Validation of a gonio-hyperspectral imaging system based on light-emitting diodes for the spectral and colorimetric analysis of automotive coatings
Abstract
In this study, a novel gonio-hyperspectral imaging system based on light-emitting diodes for the analysis of automotive coatings was validated colorimetrically and spectrally from 368 to 1309 nm. A total of 30 pearlescent, 30 metallic, and 30 solid real automotive coatings were evaluated with this system, the BYK-mac and X-Rite MA98 gonio-spectrophotometers, and also with the SPECTRO 320 spectrometer for further comparison. The results showed very precise correlations, especially in the visible range. In conclusion, this new system provides a deeper assessment of goniochromatic pigments than current approaches due to the expansion of the spectral range to the infrared.
Full text
1Validation of a gonio-hyperspectral imaging 2system based on light-emitting diodes for the 3spectral and colorimetric analysis of automotive 4coatings 5FRANCISCO J. BURGOS-FERNÁNDEZ,1,*MERITXELL VILASECA,1ESTHER PERALES,2ELÍSABET CHORRO,2 6FRANCISCO M. MARTÍNEZ-VERDÚ,2JOSÉ FERNÁNDEZ-DORADO,1AND JAUME PUJOL1 71Centre for Sensors, Instruments and Systems Development, Polytechnic University of Catalonia, Rambla Sant Nebridi 10, Terrassa 08222, Spain 82Department of Optics, Pharmacology and Anatomy, University of Alicante, San Vicente del Raspeig Road, San Vicente del Raspeig 03690, Spain 9*Corresponding author: francisco.javier.burgo[email protected] 10 Received 26 May 2017; revised 2 August 2017; accepted 7 August 2017; posted 7 August 2017 (Doc. ID 296812); published 0 MONTH 0000 11 In this study, a novel gonio-hyperspectral imaging system based on light-emitting diodes for the analysis of 12 automotive coatings was validated colorimetrically and spectrally from 368 to 1309 nm. A total of 30 pearlescent, 13 30 metallic, and 30 solid real automotive coatings were evaluated with this system, the BYK-mac and X-Rite 14 MA98 gonio-spectrophotometers, and also with the SPECTRO 320 spectrometer for further comparison. 15 The results showed very precise correlations, especially in the visible range. In conclusion, this new system 16 provides a deeper assessment of goniochromatic pigments than current approaches due to the expansion of 17 the spectral range to the infrared. © 2017 Optical Society of America OCIS codes: (010.1690) Color; (110.4234) Multispectral and hyperspectral imaging; (230.3670) Light-emitting diodes; (160.4760) 18 Optical properties. 19 https://doi.org/10.1364/AO.99.099999 20 1. INTRODUCTION 21 Goniochromatic pigments, also called effect or gonioapparent 22 pigments, are known by their appearance changes as a function 23 of the angles of illumination and observation [1–3]. These 24 changes can affect both color and texture, resulting in very 25 diverse visual impressions. Accordingly, they are classified as 26 pearlescent pigments, which mainly exhibit hue and chroma 27 shifts, and metallic pigments, which show lightness variations; 28 the traditional absorption pigments with a constant angular 29 appearance are designated as solids. These pigments have 30 acquired relevance in the industry during the last two decades, 31 specifically in the automotive, cosmetic, printing, and packag- 32 ing sectors [2,4]. 33 The angle-dependent coatings prompted the definition of a 34 new angular concept, the aspecular angle, which corresponds to 35 the angle between the specular reflection and the direction of 36 observation. Two additional concepts, the interference and 37 aspecular lines, were subsequently generated (Fig. 1). These 38 lines are used in the representation of the visual appearance 39 of goniochromatic samples in a color space. The aspecular angle 40 remains constant along the interference line, while the illumi- 41 nation angle is variable. However, in the aspecular line the 42 illumination angle remains invariable. The main function of 43 the interference line is to reveal abrupt shifts in hue and chroma 44 for different illumination angles (pearlescent pigments), in 45contrast to the aspecular line, where lightness shows the most 46 significant variation when moving far away from the specular 47reflection (metallic pigments). 48 The growing use of coatings based on goniochromatic 49 pigments called for the characterization at different illumina- 50 tion and observation geometries. Since traditional devices 51 included just one fixed configuration, the multi-angle or gonio- 52 spectrophotometers were created to provide more accurate 53 assessments of effect pigments through a larger variety of 54 measurement geometries [5]. At the same time, new standards 55 for the complete evaluation of these pigments were issued by 56 the German Institute for Normalization (Deutsches Institut für 57 Normung, DIN) and the American Society for Testing 58 Materials (ASTM) [6–8]. First, the guidelines for the analysis 59 of metallic pigments were detailed in DIN 6175-02 and ASTM 60 E2194. The measurement geometries described the fixed angle 61 of illumination for the correct assessment of lightness variations 62 along the observations angles. Since these two standards could 63 not accurately characterize pearlescent pigments, a specific 64 standard, ASTM E2539, was later issued. This new standard 65recommends two illumination angles since chroma and hue Research Article Vol. 56, No. 29 / /Applied Optics 1 1559-128X/17/290001-01 Journal © 2017 Optical Society of America
66 shifts of pearlescent pigments mostly appear for different direc- 67 tions of illumination. Table 1summarizes the measurement 68 geometries recommended by these three standards. 69 In addition, values were established for the aperture angle 70 (2σ) between the sample and the illumination and observation 71 directions, defined as the angle subtended by the sensor or the 72 light source with respect to the center of the sample. For ASTM 73 E2194 and ASTM E2539, the angle should be 2σ<8° and 74 equal for both directions, while DIN 6175-2 specifies the value 75 of these angles for each measurement geometry. Furthermore, 76 this standard differentiates between constant and variable direc- 77 tions, which can refer to the illumination or observation beams, 78 depending on which has the fixed element. Therefore, the 79 aperture angle for the constant direction must be 2σ≤5° 80 and for variable directions 2σ−20°≤4°, 2σ0≤4°, 2σ30 ≤10°, 81 and 2σ65 ≤10°. Another condition is the CIE 1964 standard 82 observer (10°) proposed by the Commission Internationale 83 del’Éclairage (CIE). The colorimetric features of goniochro- 84 matic pigments require a large field of view evaluation due 85 to the broad spatial distribution of pearlescent and metallic 86 particles over the sample. 87 In view of these developments, all large companies 88 involved in color measurement (Datacolor, Konica Minolta, 89 GretagMacbeth, X-Rite, BYK Additives & Instruments, 90 Hunterlab, etc.) have launched desktop or portable gonio- 91 spectrophotometers. The major driving force has been the 92 interest of the automotive industry in improving the visual 93 quality control of car finishes. Indeed, the automotive industry 94 is still the sector with a stronger demand for goniometric 95analysis. 96 Experimental gonio-spectrophotometers have also been 97 developed with diverse operating principles and providing 98 additional parameters: gonio-spectro-photometry based on 99 the spectral bidirectional reflectance distribution function 100 (sBRDF) [9,10], direct measure of the gloss at various 101 geometries [11,12], and versatile optomechanical designs 102 for measurements at several geometries [13]. Despite the 103 good performance of commercial and experimental gonio- 104 spectrophotometers, the data generally result from integrating 105 the information of a few square millimeters of the target object. 106 Consequently, they lack spatial resolution. For this reason, 107 these devices can be considered to not properly estimate per- 108 ceptual aspects of a surface as psychophysical experiments with 109 real observers. At this point, the combination in hyperspectral 110 imaging of colorimetric and textural assessments required for 111 the special visual attributes of these coatings was considered 112to be the best approach to overcome these constraints. 113 The combination of gonio-spectrophotometers and spectral 114 imaging systems has generated a new kind of instruments 115 known as gonio-spectral imaging systems or multi-angle 116 spectral imaging systems, with the main objective to obtain 117 spectral images at different geometries. The first approaches 118 to gonio-spectral imagery focused on the analysis of three- 119 dimensional pieces of artwork [14–16], and later extended 120 to two-dimensional objects such as paper and cloth samples 121 [17,18]. The illumination system consisted of a wide spectrum 122 light source and sets of interference filters resulting in a large 123 and mechanically complex structure. However, it is now pos- 124 sible to obtain the spectral sampling directly from the light 125 source without added elements by means of light-emitting di- 126 odes (LEDs). The advantages of LEDs are widely known: they 127 are very efficient, present a long life cycle, evolve constantly, are 128 comparatively economical, and have a small size. In addition, 129 these diodes have a narrow spectral emission, and they are avail- 130 able at several wavelength peaks along different spectral ranges 131 (ultraviolet, UV; visible, VIS; and infrared, IR). Another rel- 132 evant advantage of LED technology is the modulation of their 133emission as a function of the forward current. 134 In the existing literature, goniochromatic pigments have 135 only been evaluated in the VIS range of the electromagnetic 136 spectrum, since it is the region where the main appearance 137 shifts occur [19–21]. However, an analysis of the angular 138 dependence at deeper layers through IR illumination would 139 further elucidate the behavior of these pigments. Furthermore, 140 ultraviolet lighting can contribute to a better insight into 141 possible fluorescence. Thus, an extended spectral inspection 142 might provide valuable information on the performance of these 143 coatings in a broad range of bands of the electromagnetic 144spectrum. 145 This paper presents the validation of GOHYLED, a new 146 gonio-hyperspectral imaging system based on LEDs, developed 147 for the spectral and colorimetric evaluation of automotive 148 coatings. This initial verification will establish the basis for 149 future work on spatially resolved assessments. GOHYLED is 150 specifically intended to work in an extended range from the 151UV to IR by using commercial, high power LEDs, in contrast F1:1 Fig. 1. Interference and aspecular lines: (a) Cand Ldiagram F1:2 showing the behavior along the different measurement geometries. F1:3 (b) Measurement geometries for the interference (top) and aspecular F1:4 lines (bottom). Table 1. Measurement Geometries of DIN 6175-02, ASTM E2194, and ASTM E2539 Standards ([Angle of Illumination with Respect to the Normal] x: [Angle of Observation with Respect to the Normal]) T1:1 Standards Measurement Geometries T1:2 DIN 6175-02, ASTMs 45°×: -20°45°×:0°45°×:30°45°×:65° T1:3 ASTM E2194, E2539 45°×: -30° T1:4 ASTM E2539 45°×: -60°15°×:0°15°×: -30° 2 Vol. 56, No. 29 / / Applied Optics Research Article
152 with conventional gonio-spectral systems that only use wide 153 spectrum light sources that exclusively cover the VIS range. 154 2. EXPERIMENTAL SETUP 155 The experimental setup was based on the requirements for the 156 study of automotive coatings in a wide spectral and angular 157 range specified by the DIN 6175-2, ASTM E2194-09, and 158 ASTM E2539-08 standards [6–8]. However, it might be useful 159 for other applications and configurations. Essentially, the 160 GOHYLED is composed of a spectral light source, two cameras 161 (UV-VIS and IR) that cover the 368–1309 nm range, and two 162 rotation stages with a total rotation angle of 180° in both 163 observation and illumination directions (Fig. 2). All elements 164 were controlled through a purpose-built graphical user interface 165 developed with MATLAB R2013a, which allows the individual 166 and synchronized use of the light source, cameras, and motor- 167 ized rotation stages. 168 A. Light Source 169 The light source consists of 28 high power surface-mounted- 170 device (SMD) LED clusters of different peak wavelengths that 171 spectrally illuminate the sample and a linear actuator that 172 sequentially positions each cluster in front of it. A white 173 LED cluster was also incorporated for the subsequent spatial 174 evaluation of texture, which is not included in this work. 175 The clusters are composed of three or six LEDs, depending 176 on the radiance emitted. A lens to increase the amount of light 177 over the sample was also incorporated. Figure 3shows the 178 normalized spectral radiance, and Table 2contains the part 179number and peak wavelengths of the whole set of LEDs. 180 The LEDs were purchased from Roithner LaserTechnik 181 GmbH (Part numbers RLCU and SMB1N) and Lumileds 182 Holding B.V. (Part numbers LX). The UV-VIS LEDs 183 (368–774 nm) presented a mean spectral step between con- 184 secutive peak wavelengths of around 20 nm, whereas in the 185 IR (807–1309 nm), this value increased to 40 nm. The 186 FWHM in the UV-VIS range varies from 10 to 37 nm and 187 from 30 to 80 nm in the IR. The higher FWHM and the sub- 188 sequent overlapping of the spectra in the IR range partially 189 compensated the larger distance among peak wavelengths. 190 Even though it causes a coarser sampling, the broader 191 FWHM is a typical feature of current IR solid-state technology 192 that is still unsolved. Moreover, IR LEDs present less variety of 193 wavelengths, especially those of high power—thus the signifi- 194 cant gap before the spectrum with a peak wavelength of 1951309 nm in Fig. 3. F2:1 Fig. 2. Developed gonio-hyperspectral imaging system based on F2:2 LEDs represented in (a) a layout and in (b) a picture: 1, UV-VIS cam- F2:3 era; 2, sample holder; 3, 8MR191-30 rotation stage; 4, 8MR191-30- F2:4 28 rotation stage; 5, rotation stages controller; 6, linear actuator; 7, F2:5 lens; 8, LED clusters; 9, light source controller and power supply; F2:6 10, IR camera, which is not represented in (a). θrefers to the illumi- F2:7 nation angle controlled by the 8MR151-30 rotation stage and αto the F2:8 observation angle handled by the 8MR191-30-28 rotation stage. F3:1Fig. 3. Normalized spectral radiance of the LEDs included in the F3:2GOHYLED system. Table 2. Part Numbers and Peak Wavelengths (λp) of the 28 Types of LEDs T2:1Part Number λp[nm] Part Number λp[nm] T2:2RLCU-440-365 368 SMB1N-700 709 T2:3RLCU-440-390 397 RLCU-440-720 724 T2:4RLCU-440-410 411 SMB1N-735 735 T2:5LXML-PR01-0500 446 SMB1N-750 748 T2:6LXML-PB01-0040 476 SMB1N-770 774 T2:7SMB1N-490 489 SMB1N-810D 807 T2:8LXML-PE01-0070 498 SMB1N-850D 867 T2:9LXML-PM01-0100 528 SMB1N-870 885 T2:10LXML-PL01-0060 601 SMB1N-940D 952 T2:11LXM2-PH01-0070 622 RLCU-440-970 984 T2:12LXM2-PD01-0050 634 RLCU-440-1020 1033 T2:13LXM3-PD01 663 RLCU-440-1050 1052 T2:14SMB1N-670D 678 SMB1N-1300 1309 T2:15SMB1N-690D 693 LXW8-PW40 449/599a aPeak wavelengths of the two peaks for the white LEDs. Research Article Vol. 56, No. 29 / / Applied Optics 3
196 The use of high power LEDs was driven by the aperture 197 angles recommended by the standards. To achieve such small 198 angles, the illuminating and sensing areas should have been 199 extremely reduced, limiting the evaluation to a minute fraction 200 of the samples. However, one of the objectives of this work was 201 to evaluate bigger areas than current devices and to obtain spa- 202 tially resolved images. To this end, the LEDs were placed at a 203 distance of 580 mm in order to leave enough free space for the 204 rotation of the acquisition arm around the sample. But the long 205 distance and the angle of emission of the LEDs (120°) resulted 206 in a low illumination of the sample. To solve this issue, a plano- 207 convex lens from Edmund Optics (stock #67-228) with a focal 208 length of 75 mm, a diameter of 50 mm, and a transmittance of 209 90% from 200 to 1300 nm was selected. The relative position 210 between the LEDs and the lens was out of focus to avoid an 211 image of the LED chips to form on the sample. The lens back 212 surface was finally placed 25 mm from the LEDs; this was 213 experimentally determined and ensured a uniform illumination 214 over the sample. This configuration of the light source pro- 215 duced an aperture angle of 4.69°, which was below the maxi- 216 mum tolerance of the ASTM E2194 and E2539 (8°) and DIN 217 6175-2 (5°) standards, considering for the latter the illumina- 218 tion beam as the constant direction. 219 The linear actuator ZLW-0630-02-B-60-L-1000 from 220 Igus S.L., and a stepper motor SY42STH47-1206A from 221 Changzhou Songyang Machinery & Electronics Co., Ltd., 222 executed the sequential positioning of the LED clusters, result- 223 ing in a minimum step of 140 μm. Each LED cluster was 224 switched on only when it was placed in front of the sample 225 and behind the lens. The images were then acquired by the 226 camera, the LEDs were switched off, and the next cluster 227 was placed in the same position to repeat the process. With 228 this method, only one LED cluster was switched on at a time. 229 B. Imaging Sensors 230 Table 3contains the technical features of the two cameras used. 231 The acquisition unit included a CM-140 GE-UV monochro- 232 matic CCD camera from JAI AS, with enhanced UV sensitivity, 233 and a GMUV42528C 2.8/25 mm quartz lens from GOYO 234 OPTICAL Inc. selected specifically for its high transmittance 235 from 200 to 1300 nm. The InGaAs camera was a Hamamatsu 236 C10633-23 with a lower spatial resolution and exposure time 237 and no available gain factor. A Kowa LM12HC-SW 1.4/ 238 12.5 mm SWIR lens with high transmission from 800 to 239 2000 nm was coupled to this camera. The spatial resolution 240 of the IR camera did not pose a problem since the validation 241 of the proposed system was performed by means of averaged 242 reflectances of the images. However, it can limit future spatial 243 evaluations. 244 The acquisition was made in a two-step process. The 245 UV-VIS camera was employed to acquire images when using 246 the LED clusters with peak wavelengths from 368 to 952 nm, 247 and the IR camera for peak wavelengths from 984 to 1309 nm. 248 The images for the whole set of measurement geometries were 249first captured with the UV-VIS and second with the IR camera. 250 The aperture angles of the observation stage also conformed 251 to the standards. The minimum distance between the sample 252 and the cameras was based on the maximum diameter of the 253 entrance pupil. Since it can be obtained dividing the focal 254 length by the minimum F-number, the lenses of both cameras 255 presented the same value of 8.93 mm. However, the distance to 256 the sample was different for each camera because the position of 257 the entrance pupil was different for each lens, at 14.1 mm for 258 the UV-VIS lens and at 10.7 mm for the IR lens, both inward. 259 Consequently, the minimum distances between the sample and 260 the outer part of the cameras’lenses were 113.8 mm for the 261 UV-VIS and 117.1 mm for the IR. Finally, the UV-VIS camera 262 was positioned at 218 mm from the sample to increase the area 263 observed; this resulted in one aperture angle of 2.20°. The IR 264 camera was located at 129 mm from the sample to avoid any 265 collision with the sample holder; this distance variation gener- 266 ates an aperture angle of 3.66°. A longer distance was not used 267 for this camera because of its wider field of view and lower spa- 268 tial resolution. Both cameras conformed to the tolerance of the 269 ASTM E2194 and E2539 (8°) and DIN 6175-2 (4°) standards, 270 considering for the latter the observation beam as the variable 271 direction. Although these standards only regulate the measure- 272 ment of goniochromatic pigments in the VIS range, the same 273 conditions were applied to the IR due to the lack of regulations 274in this range. 275 To capture images within the dynamic range of the 276 cameras, exposure time and gain varied throughout the mea- 277 surement geometries and for the different acquisition channels. 278 Besides the emission power of each LED cluster, the spectral 279 sensitivity of the cameras and the angular distance between 280 the illumination and observation direction also conditioned 281these parameters. 282C. Motorized Rotation Stages 283 The changeable angular positioning was obtained using two 284 motorized rotation stages manufactured by STANDA Ltd, 285 the 8MR151-30 and the 8MR191-30-28. Both have a rotation 286 range of 360°, an angular resolution of 0.01°, and a central 287 aperture of 30 mm. At the top, the 8MR151-30 stage controls 288 the illumination angle by rotating the sample, while at the bot- 289 tom, the 8MR191-30-28 controls the observation direction by 290 positioning the cameras around the sample. Figure 2shows a 291graphical description of the angular positioning of each stage. 2923. METHODS 293A. Samples 294 A total of 30 pearlescent, 30 metallic, and 30 solid samples were 295 selected; they presented a rectangular size between 90 mm × 296 150 mm and 100 mm ×200 mm. The 90 samples originated 297 from AUDI AG, BASF SE, and PPG Industries Inc. commer- 298 cial car paints. In addition to the measurements carried out 299 with the GOHYLED system, these samples were measured 300 with the BYK-mac (BYK-Gardner GmbH) and X-Rite 301MA98 (X-Rite Inc.) gonio-spectrophotometers and also the Table 3. Technical Features of the Cameras T3:1 Camera UV-VIS IR T3:2 Spectral range [nm] 200–1000 900–1650 T3:3 Sensor size [px] 1392 ×1040 320 ×256 T3:4 Bit depth 10 14 4 Vol. 56, No. 29 / / Applied Optics Research Article
302 SPECTRO 320 spectrometer (Instruments Systems GmbH) 303 for further comparisons. 304 B. Multi-Angle Color Measurement 305 The GOHYLED system was designed and built considering 306 the measurement geometries, aperture angles, and CIE stan- 307 dard observer recommended by the DIN 6175-2, ASTM 308 E2194-09, and ASTM E2539-08 standards (Table 4). The 309 aperture angle for the UV-VIS camera was eventually smaller 310 because the diaphragm size of the camera lens was diminished 311 (f∕11) to enhance the focus at the 368–952 nm spectral range. 312 In contrast, the IR camera used a greater diaphragm (f∕2.8)to 313 ensure a good signal-to-noise ratio, since IR LEDs had lower 314 power than VIS LEDs. Furthermore, the focus was less critical 315 than in the UV-VIS region because the IR camera operated 316 with LEDs in a range of only 300 nm. The measurement area 317 or region of interest (ROI) was determined for the UV-VIS 318 camera because its field of view was more limited. 319 The major benefit of conforming to ASTM and DIN spec- 320 ifications is the possibility of performing a precise comparison 321 of readings among instruments of different optical design. The 322 measurements obtained with the GOHYLED system were 323 compared to those of the BYK-mac and X-Rite MA98 gonio- 324 spectrophotometers. These devices are calibrated for all the 325 geometries against ceramic standards of the British Ceramic 326 Research Association, known as BCRA tiles. The BYK-mac 327 conforms only to the DIN 6175-2 and ASTM E2194 stan- 328 dards, whereas the X-Rite MA98 and the GOHYLED conform 329 to the three standards (DIN 6175-2, ASTM E2194, and 330 ASTM E2539). With a fixed illumination direction, the 331 BYK-mac is particularly suitable for the evaluation of metallic 332 pigments, but it cannot characterize the changes of hue and 333 chroma that appear in pearlescent pigments for different illu- 334 mination directions. Similarly to the GOHYLED, the BYK- 335 mac performs the spectral sampling by means of LEDs and 336 can evaluate sparkle and graininess, both textural effects of 337 goniochromatic pigments. On the other hand, the X-Rite 338 MA98 can produce complete assessments of pearlescent and 339 metallic pigments since it includes the 45° and 15° illumination 340 angles required by the standards. It also includes other obser- 341 vation angles outside the conventional plane of illumination/ 342 observation, and a diffraction grating performs the spectral 343 sampling. Besides the standard, the BYK-mac and X-Rite 344 MA98 also perform evaluations at the 45°×: -60° measure- 345 ment geometry. Therefore, the spectral and colorimetric results 346 obtained with the GOHYLED system were fully comparable to 347 those of the X-Rite MA98, whereas only the geometries sug- 348 gested by the DIN 6175-2 and ASTM E2194 standards, 349 45°×: -30°, 45°×: -20°, 45°×:0°, 45°×:30°, and 45°×:65°, 350 plus 45°×: -60°, were used to compare with the results of the 351BYK-mac. 352C. Measurement Stability and Repeatability 353 The stability and repeatability of the measurement procedure 354 were also evaluated through the acquisition of images of a white 355 reference BN-R98-SQ10C from Gigahertz Optik GmbH at 356 the geometry of 45°×:0° with both cameras and the quantifi- 357cation of the mean digital level variation. 358 First, the stability of all the spectral channels was examined 359 by switching every LED cluster on and off 10 times and each 360 time capturing 10 images, which were subsequently averaged. 361 The purpose of these tests was to accurately reproduce the final 362 measurement mechanism, in which the LED clusters were used 363 as flashes of light. Next, the influence of the sample positioning 364 on the holder was evaluated. This study encompassed the com- 365 parison of measurements with and without the replacement of 366the sample among repetitions. 367 In order to reduce other sources of bias such as temporal 368 noise, sets of three, five, and 10 averaged frames were tested 369 for the two cameras at the geometries of 45°×: -60°, 370 45°×: -30°, 45°×: -20°, 45°×:0°, 45°×:30°, and 45°×:65°, 371using the white reference mentioned before. 372 The proposed system also dealt with spatial noise removal, 373 in this case by applying Eq. (1) to each acquired image: ICi;jkIi;j−IBi;j IUi;j−IBi; j;(1) 374 where ICi;j,Ii;j,IBi; j, and IUi;jcorrespond to the 375 digital levels of the corrected, raw, background or dark, and 376 uniform field or white reference images, respectively; kis 377 the calibrated reflectance of the reference provided by the 378 manufacturer. Ii;j,IBi;j, and IUi;jimages were always 379 acquired in the same conditions of exposure time and gain, 380 which change for each acquisition channel and measurement 381 geometry. Images of the calibrated white reference were cap- 382 tured and used as uniform field targets, as well as dark or back- 383 ground images with all LEDs turned off. The results derived 384from Eq. (1) were directly expressed as reflectance images. 385 The evaluation of the positioning repeatability of the rota- 386 tion stages was carried out by means of the UV-VIS camera 387 attached to the 8MR191-30-28 stage through the acquisition 388 arm at the distance previously calculated (218 mm from the 389 sample). A graph paper divided into 1 mm squares was used 390 as the sample to quantify the deviation of the image recorded. 391 Positioning tests were performed for rotations of 30°, 10°, and 392 5°, first with one of the two elements fixed and afterwards mov- 393ing both; each rotation displacement was assessed three times. 394D. Extraction of Reflectance Spectra and 395Colorimetric Data 396 The spectral and color appearance data were retrieved by means 397 of a purpose-developed image processing program using 398 MATLAB R2013a. Its main functions were to calculate the 399 reflectance spectra, spectral fitting metrics, chromatic coordi- 400nates, and color differences. 401 Reflectance images were obtained by means of Eq. (1). 402 Throughout the whole spectral range of the GOHYLED sys- 403tem (368–1309 nm), the average reflectance values were also Table 4. F-Number, Aperture Angles (2σ), Spatial Resolutions (SR), and Regions of Interest (ROI) for Each Camera T4:1 Camera F-Number 2σ[°] SR [px/mm] ROI [mm] T4:2 UV-VIS f∕11 0.56 28 50 ×37 T4:3 IR f∕2.8 1.83 3 50 ×37 Research Article Vol. 56, No. 29 / / Applied Optics 5
404 determined for three ROIs: the ROI detailed in Table 4and the 405 ROIs of the BYK-mac and the X-Rite MA98 gonio-spectro- 406 photometers with diameters of 23 and 12 mm, respectively. 407 Although the ROIs of the GOHYLED system are square or 408 rectangular and those of the two gonio-spectrophotometers 409 are circular, the differences were considered negligible. These 410 spectra were finally interpolated from 400 to 700 nm with a step 411 of 10 nm by applying a cubic spline algorithm [22,23], on 412 account of the input spectral range needed for calculating the 413 colorimetric values and color differences with respect to those 414 of the BYK-mac and X-Rite MA98. Regarding the IR range, 415 the spectra were validated with the measurements carried out 416 with the SPECTRO 320 spectrometer and the telescopic optical 417 probe TOP 100 from InstrumentSystems GmbH and a halogen 418 lamp at the geometries of 45°×:30°, 45°×:0°, and 15°×:0°. 419 This device and all its accessories are calibrated against standards 420 of the National Institute of Standards and Technology (NIST) 421 and the Physikalisch-Technische Bundesanstalt (PTB). The 422 spectral range went from 700 to 1309 nm with a step of 423 10 nm. No additional geometries were assessed since these were 424 considered sufficiently representative and also due to the lack of 425 automation of the measurement process. 426 Additionally, the comparison of the spectra between the 427 GOHYLED system and the BYK-mac, the X-Rite MA98, 428 and the SPECTRO 320 was completed through the mean ab- 429 solute error (MAE), root mean square error (RMSE) [24,25], 430 and goodness-of-fit coefficient (GFC) [26,27]. These metrics 431 were chosen because they are commonly used when comparing 432 reconstructed spectral data [25,27,28]. Moreover, the combi- 433 nation of the MAE and RMSE is useful to detect whether 434 values are caused by generalized large errors (RMSE ∼MAE) 435 or by a few errors greater than MAEs, such as outliers 436 (RMSE >MAE); both parameters are in the same units as 437 the analyzed data. On the other hand, the GFC is a relative 438 measure between 0 and 1 based on Schwartz’s inequality, com- 439 monly used when comparing spectra and, especially, color 440 differences [Eq. (2)]. According to the scale commonly linked 441 to this parameter [27], a GFC ≥0.9950 is considered 442 “colorimetrically accurate,”aGFC ≥0.9990 a“good spectral 443 fitting,”and a GFC ≥0.9999 an “excellent spectral fitting”: GFC jPn iroλireλij ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi Pn iroλi2Pn ireλi2 p;(2) 444 where roλiis the original spectrum and reλiis the estimated 445 spectrum, both evaluated at the same wavelength λi;nis the 446 total amount of wavelengths considered. 447 The colorimetric evaluations were performed under the 448 CIELAB color space [29,30]. The CIE standard illuminant se- 449 lected was D65, similarly to BYK-mac and X-Rite MA98. 450 Color differences between these devices and the GOHYLED 451 were calculated with the CIEDE2000 [31,32] and AUDI2000 452 [33] color-difference formulas. The CIEDE2000 color- 453 difference formula is defined as ΔE00 ΔL0 kLSL2 ΔC0 kCSC2 ΔH0 kHSH2 RTΔC0 kCSCΔH0 kHSH1 2;(3) 454where ΔL0,ΔC0, and ΔH0are differences of lightness, chroma, 455 and hue between two colors, respectively. This formula 456 included parametric factors kL,kC, and kHfor weighting 457 lightness, chroma, and hue differences, respectively, according 458 to the application. These three parameters are equal to 1 459 under the reference conditions proposed by the CIE [34,35]. 460 In addition, three weighting functions, SL,SC, and SHfor each 461 lightness, chroma, and hue, were also included. Finally, the 462 rotation term RTwas added to improve the performance for 463 blue colors. The improved performance for objects that exhibit 464 lightness values higher than 100 was a decisive factor, since this 465 phenomenon is very common in goniochromatic pigments, 466especially under direct illumination conditions. 467On the other hand, the AUDI color-difference formula was 468 specially developed for the approval of goniochromatic paint 469batches as follows: dE0 γ dL γ kdLsdLγ2 dC γ kdCsdCγ2 dH γ kdHsdHγ21 2:(4) 470 The main difference with respect to the CIEDE2000 is that, 471in this case, the weighting functions (sdLγ,sdCγ,sdHγ) depend 472on the measurement geometry. 4734. RESULTS AND DISCUSSION 474The spectral reconstruction of the 90 samples comprised the 475 entire spectral range of the GOHYLED system, from 368 476to 1309 nm, with 28 acquisition channels, eight color measure- 477 ment geometries, and three ROIs. Figure 4shows two reflec- 478tance images of a pearlescent pigment at the geometries of 479 45°×: -30° and 45°×: -20° and for the LED cluster with a 480peak wavelength of 476 nm. The illumination gradient was 481 not removed when dividing by the image of the white reference 482 because it has a matte finish, whereas the automotive coatings 483 were glossy. At this stage, a glossy white reference was not used 484 to reach a more precise comparison with the BYK-mac and the 485X-Rite MA98, which also employ a matte white reference. 486 Figure 5exhibits the reflectance spectra, and Fig. 6the C 487versus Lof three samples where the pearlescent, metallic, and 488 solid effects were more noticeable. Measured samples showed 489the expected theoretical behavior that was also confirmed by 490 the commercial devices. Chroma shifts were more evident in 491pearlescent samples due to variations of the reflectance spec- 492 trum in relation to wavelength and to the interference and 493aspecular lines. Luminance changes in goniochromatic samples F4:1Fig. 4. Reflectance images of a pearlescent pigment at the geom- F4:2etries of (a) 45°×: -30° and (b) 45°×: -20° for the LED cluster with F4:3a peak wavelength of 476 nm. 6 Vol. 56, No. 29 / / Applied Optics Research Article
494 were recorded as variations in reflectance [Figs. 5(a) and 5(b)] 495 and lightness [Figs. 6(a) and 6(b)]. Although this phenomenon 496 is the main attribute of metallic samples, it was also observed in 497 pearlescent samples because pearlescent pigments employed in 498 automotive coatings are commonly mixed with metallic par- 499 ticles to integrate the two effects in one single paint. With 500 regard to solid samples, the spectral reflectance, chroma, and 501 luminance remained stable in the eight geometries with only 502 minor fluctuations [Figs. 5(c) and 6(c)] 503 Figures 5(a),5(b),6(a), and 6(b) show that reflectance and 504 lightness values, respectively, were higher at the geometries 505 adjacent to the specular reflection (45°×: -60°, 45°×: -30°, 506 15°×: -30°, and 15°×:0°) and decreased drastically in the case 507 of the furthest geometries. On the other hand, the most signifi- 508 cant changes in the spectral profile emerged when comparing 509 the measurements of geometries with the illuminant at differ- 510 ent positions: 45°×: -60° and 45°×: -30° versus 15°×: -30° 511 and 15°×:0°. These geometries are recommended for the 512 evaluation of color change due to pearlescent pigments [8]. 513 The following results are based on the average reflectance of 514 the images for a more precise comparison with the BYK-mac 515 and X-Rite MA98, since they do not offer pixel-wise reflectance 516 data. At this stage, the spatially resolved images were not 517 exploited because the main objective of this work was to 518 perform a preliminary spectral and colorimetric validation of 519 performance by means of comparing the new system against 520 two commercial gonio-spectrophotometers. The next stage will 521 focus on spatial features of goniochromatic pigments such as 522 sparkle, graininess, and mottling under broad-band and nar- 523 row-band light sources. The deeper penetration of IR light will 524 provide a more accurate analysis of the goniochromatic particles 525 located at deeper layers of the coating and will also contribute 526 to quantify their influence on the total appearance of the pig- 527 ment. Preliminary results have revealed a better discrimination 528 of some spatial attributes in the IR range due to the removal of 529 sparkle at this spectral region and for coatings of different 530thickness. 531A. Stability and Repeatability Performance 532 The results for stability assessments demonstrated excellent 533 performance of all spectral channels with differences among 534 repetitions below 5.75% for the UV-VIS camera and under 535 1.97% for the IR camera. The most outstanding shifts due 536 to the influence of the sample positioning appeared when 537 replacing it, but they were considered minor fluctuations 538because the deviation values never exceeded 4%. 539 Temporal noise analysis led to maximum differences 540 below 1% when comparing the mean digital levels of the three 541 sets of frames. This percentage was calculated with respect 542 to the maximum digital level of each camera: 1023 for the 543 UV-VIS and 16383 for the IR. The minimum number of 544frames (three) was selected for each acquisition. 545 The positioning repeatability evaluation of the rotation 546 stages revealed a maximum positioning deviation under 1%, 547which confirmed the accuracy of both stages. F5:1 Fig. 5. Reflectance spectra of (a) pearlescent, (b) metallic, and F5:2 (c) solid samples at 45°×: -60°, 45°×: -30°, 45°×: -20°, 45°×:0°, F5:3 45°×:30°, 45°×:65°, 15°×: -30°, and 15°×:0°. F6:1Fig. 6. Cand Ldiagrams of (a) pearlescent, (b) metallic, and F6:2(c) solid samples at 45°×: -60°, 45°×: -30°, 45°×: -20°, 45°×:0°, F6:345°×:30°, 45°×:65°, 15°×: -30°, and 15°×:0°. Research Article Vol. 56, No. 29 / / Applied Optics 7
548 B. Spectral Performance in the Visible Range 549 The first data evaluated were the spectral reconstructions from 550 400 to 700 nm by means of the MAE, RMSE, and GFC. We 551 should stress that the MAEs and the RMSEs are expressed as 552 reflectance differences. These three metrics were computed for 553 the three ROIs and the three pigment categories. The perfor- 554 mance among ROIs was very similar and validated the use 555 of the maximum measurement area without undermining 556 the fitting; therefore, only the results for the full ROI are 557 shown. Figure 7plots the mean values considering the different 558 geometries for the three types of samples; each bar includes the 559 standard error as a measure of uncertainty. 560 The precise reconstructions obtained with the developed 561 system revealed that all the mean values of MAE and 562 RMSE were below 0.03 and very similar, which means that 563 very few outliers were found. The mean GFCs, all above 564 0.9900, also proved the color accuracy. The GOHYLED sys- 565 tem and the BYK-mac exhibited slightly more similar behavior 566 than the X-Rite; in terms of MAE and RMSE this result was 567 similar to the result of comparing the two commercial devices. 568 With regard to the scale for the GFC, the evaluation of pearl- 569 escent and solid pigments can be considered “colorimetrically 570 accurate”since the mean values of this parameter were above 571 0.9950, while metallic pigments attained “good spectral fitting” 572with values slightly above 0.9990. These last subsets of pig- 573 ments also achieved correlations very similar to the correlations 574 between the two gonio-spectrophotometers. In agreement with 575 previous studies [36,37], the measurement geometries closer to 576 the specular reflection (45°×: -60°, 45°×: -30°, 15°×:30°, 577 and 15°×:0°) resulted in larger differences. At these geometries, 578 the amount of light reflected is very high and any small varia- 579tion in position or configuration can produce substantial 580deviations of the results. 581 Color differences again showed better correlations between 582 the GOHYLED system and the BYK-mac, although the devi- 583 ations with respect to the X-Rite MA98 were less noticeable 584 than in the fitting metrics (Fig. 8). The greatest spectral and 585 colorimetric match obtained for the GOHYLED system and 586 the BYK-mac was probably due to the use of the same spectral 587filtering technique, i.e., LEDs, with similar spectral features 588such as the peak wavelength and the FWHM. 589 In general, the mean CIEDE2000 values were below 2 units 590 with all instruments and for the three kinds of samples. These 591colorimetric differences remained within the ranges proposed 592 in a previous study [38], where the authors suggested to widen 593 color tolerances when using newer versions of standardized 594 color-difference formulas such as the CIEDE2000; this study 595 did not cover non-standard color-difference equations such as 596 the AUDI2000. This tolerance proposal was based on the large 597 numbers of factors that determine the reproduction of automo- 598 tive coatings: add-on parts, closeness to the specular angle, 599 L>100, and visual texture. Considering all these circumstan- 600 ces, they recommended to apply a multiplication factor 601 between 1.5 and 3 with respect to the visual discrimination 602threshold, traditionally equal to 1. 603 In contrast, the AUDI2000 color differences were larger, 604 particularly for solid samples. It is known that this color- 605 difference formula is stricter than the CIEDE2000, particularly 606for solid pigments, because their appearance is considered less 607 complex and easier to reproduce than for goniochromatic 608pigments. Since the solid samples employed in this study F7:1 Fig. 7. Mean values of (a) MAEs and RMSEs and (b) GFCs for the F7:2 spectral reconstruction of the GOHYLED system with respect to the F7:3 BYK-mac and the X-Rite MA98 and the comparison between two F7:4 gonio-spectrophotometers. The three metrics were plotted for pearl- F7:5 escent (P), metallic (M), and solid (S) samples. The error bars represent F7:6 the standard error. F8:1Fig. 8. Mean CIEDE2000 and AUDI2000 color differences F8:2of the GOHYLED system with respect to the BYK-mac and the F8:3X-Rite MA98 and also the comparison between the two gonio- F8:4spectrophotometers for the full ROI. Both metrics were plotted for F8:5pearlescent (P), metallic (M), and solid (S) samples. The error bars F8:6represent the standard error. 8 Vol. 56, No. 29 / / Applied Optics Research Article
609 showed surfaces with more gloss than the pearlescent and 610 metallic, any minor variation in the configuration could cause 611 considerable variations among instruments. With regard to 612 Fig. 8, the values shown were calculated considering the whole 613 set of measurement geometries. 614 The comparisons with the GOHYLED system were very 615 similar to the comparison between the two commercial 616 gonio-spectrophotometers, especially for pearlescent and met- 617 allic pigments in terms of MAE, RMSE, and CIEDE2000. 618 We can thus infer that the BYK-mac and X-Rite MA98 present 619 differences of a magnitude similar to the GOHYLED. 620 However, the correlation of their measurements for solid pig- 621 ments was better than with the proposed instrument. Indeed, 622 solid pigments showed a very glossy surface, to which the 623 GOHYLED system seemed to be more sensitive than the 624 gonio-spectrophotometers. Since the ROIs employed were 625 the same as those of the commercial devices, the illumination 626 path might be considered responsible for these differences. A 627 collimated beam and/or a smaller aperture could increase the 628 fitting with the reference devices. Finally, even though the cor- 629 relations presented provide valuable information, further vali- 630 dation by means of psychophysical experiments is still needed. 631 C. Spectral Performance in the Infrared Range 632 Figure 9shows the results of the validation of the IR range from 633 700 to 1309 nm. In this case, only the spectral fitting metrics 634 (MAE, RMSE, and GFC) were computed. Besides the full 635 ROI, the ROI of 12 mm was also analyzed, since it was the 636 most appropriate according to the measurement area of the 637 TOP 100 from the SPECTRO 320 spectrometer. Similarly 638 to the performance in the VIS range, no remarkable differences 639 were detected between the different ROIs in the IR range. 640 Consequently, only the results related to the full ROI 641 are shown. 642 These results also proved the good performance of the de- 643 veloped system in the IR range, although they were less accurate 644 than in the VIS range due to the lower density of spectral bands, 645 the greater spacing among them, and their wider FWHMs. 646 Focusing on the results, the mean values of the MAEs and 647 the RMSEs were below 0.07 and especially small in the case 648 of the solid samples. The mean GFCs were above 0.9500, with 649 the higher values obtained for the metallic samples just below 650 0.9950; therefore, the highest categories of the GFC scale were 651 not reached in the IR. Results from the two ROIs revealed 652similar outcomes. Additionally, the geometries nearer to 653 the specular reflection again caused the larger differences 654 (45°×: -30°, 15°×:0°). Similarly, the values shown in Fig. 9 655were calculated from the three measurement geometries. 656 Although the evaluation of goniochromatic pigments in the 657 IR range is not common practice, it can provide useful spectral 658data on deeper layers of coatings, as well as additional informa- 659 tion about visual effects that do not produce glint, such as 660 graininess and mottling. Preliminary data obtained with the 661developed system have revealed that sparkling effects are 662 extremely diminished in this region of the electromagnetic 663 spectrum, and, therefore, they would not interfere with the as- 664 sessment of other spatial phenomena such as those previously 665 mentioned. For instance, IR light could also be used for the 666study of graininess instead of or in addition to diffuse VIS light. 6675. CONCLUSIONS 668 A gonio-hyperspectral imaging system based on LEDs has been 669 developed and validated for the analysis of automotive coatings. 670 The design conforms to the standards DIN 6175-2, ASTM 671 E2194, and ASTM E2539 for multi-angle color measurement 672of goniochromatic pigments. 673The system provided precise reconstructions of spectral re- 674 flectances when compared to commercial devices in the IR 675 (SPECTRO 320) and especially in the VIS range (BYK-mac 676 and X-Rite). The differences would decrease if both spectral 677 ranges had similar density of spectral bands, similar spectral 678steps, and similar FWHMs. Unfortunately, current solid-state 679 technology does not provide LEDs above 800 nm with these 680 features. Similarly, color differences were smaller compared 681with the BYK-mac, although deviations with respect to the 682 X-Rite were lower than for the previous metrics. In general, 683 the mean CIEDE2000 values were within recommended tol- 684 erance ranges and smaller than the higher values of the 685AUDI2000 due mainly to the solid pigments. 686 A thorough examination of the correlations obtained with 687the whole spectral range revealed that the different ROIs 688 showed analogous results. Another shared outcome was that 689 the most sensitive geometries were those closer to the specular 690 reflection, 45°×: -60°, 45°×: -30°, 15°×: -30°, and 15°×:0° 691for the VIS and 45°×: -30° and 15°×:0° for the IR. 692 Lastly, one of the main advantages of the GOHYLED 693system is the inclusion of the spectral analysis in the two ranges 694 of the electromagnetic spectrum, VIS and IR, and for all 695 the measurement geometries recommended for the characteri- 696 zation of goniochromatic pigments [6–8]. Currently, the 697 GOHYLED system is the only instrument with these attrib- 698utes. Ongoing research focuses on the assessment of spatial 699 effects such as sparkle, graininess, and mottling under white 700 light and narrow-band light sources. The higher spatial resolu- 701tion offered by the GOHYLED system in comparison with 702 other instruments makes it particularly suitable for this task. 703 Further studies will compare these results with those obtained 704 by means of psychophysical experiments to accurately validate 705the performance of the GOHYLED system. F9:1 Fig. 9. Mean values of (a) MAEs and RMSEs and (b) GFCs for the F9:2 spectral reconstruction of the GOHYLED system for the full ROI F9:3 with respect to the SPECTRO 320. The three metrics were plotted F9:4 for pearlescent (P), metallic (M), and solid (S) samples. The error bars F9:5 represent the standard error. Research Article Vol. 56, No. 29 / / Applied Optics 9