycles. After mechanical tests, samples were sectioned along the long axis and analyzed under SEM. Five implants with internal hexagon and five implants with internal octagon were collected for photoelastic analysis. The loss of preload was evaluated through the presence or absence of abutment mobility SEMPhotoelastic analysis The design of the abutment connection area affects the functional load transfer to the fixture and connection screw; screw-retained abutment based on an internal octagonal connection is less likely to come loose after cyclic load. Jorge VA et al. [38] 2013 In Vitro Evaluate the role of the implant/abutment system on torque maintenance of titanium retention screws and vertical misfit of screw-retained implant-supported crowns before and after mechanical cycling. MORSE TAPER EXTERNAL HEXAGON (n=30) 20 Ncm 30 Ncm Retention screws received insertion torque and after 3 min, initial detorque was measured. Crowns were retightened and submitted to cyclic loading under 30 degrees of 130 ±10 N, 2 Hz, 1 ×106cycles. Final detorque was measured and vertical misfit using a stereomicroscope. Stereomicroscope Torque gauge Electromechanical equipment for mastication fatigue All groups presented a significant decrease in torque before and after mechanical cycling. The Morse taper connection promoted the highest torque maintenance. Mechanical cycling reduced the vertical misfit of all groups.
Int. J. Environ. Res. Public Health 2020,17, 8685 7 of 20 Table 1. Cont. Author Year Type of Study Purpose Type of Connection Number of Implants Torque Used Intervention Evaluation Outcome of the Study Butignon LE et al. [39] 2013 In Vitro Evaluate the effectiveness of 3 types of abutments in the maintenance of screw joint preload before and after cyclic loading and observe possible microdamage in the structure. EXTERNAL HEXAGON (Neodent) (n=45) 1. machined titanium (Ti) abutment 2. premachined gold (Au] abutments 3. machined zirconia (ZrO2) abutments 20 Ncm; 32 Ncm Static vending test using five specimens of each group was conducted to determine load applied in the cyclic loading test. Ten specimens of each group measured the RTV. A cyclic loading (0.5 ×106cycles, 15 Hz) between 11 and 211 N, angle 30◦, was applied. Postload RTV was measured after cyclic loading. SEM was used to detect possible microdamage in the structure of components. Torque gauge Fatigue test machine SEM The load application reduced the preload means significantly in all groups. SEM images showed evident structural changes in the mating surface of abutments, which are related to the loss of retained preload in all groups. Gil FJ et al. [29] 2014 In Vitro Evaluate microgap size and fatigue behavior of external and internal connections. INTERNAL HEXAGON EXTERNAL HEXAGON (Klockner SK2 and Essential) (n=100) 45 Ncm Each specimen was sectioned along the longitudinal axis in a total of three slides for SEM.The aim was to find the level of stress at which the sample supported 5 million cycles at 1450 N (15 Hz), which will be considered the fatigue limit. SEM Servo-hydraulic testing machine Internal connection presented lower microgap. Very good adaptation between the implant and the screw-retained abutment, preventing infiltration of microorganisms. The fatigue behavior of the external hexagon interface showed superior results compared to internal hexagon interfaces due to better load distribution. Shin HM et al. [40] 2014 In Vitro Evaluate the influence of the implant–abutment connection design and diameter on the screw joint stability EXTERNAL HEXAGON INTERNAL CONE (n=35) 30 Ncm The initial and postload RTV were measured after 100,000 cycles of a 150 N and 10 Hz cyclic load. The rates of the initial and postload removal torque losses were calculated to evaluate the effect of the joint connection design and diameter on the screw joint stability. Digital torque gauge External butt joint was more advantageous than the internal cone in terms of the postload removal torque loss. A wide diameter was more advantageous in terms of the torque loss rate. Bernardes SR et al. [20] 2014 In Vitro Determine whether abutment screw tightening and untightening influenced loss of preload in 3 different implant/abutment interfaces or on the implant body EXTERNAL HEXAGON INTERNAL HEXAGON INTERNAL CONICAL (Neodent) (n=25) 20 Ncm 32 Ncm Each sample was submitted to five tightening/untightening sequences, with a pause of 5 min after tightening before the screw was loosened. This resulted in 25 tightening/untightening sequences of each design. Custom benchtop screw-tightening machine (Prosthetic Dept, UCL, Eastman Dental Institute) There was no immediate significant loss of preload after screw tightening. Tightening/untightening sequences did not result in any significant loss of preload. Conical implant connections demonstrated greater structural reinforcement within the internal connection. Xia D et al. [41] 2014 In Vitro Evaluate the dynamic fatigue performance of implant–abutment assemblies with different tightening torque values. INTERNAL CONNECTION (Zimmer Biomet Dental) (n=30) Three tightening groups: 24, 30, 36 Ncm. Five specimens of each group were unscrewed and RTV recorded. Another five specimens were subjected to load between 30 and 300 N, 15 Hz for 5 ×106cycles. RTV postfatigue was recorded if available. Surfaces of specimens were observed with SEM. Digital torque meter Fatigue testing machine SEM The specimens that went through fatigue loading had decreased RTV. Insufficient torque will lead to poor fatigue performance of dental-abutment assemblies. Screws should be tightened to the torque recommended by the manufacturer.
Int. J. Environ. Res. Public Health 2020,17, 8685 8 of 20 Table 1. Cont. Author Year Type of Study Purpose Type of Connection Number of Implants Torque Used Intervention Evaluation Outcome of the Study Al-Otabi HN et al. [42] 2017 In Vitro Examine the effect of different torque application techniques on the removal torque of implant-supported fixed complete dental prostheses. INTERNAL CONNECTION (Nobel Biocare-Replace) (n=4) 35 Ncm The torque experiment consisted of three protocols: 1. Torqueing screws to 35 Ncm once. 2. Torqueing the screws to 35 Ncm and then immediately retorquing the same screws to the same value. 3. Torqueing the same screws to 35 Ncm three consecutive times. Removal torque was recorded. Digital torque meter Retightening abutment screws once after the initial torquing could enhance the removal torque of the screw. Care must be taken when retorquing the screws more than once because this may inversely affect the removal torque. Tsuruta K et al. [43] 2018 In Vitro Evaluate three kinds of connection systems from the point of view of microleakage from the gap between the implant and the abutment. EXTERNAL CONNECTION INTERNAL CONICAL CONNECTION INTERNAL PARALLEL CONNECTION (Nobel Biocare)(n=21) 35 Ncm Dye leakage was observed from the abutment screw hole to the outside through the microgap under the excessive compressive and tensile load. Each cycle, one compressive load and one tensile load (10 N each), was applied per 1 s and 2000 cycles of loading were performed. Every 500 cycles, the amount of microleakage was statistically compared. After the completion of 2000-cycle loading, RTV of the abutment screw was measured. Torque wrench (Nobel Biocare) Universal test machine Spectrophotometer The amount of microleakage from implant–abutment interface was smaller in conical connection than in internal parallel connection. The increase in the amount of microleakage was observed in all three groups. Removal toque of abutment screw after the cyclic loading showed no statistically significant difference among the groups. Al-Otaibi HN et al. [44] 2018 In Vitro Examine the effect of different maintenance time of torque application on detorque values of implant abutment screw. INTERNAL HEXAGON (n=4) 35 Ncm The abutment screws were subjected to different maintenance time of torque application Protocol A: 35 Ncm “instant” torque application. Protocol B: 35 Ncm torque maintained for 10 s. Protocol C: 35 Ncm torque maintained for 30 s. The procedure was repeated for each protocol five times, in which new sets of screws were used, with a total of 60 new screws. Digital torque meter The application of 35 Ncm for different maintenance times of torque application did not appear to affect the detorque value. Maintaining the torque for a prolonged time (10 s or 30 s) was not significantly associated with a higher preload than instant torque application.
Int. J. Environ. Res. Public Health 2020,17, 8685 9 of 20 Table 1. Cont. Author Year Type of Study Purpose Type of Connection Number of Implants Torque Used Intervention Evaluation Outcome of the Study Arshad M et al. [45] 2018 In Vitro Evaluate the effect of repeated screw joint closing and opening cycles and cycling loading on abutment screw removal torque and screw thread morphology. INTERNAL CONICAL CONNECTION (Dentium) (n=30) Three groups 12 and 30 Ncm Abutments screw were tightened (12 Ncm), removed, and RTV-recorded. This sequence was repeated five times for G1 and 15 times for G2 and G3. The same screws for G1 and G2 and new screws for G3 were tightened to 12 Ncm; this was also followed by screw tightening to 30 Ncm and retightening to 30 Ncm, 15 min later. RTVs were taken after cyclic loading 0.5 ×106, 1 Hz 75 N. Surface topography of one screw in each group was evaluated with SEM and compared with an unused screw. Electronic torque meter Chewing simulator SD mechatronic SEM Using a new screw did not significantly increase the value of removal torque. Restricting the amount of screw tightening was more important than replacing the screw. Torque loss values after loading were not shown to be significantly different from each other. He Y et al. [46] 2019 In Vitro Investigate the formation of microgaps and the change in the contact area at the implant–abutment interface of two different connection designs under oblique cyclic loading. CONICAL CONNECTION EXTERNAL HEXAGON CONNECTION (n=10)20 Ncm After loading, the samples were scanned using micro-CT, with silver nitrate as a high contrast penetrant, and the level of leakage was assessed.Three-dimensional finite element (FE) analyses were conducted to reveal the microgap formation process. Micro CT with silver nitrate Fatigue machine The conical connection showed more resistance against the formation of microgaps at the implant–abutment interface than the external hexagonal connection, although the minimum load required to bridge the internal implant space was within the range of human biting force. Kim KS et al. [47] 2020 In Vitro Examine the settling of abutments into implants and the removal torque value under static loading. EXTERNAL CONNECTION INTERNAL HEXAGON INTERNAL OCTAGON (n=50) 30 Ncm Ten implant–abutment assemblies were loaded vertically downward with a 700 N load cell at a displacement rate of 1 mm/min. The settling of the abutment was obtained from the change in the total length of the entire implant–abutment unit using an electronic digital micrometer. The postloading RTV was compared to the initial torque value. Universal testing machine electronic digital micrometer Digital torque gauge The loss of the preload due to the settling effect can lead to screw loosening during a clinical procedure in the molar region, where masticatory force is relatively greater.
Int. J. Environ. Res. Public Health 2020,17, 8685 16 of 20 propagation), whereas implant assemblies in the recommended and over-tightened torque groups (30 and 36 Ncm, respectively) had intact implant–abutment interfaces, as proven by SEM. The surface topography of one screw in each group, before and after cyclic loading (0.5 × 10 6 ), was evaluated by Arshad et al. [ 45 ] and compared with an unused screw. SEM analysis after loading displayed destruction of the thread abutment screw surface (desquamation and destruction of the superficial layer). In general, it could also be seen that even a precisely machined new screw was not highly smoothed. A single study contradicts all of these findings: Murmura et al. [ 37 ] used SEM and demonstrated the absence of gaps or mechanical deformations at the stump’s closing edge on its implant after the application of the cyclic load (1 ×106) in internal hexagon and internal octagon connections. 4.3. Evaluate the Sealing Capability of Different Implant Connections against Microleakage In the systematic review of Mishra et al. (2017) [ 53 ], a maximum study showed that there was some amount of microleakage at the abutment implant interface. External hexagon implants failed to completely prevent microleakage in both static and dynamic loading conditions of implants. Internal hexagon implants, particularly internal conical (Morse taper) implants, are highly promising in the case of static loading and showed less microleakage in dynamic loading conditions. Torque values recommended by the manufacturer should be strictly followed to achieve a better seal at the abutment–implant interface. Zirconia abutments are more prone to microleakage than titanium abutments, and their use should be discouraged. Zirconia abutments should only be restricted to cases where there is a high demand for aesthetics. These results corroborate the study of He et al. [ 46 ] (2019) in which the conical connection showed more resistance against the formation of microgaps at the implant–abutment interface than the external hexagonal connection. Additionally, Gil et al. [ 28 ] concluded that internal connections had a smaller microgap than external connections, with significant statistical differences. Very good adaptation between the implant and the screw-retained abutment was observed; in many cases, the distances were smaller than the bacteria diameter, thus preventing infiltration of microorganisms. In contrast, Ricomini Filho et al. [ 34 ] observed a better bacterial seal in the group with an external hexagon with a universal post than in groups with conical connections. These authors found that the external hexagon connection could have acted as a physical barrier, blocking bacterial penetration toward the inner part of the implant. SEM micrographs show no bacterial cells on the surface of the external hexagon abutment screw, thus confirming the microbiological assay. The methodology of rubbing a paper point on the inner part of the implant was probably unable to assess the microbial colonization on the implant platform, justifying the need for future studies to confirm these findings. In vitro investigations showed that a major portion of conical connection systems presents a microgap under static forces smaller than 10 µ m [ 63 ], demonstrating a better fit of the abutment into the fixture but not eliminating it completely. Other authors have shown minimal abutment movement and microgap formation under axial and oblique forces but good resistance to torque loss and screw loosening [ 64 ]. Internal cone implants have interface force transfer characteristics similar to those of a one-piece implant, but an absolute bacterial seal cannot be achieved in a two-piece implant system. For these reasons, conical abutment should be preferred to other connection systems to minimize bacterial microleakage [ 65 ]. Corroborating these findings, Gherlone et al. [ 66 ] tested, in an in-vitro study, a new internal conical connection design characterized by a double-taper principle. The authors evaluated and compared a new connection design, named double-action tight (DAT), with other internal connections. To investigate bacterial microleakage, the inner part of each system was inoculated with an Escherichia coli suspension. They found that in the DAT connection group, 7 of 10 total implants showed no bacterial infiltration at 96 h. This new internal conical design should reduce bacterial infiltration by constructing a physically tight connection with a high level of precision in the submicrometer range. Additional studies are necessary to better understand the stability of
Int. J. Environ. Res. Public Health 2020,17, 8685 17 of 20 this new type of internal connection over a longer period, with different bacteria and subject to the mastication function. 5. Conclusions This review found that different studies have been performed using a variety of approaches, thus often making the studies difficult to compare. As a result, it is difficult to draw conclusions about which abutment system behavior is optimal. Considering the proposed objectives, we can draw the following conclusions: 1. Maintenance of the preload: Internal connections have a higher preload value than that of the external hexagon design. The conical configuration can spread the load along the fixture and the surrounding bone more homogeneously than both the external hexagon and traditional internal connections. 2. Assessment of possible deformations at different interfaces after repeated application of tightening torque: Damage was observed in the threads of the abutment screws, before and after loading, in external and internal implant–abutment connections. 3. Evaluation of the sealing capability of different implant connections against microleakage: All connections presented some microgaps and bacterial microleakage. However, the performance of the conical connection systems appeared to be superior to that of other systems. Further in-vivo prospective studies are needed to build evidence of the best-performing connection over the long-term while bearing in mind the other factors that can affect clinical results. Author Contributions: Conceptualization, A.S.V., C.A. and M.H.-C.; methodology, A.S.V. and C.A.; software, F.S.; validation, F.S. and J.V.R.-S.; formal analysis, P.L.-J. and F.S.; investigation, A.S.V., C.A. and P.L.-J.; resources, A.S.V. and C.A.; data curation, F.S. and P.L.-J.; writing—original draft preparation, A.S.V., C.A. and F.S.; writing—review and editing, J.V.R.-S. and M.H.-C.; visualization, A.S.V., C.A. and M.H.-C.; supervision, M.H.-C. and J.V.R.-S. All authors have read and agreed to the published version of the manuscript. Funding: This research received no external funding. Conflicts of Interest: The authors declare no conflict of interest. References 1. Goiato, M.C.; Pellizzer, E.P.; da Silva, E.V.; Bonatto Lda, R.; dos Santos, D.M. Is the internal connection more efficient than external connection in mechanical, biological, and esthetical point of views? A systematic review. Oral Maxillofac. Surg. 2015,19, 229–242. [PubMed] 2. Siamos, G.; Winkler, S.; Boberick, K.G. Relationship between implant preload and screw loosening on implant-supported prostheses. J. Oral Implantol. 2002,28, 67–73. [PubMed] 3. Branemark, P.I.; Hansson, B.O.; Adell, R.; Breine, U.; Lindstrom, J.; Hallen, O.; Ohman, A. Osseointegrated implants in the treatment of the edentulous jaw. Experience from a 10-year period. Scand. J. Plast. Reconstr. Surg. Suppl. 1977,16, 1–132. 4. Adell, R.; Eriksson, B.; Lekholm, U.; Branemark, P.I.; Jemt, T. Long-term follow-up study of osseointegrated implants in the treatment of totally edentulous jaws. Int. J. Oral Maxillofac. Implants 1990 ,5, 347–359. [PubMed] 5. Becker, W.; Becker, B.E. Replacement of maxillary and mandibular molars with single endosseous implant restorations: A retrospective study. J Prosthet. Dent. 1995,74, 51–55. 6. Jemt, T.; Laney, W.R.; Harris, D.; Henry, P.J.; Krogh, P.H., Jr.; Polizzi, G.; Zarb, G.A.; Herrmann, I. Osseointegrated implants for single tooth replacement: A 1-year report from amulticenter prospective study. Int. J. Oral. Maxillofac. Implants 1991,6, 29–36. 7. Pardal-Pelaez, B.; Montero, J. Preload loss of abutment screws after dynamic fatigue in single implant-supported restorations. A systematic review. J. Clin. Exp. Dent. 2017,9, e1355–e1361. 8. Binon, P. The external hexagonal interface and screw joint stability: A primer on threaded fasteners in implants dentistry. Quintessence Dent. Technol. 2000,23, 91–105.
Int. J. Environ. Res. Public Health 2020,17, 8685 18 of 20 9. Cranin, N.A.; Klein, M.; Simons, A. Root form implant surgery abutments. In Atlas of Oral Implantology; Thieme Publishing Group: New York, NY, USA, 1993. 10. Binon, P.P. Implants and components: Entering the new millennium. Int. J. Oral Maxillofac. Implants 2000 ,15, 76–94. 11. Finger, I.M.; Castellon, P.; Block, M.; Elian, N. The evolution of external and internal implant/abutment connections. Pract. Proced. Aesthet. Dent. 2003,15, 625–632. 12. Merz, B.R.; Hunenbart, S.; Belser, U.C. Mechanics of the implant-abutment connection: An 8-degree taper compared to a butt joint connection. Int. J. Oral. Maxillofac. Implants 2000,15, 519–526. [PubMed] 13. Norton, M.R. An in vitro evaluation of the strength of an internal conical interface compared to a butt joint interface in implant design. Clin. Oral. Implants Res. 1997,8, 290–298. [PubMed] 14. Sutter, R.; Weber, H.P.; Sorensen, J.; Belser, U. The new restorative concept of the ITI dental implant system: Design and engineering. Int. J. Periodontics Restorative Dent. 1993,13, 409–431. 15. Schwarz, F.; Hegewald, A.; Becker, J. Impact of implant-abutment connection and positioning of the machined collar/microgap on crestal bone level changes: A systematic review. Clin. Oral. Implants Res. 2014 ,25, 417–425. 16. Park, J.K.; Choi, J.U.; Jeon, Y.C.; Choi, K.S.; Jeong, C.M. Effects of abutment screw coating on implant preload. J Prosthodont. 2010,19, 458–464. 17. Albrektsson, T. A multicenter report on osseointegrated oral implants. J. Prosthet. Dent. 1988,60, 75–84. 18. Oh, T.J.; Yoon, J.; Misch, C.E.; Wang, H.L. The causes of early implant bone loss: Myth or science? J. Periodontol. 2002,73, 322–333. 19. Winkler, S.; Ring, K.; Ring, J.D.; Boberick, K.G. Implant screw mechanics and the settling effect: Overview. J. Oral Implantol. 2003,29, 242–245. 20. Bernardes, S.R.; da Gloria Chiarello de Mattos, M.; Hobkirk, J.; Ribeiro, R.F. Loss of preload in screwed implant joints as a function of time and tightening/untightening sequences. Int. J. Oral. Maxillofac. Implants 2014,29, 89–96. 21. Piermatti, J.; Yousef, H.; Luke, A.; Mahevich, R.; Weiner, S. An in vitro analysis of implant screw torque loss with external hex and internal connection implant systems. Implant Dent. 2006,15, 427–435. 22. Jaarda, M.J.; Razzoog, M.E.; Gratton, D.G. Geometric comparison of five interchangeable implant prosthetic retaining screws. J. Prosthet. Dent. 1995,74, 373–379. [PubMed] 23. Papaspyridakos, P.; Chen, C.J.; Chuang, S.K.; Weber, H.P.; Gallucci, G.O. A systematic review of biologic and technical complications with fixed implant rehabilitations for edentulous patients. Int. J. Oral. Maxillofac. Implants 2012,27, 102–110. 24. Kazemi, M.; Rohanian, A.; Monzavi, A.; Nazari, M.S. Evaluation of the accuracy and related factors of the mechanical torque-limiting device for dental implants. J. Dent. 2013,10, 112–118. 25. Bulaqi, H.A.; Mousavi Mashhadi, M.; Safari, H.; Samandari, M.M.; Geramipanah, F. Dynamic nature of abutment screw retightening: Finite element study of the effect of retightening on the settling effect. J. Prosthet. Dent. 2015,113, 412–419. 26. McGlumphy, E.A.; Mendel, D.A.; Holloway, J.A. Implant screw mechanics. Dent. Clin. North Am. 1998 ,42, 71–89. [PubMed] 27. Burguete, R.L.; Johns, R.B.; King, T.; Patterson, E.A. Tightening characteristics for screwed joints in osseointegrated dental implants. J. Prosthet Dent. 1994,71, 592–599. [PubMed] 28. Tallarico, M.; Fiorellini, J.; Nakajima, Y.; Omori, Y.; Takahisa, I.; Canullo, L. Mechanical outcomes, microleakage, and marginal accuracy at the Implant-abutment interface of original versus nonoriginal implant abutments: A Systematic REVIEW of in vitro studies. Biomed. Res. Int. 2018. [CrossRef] 29. Gil, F.J.; Herrero-Climent, M.; Lazaro, P.; Rios, J.V. Implant-abutment connections: Influence of the design on the microgap and their fatigue and fracture behavior of dental implants. J. Mater Sci. Mater. Med. 2014 ,25, 1825–1830. 30. Cooper, L.F.; Tarnow, D.; Froum, S.; Moriarty, J.; De Kok, I.J. Comparison of marginal bone changes with internal conus and external hexagon design implant systems: A prospective, randomized study. Int. J. Periodont. Restor. Dent. 2016,36, 631–642. 31. Vigolo, P.; Gracis, S.; Carboncini, F.; Mutinelli, S.; Group ACR. Internal vs. external-connection single implants: A retrospective study in an Italian population treated by certified prosthodontists. Int. J. Oral Maxillofac. Implants 2016,31, 1385–1396.
Int. J. Environ. Res. Public Health 2020,17, 8685 19 of 20 32. Khraisat, A.; Hashimoto, A.; Nomura, S.; Miyakawa, O. Effect of lateral cyclic loading on abutment screw loosening of an external hexagon implant system. J. Prosthet. Dent. 2004,91, 326–334. [PubMed] 33. Tsuge, T.; Hagiwara, Y. Influence of lateral-oblique cyclic loading on abutment screw loosening of internal and external hexagon implants. Dent. Mater. J. 2009,28, 373–381. [PubMed] 34. Ricomini Filho, A.P.; Fernandes, F.S.; Straioto, F.G.; da Silva, W.J.; Del Bel Cury, A.A. Preload loss and bacterial penetration on different implant-abutment connection systems. Braz. Dent. J. 2010,21, 123–129. 35. Cashman, P.M.; Schneider, R.L.; Schneider, G.B.; Stanford, C.M.; Clancy, J.M.; Qian, F. In vitro analysis of post-fatigue reverse-torque values at the dental abutment/implant interface for a unitarian abutment design. J. Prosthodont. 2011,20, 503–509. [PubMed] 36. Ferreira, M.B.; Delben, J.A.; Barao, V.A.; Faverani, L.P.; Dos Santos, P.H.; Assuncao, W.G. Evaluation of torque maintenance of abutment and cylinder screws with Morse taper implants. J. Craniofac. Surg. 2012 ,23, e631–e634. 37. Murmura, G.; Di Iorio, D.; Cicchetti, A.R.; Sinjari, B.; Caputi, S. In vitro analysis of resistance to cyclic load and preload distribution of two implant/abutment screwed connections. J. Oral. Implantol. 2013 ,39, 293–301. 38. Jorge, J.R.; Barao, V.A.; Delben, J.A.; Assuncao, W.G. The role of implant/abutment system on torque maintenance of retention screws and vertical misfit of implant-supported crowns before and after mechanical cycling. Int. J. Oral Maxillofac. Implants 2013,28, 415–422. 39. Butignon, L.E.; Basilio Mde, A.; Pereira Rde, P.; Arioli Filho, J.N. Influence of three types of abutments on preload values before and after cyclic loading with structural analysis by scanning electron microscopy. Int. J. Oral Maxillofac. Implants 2013,28, e161–e170. 40. Shin, H.M.; Huh, J.B.; Yun, M.J.; Jeon, Y.C.; Chang, B.M.; Jeong, C.M. Influence of the implant-abutment connection design and diameter on the screw joint stability. J. Adv. Prosthodont. 2014,6, 126–132. 41. Xia, D.; Lin, H.; Yuan, S.; Bai, W.; Zheng, G. Dynamic fatigue performance of implant-abutment assemblies with different tightening torque values. Biomed. Mater. Eng. 2014,24, 2143–2149. 42. Al-Otaibi, H.N.; Almutairi, A.; Alfarraj, J.; Algesadi, W. The effect of torque application technique on screw preload of implant-supported prostheses. Int. J. Oral Maxillofac. Implants 2017,32, 259–263. [PubMed] 43. Tsuruta, K.; Ayukawa, Y.; Matsuzaki, T.; Kihara, M.; Koyano, K. The influence of implant-abutment connection on the screw loosening and microleakage. Int. J. Implant Dent. 2018,4, 11. [PubMed] 44. Al-Otaibi, H.N.; Al-Fouzan, A.F.; Al-Mufleh, T.S.; Labban, N. Effect of different maintenance time of torque application on detorque values of abutment screws in full-arch implant-supported fixed prostheses. Clin. Implant Dent. Relat. Res. 2018,20, 848–851. [PubMed] 45. Arshad, M.; Mahgoli, H.; Payaminia, L. Effect of repeated screw joint closing and opening cycles and cyclic loading on abutment screw removal Torque and screw thread morphology: Scanning electron microscopy evaluation. Int. J. Oral Maxillofac. Implants 2018,33, 31–40. [PubMed] 46. He, Y.; Fok, A.; Aparicio, C.; Teng, W. Contact analysis of gap formation at dental implant-abutment interface under oblique loading: A numerical-experimental study. Clin. Implant Dent. Relat. Res. 2019 ,21, 741–752. [PubMed] 47. Kim, K.S.; Lim, Y.J. Axial displacements and removal torque changes of five different implant-abutment connections under static vertical loading. Materials Basel 2020,13, 699. [CrossRef] 48. Jorneus, L.; Jemt, T.; Carlsson, L. Loads and designs of screw joints for single crowns supported by osseointegrated implants. Int. J. Oral Maxillofac. Implants 1992,7, 353–359. 49. Goheen, K.L.; Vermilyea, S.G.; Vossoughi, J.; Agar, J.R. Torque generated by handheld screwdrivers and mechanical torquing devices for osseointegrated implants. Int. J. Oral Maxillofac. Implants 1994,9, 149–155. 50. Kim, K.S.; Han, J.S.; Lim, Y.J. Settling of abutments into implants and changes in removal torque in five different implant-abutment connections. Part 1: Cyclic loading. Int. J. Oral. Maxillofac. Implants 2014 ,29, 1079–1084. 51. Do Nascimento, C.; Miani, P.K.; Pedrazzi, V.; Goncalves, R.B.; Ribeiro, R.F.; Faria, A.C.; Macedo, A.P.; de Albuquerque, R.F., Jr. Leakage of saliva through the implant-abutment interface: In vitro evaluation of three different implant connections under unloaded and loaded conditions. Int. J. Oral. Maxillofac. Implants 2012,27, 551–560. 52. Tripodi, D.; Vantaggiato, G.; Scarano, A.; Perrotti, V.; Piattelli, A.; Izzi, G.; D’Ercole, S. An in vitro investigation concerning the bacterial leakage at implants with internal hexagon and Morse taper implant-abutment connections. Implant Dent. 2012,21, 335–339. [PubMed]
Int. J. Environ. Res. Public Health 2020,17, 8685 20 of 20 53. Mishra, S.K.; Chowdhary, R.; Kumari, S. Microleakage at the different implant abutment interface: A systematic review. J Clin. Diagn. Res. 2017,11, ZE10–ZE15. [PubMed] 54. Breeding, L.C.; Dixon, D.L.; Nelson, E.W.; Tietge, J.D. Torque required to loosen single-tooth implant abutment screws before and after simulated function. Int. J. Prosthodont. 1993,6, 435–439. 55. Bozkaya, D.; Muftu, S. Mechanics of the tapered interference fit in dental implants. J. Biomech. 2003 ,36, 1649–1658. 56. Byrne, D.; Jacobs, S.; O’Connell, B.; Houston, F.; Claffey, N. Preloads generated with repeated tightening in three types of screws used in dental implant assemblies. J. Prosthodont. 2006,15, 164–171. 57. Delben, J.A.; Gomes, E.A.; Barao, V.A.; Assuncao, W.G. Evaluation of the effect of retightening and mechanical cycling on preload maintenance of retention screws. Int. J. Oral. Maxillofac. Implants 2011 ,26, 251–256. [PubMed] 58. Saboury, A.; Neshandar Asli, H.; Vaziri, S. The effect of repeated torque in small diameter implants with machined and premachined abutments. Clin. Implant Dent. Relat. Res. 2012,14 (Suppl. 1), e224–e230. 59. Weiss, E.I.; Kozak, D.; Gross, M.D. Effect of repeated closures on opening torque values in seven abutment-implant systems. J. Prosthet Dent. 2000,84, 194–199. 60. Pjetursson, B.E.; Tan, K.; Lang, N.P.; Bragger, U.; Egger, M.; Zwahlen, M. A systematic review of the survival and complication rates of fixed partial dentures (FPDs] after an observation period of at least 5 years. Clin. Oral. Implants Res. 2004,15, 625–642. 61. Steinebrunner, L.; Wolfart, S.; Ludwig, K.; Kern, M. Implant-abutment interface design affects fatigue and fracture strength of implants. Clin. Oral Implants Res. 2008,19, 1276–1284. 62. Jaarda, M.J.; Razzoog, M.E.; Gratton, D.G. Effect of preload torque on the ultimate tensile strength of implant prosthetic retaining screws. Implant Dent. 1994,3, 17–21. [PubMed] 63. Schmitt, C.M.; Nogueira-Filho, G.; Tenenbaum, H.C.; Lai, J.Y.; Brito, C.; Doring, H.; Nonhoff, J. Performance of conical abutment (Morse Taper] connection implants: A systematic review. J. Biomed. Mater. Res. A 2014 , 102, 552–574. [PubMed] 64. Coppede, A.R.; Bersani, E.; de Mattos Mda, G.; Rodrigues, R.C.; Sartori, I.A.; Ribeiro, R.F. Fracture resistance of the implant-abutment connection in implants with internal hex and internal conical connections under oblique compressive loading: An in vitro study. Int. J. Prosthodont. 2009,22, 283–286. [PubMed] 65. Baj, A.; Bolzoni, A.; Russillo, A.; Lauritano, D.; Palmieri, A.; Cura, F.; Silvestre, F.J.; Gianni, A.B. Cone-morse implant connection system significantly reduces bacterial leakage between implant and abutment: An in vitro study. J. Biol. Regul. Homeost. Agents 2017,31 (Suppl. 1), 203–208. 66. Gherlone, E.F.; Cappare, P.; Pasciuta, R.; Grusovin, M.G.; Mancini, N.; Burioni, R. Evaluation of resistance against bacterial microleakage of a new conical implant-abutment connection versus conventional connections: An in vitro study. New Microbiol. 2016,39, 49–56. Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations. © 2020 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).