Portable low-power system for continuous monitorization of the cardiovascular system

Mansergas Fernández, Ivan

Abstract

In the last year, the appearance of a global pandemic due to the COVID-19 virus evidenced one problem: the medical system is not prepared to treat a lot of people at the same time, as all the clinical environments collapsed during practically the whole year. This is where performing ambulatory measurements plays an important role, because it would allow to carry out medical diagnosis without the need of assisting at the hospital; everyone could have a device at home which would do a first health analysis to see if it would be necessary to visit the doctor. Nevertheless, recent medical technologies are very expensive and many people could not afford them. Thus, the aim of this project is to design and implement a portable low-cost and low-power consumption system capable of acquiring biomedical signals from which different physiological parameters will be extracted to give information about the health status of the patient. Although the project will be focused on the electronics with the hardware and software, a series of measurements will be carried out as well to show the feasibility of the data acquired. Different measurements of the cardiovascular system have been carried out, obtaining the electrical signal: 12 ECG leads from the standard positions (10 electrodes), and a subset of leads with less electrodes and a not-standard position, to ease the measurements outside the hospital. For its importance, something that will be achieved as well is the possibility of synthesizing the 12 ECG leads from a subset of less electrodes incorporating also a not-standard placement of the electrode, so that it will not be necessary to place all the 10 electrodes on the body, making the measurement more comfortable. In addition, pulse wave signals have been measured in different positions. The signal acquisition is complemented with the development of a software capable of obtaining automatically the important parameters. Finally, relating the pulse wave signals with the electrical signals of the ECG, we will be able to monitor the circulation of blood and then have information about the health status of the cardiovascular system.

Full text

Escola Tècnica Superior d’Enginyeria de Telecomunicació de Barcelona Grau en Enginyeria Física Treball Final del Grau en Enginyeria Física Portable low-power system for continuous monitorization of the cardiovascular system Iván Mansergas Fernández Thesis Director Jaime Oscar Casas Piedrafita Instrumentation, Sensors and Interfaces Group Universitat Politècnica de Catalunya June 28, 2021 Iván Mansergas Fernández Portable low-power system for continuous monitorization of the cardiovascular system June 28, 2021 Thesis Director: Jaime Oscar Casas Piedrafita Universitat Politècnica de Catalunya Grau en Enginyeria Física Escola Tècnica Superior d’Enginyeria de Telecomunicació de Barcelona C/Jordi Girona, 1 08034 and Barcelona ii Acknowledgement It all started with an email, where the subject was the possibility of doing a project related to electronics, in which I will focus my career. Prof. Oscar Casas kindly presented to me the line of work of his research group, and proposed to me this interesting project. From the beginning to the end I have never felt alone in this adventure, as he provided me with all the knowledge, tools and learning that I needed at every stage of the work done, being always available and interested in my advance day over day. This is why I would like to express my most sincere gratitude to all he has done for me. I would like to thank as well the rest of the ISI group, for their availability and disposal for everything that I could have requested. In particular, to Francis López, from whom I learned very interesting things about the most technical electronic aspects of the project. Thanks to him, the devices designed have seen the day of light and with a professional quality. In addition, I would like to mention my laboratory colleagues, Marc and Marta, which apart from helping me in the biological aspects of measurements and acquisition software, they made my days more bearable, with conversations about life that made me grow up in mind, specially for the laboral life. Last but not least, I am very thankful for my family and friends, which not only during the project, but throughout the degree, have been my stress canalization, as they have taken me out of the hard times to have great times that made me recover the happiness that sometimes the studies stole. iii Abstract In the last year, the appearance of a global pandemic due to the COVID-19 virus evidenced one problem: the medical system is not prepared to treat a lot of people at the same time, as all the clinical environments collapsed during practically the whole year. This is where performing ambulatory measurements plays an important role, because it would allow to carry out medical diagnosis without the need of assisting at the hospital; everyone could have a device at home which would do a first health analysis to see if it would be necessary to visit the doctor. Nevertheless, recent medical technologies are very expensive and many people could not afford them. Thus, the aim of this project is to design and implement a portable low-cost and low-power consumption system capable of acquiring biomedical signals from which different physiological parameters will be extracted to give information about the health status of the patient. Although the project will be focused on the electronics with the hardware and software, a series of measurements will be carried out as well to show the feasibility of the data acquired. Different measurements of the cardiovascular system have been carried out, obtaining the electrical signal: 12 ECG leads from the standard positions (10 electrodes), and a subset of leads with less electrodes and a not-standard position, to ease the measurements outside the hospital. For its importance, something that will be achieved as well is the possibility of synthesizing the 12 ECG leads from a subset of less electrodes incorporating also a not-standard placement of the electrode, so that it will not be necessary to place all the 10 electrodes on the body, making the measurement more comfortable. In addition, pulse wave signals have been measured in different positions. The signal acquisition is complemented with the development of a software capable of obtaining automatically the important parameters. Finally, relating the pulse wave signals with the electrical signals of the ECG, we will be able to monitor the circulation of blood and then have information about the health status of the cardiovascular system. iv Contents 1 Introduction 1 1.1 History and motivation . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.2 Current technologies and problem statement . . . . . . . . . . . . . . 2 1.3 Objectives and thesis approach . . . . . . . . . . . . . . . . . . . . . 3 1.4 ThesisStructure.............................. 4 2 Theoretical framework 6 2.1 The heart as an electrical source . . . . . . . . . . . . . . . . . . . . . 6 2.1.1 The heart and the heartbeat . . . . . . . . . . . . . . . . . . . 6 2.2 Electrocardiogram (ECG) . . . . . . . . . . . . . . . . . . . . . . . . 7 2.2.1 Leads ............................... 7 2.2.2 TheECGwave .......................... 10 2.3 Pressuresignals.............................. 12 2.3.1 The pulse waveform . . . . . . . . . . . . . . . . . . . . . . . 13 3 Hardware development 15 3.1 ECG acquisition system . . . . . . . . . . . . . . . . . . . . . . . . . . 15 3.2 General acquisition system . . . . . . . . . . . . . . . . . . . . . . . . 18 3.3 Powersupply ............................... 20 4 Software development 22 4.1 Programming the microcontroller . . . . . . . . . . . . . . . . . . . . 22 4.2 Data acquisition software . . . . . . . . . . . . . . . . . . . . . . . . 24 4.3 Data processing, representation and parameters . . . . . . . . . . . . 25 4.3.1 Signalfiltering .......................... 26 4.3.2 Signal representation and automatic parameter extraction . . 28 4.3.3 Synchrony between devices . . . . . . . . . . . . . . . . . . . 32 4.3.4 Transform matrices . . . . . . . . . . . . . . . . . . . . . . . . 33 4.3.5 Finalprogram........................... 37 5 Measurements, results and discussion 39 5.1 Power consumption test . . . . . . . . . . . . . . . . . . . . . . . . . 39 5.2 Materialcosts............................... 41 5.3 ECGmeasurements............................ 41 5.3.1 12 ECG leads acquisition . . . . . . . . . . . . . . . . . . . . . 41 v 5.3.2 Reconstruction of the 12 ECG leads . . . . . . . . . . . . . . . 45 5.4 Pressure signal measurements . . . . . . . . . . . . . . . . . . . . . . 46 6 Conclusions 48 6.1 Future work and perspectives . . . . . . . . . . . . . . . . . . . . . . 49 Bibliography 50 Acronyms 54 Appendices 55 A Additional Figures 55 A.1 SchematicsandPCBs........................... 55 A.1.1 ECGdevice ............................ 55 A.1.2 Genericdevice .......................... 58 A.2 Finalprogram............................... 61 A.3 Measurements .............................. 65 A.3.1 ECG measurements . . . . . . . . . . . . . . . . . . . . . . . . 65 A.3.2 Pressure signal measurements . . . . . . . . . . . . . . . . . . 67 A.3.3 ECG reconstruction . . . . . . . . . . . . . . . . . . . . . . . . 69 B Codes 70 B.1 Microcontroller codes . . . . . . . . . . . . . . . . . . . . . . . . . . 70 B.1.1 STM32maincode ........................ 70 B.1.2 ADS initialization . . . . . . . . . . . . . . . . . . . . . . . . . 72 B.2 Matlabcodes ............................... 74 B.2.1 ECG Pan Tompkins algorithm . . . . . . . . . . . . . . . . . . 74 B.2.2 Filtering.............................. 76 B.2.3 PQRSTdetection ......................... 77 B.2.4 Pulse wave signal detection . . . . . . . . . . . . . . . . . . . 80 vi 1 Introduction 1.1 History and motivation Ever since ancient times, medical diagnosis have been of great importance for civilizations, as physicians were able to describe different disturbances and diseases, based on the observation of clinical symptoms [1]. The history tells how diagnosis has been improved through the development of new methodologies which allowed to gather valuable knowledge about symptoms and illnesses. This knowledge has been widely broadened over time, but it is never a closed book, due to the appearance of new diseases such as the recent COVID-19 of last year. In the last century, it has been possible to obtain biological signals from the body which can give medical information about the health status of a subject. Nowadays, with the use of the latest technologies, this data can be easily acquired with great quality. These devices are able to capture physiological signals, which are then processed in order to obtain parameters and information about the state of the patient, and therefore prevent any complications. The literature suggests two main types of biosignals for the cardiovascular system: electrical signals of the heart, and pressure signals of the blood circulation. The first one is typically performed through the electrocardiogram (ECG), used for measuring the electrical activity of the heart. It has been the most studied, as well as the electrode positioning around the body, and technology has allowed it to be the best non-invasive procedure to detect heart abnormalities. The second type of biosignal can be measured through different procedures such as photoplethysmography (PPG), impedance plethysmography (IPG), or tonometry, which all have similar waveforms. The motivation of this project is to develop the electronics of a device capable of performing the acquisition of these cardiovascular signals, in order to characterize the heart and the circulation of blood. Nevertheless, we want our system to be flexible and measure any biological signal or parameter, some of them measured directly as biopotentials like ECG, electroencephalogram (EEG) or electromyogram (EMG), and other signals will be acquired with front-ends. For this, it will be necessary to 1 design the hardware of the device, as well as the software that will control it, with the algorithms for the automatic extraction of the important parameters. 1.2 Current technologies and problem statement As mentioned before, technology has allowed to greatly advance and improve the medical diagnosis methods, to the point that many commercial devices are in the market offering different functionalities for the biosignals acquisition. However, most of these devices could present some lacks, such as some minor distortions, the consumption and market prices, or even the incommodity of the electrodes placement. Nowadays, for a conventional ECG measurement device, a set of 10 electrodes are needed to get the complete information of the heart status with a good quality. These electrodes are placed at the standard positions of the limbs (arms and legs) and the chest, so it is clear that it could be an inconvenience for the patient being connected to 10 electrodes at the same time, as well as it is not practical. In order to solve this problem, in the literature [2] [3] [4] it is found that some devices have been invented in order to obtain the required 12 ECG lead signals from a subset of leads, allowing this way to use less electrodes and to place them in comfortable positions of the body, even if they are not standard. Nevertheless, using this kind of techniques is that the correlation between the real and the synthesized one is not enough due to a loss of information, so these procedures cannot be used for diagnostic purposes, but it can be useful for monitoring. The commercial devices do not usually design their hardware in order to provide a low-power consumption, they prefer to develop a functional device which can offer as many functionalities as possible in order to have a competitive product. In addition, most of these devices can be very expensive because of the costs of manufacture and branding; the prices in the market are around thousands of dollars. Another problem of current technologies is their lack in functionalities, as some devices offer a very good quality in some aspects such as signal processing, but they are limited in other things like the number of channels to connect signals. The device in which this whole project is inspired is the one owned by the Instrumentation, Sensors and Interfaces research group: the BIOPAC Lab System [5]. This commercial device includes a user-friendly software that is capable of acquiring in real time different biosignals (muscular, cardiovascular and neurophysiological). However, the limitations presented by this device are the number of channels, the connection for these channels, and depending on the model the sampling resolution. 2Chapter 1 Introduction In conclusion, what is needed to develop is a low-cost and low-power consumption device which can compete in the market with the commercial ones, being it even improved to solve the limitations of them and provide many functionalities at the same time. Another important thing to take into account is the possibility of reaching not only a home device, but also to reach anywhere in the world. 1.3 Objectives and thesis approach As it has been stated, the aim of this project is the development of a functional low-cost and low-power consumption portable device, capable of monitoring physiological parameters from a limb. The project will be carried out under three main objectives. First of all, as a prelude, some research will be necessary to learn about the biological processes that generate the different types of signals, as well as the mechanisms of recent technologies for the data acquisition. In addition, a further study on the typical parameters of each signal will be useful in order to implement the different algorithms for our software. After this, the first goal is to design the low-cost and low-power consumption hardware for multiparametric measurements. The starting point is the commercial BIOPAC system mentioned before, as it will be the inspiration for the development of a device that solves its limitations. The idea is to design the hardware of two different acquisition devices, one focused on ECG signals, and the other used for generic signals such as IPG, PPG or ECG as well. All of this will be controlled by a microcontroller that will receive the data with SPI communication protocol, and then transmit the information by USB to a computer. A final PCB board will be designed in order to gather all the different hardware components necessary for our purpose. Finally, the developed hardware must be able to perform good quality measurements in ambulatory scenarios. The second objective is the development and implementation of the software. This comprises different aspects; first of all, the microcontroller program capable of controlling the whole system. Then, a program that receives the data and saves the information so that the data can be processed. Finally, the different data processing algorithms, plot representations and parameters must be obtained with another program as well; the main idea is to use MATLAB, and create an application program like the one provided for the BIOPAC system. 1.3 Objectives and thesis approach 3 Fig. 2.3.: Visual representation of the limb and precordial leads. 2.2.2 The ECG wave The wave considered in an electrocardiogram is related to the heartbeat cycle mentioned before. Each phase of the cycle produces an electric potential, and the combination of all parts form the ECG signal, as it is shown in the next figure. We can see that different parts appear in this signal, labelled with the letters P, Q, R, S and T. Each part corresponds to a different stage in the depolarization of the hearth. Fig. 2.4.: Representation of the different waves corresponding to the propagation of the electrical potential over different instants of time. By inspection of one or multiple leads of the ECG it is possible to obtain a huge amount of information such as if the patient has suffered from a myocardial infarction or any kind of arrhythmias. Actually, the goal of the ECG interpretation is the ability 10 Chapter 2 Theoretical framework to determine whether the ECG waves and intervals are normal or pathological. Electrical signal interpretation gives a good approximation of heart pathology [8]. All of the waves on an ECG tracing and the intervals between them have a predictable time duration, a range of acceptable amplitudes (voltages), and a typical morphology. Any deviation from the normal tracing is potentially pathological and therefore of clinical significance. Some of the most important information to take into account from an ECG wave are the following points. • P wave: it represents the depolarization of the atria, where the contraction of the atria is produced. Atrial depolarization spreads from the SA node towards the AV node (see Fig. 2.4), and from the right atrium to the left atrium. • PR interval: it reflects the time the electrical impulse takes to travel from the sinus node through the AV node. • QRS complex: it represents the rapid depolarization of the right and left ventricles. The ventricles have a large muscle mass compared to the atria, so the QRS complex usually has a much larger amplitude than the P wave. At this moment, systole is produced, where there is a ventricular polarization and contraction. • ST interval: it represents the period when the ventricles are depolarized. • T wave: it represents the repolarization of the ventricles. It is generally upright in all leads except aVR and lead V1. A relaxation of the ventricles occurs. It is interesting to notice that there is no wave corresponding to atrial repolarization, this is due to the fact that it is masked by the QRS complex. Another interesting parameter to take into account is the distance between two consecutive R peaks, the RR interval, which indicates the time between heartbeats. With this information, we can also obtain the tachogram, which compares the time intervals between heartbeats. 2.2 Electrocardiogram (ECG) 11 Fig. 2.5.: Illustration of the ECG wave with the important peaks and intervals One of the current main limitations in the measurement technique of this signal is that it takes a long time to carry out, because as it has been explained, a total of 10 electrodes (2.1) are needed to be located in some places that are not directly accessible, specially in technical placements of the electrodes (for instance, at the torax). This can also be a problem in clinical environments where time is gold in order to attend all the patients, and therefore the importance of optimizing the measurements with less and more comfortable electrodes, or even the possibility to make the measurements at home. 2.3 Pressure signals There are many pulse wave techniques used to detect blood volume changes in the microvascular bed of tissue. As they all reach the same type of signals, this project will explain Photoplethysmography (PPG), which is a simple and low-cost optical technique. Unlike ECG that uses the electrical properties to obtain a signal, PPG is focused on pressure signals of the body. PPG is a non-invasive technology that uses a light source and a photodetector at the surface of the skin to measure the volumetric variations of blood circulation [10], often used for heart rate monitoring purposes. 12 Chapter 2 Theoretical framework The measurement of blood pressure is critical to the treatment and management of many medical conditions [11]. High blood pressure is associated with many chronic disease conditions, and is a major source of mortality and morbidity around the world. Green and red infrared light are often used to obtain PPG signals because of the difference in the wavelength; each light penetrates human tissue differently. Infrared light has the deepest penetration ability, and it can reflect the blood pulse from deep tissue. This is why the detection of blood pressure, atherosclerosis, blood sugar, and other physiological parameters uses the infrared light to obtain PPG signals. Nevertheless, recent technologies are not very reliable in blood pressure measurements from a pulse waveform, so this project will not study the blood pressure, but the pulse waveform propagation through the body. 2.3.1 The pulse waveform The PPG waveform comprises a pulsatile (AC) physiological waveform attributed to cardiac synchronous changes in the blood volume with each heartbeat, and is superimposed on a slowly varying (DC) baseline with various lower frequency components attributed to respiration, sympathetic nervous system activity and thermoregulation [12]. The AC component depicts changes in blood volume, which are caused by cardiac activity and depend on the systolic and diastolic phases [10]. The systolic phase (also known as rise time) starts with a valley and ends with the pulse wave systolic peak. The pulse wave end is marked by another valley at the end of the diastolic phase. Some features such as rise time, amplitude, and shape can predict vascular changes in the bloodstream. Additionally, PPG can be used to measure the variations between heartbeat time intervals (Peak-to-Peak or P-P Interval). The variation can be due to many factors such as the individual’s age, heart conditions, and physical fitness. The PPG signal is divided into two unique phases: the rising edge of the pulse called anacrotic, which primarily describes the systole, and the falling edge of the pulse called the catacrotic, which represents the diastole. Additionally, a dicrotic notch is typically visible at the catacrotic phase. The second derivative of this signal, known as acceleration photoplethysmogram (APG), is typically used because it is an indicator of the acceleration of the blood. There are also a number of critical points that can be extracted from the second derivative wave of a PPG signal. These critical points can be used to detect and diagnose cardiac abnormalities. One last thing that can be done is that with the ECG and pulse waves signals, we can measure some important time intervals between waves that give information about 2.3 Pressure signals 13 Fig. 2.6.: On the left, a typical PPG signal. On the right, the comparison between the PPG and APG key points. the circulation of blood and the cardiovascular system. On each heartbeat, the blood ejected by the heart causes an increase of arterial pressure that propagates through the arterial tree from proximal to distal sites [13]. The velocity of this pulse is called pulse wave velocity (PWV), and it is related to the stiffness of the arterial walls, measured by dividing the estimated distance between these points by the difference of time arrival at two sites; this is the pulse transit time (PTT). As volume and flow are in phase on early waveform features, the arrival of the pressure pulse can be measured with the PPG, as it is the most widely used signal for that purpose due to its simplicity. Fig. 2.7.: Representation of the important time intervals between ECG and PPG signals. These three time intervals, the PTT, PAT, and the PEP, are widely used in research and clinical practice since several cardiovascular parameters of interest can be obtained from them. The PTT is a direct indicator of arterial stiffness, which is related to several factors such as ageing or hypertension and associated to many cardiovascular diseases including stroke and myocardial infarction. The PAT has proved to be useful for the detection of sleep disorders and autonomic nervous system activation. The PEP can be used to monitor left ventricular contractility, which provides valuable information about myocardial performance, and sympathetic activity under stable conditions of preload and afterload. 14 Chapter 2 Theoretical framework 3 Hardware development As it has been exposed in the introduction section, the aim of this project is to develop two acquisition systems: one specific for ECG signals and the other for general purpose. The idea is to design and implement low-cost and low-power devices which make a difference with respect to the commercial devices that cost thousands of dollars and have high consumption. Both systems have been designed under the same main structure. Each one will have a specific integrated circuit which will acquire the signal from the body. This signal will be transferred to a microcontroller that will be programmed to receive and then send the data to the acquisition software, so that it can be processed. Fig. 3.1.: Block diagram of the acquisition system to be implemented. First, prototypes of both systems will be done with commercial development kits, so that they can be studied and improved to the project’s objective. Then, a PCB circuit board will be designed to replace the prototypes and get a final hardware. 3.1 ECG acquisition system In order to have a final design, the start point is to work with the commercial development kit of the specific integrated circuit for ECG signals: the ADS1298 from Texas Instruments [14]. The first thing is to study the schematic and the jumpers setting from the user’s guide provided by Texas Instruments [15]. This way, the different functionalities of the acquisition integrated circuit have been learned, which are the following. 15 – 8 channels with low-noise PGAs and high resolution 24-bit ADCs, with lowpower consumption of 0.75mW/channel. – Configurable data sampling rate from 250 SPS to 32 kSPS. – Programmable gain of 1, 2, 3, 4, 6, 8 and 12. – Supports systems meeting AAMI EC11, EC13, IEC60601-1, IEC60601-2-27, and IEC60601-2-51 Standards. – Functionalities: Built-In Right Leg Drive Amplifier, Lead-Off Detection, Wilson Center Terminal, Pace Detection, Test Signals. – SPI compatible Serial Interface. The way of transmitting the ADS1298 data to a microcontroller is through SPI communication, so it is needed a microcontroller capable of this. It has been chosen an ultra-low-power and high performance microcontroller: the STM32L4R5 from the L4 family of STMicroelectronics [16] [17], which is precisely the low-power consumption family. For our project, it is a more than enough choice, because most of the microcontroller functionalities will be unused, so there exists the possibility to easily extend the functionalities of our platform. The microcontroller will be programmed in order to receive the signal data from ADS1298 through SPI communication, prepare the data to be sent including some headers for each channel so that they are separated, and transmit the data through UART communication to an Arduino circuit board. An external circuit board which converts from UART to USB will be only used for the prototype, as a way of testing if serial USB communication is enough to transmit data to the computer software. Fig. 3.2.: ECG acquisition system prototype. 16 Chapter 3 Hardware development Once we prove that the whole system works and data is received correctly, the next step is to design the final electronic circuit. It has been decided to build the hardware on the STM32 device, so that the designed PCB board can be placed over the microcontroller board as a shield. This way, the dimensions of the device will be those of the STM32 board, which are good enough. Therefore, the design of the PCB board must have the ADS1298 integrated circuit and the serial communication port. The electronic circuit for the ADS1298 part must be adapted from the commercial development kit, as it will have a specific configuration and some parts of the original circuit will be removed. After a deep look at the schematics, a final design has been made, as we can see in A.1.1. This schematic has been done with Eagle, a software of PCB designs. With the bill of materials of the circuit provided by the user’s guide [15], the dimensions of the components used can be known in order to make a physical design of the PCB board (Fig. A.2). Once it is all designed, the next step is to print the PCB board on a copper plate; for this, the information of the design is passed to a machine that is capable of “printing” all the tracks and holes of the PCB board, using drills of different sizes. Finally, the last step is to weld all the components in their respective positions, so that we end up with our final device. (a) Upper view. (b) Side view. Fig. 3.3.: Pictures of the ECG acquisition device. In this device, the data must be collected through a DB15 connector; this is because a specific cable is used, which has this type of connector, as well as the connection to the 10 electrodes necessary for a 12 ECG lead measurement. 3.1 ECG acquisition system 17 Fig. 3.4.: Illustration of the DB15 connector for the 10 ECG electrodes. 3.2 General acquisition system The procedure of prototyping and ending up with a final design will be the same as explained for the ECG acquisition system. The difference between the previous one is that now we will use another integrated circuit: the ADS1299, also from Texas Instruments [18]. This integrated circuit is not specific of any particular signal, so it is ideal for generic use; with this one could measure ECG, PPG and other biological signals of the body. Taking a look into the user’s guide of the development kit [19], the following characteristics are found. – 8 channels with low-noise PGAs and high resolution 24-bit ADCs, with lowpower consumption of 0.75mW/channel. – Less input-referred noise than the ADS1298. – Configurable data sampling rate from 250 SPS to 16 kSPS. – Programmable gain of 1, 2, 4, 6, 8, 12 and 24. As it reaches a greater gain, an electroencephalogram signal can be acquired, which is a very small signal. – Supports systems meeting AAMI EC11, EC13, IEC60601-1, IEC60601-2-27, and IEC60601-2-51 Standards. – Functionalities: Built-In Bias Drive Amplifier, Lead-Off Detection, Test Signals. – SPI compatible Serial Interface. Another STM32L4R5 microcontroller will be used, as well as the same SPI and UART communication for data transmission. The difference between the two systems will 18 Chapter 3 Hardware development be how the microcontroller is programmed in order to use the integrated circuit, as it will be explained in the next section. For the prototype, the external UART/USB conversion board will be still used so that we can prove that the system works correctly. Fig. 3.5.: Generic acquisition system prototype. As we proceeded in the ECG system, the next step is to design the final circuit by examining the original schematic of the development kit. Luckily, most of the circuit is practically the same as the previous system, so we only had to do some modifications of the final schematic of the ECG system, resulting in the schematic seen in A.1.2. Like before, we get the bill of materials from the user’s guide [19] so that we know the dimensions of the components and we make a physical design of the PCB board (Fig. A.3). Then, we print the design in a copper plate, and finally we weld all the components in their respective places. All in all, we end up having our final device. (a) Upper view. (b) Side view. Fig. 3.6.: Pictures of the Generic acquisition device. 3.2 General acquisition system 19 4.3.1 Signal filtering The first thing to do with the acquired data is to pre-process the information, due to the fact that the real world adds some undesired artifacts and noise which should be removed. As the ECG is a cyclic signal, it is governed by a set of frequencies that correspond to the different parts of the heart cycle. For instance, comparing the ECG frequencies with the common frequencies of artifacts and noise, it can be seen how these interferences affect directly to the signal. Source Frequency Heart rate 0.67 - 5 Hz (40 - 300 bpm) P-wave 0.67 - 5 Hz QRS 10 - 50 Hz T-wave 1 - 7 Hz High frequency 100 - 500 Hz potentials (a) Common ECG frequencies Source Frequency Muscle 5 - 50 Hz Respiratory 0.12 - 0.5 Hz (8 - 30 bpm) External electrical 50 - 60 Hz Other electrical typically > 10 Hz (b) Common noise and artifact frequencies Tab. 4.1.: The most important frequencies of an ECG signal In order to see the spectrum of a signal, MATLAB offers a Fast Fourier Transform function, which represents the different existing frequencies of the signal. With this, we can see if the undesired frequencies are present, so that we can apply filters for at least reducing these interferences, and try to obtain the cleanest signals possible. Doing some research, we find out that the typical filters used for biological signals are high and low pass filters, and notch filters [21]. High and low pass filters On the one hand, low-pass filters on the ECG are used to remove HF muscle artifacts and external interference. They typically attenuate only the amplitude of higher frequency components. On the other hand, high-pass filters remove LF components such as motion artifact, respiratory variation, and baseline wander. The combination of both the high and low pass filters actually result in a bandpass filter, allowing only a certain frequency band to pass through. In MATLAB, there are many ways to implement this kind of filters, and the one chosen is through a Butterworth filter, as it is done in [6]. This filter is characterized by offering a frequency response as flat as possible in the passband, as the filter order is increased. In our project, 5th order Butterworth filters have been used as high and low pass filters, with 0.5 Hz and 45 Hz cut-off frequencies, respectively. With this, we get rid off respiratory and other electrical noises in higher frequencies, but doing 26 Chapter 4 Software development (a) Low-pass filter. (b) High-pass filter. Fig. 4.3.: The combination of (a) and (b) result in the Band-pass filter. again the spectral representation with the Fast Fourier Transform, we still see that the external electrical frequency at 50 Hz has not been entirely attenuated. Notch filter The purpose of this filter is to resolve the 50 Hz interference. It is basically a stop band filter which attenuates a specific frequency, removing then the external electrical frequency, or any other undesired interference that could be seen in the spectral representation. In particular, MATLAB offers a 2nd order Infinite Impulse Response (IIR) Notch filter, which is the one used in the project. Fig. 4.4.: Notch filter at the 50 Hz frequency. It is important to take into account that all filters introduce distortion in the resulting output signal. The distortion can be in amplitude or phase; basically, they distort different frequencies differently causing this distortion. This is why it is not always convenient to apply all the filters, as sometimes applying only the bandpass, the notch, or none, one can obtain clearer signals. The idea of the pre-processing is to analyze which case gives the best performance, so that accurate information can be then obtained from the signals. 4.3 Data processing, representation and parameters 27 4.3.2 Signal representation and automatic parameter extraction After the filtering and synchronization processes, the signals can be finally represented in a MATLAB plot. Recall that the data acquired from the LabVIEW program is given as a value for every sample, and this value corresponds to the electrical potential in a magnitude of Volts. Therefore, if we represent directly these data, we would obtain a plot of the electrical signal but with respect to the samples. The idea is to get the plot with respect to the time, so what is done is the conversion of samples into time, by means of the sampling frequency. This is: t=N Fs (4.1) where N is the sample, Fs the sampling frequency, and t the time associated with the sample. Thus, we obtain the graphical representation of the electrical signal with respect to the time of the measure. The next step is to get the typical biological parameters of the electrical signals, which can give to the user some information about the biological status of the body. For this, a series of algorithms will be used, as we can find in the literature of this topic. Pan-Tompkins The first algorithm to take into account is the Pan-Tompkins algorithm, which is commonly used to detect QRS complexes in ECG signals. The reason for finding this particular part is that it is the main peak of the signal, so it is easily detected, as well as it is commonly used for heart rate monitoring. [22] The Pan-Tompkins algorithm applies a series of filters to highlight the frequency content of this rapid heart depolarization and removes the background noise. For this, high and low pass filters are needed, which have been already implemented, but a derivative filter is also required. This last filter accentuates QRS complexes relative to P and T waves, providing information of the QRS complex slope. y(n) = 1 8[2x(n) + x(n−1) −x(n−3) −2x(n−4)] (4.2) 28 Chapter 4 Software development Then, the algorithm squares the signal to amplify the QRS contribution. This is done to amplify the output of the derivative filter, emphasizing the higher frequencies and setting all values into positive ones. Finally, it applies adaptive thresholds to detect the peaks of the filtered signal; this is done with a moving-window integrator. It smooths the spiky signal coming from the squaring output, and it is applied to obtain waveform feature information. The window has the following equation: y(n) = 1 N[x(n−(N−1)) + x(n−(N−2)) + ... +x(n)] (4.3) where N is the number of samples in the width of the integration window, which should be the width of the QRS complex (about 150 ms). This algorithm disposes of a decision rule in order to validate if the QRS complex found is valid, which is performed with an adaptive thresholding operation [6]. First of all, a set of samples are produced at the location of the maxima in order to localize the candidates to QRS complexes. Then, the algorithm uses two threshold values (signal and noise thresholds) to evaluate whether the maxima is or not a QRS complex. The algorithm also eliminates physically impossible events (such as two QRS complexes within less than 200 ms), and therefore removing false positives. Additionally, if an unreasonably long period has expired without an above threshold peak, the algorithm will assume a QRS has been missed and perform a search back. Finally, if a QRS candidate occurs after the 200 ms period but within 360 ms of the previous QRS, the algorithm determines whether it is the QRS complex of the next heartbeat, or a possible anomaly of the T wave. P, Q, S and T peak detection With the Pan-Tompkin algorithm, the QRS complexes are determined by finding the maxima of the QRS peaks, so in practice only the R peak indexes are found. The objective of the biological parameters is to evaluate the different intervals of time between all the signal peaks, so we should as well find the P, Q, S and T peaks. This code is based on [23]. The idea of these detections is to find the maxima or minima in strategic windows of the signal, starting from the Pan-Tompkin algorithm in which we can know the position of the R peaks. Doing some research about the typical position of each peak with respect to the R peak, the windows can be easily selected, and once we are in the specific region of each peak, we can find the maxima or minima; for the P and T peaks the maxima are found, whereas for the Q and S peaks the minima are seeked. 4.3 Data processing, representation and parameters 29 Fig. 4.5.: P, Q, R, S and T indexes found with the algorithm. At the end, all the peak position indexes are found, which can be taken as samples or times (it can be converted from one to another by means of the sampling frequency). The only thing left to do is the time intervals, which is easily done once we obtain all the indexes; these intervals are directly the difference of time between certain peaks. The typical intervals which are found for medical information are the ones presented in Fig. 4.6. The main interval could be the R-R interval, as it takes into account the time between heartbeats, and therefore with which one can know the heart rate in beats per second. Fig. 4.6.: Representation of the important intervals of time in the ECG signal. Pressure signal analysis Apart from measuring ECG signals, the aim of the generic device is to be capable of measuring other biological signals. This project will focus on the PPG signal, whose waveform is different from the ECG, so another algorithm will be needed in order to find the key features of this signal. These key points will be the systolic peak, the dicrotic notch, the maximum slope and the valley where the systolic peak begins. 30 Chapter 4 Software development As for the ECG signal, the first step is to seek a relevant point for each heartbeat and then find the others with the help of windows. The first tests tried to find the systolic peak, so that it was similar to the Pan-Tompkin algorithm. However, it was found out that depending on the PPG signal, it was difficult to find some of these peaks because some of them did not reach the same amplitude as the others, and they were indetectable. Therefore, another strategy was thought: finding the maximum slope points, which are always just before the systolic peak. For this, the first derivative is needed, because finding the maxima of the derivative will directly provide the maximum slope points of the PPG signal, and then a reliable starting point to find the rest of the key points. Once the maximum slope indexes are found, a series of windows are used to find the other key features. The systolic peak can be easily found by seeking for the maxima of the particular window just after the maximum slope point; the same for the valley where this peak starts, by finding the minima. The trickiest one is the dicrotic notch, which will use the first derivative as well. The literature [24] suggests that the dicrotic notch can be detected as follows: taking the first derivative, from the maxima point seeked for the maximum slope, a window is used in order to find the minima just after this maxima point, which will correspond to the descendent slope after the systolic peak. From this minima, another window is taken in order to find the following maxima; this new peak in the derivative corresponds to the dicrotic notch point. Fig. 4.7.: First derivative of the PPG signal. In black, the peak corresponding to the dicrotic notch. With all the indexes found for the different key features, the last thing to do is to get the parameters. In this case, the time intervals are found as well, but for the PPG signal it is also important to get the amplitudes of the points, because the areas are directly related to the blood pressure and they give meaningful medical information about the patient. The time and amplitude intervals considered are: from minimum to maximum slope, from minimum to maximum, from slope to maximum, and from maximum to dicrotic notch. The intervals between the same points but from two consecutive heartbeats are also considered. 4.3 Data processing, representation and parameters 31 Fig. 4.8.: Representation of the important points in the PPG signal. 4.3.3 Synchrony between devices For the ECG processing, before getting the parameter information, there is a previous step to be done. One of the approaches of the project is to be able to obtain the 12 ECG leads from a subset of some of these leads, by means of transform matrices as it will be explained in further sections. For this, we need to acquire the data from both devices, and this could add some time delay between the signals, as they could be acquired at different instants of time. (a) Tachogram with one heartbeat delay. (b) Synchronized tachogram. Fig. 4.9.: Representation of the tachogram synchrony. In order to solve this problem, a synchronization process is carried out, which uses the tachogram of the signals of both devices. The tachogram is built by finding the heartbeat peaks (R peak) with the Pan-Tompkin algorithm explained before, as it is the clearest peak of any biosignal, so that the difference between two consecutive 32 Chapter 4 Software development heartbeats can be computed. This tachogram signal is the same for the acquired data from both devices, so the way of solving the delay problem is to represent both tachogram signals and see how much delay there is between them. Once we know the delay, represented by a number of heartbeats, we update the original signals so that each of them start from the same heartbeat. Therefore, we will have the ECG signals synchronized and ready to be compared. 4.3.4 Transform matrices One of the main and most interesting objectives of this project is to be capable of obtaining the 12 conventional ECG leads from a standard, or even from a notstandard subset. Then, it is necessary to pass from a smaller dimension of data to a greater one; in algebra, this is the same as thinking of certain matrices that contain the required information to reach this dimension increment. Thus, there must exist transformation matrices which relate the source leads (the acquired subset) with the target leads (the 12 ECG reconstructed leads). This project will use three different methods to obtain these matrices, based on [6]. This is one of the advantages of working with the same research group, as the codes are inherited and it is not necessary to start from nothing, but only adapt the codes to our purposes. Types of matrices First of all, it is interesting to know about the different types of matrices that can be obtained. Most of them are fixed and do not adapt to changes over time, but there are also adaptive matrices. However, the latter ones do not manifest any improvement with respect to the non-adaptive matrices, but in practice they are better to deal with measures that change in the heart rate or movement. The first type of matrix is the universal transform based on statistical data or models, which can be applied to any patient regardless of their condition. It is logical to think that these matrices are not very reliable due to the variance that can produce a generalization across patients, as there will be differences between them such as gender, age, or diseases. Secondly, there are population-specific transforms, which are more effective than the universal ones. The idea is to separate the patients into different groups that will have similar characteristics, depending on the age, gender or possible diseases. Nevertheless, other important properties of the heart position or orientation can not 4.3 Data processing, representation and parameters 33 be taken into account, which are directly related to the signals received, so there are still some uncertainties about the reliability of these matrices. The last type is patient-specific transforms, which are the ones with the best performance as one could expect. This is the same as considering that every person is different and with unique characteristics, and therefore each one must have their own matrix. The project attaches to this last type of transforms, so that each of the three methods obtain the specific matrices for each patient. Methodology Three different methods will be used in order to see which of them give the best results. They all take 3 beats to synthesize the matrices needed later to reconstruct the signal. The methods are the following. • Method 1 This method will use the Singular Value Decomposition (SVD) in order to obtain the transform matrices. The SVD is useful to decompose large n x m matrices into three matrices, so the first step is to decompose the matrix containing both the source leads and the target leads measured. In the same way, the matrix containing only the source leads is also decomposed, so that we end up with: An×m=ˆ Un×mˆ Σm×mˆ VT m×m(4.4) A0 n×p=ˆ U0 n×pˆ Σ0 p×pˆ V0T p×p(4.5) where n is the number of samples, m the number of total leads and p the number of source leads (the subset). U and V matrices, corresponding to the entire set of leads, must be kept as they carry the information of the 12 leads, specifically information about the shape and the leads themselves. In order to pass the subset of leads into a higher dimension, a matrix T1 is defined to relate the information of ˆ Σwith ˆ Σ0: ˆ Σm×m=T1 m×pˆ Σ0 p×m(4.6) 34 Chapter 4 Software development Notice that ˆ Σ0 is presented with a different number of columns so that it fits to the dimensions of ˆ Σ. With this relation, we can obtain T1as follows: T1 m×p=ˆ Σm×mˆ Σ0−1 p×m(4.7) The last step is to reconstruct the 12 leads using the U , V , and T1 matrices, as well as the ˆ Σ0 matrix to transmit the data from the source leads. It goes like this: An×m=ˆ Un×m(T1 m×pˆ Σ0 p×m)ˆ VT m×m(4.8) • Method 2 This second method directly relates both dimensions of the source and target leads, so that the target leads are a linear combination of the source leads. The direct transform matrix T2 makes the following relation. An×m=A0 n×pT2 p×m(4.9) This equation is solved by applying a Least Squares Method (LMS) that uses SVD as well. Notice that for this method it is not necessary to reconstruct the leads with 3 beat periods, but it is applied in a way that the correlations at intervals of 3 beats can be compared to other methods. • Method 3 The last method is a combination of the previous two methods; the source and target leads matrices are still being decomposed by SVD as in method 1, but now the diagonal matrices ˆ Σ and ˆ Σ0 are related by a linear combination of their singular values. σ=σ0tp×m(4.10) 4.3 Data processing, representation and parameters 35 excel file. For the ECG acquisition device, it is necessary to locate the 10 electrodes as it has been explained in the ECG section (Tab. 2.1). These 10 electrodes will be connected directly to the ECG system with the particular connector. For the Generic acquisition device we will use a series of front-ends provided by the research group, which are prepared to amplify the electrical signal. One of these front-ends will be the not-standard lead, where the electrodes can be placed anywhere in the body; we will see the cases for the neck and the waist. Another three front-ends will be used to measure the 3 first standard leads: Lead I, II and III; these 3 leads will be measured with a system developed by the research group that use two metallic balls for the hands and two metallic plates for the feet, that will work as electrodes but being them more comfortable to use. Fig. 5.1.: Measurements on a patient. Electrode position: 1. RA; 2. LA; 3. RL; 4. LL; 5. Chest electrodes; 6. Not-standard position. Once the measurements have been performed and the data has been saved, we can now go to the processing program in which we will represent the signals and extract some parameter information. We load the excel files in the program for the 12 ECG leads and the subset, and applying the filtering processes we end up with the results seen in annex A.3.1. 42 Chapter 5 Measurements, results and discussion Fig. 5.2.: Some of the 12 ECG leads acquired with the ECG device. As it can be seen, good enough quality signals are obtained. It is true that for some of them a bit of noise or interferences are observed, but this can be due to possible mechanical imperfections such as bad contacts with the electrodes or movement of the subjects. It is also important to remark that the signals can differ from different patients, as their conditions are not the same; the age, physical condition, or heart position can infer in these results, as for some subjects the signals are better and clearer than for others. The parameters of these signals have been obtained as well, following the criteria seen in Fig. 4.6. The parameters are represented in tables, where we obtain the mean and standard deviation of all the intervals. (a) Some of the parameters extracted from ECG leads. (b) Parameters extracted from the subset of leads. Fig. 5.3.: Parameters of the ECG measurements. It is observed that all the parameters in A.3.1 make sense between them, as for a same interval they are all of the same order of magnitude, and we even see small deviations of the order of the resolution error of our devices (around 2 ms). There 5.3 ECG measurements 43 are some punctual signals that have a worse performance, but as mentioned before, not all the signals can be perfect for a particular subject. In order to see the differences between not-standard locations of the electrodes, we performed two different measurements for a patient: one with the not-standard electrode in the nape, where the Wilson’s Central Terminal (WCT) is located, and the other one in the waist. The goal of these measurements is to find a comfortable electrode position for the patient, as the idea is to then reconstruct the 12 ECG leads from the subset without the need of using all the 10 uncomfortable electrodes. The results are the following. (a) Not-standard plot. (b) Mean and standard deviation. Fig. 5.4.: Results of the not-standard electrode placed at the nape. (a) Not-standard plot. (b) Mean and standard deviation. Fig. 5.5.: Results of the not-standard electrode placed at the waist. Some differences are spotted; the signal of the waist measurement seems to be clearer than the nape signal, as this last one presents some noise. In contrast, the P and T waves are better seen in the nape signal, this is why we observe greater deviations in the waist signal parameters. Nevertheless, both measurements give similar parameters, with deviations that are typical for these signals, so we can conclude that the placement of the not-standard electrode is insignificant for the nape and the waist, allowing to decide wherever is more comfortable. 44 Chapter 5 Measurements, results and discussion 5.3.2 Reconstruction of the 12 ECG leads The last thing left to do is the reconstruction of the 12 ECG leads from the subset of leads acquired. As explained in the transform matrices section, three different methods will be used in order to see which one gives the best performance for our data. The three are carried out at the same time in the code, so a total of 5 matrices are obtained: the U and V matrices which are equal for the three methods, and T1, T2 and T3, corresponding to the matrices of methods 1, 2 and 3, respectively. Using the application program, first the training phase is done in order to see which method is the best, by looking at the correlations between the 12 ECG original signal and the 12 reconstructed signals, for each one of the methods. To see rapidly the best performance, the program provides the mean of these correlations, and we get the following results for different measurements on subjects. Subjects Not-standard placement Method 1 Method 2 Method 3 Patient 1 nape 75.65% 66.01% 91.88% Patient 2 nape 58.92% 62.13% 92.26% Patient 2 waist 86.18% 90.45% 92.26% Patient 3 nape 27.94% 30.41% 90.05% Tab. 5.3.: Performance of the different methods for the measurements. As we can see, for the three measurements where the not-standard electrode was placed at the nape, method 1 and 2 do not give good results, but method 3 does. However, for the measurement where the not-standard electrode was placed at the waist, all three methods give good enough results, being method 3 better. All in all, we can conclude that method 3 gives the best performance in all cases, but for measurements where the not-standard electrode is placed at the waist the three methods give good results as well. After performing the training phase, the matrices can be saved for every patient, as they will be specific for each one. This will allow that for future measurements it will not be necessary to do the annoying 10 electrodes placement for the 12 ECG leads, as these matrices can reconstruct the 12 leads with a subset of leads measured with a comfortable placement of electrodes. We can try this with the building phase, as we can load the matrices of the subject, as well as the subset of leads measured, and reconstruct the 12 ECG leads (see annex A.3.3). This was done with Patient 1, who did a first test measurement obtaining the 12 ECG leads and the subset, and the following week performed a second measurement. With the new subset measured and the matrices obtained in the first measurement, we were able to reconstruct the 12 ECG leads and we saw that the differences in the parameters were minimum. 5.3 ECG measurements 45 Fig. 5.6.: Some of the 12 reconstructed ECG leads. Finally, we have performed the reconstruction for both not-standard placements (nape and waist), and we have seen small variations between the parameters obtained, being them practically the same. This leads to the possibility of being able to choose where to place the not-standard electrode, which could be in the most convenient position for the patient. 5.4 Pressure signal measurements In order to show that the generic device is capable of acquiring any kind of signal, some pressure signal measurements will be carried out with the devices provided by the research group. In the theoretical framework section the PPG has been explained, but we will perform IPG and tonometer measurements. Nevertheless, the theory and algorithms explained for the PPG signals are valid as well for IPG and tonometer, as they are the same type of signal. Fig. 5.7.: Measurements on a patient. Signals acquired: 1. ECG; 2. IPG; 3. Tonometer 1 and 2. 46 Chapter 5 Measurements, results and discussion Fig. 5.8.: Pressure signals and ECG. The signals acquired (Fig. 5.8) and the parameters obtained (Fig. 5.9) can be completely seen in annex A.3.2. The device provided by the research group does the measurement of the ECG, IPG (which corresponds to the aortic blood pulse) and two tonometer (which correspond to the finger blood pulse) signals at the same time, so the application program is prepared for this particular device. This way, as it has been exposed in section 2.3, we can perform the acquisition of different important time intervals between the ECG and pressure signals, in particular between the peaks of the ECG and IPG, and between the peaks of the ECG and the tonometer signals. Fig. 5.9.: Table with the time intervals of study. In figure 5.9 are represented the time intervals for each peak of the signals, with the following means: 244 ms for the ECG-IPG interval, 316 ms and 318 ms for the ECG-Tonometers intervals. As we can see, the time from which the blood exits the heart (ECG) and the instant of maximum aortic blood pressure (IPG) is greater than the interval of time from the aortic zone to the finger (tonometer). This means that the blood goes faster trough the arm, as it should be in a healthy person. 5.4 Pressure signal measurements 47 6 Conclusions The design and implementation of a functional low-cost and low-power portable system has been carried out successfully. The system finally consists of two electronic devices, one capable of acquiring the complete set of 12 ECG leads, and the other with the possibility of getting any existing biosignal. In addition, apart from monitoring signals, the software implemented computes the important parameters for each measurement, so that a medical analysis can be done for the patient. Different measurements and results have been performed during the project, where we have seen that for the two different placements of the not-standard electrode (nape and waist) there are not many differences with the parameters obtained, as they are in the range of typical values for these intervals. This means that the not-standard electrode could be placed wherever the patient feels more comfortable, something that we were seeking in this project. Moreover, the reconstruction matrices worked as expected and we are able to synthesize the 12 ECG leads from a subset and the patient-specific matrices, with correlations between the original and reconstructed leads of more than 90%. In fact, these matrices need to be obtained just once, as they reach the same results for subsets measured in different times. Thus, it is not necessary to place all the electrodes to perform future measurements, allowing more comfortable and time-saving measurements. Finally, we were able to obtain the parameters related to pressure signals, and then compute the different important time intervals between ECG and pulse wave signals. These parameters give information about the cardiovascular system and circulation of blood, so this is the last step to reach a complete medical diagnosis. Therefore, with our devices and software programs we have been able to perform a complete analysis of the cardiovascular system. It is important to remark the efficiency in the low-power consumption test with respect to the BIOPAC commercial system, with an autonomy of around 180 hours for continuous measurements and a power bank of 20000 mA. In addition, the low costs of the devices means a greater accessibility to this kind of system, because the devices designed can be affordable for anyone. Therefore, it is not crazy to think of having this system at homes, where any subject could monitor their health status constantly, and in the case something is wrong they could send the data to a 48 doctor in order to carry out a medical diagnosis. This would be a great advantage for clinical environments, which would be less congested of people and presenciality would not be necessary for minor cases or measurements. In conclusion, all of our main objectives have been reached successfully. The hardware accomplishes the specifications that we wanted in terms of consumption and good quality acquisition. The software captures, saves and processes the data so that the signals are effectively represented and the parameters are automatically extracted in order to give information about the health status of the subject. Finally, the algorithms that were proposed can be carried out in a final application program that comprises all the functionalities of our system; thus, our devices are fully competitive in the market. 6.1 Future work and perspectives Although our system is already functional in all the topics described above, there are a series of aspects that can be improved or extended, in order to have a better product. Firstly, the communication used to transmit the data could not be the fastest one, so there is the possibility to improve the communication system with other functionalities provided by the microcontroller. This is possible because the shield designed for our hardware connects all the pins of the STM32 to the PCB board designed, with the thought of using in the future some of the unused pins that can configure the other communication systems. With this we could gain some data transmission speed, and think of the possibility of representing the signals in real time. Secondly, another thing that can be done in the future is to compact the LabVIEW and MATLAB programs in just one software program, so that it would be more comfortable to acquire the data and then process the information. In addition, it would be convenient to optimize some of the codes and algorithms so that they could be useful even for bad quality signals where it is difficult to detect the important peaks or points. Finally, as an aesthetic improvement, we could compact all our hardware systems in one box with the corresponding inputs and outputs visible. Thus, we would have a final physical product that can be presented in any interested company. 6.1 Future work and perspectives 49 Bibliography [1] Darlene Berger. „A brief history of medical diagnosis and the birth of the clinical laboratory Part 1-Ancient times through the 19 th century“. In: (1999) (cit. on p. 1). [2] Philips Medical Systems. PAD 12-Lead ECG Monitoring with EASI ™ Lead System. 2007 (cit. on p. 2). [3] Sibasankar Padhy and S Dandapat. „Synthesis of 12-Lead ECG from a Reduced Lead Set using Singular Value Decomposition“. In: IEEE Power, Communication and Information Technology Conference (PCITC) 322.10 (2015) (cit. on p. 2). [4] Julie A. Scherer and John M. Nicklas. „Synthesis of the 12 lead electrocardiogram from a 3 lead semi-orthogonal subset using patient-specific linear transformation arrays“. In: 1988 (cit. on p. 2). [5]Biopac Systems. „BIOPAC Product Sheet“. Pat. 2021 (cit. on pp. 2, 40). [6] David Cano. „Synthesis of the standard 12 leads from atypical leads“. Universitat Politècnica de Catalunya, June 2018 (cit. on pp. 4, 25, 26, 29, 33). [7]Wikipedia. Cardiac Cycle. 2021 (cit. on p. 6). [8] Yasar Sattar and Lovely Chhabra. Medical Reference - Electrocardiogram. Mar. 2021 (cit. on pp. 7, 11). [9]Wikipedia. Electrocardiography. June 2021 (cit. on pp. 7, 9). [10] Mohammad Ghamari, Denisse Castaneda, Aibhlin Esparza, Cinna Soltanpur, and Homer Nazeran. „A review on wearable photoplethysmography sensors and their potential future applications in health care“. In: International Journal of Biosensors Bioelectronics 4 (4 2018) (cit. on pp. 12, 13). [11] Mohamed Elgendi, Richard Fletcher, Yongbo Liang, et al. „The use of photoplethysmography for assessing hypertension“. In: npj Digital Medicine 2 (1 2019) (cit. on p. 13). [12] Allen John. „Photoplethysmography and its application in clinical physiological measurement“. In: Physiological Measurement 28 (3 2007) (cit. on p. 13). [13] Joan Gómez Clapers and Ramon Casanella Alonso. „Assessment of trends in the cardiovascular system from time interval measurements using physiological signals obtained at the limbs“. Universitat Politècnica de Catalunya, July 2015 (cit. on p. 14). [14] Texas Instruments. „ADS129x Low-Power, 8-Channel, 24-Bit Analog Front-End for Biopotential Measurements“. Pat. 2010 (cit. on p. 15). 50 [15] Texas Instruments. „User’s Guide ADS1298ECG-FE/ADS1198ECG-FE“. Pat. 2010 (cit. on pp. 15, 17). [16] STMicroelectronics. „STM32L4R5xx Ultra-low-power Arm ® Cortex ® -M4 32-bit MCU+FPU, 150DMIPS, up to 2MB Flash, 640KB SRAM, LCD-TFT MIPI DSI, ext. SMPS“. Pat. 2017 (cit. on p. 16). [17]STMicroelectronics. „Nucleo-144 board User Manual“. Pat. 2015 (cit. on p. 16). [18] Texas Instruments. ADS1299-x Low-Noise, 4-, 6-, 8-Channel, 24-Bit, Analog-to-Digital Converter for EEG and Biopotential Measurements. 2012 (cit. on p. 18). [19] Texas Instruments. User’s Guide EEG Front-End Performance Demonstration Kit. 2012 (cit. on pp. 18, 19). [20]Serj1032. STM32 ADS1298 codes. Apr. 2017 (cit. on p. 23). [21]Christopher Watford. Understanding ECG Filtering. Mar. 2014 (cit. on p. 26). [22]Wikipedia. Pan–Tompkins algorithm. 2021 (cit. on p. 28). [23] Asadur Rahman. MATLAB based Functions for ECG Signal Preprocessing. Oct. 2018 (cit. on p. 29). [24] Sanjeev Nara, Manvinder Kaur, and Kundan Lal Verma. „Novel Notch Detection Algorithm for Detection of Dicrotic Notch in PPG Signals“. In: International Journal of Computer Applications 86 (17 2014) (cit. on p. 31). [25] Guillermo Cid Espinosa. Programación de interfaz gráfica en App Designer para el control vectorial de motores de imanes permanentes. Universidad Politécnica de Madrid, July 2018 (cit. on p. 37). Bibliography 51 A.1.2 Generic device 58 Chapter A Additional Figures A.1 Schematics and PCBs 59 60 Chapter A Additional Figures Fig. A.3.: PCB board of the generic system. A.2 Final program Menu Fig. A.4.: Main menu interface. A.2 Final program 61 ECG menu Fig. A.5.: Tachogram tab. (a) Limb leads. (b) Precordial leads. (c) Generic subset of leads. Fig. A.6.: ECG plots. 62 Chapter A Additional Figures (a) Parameters 12 ECG leads. (b) Parameters subset. Fig. A.7.: Parameter tabs. Pressure signals menu (a) Plots. (b) Parameters (the same for Tonometer 1 and 2). Fig. A.8.: Pressure signal representation and parameters. A.2 Final program 63 Fig. A.9.: Time intervals tab. ECG Reconstruction menu Fig. A.10.: ECG Reconstruction tab. 64 Chapter A Additional Figures A.3 Measurements A.3.1 ECG measurements Fig. A.11.: ECG Limb Leads. Fig. A.12.: ECG Precordial Leads. A.3 Measurements 65 Fig. A.13.: ECG Subset with not-standard lead. Fig. A.14.: 12 ECG Lead parameters. 66 Chapter A Additional Figures Fig. A.15.: ECG Subset paramaters. A.3.2 Pressure signal measurements Fig. A.16.: Pressure signals and ECG. A.3 Measurements 67 HAL_Delay(10); ADS_RREGS(0,17); HAL_Delay(1000); ADS_START(); HAL_Delay(100); } B.2 Matlab codes B.2.1 ECG Pan Tompkins algorithm function [signal,tachogram,Indexes,Frr]=PT_ECG(InECG,Fs) x=detrend(InECG); M=length(x); intrange=round(5*Fs/20); der=zeros(1,M); Pulse=zeros(1,M); timevec=linspace(0,M/Fs,M); hb=5;%Number of heartbeats to average ldelay=.13;%aprox Delay of the negtive slope with respect to the r wave udelay=.05;%aprox Delay of the positive slope with respect to the r wave confi=.3;%Maximum variation from one hearbeat with respect to the others Frr=8;%Sampling frequency of the tachogram Wn=45/Fs; NN=3;% order of 3 less processing [a,b] = butter(NN,Wn,'low'); %bandpass filtering y = filtfilt(a,b,x); y=y/max(abs(y)); Wn = .5/Fs; NN=3;% order of 3 less processing [a,b] = butter(NN,Wn,'high'); %bandpass filtering fECG = filtfilt(a,b,y); fECG = fECG/max(abs(fECG)); for i=5:M der(i)=(2*fECG(i)+fECG(i-1)-fECG(i-3)-2*fECG(i-4))/8;%derivative operator end 74 Chapter B Codes derr=detrend(der).*detrend(der);%squaring operator for n=intrange:M Pulse(n)=sum(derr(n-intrange+1:n))/intrange;%Integrator end %setting the treshold for detecting the rising and falling slopes treshold=mean(ecg); QRSStart=[]; QRSEnd=[]; for i=1:length(ecg)-1 %detect if the signal crosses the slope and save its index if Pulse(i)<= treshold && Pulse(i+1)>treshold QRSStart = [QRSStart i]; end if Pulse(i)<= treshold && Pulse(i+1)>treshold QRSEnd = [QRSEnd i]; end end Indexes=zeros(1,length(QRSStart)); for i=1:length(QRSStart) %search for the maximum point arround the tresholde crossing points [~,ind]=max(fECG(QRSStart(i)-round(ldelay*Fs):QRSStart(i)+round(udelay*Fs))); Indexes(i)= ind+QRSStart(i)-round(ldelay*Fs)-1; end %define the interval vector rr=[Indexes(2)-Indexes(1)]; truetime=[rr(1)]; i=2; Indexes=Indexes'; while i<length(Indexes) if i>=hb+1 bb=hb; else bb=i-1; end %define the tresholds avgbeat=sum(rr(end-bb+1:end))/bb; uppbound =(1+confi)*avgbeat; lowbound =(1-confi)*avgbeat; %check the RR intervals B.2 Matlab codes 75 [rr,Indexes,i]=tach_check(uppbound,lowbound,rr,Indexes,fECG,Fs,i); end %interpolate the tachogram for evenly spaced data MM=round(Frr*truetime(end)/(2*Fs))*2; truetime2=linspace(truetime(1),truetime(end),MM); tacho = interp1(truetime,rr,truetime2, 'pchip'); %store the signal and the tachogram signal=[timevec',fECG]; tachogram=[truetime2'/Fs,tacho'/Fs]; return B.2.2 Filtering function [fSig2]=Filter_BPF(InSig,Fs,Fh,Fl,order) M=length(InSig); tv=linspace(0,M/Fs,M); x=InSig; % x=detrend(x); %%Interference filtering Wn = (Fl*2)/Fs; NN = order; % order of 3 less processing [a,b] = butter(NN,Wn,'low'); %lowpass filtering y = filtfilt(a,b,x); fSig = y; Wn = (Fh*2)/Fs; NN = order; % order of 3 less processing [a,b] = butter(NN,Wn,'high'); %highpass filtering fSig2 = filtfilt(a,b,fSig); fSig2=[tv',fSig2]; 76 Chapter B Codes B.2.3 PQRST detection function out = P_QRS_Tdetect(y, Fs, peak_threshold) Q_th = round((20/360)*Fs); Q_th_1st = round((60/360)*Fs); Q_th_2nd = round((5/360)*Fs); P_th_1st = round((40/360)*Fs); P_th_2nd = round((1/360)*Fs); S_th_1st = round((5/360)*Fs); S_th_2nd = round((60/360)*Fs); T_th_1st = round((40/360)*Fs); T_th_2nd = round((120/360)*Fs); % ------------------QRS detection by FS1 and FS2 Algorithm ---------% y0 = []; y1 = []; y2 = []; y3 = []; y0 = diff(y); for i=2:length(y)-2 y1(i) = (y0(i-1)+ 2*y0(i)+y0(i+1))/4; end y1(1)=0; y2 = diff(y0); for i=1:length(y)-2 y3(i) = y1(i)+y2(i); end th = peak_threshold*max(y3); QRS = []; for i=2:length(y3)-6 if (y3(i)>th) && y3(i)>y3(i+1)&& y3(i)>y3(i-1) QRS = [QRS i]; end end QRS_Actual = []; for i=2:length(QRS)-1 if QRS(i+1)-QRS(i)>(Fs*0.6) QRS_Actual = [ QRS_Actual QRS(i) ]; end end % -------------------------------------------------------------% % ---------- R peak detection in original signal ------------------% R_L = []; B.2 Matlab codes 77 for i=2:length(QRS_Actual)-1 a = QRS_Actual(i)-Q_th: QRS_Actual(i)+S_th_2nd; m = max(y(a)); b=find(y(a)== m); if numel(b) ~= 0 b = b(1); b = a(b); R_L = [R_L b]; end end % -------------------------------------------------------------------% % ----------- Q peak detection signal -------------------------------% Q_L = []; for i=1:length(R_L) a = R_L(i)-P_th_1st: R_L(i)-P_th_2nd; m = min(y(a)); b=find(y(a)== m); if numel(b) ~= 0 b = b(1); b = a(b); Q_L = [Q_L b]; end end % ------------------------------------------------------------------ % % -------------- S peak detection -----------------------------------% S_L = []; for i=1:length(R_L) x = R_L(i)+S_th_1st:R_L(i)+S_th_2nd; m = min(y(x)); p=find(y(x) == m); if numel(p) ~= 0 p = p(1); p = x(p); S_L = [S_L p]; end end % -----------------------------------------------------------------% % ------------- P Peak detection ----------------------------------------% P_L = []; for i=1:length(Q_L) % He cambiado de coger a partir del 2do indice porque sino daban indices negativos a = Q_L(i)-Q_th_1st: Q_L(i)-Q_th_2nd; if a(1)<0 78 Chapter B Codes b=2; P_L = [P_L b]; else m = max(y(a)); b=find(y(a)==m); b = b(1); b = a(b); P_L = [P_L b]; end end % ----------------------------------------------------------------------- % % ----------------- T peak detection ------------------------------------% T_L = []; for i=1:length(S_L) a = S_L(i)+T_th_1st: S_L(i)+T_th_2nd; m = max(y(a)); b=find(y(a)==m); b = b(1); b = a(b); T_L = [T_L b]; end %-----------------------------------------------------------------------% % -----------------------------------------------------------------% T=1/Fs;L=length(y); t = (0:L-1)*T; % Suponemos que este codigo pierde una muestra P_Lt = (P_L-1)*T; Q_Lt = (Q_L-1)*T; R_Lt = (R_L-1)*T; S_Lt = (S_L-1)*T; T_Lt = (T_L-1)*T; out = [P_Lt; Q_Lt; R_Lt; S_Lt; T_Lt]; % figure % plot(t,y); % hold on; % grid on; % plot(P_Lt, y(P_L), 'rs','MarkerFaceColor','b'); % plot(Q_Lt, y(Q_L), 'ro','MarkerFaceColor','g'); % plot(R_Lt, y(R_L), 'rv','MarkerFaceColor','r'); % plot(S_Lt, y(S_L), 'ro','MarkerFaceColor','b'); % plot(T_Lt, y(T_L), 'rs','MarkerFaceColor','g'); % % plot(S_L, y(S_L), 'rs','MarkerFaceColor','b'); % title('QRS Complex, P & T waves'); B.2 Matlab codes 79 % legend('ECG Signal','P peak','Q peak','R peak','S peak','T peak'); % %axis([1000 29500 -1 1.0]); % %disp(out) end B.2.4 Pulse wave signal detection function [times, amps] = pulsedetect(data, Fs) T=1/Fs;L=length(data); t=(1:L)*T; der = diff(data); der2 = diff(der); window = round(0.5*Fs); window2 = round(0.2*Fs); % Max slope [pks,locsslope] = findpeaks(der,'MinPeakDistance',450); tiempo3 = locsslope*T; amp3 = data(locsslope); % Max peak detection locsmax = []; for i=2:length(locsslope)-1 a = locsslope(i):locsslope(i)+window2; m = max(data(a)); b=find(data(a)==m); if numel(b) ~= 0 b = b(1); b = a(b); locsmax = [locsmax b]; end end tiempo = locsmax*T; amp = data(locsmax); % Min peak detection locsmin = []; for i=2:length(locsslope)-1 80 Chapter B Codes a = locsslope(i)-window2:locsslope(i); m = min(data(a)); b=find(data(a)==m); if numel(b) ~= 0 b = b(1); b = a(b); locsmin = [locsmin b]; end end tiempo2 = locsmin*T; amp2 = data(locsmin); % Punto inflexion (notch) locsnotch = []; for i=2:length(locsslope)-1 a1 = locsslope(i):locsslope(i)+window; a1 = a1(2:round(0.9*length(a1))); m = min(der(a1)); c=find(der(a1)==m); c = a1(c); a2 = c:locsslope(i)+window; a2 = a2(1:round(0.4*length(a2))); % a2 = a2(1:round(0.3*length(a2))); %[pksa,locsa] = findpeaks(derPPG(a2)); m2 = max(der(a2)); %b = locsa(1); b=find(der(a2)==m2); if numel(b) ~= 0 b = b(1); b = a2(b); locsnotch = [locsnotch b]; end end tiempo4 = locsnotch*T; amp4 = data(locsnotch); lmin = min([length(tiempo),length(tiempo2),length(tiempo3),length(tiempo4)]); tiempo1 = tiempo(2:lmin);tiempo2 = tiempo2(2:lmin);tiempo3 = tiempo3(3:lmin+1);tiempo4 = tiempo4(2:lmin); times = [tiempo1'tiempo2'tiempo3 tiempo4']; amp1 = amp(2:lmin);amp2 = amp2(2:lmin);amp3 = amp3(3:lmin+1);amp4 = amp4(2:lmin); B.2 Matlab codes 81 amps = [amp1 amp2 amp3 amp4]; locsmax = locsmax(2:lmin);locsmin = locsmin(2:lmin);locsslope = locsslope(3:lmin+1);locsnotch = locsnotch(2:lmin); figure() plot(t,data); hold on; grid on; plot(tiempo1, data(locsmax), 'rv','MarkerFaceColor','r'); plot(tiempo2, data(locsmin), 'rs','MarkerFaceColor','b'); plot(tiempo3, data(locsslope), 'ro','MarkerFaceColor','g'); plot(tiempo4, data(locsnotch), 'ro','MarkerFaceColor','r'); title('PPG peak detection'); legend('PPG Signal','Max peak','Min peak','Max slope','Notch'); end 82 Chapter B Codes