Improvement of the accuracy of a Fizeau Interf by using LS
Full text
MSc in Photonics PHOTONICSBCN Universitat Politècnica de Catalunya (UPC) Universitat Autònoma de Barcelona (UAB) Universitat de Barcelona (UB) Institut de Ciències Fotòniques (ICFO) http://www.photonicsbcn.eu Master in Photonics MASTER THESIS WORK IMPROVEMENT OF THE ACCURACY OF A FIZEAU INTERFEROMETER BY USING LATERAL SHEARING Josep Vidal González Supervised by Dr. Juan Campos, (UAB) Presented on date 8 th September 2008 Registered at
Improvement of the accuracy of a Fizeau Interferometer by using Lateral Shearing Josep Vidal ALBA Synchrotron Light Source Facility, Bellaterra 08193, Spain E-mail: [email protected] Abstract. Quality of optical surfaces is currently limited by the accuracy of the metrology used for their characterization. Synchrotron radiation facilities require even more precise surface measurement techniques. The reconstruction of measurement functions by using Lateral Shearing technique is proposed in this paper. Thanks to this method, measurement accuracy given by a Fizeau interferometer is not limited by the reference surface, but by imprecisions in linear stage. Two of the errors caused by such imprecisions, calibration and tilt, are tackled in this study. A tilt estimation method is proposed at the same time that numerical results are detailed below. They demonstrate how these methods do improve the accuracy of optical surface measurements. PACS numbers: 06.00.00, 42.87.-d, 42.87.Bg, 07.85.Fv 1. Introduction Third generation synchrotron sources that have appeared and developed during the last twenty years are X-ray sources that produce high brilliance beams. Brilliance is defined as the spectral flux density per unit phase space. The typical source size is in the order of few microns, and the collimation of the beam is in the order of 10µrad. In the near future the free electron lasers will produce even more brilliant beams. In order to profit this brilliance characteristic, the beamline optics must be polished with accuracies better than 1µrad for mirrors up to one meter long. The quality of the optical surfaces is currently limited by the accuracy of the metrology used for their characterization during the manufacture. To reach the mirror quality required in future applications, the development of new technologies to improve the accuracy in optical metrology is needed [1]. Usually there are two kind of instruments used to characterize synchrotron optics: long trace profilometers (LTPs) [2, 3] and Fizeau interferometers [4]. While the LTP provides high accuracy, it is very sensitive to the variations of the environment, due to the low scan speed. On the other hand, the Fizeau interferometer is much more robust to the environment perturbations, although its absolute accuracy is limited by the quality of the reference optics, often worse than the mirrors being characterized. Such a limit in the accuracy is generally overcome either by a proper calibration of the optics, or by combining different datasets, either by rotating the surface [5] or by using different references [6]. In this paper we propose to remove the error introduced by the reference surface by applying the Lateral Shearing (LS) Interferometry technique [7, 8]. It consists in
Improvement of the accuracy of a Fizeau Interferometer by using Lateral Shearing 2 subtracting the datasets corresponding to two different positions of the sample surface with respect to the reference surface, and reconstructing the mirror function from its finite differences. In this case, the accuracy of the measurement is not limited by the reference surface, but by errors of the linear stage. 1.1. Fizeau Interferometer as a tool for ultra precise metrology Most of two-beam interferometers, which two wavefronts follow different paths, are very sensitive to external conditions, such as temperature fluctuations. The Fizeau configuration is one of the best, because the test and reference beams are split by the final surface of the interferometer optics. Hence all internal optical components are traversed by the two wavefronts and they are in the common path of the two beams. In this way, the potential sources of error are reduced, and as a result the Fizeau Interferometer achieves high resolution and high repeatability. It is assumed that the accuracy is basically limited by the accuracy of the reference surface. The calibration of the phase-shift mechanics, the non-linearities of the CCD camera, the diffraction effects (dust, speckle) also contribute to these errors, but the reference surface is the main source of error for a well calibrated instrument in a controlled environment. 1.2. Lateral Shearing Interferometry LS interferometers are commonly used in a broad number of applications: optical testing, active or adaptive control, atmospheric turbulence measurements. Their main advantage lies in the fact that the analyzed wavefront interferes with itself, thus eliminating the need for a separate ideal reference wavefront. That is, two replicas of the analyzed wavefront shifted one each other interfere, giving a fringe pattern related to the measurement phase. However, lateral shearing interferometers are not available for large apertures. This overlapping principle is used in our setting, but in this case the shifted wavefronts do not interfere directly one each other, but are measured sequentially with respect to a common reference. The description of the new technique for continuous infinite domain functions is given in section 2. Problems introduced when using sampled signals with finite domain are explained in section 3. Also the solution to these problems by means of the natural extension is given in this section. Then, the influence of different errors introduced by the setup is studied in section 4, and a method to minimize the influence of the errors is proposed. Finally, in section 5 a summary is presented. For the sake of simplicity only one dimensional analysis is given in this paper. 2. Setup and procedure The proposed setup consists on installing the sample mirror, the surface under test, on top of a linear stage, in front of the Fizeau interferometer (see fig. 1). The sample mirror is measured at two or more different positions of the linear stage. At each position dof a the linear stage, the Fizeau provides an accurate measure of the optical path difference t(x) between the reference surface r(x) and the sample s(x). They are related as follows t(x) = s(x)−r(x) t1(x) = s(x−d)−r(x)(1)
Improvement of the accuracy of a Fizeau Interferometer by using Lateral Shearing 3 By subtracting datasets acquired at different positions of linear stage, one gets w(x) = t(x)−t1(x) = s(x)−s(x−d) (2) which does not depend on the reference surface. Reference Surface under test Surface under test Linear stage Figure 1. Scheme of the configuration. The sample is installed in a linear stage to permit its displacement To recover the sampled function s(x) from the difference w(x), the Fourier transform (FT) is applied to w(x). As the FT is a linear operator, it can be obtained W(u) = FT [w(x)] = a z}| { FT [s(x)] − b z}| { FT [s(x−d)] (3) According to Shift Theorem [9], (a) and (b) are only differing by a linear phase which depends on the shift d. Then, S(u) =W(u)T(u), T(u) = 1 1−exp (−2πiud) (4) where S(u) = FT [s(x)] and T(u) represents the so called shearing transfer function [10, 11], which is undefined when uis multiple of 1/d. Then, a new shearing transfer function is defined as ˆ T(u) = T(u),u6=k d,k∈Z 0,otherwise (5) Applying the Inverse Fourier Transform, s(x) can be obtained, s(x) = FT−1hW(u)ˆ T(u)i.(6) This procedure, combining spatial and Fourier representations, allows w(x) to obtain the original function s(x), free from the errors induced by the reference. Note that the reconstruction has periodical errors of period dproduced by the new shearing transfer function. However, these errors can be avoided by combining three datasets with d= 0, d1and d2where d1and d2are coprimes. The actual application of the technique requires that the method must be discretized, i.e. finite domain functions have to be used, and that the influence on the accuracy of the translation stage errors should be studied. This is done in the following sections.
Improvement of the accuracy of a Fizeau Interferometer by using Lateral Shearing 4 3. Discretization with Natural Extension The problem introduced when using finite domain functions is sketched in fig. 2(a). Each measure has a length η, then, in the first measure, s(xi), the captured values are in the domain i∈[d, η +d] (represented in blue in figure 2(b)), while in the second measurement, s(xi−d), the captured values are in the domain i−d∈[0, η] (represented in red in figure 2(b)). From now, dand ηrepresents the number of pixels of the shift and of the entire frame. Note that the sampled finite domain functions are not longer the sames functions, one shifted with respect to the other, and in consequence the method described in section 2 will give errors [12, 13]. These errors are related with ǫd(s)∝ d X i=1 s(xi)−s(xi+η) (7) In order to solve this problem one constructs a periodic function which is equal to the target function in the region of interest [0, η] (see figure 2(b)). By means of the Natural Extension (NE) [14, 15] it is possible to obtain the extended difference function wd(xi) by using only measured values of w(xi). 10 30 (a) [mm] [nm] d 0η η η η+d 0 5 30 [mm] [nm] (b) ηη+d d η Figure 2. With two frames, a full period of a η+dsawtooth function is captured. These acquisitions could be obtained with two analog processes: (a) displacing the domain or (b) displacing the function. In figure (b) extended values of wd(xi) are obtained by subtraction of the measurement function and its copy. If at each acquisition, shifted dpixels one each other, the function is sampled at ηpoints, η+ddifferent values are included. Then, the values of a full period from the sample s(xi) are included in the difference w(xi). It is of note that there is no periodical relation between both η-frames, as it happens with any tested function s(xi) in a real world. Consequently, the corresponding w(xi) in both situations have no periodical relation between their points. The two η-pixels sampled functions s(xi) and s(xi−d) are not the same functions because they contain different values, and then they do not fulfill the Shift Theorem. The next step consists of conveniently extending w(xi) from no inherent periodic functions s(xi), into wd(xi), as it were η+dperiodic. The NE allows to obtain η+d pixels from the η-values of the differences between the two shifted functions assuming that s(xi) in η+dperiodic. Then, values from next period are included, and thus, the FT of wd(xi) fulfills the Shift Theorem. The main problem for this purpose is that s(xi) is unknown. However, the NE extends w(x) by dpoints with a linear combination of certain ηvalues of itself, given
Improvement of the accuracy of a Fizeau Interferometer by using Lateral Shearing 5 by the following expression wd(xi) = w(xi) if i∈[0, η] −Pη d l=1 w(xi−ld) if i∈[η+ 1, η +d] (8) To check that wd(xi) corresponds to the differences of a η+dperiodic function s(xi) in the extended domain, we compare the values of η+dperiodic function such as in fig. 2(b) with the function obtained by the NE. For this purpose, we develop eq.8(b) − η d X l=1 w(xi−ld) = − η d X l=1 s(xi−ld)−sxi−(l+1)d =s(xi−η−d)−s(xi−d). (9) which in the extended domain, [η+ 1, η +d], is w(xη+1) = s(x1−d)−s(xη−d+1) . . . w(xη+d) = s(x0)−s(xη) (10) Comparing equation 10 with w(xi) values of a η+dperiodic function (see fig. 2), it is clear that the NE effectively periodizes the tested function in the extended domain, using only a linear combination of w. The reconstruction of a function by the LS method, using NE or not, is shown in fig. 3. Note that the errors achieved with the NE are smaller than those obtained without using it. It is also remarkable that when NE is not used, the reconstruction errors at both ends of the function, known as boundary errors, are larger than central reconstruction errors, due to the fact that Shift Theorem is not fulfilled there. Furthermore, the bigger is the displacement d, the bigger the error of the reconstruction is. As one can see in figure 3(d) the error when using NE is a periodic function of period d, as commented above this error can be reduced by using two differents wd1(xi) and wd2(xi) 50 150 0 15 [mm] [nm] (a) 50 150 −10 10 [mm] (b) 50 150 0 15 [mm] (c) 50 150 −15 −10 −5 0 5 10 15 [mm] (d) Figure 3. In fig (a) the original function in dashed red and reconstructed in blue, with a displacement dof 10mm. In fig. (b) the error of this reconstruction. In fig. (c) the original function in dashed red and reconstructed in blue, using NE with same displacement. In fig. (d) the error of this reconstruction. 4. Linear stage errors As described in section 2 the sampled mirror is shifted with a linear stage that has both guidance and positioning errors. These movement imprecisions become the limit of the reconstruction accuracy for the proposed technique.
Improvement of the accuracy of a Fizeau Interferometer by using Lateral Shearing 6 4.1. Position errors One of the most important parameters for the reconstruction of a function using LS is the precision in the displacement between the two frames. As computational work after function reconstruction by the Fizeau interferometer is required, it is crucial to have good equivalence between the Fizeau and the computational coordinate systems. Once the relation between computer and Fizeau pixel size is established, it is important to know which is the difference between the real displacement of the linear stage and the displacement given by the computer. To perform such calibration, a software based on multiple line 1D correlation is developed. The encoded displacement of the linear stage is compared to the position of the mirror in the screen of the Fizeau, which is given by the maximum of the cross-correlation to the image acquired at reference position. fizeau position (pixels) 5 0 5000 motor position (steps) fizeau 0 5000 -0.1 10 30 60 Location Error (pixels) 0.1 Location Error (pixels) 30 100 200 600 800 location error ( px /100) 0 Q 30 location error ( px /100) 0 Figure 4. Chart (a) represents the fizeau position as a function of the motor position (in pixels.) Chart (b) represents the number of events as a function of the location error (in pixels). In chart (c) the quality factor Qas a function of the location error is represented. The location method has been tested in the laboratory and the positions in pixels given by the algorithm have been compared to those given by the encoder of the linear stage. The dependence is plotted in figure 4 (a). The uncertainty of the location provided by the correlation algorithm is limited by its repeatability, due to the noise of the acquisitions. The histogram of such repeatability is given in figure 4 (b). It shows that the root mean square uncertainty is below 1.6% of a pixel. In figure 4 (c) the quality factor Q, defined as Q=λHeNe/∆swhere ∆sis the peak to valley of the error of the reconstruction, decreases dramatically with the location error. However, the callibration algorithm allows to reconstruct profiles with peak to valley errors about λ/700, i.e. Q≈700. 4.2. Tilt error Another error introduced by the linear stage is the pitch error. It is when the guide in which the stage is moved is not perfectly straight. This error is described by a linear function added to the shifted profile as follows ˆ t1(xi) = s(xi−d)−r(xi) + axi,(11) and therefore to the function of differences ˆw(xi) = s(xi)−s(xi−d)−axi(12) This is a critical error, not only because a change in the optical path difference induces error in the LS reconstruction, but also because the NE is extremely sensitive to it, giving poor measurement reconstructions. As stated before, this sensitivity
Improvement of the accuracy of a Fizeau Interferometer by using Lateral Shearing 7 comes from the fact that when periodizing s(xi) in the extended domain, this method only uses w(xi) values. Consequently, if these w(xi) values are modified by a given tilt, the NE can not obtain a correct wd(xi). 50 150 −20 20 [nm] (a) 50 150 −20 20 (b) 50 150 −20 20 (c) 50 150 −15 10 [mm] [nm] 50 150 −15 10 [mm] 50 150 −15 10 [mm] Figure 5. Reconstruction error for different values of the tilt introduced by the platform. On the top row the original function is represented in dashed red, while the reconstruction is represented in blue. Bottom row represents the reconstruction errors. The used tilts are (a) a= 0.005µrad, (b) a= 0.015µrad, and (c) a= 0.03µrad. Reconstruction errors caused by different tilts are shown in figure 5. Notice that in fig. 5(a), the maximum error between t(xi) and t1(xi) induced by the tilt is 0.5 nanometers. With such a sub-nanometer error, an imprecise reconstruction would already be obtained, and the larger tilt error the worse final reconstruction. It can also be seen in fig. 5 that, as tilt increases, the reconstruction error becomes quadratic. In other words, the tilt error provokes that the curvature of the reconstructed function is very different to that of the real function. Aside from this, the function is well reconstructed (i.e. the higher orders of the taylor series are not affected by the tilt error). Therefore, a method for removing this tilt reconstruction error before working in the Fourier domain is required. 4.3. Tilt error estimation We propose a method for estimating the tilt error. As t(xi) and t1(xi) implicitly have similar information, hence both have repeated values of the measurement function, the tilt can be estimated through w(xi). If a perfect displacement without tilt is assumed, values of w(xi) are I w(x1) = s(x1)−s(x1−d) . . . w(xd) = s(xd)−s(x0)
Improvement of the accuracy of a Fizeau Interferometer by using Lateral Shearing 8 II w(xd+1) = s(xd+1)−s(x1) . . . w(xη−d) = s(xη−d)−s(xη−2d) (13) III w(xη−d+1) = s(xη−d+1)−s(xη−2d+1) . . . w(xη) = s(xη)−s(xη−d) When all values of w(xi) are added up, central points (II) are canceled each other, and the only values left are the negative terms from initial points (I) and the positive terms from final ones (III), as follows A= η X i=1 w(xi) = − d−1 X i=0 s(x−i) + η X i=η−d+1 s(xi) = d−1 X i=0 s(xη−i)−s(x−i) (14) If instead of having a perfect displacement (eq.13), a tilt is present (eq.11), the sum (eq.14) becomes η X i=1 ˆw(xi) = A−a η X i=1 xi(15) Then, tilt acan be written as a=−Pη i=1 ˆw(xi) + A Pη i=1 xi (16) where ˆw(xi) and all xivalues are known. Aonly depends on s(xi), which remains unknown, and thus, it becomes the unique error source for tilt estimation. In order to determine the slope aof the error, Aterm could be in different ways, here we present two of them: (i) A= 0 With this statement, it is assumed that dinitial values of s(xi) are equal to d final values of s(xi−d). If Ais not equal to zero, the tilt estimation error is δa=A Pη i=1 xi (17) (ii) A=− d X i=1 t1(xi) + η X i=η−d+1 t(xi) With this statement, it is assumed that dinitial values of s(xi) are equal to its respective t(xi) values and that dfinal values of s(xi−d) are equal to its respective t1(x) values s(xi)≈ t1(xi+d) if i∈[−d+ 1,0] t(xi) if i∈[η−d+ 1, η] (18)