Research Article Brno University of Technology Smartphone PPG Database (BUT PPG): Annotated Dataset for PPG Quality Assessment and Heart Rate Estimation Andrea Nemcova , 1 Enikö Vargova, 1 Radovan Smisek , 1,2 Lucie Marsanova , 1 Lukas Smital , 1 and Martin Vitek 1 1 Department of Biomedical Engineering, Faculty of Electrical Engineering and Communication, Brno University of Technology, Technicka 12, Brno 616 00, Czech Republic 2 Institute of Scientific Instruments, The Czech Academy of Sciences, Kralovopolska 147, Brno 612 64, Czech Republic Correspondence should be addressed to Andrea Nemcova; [email protected] Received 8 July 2021; Accepted 26 August 2021; Published 7 September 2021 Academic Editor: Mihajlo Jakovljevic Copyright © 2021 Andrea Nemcova et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. To the best of our knowledge, there is no annotated database of PPG signals recorded by smartphone publicly available. This article introduces Brno University of Technology Smartphone PPG Database (BUT PPG) which is an original database created by the cardiology team at the Department of Biomedical Engineering, Brno University of Technology, for the purpose of evaluating photoplethysmographic (PPG) signal quality and estimation of heart rate (HR). The data comprises 48 10-second recordings of PPGs and associated electrocardiographic (ECG) signals used for determination of reference HR. The data were collected from 12 subjects (6 female, 6 male) aged between 21 and 61. PPG data were collected by smartphone Xiaomi Mi9 with sampling frequency of 30 Hz. Reference ECG signals were recorded using a mobile ECG recorder (Bittium Faros 360) with a sampling frequency of 1,000 Hz. Each PPG signal includes annotation of quality created manually by biomedical experts and reference HR. PPG signal quality is indicated binary: 1 indicates good quality for HR estimation, 0 indicates signals where HR cannot be detected reliably, and thus, these signals are unsuitable for further analysis. As the only available database containing PPG signals recorded by smartphone, BUT PPG is a unique tool for the development of smart, user-friendly, cheap, on-the-spot, self-home-monitoring of heart rate with the potential of widespread using. 1. Introduction PPG signals are commonly used nowadays because the optical sensors can be integrated in almost any device (smartphone, smartwatch, smartring, …). The recording of PPG is cheap, simple, noninvasive, and comfortable, and it can be done on-the-spot. In case of using any smartphone, no additional device is necessary. The PPG signal is commonly used for heart rate estimation. Evaluation of PPG signal quality and estimation of HR from PPG signal have become a popular research topic [1–6], driven in part by the increased use of smartphones and wearables for health monitoring. A majority of HR estimation smartphone applications are neither tested, nor certified and moreover are not verified by medical experts (except, e.g., Preventicus Heartbeats [4]). Commercial smart watches such as Garmin [7] and Polar [8] also report problems with accuracy of HR estimation (caused, e.g., by physical activities, worsen blood flow in periphery, improperly weared watches, tattoo, measuring in water). Thus, HR estimation from mobile devices and wearables is still a challenging problem. It is logical for health monitoring algorithms or applications to carry out signal quality estimation before HR estimation. This approach may lead to greater robustness and reliability. For example, if a poor PPG signal quality is detected by a monitoring device, the signal can be discarded (preventing distorted and unreliable results) and a new measurement can be requested. Hindawi BioMed Research International Volume 2021, Article ID 3453007, 6 pages https://doi.org/10.1155/2021/3453007
There exist some PPG database. However, to the best of our knowledge, there is no annotated database of PPG signals recorded by smartphone publicly available. The BUT PPG database is unique in this sense. There are some benefits of PPG recording by smartphone comparing to the smartwatch/smartrings: it is widely available and cheap; the quality of PPG is satisfactory; application for PPG-based HR estimation and evaluation can be simply integrated. As the data acquisition as well as processing can be done in one device, there is no problem with (wireless) data transmission. BUT PPG database was created for the purpose of developing and evaluating algorithms designed to assess the quality of PPG records and algorithms designed for estimation of HR from PPG. Study design overview is shown in Figure 1. It starts with the subjects (on the left side)—volunteers on which the signals were acquired. Altogether, 6 men (light blue color) and 6 women (pink color) of various ages (number in each block) were measured. Two signals were acquired simultaneously—ECG signal (red color in Figure 1) using mobile ECG recorder and PPG signal (blue color in Figure 1) in a form of video using smartphone. Each subject was measured four times—three times in rest and once in move. The measured video was processed in three steps to obtain the PPG signal. The red video channel was selected, and each video frame was averaged to obtain a single value. The calculated PPG signal was inverted. Then, the ECG and PPG signals were manually synchronized, and the middle 10 seconds were selected to eliminate transition artifacts. The PPG quality and HR annotations were calculated. The HR annotations were created using the ECG signal. The QRS complexes were automatically detected and manually verified. Then, the median HR was calculated, and the output is the reference HR. The annotations of PPG signal quality were created by 5 experts using tailor-made annotating software. The experts estimated the HR from the PPG signals. Thereafter, the consensus of experts was made, and each signal was classified into good quality class (1) or poor quality class (0). All parts of the study are described below. 2. Materials and Methods 2.1. Signal Acquisition. All human studies were approved by the Institutional Review Board of DBME Faculty of Electrical Engineering and Communication, Brno University of Technology, on July 27, 2018 (IRB Protocol EC:EK:05b/2018). Informed written consent was obtained from all subjects prior to the studies. We recorded 48 10-second PPG signals using the Xiaomi Mi9 smartphone. The measured subject attached his/her index finger to the rear side of the smartphone to cover the camera and LED light. The light was turned on, and a 30second video of the finger was recorded. Concurrently, the single-lead ECG signal was recorded as a reference for HR. The Bittium Faros 360 device was used for this purpose. The electrodes were attached to the chest according to the device manual. The database is balanced in terms of gender (6 female records and 6 male records) and age (21 to 61 years, mean 36 years, median 24 years). Each subject was measured four times. The first three measurements were in rest while the subject was sitting. During the fourth measurement, the subject was walking and/or moving with the finger on the lens. For the database, only the middle 10 seconds of recordings were included to eliminate signal fluctuations. From the video with resolution of 720 × 1280 px, only the red color channel [9] was considered, and the PPG signal was created using averaging of each frame. Finally, the PPG signal was inverted. The sampling frequency was 30 Hz (frames per second). The sampling frequency of the ECG signal was 1,000 Hz. PPG and ECG signals were synchronized manually. The example of PPG signals of good and poor quality together with the reference ECG signals are shown in Figure 2. 2.2. Annotating. In the BUT PPG database, the HR and the quality of the signals are annotated. The annotations are provided for the whole 10-second signals (one signal, one number of HR, and one quality label). The reference HR was calculated from the ECG signals which are considered as the gold standard. The QRS complexes were found using a robust QRS detector based on a combination of three independent methods [10] and manually verified. Then, the median HR was calculated (one number for the whole 10 second signal), and this number was considered as the reference HR for the PPG signal. PPG signals were annotated in terms of their quality. The database contains binary signal-quality labels: (i) 1: indicates good quality for HR estimation (ii) 0: indicates poor-quality signals where the HR cannot be detected reliably and thus unsuitable for further analysis Quality annotations are based on HR estimation using PPG signals and comparison with reference (ECG) value of HR. Each PPG signal was annotated by 5 experts in terms of HR. Annotators were provided with tailor-made software (described below) for the purpose of annotating. Annotators also had the opportunity to decline to estimate the HR when the estimate would not be reliable. Annotators saw only the PPG signals (ECG signals were not provided for the objectivity of the annotations). Moreover, the PPG signals appeared in a random manner. The annotators did not know whether the signal was recorded in rest conditions or under the artifact scenario. Signals from one subject were not annotated consecutively (this could influence the annotators because HR is similar for resting PPGs of one subject). Once all the experts completed the HR estimations, we attempted to reach consensus for quality annotations. Only the “good”annotations were considered, meaning those HR values that differed from the reference HR by less than or equal to 5 bpm. The maximal error value of ±5 bpm is based on the international standard IEC 60601-2-27 [11]. The norm enables accuracy of ±10% or ±5 bpm (higher of these values) [11]. In the BUT PPG database, all the HR values are higher than 60 bpm; thus, the value of 5 bpm selected by us is even more strict than the norm. When at least three annotators provided good annotations, the signal 2 BioMed Research International
was labelled as 1 (“good quality”). When this condition was not met, the signal was labelled as 0 (“poor quality”). 2.3. Annotating Software. The main window of the tailormade annotating software is shown in Figure 3. The software includes the picture of the original PPG signal (left and right upper graphs), decomposition of the PPG signal using stationary wavelet transform (SWT) (five graphs on the left side), mean and median HR estimated from each frequency band, spectrum created using fast Fourier transform (right lower graph), and HR estimated from spectrum. In Figure 3, the 2nd to the 6th frequency bands after SWT are shown. In each of them, the peaks are automatically found and marked by the red circles. Automatic detection of peaks was provided by finding local maxima using minimum peak distance rule (set to 8) and minimum peak prominence rule (set to 0.01). Using the distance between marked peaks, the mean and median HR are calculated and shown in the title of each graph. All the graphs (except for spectrum) are equipped with a ruler (black line with the number in the middle) for the manual measurement of the distances between peaks within the signals. From the spectrum, the HR was determined automatically as the maximum in the frequency range of 0.5-4 Hz, which corresponds with HR of 30-240 bpm, and it is shown in the title of the spectrum. 3. Results 3.1. Data Records. The data are publicly available on PhysioNet [12, 13]. Each record contains PPG signal and one-lead ECG signal recorded with a sampling frequency of 30Hz and 1,000Hz, Signal acqusition 23 51 54 61 24 21 59 23 24 21 50 21 1 rest 2 rest 3 rest 4 move ECG 1 rest 2 rest 3 rest 4 move PPG Selection of red channel Frames averaging Inversion manualsynchronization selectionof the middle10 s Annotation of HR Consensus error ≤ 5 bpm Quality 1 Quality 0 QRS detection& manual verification Median HR calculation Reference HR Subjects Tailor-made SW E1 E2 E3 E4 E5 Experts + – Figure 1: Schematic overview of the study. The diagram shows the process of creating the BUT PPG database. 0 1000 2000 3000 4000 ECG (𝜇V) PPG (–) 0123456789 Time (s) Time (s) –254.6 –254.4 –254.2 1000 2000 3000 4000 5000 0123456789 –250 –200 –150 1010 Good quality Poor quality Figure 2: Examples of good (signal no. 6) and poor (signal no. 28) quality PPG signals. In both cases, the PPG signals are shown together with the reference ECG signals. 3BioMed Research International
respectively. All signals are provided in the WaveForm Database (WFDB) format. The names (IDs) of the recordings are six-digit numbers where the first three numbers are unique subject identifiers and the next three numbers indicate the measurement number of this subject. The PPG and ECG signals are in two separate files: ∗_PPG.dat,∗_PPG.hea,and∗ _ECG.dat,∗_ECG.hea, respectively. The annotations QUALITY-HR-ANN.csv are recorded in a CSV file with three columns. The first column contains signal IDs. The second column contains a binary quality indicator (1: “good quality”;0:“poor quality”). The third column contains the reference HR. The raw HR estimations from each annotator are available in raw-HR-annotations.csv. Patient demographics (gender, age, and weight) and motion information are provided in subject-info.csv. 3.2. Technical Validation. In Table 1, there are shown median absolute errors between experts’PPG HR estimations and ECG HR reference. Two experts have median error of 1.5 bpm, and three of them have 2 bpm. Moreover, Table 1 shows the number of signals with HR error equal or lower than 5 bpm (based on the international standard IEC 60601-2-27 [11]) for each expert. In Table 2, it is shown how annotators’results match with the reference. Altogether, 31 signals were classified as quality 1 (“good quality”) according to all annotators (the error of annotators’HR estimation was lower than or equal to 5 bpm in comparison with the reference). Three signals were classified as quality 1 (“good quality”), when 4 annotators’HR estimations match with the reference. Threeannotator match was reached for only one signal no. 12 (the lowest certainty). In two signals, only 2 annotators match with reference; thus, the signal was classified as quality 0 (“poor quality”) with the lowest certainty. In three signals, the HR was correctly estimated by one expert; thus, the signals belong to quality 0 (“poor quality”) class. In 8 signals, none of the experts estimated HR correctly; thus, these signals belong unambiguously to quality 0 (“poor quality”). From this table, it is evident that in case of 39 of 48 signals, the signals were classified into one of the two groups according to all of the experts. In 6 signals, only one expert was not in accordance with others. Only three signals were classified into the particular group according to three experts. 12345678910 -0.5 0 0.5 PPG no. 1 4.98 12345678910 –0.4 –0.2 0 0.2 0.4 SWT–2nd band; HRmean= 119 bpm; HRmedian= 106 bpm 4.98 12345678910 –0.5 0 0.5 SWT–3rd band; HRmean= 117 bpm; HRmedian= 106 bpm 4.98 12345678910 –0.5 0 0.5 SWT–4th band; HRmean= 96 bpm; HRmedian= 95 bpm 3.07 12345678910 –1 –0.5 0 0.5 SWT–5th band; HRmean= 54 bpm; HRmedian= 46 bpm 4.98 12345678910 Time (s) Time (s) –1 0 1 SWT–6th band; HRmean= 30 bpm; HRmedian= 30 bpm 4.98 0 5 10 15 20 25 30 0 5 10 15 20 25 30 35 40 Spectrum of PPG signal; HR = 96 bpm 12345678910 –0.6 –0.4 –0.2 0 0.2 0.4 0.6 PPG (–) PPG 3.76 0.23-0.47 Hz 14-28 bpm 0.47-0.94 Hz 28-56 bpm 0.94-1.9 Hz 56-113 bpm 1.9-3.8 Hz 113-225 bpm 3.8-7.5 Hz 225-450 bpm Frequency (Hz) Figure 3: The main window of the tailor-made annotating software. (a) The original PPG signal. The five graphs on (b) show the selected frequency bands of the signal decomposed by the SWT. (c) The original PPG signal plotted especially for the manual measurement of the distances between peaks. (d) The spectrum of the signal. Red circles represent automatically found peaks of the signals. Black lines with numbers in the middle are the rulers for the manual measurement of distances between peaks within the signals. 4 BioMed Research International
Both tables show that there is only a small variance between annotators. Tables 1 and 2 prove that the annotations of PPG signal quality are reliable. 4. Discussion The reference HR annotations were made by ECG expert with the use of an advanced QRS complex detection algorithm [10]. All the ECG signals were manually checked whether the median HR can be determined from them. Thus, the reference HR annotations can be considered as reliable. All five ECG experts who annotated the PPG signals are biomedical engineers with master’s or doctoral degrees. All of them have at least 10-year experience with processing and analysis of biomedical signals. Two of them have a 7year practice from a cardiology clinic and have experience with annotating biomedical signals [14, 15]. According to our results, 10 s PPG signal is appropriate to precisely estimate the HR (analogous to HR estimation during standard clinical ECG procedure). Another advantage of 10 s signals is that there is minimal risk of data loss during measurement and presence of artifacts (when they were not evoked) unlike long-time measurements. Shorttime measurements are also comfortable for the measured subjects and do not limit them. Short-time measurement has low demands on energy, battery capacity, and memory. Moreover, BUT PPG database is quite diverse—the data were measured both in rest and in move, subjects aged 2163 participated. It includes HR annotations and annotations of PPG signal quality. The described results and annotations contribute to credibility of smartphone use for medical purposes—home self-monitoring of HR (by anyone—children, adults, seniors, athletes, …). 4.1. Limitation of the Study. BUT PPG database includes 48 10 s signals, which can be insufficient, e.g., PPG signal analysis based on artificial intelligence. This paper includes enough methodological information to enable future extensions of the database by our team (which is planned) or even by other teams. 5. Conclusion As the only available database containing PPG signals recorded by smartphone, BUT PPG is a unique tool for the development of smart, user-friendly, cheap, on-the-spot, self-home-monitoring of HR with the potential of widespread using. A benefit is that it includes PPG quality annotations which can enhance the robustness and reliability of HR estimation algorithms. BUT PPG with PPG quality annotations can be used also separately for developing and testing quality estimation algorithms. One of the unique features of the database is that the quality annotation of each PPG signal is based on 5 experts and their consensus, meeting international standard guidelines, and moreover meeting the more strict variant. This database has a great potential to contribute in the swiftly developing area of well-being, telemedicine, and self-monitoring in home environment by the omnipresent cell phone. Data Availability The data are publicly available on PhysioNet: https:// physionet.org/content/butppg/1.0.0/. Disclosure The funders had no role in the design of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript; or in the decision to publish the results. Conflicts of Interest The authors declare no conflict of interest. Acknowledgments This work has been funded by the United States Office of Naval Research Global, award number N62909-19-1-2006. The authors wish to thank LCDR Joshua Swift from ONR Code 342 and Dr. Stephen O’Regan from ONR Global Central and Eastern European Office for their support. References [1] C. Orphanidou, Signal Quality Assessment in Physiological Monitoring State of the Art and Practical Considerations, Springer, Cham, Switzerland, 1st edition, 2018. [2] E. K. Naeini, I. Azimi, A. M. Rahmani, P. Liljeberg, and N. Dutt, “A real-time PPG quality assessment approach for healthcare internet-of-things,”Procedia Computer Science, vol. 151, pp. 551–558, 2019. [3] A. Nemcova, I. Jordanova, M. Varecka et al., “Monitoring of heart rate, blood oxygen saturation, and blood pressure using a smartphone,”Biomedical Signal Processing and Control, vol. 59, 2020. [4] N. Koenig, A. Seeck, J. Eckstein et al., “Validation of a new heart rate measurement algorithm for fingertip recording of video signals with smartphones,”Telemedicine and e-Health, vol. 22, no. 8, pp. 631–636, 2016. Table 1: Successfulness of PPG expert annotators. E1 E2 E3 E4 E5 Median error (bpm) 1.5 1.5 2 2 2 No. of errors ≤5 bpm 36 36 34 35 36 Table 2: Match rate of experts with the reference ECG HR. Match of experts with reference ECG HR 5×4×3×2×1×0× No. of signals 31 3 1 2 3 8 5BioMed Research International
[5] S. A. Siddiqui, Y. Zhang, Z. Feng, and A. Kos, “A pulse rate estimation algorithm using PPG and smartphone camera,” Journal of Medical Systems, vol. 40, no. 5, 2016. [6] F. Tabei, R. Zaman, K. H. Foysal, R. Kumar, Y. Kim, and J. W. Chong, “A novel diversity method for smartphone camerabased heart rhythm signals in the presence of motion and noise artifacts,”PLoS One, vol. 14, no. 6, p. e0218248, 2019. [7] Garmin, The heart rate sensor on my Garmin watch is not accurate, Garmin Support Center, United States, 2021, https://support.garmin.com/en-US/?faq= xQwjQjzUew4BF1GYcusE59. [8] Polar, “Training with wrist-based heart rate,”2021, M430 User Manual https://support.polar.com/e_manuals/M430/Polar_ M430_user_manual_English/Content/Training-with-wristbased-heart-rate.htm. [9] R. Peng, X. Zhou, W. Lin, and Y. Zhang, “Extraction of heart rate variability from smartphone photoplethysmograms,” Computational and Mathematical Methods in Medicine, vol. 2015, Article ID 516826, 11 pages, 2015. [10] L. Smital, L. Marsanova, R. Smisek, A. Nemcova, and M. Vitek, “Robust QRS detection using combination of three independent methods,”in 2020 Computing in Cardiology, pp. 1–4, Rimini, Italy, 2020. [11] International Electrotechnical Commission, “Medical Electrical Equipment,”Particular Requirements for the Basic Safety and Essential Performance of Electrocardiographic Monitoring Equipment (IEC 60601-2-27), 2015. [12] A. Nemcova, R. Smisek, E. Vargova, L. Maršánová, M. Vitek, and L. Smital, “Brno University of Technology Smartphone PPG Database (BUT PPG),”PhysioNet, 2021. [13] A. Goldberger, L. A. Amaral, L. Glass et al., “PhysioBank, PhysioToolkit, and PhysioNet: components of a new research resource for complex physiologic signals,”Circulation, vol. 101, no. 23, pp. e215–e220, 2000. [14] R. Smíšek, L. Maršánová, A. Němcová, M. Vítek, J. Kozumplík, and M. Nováková, “CSE database: extended annotations and new recommendations for ECG software testing,”Medical & Biological Engineering & Computing, vol. 55, no. 8, pp. 1473– 1482, 2017. [15] A. Nemcova, R. Smisek, K. Opravilová, M. Vitek, L. Smital, and L. Maršánová, “Brno University of Technology ECG Quality Database (BUT QDB),”PhysioNet, 2020. 6 BioMed Research International