Analysis of the 40K contamination in NaI(Tl) crystals from different providers in the frame of the ANAIS project

Cuesta, Clara; Villar, José Ángel; Martínez, María; Ginestra, Carlos; Pobes, Carlos; Ortigoza, Ysrael; Puimedón, Jorge; Amaré, Julio; Cebrián, Susana; García, Eduardo; Ortíz de Solórzano, Alfonso; Oliván, Miguel A; Villar, Patricia; Sarsa, María Luisa

Abstract

NaI(Tl) large crystals are applied in the search for galactic dark matter particles through their elastic scattering off the target nuclei in the detector by measuring the scintillation signal produced. However, energies deposited in the form of nuclear recoils are small, which added to the low efficiency to convert that energy into scintillation, makes that events at or very near the energy threshold, attributed either to radioactive backgrounds or to spurious noise (nonbulk NaI(Tl) scintillation events), can compromise the sensitivity goals of such an experiment. DAMA/LIBRA experiment, using 250 kg NaI(Tl) target, reported first evidence of the presence of an annual modulation in the detection rate compatible with that expected for a dark matter signal just in the region below 6 keVee (electron equivalent energy). In the frame of the ANAIS (Annual modulation with NaI Scintillators) dark matter search project a large and long effort has been carried out in order to understand the origin of events at very low energy in large sodium iodide detectors and develop convenient filters to reject those nonattributable to scintillation in the bulk NaI(Tl) crystal. 40K is probably the most relevant radioactive contaminant in the bulk for NaI(Tl) detectors because of its important contribution to the background at very low energy. ANAIS goal is to achieve levels at or below 20 ppb natural potassium In this paper we will report on our effort to determine the 40K contamination in several NaI(Tl) crystals, by measuring in coincidence between two (or more) of them. Results obtained for the 40K content of crystals from different providers will be compared and prospects of the ANAIS dark matter search experiment will be briefly reviewed Cuesta, Clara; Amaré, Julio; Cebrián, Susana; García, Eduardo; Ginestra, Carlos; Martínez, María; Oliván, Miguel A; Ortigoza, Ysrael; Ortíz de Solórzano, Alfonso; Pobes, Carlos; Puimedón, Jorge; Sarsa, María Luisa; Villar, José Ángel; Villar, Patricia

Full text

July 30, 2014 9:7 WSPC/139-IJMPA S0217751X14430106 International Journal of Modern Physics A Vol. 29, No. 19 (2014) 1443010 (20 pages) c World Scientific Publishing Company DOI: 10.1142/S0217751X14430106 Analysis of the 40K contamination in NaI(Tl) crystals from different providers in the frame of the ANAIS project C. Cuesta,∗J. Amar´e, S. Cebri´an,E.Garc´ıa, C. Ginestra, M. Mart´ınez,† M. A. Oliv´an, Y. Ortigoza, A. Ortiz De Sol´orzano,C.Pobes, ‡ J. Puimed´on, M. L. Sarsa, J. A. Villar and P. Villar Laboratorio de F´ısica Nuclear y Astropart´ıculas, Universidad de Zaragoza, C/Pedro Cerbuna 12, 50009 Zaragoza, Spain Laboratorio Subterr´aneo de Canfranc, Paseo de los Ayerbe s.n., 22880 Canfranc Estaci´on, Huesca, Spain ∗ccuest[email protected]du Received 24 January 2014 Accepted 1 March 2014 Published 15 July 2014 NaI(Tl) large crystals are applied in the search for galactic dark matter particles through their elastic scattering off the target nuclei in the detector by measuring the scintillation signal produced. However, energies deposited in the form of nuclear recoils are small, which added to the low efficiency to convert that energy into scintillation, makes that events at or very near the energy threshold, attributed either to radioactive backgrounds or to spurious noise (nonbulk NaI(Tl) scintillation events), can compromise the sensitivity goals of such an experiment. DAMA/LIBRA experiment, using 250 kg NaI(Tl) target, reported first evidence of the presence of an annual modulation in the detection rate compatible with that expected for a dark matter signal just in the region below 6 keVee (electron equivalent energy). In the frame of the ANAIS (Annual modulation with NaI Scintillators) dark matter search project a large and long effort has been carried out in order to understand the origin of events at very low energy in large sodium iodide detectors and develop convenient filters to reject those nonattributable to scintillation in the bulk NaI(Tl) crystal. 40K is probably the most relevant radioactive contaminant in the bulk for NaI(Tl) detectors because of its important contribution to the background at very low energy. ANAIS goal is to achieve levels at or below 20 ppb natural potassium. In this paper we will report on our effort to determine the 40K contamination in several NaI(Tl) crystals, by measuring in coincidence between two (or more) of them. Results ∗Presently at Center for Experimental Nuclear Physics and Astrophysics, and Department of Physics, University of Washington, Seattle, WA, USA. †Fundaci´on ARAID, Mar´ıa de Luna 11, Edificio CEEI Arag´on, 50018 Zaragoza, Spain. ‡Presently at Instituto de Ciencia de Materiales de Arag´on, University of Zaragoza – CSIC. 1443010-1 Int. J. Mod. Phys. A 2014.29. Downloaded from www.worldscientific.com by UNIVERSITY OF ZARAGOZA on 11/06/15. For personal use only. July 30, 2014 9:7 WSPC/139-IJMPA S0217751X14430106 C. Cuesta et al. obtained for the 40K content of crystals from different providers will be compared and prospects of the ANAIS dark matter search experiment will be briefly reviewed. Keywords: Sodium iodide; scintillation; potassium; dark matter search; annual modulation. PACS numbers: 29.40.Mc, 29.40.Wk, 95.35.+d 1. Introduction ANAIS project aims at the study of the annual modulation signal attributed to galactic dark matter particles using 250 kg NaI(Tl) scintillators at the Canfranc Underground Laboratory (LSC), in Spain. NaI(Tl) large crystals have been applied for a long time in the search for galactic dark matter particles through their elastic scattering off the target nuclei in the detector by measuring the weak scintillation signal produced.1–8However, energies deposited in the form of nuclear recoils are small, which added to the low efficiency to convert that energy into scintillation, makes that events at or very near the energy threshold, attributed either to radioactive backgrounds or to spurious noise (nonbulk NaI(Tl) scintillation events), can compromise the sensitivity goals of such an experiment. DAMA experiment, at the Laboratori Nazionali del Gran Sasso, in Italy, and using 100 kg NaI(Tl) target, reported first evidence of the presence of an annual modulation in the detection rate compatible with that expected for a dark matter signal just in the region below 6 keVee (electron equivalent energy) with a high statistical significance.1This signal was further confirmed by LIBRA experiment, using 250 kg of more radiopure NaI(Tl) detectors.2Using the same target than DAMA/LIBRA experiment, which accumulates by now fourteen annual cycles, makes possible for ANAIS to confirm DAMA/LIBRA results in a model-independent way. To achieve such a goal ANAIS detectors should be as good (or better) as (than) those of DAMA/LIBRA in terms of energy threshold and radioactive background below 10 keVee (electron equivalent energy). In this paper we will present some of the past and recent efforts to determine and reduce the background related to 40K contamination in several prototypes, as well as to determine the achievable threshold profiting from the low energy events population, conveniently tagged, that such a contamination provides. We will start by presenting the ANAIS project and the different crystals and experimental setups studied; then, we will move to understand the importance of the background due to 40K contamination and the technique used to determine its content in the different crystals, in particular, in ANAIS-25 modules, as well as the results derived. 2. ANAIS Project and NaI(Tl) Crystals Studied ANAIS project was conceived to use 10 hexagonal NaI(Tl) crystals of 10.7 kg each made by BICRON (now Saint Gobain) in the eighties. These detectors were part of those used in an experiment which looked for the 76Ge ββ decay to the first excited 1443010-2 Int. J. Mod. Phys. A 2014.29. Downloaded from www.worldscientific.com by UNIVERSITY OF ZARAGOZA on 11/06/15. For personal use only. July 30, 2014 9:7 WSPC/139-IJMPA S0217751X14430106 Analysis of the 40K contamination in NaI(Tl)crystals state at the Modane Underground Laboratory first,9andthenattheLSC. 10 They have been stored underground since the late eighties. After that, the NaI32 experiment searched for dark matter at the LSC with 3 of these detectors (amounting 32.1 kg) and accumulating two years of data taking. Bounds on WIMP masses and cross-sections were derived from the absence of positive hints, both in the usual analysis of the total rate, and in a pioneer modulation analysis.3Then, one of the NaI crystals was chosen to be further studied and modified: it was used to build ANAIS Prototype I,11 and after decoupling the PMT and removing the original stainless-steel encapsulation it was used in ANAIS Prototypes II12 and III (see Fig. 1(left)). In Prototype II a copper box was used to allow testing easily different light guides geometries and lengths, optical couplings and reflector/diffuser materials, whereas in Prototype III (PIII) a tight copper encapsulation was designed and PMTs and optionally light guides were coupled in a second step. The 40K content of these crystals (see Subsec. 5.1) was too high to allow their use in a dark matter search experiment, which implied a significant change in the ANAIS experiment time-line. The ANAIS collaboration started to look for new NaI(Tl) crystals with very low content in potassium (less than 20 ppb). Saint Gobain13 was contacted as the first option to produce the new crystals, as it had low radioactive background state- of-the-art NaI(Tl) detector technology: they had previously built DAMA/LIBRA detectors with similar background requirements. A 9.7 kg NaI(Tl) crystal had been bought previously to that company to be tested as first step in this direction. This crystal (see Fig. 1(center)) was used to build the ANAIS-0 module. The crystal was encapsulated at the University of Zaragoza in ETP (Electrolytic Tough Pitch) copper, closing tightly the detector, and using two synthetic quartz windows to get the light out to the photomultiplier tubes. A Teflon sheet as diffuser and a VikuitiTM reflector layer wrapped the crystal to increase light collection efficiency. ANAIS-0 module was designed to characterize and understand ANAIS background at low energy, optimize NaI scintillation events selection, fix the calibration method and test the electronics. ANAIS-0 module took data in different configurations: with or without 10 cm light guides and using different PMT models.14 A detailed background study of ANAIS-0 module in the different configurations tested has been recently published.15 Levels of 40K of that crystal were not enough for ANAIS requirements, as will be shown in Subsec. 5.1.1. Unfortunately, the contacts with Saint Gobain to develop more radiopure crystals were not successful. The ANAIS experiment had then to be redefined and, in order to reach the experimental goals, the final proposal consisted of 250 kg of ultrapure NaI(Tl) crystals to study the expected annual modulation in the galactic dark matter signal. An ultrapure powder provider and manufacturer of the ultra-low-background crystals had to be searched for. A NaI powder having a potassium level under the limit of the analytical techniques used for the radioactivity screening (less than 90 ppb) was found, and then, two crystals were grown by Alpha Spectra16 to determine, afterwards, more precisely the potassium content using the coincidence technique 1443010-3 Int. J. Mod. Phys. A 2014.29. Downloaded from www.worldscientific.com by UNIVERSITY OF ZARAGOZA on 11/06/15. For personal use only. July 30, 2014 9:7 WSPC/139-IJMPA S0217751X14430106 C. Cuesta et al. Fig. 1. Different NaI(Tl) crystals studied at the LSC in order to determine their potassium content by the coincidence technique: (left) 10.7 kg crystal from BICRON (PIII), (center) 9.7 kg crystal from Saint Gobain (ANAIS-0), and (right) 12.5 kg crystals from Alpha Spectra (ANAIS-25). presented in Sec. 3. These two NaI(Tl) crystals (12.5 kg each) form the ANAIS-25 setup, and are named detector 0 (D0) and 1 (D1), respectively (see Fig. 1(right)). The crystals were encapsulated in OFHC copper with two synthetic quartz windows allowing the PMTs coupling in a second step, to be done at LSC clean room. Only white Teflon was used as light diffuser, wrapping the crystal, inside the copper encapsulation. After the determination of the corresponding potassium level (presented in Subsec. 5.1.2), these detectors are still taking data at the LSC, remaining as main goal an overall background assessment.14 3. Determination of the 40KBulkContentinNaI(Tl)by Measuring in Coincidence 40K is probably the most relevant radioactive contaminant in the bulk for NaI(Tl) detectors. It is specially important its contribution to the background at very low energy, as it will be shown below. Because of that, a very good knowledge of such a contamination is required in order to properly estimate the sensitivity prospects achievable in the frame of the ANAIS project. The technique used to estimate the 40K activity in the bulk of the NaI(Tl) crystal is the measurement in coincidence: one detector measures the energy released by the X-ray/Auger electrons emissions of argon, amounting a total energy release of 3.2 keV, following the K-shell EC of 40K, while another detector measures the high energy gamma at 1460.8 keV, escaping from the former, and being fully absorbed in the latter. Besides a few accidental coincidences, the 3.2 keV peak can be clearly observed in all the studied crystals. From the measured coincidence rates and the corresponding efficiencies estimated with Geant4 package, the 40K activity in the crystal can be deduced. In Fig. 2a simplified 40K decay diagram is shown and more details on the 40K decay channels can be found in Table 1. Concerning the contribution to the background at low energy from the decay of this isotope, it is precisely the X-ray/Auger electrons emission of argon after K-shell EC in 40K, amounting a total energy release of 3.2 keV, which is more dangerous. 1443010-4 Int. J. Mod. Phys. A 2014.29. Downloaded from www.worldscientific.com by UNIVERSITY OF ZARAGOZA on 11/06/15. For personal use only. July 30, 2014 9:7 WSPC/139-IJMPA S0217751X14430106 Analysis of the 40K contamination in NaI(Tl)crystals Fig. 2. Simplified 40K decay diagram. Data have been taken from Ref. 18. More details about branching ratios and decay channels are given in Table 1. Table 1. Information about the different decay channels available for the 40K decay: EC to the first excited state (0,1), EC to the ground state (0,0), β+,andβ−.OnlyECto the first excited state is followed by a gamma emission. In the case of the EC channels the relative probabilities of K/L/M capture are also given: PK,PL,andPM. Energy (keV) Probability (%) PKPLPM 0,144.0±0.310.55 ±0.11 0.763 0.209 0.027 0,01504.69 ±0.19 0.2±0.1 0.880 0.086 0.013 β+ 0,0489.3±0.3(1.00 ±0.12) 10−3 β− 0,01311.07 ±0.11 89.25 ±0.17 A fraction of these events is moved to higher energies by the simultaneous interaction of the high energy gamma in the same detector, another is vetoed by an interaction of the high energy gamma in a different detector, however, a residual peak at 3.2 keV jeopardizes the search for dark matter if 40Kcontentisnotata few ppb level. At this point, it is worth noticing that the rate of the direct decay of 40K to the ground state of 40Ar through EC (at the 0.2% level) has not been yet experimentally measured, and hence, the uncertainty on the corresponding branching ratio is really high.17 This decay is important for the final ANAIS background because those events cannot be rejected by the coincidence method. At this moment, and up to our knowledge, there is not a feasiblewaytodeterminemoreprecisely such a branching ratio. 4. Brief Description of the Experimental Setups Different dedicated setups were operated at the LSC with the objective of determining the 40K content of all the NaI(Tl) available crystals. Pictures of those crystals have been shown in Fig. 1and their main features are summarized in Table 2.We describe below briefly the most relevant aspects of the different experimental setups, corresponding pictures are shown in Fig. 3, in which the coincidence measurements were carried out. All the setups have been operated at the LSC facilities (under 2450 m.w.e.), inside a shielding consisting of 10 cm archaeological lead plus 20 cm low activity lead, all enclosed in a PVC box tightly closed and continuously flushed with boil-off nitrogen. 1443010-5 Int. J. Mod. Phys. A 2014.29. Downloaded from www.worldscientific.com by UNIVERSITY OF ZARAGOZA on 11/06/15. For personal use only. July 30, 2014 9:7 WSPC/139-IJMPA S0217751X14430106 C. Cuesta et al. Fig. 3. Different setups operated at LSC to measure the 40K content of the available NaI(Tl) crystals and described in more detail in the text. (Left) BICRON setup, (center) ANAIS-0 and PIII setup, and (right) ANAIS-25 setup. Complementary information is also given in Table 3. Table 2. Main features of the different crystals studied in this work: setup where they have been operated, manufacturer, mass, shape and dimensions∗. Setup Detector Manufacturer Mass Shape Dimensions 1 EP054, EP055, BICRON 10.7 kg hexagonal prism 15.94 cm×20.32 cm EP056, EP057, EP058, EP059, EL214, EM301, EL604, EL603, EL607 2 PIII (EL607) BICRON 10.7 kg hexagonal prism 15.94 cm×20.32 cm 2 ANAIS-0 Saint Gobain 9.6 kg parallelepiped prism 10.16×10.16×25.40 cm3 3 ANAIS-25 D0 Alpha Spectra 12.5 kg cylinder 4.75 (Φ)×11.75 ANAIS-25 D1 ∗For the cylinders are given diameter and length, and for the hexagonal prism, distance between opposite vertices in the hexagonal face and length. (1) BICRON setup. In this setup the BICRON detectors were measured without modification with respect to their original assembly. Because of that, only one PMT per crystal was recording the scintillation signal. In order to mostly profit from the large number of crystals available, six and seven crystal units were measured simultaneously in two runs (labeled as (a) and (b) in Table 3). Although using only one PMT signal was quite limiting to lower the experimental threshold, the tagging with the high energy gamma from potassium allowed to distinguish clearly the 3.2 keV peak in all the crystals and to derive the results shown in Subsec. 5.1. A picture of the setup is shown in Fig. 3(left). (2) ANAIS-0 and Prototype III setup. The estimate of the 40K bulk content of the ANAIS-0 crystal, was done by measuring in coincidence with the previously studied Prototype III in a dedicated setup, shown in Fig. 3(center). Measurements were carried out in several phases, see Table 3, and although Phase III had another purpose,19 the corresponding data will be revised too in Subsec. 5.1.1. Data have been analyzed separately for each phase and 40K bulk content of the PIII has also been determined as cross-check. 1443010-6 Int. J. Mod. Phys. A 2014.29. Downloaded from www.worldscientific.com by UNIVERSITY OF ZARAGOZA on 11/06/15. For personal use only. July 30, 2014 9:7 WSPC/139-IJMPA S0217751X14430106 Analysis of the 40K contamination in NaI(Tl)crystals Table 3. Information about the different setups in which the studied crystals have been operated: PMT model coupled and if light guides (LG) have been or not used in each detector, and available live time (LT) of data corresponding to each setup and phase. Setup Phase PMTs LT (days) 1 a ETL 9765 65.59 b ETL 9765 73.49 2 I ETL 9302B (ANAIS-0) 51.20 ETL 9302B and LG (PIII) II Ham. R6233-100MOD (ANAIS-0) 57.82 ETL 9302B and LG (PIII) III Ham. R6956MOD (ANAIS-0) 39.30 Ham. R11065SEL (PIII) 3 Ham. R6956MOD (ANAIS-25 D0) 70.40 Ham. R11065SEL (ANAIS-25 D1) (3) ANAIS-25 setup. The main goal of the ANAIS-25 setup was precisely the determination of the potassium content of the new crystals. It consists of two cylindrical 12.5 kg NaI(Tl) detectors grown with ultrapure NaI powder (<90 ppb potassium at 95% C.L. according to results from HPGe spectrometry screening carried out at LSC) and built in collaboration with Alpha Spectra.16 A picture of the setup is shown in Fig. 3(right), and more details about it can be found in Table 3. In all cases, each PMT charge output signal is separately processed; each one is divided into a trigger signal, a signal going to the digitizer, and a number of signals differently attenuated and fed into QDC (charge-to-digital converter) module channels to be integrated in a 1 μs window. At least, a low energy and a high energy ranges were considered in all the analyzed setups. The building of the spectra was done by software (off-line) by adding the signals from both PMTs whenever possible. Data triggering is done in logical OR mode between different detectors operated at the same time and in logical AND between the two PMT signals from each detector, if available (only the BICRON setup could not use it). 5. Results and Conclusions 5.1. 40Kbulk content In this section, we will present the potassium bulk content results for all the studied crystals. The first step in order to derive the 40K activity in one crystal, is to select the corresponding energy windows containing the 1460.8 keV energy depositions in the other(s) crystal(s) sharing the experimental space. The 40K activity of every crystal can be estimated with the area of the 3.2 keV peak (Area) identified in the coincident spectra with the high energy window chosen in the other(s) detector(s), the total available live time (t), the crystal mass (m), the efficiency of the 1443010-7 Int. J. Mod. Phys. A 2014.29. Downloaded from www.worldscientific.com by UNIVERSITY OF ZARAGOZA on 11/06/15. For personal use only. July 30, 2014 9:7 WSPC/139-IJMPA S0217751X14430106 C. Cuesta et al. Table 4. Results for 40K bulk activity of the BICRON −10.7kgNaI(Tl)crystals. Detector 40K activity (mBq/kg) EP054 13.7±0.3 EP055 15.2±0.1 EP056 18.8±0.2 EP057 20.9±0.4 EP058 16.2±0.3 EP059 16.6±0.2 EL214 17.9±0.4 EM301 21.2±0.4 EL604 16.5±0.3 EL603 14.5±0.2 EL607 15.7±0.5 coincidence, determined by simulation using Geant4 package () and the fraction of events effectively selected by the coincidence window chosen (F): Activity (Bq/kg) = Area (counts) t(s) ·m(kg) ··F.(1) Only statistical errors coming from the 3.2 keV peak area determination are taken into account in the derivation of the activity errors shown in the following. First of all, eleven NaI(Tl) crystals from BICRON, 10.7 kg mass each, were measured: six and seven detectors were placed in a similar configuration in runs aand b, respectively, using two of them in both as cross-check. The results for the 40K activity corresponding to all of the old BICRON crystals are presented in Table 4. It can be observed that it is very similar for all of them, ranging from 13 to 21 mBq/kg, which corresponds to 0.42 to 0.68 ppm natural potassium in the bulk of the crystals. The procedure followed in order to derive such activity values is equivalent to those explained in detail in the following for ANAIS-0 and PIII crystals, as well as for both ANAIS-25 modules, and is described in detail in Subsec. 5.1.1. 5.1.1. ANAIS-0and Prototype III The first step in order to derive the 40K activity in both crystals, is to select the corresponding energy windows containing the 1460.8 keV energy depositions. Different window widths (1σ,a2σand 3σ) have been considered for the selection of coincident events in both detectors. High energy spectra for the two phases of measurement with ANAIS-0 and PIII detectors are shown in Fig. 4(left), and a zoom around the 1460.8 keV line, remarking the three windows considered (right). aσwill refer for each detector to the corresponding value of the standard deviation obtained in the Gaussian fit of the 1460.8 keV line. 1443010-8 Int. J. Mod. Phys. A 2014.29. Downloaded from www.worldscientific.com by UNIVERSITY OF ZARAGOZA on 11/06/15. For personal use only. July 30, 2014 9:7 WSPC/139-IJMPA S0217751X14430106 Analysis of the 40K contamination in NaI(Tl)crystals Fig. 4. (Color online) High energy spectra of PIII and ANAIS-0 detectors. (Left): The whole spectra and (Right): a zoom showing the 1460.8 keV gamma line following 40K EC decay for the two considered phases. The 1σ(red), 2σ(blue) and 3σ(green) coincidence windows are also shown. The low energy spectra in coincidence with 1σwindow around the 1460.8 keV line in the other detector are shown in Fig. 5. The gain of PIII along phase II was probably not stable enough and the 1460.8 keV line is clearly distorted; this could lead to a decrease in the efficiency of the coincidence in an undetermined way, compromising the validity of the derived result. Fig. 5. Low energy spectra (in counts/channel) in coincidence with 1σwindows around the 1460.8 keV line in the other crystal, shown in Fig. 4, for ANAIS-0 and PIII detectors in the two measurement phases considered in this study. The 3.2 keV peak is clearly visible above the baseline noise peak. 1443010-9 Int. J. Mod. Phys. A 2014.29. Downloaded from www.worldscientific.com by UNIVERSITY OF ZARAGOZA on 11/06/15. For personal use only. July 30, 2014 9:7 WSPC/139-IJMPA S0217751X14430106 C. Cuesta et al. Fig. 12. Low energy spectra (in counts/channel) in coincidence with 1σwindows around the 1460.8 keV line (shown in Fig. 11) in the other crystal for ANAIS-25 D0 (left) and D1 (right). Fig. 13. (Color online) Low energy spectra (in counts/channel) in coincidence with high energy windows in the other detector (all of them having the same width, ±σ) are shown for ANAIS-25 D0 (left) and D1 (right): centered at 1460.8 keV peak (μ) in black, centered 2σabove (μ+2σ)in blue, centered 2σbelow (μ−2σ) in red, and centered further above (μ+11σ) in green. channel 70 for the D1. The effect of changing the high energy window above and below the 1460.8 keV position has also been studied, see results in Fig. 13.Numbers of counts over these thresholds for every coincidence window are shown in Table 8. The probability that, after a 40K disintegration in one crystal, the 1460.8 keV photon escapes and releases the full energy in the other detector has been estimated with Geant4, in this case using version geant4.9.4.p01. The corresponding efficiency for the coincidences between both ANAIS-25 modules is 1.08×10−3, just a bit lower Table 8. Events above the analysis threshold (channel 60 for D0 and channel 70 for D1), among those selected by the coincidence with an event in a window centered at, below, or above the 1460.8 keV gamma position (1σwidth). Detector μμ−2σμ+2σμ+11σ 08225 16 7 17822 21 10 1443010-16 Int. J. Mod. Phys. A 2014.29. Downloaded from www.worldscientific.com by UNIVERSITY OF ZARAGOZA on 11/06/15. For personal use only. July 30, 2014 9:7 WSPC/139-IJMPA S0217751X14430106 Analysis of the 40K contamination in NaI(Tl)crystals Fig. 14. Low energy spectra (in counts/channel) in coincidence with 1σ,2σ,and3σwindows around 1460.8 keV line in the other crystal for ANAIS-25 D0 and D1. Table 9. 40K activity calculated for the two ANAIS-25 crystals using different widths coincidence windows. Detector 40K activity (mBq/kg) 1σ2σ3σ 01.34 ±0.13 1.16 ±0.11 1.31 ±0.11 11.15 ±0.18 1.08 ±0.16 1.21 ±0.20 than that obtained for ANAIS-0 and PIII because of the higher mass, that decreases the probability for the escape of the gamma without losing any energy. Then, the activities of 40K for each ANAIS-25 crystal have been estimated for the different width coincidence windows, see Eq. (1). The Gaussian fits performed to redeem the area are shown in Fig. 14. Results on the 40K content for each detector are shown in Table 9. Good agreement between results derived for both detectors is observed, as expected. Averaging the 1σwindow results for the two crystals, we can conclude that ANAIS-25 crystals have a 40Kcontent1.25 ±0.11 mBq/kg (41.7±3.7 ppb of potassium) much lower than that estimated for ANAIS-0 crystal, see Fig. 15. NaI(Tl) crystals from different manufacturers have been characterized in terms of their potassium bulk content by a measurement in coincidence, and improvement 1443010-17 Int. J. Mod. Phys. A 2014.29. Downloaded from www.worldscientific.com by UNIVERSITY OF ZARAGOZA on 11/06/15. For personal use only. July 30, 2014 9:7 WSPC/139-IJMPA S0217751X14430106 C. Cuesta et al. Fig. 15. (Color online) Low energy spectra in coincidence with 1σwindows around 1460.8 keV line in the other crystal for ANAIS-0 (black), ANAIS-25 detector 0 (blue), and ANAIS-25 detector 1 (red). of one order of magnitude in the potassium content can be reported for the ANAIS- 25 detectors, built in collaboration with Alpha Spectra. However, the 20 ppb goal has not yet been achieved and before ordering the additional 18 modules required to complete the ANAIS total detection mass, careful analysis of the situation in collaboration with Alpha Spectra is undergoing, trying to further purify the starting NaI powder. 5.2. Trigger efficiency at 3.2keV However, having a tagged population of bulk scintillation events at 3.2 keV is very useful for many other purposes related to the DM search. Just to show an example, in ANAIS-25 setup we used this 40K events to estimate the trigger efficiency of the experiment. It is shown in Fig. 16 how many of the low energy events identified by the coincidence with the high energy window around 1460.8 keV, and hence, Fig. 16. 40K events at low energy, identified by the coincidence with a high energy gamma for the ANAIS-25 D0 (left), and D1 (right). Events with T=2(T= 1) have not triggered in D0 (D1), whereas events with T= 3 have triggered in both detectors. 1443010-18 Int. J. Mod. Phys. A 2014.29. Downloaded from www.worldscientific.com by UNIVERSITY OF ZARAGOZA on 11/06/15. For personal use only. July 30, 2014 9:7 WSPC/139-IJMPA S0217751X14430106 Analysis of the 40K contamination in NaI(Tl)crystals corresponding to the decay of 40K, have effectively triggered our acquisition. A good trigger efficiency can be reported: 99% of the events above 1.5 keV are triggering in D1, and 97% in D0. It is worth noting that fortuitous coincidences in D0 arrive up to higher energies (which is related to the higher dark current of the PMTs used), and that in D1, baseline is eventually triggering. Acknowledgments This work has been supported by the Spanish Ministerio de Econom´ıa y Competitividad and the European Regional Development Fund (MINECO-FEDER) (FPA2011-23749), the Consolider-Ingenio 2010 Programme under grants MULTIDARK CSD2009-00064 and CPAN CSD2007-00042, and the Gobierno de Arag´on (Group in Nuclear and Astroparticle Physics, ARAID Foundation and C. Cuesta predoctoral grant). C. Ginestra and P. Villar have been supported by the MINECO Subprograma de Formaci´on de Personal Investigador. We also acknowledge LSC and GIFNA staff for their support. References 1.R.Bernabeiet al.,Riv. Nuovo Cimento A112,545 (1999); R.Bernabeiet al.,Phys. Lett. B 450,448 (1999). 2. R.Bernabeiet al.,Nucl. Inst. Methods A592,297 (2008); R.Bernabeiet al.,Eur. Phys. J. C 56, 333 (2008); R.Bernabeiet al.,Eur. Phys. J. C 67,39(2010); R.Bernabeiet al.,Nucl. Instrum. Methods A692,120 (2012); R.Bernabeiet al., Eur. Phys. J. C 73,2648 (2013). 3. M. L. Sarsaet al.,Nucl. Phys. B (Proc. Suppl.)35,154 (1994); M. L. Sarsaet al., Phys. Lett. B 386,458 (1996);M.L.Sarsaet al.,Phys. Rev. D 56,185 (1997). 4.K.Fushimiet al.,Phys. Rev. C 47,425(1993); K. Fushimiet al.,Astropart. Phys. 12,185 (1999). 5.G.Gerbieret al.,Astropart. Phys. 11,287(1999). 6. G. J. Alneret al.,Phys. Lett. B 616,17 (2005). 7. K. Fushimiet al.,J. Phys.:Conf. Ser. 203,012046(2010). 8.J.Cherwinkaet al.,Astropart. Phys. 35,749 (2012). 9.A.Morales et al.,Il Nuovo Cimento A100,525(1988). 10. A. Morales et al.,J. Phys. G:Nucl. Phys. 17,211 (1991); A. Morales et al.,Il Nuovo Cimento A104,1581 (1991). 11.S.Cebri´an,Estudiodelfondo radiactivo enexperimentos subterr´aneos de b´usqueda de sucesos poco probables: CUORE(CryogenicUndergroundObservatoryforRare Events) y ANAIS (Annual modulationwith NaI(Tl)), Ph.D. Dissertation,Universidad de Zaragoza, 2002. 12. M. Mart´ınez, Dise˜nodeunprototipo para unexperimento de detecci´ondirectade materia oscura mediante modulaci´onanualconcentelleadores de iodurodesodio, Ph.D. Dissertation,UniversidaddeZaragoza, 2006. 13. SaintGobain, http://www.saint-gobain.com/en. 14.C.Cuesta, ANAIS-0: Feasibility study fora250kgNaI(Tl) darkmattersearch experimentattheCanfrancUndergroundLaboratory. Ph.D. Dissertation,Universidadde Zaragoza,2013. 15.S.Cebri´an et al.,Astropart. Phys. 37,60(2012). 1443010-19 Int. J. Mod. Phys. A 2014.29. Downloaded from www.worldscientific.com by UNIVERSITY OF ZARAGOZA on 11/06/15. For personal use only. July 30, 2014 9:7 WSPC/139-IJMPA S0217751X14430106 C. Cuesta et al. 16. AlphaSpectra Inc., GrandJunction,Colorado, US, http://www.alphaspectra.com/. 17. J. Pradler,B.Singh andI.Yavin,Phys. Lett. B 720,399 (2013). 18. http://www.nucleide.org. 19.C.Cuestaet al.,Opt. Mat. 36,316(2013). 20. S. Agostinelliet al.,Nucl. Instrum. Methods A506,250 (2003). 1443010-20 Int. J. Mod. Phys. A 2014.29. Downloaded from www.worldscientific.com by UNIVERSITY OF ZARAGOZA on 11/06/15. For personal use only.