Sensors 2015, 15, 2438-2452; doi:10.3390/s150202438 sensors ISSN 1424-8220 www.mdpi.com/journal/sensors Article A 3D Microfluidic Chip for Electrochemical Detection of Hydrolysed Nucleic Bases by a Modified Glassy Carbon Electrode Jana Vlachova 1,2, Katerina Tmejova 1,2, Pavel Kopel 1,2, Maria Korabik 3, Jan Zitka 1, David Hynek 1,2, Jindrich Kynicky 4, Vojtech Adam 1,2 and Rene Kizek 1,2,* 1 Department of Chemistry and Biochemistry, Faculty of Agronomy, Mendel University in Brno, Zemedelska 1, Brno CZ-613 00, Czech Republic; E-Mails:
[email protected] (J.V.); [email protected] (K.T.);
[email protected] (P.K.); [email protected] (J.Z.); [email protected] (D.H.); vojtech[email protected] (V.A.) 2 Central European Institute of Technology, Brno University of Technology, Technicka 3058/10, Brno CZ-616 00, Czech Republic 3 Faculty of Chemistry, University of Wroclaw, 14 Joliot-Curie, Wroclaw PL-50383, Poland; E-Mail:
[email protected] 4 Karel Englis College, Sujanovo nam. 356/1, Brno CZ-602 00, Czech Republic; E-Mail:
[email protected] * Author to whom correspondence should be addressed; E-Mail: [email protected]; Tel.: +420-5-4513-3350; Fax: +420-5-4521-2044. Academic Editor: Libuse Trnkova Received: 22 October 2014 / Accepted: 4 January 2015 / Published: 22 January 2015 Abstract: Modification of carbon materials, especially graphene-based materials, has wide applications in electrochemical detection such as electrochemical lab-on-chip devices. A glassy carbon electrode (GCE) modified with chemically alternated graphene oxide was used as a working electrode (glassy carbon modified by graphene oxide with sulphur containing compounds and Nafion) for detection of nucleobases in hydrolysed samples (HCl pH = 2.9, 100 °C, 1 h, neutralization by NaOH). It was found out that modification, especially with trithiocyanuric acid, increased the sensitivity of detection in comparison with pure GCE. All processes were finally implemented in a microfluidic chip formed with a 3D printer by fused deposition modelling technology. As a material for chip fabrication, acrylonitrile butadiene styrene was chosen because of its mechanical and chemical stability. The chip contained the one chamber for the hydrolysis of the nucleic acid and OPEN ACCESS
Sensors 2015, 15 2439 another for the electrochemical detection by the modified GCE. This chamber was fabricated to allow for replacement of the GCE. Keywords: biosensor; lab-on-chip; electrochemistry; graphene; modification; hydrolysis; trithiocyanuric acid 1. Introduction The microfluidic systems called lab-on-chip devices enables one to create fast and low cost detection systems with ultralow sample consumption [1]. Due to the fact that the electrochemical detection is still very attractive for its high sensitivity and inherent simplicity [2–4], some of these are often used to construct the mentioned microfluidic systems, including those aimed at detection of viruses [5] or DNA/RNA nucleobases [6,7]. However, the sensitivity of such microfluidic devices is still an issue. In this field, the utilization of various electrode surfaces such as carbon-based modifications and boron-doped diamond electrodes have enhanced the sensitivity and reproducibility [8]. For electrochemical detection of nucleobases by a glassy carbon electrode (GCE), the modification of graphene-based compounds altered by Nafion [9] or Nafion with TiO2 [10] has been also used to increase the sensitivity. Graphene oxide (GO) is a multilayer system created by the oxidation of graphite and therefore oxygenated functionalities are introduced in the graphite structure [11]. This fact causes hydrophilicity and GO becomes very well soluble in water or other solvents utilized for numerous purposes [12–14]. GO has been widely applied for modification [15,16] of glassy carbon electrodes [17] used for the detection of metal ions, proteins, DNA mutations and other compounds [18]. Trithiocyanuric acid (ttc), also referred to as trimercaptotriazine, has a symmetric structure with three nitrogen atoms in a ring and three sulphur atoms connected to carbons, whereby these six atoms are highly reactive and can form bridges connecting central atoms. To illustrate the importance of trithiocyanuric acid, its use as a remediation agent for removal of heavy metals from industrial waters should be mentioned [19–21]. Removal of residual palladium from reaction mixtures in drug production is an especially important application of ttcNa3 [22–24]. This acid was also applied in plating processes, in production of composite materials with metals and rubbers and as an anticorrosion agent [25–28]. In this work, modified graphene oxide was used to develop a simple and sensitive method for detection of adenine in a hydrolysed miRNA sample. The modified graphene oxide was firstly dissolved in dimethylformamide and subsequently modified by Nafion and sonicated in an ultrasonic bath. The mixture was deposited on a glassy carbon surface. This way the layer for each measurement was prepared. The described modification allows adenine detection in the hydrolysed miRNA sample and it was possible to detect other nucleic bases too. Moreover, hydrolysis and electrochemical detection was carried out within a microfluidic chip with a chamber that allows repeatable input and output of the modified working electrode.
Sensors 2015, 15 2440 2. Experimental Section 2.1. Chemicals Ttc, thiodiacetic acid (tda), mercaptosuccinic acid (MSA), Nafion 117 (formulas are shown in Figure 1), titanium(IV) isopropoxide, graphite flakes, NaNO 3 , KMnO 4 , sodium acetate trihydrate, acetic acid, Hg(NO 3 ) 2 , water and other chemicals were purchased from Sigma-Aldrich (St. Louis, MO, USA) in ACS purity (chemicals meet the specifications of the American Chemical Society). MiRNA was also purchased from Sigma-Aldrich and its sequence was 5' UAA GGC ACG CGG UGA AUG CCA AAA AAA AAA 3'. Standard nucleobases were adenine and cytosine (MP Biomedicals, Santa Ana, CA, USA), guanine and uracil (Sigma-Aldrich, St. Louis, MO, USA). Figure 1. Schemes of compounds used for modification of graphene oxide. (A) ttc; (B) tda; (C) MSA; (D) Nafion (NAF). 2.2. GO Modifications 2.2.1. Preparation of GO GO was prepared from graphite flakes by the Hummers method [29]. Briefly, graphite (2 g), NaNO 3 (1 g) and KMnO 4 (6 g) were added with stirring to concentrated H 2 SO 4 (96%, w/w, 46 mL) placed in an ice bath. The mixture was stirred overnight with gradual temperature growth. Water (400 mL) was slowly added and the mixture was heated at 90 °C for 2 h. Subsequently, H 2 O 2 (3%, v/v) was slowly
Sensors 2015, 15 2441 added till colour turned to yellow. The product was decanted and washed several times with water. The final volume of the GO suspension was 200 mL. 2.2.2. Preparation of Modified GO (GO-ttc, GO-tda, and GO-MSA) In general, the samples were prepared as described in the previous paragraph by the following method: GO suspension (25 mL) in a 100 mL beaker was mixed with thiol compound (ttc, tda, or MSA in the amount of 0.5 g) dissolved in water (5 mL). Samples were diluted with water and solution was shaken on orbital shaker overnight. The next step was decantation and washing several times with water. Samples were collected by centrifugation and dried at 60 °C. 2.2.3. Preparation of GO-TiO2 Solution GO-ttc (1 mg) in water: ethanol (1:1) mixture (0.5 mL) was sonicated (1 h). Titanium (IV) isopropoxide (40 µL) was added and sonicated for 2 h. The sample was centrifuged and solvent was discarded. Dimethylformamide was added (0.85 mL) and sonicated for 1 h. Finally, Nafion 117 (100 µL) was added and sample was sonicated for 0.5 h. This solution was used for glassy carbon modification. 2.2.4. Preparation of GO-thiol-Nafion Solutions (GO-ttc-NAF, GO-tda-NAF, and GO-MSA-NAF) These solutions were prepared as follows. 1 mg of GO-thiol was suspended and sonicated in dimethylformamide (1 mL) for 2 h. Then, Nafion 117 (100 µL) was added and sonicated for 0.5 h. This solution was used for glassy carbon modification. 2.3. X-ray Fluorescence Analysis (XRF) The samples were analysed on Spectro Xepos (Spectro Analytical Instruments, Kleve, Germany) using anode X-ray tube with Pd anode working at a voltage of 44.69 kV and a current of 0.55 mA. Signals were detected with Barkla scatter aluminium oxide for 300 s. For excitation three secondary targets (Mo, Al2O3 and high-ordered pyrolithic graphite crystal) were used. The excitation geometry was 90°. The samples were measured through the polyethylene bottle side wall 20 mm above the bottom. The Spectro Xepos software and TurboQuant method were applied for data analysis. 2.4. Elemental Detection C, H, N and S detection of samples was carried out on Flash 2000 (Thermo Scientific, Waltham, MA, USA). 2.5. Fourier Transformation Infrared Spectroscopy (FTIR) FTIR study was performed over the range of wavenumber 4000–400 cm−1 by a Bruker Vertex 70 FTIR spectrometer (Bruker, Ettlingen, Germany) using KBr pellets.
Sensors 2015, 15 2442 2.6. Sample Preparation for Electrochemical Detection of Nucleic Bases 24 µL of HCl (pH = 2.9) was added to 24 µL of 22 µM miRNA and mixture was incubated at 99 °C for 1 h in Thermomixer comfort (Eppendorf Czech & Slovakia s.r.o., Ricany, Czech Republic). After hydrolysis the mixture was neutralized by equimolar quantity of NaOH. Hydrolysed samples were analysed electrochemically. 2.7. Electrochemical Detection Electrochemical detection was carried out using a CH Instruments Electrochemical Workstation (CH Instruments, Bee Cave, TX, USA) and glass cell with three electrodes. Ag/AgCl/3M KCl was used as a reference electrode and platinum as a counter electrode. As a working electrode, modified GCE was used. This GCE was modified by various types of chemically modified graphene oxide (Table 1, Figure 1). Five µL of chemically modified graphene oxide in dimethylformamide was applied on the GCE and dried with a hair drier for 15 min. For all measurement differential pulse voltammetry was used and measurement parameters were as follows: initial potential 0.2 V, final potential 1.5 V for adenine and 1.8 V for simultaneous detection of bases and hydrolysed sample, step potential 0.004, modulation amplitude 0.05 V, pulse width 0.2 s, sampling width 0.0167 s, pulse period 0.5 s, quiet time 10 s. All experiments were performed at room temperature. As the supporting electrolyte the acetate buffer (pH = 5, 0.2 M CH3COOH and 0.2 M CH3COONa) was used. 1 ml of electrolyte was used for adenine detection and 0.5 mL for simultaneous detection of bases and hydrolysed sample. The limit of detection was calculated by LOD = .× , SD = standard deviation of the response and S = slope of the calibration curve. Table 1. Chemical agents used for modification of graphene oxide. Abbreviation Applied Modifiers GO-MSA-NAF mercaptosuccinic acid and Nafion GO-tda-NAF thiodiacetic acid and Nafion GO-ttc trithiocyanuric acid GO-ttc-NAF trithiocyanuric acid and Nafion GO-TiO2 trithiocyanuric acid, Nafion and titanium dioxide 2.8. Microfluidic Detection System Electrodes were placed to a holder fabricated by using a 3D printer (Profi3Dmaker, 3D Factories, Straznice, Czech Republic) using Fused Deposition Modelling (FDM) technology. For electrochemical detection in the microfluidic system different electrodes were chosen due to their size. Standard Ag/AgCl/3M KCl was replaced with quasi-reference electrode as graphite lead and as a counter electrode a platinum electrode with diameter 0.5 mm (CH Instruments, Bee Cave, TX, USA) was used.
Sensors 2015, 15 2443 3. Results and Discussion 3.1. Characterisation of Graphene Modified Materials 3.1.1. X-Ray Fluorescence Spectroscopy and Elemental Analysis Modified GO was prepared by mixing of a water suspension of GO with sulphur-containing compounds. It was easily seen that the samples changed in colour from light brown to deep brown or black during overnight mixing. Prepared samples were decanted and washed several times with water to remove excess reagents. Solid samples were obtained by centrifugation and drying. Composition of the modified GO was proved by X-ray fluorescence spectroscopy and elemental analysis. The presence of sulphur in all the samples was proved by peak characteristic for S Kα , which appeared in all XRF spectra. The peak position was always the same, whereas the intensity of the peak was changing. That is why only one typical spectrum is shown in Figure 2A (top) for GO-ttc. XRF spectrum of GO-TiO 2 , where peak Ti Kα was also observed (Figure 2A (bottom)) and thus presence of Ti in the sample was unambiguously proved, was the only exception. C, H, N, and S elemental analysis of samples also proved the presence of sulphur and content of carbon and nitrogen in the case of trithiocyanuric acid (Table 2). Figure 2. Characterization of graphene oxide modified materials. (A) XRF analysis of GO-ttc (dashed line) and GO-TiO 2 (solid line); (B–D) IR spectra of studied materials.
Sensors 2015, 15 2444 Table 2. Elemental analysis results of various modified graphene oxides. Nitrogen (%) Carbon (%) Hydrogen (%) Sulphur (%) GO 0 84.15 0.44 0 GO-MSA 0 43.27 2.34 15.45 GO-tda 0 38.06 2.89 20.77 GO-ttc 10.14 52.95 1.35 25.95 GO-TiO2 6.56 34.12 0.78 16.02 3.1.2. FTIR Analysis FTIR spectra of GO and modified GO are shown in Figures 2B–D. The presence of oxygen containing groups such as carbonyl, carboxylic acid, epoxy and hydroxyl in the GO is proved by the presence of the corresponding characteristic peaks. These peaks can be attributed to ν(C=O) stretching (1663 cm−1), C-O-C stretching (1193 cm−1), ν(C-O) stretching (1072 cm−1) and a broad shoulder with a maximum at 3480 cm−1 can be assigned to vibration of hydroxyl groups [30]. The spectrum of GO modified with Nafion (GO-NAF) (not shown) was also measured. In the GO-NAF spectrum there is very broad band from 400–800 cm−1, probably due to CO vibrations, and very weak peaks in the 1212–1232 cm−1, 1150–1160 cm−1 and 1056–1105 cm−1 regions that can be assigned to νas(CF2), νs(CF2) and νs(SO), respectively [31]. The same peaks, of weak or medium intensity, were found in the spectra of Nafion-containing samples. In the spectra of all compounds, except for the GO spectrum, there is very strong peak near 670 cm−1 characteristic for sulphur-containing compounds, attributable to ν(C-S). In the spectra of GO-tda-NAF and GO-MSA-NAF (Figure 2C) there are, in comparison with the GO spectrum, more intense peaks near 1560 and 1388 cm−1, which can be attributed to νas(COO) and νs(COO) vibrations, respectively [32]. In the spectra of ttc-modified GO, peaks in the 1248–1265 and 1529–1589 cm−1 regions were found, which can be attributed to ν(C-N) vibrations of the heterocyclic ring and peaks near 870 cm−1 which are most probably connected with ν(C-S) vibrations [33]. In the spectrum of GO-TiO2 a shoulder from 400 to 670 cm−1 was observed, which is probably caused by the presence of the Ti-O stretching vibration. Thus, the modification of GO by TiO2 on the surface was confirmed. 3.2. Electrochemical Detection of Nucleobases in Electrochemical Cell The first step of the detection part of the experiment involved testing of five variously modified graphene oxides for adenine detection. The electrochemical detection was performed by differential pulse voltammetry using a GCE modified with specific graphene oxides. Figure 3A depicts voltammograms of these modifications with the characteristic peak for adenine, which was detected at approximately 1.0 V. The highest sensitivities were found for modifications with trithiocyanuric acid. According to [9,10], we attempted to improve the sensitivity by Nafion and TiO2, respectively. In the case of modification by TiO2, the sensitivity decreased, but for GO-ttc-NAF higher sensitivity than without Nafion was found. Voltammograms of GO-ttc, GO-ttc-NAF and GO-TiO2 show another peak at a potential of about 1.4 V. This peak is probably connected with oxidation of the material itself. The high sensitivity of trithiocyanuric acid to nucleobases was probably caused by its ability to form hydrogen bonds with compounds containing nitrogen [34]. The addition of Nafion had a few
Sensors 2015, 15 2445 supplemental positive results. It prevented graphene from aggregation and increased the stability of the immobilisation of graphene on GCE. The negative charge of Nafion can facilitate adsorption of positively charged nucleobases. Nafion also increased the oxidation signal due to the easier electron transfer [9]. Based on these advantages we assume this modification GO-ttc-NAF has great potential for wide electrochemical application. The next step was the determination of calibration curves for adenine on unmodified GCE and GCE modified by GO-ttc-NAF. Both curves fitted by linear regression are depicted in Figure 3B. Analytical parameters like regression parameters, standard deviation, limit of detection and quantification are listed in Tables 3 and 4. We observed quite similar limits of detection for unmodified and modified GCE. Although the sensitivity of applied electrochemical method using modified GCE was five times higher, the dissimilar surface of the modified electrode caused a high standard deviation which resulted in a detection limit decrease. Figure 3. (A) Detection of 200 µM adenine on GCE modified by different graphene oxide; (B) Calibration curve of adenine on unmodified GCE and GCE modified by GO-ttc-NAF; repeated in triplicates; (C) Simultaneous detection of guanine (262 µM), adenine (51 µM) and cytosine (273 µM) with uracil (255 µM) by unmodified GCE and GCE modified by GO-ttc-NAF; (D) Optimization of hydrolysis temperature measured by GCE modified with GO-ttc-NAF, concentration of adenine and guanine was 9 µM and 4 µM respectively.
Sensors 2015, 15 2446 Table 3. Analytical parameters of electrochemical detection of unmodified GCE, n = 3. Substance Regression Equation Linear Dynamic Range (μM) Linear Dynamic Range (μg/mL) R 1,2 LOD 2 (μM) LOD (μg/mL) LOQ 3 (μM) LOQ (μg/mL) RSD 4 (%) unmodified GCE y = 0.0342x − 0.0066 5.18–51.30 0.70–6.93 0.9979 0.67 0.09 2.23 0.30 3.72 1 Coefficients of determination; 2 Limits of detection of detector (3 S/N); 3 Limits of quantification of detector (10 S/N); 4 Relative standard deviations. Table 4. Analytical parameters of electrochemical detection of GCE modified by CO-ttc-NAF, n = 3. Substance Regression Equation Linear Dynamic Range (μM) Linear Dynamic Range (μg/mL) R 1,2 LOD 2 (μM) LOD (μg/mL) LOQ 3 (μM) LOQ (μg/mL) RSD 4 (%) modified GCE y = 0.1894x − 0.0074 2.59–51.30 0.35–6.93 0.9766 0.66 0.09 2.21 0.30 5.95 1 Ccoefficients of determination; 2 Limits of detection of detector (3 S/N); 3 Limits of quantification of detector (10 S/N); 4 Relative standard deviations. Furthermore, simultaneous detection of all nucleobases was performed on GCE modified by GO-ttc-NAF. Figure 3C shows a comparison between detection on this modified GCE and unmodified GCE. By using of GCE modified by GO-ttc-NAF three peaks may be observed. The peaks at potentials of 0.75 V and 1.0 V belong to guanine and adenine, respectively. The last peak at a potential of about 1.4 V was a signal of cytosine and uracil, which had peaks at the same potential and thus it is not possible to distinguish between them. The one disadvantage of GCE modified by GO-ttc-NAF was the peak of the material itself, which occurred at the same potential as the peak for cytosine and uracil. For this reason it was not possible to determine small amounts of cytosine and uracil. It should be noted that the best sensitivity was determined for adenine which shows a very high peak at a concentration of 51 µM in contrast to other nucleobases, whose concentrations were about 260 µM. 3.3. Optimization of the Hydrolysis Process The optimization was carried out for adenine and guanine detection at different temperatures of hydrolysis as follows: 25 °C, 40 °C, 60 °C, 80 °C and 99 °C (Figure 3D). At 25 °C and 40 °C no signal was observed for either adenine or guanine. At 60 °C the adenine peak occurred, but the peak for guanine was not observed. Both peaks were found for temperatures of hydrolysis of 80 °C and 99 °C. The difference between the signals was insignificant (about 0.1 µA). 3.4. Microfluidic System Connection of the hydrolysis process and electrochemical detection of nucleobases was the great challenge and therefore microfluidic system was designed and fabricated. The microfluidic system was composed of two basic parts: the first part was for hydrolysis, the second one for the electrochemical detection. The principle scheme of the suggested system is shown in Figure 4A. The fluidic system was realized by plastic tubes. The liquid movements were done using three syringe pumps, three three-step valves and three magnetic valves. The microfluidic chip was printed by a 3D printer and as