scieee AI-readable full text Open interactive document viewer

Layer-by-layer biofabrication of coronary covered stents with clickable elastin-like recombinamers

Fernández Colino, Alicia,Wolf, Frederic,Moreira, Ricardo,Rütten, Stephan,Schmitz-Rode, Thomas,Rodríguez Cabello, José Carlos,Jockenhoevel, Stefan,Mela, Petra

Abstract

Producción Científica

Full text

1 Layer-by-layer biofabrication of coronary covered stents with clickable elastin-like recombinamers Alicia Fernández-Colinoa†, Frederic Wolfa†, Ricardo Moreiraa, Stephan Rüttenb, Thomas Schmitz-Rodea, J. Carlos Rodríguez-Cabelloc, Stefan Jockenhoevela*, Petra Melaa,d * a Department of Biohybrid & Medical Textiles (BioTex), AME Institute of Applied Medical Engineering, Helmholtz Institute, RWTH Aachen University, Forckenbeckstr. 55, 52074 Aachen, Germany b Electron Microscopy Facility, Uniklinik RWTH Aachen, Pauwelsstrasse, 30, D-52074 Aachen, Germany c Bioforge Lab, University of Valladolid, CIBER-BBN, Paseo de Belen 11, 47011 Valladolid, Spain d Medical Materials and Implants, Department of Mechanical Engineering, Technical University of Munich, Boltzmannstr. 15, 85748 Garching, Germany †Equal contribution * Corresponding authors: [email protected]; jockenhoeve[email protected]aachen.de Abstract Coronary artery disease is the leading cause of death around the world. Endovascular stenting is the preferred treatment option to restore blood flow in the coronary arteries due to the lower perioperative morbidity when compared with more invasive treatment options. However, stent failure is still a major clinical problem, and further technological solutions are required to improve the performance of current stents. Here, we developed coronary stents covered with elastin-like recombinamers (ELRs) by exploiting a layer-by-layer technique combined with catalyst free click-chemistry. The resulting ELR-covered stents were intact after an in vitro simulated implantation procedure by balloon dilatation, which evidence the elastic performance of the membrane. Additionally, the stents were mechanically stable under high flow conditions, 2 which is in agreement with the covalent and stable nature of the click-chemistry crosslinking strategy exploited during the ELR-membrane manufacturing and the successful embedding of the stent. Minimal platelet adhesion was detected after blood exposure in a Chandler loop as shown by scanning electron microscopy. The seeding of human endothelial progenitor cells (EPCs) on the ELR-membranes resulted in a confluent endothelial layer. These results prove the potential of this strategy to develop an advanced generation of coronary stents, with a stable and bioactive elastin-like membrane to exclude the atherosclerotic plaque from the blood stream or to seal coronary perforations and aneurysms, while providing a non-thrombogenic luminal surface and favouring the endothelialization. Keywords: coronary stents, layer-by-layer, elastin-like recombinamers, biobased covered stents, click-chemistry, hemocompatibility. 1. INTRODUCTION Coronary artery disease is responsible for approx. 20% of all deaths in Europe [1]. Endovascular stenting is a minimally invasive procedure to restore blood flow in the coronary arteries, which offers lower perioperative morbidity when compared with more invasive treatment options. Despite the great progress made with coronary stenting [2], restenosis and thrombosis after stent implantation are still major clinical complications that lead to stent failure and often to the need for reintervention [3]. Coated stents (e.g. drug eluting polymer coated stents and biofunctionalized endothelial progenitor cell-capturing stents) [4] and covered stents [5] have been proposed as a solution for these complications. Coronary covered stents feature a membrane that covers the metallic struts and functions as a physical barrier that can i) exclude the underlying atherosclerotic plaque from the blood flow, potentially preventing the ingrowth of smooth muscle cells and therefore the reocclusion of the stented vessel [6] and ii) seal 3 coronary artery perforations [7]. The success of such an approach relies mainly on the properties of the membrane’s material. Ideally, it should be hemocompatible, mechanically stable and elastic enough to allow the stent’s configurational changes upon implantation [5]. Unfortunately, the synthetic materials originally used for covering small-caliber stents (Table 1) display limited hemocompatibility, and their use has resulted in restenosis, thrombotic [8, 9] and inflammatory events [10, 11]. Table 1: Materials employed for the fabrication of covered coronary stents and clinical outcomes. The need to address these issues has motivated the search for new membranes with nonthrombogenic properties. In this regard, the use of bio-based materials (i.e. materials that have either a biological origin or a bio-inspired chemical composition) has been proposed as Covering material Illustrative devices Limitations/ outcomes Refs Polytetrafluoroethylene (PTFE) - JOSTENT® GraftMaster Stent Graft (Abbot, USA) - Symbiot TM (Boston Scientific, USA) -No improvement in clinical outcomes when compared to bare metal stents and in many cases, association with a higher incidence of restenosis and thrombosis [12-16] Polyethylene terephthalate (PET) - MGuard Coronary Stent System (InspireMD Ltd, Israel) - High incidence of major adverse cardiac events [17] Polyurethane (PU) PK Papyrus stent (BIOTRONIK, Germany) - No differences in major adverse cardiac events in patients treated with PU-CS or PTFE-CS after 1-year follow-up [18] Equine pericardium AneugraftDx (Amnis Therapuetics Ltd, Israel) - Potential risk of disease transmission [19, 20] Autologous vein Any balloon expandable metal stent - Time consuming fabrication - Limited availability - Patient morbidity [21] 4 alternative to the synthetic ones with the rationale of eliciting a more favourable host response [22]. Equine pericardium coronary covered stents are currently available on the market (Table 1), but the xenogenic origin may represent a limitation. Autologous veins are also used in the operating room to cover bare metal coronary stents, but their use is hampered by their limited availability and the invasiveness of the harvesting procedure. Therefore, advanced materials, able to provide a non-thrombogenic surface while offering off-the-shelf availability, are required to address the unmet clinical need of coronary covered stents. The elastin-like recombinamers (ELRs) are a family of artificial polymers bioinspired by the pentapeptide VPGVG present in the natural elastin [18, 19]. The ELRs show the advantages of an engineered material because of the exhaustive control over their composition thanks to their recombinant nature, while maintaining inherent properties of the natural elastin (i.e. elastic mechanical behaviour, hemocompatibility and bioactivity) [20]. ELRs are therefore excellent candidates for the fabrication of devices intended to be in contact with blood [23, 24]. Besides the membrane material, the fabrication strategy represents a paramount aspect to be considered, as it must be suitable for covering a device with a very small diameter (e.g. 2-4 mm). Here, we show the fabrication of covered coronary stents by combining the efficiency, stability and selectivity of catalyst-free-click chemistry [25, 26], with a layer-by-layer dip-coating technique, and the hemocompatibility and elasticity of ELRs. We tested the resulting click- ELR-coronary covered stents for their ability to withstand deployment by balloon catheter and high shear stress flow. We also evaluated the hemocompatibility of the stents and their capability to support endothelialization. 5 2. MATERIALS AND METHODS 2.1. Layer-by-layer fabrication of ELR-covered coronary stents Two ELRs modified with either cyclooctyne or azide groups (Table 2) [27] were dissolved at 100 mg/mL in PBS at 4 °C. ELR-c is a structural recombinamer while ELR-a contains the RGD adhesion sequence. For the dip coating process, ELR solutions were kept at 4 °C in separate cylindrical containers and the coronary stents (custom-made by R.T.M. Rainer Trapp Medizintechnik GmbH (Germany), L 605 Co-Cr-alloy, polished, with a length of 14.9 mm, an outer diameter of 1.8 mm and a strut thickness of 90 µm) were sequentially immersed in the ELR solutions and washed in PBS to remove the excess of polymer. The dipping procedure was repeated 5 times. The covered stents were stored in ethanol 70%. Table 2: Amino acid sequences of each ELR and the corresponding reactive groups used for catalyst-free-click chemistry (specifically Huisgen 1,3-dipolar cycloaddition of azides and alkynes). The reactive groups were incorporated by chemical modification of the lysine residues. ELR Amino acid sequence Reactive group ELR-c MESLLPVGVPGVG[VPGKG(VPGVG)5]23VPGKGVPGVGV PGVGVPGVGVPGV Cyclooctyne ELR-a MGSSHHHHHHSSGLVPRGSHMESLLP[(VPGIG)2(VPGK G)(VPGIG)2]2AVTGRGDSPASS [(VPGIG)2(VPGKG)(VPGIG)2]2 Azide 6 2.2. Mechanical stability of ELR-covered coronary stents The ELR-coronary covered stents (n=3) were exposed for 24 h to high shear stress flow in a closed-loop flow system to evaluate the robustness of the polymer covering. The system consisted of two peristaltic pumps (Ismatec, MCP Process) connected in parallel to a fluid reservoir through silicone tubes (inner diameter 6.4 mm; Ismatec). The stents were fixed in the expanded state in a tube with a diameter of 3 mm, and then subjected to the pulsatile arterial pressure of 80-120 mmHg. The resulting flow rate was set to 300 mL/min resulting in a mean shear stress of 1.5 Pa, according to the equation [28]. P = (4*η*Q)/(π*r³) Eq. 1 where P is the shear stress (in Pa), η the dynamic viscosity (in Pa*s), r is the inner radius (in m) of the covered stent and Q the volumetric flow rate (in m³*s-1). The completeness of the ELR-layer was evaluated by measuring the difference of the dry weight of the ELR-covered stents before and after exposure to the flow. The dry weight was determined after storing the ELR-covered stent in a 70 % ethanol solution overnight and subsequently placing it in an oven (Binder GmbH, Germany) at 60 °C for 90 min. Values are expressed as mean ± standard deviation (SD). Wilcoxon signed rank test was used for the statistical analysis. 2.3. Balloon expansion of the ELR-covered coronary stents The stents were positioned on the balloons (TREK coronary dilation catheter, Abbot), and expanded up to a pressure of 8 atm, corresponding to a stent diameter of 3.4 mm (supplementary video S1) as advised by the manufacturer. 7 2.4. Thrombogenicity assay Human blood was drawn from healthy volunteers and mixed with 3.2% sodium citrate using an S-Monovette CPDA1 device (Sarstedt, Germany) at a volumetric ratio of 1:9 to prevent coagulation. PVC tubes (CODAN pvd Medical GmbH) were filled with 1.5 mL of human blood. The blood and the ELR-covered stent were placed into a Chandler-loop system, formed by a closed PVC tube mounted on the rotating head of a roller pump (company) [29, 30]. The rotational speed of the Chandler loop resulted in a shear rate of 429 s-1 and a shear stress of ~ 1.5 Pa (considering a blood dynamic viscosity of 3.5 x10-3 Pa*s), within the range of physiological values [31]. After 1 h of exposure to the blood flow, the stents were washed with PBS and cut into pieces for SEM visualization. The same procedure was done with the commercially available ePTFE grafts (GORE-TEX®) as controls. The platelet covered-area was assessed by image analysis with ImageJ software [32] by analyzing three different regions per sample, each with an area of 461 µm2. SEM images were coloured with MountainsMap7 SEM software courtesy of Digital Surf, France. 2.5. Scanning electron microscopy Samples for SEM investigation were fixed in 3% glutaraldehyde in 0.1 M Sorenson’s buffer (pH 7.4) at room temperature for 1h. They were rinsed with sodium phosphate buffer (0.2 M, pH 7.39, Merck) and dehydrated consecutively in 30%, 50%, 70% and 90% acetone and then three times in 100% acetone for 10 minutes. After critical-point-drying in CO2, they were sputter-coated (Leica EM SC D500) with a 20 nm gold-palladium layer. Images were obtained with an ESEM XL 30 FEG microscope (FEI, Philips, Eindhoven, the Netherlands) with accelerating voltage of 10 kV. 8 2.6. Cell isolation and culture Endothelial progenitor cells (EPCs) were isolated from peripheral blood of human adult volunteers. Anticoagulated blood of healthy donors was carefully added to the separating solution Histo-Paque-1077 (Sigma-Aldrich, St Louis, Missouri), and centrifuged at 400 g at room temperature for 30 min. The layer containing mononuclear cells was gently washed twice with PBS. The cell pellet was resuspended with 15 mL of endothelial cell growth medium (Endothelial Cell Growth Medium MV2; PromoCell, Heidelberg, Germany) containing epidermal growth factor (5 ng/mL), basic fibroblast growth factor (10 ng/mL), insulin-like growth factor (20 ng/mL), vascular endothelial growth factor 165 (0.5 ng/mL), ascorbic acid (1 µg/mL) and hydrocoecisone (0.2 µg/mL), and transferred into T-75 culture flask precoated with human fibronectin (1 mg/cm2, Sigma-Aldrich, St. Louis, MO). The cells were cultured in 5% CO2 and 95% humidity at 37 °C. EPCs colonies were trypsinized (0.25% trypsin/0.02% ethylenediaminetetraacetic acid solution (Gibco, Karlsruhe, Germany)) and transferred to T-25 culture flask precoated with type I rat-tail collagen (5 mg/cm2, BD Biosciences, San Jose, CA). Cells with 70-80% of confluence were trypsinized and transferred into T-75 culture flasks. Cells up to passage 9 were used for all experiments. To study the ability to support endothelialization, the ELR membranes were incubated with a suspension of EPCs (5x104 cells/mL) for 6 h, after which the medium was exchanged. The scaffolds were then cultured for additional 18 h and subsequently investigated by immunohistochemistry followed by confocal microscopy. 9 2.7. Immunohistochemistry ELR membranes seeded with EPCs were fixed with methanol-free formaldehyde (Roth, Karlsruhe, Germany) at 4% for 1h at room temperature. Nonspecific sites were blocked and the cells were permeabilized by incubation in 5% normal goat serum (NGS, Dako) in 0.1% Triton- PBS. The samples were incubated overnight at 4°C with a 1:100 dilution of the primary mouse anti-CD31 antibody (P8590, Sigma). After washing three times with PBS, the samples were incubated for 1 h at room temperature with a 1:400 dilution of AlexaFluor 594 goat anti-mouse antibody (A11005, Invitrogen). The samples were washed three times with PBS, and then incubated with Triton X-100 at 0.1 % in PBS for 5 min. The membranes were then incubated for 45 min at room temperature with Acti-stainTM 488 fluorescent phalloidin (7:1000 in PBS). Samples were counterstained with 4’,6- diamidino- 2-phenylindole (DAPI) nucleic acid stain (Molecular Probes). Images were acquired using a Zeiss LSM 710 confocal laser scanning microscope. 3. RESULTS 3.1. Layer-by-layer fabrication of ELR-covered coronary stents The layer-by-layer fabrication approach (Figure 1 a) resulted in a complete click-ELR membrane which uniformly covered the whole stent with no defects or voids detected by macroscopic and microscopic inspection (Figure 1 c). 16 devices [23, 24, 41, 42] and support the use of ELRs for the fabrication of covered stents that will be in direct contact with the blood before endothelialization. In addition to the intrinsic hemocompatibility shown in the blood-contacting test, the antithrombogenicity of the ELR-covered stents is further guaranteed by their ability to support the formation of a confluent endothelial layer, as shown by culturing human EPCs (Figure 4). Circulating EPCs are increasingly recognized as important contributors to vascular prosthesis endothelialization, making them important mediators of implant compatibility [43, 44]. Besides its paramount role as anti-thrombogenic surface, the endothelium is involved in several aspects of vascular biology, e.g. blood vessel tone, hemostasis, neutrophil recruitment and hormone trafficking [45, 46]. Therefore, the ability of the covered stent to support endothelialization is of key importance to achieve the success of the implant. It is important to consider that a stented artery presents a geometry based on the stent strut configuration, which contrasts with the smoothness of the healthy arteries. This inevitably leads to a disturbed local hemodynamics (i.e. turbulent flow) [47], which has been related to the stimulation of vascular smooth muscle cell proliferation [48] and platelet activation [49]. Indeed, the thickness of the stent-struts plays an important role in-stent restenosis, with thinner struts showing considerably lower restenosis rates [50-52]. Importantly, covering the stent with the ELR-membrane bridges the stent-struts, with a concomitant reduction in the height changes in between the stent struts, which may further contribute to the positive outcome of the device. Overall, we have established a strategy that combines the simplicity of the layer-by-layer technique with the biocompatibility and specificity of the catalyst-free click-chemistry to develop a new class of coronary covered stents. The elastic membrane provided physiological hemocompatibility to the implant, and enabled endothelialization, while representing a physical barrier for the atherosclerotic plaque and smooth muscle cells ingrowth. Additionally, the 17 recombinant nature of the ELR allows for the introduction of further biological cues (e.g. immunomodulatory cues, antimicrobial peptides) to develop a platform that enables stent fabrication with tailored biofunctionalities. Acknowledgements This work was funded by the Excellence Initiative of the German federal and state governments in the framework of the START-UP Program (StUpPD_330-18), and by the START-Program of the Medical Faculty of RWTH Aachen University (60/17). JCRC acknowledges the funding from the Spanish Government (PCIN-2015-010 (FunBioPlas)) and MAT2016-78903-R), Junta de Castilla y León (VA317P18) and Centro en Red de Medicina Regenerativa y Terapia Celular de Castilla y León. The authors acknowledge the support of Prof. Dr. Müller-Newen, Sabrina Ernst and the confocal microscopy facility, a core facility of the Interdisciplinary Center for Clinical Research (IZKF) Aachen within the Faculty of Medicine at RWTH Aachen University. The authors also thank Andreas Lubig for his assistance with the graphic software. Data availability The data associated with this manuscript is available upon request to the corresponding author. References 1. Townsend, N., et al., Cardiovascular disease in Europe: epidemiological update 2016. European Heart Journal, 2016. 37(42): p. 3232-3245. 2. Byrne, R.A., M. Joner, and A. Kastrati, Stent thrombosis and restenosis: what have we learned and where are we going? The Andreas Gruntzig Lecture ESC 2014. Eur Heart J, 2015. 36(47): p. 3320-31. 3. Collet, C., et al., Coronary stent thrombosis: what have we learned? Journal of thoracic disease, 2016. 8(7): p. 1398-1405. 4. Babapulle, M.N. and M.J. Eisenberg, Coated Stents for the Prevention of Restenosis: Part I. Circulation, 2002. 106(21): p. 2734-2740. 5. Farhatnia, Y., et al., Evolution of covered stents in the contemporary era: clinical application, materials and manufacturing strategies using nanotechnology. Biotechnol Adv, 2013. 31(5): p. 524-42. 6. Kilic, I.K., et al., Coronary covered stents. EuroIntervention, 2016. 12(10): p. 1288-1295. 18 7. Romaguera, R. and R. Waksman, Covered stents for coronary perforations: is there enough evidence? Catheter Cardiovasc Interv, 2011. 78(2): p. 246-53. 8. Jamshidi, P., K. Mahmoody, and P. Erne, Covered stents: a review. Int J Cardiol, 2008. 130(3): p. 310-8. 9. Takano, M., et al., Delayed endothelialization after polytetrafluoroethylene-covered stent implantation for coronary aneurysm. Circ J, 2009. 73(1): p. 190-3. 10. Nakayama, Y., et al., Fabrication of micropored elastomeric film-covered stents and acutephase performances. J Biomed Mater Res A, 2003. 64(1): p. 52-61. 11. Sato, S., et al., Development of self-expandable covered stents. J Biomed Mater Res B Appl Biomater, 2007. 83(2): p. 345-53. 12. Kawamoto, H., et al., Short-Term and Long-Term Outcomes After Polytetrafluoroethylene- Covered Stent Implantation for the Treatment of Coronary Perforation. Am J Cardiol, 2015. 116(12): p. 1822-6. 13. Stone, G.W., et al., 5-year follow-up of polytetrafluoroethylene-covered stents compared with bare-metal stents in aortocoronary saphenous vein grafts the randomized BARRICADE (barrier approach to restenosis: restrict intima to curtail adverse events) trial. JACC Cardiovasc Interv, 2011. 4(3): p. 300-9. 14. Schächinger, V., et al., A randomized trial of polytetrafluoroethylene-membrane-covered stents compared with conventional stents in aortocoronary saphenous vein grafts. Journal of the American College of Cardiology, 2003. 42(8): p. 1360-1369. 15. Turco, M.A., et al., Pivotal, randomized U.S. study of the Symbiot™ covered stent system in patients with saphenous vein graft disease: Eight-month angiographic and clinical results from the Symbiot III trial. Catheterization and Cardiovascular Interventions, 2006. 68(3): p. 379-388. 16. Stankovic, G., et al., Randomized evaluation of polytetrafluoroethylene-covered stent in saphenous vein grafts: the Randomized Evaluation of polytetrafluoroethylene COVERed stent in Saphenous vein grafts (RECOVERS) Trial. Circulation, 2003. 108(1): p. 37-42. 17. Dudek, D., et al., Mesh-covered embolic protection stent implantation in ST-segment-elevation myocardial infarction: final 1-year clinical and angiographic results from the MGUARD for acute ST elevation reperfusion trial. Circ Cardiovasc Interv, 2015. 8(2): p. e001484. 18. Hernandez-Enriquez, M., et al., Outcomes after use of covered stents to treat coronary artery perforations. Comparison of old and new-generation covered stents. J Interv Cardiol, 2018. 31(5): p. 617-623. 19. Chen, S., et al., Pericardial covered stent for coronary perforations. Catheter Cardiovasc Interv, 2015. 86(3): p. 400-4. 20. Buchanan, G.L., A. Durante, and A. Chieffo, Pericardium-covered stent: a reality for coronary interventions of the future? . Interventional Cardiology, 2012. 4(4): p. 411-418. 21. Chae, J.K., et al., Successful treatment of coronary artery perforation during angioplasty using autologous vein graft-coated stent. European Heart Journal, 1997. 18(6): p. 1030-1032. 22. Ichihashi, S., et al., Bio-Based Covered Stents: The Potential of Biologically Derived Membranes. Tissue Eng Part B Rev, 2019. 25(2): p. 135-151. 23. Jordan, S.W., et al., The effect of a recombinant elastin-mimetic coating of an ePTFE prosthesis on acute thrombogenicity in a baboon arteriovenous shunt. Biomaterials, 2007. 28(6): p. 1191- 7. 24. de Torre, I.G., et al., Elastin-like recombinamer-covered stents: Towards a fully biocompatible and non-thrombogenic device for cardiovascular diseases. Acta Biomater, 2015. 12: p. 146- 155. 25. McKay, C.S. and M.G. Finn, Click chemistry in complex mixtures: bioorthogonal bioconjugation. Chem. Biol., 2014. 21(9): p. 1075-101. 26. Jiang, Y., et al., Click hydrogels, microgels and nanogels: emerging platforms for drug delivery and tissue engineering. Biomaterials, 2014. 35(18): p. 4969-85. 27. Gonzalez de Torre, I., et al., Elastin-like recombinamer catalyst-free click gels: characterization of poroelastic and intrinsic viscoelastic properties. Acta Biomater, 2014. 10(6): p. 2495-505. 19 28. Darby, R. and R.P. Chhabra, Chemical Engineering Fluid Mechanics. 2001: CRC Press. 559. 29. Chandler, A.B., In vitro thrombotic coagulation of the blood; a method for producing a thrombus. Lab Invest, 1958. 7(2): p. 110-4. 30. Gong, J., et al., Tubing loops as a model for cardiopulmonary bypass circuits: both the biomaterial and the blood-gas phase interfaces induce complement activation in an in vitro model. J Clin Immunol, 1996. 16(4): p. 222-9. 31. Gaamangwe, T., S.D. Peterson, and M.B. Gorbet, Investigating the Effect of Blood Sample Volume in the Chandler Loop Model: Theoretical and Experimental Analysis. Cardiovascular Engineering and Technology, 2014. 5(2): p. 133-144. 32. Schneider, C.A., W.S. Rasband, and K.W. Eliceiri, NIH Image to ImageJ: 25 years of image analysis. Nature Methods, 2012. 9: p. 671. 33. Al-Mukhaini, M., et al., Coronary perforation and covered stents: an update and review. Heart views : the official journal of the Gulf Heart Association, 2011. 12(2): p. 63-70. 34. Banerjee, S., et al., Femoropopliteal Artery Stent Thrombosis: Report From the Excellence in Peripheral Artery Disease Registry. Circ Cardiovasc Interv, 2016. 9(2): p. e002730. 35. Nebeker, J.R., et al., Hypersensitivity cases associated with drug-eluting coronary stents: a review of available cases from the Research on Adverse Drug Events and Reports (RADAR) project. J Am Coll Cardiol, 2006. 47(1): p. 175-81. 36. Khan, W., S. Farah, and A.J. Domb, Drug eluting stents: Developments and current status. Journal of Controlled Release, 2012. 161(2): p. 703-712. 37. Ichihashi, S., et al., In Vitro Quantification of Luminal Denudation After Crimping and Balloon Dilatation of Endothelialized Covered Stents. CardioVascular and Interventional Radiology, 2017. 40(8): p. 1229-1236. 38. Urry, D.W., et al., Biocompatibility of the Bioelastic Materials, Poly(GVGVP) and Its γ-Irradiation Cross-Linked Matrix: Summary of Generic Biological Test Results. Journal of Bioactive and Compatible Polymers, 1991. 6(3): p. 263-282. 39. Ibanez-Fonseca, A., et al., Biocompatibility of two model elastin-like recombinamer-based hydrogels formed through physical or chemical cross-linking for various applications in tissue engineering and regenerative medicine. J Tissue Eng Regen Med, 2018. 12(3): p. e1450-e1460. 40. Hoeve, C.A. and P.J. Flory, The elastic properties of elastin. Biopolymers, 1974. 13(4): p. 677- 86. 41. Woodhouse, K.A., et al., Investigation of recombinant human elastin polypeptides as nonthrombogenic coatings. Biomaterials, 2004. 25(19): p. 4543-53. 42. Ito, S., S. Ishimaru, and S.E. Wilson, Application of coacervated alpha-elastin to arterial prostheses for inhibition of anastomotic intimal hyperplasia. ASAIO J, 1998. 44(5): p. M501-5. 43. Melchiorri, A.J., et al., In Vitro Endothelialization of Biodegradable Vascular Grafts Via Endothelial Progenitor Cell Seeding and Maturation in a Tubular Perfusion System Bioreactor. Tissue Eng Part C Methods, 2016. 22(7): p. 663-70. 44. Yu, Y., et al., Characterization of Endothelial Progenitor Cell Interactions with Human Tropoelastin. PLoS One, 2015. 10(6): p. e0131101. 45. Rajendran, P., et al., The vascular endothelium and human diseases. Int J Biol Sci, 2013. 9(10): p. 1057-69. 46. Sandoo, A., et al., The endothelium and its role in regulating vascular tone. Open Cardiovasc Med J, 2010. 4: p. 302-12. 47. Murphy, E.A. and F.J. Boyle, Reducing In-Stent Restenosis Through Novel Stent Flow Field Augmentation. Cardiovascular Engineering and Technology, 2012. 3(4): p. 353-373. 48. Shigematsu, K., et al., Direct and indirect effects of pulsatile shear stress on the smooth muscle cell. Int Angiol, 2000. 19(1): p. 39-46. 49. Shadden, S.C. and S. Hendabadi, Potential fluid mechanic pathways of platelet activation. Biomechanics and modeling in mechanobiology, 2013. 12(3): p. 467-474. 20 50. Sakamoto, A., et al., Understanding the Impact of Stent and Scaffold Material and Strut Design on Coronary Artery Thrombosis from the Basic and Clinical Points of View. Bioengineering (Basel), 2018. 5(3). 51. Kastrati, A., et al., Intracoronary stenting and angiographic results: strut thickness effect on restenosis outcome (ISAR-STEREO) trial. Circulation, 2001. 103(23): p. 2816-21. 52. Pache, J., et al., Intracoronary stenting and angiographic results: strut thickness effect on restenosis outcome (ISAR-STEREO-2) trial. J Am Coll Cardiol, 2003. 41(8): p. 1283-8.