Anterior chamber angle width, central vault and intraocular pressure changes after 12 months of Visian collamer lens implantation

Sánchez González, José María; Alonso Aliste, Federico; Perea Peña, Gracia; Rachwani-Anil, Rahul; Márquez de Aracena del Cid, Rafael; Rocha de Lossada, Carlos

Abstract

Purpose: To evaluate 1-year changes of vault, anterior chamber angle width (ACAW) and intraocular pressure (IOP) after the implantation of implantable collamer lens (ICL) V4c in high myopic patients. Methods: Sixty ICL V4c implantation surgeries were included. Patients underwent surgery between January 2014 and June 2017 at Ophthalmology Center Tecnolaser Clinic Vision®, Seville, Spain. All patients had a 12-month follow-up. A preoperative spherical equivalent between − 4.00 and − 15.00 D was established as inclusion criterion. ACAW and vault were measured with Pentacam single rotation Scheimpflug camera (Oculus Optikgeräte GmbH, Wetzlar, Germany). IOP was measured with CORVIS ST (Oculus Optikgeräte GmbH, Wetzlar, Germany). Results: One-day vault was 553.17 ± 169.80 [268–975] µm; 12-month vault was 515.67 ± 176.46 [169–950] µm (p < 0.01). Preoperative ACAW was 36.85° ± 4.60° [26.7°–46.8°]; postoperative ACAW was 25.98° ± 3.85° [16.4°–32.6°] (p < 0.01). Preoperative IOP was 15.37 ± 1.58 [12.0–18.3] mmHg, postoperative IOP was 15.61 ± 1.60 [12.0–19.0] mm Hg (p = 0.35). Conclusions: ICL V4c lens experienced a progressive reduction of vault after 12-month follow-up. ACAW decrease was significant between preoperative and postoperative values. We propose to include a minimum preoperative ACAW and an IOP correction factor for lens size.

Full text

1 Clean Manuscript Click here to access/download;Manuscript;File #3 - Clean Manuscript.docx Click here to view linked References 1 This version of the article has been accepted for publication, after peer review (when applicable) and is subject to Springer Nature’s AM terms of use, but is not the Version of Record and does not reflect postacceptance improvements, or any corrections. The Version of Record is available online at: https://doi.org/10.1007/s10792-020-01381-w 2 3 4 Title 5 Anterior chamber angle width, central vault and intraocular pressure changes after twelve months of Visian 6 collamer lens implantation. 4 5 Authors 6 1. Sánchez-González, José-María OD, PhD a b 7 2. Alonso-Aliste, Federico, MD b 8 3. Perea-Peña, Gracia, OD a 9 4. Rachwani-Anil, Rahul MD d 10 5. Márquez-de-Aracena-del-Cid, Rafael, MD PhD c 11 6. Rocha-de-Lossada, Carlos, MD d e 12 13 a Department of Physics of Condensed Matter, Optics Area. University of Seville, Seville, Spain. 14 b Department of Ophthalmology. Tecnolaser Clinic Vision, Seville, Spain. 15 c Department of Surgery, Ophthalmology Area. University of Seville, Seville, Spain. 16 d Department of Ophthalmology, Regional University Hospital of Malaga, Malaga, Spain. 17 e Department of Ophthalmology, Clinic Barcelona Hospital, Barcelona, Spain. 18 19 Corresponding author 20 Sánchez-González, José-María 21 Reina Mercedes Street, University of Seville 22 +34 618204110 / [email protected] 23 24 Conflict of Interest: Author José-María Sánchez-González declares that he has no conflict of interest. Author 25 Federico Alonso-Aliste declares that he has no conflict of interest. Author Gracia-Perea-Peña declares that she 26 has no conflict of interest. Author Rafael Márquez-de-Aracena-del-Cid declares that he has no conflict of interest. 27 Author Rahul Rachwani-Anil declares that he has no conflict of interest. Author Carlos Rocha-de-Lossada 28 declares that he has no conflict of interest. 2 29 3 30 Abstract 31 Purpose: To evaluate one-year changes of vault, anterior chamber angle width (ACAW) and intraocular pressure 32 (IOP) after the implantation of Implantable Collamer Lens (ICL) V4c in high myopic patients. 33 34 Methods: Sixty ICL V4c implantation surgeries were included. Patients underwent surgery between January 2014 35 and June 2017 at Ophthalmology Center Tecnolaser Clinic Vision®, Seville, Spain. All patients had a twelve- 36 month follow-up. A preoperative spherical equivalent between - 4.00 diopters (D) and -15.00 D was established 37 as inclusion criterion. ACAW and vault were measured with Pentacam single rotation Scheimpflug camera 38 (Oculus Optikgeräte GmbH, Wetzlar, Germany). IOP was measured with CORVIS ST (Oculus Optikgeräte 39 GmbH, Wetzlar, Germany). 40 41 Results: One day vault was 553.17 ± 169.80 [268 – 975] µm, 12 months vault was 515.67 ± 176.46 [169 – 950] 42 µm (p < 0.01). Preoperative ACAW was 36.85° ± 4.60° [26.7° – 46.8°], postoperative ACAW was 25.98° ± 3.85° 43 [16.4° – 32.6°] (p < 0.01). Preoperative IOP was 15.37 ± 1.58 [12.0 – 18.3] mmHg, postoperative IOP was 15.61 44 ± 1.60 [12.0 – 19.0] mm Hg (p = 0.35). 45 46 Conclusions: ICL V4c lens experienced a progressive reduction of vault after twelve months follow-up. ACAW 47 decrease was significant between preoperative and postoperative value. We propose to include a minimum 48 preoperative ACAW and an IOP correction factor for lens size. 49 50 Keywords 51 Implantable collamer lens; anterior chamber angle width; central vault; intraocular phakic lens; posterior chamber 52 phakic lens; intraocular pressure 4 53 Declaration 54 Funding: No funding obtained 55 Conflicts of interest/Competing interests: Authors declare no conflict of interest 56 Ethics approval: International Review Board approved this Research 57 Consent to participate: All patient read, understand and sign inform consent 58 Consent for publication: All authors declare positive consent for publication 59 Availability of data and material: Data is available on demand to corresponding 60 Code availability: N/A 61 62 Acknowledgments 63 The authors acknowledge the support offered by the members of the Faculty of Pharmacy of the University of 64 Seville as well as the staff of the Degree in Optics and Optometry. Data collection and support help: Amián- 65 Cordero, Jonatan; Hernández-Barahona, Mariano; Castellanos-Gómez, Ignacio; Socas-Priego, Luisa and López- 66 Martos, Nino 67 68 69 5 70 Introduction 71 Posterior chamber Implantable Collamer Lens (ICL), has been described as a safe and effective[1–5] surgery 72 technique in high myopia correction. Conventional ablation techniques confer excellent results in visual acuity, 73 however the risk of corneal ectasia due to excessive removal of corneal tissue remains as a potential severe 74 complication. ICL appears as an alternative in patients who are not suitable for laser ablation of the corneal 75 surface. Despite being an effective treatment and offering the advantage of being a reversible technique, there are 76 several complications that can considerably worsen the results. Increased Intraocular Pressure (IOP) as a result of 77 pupillary block[6] and the development of anterior subcapsular cataracts[7] due to mechanical friction of the ICL 78 lens with crystalline lens are some of the most frequent complications.[8, 9] 79 80 Vault is the distance between the anterior face of the crystalline lens and the posterior face of the iris. An excess 81 vault (>750 µm) contributes to angular narrowing and, subsequently, IOP increase and risk of glaucoma. An 82 insufficient vault (<250 µm) increases friction risk against the anterior face of the crystalline.[9] This parameter 83 may change over time.[5, 10–14] Previous studies had shown an inverse proportional relationship between vault 84 and iridocorneal angle.[15–17] Efficacy of ICL in the refractive correction has been evaluated, with follow-up at 85 one month, 6 months and 18 months after implantation. Recently, new ICL models have been designed with 86 advanced technology that prevent complications such as IOP increase, avoiding the performance of intraoperative 87 iridectomies.[6, 18] V4c collamer lens (ICL V4c) includes a 0.36 mm diameter central hole, called KS- 88 AquaPORT, allowing aqueous humor flow through the ICL anterior face. Currently, intraoperative complications 89 have reduced, however they are still frequent in the post-operative period.[6] 90 91 The purpose of the present study was to describe changes in vault, ACAW and IOP after the implantation of ICL 92 V4c for correcting high myopia after twelve months follow-up. 93 6 94 Materials and Methods 95 96 Design 97 Sixty ICL V4c implantation surgeries were included in this retrospective, observational and longitudinal study. 98 Patients underwent surgery between January 2014 and June 2017. All phakic lens surgeries were performed at the 99 facilities of the Ophthalmology Center Tecnolaser Clinic Vision ®, Seville, Spain. All patients had a 12-month 100 follow-up. 101 101 102 Ethical aspects 103 All patients included in this work were adequately informed verbally and in writing of the benefits, characteristics 104 and risks of surgeries. All patients signed a consent form prior to the surgery and after the interview performed 105 with the ophthalmologist. This study was conducted in accordance with the tenets of the Helsinki Declaration and 106 obtained Institutional Review Board exemption. 107 107 108 Patients 109 Patients voluntarily went to the clinic to perform the tests. Next, an ophthalmologist determined their suitability 110 for surgery, they underwent ICL refractive surgery voluntarily. The inclusion criteria were; (1) age between 20 111 and 45 years, (2) a stable refraction for at least 2 years, (3) presence of myopia in spherical equivalent between – 112 4.00 D and – 15.00 D, (4) presence of astigmatism between 0.00 D and – 5.00 D and (5) non-candidates for laser 113 refractive surgery. The exclusion criteria comprised; (6) anterior chamber depth from endothelium ≤ 2.8 mm, (7) 114 endothelial cell density ≤ 2000 cells per mm2, (8) hyperopia, (9) a family history of glaucoma, (10) inflammatory 115 eye disease (keratitis, herpes zoster, ocular herpes), (11) previous eye surgery or (12) any kind of corneal disease 116 (keratoconus, acute or chronic corneal infection, ocular herpes, Stevens-Johnson syndrome, glaucoma, cataract, 117 uveitis or retinal detachment). ICL power and size were calculated with STAAR ® Surgical software. 118 118 119 119 7 120 Preoperative Examinations 121 Before the implantation of the ICL (Visian ICL [V4c] STAAR® Surgical, Monrovia, CA), patients underwent a 122 thorough preoperative examination performed by licensed and expert optometrists, including: uncorrected 123 distance visual acuity (UDVA) and corrected distance visual acuity (CDVA) in Snellen scale, manifest refraction 124 with and without cycloplegia (refraction method of the first and second maximum positive). Astigmatism was 125 assessed by the Jackson cross cylinder method. These data were checked with a Wavefront Supported Custom 126 Ablation (WASCA) and an autorefractor-aberrometer (Carl Zeiss Meditec AG, Jena, Germany). Binocular vision 127 and accommodation were assessed in all patients. Corneal topography, white to white (WTW), anterior chamber 128 angle width (ACAW) and anterior chamber depth (ACD) were measured and later analyzed with Pentacam single 129 rotation Scheimpflug camera (Oculus Optikgeräte GmbH, Wetzlar, Germany). IOP and corneal biomechanics 130 were measured with CORVIS ST (Oculus Optikgeräte GmbH, Wetzlar, Germany). Pachymetry, epithelial 131 thickness, retinal optical coherence tomography and measurement of the distance from the posterior side of the 132 ICL to the anterior face of the crystalline lens (vault) were measured using spectral domain optical coherence 133 tomography (SD-OCT) (Optovue Inc., Fremont, CA). WTW and ACD were confirmed with a conventional optical 134 biometry (PCI Technology, IOL Master 500, Carl Zeiss Meditec AG, Jena, Germany). In uncertain ACD, a third 135 measurement was carried out with an anterior segment optical coherence tomography (Visante OCT, Carl Zeiss 136 Meditec AG, Germany). Finally, endothelial cell density (ECD) was measured with CEM-530 non-contact 137 specular microscopy (Nidek Co Ltd, Japan). 138 138 139 Surgical technique 140 All surgeries were performed by two experienced surgeons (F.A.A and A.S.J). Pupils were dilated with 141 cycloplegic and phenylephrine every 15 minutes one hour prior to surgery. Eye contour was disinfected with 5% 142 povidone-iodine (Betadine; Meda Manufacturing, Bordeaux, France) 10 minutes prior to surgery. Peribulbar 143 anesthesia was carried out. After injecting 1% sodium hyaluronate in the anterior chamber (Provisc; Alcon 144 Laboratories Inc, from Fort Worth, TX) ICL was implanted via a 3.0 mm corneal incision in the most curved 145 meridian. Sodium hyaluronate was removed before finalizing the surgery. Postoperative treatment consisted of 146 tobramycin and dexamethasone at 0.1% drops (Tobradex Alcon Laboratories Inc) and diclofenac sodium drops 147 (Voltaren; Novartis Pharmaceuticals, Basel, Switzerland) ever six hours for the first 15 days. Posteriorly, the 148 treatment was tapered. All cases were bilateral implantations. The second eye was conducted a week after the first 8 149 surgery. All ICL were spherical and, in case of residual astigmatism, femtosecond Laser-assisted in situ 150 Keratomileusis (femto-LASIK) was performed to correct the remaining refraction. Flap was performed using 151 VisuMax Femtosecond Laser System (Carl Zeiss Meditec AG, Jena, Germany). Laser pulses were applied with a 152 pulse energy of approximately 130 nJ. The frequency of the laser was 500 KHz. The line and spot distance was 153 4.5 µm. The raster pattern was circular. The estimated flap thickness was 100 µm, and the flap diameter was 8.5 154 mm. Excimer laser was performed with MEL 90 (Carl Zeiss Meditec AG, Jena, Germany), an argon-fluoride 155 solid-state laser. Excimer laser wavelength was 193 nm and fixation laser wavelength was 532 nm. Regarding 156 spot scanning parameters, beam dimensions were 0.7 mm FWHM (full width at half maximum) with a Gaussian 157 beam profile. 158 158 159 Statistical analysis 160 Statistical analysis was carried out using SPSS statistics 25.0 (IBM Corporation, Armonk, NY, USA). Visual 161 acuity data was converted into Snellen format. Effect size was calculated using Cohen's d. Paired t test was used 162 for normally distributed data. We carried out correlation analysis between preoperative and postoperative patients. 163 A P value of less than 0.05 was considered statistically significant. 164 164 165 165 9 166 Results 167 Refractive Outcomes 168 This study included 60 eyes belonging to 30 subjects (15 females and 15 males). Mean age was 30.10 ± 6.70 [20 169 – 45] years. Mean preoperative spherical refraction was -9.04 ± 2.09 [-13.25 – -4.00] D. Mean preoperative 170 cylindrical refraction was -1.20 ± 1.36 [-5.00 – -0.50] D. Mean preoperative spherical equivalent was calculated 171 as sum of the sphere and half of the cylinder and was -9.64 ± 2.26 [-15.00 – -4.75] D. Mean CDVA was 0.84 ± 172 0.14 [0.50 – 1.00] in decimal scale. 173 Mean postoperative spherical refraction was 0.08 ± 0.47 [-0.75 – + 1.00] D, with a difference of -9.12 ± 2.20 D 174 compared to the preoperative status. The reduction in spherical refraction was statistically significant (p < 0.01), 175 with a large effect size of 5.42 according to Cohen's d. Mean postoperative cylindrical refraction was -0.26 ± 0.66 176 [-2.00 – + 1.25] D, with a difference of -0.94 ± 1.60 D compared to the preoperative status. The reduction in 177 astigmatism was statistically significant (p < 0.01) with a large effect size of 0.87 according to Cohen's d. Residual 178 astigmatism after spherical ICL V4c was treated with femto-LASIK. Mean postoperative spherical equivalent was 179 -0.47 ± 0.50 [-1.00 – +1.38] D with a difference of -9.59 ± 2.45 D compared to the preoperative status. The 180 reduction in spherical equivalent was significantly statistical (p < 0.01), with a large effect size of 5.86 according 181 to Cohen's d. Decimal mean postoperative UDVA was 0.97 ± 0.16 [0.60 – 1.50]. The increase in visual acuity 182 was 0.12 ± 0.20 in decimal scale, being statistically significant (p < 0.01), with a large effect size of 0.87 according 183 to Cohen's d. The correlation analysis between these samples was not relevant nor statistically significant. 184 184 185 185 186 Anterior Chamber Changes 187 One day vault was 553.17 ± 169.80 [268 – 975] µm and decreased to 515.67 ± 176.46 [169 – 950] µm twelve 188 months after surgery. This means a reduction of 37.50 ± 29.10 µm (p < 0.01), with a small effect size of 0.22 189 according to Cohen's d. Mean preoperative ACAW was 36.85° ± 4.60° [26.7° – 46.8°] and decreased to 25.98° 190 ± 3.85° [16.4° – 32.6°] twelve months after the surgery, suffering a reduction of 10.86° ± 3.42 ° (p < 0.01), with 191 a large effect size of 2.56 according to Cohen’s d . Preoperative IOP was 15.37 ± 1.58 [12.0 – 18.3] mm Hg and 192 increased to 15.61 ± 1.60 [12.0 – 19.0] mm Hg twelve months after surgery. This implies an increase of IOP of 193 0.24 ± 1.99 mm Hg (p = 0.35), although not being statistically significant and with an insignificant effect size of 16 328 Figure legends 329 Figure 1. Preoperative and postoperative anterior chamber angle width (ACAW) of one case included in the study. 330 330 line figure 1 Click here to access/download;line figure;Figure 1.tiff