Microdosimetry applied to proton radiotherapy Alejandro Bertolet Reina Facultad de F´ ısica Universidad de Sevilla Director: Miguel Antonio Cort´ es-Giraldo, PhD Departamento de F´ ısica At´ omica, Molecular y Nuclear Universidad de Sevilla Co-Director: Alejandro Carabe-Fern´ andez, PhD Departament of Radiation Oncology University of Pennsylvania A thesis submitted for the degree of Doctor of Philosophy Sevilla, 2020
2
Acknowledgements This thesis represents the fruit of not only two years of hard work, but more than a lustrum of permanent development and accomplishments, both personal and scientific. I would like to acknowledge and thank the opportunity given by Dr. Alejandro Carabe, one of the directors of this thesis, to join his lab. The learning, knowledge and experiences I have gone through during this time in Philadelphia are countless. Also, the support, guidance and advice from him and Dr. Miguel Cort´ es, the other director of this thesis, have definitely and continuously contributed to make me a better scientist and a better person. I feel fortunate and lucky to have accomplished a PhD that has been joy every step of the way. My sincere thanks to all the collaborators: Anna Baratto, Giorgio Baiocco, Sofia Barbieri, Consuelo Guardiola, Veljko Grilj, Andrew Harken, Kevin Souris and Marie Cohilis. Thanks to all members of Carabe’s lab at the University of Pennsylvania, involved in productive discussions improving all the work developed here as well as Dr. Eric Abel and Dr. Camille Noel from Varian for their interest in this project. I also thank all the Department of Radiation Oncology, and, particularly, of Medical Physics, from the Hospital Virgen Macarena in Sevilla. I grew up there as a professional and as a scientist and this thesis would have not been possible without them. Thank you to the Spanish crew in Philadelphia for the support and fun during this time and for being, essentially, a family abroad. And last but not the least, special recognition and thanks to my family and friends in Spain, whose constant contributions along years have been invaluable to make me turn into who I am. A mi madre, mi inspiraci´ on y la raz´ on de levantarme cada d´ ıa con la humilde ambici´ on de aportar, por poco que sea, a mejorar la vida de los enfermos con c ´ ancer. Te quiero.
Abstract Proton radiotherapy as a treatment for cancer is increasingly used worldwide. The main current clinical indications for proton therapy are based on the physical properties of proton beams, assuming a constant relative biological effectiveness (RBE) of 1.1 with respect to x-ray radiotherapy. However, there is growing evidence of variable RBE depending on the linear energy transfer (LET) of the proton beam, that, in turns, depends on the relative position along the proton track. Additionally, RBE depends on the considered endpoint and the biological properties of different cells and tissues as well as on the dose fractionation scheme. Microdosimetry is a theory that studies the patterns of energy deposition in microscopic sized volumes. By determining the local concentration of the energy imparted at such dimensions, it is possible to obtain LET and other quantities characterizing the quality of the beam from this theory. Eventually, the study of the distributions of energy imparted at this scale may lead to a better understanding of the variable RBE for proton beams, allowing for potential clinical application. This thesis is a collection of studies in that direction. As a whole, they are intended to make the connection between elementary description of proton-matter interaction in microdosimetric terms and the clinical application level. In this sense, Chapter 2studies the basics of microdosimetric quantities and their interrelations, and focuses on how to produce correct microdosimetric results from Monte Carlo (MC) simulations. Chapter 3uses the methodology explained in Chapter 2to generate monoenergetic distributions for protons of energies up to 100 MeV. These distributions are used to create analytical models to derive microdosimetric quantities in polyenergetic beams instead performing dedicated MC simulations for each individual case. In chapter 4, the results produced by these analytical models for monoenergetic beams are compared to actual measurements with silicon-based microdosimeters exposed to monoenergetic proton beams.
In order to calculate spatial distributions of microdosimetric quantities with the formalism and functions developed in Chapter 3, the determination of the beam spectra at each point, or spectral fluence, is required. Chapter 5provides an independent formalism to calculate spectral fluences in clinical proton beams as a function of depth and lateral position in water. This formalism, in conjunction with the microdosimetry models proposed in Chapter 3, allows for determination of 3D distributions of microdosimetric quantities. Chapter 6illustrates the combination of formalisms from Chapters 3 and 5to produce calculation of dose distributions in a treatment planning system (TPS). These results are compared to independent MC simulations with the code MCsquare and analytical calculations clinically validated from a commercial TPS, showing consistent results. This provided a partial validation of the microdosimetric models presented in Chapter 3. Microdosimetric calculations of dose-mean lineal energy (yD), and both unrestricted and restricted dose-averaged LET can also be obtained from the combination of the formalisms presented in Chapters 3and 5, as shown in Chapter 7. Additionally, the relations between LET and lineal energy in microdosimetry are revisited, and a new equation to calculate restricted LET is proposed and tested against the previously established one. Finally, in Chapter 8, the Microdosimetric Kinetic Model (MKM) is used to calculate RBE distributions based on the physical quantities derived in previous chapters. A method for the determination of the ideal site dimension for different cell lines is proposed and its impact on the RBE results is assessed. This thesis presents a comprehensive method to calculate microdosimetric distributions in patient-specific geometries from ground microdosimetric principles. Although the formulations here shown are specifically referred to protons, microdosimetric computations for other heavier particles may follow the same procedures and formalisms, which represents one of the most interesting directions of further development from this work. iv
Report of publications This PhD thesis is presented under the modality of compendium of publications. These are the full references and quality indexes for them: •Bertolet A., Baratto-Rold´ an A., Barbieri S., Baiocco G., Carabe A., Cort´ es- Giraldo M. A. Dose-averaged LET calculation for proton track segments using microdosimetric Monte Carlo simulations. Medical Physics, 2019, 46(9), 4184-4192. doi: 10.1002/mp.13643. Impact Factor: 3.317 (2019 Journal Citation Reports) Rankings: SJR: 37/308, Q1. Category: Radiology, Nuclear Medicine and Imaging Web of Science: 33/133, T1, Q1. Category: Radiology, Nuclear Medicine and Medical Imaging Cites (Web of Science): 5. •Bertolet A., Baratto-Rold´ an A., Cort´ es-Giraldo M. A., Carabe-Fernandez A. Segment-averaged LET concept and analytical calculation from microdosimetry quantities in proton radiation therapy. Medical Physics, 2019, 46(9), 4184-4192. doi: 10.1002/mp.13673. Impact Factor: 3.317 (2019 Journal Citation Reports) Rankings: SJR: 37/308, Q1. Category: Radiology, Nuclear Medicine and Imaging Web of Science: 33/133, T1, Q1. Category: Radiology, Nuclear Medicine and Medical Imaging Cites (Web of Science): 6. •Bertolet A., Grilj V., Guardiola C., Harken A. D., Baratto-Rold´ an A., Cort´ es- Giraldo M. A., Carabe-Fernandez A. Experimental validation of an analytical microdosimetric model based on Geant4-DNA simulations by using a siliconbased microdosimeter. Radiation Physics and Chemistry, 2020, First published on 17 June. doi: v
10.1016/j.radphyschem.2020.109060 Impact Factor: 2.226 (2019 Journal Citation Reports) Rankings: SJR: 17/52, Q2. Category: Radiation Web of Science: 4/34, T1, Q1. Category: Nuclear Science and Technology •Bertolet A., Cort´ es-Giraldo M. A., Souris, K., Carabe-Fernandez A. A kernelbased algorithm for the spectral fluence of clinical proton beams to calculate dose-averaged LET and other dosimetric quantities of interest. Medical Physics, 2020, 47(6), 2495-2505. doi: 10.1002/mp.14108 Impact Factor: 3.317 (2019 Journal Citation Reports) Rankings: SJR: 37/308, Q1. Category: Radiology, Nuclear Medicine and Imaging Web of Science: 33/133, T1, Q1. Category: Radiology, Nuclear Medicine and Medical Imaging Cites (Web of Science): 2. •Bertolet A., Cort´ es-Giraldo M. A., Souris, K., Cohilis M., Carabe-Fernandez A. Calculation of clinical dose distributions in proton therapy from microdosimetry. Medical Physics, 2019, 46(12), 5816-5823. doi: 10.1002/mp/13861. Impact Factor: 3.317 (2019 Journal Citation Reports) Rankings: SJR: 37/308, Q1. Category: Radiology, Nuclear Medicine and Imaging Web of Science: 33/133, T1, Q1. Category: Radiology, Nuclear Medicine and Medical Imaging Cites (Web of Science): 2. •Bertolet A., Cort´ es-Giraldo M. A., Carabe-Fernandez A. On the concepts of dose-mean lineal energy, unrestricted and restricted dose-averaged LET in proton therapy. Physics in Medicine and Biology, 2020, 65 075011. doi: 10.1088/1361- 6560/ab730a. Impact Factor: 2.883 (2019 Journal Citation Reports) Rankings: SJR: 42/308, Q1. Category: Radiology, Nuclear Medicine and Imaging Web of Science: 46/133, T1, Q2. Category: Radiology, Nuclear Medicine and Medical Imaging vi
Cites (Web of Science): 1. •Bertolet A., Cort´ es-Giraldo M. A., Carabe-Fernandez A. Analysis of a new method to determine domain sizes to calculate RBE in proton therapy with the microdosimetric kinetic model using analytical calculations of dose-mean lineal energy in a TPS. Medical Physics, 2020, SUBMITTED. The following articles have also been developed and published along the PhD program, not being included in this thesis but related to it: •Bertolet A., Carabe-Fernandez A. Clinical implications of variable relative biological effectiveness in proton therapy for prostate cancer. Acta Oncologica, 2020. doi: 10.1080/0284186X.2020.1762928 •Bertolet A., Carabe-Fernandez A. Proton Monoenergetic Arc Therapy (PMAT) to enhance LETd within the target. Physics in Medicine and Biology, 2020. doi: 10.1088/1361-6560/ab9455 •Bertolet A., Cort´ es-Giraldo M. A., Carabe-Fernandez A. An analytical microdosimetric model for radioimmunotherapeutic alpha emitters. Radiation Research, 2020, ACCEPTED. •Carabe A., Karagounis I. V., Huynh K., Bertolet A., Franc¸ois N., Kim M. M., Maity A., Abel E., Dale R. G. Radiobiological effectiveness in proton arc beams versus conventional proton and photon beams. Physics in Medicine and Biology, 2020. doi: 10.1088/1361-6560/ab9370 •Carabe-Fernandez A., Bertolet A., Karagounis I. V., Huynh K., Dale R. G. Is there a role for arcing techniques in proton therapy? British Journal of Radiology, 2020, 93, 20190469. vii
expressed using the CGS system of units, Sel =−dE dl=4πe4Z2 mev2Z2 1log 2mev2 I−log(1 −β2)−β2−C Z2−δ 2,(1.1) where eis the electronic charge, meis the electronic mass at rest, Z1and Z2are the atomic numbers of the ion (1 for protons) and the medium, respectively, vis the speed of the ion, β=v/c where cis the speed light, Iis the mean ionization potential of the medium, the C/Z2term corresponds to shell corrections and the δ/2term is the density effect; both correction terms are calculated numerically. Thus, as the proton is slowed down along its path, the energy deposition increases and so does dose deposition. This property, combined with the finite range that protons have when traversing any material, produces a distinct maximum on the depth-dose deposition curve, usually known as “Bragg peak”. Figure 1.2: Variation of the electronic stopping power (Sel) with depth for a proton with initial kinetic energy of 125 MeV traversing liquid water, using a spatial resolution of 1 mm, according to the data published in the PSTAR database (8). 1.1.2 Relative Biological Effectiveness (RBE) Along with the different physical behavior of proton beams, summarized in Figure 1.1, the ability to produce biological damage also differs from x-ray beams for the same dose imparted. The ratio between the dose needed to produce a given effect e with a given reference radiation, Dref , and with protons (or other test radiation), Dtest =Dproton is called relative biological effectiveness (RBE). Typically, the 3
reference radiation is chosen to be x-rays coming from 60Co sources or megavoltage and orthovoltage x-ray sources (9, 10), i.e., Dref =Dx−ray: RBE|e=Dref |e Dtest|e =Dx−ray|e Dproton|e .(1.2) The RBE of proton beams increases towards the end of the beam range (11), since the interactions between protons and matter become more concentrated as the proton kinetic energy decreases, as discussed previously. A useful quantity to characterize the concentration of interactions -and, eventually, of energy imparted in themis called linear energy transfer (LET). LET subtly differs from stopping power, and it is defined as the energy lost by the proton per unit path length due to electronic collisions with the material in which the energy exchange is lower than a set threshold (12). That energy threshold is denoted with a subscript ∆. Sometimes the threshold can be expressed in terms of the range of the particle. In that case, ∆is the maximum range considered and LET∆is called spatially restricted LET. If that threshold is set to infinite energy, then LET equals the electronic stopping power and it is called unrestricted LET. Furthermore, LET is a property well defined at a macroscopic scale for a particle with a given kinetic energy, but when a variety of particles with different kinetic energies is considered, the LET ceases to be well defined. In this sense, two possible averaging processes are traditionally used: (i) weighting the number of tracks with a given LET to yield the track-averaged LET (LETt); or (ii) weighting the dose imparted by each particle with a given LET to yield the dose-averaged LET (LETd). It has been shown how protons are capable to produce more damage as their LET (either LETt or LETd) increases (13, 14). These studies are based on clonogenic assays, i.e. experiments in which the fraction of surviving cells are determined after being exposed to radiation. This survival fraction Sis usually modeled by means of the Linear-Quadratic (LQ) model (15), which assumes a dependency on dose such as S= exp −αD −βD2,(1.3) where Dis the absorbed dose to which cells are exposed, and αand βare the parameters of the LQ model. When exposing cells to high LET radiation, a lower survival fraction is usually observed, which is translated into a generally higher value of the αparameter. However, the actual value for RBE depends on multiple factors besides LET. Namely: the fractionation scheme of the dose or the α/β ratio, which in turn depends on the cell line, its environment, the stage within the cell cycle and the considered endpoint, have major impacts on RBE (9). Therefore, to 4
avoid dealing with this variability among cells added to the intrinsic experimental uncertainty, a unique constant value of RBE = 1.1has been traditionally adopted for the clinical routine of proton radiotherapy. This value represents an average of all RBE reported (6), as shown in Figure 1.3. Figure 1.3: Collection of published data points of RBE for cell survival shown as a function of the (α/β)xratio of the cell/tissue type for a given endpoint, where α and βare the parameters of the linear-quadratic (LQ) model, used to represent the survival fraction of cells exposed to the reference radiation, i.e., photons from 60Co or MV sources. LETd is ≤15 keV/µm and two doses are shown: 2 Gy (left) and 6 Gy (right). Solid lines are fits through the data included in each plot considering the published uncertainties. Dashed lines show the fit results not considering individual uncertainties. Figure reproduced from Paganetti, 2014 (16). Nonetheless, the increasing amount of data showing a variable RBE for proton beams (17, 18, 19, 20) suggests the need for a more sophisticated way to predict this radiobiological effectiveness. Several phenomenological models for the relation between RBE and LET can be found in the literature (21, 22, 23, 24), as well as more detailed mechanistic models (25, 26, 27). Additionally, several studies 5
exploring the variable RBE have also been recently published to add efforts in seeking for correct biological characterization of proton beams (28, 29, 30, 31). In this work, we follow the approach suggested by other works (32, 33, 34), and address the variable biological effectiveness of proton beams problem by taking use of the microdosimetry theory. 1.2 Introduction to microdosimetry Microdosimetry theory was developed by Rossi and Kellerer along the decades of 1960’s and 1970’s in order to study the effects of ionizing radiation in biological systems (35, 36, 37). In particular, microdosimetry studies interactions between ionizing particles and matter at a microscopic scale, as the assumed critical biological structures within the cell (e.g. DNA) are of this dimension. These critical structures are represented by volumes (tradiationally spheres) of microscopic dimensions, known as sites (36). An alternative description of the microdosimetry theory is based on the concept of track structure, that requires the knowledge of the radial profile of interactions along the particle trajectory. In the site-based microdosimetry, a radiation event is considered to enclose all the energy depositions in a site caused by the primary particle as well as all the secondary particles generated after the primary one. The energy imparted to this microscopic site after a number νof events, ε, divided by its mass mis called specific energy (z=ε/m), which is a stochastic quantity susceptible to vary from event to event. Another convenient quantity is called lineal energy (y), whose dimensions are energy imparted per unit path length. If primary particles can be considered to travel in straight lines, then the intersections between their straight trajectory and the site are called chord lengths (l). Lineal energy is defined as the energy imparted in a single event (εs) to the site divided by the mean chord length (l) among all the possible straight trajectories intersecting the site, i.e. y=εs/l. At this scale, the stochastic nature of radiation-matter interactions play a significant role, and, therefore, distributions of specific energy or lineal energy must be accounted for rather than single average values, as illustrated in Figure 1.4. In this sense, (1) the macroscopic concept of dose arises as the average value for the microscopic distribution of specific energy; and (2) the macroscopic concept of spatially restricted LET∆for a given particle with kinetic energy well defined corresponds to the average value for the microscopic distribution of lineal energy, shown in Figure 1.4, being ∆given by the dimensions of the site. For a variety of different particles, re- 6
stricted LETt also coincides with the average lineal energy combining events from all particles, also called frequency-mean lineal energy yF, while restricted LETd is related to the dose-mean lineal energy, yD, the average obtained from weighting the lineal energy of each event by the dose, or, equivalently, by the energy imparted in that event. However, LETd∆≤yD, since yDincludes extra terms due to variability of chord length and range straggling, not included in LETd∆. Table 1.1 summarizes the correspondences between macroscopic concepts used in radiation therapy and their microdosimetric counterparts. Figure 1.4: Distributions of energy imparted per unit path length for protons of 950 keV traversing thin layers of tissue equivalent gas. The thinner the layer, the fewer the interactions and hence, the greater relative variance of the distribution. Thicker layers allow for a larger number of interactions and then the distributions become narrower. Figure taken from the Report 36 of ICRU (44). Kellerer and Rossi, in their formulation of the Theory of Dual Radiation Action (TDRA) (38, 39) proposed that lethal damage produced to cells by ionizing radiation might be mediated by pairs of sublesions besides direct action, and the yield of these two types of mechanisms can be characterized by the microscopic distributions of energy imparted. Therefore, as different types of particles produce different distributions, the way in which lethal lesions are produced also differs. Kellerer’s microdosimetric formulation (39, 40, 41, 42, 43) contributes to establish a relation between the effectiveness for a type of radiation and its associated LET. Although microdosimetry theory has been succesfully used to characterize ra- 7
Table 1.1: Summary of microdosimetric quantities and their correspondences to macroscopic concepts in radiation therapy, when applicable. Microdosimetric quantities are stochastic variables with a given distribution among events or set of events, whose different averages correspond to macroscopic quantities. Symbol Microscopic quantity Macroscopic quantity Relation εsEnergy imparted per single event ε=PνεsEnergy imparted after ν events z=ε/m Specific energy (dist f(z)) z=Rzf(z)dz Dose D=z lChord length (dist f(l)) l=Rlf(l)dl Lineal energy (dist. f(y)) Restricted LET (single particle/energy) LET∆=y y=εs/l y =yF=Ryf(y)dyRestricted track-averaged LET (multiple particles/energies) LETt∆=yF yD=Ry2f(y)dy Ryf(y)dyRestricted dose-averaged LET (multiple particles/energies) LETd ∆≤yD diation quality for radiation protection purposes (44), its application to radiation therapy is challenging. Firstly, experimental measurements are not generally possible for clinical purposes as (i) clinical beams are polyenergetic and the spectra need to be known; and (ii) traditional devices such as tissue-equivalent proportional counters (TEPCs) are too expensive and bulky (45). Secondly, analytical approaches are complex and need to be simplified up to a point not as accurate as required for therapy and Monte Carlo (MC) simulations at such scale are very demanding computationally (46). However, models to relate microdosimetric quantities to RBE have been developed, such as the Microdosimetric Kinetic Model (MKM) (47, 48, 49). Also, efforts have been put on the determination of microdosimetric distributions for clinical proton beams recently (34, 50), aiming at characterizing the variability of RBE described above. This thesis is a series of articles that comprehensively address the connection between microdosimetry theory and clinical protontherapy. 8
1.3 Purpose of this thesis This thesis is structured as follows. Chapters 2,3and 4explore microdosimetry theory for protons, looking into how to perform microdosimetric MC simulations, providing an analytical framework to compute microdosimetric quantities and verifying the results against experimental measurements, respectively. Independently, chapter 5introduces an also analytical formalism to calculate spectra for clinical proton beams in treatment planning systems (TPS). Once microdosimetry has been analytically modeled and spectral fluences for proton beams are calculated, Chapters 6and 7present results for calculation of dose, LETd and yDfor clinical proton beams. Finally, Chapter 8explores a way to transform the physical quantities calculated from all the previous formalism in biologically relevant results by the MKM. References [1] G. Delaney, S. Jacob, C. Featherstone, and M. Barton. The role of radiotherapy in cancer treatment: Estimating optimal utilization from a review of evidence-based clinical guidelines. Cancer, 104(6):1129–1137, 2005. ISSN 0008543X. doi: 10.1002/cncr.21324. [2] R. Baskar, K. A. Lee, R. Yeo, and K. W. Yeoh. Cancer and radiation therapy: Current advances and future directions. International Journal of Medical Sciences, 9(3):193–199, 2012. ISSN 14491907. doi: 10.7150/ijms.3635. [3] R. R. Wilson. Radiological use of fast protons. Radiology, 47(5):487–491, 1946. [4] X. Tian et al. The evolution of proton beam therapy: Current and future status (Review). Molecular and Clinical Oncology, pages 15–21, 2017. ISSN 2049-9450. doi: 10.3892/mco.2017.1499. [5] W. P. Levin, H. Kooy, J. S. Loeffler, and T. F. DeLaney. Proton beam therapy. British Journal of Cancer, 93(8):849–854, 2005. ISSN 00070920. doi: 10. 1038/sj.bjc.6602754. [6] H. Paganetti. Proton Therapy Physics. Taylor & Francis, 2012. ISBN 9781439836453. 9
[7] J. F. Ziegler. Stopping of energetic light ions in elemental matter. Journal of Applied Physics, 85(3):1249–1272, 1999. ISSN 00218979. doi: 10.1063/1. 369844. [8] M. Berger, J. Coursey, M. Zucker, and J. Chang. ESTAR, PSTAR and ASTAR: Computer Programs for Calculating Stopping-Power and Range Tables for Electrons, Protons, and Helium Ions (version 2.0.1), 2017. [9] H. Paganetti et al. Report of the AAPM TG-256 on the relative biological effectiveness of proton beams in radiation therapy. Medical Physics, 46(3):e53–e78, 2019. ISSN 00942405. doi: 10.1002/mp.13390. [10] IAEA and ICRU. Relative Biological Effectiveness in Ion Beam Therapy. IAEA Technical Reports, (461):1–165, 2008. doi: TechnicalReportsSeriesNo.461. [11] H. Paganetti et al. Relative biological effectiveness (RBE) values for proton beam therapy. International journal of radiation oncology, biology, physics, 53(2):407–21, jun 2002. ISSN 0360-3016. doi: 10.1016/S0360-3016(02) 02754-2. [12] ICRU. Report 90: Key Data for ionizing radiation dosimetry: measurement standards and applications. Technical Report 90, 2016. [13] G. W. Barendsen. The Relationships between RBE and LET for Different Types of Lethal Damage in Mammalian Cells: Biophysical and Molecular Mechanisms. Radiation Research, 139(3):257–270, 1994. ISSN 00337587. doi: 10.2307/3578823. [14] M. Belli et al. RBE-LET relationships for cell inactivation and mutation induced by low energy protons in V79 cells: Further results at the LNL facility. International Journal of Radiation Biology, 74(4):501–509, 1998. ISSN 09553002. doi: 10.1080/095530098141375. [15] S. J. McMahon. The linear quadratic model: Usage, interpretation and challenges. Physics in Medicine and Biology, 64(1):0–24, 2019. ISSN 13616560. doi: 10.1088/1361-6560/aaf26a. [16] H. Paganetti. Relative biological effectiveness (RBE) values for proton beam therapy. Variations as a function of biological endpoint, dose, and linear energy transfer. Physics in Medicine and Biology, 59(22):R419–R472, nov 2014. ISSN 0031-9155. doi: 10.1088/0031-9155/59/22/R419. 10
[17] Y. Matsumoto et al. Enhanced radiobiological effects at the distal end of a clinical proton beam: In vitro study. Journal of Radiation Research, 55(4): 816–822, 2014. ISSN 13499157. doi: 10.1093/jrr/rrt230. [18] R. Mohan et al. Radiobiological issues in proton therapy. Acta Oncologica, 56(11):1367–1373, 2017. ISSN 1651226X. doi: 10.1080/0284186X.2017. 1348621. [19] H. Willers et al. Toward A variable RBE for proton beam therapy. Radiotherapy and Oncology, 128(1):68–75, 2018. ISSN 18790887. doi: 10.1016/j.radonc. 2018.05.019. [20] T. Friedrich. Proton RBE dependence on dose in the setting of hypofractionation. The British Journal of Radiology, (March):20190291, 2019. ISSN 0007-1285. doi: 10.1259/bjr.20190291. [21] J. J. Wilkens and U. Oelfke. A phenomenological model for the relative biological effectiveness in therapeutic proton beams. Physics in Medicine and Biology, 49(13):2811–2825, 2004. ISSN 00319155. doi: 10.1088/0031-9155/ 49/13/004. [22] A. Carabe et al. Range uncertainty in proton therapy due to variable biological effectiveness. Physics in Medicine and Biology, 57(5):1159–1172, 2012. ISSN 00319155. doi: 10.1088/0031-9155/57/5/1159. [23] A. L. McNamara, J. Schuemann, and H. Paganetti. A phenomenological relative biological effectiveness (RBE) model for proton therapy based on all published in vitro cell survival data. Physics in Medicine and Biology, 60(21): 8399–8416, nov 2015. ISSN 0031-9155. doi: 10.1088/0031-9155/60/21/ 8399. [24] M. Wedenberg et al. A model for the relative biological effectiveness of protons: The tissue specific parameter α/βof photons is a predictor for the sensitivity to LET changes. Acta Oncologica, 52(3):580–588, 2013. ISSN 0284186X. doi: 10.3109/0284186X.2012.705892. [25] S. J. McMahon, J. Schuemann, H. Paganetti, and K. M. Prise. Mechanistic Modelling of DNA Repair and Cellular Survival Following Radiation-Induced DNA Damage. Scientific Reports, 6(August):1–14, 2016. ISSN 20452322. doi: 10.1038/srep33290. 11
[26] S. J. McMahon et al. A general mechanistic model enables predictions of the biological effectiveness of different qualities of radiation. Scientific Reports, 7 (1):1–14, 2017. ISSN 20452322. doi: 10.1038/s41598-017-10820-1. [27] R. Abolfath et al. Renormalization of radiobiological response functions by energy loss fluctuations and complexities in chromosome aberration induction: deactivation theory for proton therapy from cells to tumor control. European Physical Journal D, 73(3):1–22, 2019. ISSN 14346079. doi: 10.1140/epjd/e2019-90263-5. [28] H. Paganetti and D. Giantsoudi. Relative Biological Effectiveness Uncertainties and Implications for Beam Arrangements and Dose Constraints in Proton Therapy. Seminars in Radiation Oncology, 28(3):256–263, 2018. ISSN 15329461. doi: 10.1016/j.semradonc.2018.02.010. [29] J. Unkelbach and H. Paganetti. Robust Proton Treatment Planning: Physical and Biological Optimization. Seminars in Radiation Oncology, 28(2):88–96, 2018. ISSN 15329461. doi: 10.1016/j.semradonc.2017.11.005. [30] K. Takada et al. Validation of the physical and RBE-weighted dose estimator based on PHITS coupled with a microdosimetric kinetic model for proton therapy. Journal of Radiation Research, 59(1):91–99, jan 2018. ISSN 0449-3060. doi: 10.1093/jrr/rrx057. [31] D. S´ anchez-Parcerisa et al. MultiRBE: Treatment planning for protons with selective radiobiological effectiveness. Medical Physics, 46(9):4276–4284, 2019. ISSN 0094-2405. doi: 10.1002/mp.13718. [32] Y. Kase et al. Microdosimetric Measurements and Estimation of Human Cell Survival for Heavy-Ion Beams. Radiation Research, 166(4):629–638, oct 2006. ISSN 0033-7587. doi: 10.1667/RR0536.1. [33] Y. Kase et al. Microdosimetric calculation of relative biological effectiveness for design of therapeutic proton beams. Journal of Radiation Research, 54(3): 485–493, may 2013. ISSN 0449-3060. doi: 10.1093/jrr/rrs110. [34] M. Newpower et al. Using the Proton Energy Spectrum and Microdosimetry To Model Proton Relative Biological Effectiveness. International Journal of Radiation Oncology •Biology •Physics, 104(2):316–324, 2019. ISSN 0360- 3016. doi: 10.1016/j.ijrobp.2019.01.094. 12
two different distributions: its frequency distribution and its weighted distribution. Both are here expressed in terms of their probability density functions f(x) and d(x), respectively. Then, f(x)dx represents the probability of occurrence for a value in the interval [x, x +dx]of that stochastic quantity, thus f(x) = Nx,dx/N, where Nx,dx is the number of events with values within that interval and Nthe overall number of events. The weighted probability density d(x)is obtained by weighting each event by the value of the stochastic quantity. The averages of each distribution are called frequency average ¯xFand weighted average ¯xD, which are given by: ¯xF=Zxf(x) dx ¯xD=Zxd(x) dx=Rx2f(x) dx Rxf(x) dx (2.1) As seen in Eq. (2.1), the weighted average ¯xDof a stochastic quantity can be obtained from its frequency distribution. From (2.1), it is possible to relate both averages through: ¯xD= ¯xF+σ2 x ¯xF = ¯xF(1 + Vx)(2.2) where σ2 xand Vx=σ2 x/¯x2 Fare the variance and relative variance of the frequency distribution of x, respectively. It can be seen, therefore, that the weighted average is a combination of the mean value and the variance of the produced values of the stochastic quantity. 2.2.2 Obtaining LET from microdosimetry From a macroscopic point of view, the mean energy lost by a charged particle due to electronic interactions per length unit is defined as the (unrestricted) LET (23). The LET value for a particle depends on the energy of the particle. Additionally, a beam in radiation therapy is usually composed of particles with different energies, which implies different LET. Hence, the beam has an LET frequency distribution, given by the number of particles with a given LET. This distribution can also be weighted by the absorbed dose, as described above, resulting in the so-called dose-weighted LET distribution. Therefore, the LET of a beam can be characterized by its frequency average, also known as track average LET, ¯ LT, and more interestingly from a radiobiological point of view, the dose-averaged LET, ¯ LD. The reason for using ¯ LD, according to dual radiation action theory (24), is that biological damage is related to the amount of energy imparted (i.e., dose) and the spatial concentration of this energy imparted (i.e., LET), being ¯ LDsimultaneously 19
representative of both. Furthermore, clinically, we are not interested in the total energy loss of a particle but in the energy deposited “locally”. For those cases, the concept of restricted LET, L∆, can be used, where ∆is a cutoff kinetic energy for secondary electrons, above which no energy deposit is counted. This cutoff energy is ultimately related to the distance that secondary electrons can travel from the place they are produced, the core of the proton track. It is worth to point out that all the LET values considered herein are therefore restricted and will depend on the ∆value, which is determined in our case by the site dimension. Thereby the subscript ∆for the restricted LET will be omitted in the following for notation simplicity. This work is focused on analyzing LET in proton track segments, since this approach allows for studying the microscopic behavior of a clinical beam in a specific portion of its path in tissue, which might lead eventually to calculations of LET spatial distributions. The following discussion remains valid as long as the site dimension is small enough so that the LET of the proton can be considered constant along all its travel within the site. This is called the short track segment condition. Looking at the problem microscopically, an ionizing particle can traverse the site from a random direction. The length of the particle track within the site is a stochastic quantity called chord length, l, and depends on the shape and dimension of the site. If the track is considered as a straight line, which is our case, then chord length is defined as the length of randomly oriented chords in the site (25), and its determination becomes a purely geometric problem. The energy imparted to the site during an event εsdivided by the frequency average of the chord length is also a stochastic quantity called lineal energy y=εs/¯ lF, whose dose-weighted average ¯yDis also of great interest (7). The variability of this collected energy has different causes. First, it depends on the different LET values for the composite beam that strikes the site. Second, the different chord lengths that can result in an event will also produce different εsvalues. Finally, even if the particle LET and its chord length remained constant, the energy imparted can vary. This last variability is called straggling and it is caused by the stochastic nature of the number of electronic collisions within the site and the energy imparted in each one of them. This can be expressed, for a particle of a given LET L and chord length l, as εs=L·l±δ, where δis the stochastic amount of energy to add or subtract to the expected energy ¯εs=L·ldue to straggling. Notice that the frequency average of the δvalue due to straggling is zero by definition, as the mean energy need to remain constant. The relationship between variabilities, expressed as relative variances, 20
Vx=σ2 x ¯x2 F =¯xD ¯xF−1, of each quantity is given by (26): Vεs=VLET +Vl+VLET ·Vl+Vδ(2.3) The relative variance of the straggling distribution is determined by the weighted average of the distribution of energy imparted per collision, δ2= ¯εc1 + σ2 εc/¯εc2, where ¯εcand σεcare the average and the standard deviation of the distribution of energy imparted per collision, respectively. This factor can be taken as Vδ=δ2/¯εs (27). Consequently, we have εD εF =LD LT−1lD lF+lD lF +δ2 εs (2.4) where εFand εDare the frequency average and the weighted average of the energy imparted εs, respectively. Therefore, the relation between frequency averages is εF=LT·lF=εs. According to the definition of y, the frequency-mean lineal energy and the track average LET coincide as long as the short track segment condition is met: yF=εF/lF=LT. By using these relationships, one can obtain for the dose-averaged LET LD=εD lD−δ2 lD =yD lF lD−δ2 lD (2.5) where we have used yD=εD/lF. In Eq. (2.5), macroscopic and microscopic quantities are related to each other. Therefore, to obtain the dose-averaged LET of a particular composite beam, the microdosimetric quantities yDand δ2must be obtained, besides characterizing its chord length distribution. 2.2.3 Methods of samping and their impact on chord length distributions In order to calculate yDthrough MCTS simulations, the following steps should be followed. First, distributions of the energy imparted per event should be collected in the MCTS simulations. Then, they can be transformed into the ydistributions by dividing the imparted energy distribution by the frequency average of the chord length distribution. The dose-mean lineal energy can then be calculated through Eqs. (2.1) or (2.2). In order to collect the single-event distributions, a particle track should be simulated and sampled, that is, the site should be placed, in such a way that at least one energy deposit is produced inside the site. Two methods are considered here: uniform and weighted sampling. Uniform sampling consists of randomly placing the site in the considered volume, which in track segment analysis only contains a portion of the whole track. This method 21
can be expensive in computational terms as it is likely that regions with no energy transfers are repeatedly selected. To avoid this, weighted sampling ensures that at least one energy transfer point is placed within the site. The term “weighted” refers to the fact that the probability of arriving at a certain point is not uniform anymore, but it is weighted by a factor. The “weighted” procedure consists in selecting a transfer point with equal probability and then placing the site center by sampling uniformly a point inside a sphere of radius equal to the site radius centered at the selected transfer point. Not all the elementary volumes have equal probabilities to be selected as the density of transfer points is not uniform, so a compensation factor for each event is needed to obtain the correct frequency single-event distributions. This compensation factor depends on the particular geometry and region selected for sampling in each case, but in general should be inversely proportional to the weight factor in sampling, that is, the probability of selecting a certain elementary volume. Chord length is a stochastic quantity which depends on the geometry of the site and the way it is placed (uniform or weighted randomness) in relation to the particle track. Chord length distributions for specific geometrical configurations are well known (28). In our case, we restrict our study to spherical sites and straight particle paths like proton tracks at clinically relevant energies randomly crossing it from exterior points in space. This situation in which particle tracks traverse the site in a completely random way is known as mean-free-path or µ-randomness (29). The frequency chord length distribution is given by fµ(l) = 2l/d2, whose averages are given by lF,µ = 2d/3and lD,µ = 3d/4, where dis the site diameter. Uniform randomness applies in experimental microdosimetry and in clinical radiation therapy, where targets are uniformly irradiated from the exterior. When using uniform sampling for a straight track, uniform randomness is met. However, if the same track is sampled through weighted sampling, the chord length distribution varies. Now, the track is set to pass at a specific point inside the site and it is not coming from a random point in the exterior anymore. This type of randomness has been studied previously and is known as weighted or νrandomness (28). In this case, the frequency chord length distribution is now given by fν(l)=3l2/d3and its averages are lF,ν = 3d/4and lD,ν = 4d/5. Therefore, the distribution of yvalues depends directly on the method of sampling, since the frequency distributions of chord length are different for each case. A longer mean chord length will imply a larger mean number of collisions within the site, which will be translated into a greater amount of energy imparted. This 22
can be at least partially corrected for the higher mean chord length for yF. In addition, the relationship between LDand yDgiven in Eq. (2.5) is also affected by the different distributions of chord length. In fact, when uniform sampling is employed, this relationship turns out to be LD=8 9yD−4δ2 3dfor µ−randomness (2.6) while, when a weighted sampling method is used, Eq. (2.5) becomes LD=15 16yD−5δ2 4dfor ν−randomness (2.7) where yDis now defined with the mean chord length for the ν-randomness case. Consequently, depending on the chosen kind of sampling, either Eq. (2.6) or (2.7) is used to calculate LD. 2.2.4 Obtaining δ2from track structure Monte Carlo simulations The δ2value present in the Eqs. (2.4)-(2.7) is the weighted average of the energy imparted per collision distribution, f(εc). However, this distribution is complex to study from an analytical point of view, and for this reason it is often considered the energy lost per collision instead of the energy imparted per collision. The simplest approach is to employ the free electron model, which establishes that the energy lost per collision is determined by fl(εc) = k/ε2 c(27), where kis a normalization constant and the subscript lstands for energy lost. In that case, δ2can be approximated by εmax/2 ln(εmax/I), where εmax is the maximum energy loss per collision and Iis the mean ionization potential of the medium. This expression can be adapted to take into account the distribution of the energy imparted per collision to the site by using an effective energy εd, set as the energy of an electron with range approximately equal to the site diameter. Then, δ2≈εd/2 ln(εd/I). A way of estimating the εdvalue is provided in the Appendix of Cort´ es-Giraldo and Carabe work (30). This is a suitable approach when using condensed-history MC simulations, as fεccannot be obtained from the simulations due to the inherent definition of a low energy limit for the tracking of secondary electrons. A more accurate way to obtain δ2is to collect fεcin MCTS simulations and then to obtain its weighted average. As shown below, details may differ from the simulation for obtaining the single-event distributions. The straggling distribution, f(εs, εs), corresponds to the actual energy imparted to the site es when the average energy imparted is given by εs=L·l. Thus, even when particles, with the same LET Lstrike the site, and traverse it with same chord 23
length, l, the single-event energy imparted may differ. Straggling is, therefore, an independent random factor from LET and chord length and its distribution should not be affected by them. It is not possible, consequently, to consider directly all the collisions produced in different events when simulating beams composed by different LET particles, since different events are generated by particles with diverse LET and we would spuriously introduce the variance of the LET distribution into the straggling distribution. Furthermore, the straggling distribution can be obtained from fεcthrough a compound Poisson process (27), which means that the number of collisions in an event can be considered independent and following a Poisson distribution. Nevertheless, when considering particles with different LET values, this independency breaks down and this relation is not valid anymore. 2.2.5 Monoenergetic proton beams The previous formulation has been introduced for the general case of a particle beam with a given LET distribution. If we focus on the particular case of monoenergetic proton beams, the LET distribution is represented by a delta function located at the specific LET of the particle. Therefore, both averages coincide: LT=LD. Furthermore, the relationship between LET and lineal energy is simplified as: LD=LT=yF(2.8) This result is convenient for the following reasoning. We can perform monoenergetic beam simulations and follow the procedure presented above to calculate LDthrough Eqs. (2.6) and (2.7). Then, those results can be compared with the values of yFobtained from the MC simulations according to Eq. (2.8). This represents an indirect method to validate the proposed simulations to obtain δ2: since the uniform simulation is not subject to any bias, we take as reference the value of yFresulting from that simulation, from which we also obtain yDfor the uniform case. We also perform a simulation with weighted sampling method to calculate yDfor this case. Later, we obtain δ2independently for each sampling method and use Eqs. (2.6) and (2.7) for determining LDfor both sampling methods. If results coincide with the reference yF, we can consider the δ2results as valid. 2.2.6 MC simulations for yDand δ2calculations with Geant4-DNA We have performed MCTS simulations by employing Geant4-DNA Incerti et al. (31, 32), Bernal et al. (33), Incerti et al. (34), which is available within the Geant4 Monte Carlo toolkit Agostinelli et al. (35), Allison et al. (36, 37). Monoenergetic 24
proton beams and their secondary electron tracks have been simulated in water, following the procedures in the previous section. Geant4-DNA allows the tracking of electrons down to 10 eV, case in which the energy is considered to be deposited locally. The two sampling methods previously mentioned have been used in different simulations. In both cases, a world volume made of water with margins large enough to account for all the secondary electron is employed. A compensation factor is introduced for accounting the different interaction density when the weighted sampling is used, as it is done in other works (20). Further details about our simulations can be found elsewhere (38). A number of monoenergetic proton beams have been simulatedin these conditions — 2, 5, 10, 40, 70, and 90 MeV — for spherical sites with diameters of 1 and 10 µm, in order to check out the validity of the proposed methodology in those different scenes. Frequency and weighted averages for the single-event energy imparted distribution are directly obtained from the MC calculations. The corresponding yFand yDare calculated by dividing the respective average of the energy imparted by the mean chord length corresponding to the sampling method chosen to place the site. The results are stored in ROOT histograms (39). Since the CSDA range for 2 MeV protons is around 75 µm (40), for the 10 µm site diameter, we are getting closer to the validity limit of the theory (27). This is intentionally done to explore behavior at that limit. As the reason for using weighted sampling is based on efficiency, we have measured the time required for each simulation, consisting of 250 000 proton tracks for each case. Simulations are performed by using our computing cluster hosted at CICA (Seville, Spain), consisting of 24 computational nodes with 2 ×12C AMD Abu Dhabi 6344 (2.6 GHz/16 MB L3). A series of 500 events is simulated each one in 500 cores on these nodes. Computational times presented here are mean times corresponding to the simulation performed by one core, thus, 500 events. As discussed previously, a dedicated approach must be used for obtaining δ2. For each kind of sampling used for obtaining yD, the site should be positioned in such a way that the particle traverses it with a path length equal to the mean chord length (Fig. 2.1). The corresponding position of the site for each case can be easily determined. These dedicated simulations are less expensive computationally since the site is fixed, that is, no sampling is required. Furthermore, in order to check the need to perform these dedicated simulations to calculate δ2, the distributions of the energy imparted per collision were also collected in our simulations devoted to calculate yFand yF(i.e., non-dedicated). 25
X X X X X X X X X X X X X X X X XX X X X X X X X X X X XX X X X X X X X X X X XX X X X XX X X X X X X Figure 2.1: Dedicated approach for the calculation of fεc. The site is fixed at the center of the world volume, made of water and thick enough to allow intra-track charged particle equilibrium. On the transversal plane, the site is placed at a position in which the particle traverses it describing a trajectory with length equal to the mean chord length of the random case (µ- or ν-) under study. Legends: Scoring Region (site); Scored Hits; Non-Scored Hits; Fixed Site. Uncertainty calculations for the presented results are ,derived according to the methodology explained in the Appendix A. All error bars show ±2σfor each quantity. 2.3 Results and Discussion Figure 2.2 shows the results for site diameters of 1 and 10 µm in our range of beam energies for yFand yDwith both sampling methods for the site positioning. A higher average energy imparted is obtained when employing weighted sampling with respect to the uniform sampling case. This is due to the differences 26
in chord length distribution, whose averages are larger for the weighted sampling case. When obtaining y these discrepancies tend to disappear, at least on the average value. Therefore, a good agreement between uniform and weighted yFvalues is observed for the two sampling methods and both site dimensions (Fig. 2.2). However, it is not expected to obtain the same yDvalues by employing the two different sampling approaches, even though the frequency averages do coincide. The reason for this is not only because the average value of the imparted energy distribution is affected by the different chord length distribution, but because its variance is modified. 0 20 40 60 80 0 5 10 15 20 (a) Beam Energy (MeV) Lineal energy (keV/µm) 0 20 40 60 80 0 5 10 15 20 (b) Beam Energy (MeV) Lineal energy (keV/µm) 0 20 40 60 80 −0.5 0.5 1.0 1.5 2.0 (c) Beam Energy (MeV) yF(Weighted) −yF(Uniform) (keV/µm) 0 20 40 60 80 −0.5 0.5 1.0 1.5 2.0 (d) Beam Energy (MeV) yD(Weighted) −yD(Uniform) (keV/µm) Figure 2.2: Frequency and dose mean lineal energy obtained for simulations with weighted and uniform sampling for spherical sites of: (a) 1 µm diameter and (b) 10 µm diameter. For both site dimensions, differences between weighted and uniform results are also shown: (c) for yFand (d)for yD. Legends: (a) yF (Weighted); yF(Uniform); yD(Weighted); yD(Uniform); (b) yF(Weighted); yF(Uniform); yD(Weighted); yD(Uniform); (c) 1 µm; 10 µm; (d) 1 µm; 10 µm. 27
0 20 40 60 80 100 0.5 1.0 1.5 2.0 2.5 δ2(E) for monoenergetic beams Beam energy (MeV) δ2 (keV) Weighted − 1 µm Uniform − 1 µm Weighted − 10 µm Uniform − 10 µm Figure 2.3: Calculated δ2values by using a geometry consisting of a fixed site as shown in Fig. 2.1, depending on the ν- or µrandomness present in the main simulation. Two site diameters (1 and 10 µm) and a variety of monoenergetic beams (from 2 to 90 MeV) are considered. Legends: Weighted - 1 µm; Uniform - 1 µm; Weighted - 10 µm; Uniform - 10 µm. The δ2values calculated with the approach proposed in this work (see Fig. 2.1) are shown in Fig.2.3 for the energy range and the two site diameters considered in this work. As it is not possible to obtain the same frequency distribution for yif uniform or weighted sampling is used, a method for adapting the different variances must be proposed. Thus, the differences in yD[Fig. 2.2(d)] can be compensated to obtain LDby using Eqs. (2.6) and (2.7) with their corresponding δ2value. As it is seen in Fig. 2.3,δ2values are slightly higher for weighted randomness for both site dimensions. This is explained by the longer average chord length, which implies a longer mean path for δrays to travel inside the site: thus the average energy imparted per collision becomes higher. 28
tion code PHITS. International Journal of Radiation Biology, 88(1-2):143–150, jan 2012. ISSN 0955-3002. doi: 10.3109/09553002.2011.611216. [23] ICRU. Report 90: Key Data for ionizing radiation dosimetry: measurement standards and applications. Technical Report 90, 2016. [24] A. M. Kellerer and H. H. Rossi. The Theory of Dual Radiation Action. In M. Ebert and A. Howard, editors, Current Topics in Radiation Research. Volume VIII, pages 85–156. American Elsevier Publishing Company, Manchester, U.K., 1974. [25] ICRU. Report 36. Microdosimetry. Technical report, International Commision On Radiation Units and Measurements (ICRU), 1983. [26] A. M. Kellerer. Analysis of Patterns of Energy Deposition. In H. G. Ebert, editor, Second Symposium on Microdosimetry, pages 107–136, Stresa (Italy), 1970. [27] A. M. Kellerer. Fundamentals of microdosimetry. In K. R. Kase, B. E. Bjarngard, and F. H. Attix, editors, The Dosimetry of Ionization Radiation. Volume I, chapter Fundamenta, pages 77–162. Academic Press, Inc., 1985. [28] A. M. Kellerer. Chord-Length Distributions and Related Quantities for Spheroids. Radiation Research, 98(1):425–437, 1984. [29] R. Coleman. Random Paths through Convex Bodies. Journal of Applied Probability, 6(2):430–441, 1969. ISSN 87507587. doi: 10.1111/j.1365-2672.2007. 03484.x. [30] M. A. Cort´ es-Giraldo and A. Carabe. A critical study of different Monte Carlo scoring methods of dose average linear-energy-transfer maps calculated in voxelized geometries irradiated with clinical proton beams. Physics in Medicine and Biology, 60(7):2645–2669, 2015. ISSN 13616560. doi: 10.1088/0031-9155/60/7/2645. [31] S. Incerti et al. Comparison of GEANT4 very low energy cross section models with experimental data in water. Medical Physics, 37(9):4692–4708, 2010. ISSN 00942405. doi: 10.1118/1.3476457. [32] S. Incerti et al. The Geant4-DNA project. International Journal of Modeling, Simulation, and Scientific Computing, 1(2):157, 2010. ISSN 1793-9623. doi: 10.1142/S1793962310000122. 35
[33] M. A. Bernal et al. Track structure modeling in liquid water: A review of the Geant4-DNA very low energy extension of the Geant4 Monte Carlo simulation toolkit. Physica Medica, 31(8):861–874, dec 2015. ISSN 11201797. doi: 10.1016/j.ejmp.2015.10.087. [34] S. Incerti et al. Geant4-DNA example applications for track structure simulations in liquid water: A report from the Geant4-DNA Project. Medical Physics, 45(8):e722–e739, 2018. ISSN 00942405. doi: 10.1002/mp.13048. [35] S. Agostinelli et al. Geant4—a simulation toolkit. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 506(3):250–303, jul 2003. ISSN 01689002. doi: 10.1016/S0168-9002(03)01368-8. [36] J. Allison et al. Geant4 developments and applications. IEEE Transactions on Nuclear Science, 53(1):270–278, feb 2006. ISSN 0018-9499. doi: 10.1109/ TNS.2006.869826. [37] J. Allison et al. Recent developments in Geant4. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 835:186–225, nov 2016. ISSN 0168-9002. doi: 10.1016/J.NIMA.2016.06.125. [38] A. Baratto-Rold´ an et al. Abstract ID: 196 Relation between dose average linear energy transfer and dose mean lineal energy calculated for proton therapy beams off axis: A study with the Geant4 toolkit. Physica Medica, 42(2017): 42–43, 2017. ISSN 11201797. doi: 10.1016/J.EJMP.2017.09.104. [39] R. Brun and F. Rademakers. ROOT — An object oriented data analysis framework. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 389(1-2):81–86, apr 1997. ISSN 01689002. doi: 10.1016/S0168-9002(97)00048-X. [40] M. Berger, J. Coursey, M. Zucker, and J. Chang. ESTAR, PSTAR and ASTAR: Computer Programs for Calculating Stopping-Power and Range Tables for Electrons, Protons, and Helium Ions (version 2.0.1), 2017. 36
Chapter 3 Segment-averaged LET concept and analytical calculation from microdosimetric quantities in proton radiation therapy 37
Med. Phys. 2019, 46(9). 4204-4214 Segment-averaged LET concept and analytical calculation from microdosimetric quantities in proton radiation therapy A. Bertolet1,2, A. Baratto-Rold´ an2, M.A. Cort´ es-Giraldo2, A. Carabe-Fernandez1 1Department of Radiation Oncology, Hospital of The University of Pennsylvania, Philadelphia, 19104 PA, USA 2Department of Atomic, Molecular and Nuclear Physics, Universidad de Sevilla, Seville, Spain Purpose: This work introduces the concept of segment-averaged linear energy transfer (LET) as a new approach to average distributions of LET of proton beams based on a revisiting of microdosimetry theory. The concept of segment-averaged LET is then used to generate an analytical model from Monte Carlo simulations data to perform fast and accurate calculations of LET distributions for proton beams Methods: The distribution of energy imparted by a proton beam into a representative biological structure or site is influenced by the distributions of (a) LET, (b) segment length, which is the section of the proton track in the site, and (c) energy straggling of the proton beam. The distribution of LET is thus generated by the LET of each component of the beam in the site. However, the situation when the LET of each single proton varies appreciably along its path in the site is not defined. Therefore, a new distribution can be obtained if the particle track segment is decomposed into smaller portions in which LET is roughly constant. We have called “segment distribution” of LET the one generated by the contribution of each portion. The average of that distribution is called segment-averaged LET. This quantity is obtained in the microdosimetry theory from the average and standard deviation of the distributions of energy imparted to the site, segment length, and energy imparted per collision. All this information is calculated for protons of clinically relevant energies by means of Geant4-DNA microdosimetric simulations. Finally, a set of analytical functions is proposed for each one of the previous quantities. The presented model functions are fitted to data from Geant4-DNA 38
simulations for monoenergetic beams from 100 keV to 100 MeV and for spherical sites of 1, 5, and 10 µm in diameter. Results: The average differences along the considered energy range between calculations based on our analytical models and MC for segment-averaged doseaveraged restricted LET are -0.2 ±0.7 keV/µm for the 1 µm case, 0.0 ±0.9 keV/µm for the 5 µm case, and -0.3 ±1.1 keV/µm for the 10 µm case, respectively. All average differences are below the average standard deviation (1σ) of the MC calculations. Conclusions: A new way of averaging LET for a proton beam is performed to incorporate the effects produced by the variation of stopping power of each individual proton along microscopic biological structures. An analytical model based on MC simulations allows for fast and accurate calculations of segment-averaged dose-averaged restricted LET for proton beams, which otherwise would need to be calculated from exhaustive MC simulations of clinical plans. 3.1 Introduction In radiation therapy, the absorbed dose by cells determines their probability of survival. However, the biological effectiveness of the same physical dose of particle therapy and x-rays is different (1). In particular, in vivo experiments have shown that protons are not equally effective in the entrance region, the Bragg peak or the distal falloff region of the beam (2). This difference in radiobiological effectiveness has been related to the increase of linear energy transfer (LET) towards the end of the proton range (3, 4, 5). There is, therefore, an increasing interest in using LET in the optimization process of the treatment plan (6, 7). For this reason, we need an accurate and fast method to calculate LET. The present work proposes the use of generalized microdosimetry theory to calculate LET for proton beams based on the distribution of energy imparted in microscopic (and possibly biologically relevant) structures (8, 9, 10). The concept of electronic stopping power — or unrestricted LET — could be considered to measure the energy deposited by proton beams due to electronic collisions in cellular or sub-cellular regions. However, there are at least two relevant considerations disregarded by this quantity: the energy transported out of these regions by secondary electrons and the finite range of protons (11); in 39
other words, situations of fast changes of LET in short paths, which occurs in the trajectory end of the beam. The first can be incorporated in the concept of restricted linear energy transfer(8) L∆, where the energy deposition by secondary electrons with kinetic energy beyond a cutoff value ∆is disregarded. Efforts have been devoted to calculate either unrestricted or restricted LET for clinical proton beams in acceptable times in clinical routine. Some analytical approaches have been proposed based on the proton stopping power values (12, 13, 14, 15), while other works try to include the effect of the energy carried away by secondary electrons using geometrical considerations to obtain spatially or radially restricted LET (16, 17, 18). However, how to deal with situations in which LET may change significantly across microdosimetric structures remains unclear. This is especially important since in such situations the highest values of LET are produced. In this work, we are proposing a calculation method of LET based on microdosimetry to address simultaneously the limitations referred above, that is, both situations in which restricted LET applies and when LET varies along microscopic structures. Specific details for performing microdosimetric Monte Carlo (MC) simulations for this aim were presented in 2(19). The resulting distributions from these simulations were used in this work as input data to generate analytical models to calculate LET, and the following sections describe these models in detail. The use of these models to calculated LET in clinical proton beams is expected to be a faster approach than full Monte Carlo LET calculations while keeping a comparable accuracy. 3.2 Methods and Materials 3.2.1 Definitions of relevant LET and microdosimetric concepts Linear energy transfer (LET) is a nonstochastic quantity that considers the mean energy lost by a charged particle per unit path length. When dealing with a field of charged particles with diverse energies, as in the case of a clinical beam, a different value of LET can be found for each particle. Therefore, rather than a single LET value, a spatial distribution of LET has to be considered in a patient. Thus, it is convenient to reduce that distribution to average values, which can be done in two different ways: one is the fluence-based average, traditionally called track-averaged LET, LT; while in the second one, is called dose-averaged LET, LD, where the LET of each particle is weighted by its contribution to the local dose (12). When we consider energy imparted into a certain region instead of the energy lost 40
by the particle, the concept of energy-restricted LET, L∆, becomes useful. In microdosimetry, a site is a region of the irradiated volume with a given shape and microscopic dimension (20). Although any shape can be considered, this work is restricted to spherical sites for simplicity. When a charged particle strikes a site, a stochastic amount of energy, ε, is imparted to it, carried by both the particle itself and the secondary electrons generated in electronic collisions. The term “imparted” means that energy is considered to be deposited locally, that is, within the site. Each energy deposition resulting from a different primary particle and its showers of secondary particles is called an event. As we are interested in calculating LD, only electrons are taken into consideration as secondary particles in each event; this implies that other types of secondary particles (e.g., alphas) must then be considered as separate contributors to LD. Therefore, when a number of particles impact on a site, a distribution of energy imparted per event, so-called single-event distribution (21), is originated, with average energy imparted εand variance σ2 ε. The problem studied in this work consists in a clinical proton beam irradiating a biological target. Tissues can be modeled as composed of sites, also called domains (22), which may represent cells or sub-cellular critical structures. Our interest is to calculate the distributions of energy imparted in sites by a beam of protons with a given energetic spectrum. This beam is considered to be broad enough to laterally cover all the site plus a certain margin used to ensure that all energy transfers imparted by secondary electrons generated outside the domain are accounted for. In this case, for any proton randomly selected from the incidental beam, the portion of its track inside the site is also a stochastic quantity. Protons are assumed to travel in a straight line, which is a good approximation as long as we disregard nonelectronic collisions in the analysis (23). The path of the track the proton travels within the site is called segment length (s) (24). This definition applies to both when protons stop in the site and when they completely traverse it. We use this quantity to define lineal energy (ys) as the quotient between the energy imparted to the site in an event and the mean segment length: ys=ε s(3.1) The subscript sindicates that segment length is used to define lineal energy. Lineal energy is, therefore, a stochastic quantity with its own distribution of frequencies, characterized by the frequency probability density f(ys), whose average is denoted as yF,s. A related distribution of ysconsiders the dose absorbed by 41
the tissue in each event with lineal energy ys. This is called dose distribution of ys,d(ys), and its average is called dose-mean lineal energy (21), yD,s. Since lineal energy is proportional to the dose absorbed by the site, the dose distribution of yis essentially a weighted distribution multiplied by a constant factor, that is, d(ys)∝ysf(ys). Therefore, the averages are related by the following equation, which applies to all frequency and weighted averages (9): yD,s =yF,s 1 + σ2 ys y2 F,s !=ε s1 + σ2 ε ε2(3.2) where σ2 ysis the variance of the frequency distribution of ys. The second equality is obtained from the definition of ysin Eq. (3.1). 3.2.2 Segment distribution of LET According to Eq. (3.2),yD,s is not only related to the average of the frequency distribution of lineal energy in the site but it also takes into account the variability of that distribution. It can be shown (9) that three factors are involved in the different values of ysfor each case: the different LET of the protons in the site, the different chord length of each proton and the energy straggling. We can formulate this in terms of the variance of each variability source. By considering Eq. (3.2) one can obtain the variance of the distribution of segment length in terms of the average of its weighted distribution d(s) = sf(s), also called weighted-mean segment length, sD. Additionally, the variance of the straggling distribution can be expressed as δ2/ε, where δ2is also the weighted average of the distribution of energy imparted into the site per collision (25). Here, we propose an extension of the applicability of the LET concept based on the following main idea: the variability of LET in the site can be decomposed into two different terms. In the situation in which protons have no appreciable change of LET in the site, a field of protons striking the site with different LET, or, equivalently, different energy, produce a variety of LET values. Thus, a “field distribution” of LET is produced by different protons. On the other hand, if a single proton loses an appreciable amount of its kinetic energy inside the site, its stopping power and, consequently, its LET are not unique as they can vary considerably in the proton path within the site. This case in which a proton changes its L∆within the site has not been previously considered in the microdosimetry literature, since LET is a macroscopic concept and such a change would have a stochastic nature. In order to overcome the indetermination on the LET concept, 42
we have defined a new distribution of LET values based on this variability within the site. This distribution can be understood as follows. Let dx be an element of length small enough so that L∆can be considered constant along dx for each proton energy. Then, the different values of L∆in each dx of the segment length within the site result in a new LET distribution. We call this distribution thereby “segment distribution” of LET for each proton traversing the site. This “segment distribution” is also a source of variability in ys, since depending on the actual track segment that intersects the site, the combination of L∆values for the proton in the site might change from event to event. Hence, the total distribution of LET in the considered segment length is produced by the combination of the “field distribution” and the “segment distribution”. Based on the introduction of a new level of variability in the lineal energy due to the segment distribution defined above, the segment-distribution-average value of the dose-weighted distribution of LET is here called “segment-averaged doseaveraged LET”, LD,s. This “segment-average” depends on the shape and size of the site. Again, this quantity is related to the variance of the LET distribution by means of Eq. (3.2). Considering all above, from the relation between relative variances of the sources of ysvariability, the following equation for the segmentaveraged dose-averaged restricted LET can be obtained by employing analogous arguments to those developed by Kellerer (9): L∆,D,s =s sD yD,s −δ2 sD (3.3) Here, as energy imparted into a spherical site is considered instead of energy lost, the ∆subscript is added to LD,s in Eq. (3.3) to indicate we are dealing with dose-averaged restricted LET. Now, ∆is equal to the energy for secondary electrons that makes their range match the distance they have to travel to escape from the site. Therefore, by employing Eq. 3.3, it is possible to obtain a calculation for LET that does not count the energy escaped from each site or biological domain by secondary electrons and the effects of the energy straggling. Additionally, this definition of LET applies to the situation in which the energy imparted per unit length varies appreciably within the site or the extreme situation in which protons stop inside the site. 3.2.3 A particular case of interest: constant LET in the site When protons completely traverse the site, that is, when their range is greater than the site diameter, the calculation of the distribution of segment length be- 43
comes a purely geometrical problem and it is obtained by determining the intersection between a straight line and the site. This is a well-known and solved problem for a spherical site (26). In this case, the “segment length” is called “chord length” (l). For spherical sites with diameter d, the chord length distribution is given by f(l)=2l/d2/: (0 ≤l≤d), with average l= 2d/3. This quantity is traditionally used in microdosimetry to define lineal energy (y) instead of the more general segment length (21) defined above. Therefore, in this particular case, Eq. (3.1)) becomes: y=ε l(3.4) We now consider only the case in which protons have a kinetic energy large enough to consider that their LET does not change within the site. In this case, furthermore, protons always cross the site completely thus s→l. Also, under constant LET conditions, only the “field distribution” of LET plays a role. Therefore, in this limit, L∆,D,s →L∆,D and Eq. 3.3 becomes: L∆,D =l lD yD−δ2 lD =8 9yD−4δ2 3d(3.5) where the last equality is valid only for spherical sites with diameter d. Eq. (3.5) is not new (8, 9), but now it is presented as a particular case of Eq. (3.3) to determine the dose-averaged LET of proton beams. Therefore, we extend its validity by presenting a more general theory in Eq. 3.3, in which this result is included. 3.2.4 Monte Carlo simulations for monoenergetic protons and its application to polyenergetic (clinical) beams The segment-averaged LET definition is, generally, inconvenient since it depends on the site shape and dimension and, therefore, produces different values for each site considered. Moreover, segment length for low energy protons are complicated to calculate analytically, thus using Eq. (3.3) instead of Eq. (3.5) seems to add complexity to LET calculations. However, by using Monte Carlo track structure (MCTS) simulations, all these quantities can be obtained. Furthermore, the problem of calculating yD,s and L∆,D,s in a polyenergetic (clinical) beam can be decomposed into simulations of monoenergetic protons, from which an energy kernel can be developed starting from the corresponding distributions of energy imparted. Then, if the proton clinical beam is composed by an energy spectrum given by φ(E), as the energy imparted distribution is independent for each monoenergetic proton, the average, εφ, and the variance of the total distribution of energy imparted, σ2 εφ, (i.e. considering the whole fluence) are given, respectively, by 44
10-2 10-1 100101102 ¯sF(µm) 0 0.375 0.75 1 µm 10-2 10-1 100101102 ¯sF(µm) 0 1.875 3.75 5 µm Proton Energy (MeV) 10-2 10-1 100101102 ¯sF(µm) 0 3.75 7.5 10 µm Figure 3.3: Calculations from MC simulations done with Geant4-DNA in liquid water for monoenergetic protons from 0.01 to 100 MeV (points) and corresponding fitted function, Eq. (3.11) (lines) for the average segment length described in spherical sites with diameters of 1, 5, and 10 µm, respectively. Statistical uncertainties of the MC simulation (1σ) are shown with error bars. Legends: MC Data; Equation (3.11). for spherical sites of 1 µm diameter. Residuals for this fit are shown in the Data S1 (SM-2a). Results and residuals for weighted average segment length, sD, Eq. (3.12), are also shown in the Data S1 (SM-3 and SM-2b, respectively). 51
Once obtained the previous results, it is possible to calculate yDaccording to the way specified in Eq. (3.2). Figure 3.4 shows the comparison between the calculated yDvalues following this equation by using the calculations from MC simulations and the analytical fitted functions already shown. Differences between calculated values through the MC calculations and the analytical model are shown in the Data S1 (SM-4) Figure 3.5 shows the comparison between the analytical expression shown in Eq. (3.13) and the MC data resulting from Geant4-DNA simulations for the average and the standard deviation of the distribution of imparted energy per collision, δ2(E). Results and residuals for the fits for each one of these two functions are shown in the Data S1 (SM-5 and SM-6, respectively) Differences between δ2(E) values for MC and analytical calculations are shown in the Data S1 (SM-7) as well. Combining all the results shown above, for the three site diameters considered, it is possible to calculate the segment-averaged dose-averaged restricted LET, L∆,D,s, for each proton energy by means of Eq. (3.3). Figure 3.6 shows the comparison between the results for L∆,D,s employing the data coming from the MC simulations and from all the analytical models proposed in this work. Differences between L∆,D,s values calculated from MC data and the analytical models are also shown in Fig. 3.6. The difference for protons at 0.1 MeV is found to be considerably larger than at any other energy for a site of 10 µm diameter. This is due to the low values for sFand sDat that point, which implies that small variations from the MC results for that quantities produce a large deviation for L∆,D,s. Leaving that point out, average differences (model calculations minus G4-DNA computations) are 0.2 ± 0.7 keV/mum for the 1 mum case, 0.0 ±0.9 keV/µm for the 5 µcase and 0.3 ±1.1 keV/µm for the 10 µm case, where the uncertainties indicate the standard deviation of the differences. To look at the relevance of found differences, these values can be expressed as relative to the average standard deviation for the Monte Carlo calculations in each case. The average uncertainties (1σ) for the Monte Carlo calculations, respectively, are 0.3, 0.8, and 1.4 keV/lµ. Consequently, all the average discrepancies between the model and the MC calculations are below 1σ. Finally, by interpolating the values of the fitting parameters obtained in the models given by the Eqs. (3.9)–(3.12), a prediction of the average energy imparted and L∆,D,s for intermediate site diameters can be assessed. Figure 3.7 shows predictions for spherical sites with diameters of 3 and 7 µm, respectively, compared 52
10-1 100101102 ¯yD,s (kev/µm) 0 50 100 1 µm 10-1 100101102 ¯yD,s (kev/µm) 0 45 90 5 µm Proton Energy (MeV) 10-1 100101102 ¯yD,s (kev/µm) 0 45 90 10 µm Figure 3.4: Dose-mean lineal energy obtained from MC simulations of monoenergetic protons in liquid water done with Geant4-DNA (points), and with our proposed analytical models (lines) by using Eq. (3.2), for spherical sites with diameters of 1, 5, and 10 µm, respectively. Statistical uncertainties propagated from the MC simulations (1σ) are shown with error bars. Legends: MC Data; Equation (3.4). with the calculations obtained from Geant4-DNA MC simulations. Residuals between the predicted results and the MC data are shown in the Data S1 (SM-8). 53
10-1 100101102 δ2(keV) 0 0.375 0.75 1 µm 10-1 100101102 δ2(keV) 0 0.875 1.75 5 µm Proton Energy (MeV) 10-1 100101102 δ2(keV) 0 1.125 2.25 10 µm Figure 3.5: Weighted average of the distribution of energy imparted per collision, δ2, calculated for monoenergetic protons in liquid water from Geant4-DNA MC simulations (points), and from the analytical models described in Eq. (3.13) (lines) for spherical sites with diameters of 1, 5, and 10 µm, respectively. Statistical uncertainties of the MC simulations (1σ) are shown with error bars. Legends: MC Data; Equation (3.12). The average deviation found for the L∆,D,s prediction with respect to the G4- DNA calculated points is now 1.2 ±1.8 keV/µm for the 3 µm site and 1.2 ±3.6 keV/mum for the 7 µm, where the uncertainties refer to the standard deviation of 54
100102 ¯ L∆,D,s (keV/µm) 0 45 90 1 µm (a) 100102 ¯ L∆,D,s (keV/µm) 0 35 70 5 µm Proton Energy (MeV) 100102 ¯ L∆,D,s (keV/µm) 0 35 70 10 µm 100102 Diff(keV/µm) -5 0 51 µm (b) 100102 Diff(keV/µm) -10 0 10 5 µm Proton Energy (MeV) 100102 Diff(keV/µm) -5 0 510 µm Figure 3.6: (a) Segment-averaged dose-averaged restricted LET, L∆,D,s, calculated with Geant4-DNA MC simulations of protons in liquid water (points) and with our proposed analytical models (lines) by using Eq. (3.3), for spherical site with diameters of 1, 5, and 10 µm, respectively. Statistical uncertainties propagated from the MC simulations (1σ) are shown with error bars. (b) Absolute differences between L∆,D,s calculated with our proposed analytical models and with Geant4- DNA MC simulations. The point for the 10 µm site and 0.1 MeV is not shown, being this difference equal to 22.64 keV/µm. Legends: (a) MC Data; Equation (3.3); (b) Residuals: (L)Model ∆,D,s −(L)MC ∆,D,s. 55
100102 ¯ε(keV) 0 75 150 3 µm (a) Proton Energy (MeV) 100102 ¯ε(keV) 0 165 330 7 µm 100102 ¯ L∆,D,s (keV/µm) 0 40 80 3 µm (b) Proton Energy (MeV) 100102 ¯ L∆,D,s (keV/µm) 0 35 70 7 µm Figure 3.7: Data obtained from the Geant4-DNA MC simulations of monoenergetic protons in liquid water (points) and the prediction resulting from the interpolation of the fitting parameters of our models (lines) for spherical sites with diameter of 3 and 7 µm, respectively. (a) Frequency average of single-event energy imparted to the site, ε; (b) segment-averaged dose-averaged restricted LET L∆,D,s. Statistical uncertainties propagated from the MC simulations (1σ) are shown with error bars. Legends: (a) MC Data; Model Prediction; (b) MC Data; Model Prediction. the differences. 3.4 Discussion Three characteristics make the LET calculation presented in this work different from the collision stopping power: dose-averaging, restriction, and segment averaging. Furthermore, the calculated LET depends on two different parameters: the initial proton kinetic energy and the site dimension. In our case, as we deal with spherical sites, the latter is characterized by the site diameter. The following discussion analyzes the importance of each characteristic of our LET calculation depending on the proton energy and the site diameter. 56
The dose-averaging is incorporated into the calculation by taking into account the standard deviation of the different quantities involved and by using the relative variances relation leading to Eq. (3.3). It is, therefore, relevant for all proton energies and site diameters. The restriction is performed by using a given spatial structure to score the energy imparted and, therefore, is implicit in the distributions of single-event energy imparted calculated. This spatial restriction is important in such situations in which secondary electrons generated by the proton from ionizations of molecules of the medium might escape from the site in a non-negligible proportion. This condition is met when either the site is relatively small or the kinetic energy transferred to the electrons is high enough, that is, the proton energy is high enough to make them escape. Therefore, we expect to find the main differences between restricted LET and stopping power at higher energies as the site size increases. For the same proton energy, the restricted LET value is higher for larger sites and it tends to converge to the stopping power when an infinitely large site is considered. On the other hand, the segment-averaging process becomes important in those cases in which the stopping power of the proton changes its value significantly across the site. This happens when either the site is relatively large or the proton energy is low enough. Secondary electrons do not play a role in the latter case because the amount of energy transferred to them is of the order of eV, thus they are not expected to travel long distances from the proton track. It is convenient to notice that the segment-averaging process is similar to integrate the collision stopping power curve S(E)for protons in water along the considered distance, that is the site diameter, LD,s ≈RS(E(x)) dE(x) dxdx/ RdE(x) dxdx, where E(x)is the energy of the proton at the point x. Therefore, the larger is that distance, the broader and smaller results the new curve’s peak with respect to collision stopping power. Additionally, because of the same reason, the new peak will be shifted towards higher energies. This explains why for the 1 µm diameter site such a peak is not observed. Additionally, the larger the site, the more the stopping power is able to change. This means that a longer portion of the curve of stopping power respect to the proton energy is considered to calculate the segment-average LET. This explains the different positions (on x-axis) and amplitude of the peaks obtained in the curves for the average and the standard deviation of the single-event energy imparted distribution shown in Fig. 3.2 for different site diameters. As the site becomes larger, the peak gets higher and shifted towards higher initial proton energies because a 57
longer portion of the track is considered, thus a proton is able to stop and deposit all its energy in the site entering with higher initial energies. The reason for the spike observed in Fig. 3.2(b) for σεat large sites (diameter of 5–10 µm) is based on the fact that distributions of energy imparted by protons at very low energies have a tailed shape at the high energy part of the distribution, being this tail due to the events produced by “stoppers”. Then, two competitive effects take place when very low energies are considered: as the proton energy increases, (a) the absolute energy imparted grows so that the absolute standard deviation of the distribution does too; but (b) the number of events produced by “stoppers” decrease (their range become larger), thus the mentioned tail fades away, which makes the standard deviation to decrease. For the 1 µm diameter site, the second effect occurs already below 100 keV, which is why this effect has practically no impact on the curve. However, the larger the site, the higher the energy at which the tail disappears so that, at a certain point, the second effect starts to be noticed, making the standard deviation decrease. Once the tail vanishes, σεfollows the dependence of εwith the proton energy. By means of Eq. (3.3), the effects of segment length variability and straggling on the distribution of ysmay be disregarded in the L∆,D,s value. This is convenient because these two sources of variability for yscarry all the dependencies with the site shape and, as LET is a macroscopic concept, it should not depend on this site shape. This dependency for the segment length is straightforward. In the case of the straggling effect, as the energy imparted per collision includes the energy transferred by secondary electrons, the site shape influences whether an electron with a given kinetic energy can leave the site or not. According to all discussed above, therefore, L∆,D,s does not depend on the site shape but a spatial parameter given by the site diameter d. Furthermore, the term “restricted LET” refers to a spatial restriction of the energy imparted given by the parameter ∆, related somehow to the site diameter d, that is, ∆ = ∆(d), and it is relevant in those cases in which secondary electrons have ranges higher or comparable to d. On the other hand, the term “segment-averaged LET” refers to a longitudinal averaging of the energy imparted given by the mean segment length, s(d). In this case, this becomes important when LET changes significantly along that track segment. The adequate values for the parameters ∆(d)and s(d)should be related to different clinical endpoints and the effect of the distribution of energy imparted in certain biological structures on those endpoints. Therefore, it is desirable to have 58
a tool to calculate spatial segment-averaged LET distributions for different sites in order to look for correlations between L∆,D,s values and the studied endpoints. By studying if these correlations exist, the quality of a restricted LET based on a given dimension d as an indicator of some biological effect can be assessed. In other words, LET restriction may need to be considered to explain the relation between RBE and LET. This work introduces this degree of freedom and proposes how to handle it in order to find such correlations. The concepts of unrestricted and segment-averaged LET make this calculation expected to be different from the usual LET calculations published in literature, which employ a concept of LET based on the (unrestricted) electronic stopping power values for protons in water (12, 13, 14, 15). These differences are expected to occur both at the entrance region, where the restriction effect plays a role, and at the proximities of the Bragg peak, where the segment averaging becomes relevant. In all cases, L∆,D,s is expected to be lower than the LDcalculated by those methods. Regarding the models proposed in this work, a remarkable agreement between the analytical functions and the MC-based data can be found in Figs. 3.1-3.5. This provides a double advantage: (a) calculations of L∆,D,s values can be performed much faster than MC simulations by following this approach, since simple analytical functions are only used; and (b) precision comparable to MC calculations, which resolves some of the issues of the current analytical models for LET calculations (14, 15), besides the restricted and segment-averaged characteristics, which, as discussed above, might link LET with clinical endpoint. For example, performing LET calculation with this approach does not need to account for primary and secondary protons separately, as many of previous works do (12, 13, 14). As, according to Eqs. (3.6) and (3.7) we need to use the clinical beam spectral fluence φ(E)to convolve the presented models and to finally obtain the corresponding LET values, it is enough with calculating the contribution from the secondaries in terms of fluence instead of in terms of LET. This may represent a real advantage when calculating lateral distribution of LET, as they only depend on the lateral beam fluence. To perform clinical calculations of L∆,D,s using the proposed method, each domain has to have a single value of L∆,D,s which is obtained by averaging the modeled quantities respect to the spectral fluence of the beam, as shown in Eqs. (3.6) and (3.7), that is, by considering the distribution of entrance proton energies into the domain. Because of this polyenergetic averaging process, discrepancies observed in Fig. 3.6 and even in Fig. 3.7, which shows the prediction for site di- 59
mensions whose data has not been employed in the model process, will probably diffuse. This is especially likely for the high discrepancies in low energies since in the clinical situation, relatively broad spectra are present and usually several beams converge at the same point. This means that the fraction of protons with low energy becomes small, which makes the contribution of low-energy protons to the integrals used in Eqs. (3.6) and (3.7) to be irrelevant. We would also like to remark that this work does not aim at determining the spectral fluence employed in Eqs. (3.6) and (3.7), which is a separated problem to be addressed in separate chapters. It should be noticed that the actual accuracy of this methodology strongly depends on how accurate is the determination of the spectral fluence. Essentially, this determination consists in transporting protons in the actual media they traverse, so that either calculations based on MC simulations or a fluence-based kernel can be employed. As a consequence, resolution of CT scans, for example, becomes an issue for this spectral fluence determination. However, the intrinsic microscopic distributions are carried by our models regardless how accurate the spectral fluence is. On the other hand, as our microdosimetric models represent the way in which energy is imparted to water, the L∆,D,s calculated distributions are referred to water. This seems convenient considering that the usual way to report dosimetric distributions in radiation therapy is dose-to-water. The presence of non-homogeneities, therefore, affects the proton transport problem (at macroscopic scale) rather than the microdosimetric energy deposition. Calculations in this work need to be extended beyond the current highest proton energy simulated in Geant4-DNA, currently fixed at 100 MeV, in order to simulate protons at the full clinical energy range. However, the contribution of protons with energy higher than 100 MeV to the LET value is expected to be low. Additionally, if we restrict our calculation to the LET distribution in a tumor, those energies above 100 MeV do not have much relevance. Nonetheless, we do not expect an appreciable change of behavior in LET at 100 MeV along the rest of the clinical energy ranges, so the model presented here could be extrapolated to higher energies with probable limited error. In any case, Monte Carlo calculations should be performed to confirm this. 60
for yFand 1.17% for yD. Our analytical method generates maximum deviations of 1.29% and 3.33%, respectively, with respect to experimental results. Conclusions: Simulations in Geant4-DNA with ideal cylindrical sites in liquid water produce similar results to the measurements in an actual silicon-based cylindrical microdosimeter properly calibrated. The found agreement suggests the possibility to experimentally verify the calculated clinical yDwith our analytical method. 4.1 Introduction Protons present increased values of linear energy transfer (LET) (1, 2, 3, 4, 5, 6) which have been correlated with an increased biological effectiveness at the distal edge of clinical beams (7, 8). For this reason, LET calculations capabilities are being introduced in treatment planning systems (TPS) for research purposes (9, 10, 11). Additionally, in the case of further validation of that correlation, clinical treatments might need to consider LET besides dose to guide the treatment design. If that is the case, such LET distributions calculated by the TPS should be verified as part of the quality control program of the proton clinic. However, distributions of LET of the particles in a volume are challenging to be measured in a patient, so that measuring a closely related quantity to LET such as the lineal energy (y) becomes more convenient. yis the microdosimetric equivalent to LET (12, 13), and is also correlated to biological effectiveness (14, 15, 16, 17, 18, 19, 20, 21). Lineal energy, unlike LET, is a stochastic quantity, meaning that distributions of occurrence are obtained for a given irradiation instead of deterministic values. Hence, yis capable of determining the volumetric patterns of energy deposition in microscopic structures (22, 23). Lineal energy can be computed with dedicated microdosimetric Monte Carlo (MC) simulations (24, 25, 26, 27), usually slower than analogous macroscopic MC simulations to compute LET. Nonetheless, it can be experimentally determined using tissue-equivalent proportional counters (TEPCs), which have been traditionally used (28). These instruments, however, are filled with gas to simulate an equivalent microscopic site, which means that their actual size is macroscopic, producing some flaws that need to be corrected. To overcome this, mini-TEPC has been recently developed and tested to measure more accurate microdosimetric distributions for proton beams (29, 30). A different technology to address microdosimetric 67
measurements is based on silicon detectors (31, 32, 33, 34). Unlike TEPC, new silicon-based 3D microdetectors do not need a gas supply, can reach microscopic size, and their ability to reproduce accurate microdosimetric spectra has been checked (35, 36, 37). In the previous chapter, we have developed a methodology able to compute 3D distributions of the dose-mean lineal energy within a patient (38). In this work, we measure lineal energy distributions of low energy proton monoenergetic beams and compare the results with the mean values calculated analytically with our method, as well as MC-simulated distributions. Our goal is, to validate our analytical calculation methodology (38) in the low proton energy regionto then assess its potential for experimental measurements in clinical scenarios. 4.2 Methods and Materials Microdosimetry deals with the interactions between radiation and matter at microscopic level, in which the stochastic nature of these interactions become prominent. In this context, the concept of microdosimetric site arises in order to define a region in which the energy deposited by radiation is accounted for. Thus, the energy imparted in a site, ε, is a stochastic quantity that varies from experiment to experiment or, in microdosimetric terms, from event to event, depending on the dimensions of the considered site. In the case of hadron beams, lineal energy (y), which represents the amount of energy imparted per unit chord length, might be a more meaningful quantity to characterize microdosimetric interactions. Lineal energy is defined as (12) y=ε l(4.1) where lis the mean chord length for the particle track within the site. This is a solved geometrical problem for sites of simple geometry (39, 40), with l= 2d/3for spheres of diameter dfor isotropic radiation. The resultant distribution of y,f(y), obtained from the variable value of yper event, can be characterized by its mean value, yF. Nonetheless, this quantity disregards that high values of y, i.e., events with large energy imparted, are associated with harder damage to cells according to the theory of dual radiation action (TDRA) (41). Therefore, the dose-weighted distribution of y, defined as d(y) = yf(y)/yFrepresents a more convenient way to deal with lineal energy. Therefore, the average of d(y), so-called dose-mean lineal energy yDrepresents a single parameter that best characterizes the efficiency of a given radiation for producing biological damage. 68
4.2.1 Analytical models for microdosimetric calculations A new analytical methodology to obtain mean values of microdosimetric distributions in clinical beams has been recently developed and published in Bertolet et al., 2019 (38). In this approach, polyenergetic clinical beams, characterized by a spectral fluence φE(E), are decomposed into their spectral components. The mean energy imparted to a microscopic site for the polyenergetic beam ¯εcan be expressed in terms of the mean energy imparted by monoenergetic beams of energy E,¯ε(E), by integrating as follows: ¯ε(E) = RφE(E)¯ε(E) dE φE(E) dE(4.2) Similarly, the variance of the energy imparted σ2 εfor a polyenergetic beam can be expressed in terms of the variances of energy imparted for monoenergetic beams σ2 ε(E),¯ε(E)and ¯εas (38): σ2 εφ=Rφ(E)σ2 ε(E) dE Rφ(E) dE+Rφ(E) (ε(E)−εφ)2dE Rφ(E) dE(4.3) Therefore, by characterizing the values ¯ε(E)and σ2 ε(E)for monoenergetic beams it is possible to obtain results applied to polyenergetic, i.e. clinical, beams. To do so, monoenergetic beams imparting energy to a spherical site were simulated using Geant4-DNA (42, 43, 44, 45) with a code described in previous works (27, 38) and analytical functions for ¯ε(E)and σ2 ε(E)were developed, among other microdosimetric quantities. All of this allows for an analytical calculation of dose-averaged LET, ¯yFand ¯yDfor both monoenergetic and clinical beams. Here, we validate both the Geant4-DNA code used and the analytical functions to model ¯ε(E)and σ2 ε(E) for a set of experimentally available monoenergetic beams. These calculations are based on spherical sites whereas our silicon-based microdosimeter is represented by a cylindrical site of 5.5 µm in length and 16 µm in diameter whose axis is oriented in parallel to the particle tracks. In order to assess whether it is possible to compare spherical-based results with experimental measurements with this device, we have considered the concept of length-equivalent sphere. As ¯yFis specific to the mean chord length, the hypothesis of equal ¯yF for equal site mean chord lengths is tested. This would imply that mean energies imparted to length-equivalent sites are the same. In the sphere case, the mean chord length for a site with diameter dis given by ¯ l= 2d/3so that we consider a sphere with 8.25 µm in diameter, which corresponds to a mean chord length of 5.5 µm. Nonetheless, the distribution of lineal energy f(y)for that spherical site is expected to be wider than for its analogue cylindrical geometry due 69
to the variability in the chord length happening in the first case but not in the second one. This means that, even though ¯yFcoincided for both geometries, ¯yD would be larger for the spherical case than for the cylindrical one, as ¯yDnot only takes into account the mean energy imparted but also the variability in the distribution of energy imparted. According to Kellerer (46, 47), the impact on ¯yDof variability in the distribution of chord length in spherical sites can be approximated by a multiplicative factor given by the quotient between the mean and the weighted mean chord length for the spherical case. These means are defined as ¯ l=Rlf(l)dl and ¯ lD=Rl2f(l)dl/ Rlf(l)dl, respectively, being the chord length distribution f(l)=2l/d2for a spherical site with diameter d(39). The mentioned quotient is then ¯ l/¯ lD= 8/9for the spherical case. Thus, in this work, calculations of ¯yDcarried out with our analytical model for a spherical site of 8.25 µm in diameter already contains the multiplicative factor 8/9 to be comparable to the results obtained with cylindrical sites. 4.2.2 Experimental setup Proton monoenergetic microbeams of a set of different nominal energies (2.0 MeV, 2.7 MeV, 3.4 MeV, 4.0 MeV and 4.5 MeV) were generated at the Radiological Research Accelerator Facility (RARAF) of the University of Columbia. The experimental setup can be described as follows: vertical monoenergetic proton microbeams are magnetically collimated to several µm of diameter. Protons leave the vacuum system through a 2 µm-thick Mylar layer and a 2.89 µm-thick havar foil and traverse an air gap of 6 mm before reaching any device. There, a wheel is able to distribute the beam through three possible detectors independently (see Figure 4.1(a)), controlling the time of irradiation allocated to each one through the number of monitor counts. These three detectors are: (a) an ionization chamber in current mode that serves as a monitor chamber; (b) a solid-state detector (SSD) to measure the energy spectrum of the incoming beam; and (c) a TEPC in pulse mode, filled with gas to represent 6 µm tissue equivalency in a slab-like site that collects the electrons produced by incoming particles, i.e., measures events of energy deposition, ε. Lineal energy at the TEPC can be then obtained, according to equation (1), as y=ε/l =ε/(6 µm).Note that, as the TEPC is a slab perpendicularly traversed by protons, its thickness, 6 µm, corresponds to the single chord length for any track, with in turn coincides with the mean chord length. Besides the 6 mm air gap, the beam needs to travel through 6 µm more of Mylar to arrive at the TEPC, which produces a certain loss on the kinetic energy of the beam. 70
Figure 4.1: Pictures of the experimental setup for: (a) TEPC measurements and (b) silicon-based 3D microdosimeter measurements. Independently, the silicon-based 3D microdosimeter (32, 33, 35) was placed just at the exit of the 6 mm air gap (see Figure 4.1(b)) to measure the microdosimetric spectra of the nominal energies previously considered: 2.0 MeV, 2.7 MeV, 71
3.4 MeV and 4.5 MeV. This device represents a cylindrical microdosimetric site with dimensions of 5.5 µm in length and 16 µm in diameter. Since the pathway beams varies lightly from the TEPC to the microdosimeter, the actual kinetic energy for the particles conforming the beams is different at each detector. To overcome this, Monte Carlo simulations with the SRIM code (48) were performed to determine the real energy of the protons arriving at both TEPC and silicon-based 3D microdosimeter. 4.2.3 Calibration of the microdosimeter The customized readout-electronic system provides a calibration curve between ADC channels and energy imparted, given by ε(keV)=0.253 ×ADC Channel+26.161. However, that calibration has been performed for thicker silicon detectors and, therefore, a calibration readjustment can be required for small sensitive volumes as that of the microdosimeter used herein. That calibration correction factor was performed by employing the measurements done with the TEPC. For this purpose, the ADC channel corresponding to the measured peak lineal energy, ˆyTEPC, for each one of the employed beams was recorded so that four points (ADC channel, ˆyTEPC) were obtained. Also, the TEPC measures energy deposition in tissue-equivalent gas while the microdosimeter does it in silicon, so the following transformation is required in order to obtain the equivalent TEPC peak lineal energy value in silicon. As we deal with monoenergetic beams, the peak lineal energy referred to silicon, ˆySi, can be obtained by applying directly the correction factor regarding the stopping power ratio of silicon to tissue (28): ˆySi = ˆyTEPC ·SSi(E) STE(E)(4.4) where Eis the energy of the monoenergetic microbeam within the microdosimeter, and SSi(E)and STE(E)are, respectively, the stopping power of silicon and tissue for that energy, which we obtained from the PSTAR database of the National Institute of Standards and Technology (NIST). This simple transformation when mono-energetic beams are used for the calibration, as long as the stopping power ratio can be considered relatively constant along the active volume of the detector. Note that the same process for polyenergetic beams would require the knowledge of the energetic spectrum, which could be obtained by using telescopic detection systems (49). δ-ray equilibrium was assumed to be immediately achieved in both detectors as secondary electrons released in ionization collisions of protons with the energies considered in this work have ranges much lower than the longitudinal 72
dimension of both detectors. As mentioned above, the path that beams have to travel to reach the TEPC or the microdosimeter are different, so that the actual energy of the particles arriving to them from a beam is different. This means that a direct relation between peak energies as in equation (4.2) is not possible, since Echanges from one detector to the other. Our approach then to obtain the values ˆySi for the calibration correction was to use a cubic-spline interpolation of the corresponding value ˆyTEPC between the actual energies registered at the TEPC and then use equation (4.4). To obtain the peak energies ˆεSi to finally perform the calibration, according to equation (4.1), lineal energy is multiplied by the mean chord length in the microdosimeter, which is a cylinder of 5.5 µm in length oriented parallel to the beam axis, so that l= 5.5 µm. 4.2.4 Converting silicon readouts into tissue-equivalent Raw measurements from the calibrated microdosimeter provide f(εSi), i.e. the distribution of energy imparted in the silicon site, εSi. To obtain the corresponding distribution of lineal energy, f(ySi), it is necessary to divide εSi by the mean chord length in the site, l. However, the non-tissue equivalence of the silicon of which the cylindrical site is made needs to be corrected by yTE =ySi ·STE(E)/SSi(E)in a similar way as done in equation (4.4). 4.2.5 Calculating yFand yD Finally, the values for the averages of frequency and dose-weighted distributions for y can be calculated from the measured distribution referred to tissueequivalent as follows: yF=Pyf(y) Pf(y)(4.5) and yD=Pyd(y) Pd(y)=Py2f(y) Pyf(y)(4.6) To estimate the uncertainties for these two experimental results, we assume that the effect of the uncertainty in the TEPC peak ˆyTEPC is smaller than the uncertainty provided by the calibration curve for the energy deposited in silicon. This uncertainty can be estimated by means of the root mean square of the residuals between the data points from the TEPC and the fitted function. Then, it can be propagated according to equation (4.1) and added to the intrinsic uncertainty for 73
the expressions in equations (4.5) and (4.6) assuming a Poisson distribution for f(y). 4.2.6 Geant4-DNA simulations The same monoenergetic proton beams, with the actual energy arriving at the microdosimeter, were simulated in liquid water using an application developed in Geant4-DNA and converted to tissue-equivalent in a similar way than equation (4.4) shows in order to check the validity of the Geant4-DNA code employed to develop our analytical model. Instead of spherical sites, we used a cylindrical site with dimensions of 5.5 µm in length and 16 µm as basis diameter. Tracks for primary protons were originated from a point source to penetrate a liquid water box with an inner box to score energy deposition inside, extended with margins -upstream, downstream and lateralequal to the maximum range of secondary electrons for each beam according to Tabata’s formula (50). The cylinder axis is oriented parallel to the beam axis, as it happens in the experiment. The position for the cylinder was sorted uniformly in the inner box for each primary proton track, i.e., for each event to reproduce the situation in which the cylinder is irradiated by a uniform proton beam. The default physics list constructor of Geant4-DNA (45) was employed, which simulates protons and electrons down to 1 keV and 7.4 eV, respectively. Simulations of track structure included secondary electrons in the entire liquid water box. The cross sections for ionization and excitation processes were calculated using the First Born Approximation and the Emfietzoglou dielectric model (42, 45, 51). Geant4-DNA provides the user with various options for physics lists which incorporate different cross-section models and data relevant for electron transport. However, according to dose-point kernels calculated for monoenergetic electrons by Kyriakou et al. (52), it seems that their influence would become noticeable for sensitive volumes of the order of tens of nanometers Further details on the Geant4-DNA application can be found in previous works (27, 38, 53). Uncertainties for yFand yDare calculated following the procedure exposed in the Appendix A. 4.3 Results Table 4.1 shows the results obtained from the SRIM simulations to determine the beam energy arriving at each considered detector. Beam energies are higher at the entrance of the microdosimeter than the TEPC entrance as the thickness 74
particles need to go through is thinner in the former. The measured values for the peak of lineal energy at the TEPC, ˆyTEPC, for all these beams are also shown. Table 4.1: Mean energies at the entrance of the TEPC and the microdosimeter for each nominal energy according to simulations with the MC code SRIM. The last column shows the peak in terms of lineal energy ˆyTEPC measured at the TEPC. Nominal Energy (MeV) Mean energy at TEPC (MeV) Mean energy at microdosimeter (MeV) ˆyTEPC at TEPC (keV/µm) 2.00 1.39 1.55 21.73 2.70 2.18 2.30 15.22 3.40 2.92 3.01 11.98 3.50 4.04 4.12 9.67 Figure 4.2 shows the obtained calibration between ADC channels and energy imparted in silicon converted from measured peak lineal energies at the TEPC as described in section ‘Calibration of the microdosimeter’. The following relation is obtained by fitting a straight line to the measured data: ε(keV) = 0.456 ×C+ 13.55 (4.7) where Crepresents the ADC channel. The goodness of the fit can be assessed through the R2value, that yields the value 0.997. Figure 4.2: Calibration curve obtained for the relation between ADC channel and energy imparted into the silicon-based microdosimeter. The resulting fit yields the relation ε(keV) = 0.456 ×C+ 13.55, with R2= 0.997. 75
Measured distributions with the calibrated microdosimeter compared to the obtained from simulations with cylindrical sites in Geant4-DNA are shown in Figure 4.3 for beams with energies equal to the obtained in Table 4.1 to arrive at the microdosimeter. The distributions yd(y)is represented in logarithmic scale for the four beam energies considered at the entrance of the microdosimeter. Figure 4.3: Microdosimetric spectra in tissue represented as yd(y)in logarithmic scale for microdosimeter measurements and simulations with Geant4-DNA for beams with energies: (a) left upper: 1.55 MeV; (b) right upper: 2.30 MeV; (c) left bottom: 3.01 MeV; and (d) right bottom: 4.12 MeV. Values express the mean energy at entrance in the sensitive volume. Measured spectra in silicon with the microdosimeter are transformed into tissue by multiplying by the stopping power ratio of these media. 76
ological Systems to Low-and High-LET Radiations, pages 1–73. 2006. ISBN 9780444516435. doi: 10.1016/B978-044451643-5/50013-7. [19] R. D. Stewart et al. Effects of Radiation Quality and Oxygen on Clustered DNA Lesions and Cell Death. Radiation Research, 176(5):587–602, 2011. ISSN 0033-7587. doi: 10.1667/rr2663.1. [20] L. Lindborg, M. Hultqvist, ˚ A. Carlsson Tedgren, and H. Nikjoo. Lineal energy and radiation quality in radiation therapy: Model calculations and comparison with experiment. Physics in Medicine and Biology, 58(10):3089–3105, 2013. ISSN 00319155. doi: 10.1088/0031-9155/58/10/3089. [21] H. Nikjoo et al. Radiation track, DNA damage and response - A review. Reports on Progress in Physics, 79(11), 2016. ISSN 00344885. doi: 10.1088/ 0034-4885/79/11/116601. [22] D. Chmelevsky and A. M. Kellerer. Computation of microdosimetric distributions for small sites. Radiation and Environmental Biophysics, 14(2):123–136, 1977. ISSN 0301634X. doi: 10.1007/BF01332148. [23] C. Grassberger and H. Paganetti. Elevated LET components in clinical proton beams. Physics in Medicine and Biology, 56(20):6677–6691, 2011. ISSN 00319155. doi: 10.1088/0031-9155/56/20/011. [24] D. J. Brenner and M. Zaider. The Application of Track Calculations to Radiobiology: II. Calculations of Microdosimetric Quantities. Radiation Research, 98(1):pp.14—-25, 1984. ISSN 00337587. doi: 10.2307/3576047. [25] F. Villegas, N. Tilly, and A. Ahnesj¨ o. Target Size Variation in Microdosimetric Distributions and its Impact on the Linear-Quadratic Parameterization of Cell Survival. Radiation Research, 190(5):504–512, 2018. ISSN 0033-7587. doi: 10.1667/rr15089.1. [26] M. Newpower et al. Using the Proton Energy Spectrum and Microdosimetry To Model Proton Relative Biological Effectiveness. International Journal of Radiation Oncology •Biology •Physics, 104(2):316–324, 2019. ISSN 0360- 3016. doi: 10.1016/j.ijrobp.2019.01.094. [27] A. Bertolet et al. Dose-averaged LET calculation for proton track segments using microdosimetric Monte Carlo simulations. Medical Physics, 46(9):4184– 4192, jul 2019. ISSN 00942405. doi: 10.1002/mp.13643. 83
[28] L. Lindborg and A. Waker. Microdosimetry. Experimental Methods and Applications. CRC Press, 2017. ISBN 9781482217407. [29] V. Conte et al. Microdosimetry at the CATANA 62-MeV proton beam with a sealed miniaturized TEPC. Physica Medica, 64(June):114–122, 2019. ISSN 1724191X. doi: 10.1016/j.ejmp.2019.06.011. [30] D. Mazzucconi et al. Nano-microdosimetric investigation at the therapeutic proton irradiation line of CATANA. Radiation Measurements, 123(February): 26–33, 2019. ISSN 13504487. doi: 10.1016/j.radmeas.2019.02.012. [31] A. B. Rosenfeld et al. A new silicon detector for microdosimetry applications in proton therapy. IEEE Transactions on Nuclear Science, 47(4 PART 1):1386– 1394, 2000. ISSN 0018-9499. doi: 10.1109/23.872983. [32] C. Guardiola et al. Silicon-based three-dimensional microstructures for radiation dosimetry in hadrontherapy. Applied Physics Letters, 107(2):1–6, 2015. ISSN 00036951. doi: 10.1063/1.4926962. [33] C. Guardiola et al. Preliminary microdosimetric measurements with ultrathin 3D silicon detectors of a 62MeV proton beam. Journal of Instrumentation, 10(1):P01008, 2015. ISSN 17480221. doi: 10.1088/1748-0221/10/ 01/P01008. [34] S. E. Anderson et al. Microdosimetric measurements of a clinical proton beam with micrometer-sized solid-state detector. Medical Physics, 44(11): 6029–6037, 2017. ISSN 00942405. doi: 10.1002/mp.12583. [35] F. G´ omez et al. Measurement of carbon ion microdosimetric distributions with ultrathin 3D silicon diodes. Physics in Medicine and Biology, 61(11): 4036–4047, 2016. ISSN 13616560. doi: 10.1088/0031-9155/61/11/4036. [36] J. Prieto-pena et al. Microdosimetric spectra measurements on a clinical carbon beam at nominal therapeutic fluence rate with silicon cylindrical microdosimeters. IEEE Transactions on Nuclear Science, 66(7):1–1, 2019. ISSN 0018-9499. doi: 10.1109/tns.2019.2921453. [37] D. Bolst et al. Correction factors to convert microdosimetry measurements in silicon to tissue in 12C ion therapy. Physics in Medicine and Biology, 62(6): 2055–2069, 2017. ISSN 13616560. doi: 10.1088/1361-6560/aa5de5. 84
[38] A. Bertolet, A. Baratto-Rold´ an, M. A. Cort´ es-Giraldo, and A. Carabe- Fernandez. Segment-averaged LET concept and analytical calculation from microdosimetric quantities in proton radiation therapy. Medical Physics, 46 (9):4204–4214, 2019. ISSN 00942405. doi: 10.1002/mp.13673. [39] A. M. Kellerer. Chord-Length Distributions and Related Quantities for Spheroids. Radiation Research, 98(1):425–437, 1984. [40] G. Magrin. A method to convert spectra from slab microdosimeters in therapeutic ion-beams to the spectra referring to microdosimeters of different shapes and material. Physics in Medicine and Biology, 63(21), 2018. doi: 10.1016/j.pneurobio.2007.06.001. [41] A. M. Kellerer and H. H. Rossi. The Theory of Dual Radiation Action. In M. Ebert and A. Howard, editors, Current Topics in Radiation Research. Volume VIII, pages 85–156. American Elsevier Publishing Company, Manchester, U.K., 1974. [42] S. Incerti et al. Comparison of GEANT4 very low energy cross section models with experimental data in water. Medical Physics, 37(9):4692–4708, 2010. ISSN 00942405. doi: 10.1118/1.3476457. [43] S. Incerti et al. The Geant4-DNA project. International Journal of Modeling, Simulation, and Scientific Computing, 1(2):157, 2010. ISSN 1793-9623. doi: 10.1142/S1793962310000122. [44] M. A. Bernal et al. Track structure modeling in liquid water: A review of the Geant4-DNA very low energy extension of the Geant4 Monte Carlo simulation toolkit. Physica Medica, 31(8):861–874, dec 2015. ISSN 11201797. doi: 10.1016/j.ejmp.2015.10.087. [45] S. Incerti et al. Geant4-DNA example applications for track structure simulations in liquid water: A report from the Geant4-DNA Project. Medical Physics, 45(8):e722–e739, 2018. ISSN 00942405. doi: 10.1002/mp.13048. [46] A. M. Kellerer. Analysis of Patterns of Energy Deposition. In H. G. Ebert, editor, Second Symposium on Microdosimetry, pages 107–136, Stresa (Italy), 1970. [47] A. M. Kellerer. Fundamentals of microdosimetry. In K. R. Kase, B. E. Bjarngard, and F. H. Attix, editors, The Dosimetry of Ionization Radiation. Volume I, chapter Fundamenta, pages 77–162. Academic Press, Inc., 1985. 85
[48] J. F. Ziegler, M. D. Ziegler, and J. P. Biersack. SRIM – The stopping and range of ions in matter (2010). Nuclear Inst. and Methods in Physics Research, B, 268(11-12):1818–1823, 2010. ISSN 0168-583X. doi: 10.1016/j.nimb.2010. 02.091. [49] S. Agosteo et al. Study of a monolithic silicon telescope for solid state microdosimetry: Response to a 100 MeV proton beam. Radiation Measurements, 46(12):1529–1533, 2011. ISSN 13504487. doi: 10.1016/j.radmeas.2011.03. 040. [50] T. Tabata, R. Ito, and S. Okabe. Generalized semiempirical equations for the extrapolated range of electrons. Nuclear Instruments and Methods, 103:85– 91, 1972. [51] D. Emfietzoglou. Inelastic cross-sections for electron transport in liquid water: A comparison of dielectric models. Radiation Physics and Chemistry, 66(6): 373–385, 2003. ISSN 0969806X. doi: 10.1016/S0969-806X(02)00504-2. [52] I. Kyriakou, M. ˇ Sefl, V. Nourry, and S. Incerti. The impact of new Geant4-DNA cross section models on electron track structure simulations in liquid water. Journal of Applied Physics, 119(19), 2016. ISSN 10897550. doi: 10.1063/1. 4950808. [53] A. Baratto-Rold´ an et al. Abstract ID: 196 Relation between dose average linear energy transfer and dose mean lineal energy calculated for proton therapy beams off axis: A study with the Geant4 toolkit. Physica Medica, 42(2017): 42–43, 2017. ISSN 11201797. doi: 10.1016/J.EJMP.2017.09.104. [54] A. Bertolet, M. Cort´ es-Giraldo, K. Souris, and A. Carabe. A kernel-based algorithm for the spectral fluence of clinical proton beams to calculate doseaveraged LET and other dosimetric quantities of interest. Medical Physics, pages 1–11, 2020. ISSN 0094-2405. doi: 10.1002/mp.14108. 86
Chapter 5 A kernel-based algorithm for the spectral fluence of clinical proton beams to calculate dose-averaged LET and other dosimetric quantities of interest 87
Med. Phys. 2020, 47(6), 2495-2505. doi: 10.1002/mp.14108 A kernel-based algorithm for the spectral fluence of clinical proton beams to calculate dose-averaged LET and other dosimetric quantities of interest A. Bertolet1,2, M.A. Cort´ es-Giraldo2, K. Souris3, A. Carabe1 1Department of Radiation Oncology, Hospital of The University of Pennsylvania, Philadelphia, 19104 PA, USA 2Department of Atomic, Molecular and Nuclear Physics, Universidad de Sevilla, Seville, Spain 3Universit´ e Catholique de Louvain, Center for Molecular Imaging and Experimental Radiotherapy, Louvain, Belgium Purpose: To introduce a new analytical methodology to calculate quantities of interest in particle radiotherapy inside the treatment planning system. Models are proposed to calculate dose-averaged LET (LETd) in proton radiotherapy. Methods: A kernel-based approach for the spectral fluence of particles is developed by means of analytical functions depending on depth and lateral position. These functions are obtained by fitting them to data calculated with Monte Carlo (MC) simulations using Geant4 in liquid water for energies from 50 MeV to 250 MeV. Contributions of primary, secondary protons and alpha particles are modeled separately. Lateral profiles and spectra are modeled as Gaussian functions to be convolved with the fluence coming from the nozzle. LETd is obtained by integrating the stopping power curves from the PSTAR and ASTAR databases weighted by the spectrum at each position. The fast MC code MCsquare is employed to benchmark the results. Results: Considering the nine energies simulated, fits for the functions modeling the fluence in-depth provide an average R2equal to 0.998, 0.995 and 0.986 for each one of the particles considered. Fits for the Gaussian lateral functions yield average R2of 0.997, 0.982 and 0.993, respectively. Similarly, the Gaussian functions fitted to the computed spectra lead to average R2of 0.995, 0.938 and 88
0.902. LETd calculation in water shows mean differences of −0.007±0.008 keV/µm with respect to MCsquare if only protons are considered and 0.022 ±0.007 keV/µm including alpha particles. In a prostate case, mean difference for all voxels with dose greater than 5% of prescribed dose is 0.28 ±0.23 keV/µm. Conclusions: This new spectral fluence-based methodology allows for simultaneous calculations of quantities of interest in proton radiotherapy such as dose, LETd or microdosimetric quantities. The method also enables the inclusion of more particles by following an analogous process. 5.1 Introduction Analytical pencil beam algorithms in proton therapy are now widely used and accepted as a reliable method for dose calculation due to their accuracy and computational efficiency (1, 2, 3). This kind of algorithms use generally a model for the response (absorbed dose) for a proton beamlet, i.e., a kernel, to be convolved with the input beam, computed from a source beam model (4, 5). Similarly, the kernel approach is also commonly used for dose-averaged LET (LETd) calculations (6, 7, 8). Most of these kernel approaches, although keep accuracy in central axis calculations, are not able to describe accurately the lateral profile observed when calculating LETd in proton beams using MC. More generally, pencil beam algorithms are not as accurate as Monte Carlo (MC) techniques, which can be considered as the gold standard (6), especially when dealing with inhomogeneities (7, 8, 9). These shortcomings of the kernel based models makes necessary to add ad-hoc corrections or to use more complex models (9). Pencil beam algorithms also require dedicated functions for each quantity to calculate, doubling the workload if both dose and LETd need to be determined. Again, MC methods are able to overcome this limitation since in a single simulation it is possible to score several quantities in a volume of interest. However, the computation requirements for MC simulations in terms of time and resources are clearly higher than for analytical algorithms. In this work, a different approach is proposed to address calculations of different quantities of interest (QOIs) in proton therapy, and particularly, LETd. Instead of using specific quantity-depending kernels (such as dose kernels or LET kernels), a more general kernel for the spectrum of proton beams (in terms of kinetic energy or, equivalently, range of the particles) at each spatial point, i.e. spectral fluence, 89
is proposed. This allows for a more MC-alike approach since the information about the density of particles present and their energy is taken into account along their whole path. Then, simultaneous calculations for polyenergetic beams can be carried out by integrating the corresponding function on the energy for each QOI. Here, we present the particular results for LETd by using the stopping power data from NIST. Additionally, in previous works we have developed some functions dependent on the kinetic energy of protons to calculate dose-mean linear energy (yD) and LETd from microdosimetry by using MC simulations (10, 11). The method presented here also enables calculations of three-dimensional distributions of these microdosimetric QOIs. 5.2 Methods and Materials 5.2.1 General calculations of QOIs Stopping power for protons and other particles is generally available in terms of the kinetic energy of the particle from different databases (12, 13). Such energydependent functions need the knowledge of the particle spectrum in order to be used. In this way, if φ(r)is the fluence at a point r, given the spectral fluence φE(E;r)≡dφ(r)/dEat a certain point r, the average value for a generic energydependent function q=q(E)at that point will be given by q(r) = Rq(E)φE(E;r) dE RφE(E;r) dE(5.1) Possible simple examples of quantities based on electronic stopping power S(E) that can be calculated with equation (5.1) are: (i) dose, by introducing q(E) = S(E)/(Aρ)in the integration, where Ais the cross-sectional area of a voxel and ρis the physical density of water; or (b) track-averaged LET, LT, by simply exchanging q(E) = S(E). Furthermore, more complex quantities can be calculated as well, such as dose-averaged LET (LD), by computing separately equation (1) with q(E) = S2(E)and q(E) = S(E)and dividing the final results. The latter example is used in this work to show results. Therefore, this work is centered on how to determine φE(E;r)as this quantity allows for a variety of different calculations of interest. In equation (5.1), the spectral fluence φE(E)acts as a weight for each spectral component in the energy-dependent function q(E). Equivalently, if the relation between energy and range in the continuous slow-down approximation (CSDA) is known, R=R(E), then the average value for qat a point rcan be alternatively 90
computed as q(r) = Rq(R)φR(R;r) dR RφR(R;r) dR(5.2) where φR(R;r)≡dφ(r)/dRis the fluence range-density, i.e. the relative number of particles per cross-sectional unit area with CSDA range equal to Rwith respect to the total number of particles per cross-sectional unit area at r. This second approach is preferred because of practical reasons. Concretely, the variation of stopping power and other QOIs with the CSDA range is smoother than with the particle energy, assuming as this relation R(E) = αEp, i.e. the Bragg-Kleeman rule (14). Therefore, if the energy-dependent functions qj(E)and the transform rule to CSDA range R=R(E)are known, according to equation (5.2), our problem is reduced to the determination of φR(R;r)at each point of interest. 5.2.2 Application to clinical beams A clinical beam is generally composed of the superposition of several layers, being each one of them calculated independently. In a pencil beam scanning (PBS) system, a beam is composed of spots that are magnetically scanned to achieve a conformal dose distribution. Note that a layer is here defined by a nominal energy, a reference system (beam angle and isocenter position) and a two-dimensional map of spot-fluence weights. In this context, φR(R;r)is the total resulting fluence range-density and is obtained as a combination of the fluence range-density for each beam, which, in turn, is a combination of the individual fluence coming from each spot. Let NLbe the number of layers present in a given beam and φR,i(R;r) the fluence range-density for the i-th layer. If φi(r)is the total fluence that the i-th layer carries to the point r, then φR(R;r) = PNL i=1 φi(r)φR,i(R;r) PNb i=1 φi(r)(5.3) Eq. (5.3) represents a weighted average for the different spectral fluence converging at a point r, in which the weights are their corresponding fluences. Consequently, it becomes necessary to determine the number of particles per area unit carried by each layer at each point. This determination, however, is a problem frequently solved in pencil beam algorithms as follows. A treatment planning system (TPS) with pencil beam scanning (PBS) system, should provide the following information of a layer: (i) its nominal energy with a certain initial spread τini; and (ii) the lateral width of the spots σlat,spot(z)on that layer at a certain reference depth, normally the isocenter (15). In this sense, φR,i(r)can be obtained by summing the 91
contributions from every individual spot composing it and, in turn, that contribution for each spot is obtained from the composition of different infinitely narrow beam, usually called beamlets. Then, it is convenient to separate the problem of determining φbeamlet(r)for each beamlet into two different functions: one depending only on the depth in water (integrated fluence, i.e., total number of particles at each depth) and the other depending on the lateral distance to the beam propagation axis ρat each depth z,φbeamlet(r) = (Φint(z)/Φ0)φlat,beamlet(ρ;z), where Φint(z)/Φ0 represents the fraction of particles arriving at each depth with respect to the total number of particles coming from the water surface. Now, σlat,spot(z)refers to the width of the spot’s lateral profile for the fluence at a depth zfor a layer, which becomes larger as it goes away from the source due to the beam divergence. Furthermore, when those particles travel through matter (particularly, water), they undergo scattering processes thus the lateral distribution of particles becomes broader. This can be incorporated by generating a lateral kernel for a beamlet, which, essentially, represents the probability density for a particle to reach each lateral position. In other words, if at a certain depth z the probability density for a single particle to be scattered up to a lateral distance ρis given by Kw(ρ;z) = φlat,beamlet(ρ;z), then the fluence for a spot at that point will be given by the convolution φspot(r) = Φint(z) Φ0Zφlat,spot(ρ0;z)Kw(ρ0−ρ;z) dρ0(5.4) Here, again, φlat,spot(ρ;z)represents the fluence for each spot coming from the nozzle taking into account the beam divergence and it is usually provided by the TPS and modeled as a single Gaussian function thus φlat,spot(ρ;z) = exp(−ρ2/2σ2 lat,spot(z)). Finally, the fluence for the overall i-th layer is obtained by convolving again the fluence for a single spot at each depth, φspot(r), with its twodimensional map of spot weights produced by the PBS system, w(x, y): φi(r) = Zφspot(x0, y0;z)w(x−x0, y −y0) dx0dy0(5.5) where now lateral coordinates have been transformed from cylindrical to Cartesian. This way equation (5.5) provides the weights in the weighted average of spectral fluences shown in equation (5.3). 5.2.3 Beam components Generally, we can distinguish between two different types of particle components present in a beam according to their different behavior in depth and laterally. In this work, we use the distinction provided by the MC package Geant4 92
for the functions shown in equations (5.7), (5.9) and (5.10) are also shown. Results for only three energies (125 MeV, 175 MeV and 225 MeV) are shown to ease the visualization. Similar trends are obtained for the rest of the energies. As a measurement of the goodness of the fits, for primary protons the nine simulated energies provide a mean R2= 0.998, for secondary protons R2= 0.995 and for alpha particles R2= 0.986. Figure 5.1: Fraction of (a) primary protons, (b) secondary protons and (c) alpha particles in depth respect to the number of protons arriving at water surface (N0). Dashed lines represent the results obtained from Geant4 simulations for protons of 125 MeV (blue), 175 MeV (red) and 225 MeV (black). Solid lines are the fits obtained with the functions shown in equations (5.7), (5.9) and (5.10) for the panels (a), (b) and (c), respectively. Legends: (a) Geant4 simulation; Equation (5.7); (b) Geant4 simulation; Equation (5.9); (c) Geant4 simulation; Equation (5.10). Lateral distributions for primary protons and secondary alphas are modeled 99
as single Gaussian functions at every individual depth considered. Whereas, for secondary protons, a double Gaussian function is employed. The goodness of these fits can be evaluated through the mean R2for each one of the nine energies employed. The averages and standard deviations of these sets of nine mean R2are 0.997 ±0.0004 for primary protons, 0.982 ±0.0007 for secondary protons and 0.993 ±0.003 for alpha particles. Fig. 5.2 shows the dependency of the lateral σ for each case for the same three selected energies than in Fig. 5.1. Figure 5.2: Parameters to characterize the lateral distribution for (a) primary protons, (b) secondary protons and (c) alpha particles for three different energies: 125 MeV (blue), 175 MeV (red) and 225 MeV (black). Polynomial functions are fitted to the points, obtained, in turn, by fitting a single Gaussian functions in the cases (a) and (c) and Double Gaussian functions for the case (b) to the lateral profile at each independent depth. These parameters correspond to those in equations (5.11), (5.13) and (5.14) and are used to build the lateral kernels at each depth and for each energy. Legends: (a) σlat,P P obtained from Single Gaussian fit to Geant4 data; Polynomial fit, Equation (5.11); (b) σl1,SP and wSP obtained from Double Gaussian fits to Geant4 data; σl2,SP obtained from Double Gaussian fits to Geant4 data; Polynomial fits; (c) σlat,AP obtained from Single Gaussian fit to Geant4 data; Polynomial fit. 100
The spectrum in terms of CSDA range for each component is modeled as a single Gaussian function. Fig. 5.3(a) shows an example of the spectrum obtained from Geant4 MC simulations and the fitted functions (beam energy of 150 MeV at 10 cm in depth). The overall goodness of this fits is evaluated through the average of the mean R2parameters obtained for each energy along all the considered depths: 0.995 ±0.001 for primary protons, 0.938 ±0.025 for secondary protons and 0.90 ±0.02 for alpha particles. Fig. 5.3(b) shows the dependency of the mean range on the depth for the same beam energies. Note that the relation between CSDA range and kinetic energy is different for protons and for alpha particles. Fig. 5.3(c) shows the standard deviation for the Gaussian functions for the same beam than in Fig. 5.3(a) and (b). Figure 5.4 shows the comparison between the calculations for LETd by using the fluence kernel method exposed in this work and by MCsquare only considering primary and secondary protons. Results at the central axis are shown in Fig. 5.4(a) whereas Fig. 5.4(b) and (c) show two lateral profiles at different depths. The mean difference between two results along the central axis, taking samples separated by 1 mm, is -0.29 ±0.15 keV/µm, where uncertainty is expressed as one standard deviation of the mean difference. For the whole three-dimensional distribution, we find a mean difference of -0.007 ±0.008 keV/µm. Points with a number of particles less than 1% of the total number of particles coming for each layer are excluded from this computation. The same case but adding the alpha particles’ contribution to the fluence kernel and all the secondary heavier particles in MCsquare is shown in Figure 5.5. In this case, the mean deviation at the central axis is found as -0.24 ±0.14 keV/µm, meanwhile for the complete distribution it now goes to -0.022 ±0.007 keV/µm. Results of the calculations for LDin a prostate case are shown in Figure 5.6. Besides the LDdistributions calculated with MCsquare and our method and their differences, dose distribution obtained with MCsquare is also provided to contrast areas in which LDmay result high but dose is very low. Mean difference between MCsquare and our method in all the volume irradiated above 5% of the dose is equal to 0.28 keV ±0.23 keV/µm. However, for the voxels contained in the PTV, this mean difference decreases to 0.07 ±0.11 keV/µm. In the rectum, in turn, mean difference is 0.0 ±0.8 keV/µm. 101
Figure 5.3: (a) Examples of spectra for primary protons, secondary protons and alpha particles in terms of CSDA range for a single-layer proton beam of 150 MeV at a depth equal to 10 cm. (b) Mean of the spectra shown in (a) in depth for each type of particle for a single-layer proton beam of 150 MeV. For primary protons (left), this is modeled as the residual range, whereas for secondary protons (middle) and alpha particles (right) a third-grade polynomial function is used. (c) Standard deviation of the spectra shown in (a) considered as a Gaussian function for a single-layer proton beam of 150 MeV for primary protons (left), secondary protons (middle) and alpha particles (right). Legends: (a) Geant4 simulation; Single Gaussian fits; (b) Mean µR,P of the Gaussian function fitted to Geant4 simulations; Polynomial fits; (b) Sigma σR,P of the Gaussian function fitted to Geant4 simulations; Polynomial fits. 102
Figure 5.4: Calculations of LETd considering only protons with our fluence kernelbased method and with MCsquare (109protons simulated) for a spread-out Bragg peak with maximum range of 20 cm and 10 cm of modulation at (a) the central axis, (b) a lateral profile at depth = 11.5 cm and (c) another lateral profile at depth = 17.44 cm. The positions of the lateral profiles are marked in (a) with gray vertical lines. Dose calculations with MCsquare are shown in gray dashed lines. Legends: Fluence kernel-based calculation; MCsquare. 5.4 Discussion Results for LETd calculation show a good performance of our fluence model, especially when only protons are considered (see Fig. 5.4), in which not only central axis profiles tend to coincide but also the characteristic lateral curvature is met at any depth. We have chosen specifically LETd to assess our method for two reasons: (i) the values for the stopping power are extensively validated and (ii) it is a quantity strongly dependent on the spectrum and the relative number of particles from each category rather than on the overall number of particles, 103
Figure 5.5: Calculations of LETd considering protons and alpha particles with our fluence kernel-based method and a specially compiled version of MCsquare that includes only primary protons, secondary protons and secondary alpha particles (109 primary particles simulated) for a spread-out Bragg peak with maximum range of 20 cm and 10 cm of modulation at (a) the central axis, (b) a lateral profile at depth = 11.5 cm and (c) another lateral profile at depth = 17.44 cm. The positions of the lateral profiles are marked in (a) with gray vertical lines. Dose calculations with MCsquare are shown in gray dashed lines. Legends: Fluence kernel-based calculation; MCsquare. which would be the case of quantities as energy or dose absorbed. Because of that dependency, every part of the method needs to be correctly modeled so that acceptable agreements with MC calculations can be found. Even though our methodology uses convolution-superposition principles, as pencil beam algorithms do, the main difference here resides in the quantities to be superposed. While dose or LETd-specific kernels directly convolve and superpose curves of those quantities, we sum, on the one hand, the number of particles produced by each beamlet in a spot, and average, on the other hand, the mean 104
Figure 5.6: Calculations on a central axial plane for prostate case with two lateral beams and a prescribed dose of 70.2 Gy. (a) LDcalculated with MCsquare. (b) LD calculated with our method. (c) Difference between LDcalculated with MCsquare and with our method. (d) Dose distribution calculated with MCsquare. For (a), (b) and (c), color bars are specified in keV/µm. For (d), values on the color bar indicate Gy. energy of the particles traveling in that beamlet. These quantities can naturally be superposed under any condition whereas for dose, some assumptions need to be done: the energy deposition needs to be concentrated close around the particle track. The case of LETd is even worse: the non-linear component on the squared stopping power cannot be superposed in this way and artificial workarounds need to be incorporated to the lateral kernels. However, some major deviations from the fluence data computed by Geant4 are observed in some cases. This is especially the case for the number of alpha particles in depth, as seen in Fig 5.1(c), in which equation (5.9) appreciably deviates from the curve obtained from Geant4 around the end of the beam range. Nonetheless, the impact of these deviations is reduced since, as protons’ stopping power increases in that area, the relative contribution to the LETd of alpha particles becomes small. Our models for secondary protons do not seem accurate at 105
the entrance region (see Fig. 5.2(b) and Fig. 5.3(b) and (c)). This can be explained by the small number of secondary protons produced there (see Fig. 5.1(b)), which makes unstable the Gaussian fits and, in turn, makes the values erratic for the parameters. Yet, precisely due to its relatively small number, the weight of these deviations in the calculations is limited. Finally, the relatively low R2values for the fits of Gaussian functions to the alpha particles are acceptable as a compromise for a simple meaningful function with few parameters, with the mean allowed to vary laterally. We have included alpha particles in our formalism as they represent the main secondary contributor, besides protons, in clinical proton beams. Deuterons, for example, contributes to dose deposition by less than 0.1%, roughly an order of magnitude less than alpha (19). However, deuterons and other particles might be incorporated following the same methodology to develop kernels for their fluence and spectral fluence. The contribution of alpha particles makes LETd increase especially at the entrance region, although our method seems to slightly underestimate this compared to MCsquare (see Fig. 5.5). However, the general agreement is remarkable and the entrance area is not expected to be particularly relevant to assess biological effect since LETd is relatively low there. In any case, as shown with alpha particles as an example, the contribution of any other particle can be added to the model by following a similar process: (i) obtaining the 3D distributions (number of particles and energy spectrum) for production of that particle in proton beams of a range of energies; (ii) model the functions for the number of particles in depth and laterally; and (iii) model the functions for the spectrum at the central axis and laterally. It is important to remark that the method to consider inhomogeneous media here presented is a simple displacement of the calculated fluence according to the water-equivalent depth. This may result in errors for the lateral contribution of each beamlet to an adjacent point, especially around edges between two media highly inhomogeneous. Therefore, a more sophisticated method to overcome these situations is desirable. However, as shown in Figure 5.6, average differences in LDcalculation as a surrogate of spectral fluence do not tend to be high, especially in areas in which dose, i.e. fluence, is large. Only LETd calculations are shown in this work as an example of the potential of this method. It should be noted, though, that an advantage here is the ability to simultaneously calculate a number of QOIs. For instance, if a function for the dependency of the dose on the energy of a particle, D=D(E), can be found, 106
then the same framework can be used to compute dose and LETd, which may lead to an efficient optimization process based on distributions of both quantities. Additionally, as mentioned before, our published microdosimetric functions (11) can be employed in order to generate three-dimensional distributions for them in commercial TPS. 5.5 Conclusion A new analytical method to reproduce the three-dimensional distributions of spectral fluence in particle therapy is presented and particularized for the proton therapy case. The advantage of this spectral fluence-based approach is that several quantities of interest can be derived from the spectral fluence by means of a convolution integral as long as the dependency of those with the energy of the involved particles is known. As a specific application, LETd results with this approach look remarkably good even though only protons and alpha particles are taken into consideration. The philosophy of the method includes the potential of extending the considered particles by following a process similar to presented here. References [1] L. Hong et al. A pencil beam algorithm for proton dose calculations. Physics in Medicine and Biology, 41:1305–1330, 1996. [2] M. Soukup, M. Fippel, and M. Alber. A pencil beam algorithm for intensity modulated proton therapy derived from Monte Carlo simulations. Physics in Medicine and Biology, 50(21):5089–5104, 2005. ISSN 00319155. doi: 10. 1088/0031-9155/50/21/010. [3] E. Pedroni et al. Experimental characterization and physical modelling of the dose distribution of scanned proton pencil beams. Physics in Medicine and Biology, 50(3):541–561, 2005. ISSN 00319155. doi: 10.1088/0031-9155/ 50/3/011. [4] W. Ulmer. Theoretical aspects of energy-range relations, stopping power and energy straggling of protons. Radiation Physics and Chemistry, 76(7):1089– 1107, 2007. ISSN 0969806X. doi: 10.1016/j.radphyschem.2007.02.083. 107
[5] P. Kimstrand et al. A beam source model for scanned proton beams. Physics in medicine and biology, 52(11):3151–68, 2007. ISSN 0031-9155. doi: 10. 1088/0031-9155/52/11/015. [6] J. J. Wilkens and U. Oelfke. Analytical linear energy transfer calculations for proton therapy. Medical Physics, 30(5):806–815, 2003. ISSN 00942405. doi: 10.1118/1.1567852. [7] F. Marsolat, L. De Marzi, F. Pouzoulet, and A. Mazal. Analytical linear energy transfer model including secondary particles: Calculations along the central axis of the proton pencil beam. Physics in Medicine and Biology, 61(2):740– 757, 2016. ISSN 13616560. doi: 10.1088/0031-9155/61/2/740. [8] D. Sanchez-Parcerisa et al. Analytical calculation of proton linear energy transfer in voxelized geometries including secondary protons. Physics in Medicine and Biology, 61(4):1705–1721, feb 2016. ISSN 0031-9155. doi: 10.1088/0031-9155/61/4/1705. [9] S. Hirayama et al. An analytical dose-averaged LET calculation algorithm considering the off-axis LET enhancement by secondary protons for spotscanning proton therapy. Medical Physics, 45(7):3404–3416, 2018. ISSN 00942405. doi: 10.1002/mp.12991. [10] A. Bertolet et al. Dose-averaged LET calculation for proton track segments using microdosimetric Monte Carlo simulations. Medical Physics, 46(9):4184– 4192, jul 2019. ISSN 00942405. doi: 10.1002/mp.13643. [11] A. Bertolet et al. Calculation of clinical dose distributions in proton therapy from microdosimetry. Medical Physics, 46(12):5816–5823, oct 2019. ISSN 00942405. doi: 10.1002/mp.13861. [12] M. Berger, J. Coursey, M. Zucker, and J. Chang. ESTAR, PSTAR and ASTAR: Computer Programs for Calculating Stopping-Power and Range Tables for Electrons, Protons, and Helium Ions (version 2.0.1), 2017. [13] ICRU. Report 90: Key Data for ionizing radiation dosimetry: measurement standards and applications. Technical Report 90, 2016. [14] W. Ulmer and E. Matsinos. Theoretical methods for the calculation of Bragg curves and 3D distributions of proton beams. European Physical Journal: 108
The average energy imparted to a site at each particle energy depends on three factors: (a) the stopping power of the particle, (b) how long is the particle track within the site -segment length-, and (c) the probability that secondary electrons generated by the particle leave the site. The trial function for εs(E)should be, firstly, proportional to the stopping power, which in turn depends on the particle energy. It also should contain a term to consider the losses of energy imparted when the range of secondary electrons is larger than the site dimensions. And finally, it needs to decrease as the particle track becomes shorter. In previous chapters we have explored how to use Monte Carlo simulations to produce microdosimetric data (13) and what functions can be used to characterize microdosimetric quantities (14). A function meeting all the referred requirements is: εs(E) = C·erf(kEq)·log(hP,mE+b)αP,m ·E(pP,m−1) +e αP,m ·EpP,m + 2Ds/3(6.5) where C,band eare free parameters of the function whereas hP,m,αP,m and pP,m depend on the considered particle and medium and Dsdepends on the site dimension. The mass of the site can be obtained from its volume and the density of the considered medium. Since dose to water is the quantity normally used in a TPS for dose calculations, the site is considered to be made of liquid water. 6.2.4 Number of events per energy, ν(E): a model for the voxel-site relation By definition, the distribution of average number of events per energy ν(E) should coincide with the spectral distribution at the site, i.e., the number of particles of each energy impinging on the site. However, we are normally interested in calculating dose in a voxelized geometry, being the voxel considerably larger than the typical microscopic sites. Therefore, we need to firstly consider the number of events occurring in the voxel. For this, we assume the voxel is represented by an arrangement of sites in a rectangular grid, as Figure 6.1 shows for the case of spherical sites. Let Ntbe the number of sites in the voxel contained on a layer perpendicular to the particle track, where t stands for transversal. Then, if the transversal cross section of a site is σs, the total cross section in which events are produced is Ntσsso that the number of events in the voxel is proportional to φE(E)Ntσs, where φE(E)is the spectral fluence of particles. Note that this is valid as long as we can assume that the number of indirect events (i.e. produced just by secondary particles) is negligible with respect to the number of direct events 115
(i.e. produced by primary protons), which is expected to be the case for any voxel with dose relevant in radiotherapy. Figure 6.1: Model for the internal structure of a voxel composed by spherical sites. Protons characterized by a given fluence density impinge on the frontal face of the voxel and particles interact with the row of sites along their track. An additional aspect needs to be taken into consideration. Let dVbe the dimension of the voxel parallel to the direction of the particle tracks and dsthe maximum dimension of the site in the same direction. Then, the number of sites aligned within the voxel will be Nl=dV/ds, where lstands for longitudinal. If the incidental particle energy is such that its range is longer than dV, then it will exit the voxel producing an event in each site on its way. Therefore, the number of events will be ν=Nl. However, if the range of the particle is shorter than dV, the number of events produced by that particle is fewer since the proton stops before reaching the final sites. We assume, in that case, that the number of events produced by that particle is proportional to its CSDA range, R(E), so the number of events in the voxel per energy, νV(E), is: νV(E) = φE(E)Ntσsmax dV ds ,R(E) ds (6.6) where dxemeans the least integer greater than or equal to x, since the number of events is an integer quantity. However, according to equation (6.2), we are interested in the average number of events per site rather than in the complete voxel. As the total number of sites in the voxel is NtNl, the average number of events per 116
site is given by ν(E) = νV(E) NtNl =1 Nl φE(E)σsmax dV ds ,R(E) ds (6.7) 6.2.5 A particular case: protons in spherical sites All shown so far remains as valid regardless the particle species, and the site’s shape or dimension considered. Now, we introduce a particular case to illustrate the performance of the method. The analysis beyond this point is focused on proton clinical beams and spherical sites with diameter d. For this specific case, the longitudinal length of the site equals to the site diameter, ds=d, and so it does the site-dependent parameter on the expression for εsin equation (6.5)Ds=d(14). In order to obtain the rest of parameters in equation (6.5), we performed simulations with Geant4-DNA (17, 18, 19, 20), of mono-energetic protons in spherical sites made of liquid water up to 100 MeV, which is the upper limit the code has for proton transport. To assess the influence of the site dimension in the calculation, we have considered three different cases for the site diameter: 1 µm, 5 µm and 10 µm. To calculate εs(E), for proton energies higher than 100 MeV we assume the validity of equation (6.5) can be extrapolated to those energies, as our trial function has a smooth behavior at such energies. 6.2.6 Kernel for the spectral fluence of proton beams Once εs(E), and therefore zF, are obtained, the only quantity remaining to be determined is the spectral fluence density, φE(E). Again, this means the number of protons of each energy per unit area traversing each voxel in which the CT image of a patient is divided. This information can be obtained through Monte Carlo simulations of the proton transport in the patient. Some TPS are able to provide that information (21, 22). However, we have developed a kernel-based algorithm for the spectral fluence at each point given a proton arriving to the patient surface. Here we provide an outline of this, although it is fully presented in Chapter 5(23). Different simulations of the transport of protons in water were carried out by using an application based on the Monte Carlo toolkit Geant4 (24, 25, 26). Several values for the energy have been selected in the range of clinical use in proton radiotherapy: from 50 MeV up to 250 MeV. A distinction between primary and secondary protons is made to develop independent models for their transport in water. Primary protons are those which only undergo elastic collisions with the atoms of matter meanwhile secondary protons 117
are those coming from inelastic collisions with nuclei. For each one, the model is composed by three elements: (i) given an initial fluence of protons Φ0penetrating in water, the fraction of primary ΦP P (d)/Φ0and secondary protons ΦSP (d)/Φ0in a whole transversal plane at a given depth d; (ii) the lateral distribution of primary dφP P (r;d)/drand secondary protons dφSP (r;d)/dr at each depth and each distance rto the central axis of the beam; and (iii) the energy spectrum for primary dφP P (E;r, d)/dEand secondary dφSP (E;r, d)/dEat each lateral position and each depth. For the first case, we use the function proposed in the work of Ulmer (27) for the primary protons, whereas the proportion of secondary protons at each depth ΦSP (d)is modeled by a phenomenological function (23): ΦSP (d) Φ0 = (ad −bdc)1 21 + erf R(E)−d √2τ (6.8) where a,band care parameters to be fitted dependent on the beam energy, R(E) is the CSDA range corresponding to the initial beam energy and τis the same parameter than used by Ulmer (27), related to the range straggling of the beam. For the lateral distribution, we assume that a single Gaussian centered at the beam axis is enough to model the spread in the fluence of primary protons, but not for the secondary ones. Hence, a Double Gaussian is proposed as a better function to model the latter case. Finally, for the sake of simplicity, the energy spectrum for both primaries and secondaries is modeled as a single Gaussian whose mean energy is given by the residual range (i.e., R(E)−d). Then, the kernel of spectral distribution at a voxel with entrance surface dSat the position (d, r)is given by φKernel E(E;d, r) dS=X i=P P,SP Φi(d) Φ0 dφi(r;d) dr dφi(E;r, d) dEdS(6.9) where irefers to protons and secondary protons. Note that both dφi(r;d)/drand dφi(E;r, z)/dEare normalized to the unit so that the argument in the summation provides the probability for a proton to have an energy Eat the position (d, r). Finally, to obtain the spectral fluence density for a broad beam it is necessary to convolve two-dimensionally the spectral fluence density of the broad beam coming onto the patient surface, φSurface E(E;r), with the kernel presented in equation (6.9): φE(E;d, r) = ZφSurface E(E;r0)φKernel E(E;d, r −r0) dr0(6.10) This model for the proton transport only considers liquid water. Transport through other media is here considered by replacing in equation (6.9) the depth in water, d, by the corresponding water-equivalent thickness dW ET resulting from the beam raytracing. 118
6.2.7 Dose distributions comparisons All above have been implemented in a script for the Eclipse TPS v15.6 (Varian Medical Systems, Palo Alto, CA, USA) for Proton Beam Scanning (PBS) treatments, called hereinafter ‘MicroCalc’ or ‘µCalc’. This allows us for producing actual dose calculations in three-dimensional geometries. The TPS provides the information about the beam lateral width –to be convolved with our lateral model–. On the other hand, the number of protons prior to arrive to the patient surface can be estimated according to the number of monitor units (MU) for each spot (21) so that we can obtain φSurface E(E;r). We present comparisons of dose distributions obtained by: (a) the microdosimetric method presented in equations (6.3)-(6.7) together with the kernel for the proton fluence summarized in equation (6.9); (b) the analytical Proton Convolution-Superposition (PCS) algorithm (28) provided by Eclipse; and (c) Monte Carlo simulations with the open source fast MC code MC- square (29), which is used as a benchmark in this work as it has been validated elsewhere (30, 31). We have considered three different situations with different complexities: (1) single energy beams broad enough to provide lateral equilibrium at the central axis, computed in water; (2) a single beam conforming a spread-out Bragg peak (SOBP) with energy layers from 118 MeV to 175 MeV in a water box; and (3) clinical plans for a brain cancer case with three beams and a prostate case with two lateral beams. In the last cases, global 3D gamma analyses in the PTV structures with interpolation are performed to evaluate the differences between analytical methods (MicroCalc and PCS) and the MC calculation (MCsquare) with the MATLAB-based environment OpenREGGUI. The criterion for those gamma analyses is (3%, 3 mm) considering the MCsquare distribution as the reference in all cases. To evaluate the accuracy of the dose calculations in organs at risk, we select an adjacent structure to the PTV in each case to compute differences with respect to MCsquare: brainstem in the brain case and rectum in the prostate case. 6.3 Results Three different beams are considered to show their dose profiles at central axis in Fig. 6.2 single energy layers of 110 MeV, 160 MeV and 210 MeV. Fig. 6.3 shows the dose profiles at the central axis for a beam with energy layers from 118 MeV to 175 MeV optimized in EclipseTM to produce a uniform dose distribution in a target of 10 cm in length and whose distal edge is placed at 20 119
Depth (cm) 5 10 15 20 25 30 Dose (a.u.) 0 0.5 1 1.5 2 2.5 3 3.5 110 MeV 160 MeV 210 MeV Figure 6.2: Dose profiles at the central axis for single energy layers beams with energies 110 MeV, 160 MeV and 210 MeV calculated by employing the MicroCalc script, the PCS algorithm in Eclipse and the MC code MCsquare. Legends: µCalc; PCS Algorithm; MCsquare. cm in depth. The mean absolute difference along the profile between MCsquare and MicroCalc is 0.86% while the mean difference between MCsquare and PCS algorithm is 1.03%. A clinical case for brain cancer is presented in Fig. 6.4. Dose distributions for MCsquare, MicroCalc and PCS on an axial plane located at the middle of the PTV are shown in Fig. 6.4(a), (b) and (c), respectively. Differences between MC- square and (i) MicroCalc and (ii) PCS can be seen in Fig. 6.4(d) and (e), respectively. Three-dimensional gamma analysis with criterion (3 mm, 3%) in the PTV between MicroCalc and MCsquare shows a gamma index of 95.07% while the gamma analysis for differences between PCS and MCsquare is 94.50%. The same procedure in the CTV produces results of 95.10% and 95.03%, respectively. Mean differences in the dose distribution for the brainstem structure are found to be 0.25 Gy between MicroCalc and MCsquare and 0.56 Gy between PCS and MCsquare, respectively. A similar analysis is shown in Figure 6.5 for a prostate case. Results for dose distributions for an axial plane, located at the middle of the PTV are shown for MC- square, MicroCalc and PCS calculations, respectively. Again, 3D gamma analysis in the PTV contour with criterion (3%, 3mm) is performed by comparing MCsquare to MicroCalc and PCS, respectively. The gamma index is 99.90% for the comparison MCsquare-MicroCalc and 100% for the comparison MCsquare-PCS. In the rectum, the mean differences in dose between MicroCalc or PCS and MCsquare are 0.05 Gy and 0.07 Gy. 120
Depth (cm) 2 4 6 8 10 12 14 16 18 20 22 Dose (a.u.) 0 0.2 0.4 0.6 0.8 1 1.2 Figure 6.3: Dose profiles at central axis for a spread-out Bragg peak with maximum range at 20 cm in depth and modulation of 10 cm calculated by employing the MicroCalc script, the PCS algorithm in Eclipse and the MC code MCsquare. Legends: µCalc; PCS Algorithm; MCsquare. 6.4 Discussion Whereas the number of results presented here is not large enough to assess the significance of the goodness for MicroCalc dose calculation, the performance of the proposed model is found comparable to the differences found between MC- square and the clinically validated algorithm PCS. The ability to reproduce dose ratio between Bragg peak and plateau zone, as well as the beam range, is clearly observable in Fig. 6.2. When not considering only pristine Bragg peaks, a better agreement with MCsquare is found for MicroCalc at the SOBP than PCS. This better agreement shows up both in the mean difference along the profile at the central axis and the shape of the curve itself, slightly non-uniform as opposed to the PCS algorithm shows. Results shown in Fig. 6.3 are reproduced in both clinical cases presented (Fig. 6.4 and Fig. 6.5). Comparable (3%, 3mm) gamma index are found 121
Figure 6.4: Dose distributions on an axial plane for a brain case for (a) MCsquare, (b) MicroCalc and (c) PCS algorithm. The yellow contour depicts the PTV, with prescription dose of 54 Gy. On the color map, blue color represents low dose (near 0 Gy) while most intense red color represents 54 Gy. Differences between (d) MicroCalc and MCsquare and (e) PCS and MCsquare are also shown. In these color maps, most intense blue shows MCsquare dose is higher by 5 Gy, most intense red means MCsquare dose is lower by 5 Gy and translucent means no difference. for the 3D analysis inside the PTV. Also, fewer hot spots are produced beyond that structure for MicroCalc than for PCS, as observable by comparing Figs. 6.4 and 6.5(d) and (e), even though both analytical algorithms tend to underestimate the dose beyond distal edges. On the other hand, an underestimation of the dose is observed at the penumbra of the beams for both analytical algorithms (Fig. 6.5(d) and (e)). However, the beam model employed in Eclipse to analytically compute dose uses a single Gaussian to characterize the lateral profile whereas it has been pointed out that might be insufficient to reproduce adequately transversal dose distributions (32). As pointed out, the function in equation 6.5 that characterize the energy de- 122
Figure 6.5: Dose distributions on an axial plane at the middle of the PTV for a prostate case for (a) MCsquare, (b) MicroCalc and (c) PCS algorithm. The yellow contour depicts the PTV, with prescription dose of 68 Gy. On the color map, blue color represents low dose (near 0 Gy) while most intense red color represents 68 Gy. Differences between (d) MicroCalc and MCsquare and (e) PCS and MCsquare are also shown. In these color maps, most intense blue shows MCsquare dose is lower by 3.5 Gy, most intense red means MCsquare dose is higher by 3.5 Gy and white means no difference. position on the proton energy is extrapolated beyond 100 MeV (14) because of the limitation existing in the current version of Geant4-DNA. That lack of actual data for higher energies than this limit is a potential explanation for the observed differences at the entrance with MCsquare in Fig. (6.3). Besides that, in our current calculations, only protons and alpha particles are considered thus dose coming from heavy particles produced in heavy nuclear interaction is not considered. According to Grassberger and Paganetti (33), the contribution of those is relatively higher at the entrance so that this effect may also participate on the found differences at that area. An important advantage of MicroCalc is that it allows for incorporating the contributions of different particles independently. This can be done by characterizing the microdosimetric energy depositions in a similar way as for protons. Even in complex non-homogeneous cases as shown in Fig. 6.4, the dose calculation provided by MicroCalc is able to produce results close to those provided by MCsquare. However, larger inaccuracies may appear when considering greater tissue variabilities along the path beam, as seen at the left posterior cavity in Fig. 6.4(d). Nonetheless, the approach employed to take into consideration lack of homogeneity is relatively simple, i.e., only a displacement of the calculated flu- 123
ence according to the water-equivalent depth. Therefore, a potential improvement of this work consists of implementing a more sophisticated method to overcome large inhomogeneities situations. Besides the accuracy of the calculation, at least comparable to other analytical methods, two main reasons make MicroCalc a powerful tool compared to pencil beam algorithms: (i) it can calculate several quantities simultaneously (23); and (ii) it is modular, i.e. different particles can be added independently. Both reasons are based on the idea of transporting protons in water while keeping their frequency and spectrum information, which makes MicroCalc closer to actual MC methods. As shown in previous works (14), not only dose but also three-dimensional distributions of microdosimetric quantities might be obtained in combination with modeled energy-dependent functions in a similar way than shown here. Furthermore, other macroscopic quantities as dose-averaged LET can be also calculated (23) based on stopping power curves, and even combinations of all those functions for optimization purposes (e.g. dose-LET product) may be computed in a straightforward way. 6.5 Conclusions A new analytical algorithm –MicroCalc– to calculate dose is explicitly introduced and detailed in this work. The accuracy of this method is, at least, not inferior to the Proton Convolution Superposition (PCS) algorithm, already clinically validated in Eclipse. This seems to corroborate the methodology employed in this work as well as, partially, the microdosimetric functions developed in our previous work (14). On top of that, MicroCalc provides the possibility of separating different particles contributions and simultaneously computing several quantities of interest, which conceptually makes it similar to Monte Carlo techniques. References [1] H. H. Rossi. Spatial Distribution of Energy Deposition by Ionizing Radiation Source. Radiation Research, 2(Suplement):290–299, 1960. [2] H. H. Rossi. Microdosimetry. Biophysical Aspects of Radiation Quality. IAEA, Technical Reports Series, 58, 1966. [3] A. M. Kellerer and H. H. Rossi. The Theory of Dual Radiation Action. In M. Ebert and A. Howard, editors, Current Topics in Radiation Research. Vol- 124
Conclusions: Unrestricted LETd, restricted LETd and yDcan be analytically computed and compared for clinical plans. Two important consequences of the calculation of yDare: 1) its distribution can be verified by directly measuring it in clinical beams; and 2), optimization of proton treatments based on these quantities is enabled as well as future developments of RBE models based on them. 7.1 Introduction The importance of the biophysical properties of proton therapy beams have been substantially considered due to the variability of the relative biological effectiveness (RBE) (1, 2) and its potential clinical consequences. The use of intensity modulated techniques in proton therapy (IMPT) is becoming the standard of care, and these types of techniques produce non-homogeneous dose-averaged LET (LETd) distributions (3) that have been shown to be correlated with clinical outcomes (4). While the large variability of RBE in vitro experiments makes its determination, in general, indistinguishable from its associated uncertainty (2), LETd is a physical quantity accurately computable (5, 6) and directly correlated with RBE. This correlation may lead to a practical biophysical optimization of proton therapy treatments (7, 8, 9, 10, 11, 12). The concept of LETd, nonetheless, contains some simplifications and disregards relevant aspects of the patterns of energy imparted in matter. LETd is traditionally calculated by employing data for collision stopping power or unrestricted LET, which is defined as the mean energy lost by the proton in electronic collision processes per unit path length (13). Nevertheless, to better characterize the concentration or density of energy deposition in biological structures, energy carried out away from them should be taken off this computation. In this context, the concept of restricted LET, L∆arises, normally defined by using a threshold kinetic energy for secondary delta-rays, ∆, above which those are disregarded (13). By using the microdosimetry formalism (14, 15) and the concept of site, spatially-restricted LETd can be calculated. Furthermore, the microdosimetric concept of lineal energy (y) and its equivalent version of LETd, called dose-mean lineal energy (yD) might better represent microscopic distribution of the energy depositions since they contain the information of the imparted energy patterns in small volumes instead of per track length as done with the macroscopic concept of LET. In this sense, yDhas been also related to RBE by means of the Microdosimetric- Kinetic Model (MKM) (16, 17) and their subsequent modifications (18, 19). In fact, 131
efforts have lately been put to establish this connection (20, 21) and enable the biophysical optimization of proton radiotherapy treatments in terms of yD. Nevertheless, determination of yDin clinical cases would require extremely lengthy Monte Carlo (MC) simulations down to microscopic level, which has made its use impractical so far. In previous articles, we have presented models to calculate microdosimetric quantities such as yDand restricted LETd in monoenergetic beams (22). Independently, we have developed and introduced an analytical algorithm to determine the energetic spectrum for proton therapeutic beams in each voxel inside a calculation volume within a patient (23). Here, we combine these two works to produce distributions of restricted LETd and yDin clinical proton therapy cases. This has led us to propose a simplified and optimized expression for restricted LETd. Analogies and differences between distributions of unrestricted LETd, restricted LETd and yDare here explored. 7.2 Methods and Materials In order to clarify the notation employed throughout this Chapter and make the reading easier, Table 7.1 shows a list of relevant variables, their symbols and their nature. 7.2.1 LET and its relation with microdosimetry A site is a certain region of interest, with a given shape and size, in a particular medium in which the energy imparted by a particle in its electronic collisions is scored. Each incident charged heavy particle has a defined track, composed of its trajectory as well as those for the secondary electrons generated along its way. An interaction between a track and a site is called event. The amount of energy imparted to the site in a single event is notated by εs. This is a stochastic quantity, i.e. it varies from event to event even when the experimental conditions are the same. Therefore, εsis characterized by a probability distribution f(εs), with certain mean value εsand variance σ(εs)2, which provides a measurement of the variability for the value of εsin an experiment. The variability on εsis influenced by three sources (14): (i) the variability on the average energy imparted per unit length, characterized by means of the linear energy transfer (LET), which represents the ability of the particle to interact with the medium (i.e. interaction cross section); (ii) the variability on the segment length s, which is the distance 132
Table 7.1: List of relevant quantities employed in the mathematical development throughout this work, with their symbols and nature. For LET-wise quantities, the combination of the subscripts here shown combine their nature: for example, L∆,D is the dose-average spatially restricted linear energy transfer. Symbol Quantity Nature εsEnergy imparted per event Stochastic (event-wise) εsMean energy imparted per event Average of distribution of εs per event σ2 εsVariance of the energy imparted per event Variance of distribution of εsper event sSegment length for an event Stochastic (event-wise) sMean segment length Average of distribution of s per event sDWeighted average segment length Weighted average of distribution of sper event yLineal energy for an event Stochastic (event-wise) yDDose-mean lineal energy Weighted average of distribution of yper event εcEnergy imparted per collision Stochastic (collision-wise) εc≡δ1Mean energy imparted per collision Average of distribution of εc per collision σ2 εc≡σ2 δVariance of the energy imparted per collision Variance of distribution of εcper collision δ2Weighted average energy imparted per collision Weighted average of distribution of εcper collision S(E)Stopping power for protons of energy EFunction of proton energy εs(E)Mean energy imparted per event for protons of energy E Function of proton energy σ2 εs(E)Variance of the energy imparted per event for protons of energy E Function of proton energy φE(E)Differential spectrum for a beam: distribution of protons with energy E Distribution per proton energy εs≡εs(E)Average of mean energy imparted per event for a polyenergetic beam Average of distribution of εs per proton energy σ2 εs≡σ2 εs(E)Average of the variance of the energy imparted per event for protons of energy E Average of distribution of σ2 εsper proton energy σ2 εs≡σ2 εs(E)Variance of the mean energy imparted per event Variance of distribution of εsper proton energy LTTrack-average linear energy transfer Average of distribution of LET per particle LDDose-average linear energy transfer Weighted average of distribution of LET per particle L∞Unrestricted linear energy transfer Deterministic L∆Spatially restricted linear energy transfer Deterministic traveled by the particle within the site (22); and (iii) the energy straggling, i.e. given an average energy imparted per unit length, the variability on the number of collisions and energy deposited in each one of them due to stochastic nature of 133
interactions between radiation and matter. The quotient between the energy imparted in a single event εsand the mean segment length sof tracks in the site is called lineal energy: y=εs s(7.1) Consequently, lineal energy is also a stochastic quantity, characterized by a distribution f(y)that includes information about the volumetric pattern of energy deposition. However, the deterministic concept of LET is generally employed instead, since it allows for a more simplistic approach in which track structure can be disregarded. In this context, when dealing with a polyenergetic or multi-particle beam, a distribution of different values for LET arises. In such a situation, it has been shown (24) that for typical doses in proton therapy the dose-weighted average or dose-averaged LET, LD, is more representative of the biological effect than the simple average LET, traditionally called track-averaged or fluence-averaged LET, LT. Dose-averaged restricted LET is related to dose-mean lineal energy yDby (22) L∆,D =s sD yD−δ2 sD (7.2) where sand sDare the average and the weighted average of the distribution of segment length, respectively, and δ2is the weighted average of the distribution of the energy imparted per electronic collision εc(25). Additionally, the subscript ∆is added to LDto indicate that, by using this formalism, the obtained LET is restricted to a site dimension noted by ∆. In previous works, we have developed analytical functions on the proton energy for the averages and variances of the microdosimetric distributions (22). By using these functions, the mean energy imparted, and the variance of the energy imparted for a polyenergetic proton beam can be obtained, respectively, as εs=Rεs(E)φE(E) dE RφE(E) dE(7.3) and σ2 εs=Rσ2 εsφE(E) dE φE(E) dE+Rεs(E)−εs2φE(E) dE RφE(E) dE(7.4) where φE(E)represents the spectral fluence of the beam as a function of the energy, εs(E)and σ2 εs(E)are the average and variance of εscalculated for monoenergetic protons of energy E. Similarly, averages and variances of the overall distributions of segment length sand energy imparted per electronic collision can be obtained. In equation (7.3) note the double average since the obtained mean energy is the averaged value over the different energies present in the beam, which, 134
in turn, have their own average value for the microdosimetric distribution of energy imparted. Equation (7.4), on the other hand, is composed by two terms: the first term is the mean variance across the energies present in the beam, σ2 εs, and the second is the variance of the averages of the microdosimetric distributions of energy imparted, σ2 εs. 7.2.2 Unrestricted LETd Dose-averaged LET can be obtained regardless microdosimetric considerations by employing the concept of collision or electronic stopping power Sor unrestricted LET, L∞, in which the energy lost by the particle in electronic collisions is considered instead of the energy imparted in a certain region as microdosimetry does. Stopping power is a deterministic concept, which means it has no stochastic variability. Therefore, given a particle with energy E, we can note it as S(E) = dε∞(E) ds(7.5) where dε∞(E)is the mean energy lost by the particle in an infinitesimal distance ds. Note that dε∞here is a deterministic quantity. As the dose imparted by a particle with a given stopping power Scan be considered proportional to S, the dose-averaged electronic stopping power or unrestricted dose-averaged LET can be obtained by weighting the electronic stopping power by itself so that L∞,D =RS2(E)φE(E) dE RS(E)φE(E) dE(7.6) where φE(E)represents again the spectral fluence. 7.2.3 Connection between restricted and unrestricted dose-averaged LET If we consider a virtual site that is infinitely large to contain all the energy lost by the particle, then ∆→ ∞, and dε∞(E)in equation (7.5) coincides with the microdosimetric mean energy imparted, εs→dε∞(E). This situation occurs as long as secondary electrons do not travel a distance from the primary track comparable with the dimensions of the site. Equivalently, if a virtual site infinitely small is considered, then s→0. In practice, this happens if stopping value does not appreciably change along the path of the particle within the site. Now, in the case of this double-ideally virtual site, equation (7.6) can be expressed in terms of 135
microdosimetric quantities means as L∞,D = lim ∆→∞ s→0 1 sRεs2φE(E) dE RεsφE(E) dE= lim ∆→∞ s→0 1 s σ2 εs+εs2 εs = lim ∆→∞ s→0 εs s 1 + σ2 εs εs2!(7.7) where we have used the general relation σ2 x=x2−x2, applied for the case x=εs, so that we have εs2=σ2 εs+εs2. As, by definition, L∞,D = lim∆→∞ s→0L∆,D, it follows L∆,D =εs s1 + σ2 εs εs2(7.8) Note that the microscopic spatially-restricted L∆,D from equation (7.8) will differ from the macroscopic unrestricted L∞,D in equation (7.7) as much as the real microdosimetric site considered deviates from the doubly ideal situation represented by the twofold limit in equation (7.7). The expression in equation (7.8) according to the equations (7.3) and (7.4), only needs two microdosimetric energy-dependent functions: the average energy imparted per event ε(E)and the average segment length s(E), instead of the six functions necessary to compute L∆,D through equation (7.2). Note that, according to y definition in equation (7.1), the expression for yD≡yF(1 + σ2 y/y2 F)is quite similar to equation ((7.8) yD≡εs s1 + σ2 εs εs2=εs s 1 + σ2 εs+σ2 εs εs2!(7.9) where the second equality is obtained directly from equation (7.4) and segment length is now double-averaged to take into account the most general case, in which it can depend on the particle energy: s=Rs(E)φE(E) dE. However, this can be approximated by simply sas long as the fraction of particles with range shorter than the site dimension is negligible, which turns out to be most of the cases for typical energies and site dimensions in proton radiotherapy. Now, as seen in equation (7.8) a third microdosimetric energy-dependent function needs to be incorporated to consider the spread of the energy imparted around its mean value: the variance of the energy imparted per event δ2 ε(E)–or, equivalently, its squared root-. 7.2.4 Revision of the needed microdosimetric functions to be modeled In Bertolet et al, 2019a (22) (Chapter 3) we presented analytical models to recreate the microdosimetric quantities correspondent to different monoenergetic proton beams. The microdosimetric patterns of energy deposition were obtained by 136
using MC simulations, done with Geant4-DNA, of monoenergetic proton beams up to 100 MeV penetrating into spherical sites of 1 µm, 5 µm and 10 µm in-diameter. From these results, a series of analytical functions of the beam energy were derived. Namely, analytical forms for those site dimensions were obtained for the quantities εs(E),σεs(E),s(E),sD(E),δ1(E)and σδ(E), where the two latter are the average and standard deviation of the distribution of energy imparted per collision, respectively. Note that, as shown in that work, δ2=δ1(1 + σ2 δ/δ2 1)and, therefore, from the six previous functions, it is possible to calculate L∆,D according to equation (7.2). According to equation (7.8), three functions modeled in that work are not necessary anymore: sD(E),δ1(E)and σδ(E), being the latter two the average and standard deviation of the distribution of energy imparted per collision, respectively. This is especially convenient since that distribution needs to be obtained from monoenergetic independent simulations (25). In this work, we show that equation (7.8) is equivalent to equation (7.2) to obtain restricted LETd from microdosimetry in a more efficient and compact way. Furthermore, we calculate three-dimensional distributions of yDaccording to equation (7.9). The interest on yDis twofold: first, yDis a measurable quantity as long as a microdosimeter capable to store energy deposition distributions is available, while L∆,D) is not. This can be understood just looking at the difference between equations (7.8) and (7.9): σ2 εsis the actual variance of the measured distribution whereas σ2 εsis the variance of the mean energies imparted if the problem is decomposed in its spectral components, which needs the knowledge of the specific energetic spectrum measured by the microdosimeter. Second, as it has been pointed out (24), yDcan be a more representative quantity to characterize the energy imparted and eventual biological damage in patients than L∆,D, which, indeed, disregards the volumetric pattern of deposition of energy reducing the problem to a unidimensional one. Nevertheless, as multiple models for biological effect are based on LETd, we keep calculating this quantity in its two versions, unrestricted and restricted; their differences between each other and with respect to yDare here explored. 7.2.5 MicroCalc: an analytical algorithm based on spectral fluence models In order to obtain the integrations over the spectrum given in equations (7.3), (7.4) and (7.6), we have developed an analytical algorithm to reproduce the spectra 137
at the voxel level in proton particle therapy. This algorithm has been presented in previous articles (23, 26), in which calculations of unrestricted LETd as well as dose are shown and benchmarked against Monte Carlo computations, and can be briefly summarized as follows: Transport of monoenergetic proton beams with energies in the clinical range (50-250 MeV) in water was calculated by means of MC simulations. Different types of particles are separated depending on their spectra: primary protons, secondary protons and other secondary particles. The dependency of the fluence of each one both on depth and laterally was modeled using analytical functions. At the same time, spectra for each one is assumed to be Gaussian and the mean and the standard deviation of these Gaussian spectra are, respectively, analytically modeled as functions of the depth and the lateral distance to the beam axis. The combination of these two models yields the spectral fluence for each species. Then, the integration given in equation (7.6) can be performed for each one of the species. The average LET is obtained as the fluence-weighted average of these results. More details can be obtained elsewhere(23). Here, we reproduce the unrestricted LETd calculation done in Bertolet et al, 2019b (23) (Chapter 5). The above algorithm is now used to obtain φE(E)at each position to perform the integrations in equations (7.3) and (7.4) to eventually calculate restricted LETd and yDfor spherical sites of 1 µm and 10 µm using the microdosimetric functions developed in Bertolet et al, 2019a (22) (Chapter 3). These functions are valid up to a beam energy of 100 MeV, being extrapolated beyond this point. The algorithm is already deployed as a script in the treatment planning system (TPS) Eclipse v15.6 of Varian Medical Systems, which allows for calculations in clinical cases. This script is used to produce results for: (i) L∆,D according to equation (7.2) and to equation (7.8); (ii) L∞,D according to equation (7.6); and (iii) yDaccording to equation (7.9). To assess the equivalency between equations (7.2) and (7.8), a simple case is evaluated: a pristine Bragg peak for a beam with nominal energy equal to 100 MeV produced with the Ion Beam Applications (IBA) Pencil Beam Scanning (PBS) system installed in the Roberts Proton Center at the Hospital of the University of Pennsylvania. This energy is selected to disregard possible discrepancies coming from the extrapolation of the microdosimetric functions for beam energies beyond 100 MeV. On the other hand, a prostate cancer clinical case is considered to illustrate and analyze the different distributions of unrestricted LETd, restricted LETd and 138
yD. The calculated plan consists of two lateral beams optimized individually to produce a uniform dose distribution over the target, with a prescription dose of 70.2 Gy in 39 fractions. Although MicroCalc considers non-homogeneities in the fluence model, results for L∆,D,L∞,D and yDare referred to water. Additionally, results shown here only encompass primary and secondary protons. 7.3 Results Figure 7.1 shows the profiles for LETd at the central axis for a beam of 100 MeV as nominal energy as well as for a SOBP beam with range 20 cm and 10 cm of modulation. Three different calculations are presented: unrestricted LETd or L∞,D according to equation (7.6) and restricted L∆,D for two different microdosimetric sites, both spherical and with diameter of 1 µm and 10 µm, respectively. Results corresponding to equations (7.2) and (7.8) are compared to assess the equivalency of the methods. The mean difference along the profiles taking samples stepped by 1 mm, between these two curves, are 0.05 ±0.04 keV/µm for the 1-µm case and 0.05 ±0.02 keV/µm for the 10-µm case, respectively, where the uncertainties correspond to the standard deviation of the difference samples. Differences between yDand both versions of LETd are illustrated in Figure 7.2 for the same 100-MeV proton beam. While L∆,D remains almost independent of the site dimension, yDis clearly higher for the 1 µm case than for the 10 µm case. An axial plane at the middle of the prostate is selected to show the computed distributions. Dose and unrestricted LETd are the same for the two considered sites, meanwhile restricted LETd and yDdistributions are different depending on the site dimension selected. Figure 7.3 shows the projection of the mentioned distributions upon the chosen axial plane. The profiles along the dashed line marked in Figure 7.3(a) for these quantities are shown in Figure 7.4. Two types of comparisons are shown: on the one hand, restricted and unrestricted LETd are displayed in Figure 7.4(a). Unrestricted LETd is markedly higher along the whole profile than both restricted versions of LETd, being the 10 µm-site-restricted LETd curve slightly above the 1 µm-site- restricted LETd. On the other hand, Figure 7.4(b) shows the differences between unrestricted LETd and yD. In this case, yDfor the site of 1 µm in diameter is clearly superior to yDfor the site of 10 µm. As expected according to equations (2) and (9), yDcurve is above LDcurves at every point. To evaluate the relative shape 139
Figure 7.1: Profiles at the central axis for unrestricted LETd and restricted LETd calculations using the MicroCalc algorithm and considering only primary and secondary protons, using equations (7.6) for unrestricted LETd, and (7.2) and (7.8) for restricted LETd, respectively for a beam with nominal energy equal to 100 MeV. Left panel is obtained using the microdosimetric models for spherical sites of 1 µm in diameter meanwhile right panel shows the same results for spherical sites of 10 µm in diameter. Solid line: L∞,D. Dotted line: L∆,D from equation (7.2). Dashed line: L∆,D from equation (7.8) . of each distribution, the ratio between the peaks and the plateau, at the center of the profiles, yields the values: (a) 1.90 for L∞,D; (b) 2.17 for L∆,D (1 µm); (c) 2.15 for L∆,D (10 µm); 1.78 for yD(1 µm); and 2.01 for yD(10 µm). 7.4 Discussion Results in Fig. 7.1 seem to endorse the expression for restricted LETd from microdosimetry given in equation (7.8). Although both calculations share, essentially, the models for single-event average energy imparted per event ε(E)and mean segment length s(E), the use of equation (7.2) includes the other four quantities modeled in Bertolet et al, 2019a (22) (Chapter 3), i.e. standard deviation for the single-event energy imparted in the site σε(E), weighted-mean segment length sD(E)and average and standard deviation for the energy imparted in the site per 140