A Feasibility Study of the Suitability of an AD5933-based Spectrometer for EBI Applications

Ansede Pena, Antonio

Abstract

Electrical Bioimpedance (EBI) measurements have proven their validity in several medical applications like body composition analysis and detection of melanoma among others. The successful application of EBI technology on the field of medicine has lead the way for applications in the field of personal healthcare and body performance in the field of sports. Due to the widespread use of the EBI technology and rising of new EBI applications requiring system portability or even suitable to wear, the manufacturer Analog devices has introduced in the market the first integrated system dedicated to measure EBI, the impedance network analyzer AD5933. The availability of this EBI spectrometer device opens up new horizons for the integration of the measurement systems to meet the demands of new EBI applications and allowing the development of portable and even wearable measurement systems. This project is focused on the AD5933 impedance network analyzer, and it aims to identify the EBI applications in which, the use of an AD5933 device is suitable. To adapt the AD5933 device for biomedical measurements an Analog Front-End (AFE) has been used to enable the system for 4-electrodes measurements. In order to evaluate the performance of AD5933 with the AFE, experimental measurements on electrical equivalent models have been taken with the AD5933+4E-AFE system and the EBI spectrometer Impedimed SFB7. The obtained impedance spectral data have been used to estimate the values of the equivalent circuit under measurement and the estimated values have been mutually compared in terms of estimation accuracy.

Full text

Final Degree Thesis A Feasibility Study of the Suitability of an AD5933-based Spectrometer for EBI Applications By Antonio Ansede Pena FINAL DEGREE THESIS 30 ECTS, ERASMUS 2009-10 SWEDEN ELECTRICAL ENGINEERING SPECIALIZATION IN COMMUNICATIONS & SIGNAL PROCESSING THESIS Nº9/2009 ii A Feasibility Study of the Suitability of an AD5933-based Spectrometer for EBI Applications Antonio Ansede Pena Final Degree Thesis Subject Category: Medical Technology, Electronic Instrumentation University of Borås School of Engineering SE- 501 90 BORAS Telephone +46 33 435 4640 Examiner: Fernando Seoane Martínez Supervisor: Fernando Seoane Martínez Date: 2009 Sept 28th Keyword: Electrical Bioimpedance, Electrical Bioimpedance Spectroscopy, AD5933, SFB7, Analog Frond End, Statistical Analysis, Wearable Medical Devices. iii ABSTRACT Electrical Bioimpedance (EBI) measurements have proven their validity in several medical applications like body composition analysis and detection of melanoma among others. The successful application of EBI technology on the field of medicine has lead the way for applications in the field of personal healthcare and body performance in the field of sports. Due to the widespread use of the EBI technology and rising of new EBI applications requiring system portability or even suitable to wear, the manufacturer Analog devices has introduced in the market the first integrated system dedicated to measure EBI, the impedance network analyzer AD5933. The availability of this EBI spectrometer device opens up new horizons for the integration of the measurement systems to meet the demands of new EBI applications and allowing the development of portable and even wearable measurement systems. This project is focused on the AD5933 impedance network analyzer, and it aims to identify the EBI applications in which, the use of an AD5933 device is suitable. To adapt the AD5933 device for biomedical measurements an Analog Front-End (AFE) has been used to enable the system for 4-electrodes measurements. In order to evaluate the performance of AD5933 with the AFE, experimental measurements on electrical equivalent models have been taken with the AD5933+4E-AFE system and the EBI spectrometer Impedimed SFB7. The obtained impedance spectral data have been used to estimate the values of the equivalent circuit under measurement and the estimated values have been mutually compared in terms of estimation accuracy. iv ACKNOWLEDGEMENTS First I would like to thank specially my thesis supervisor, Dr. Fernando Seoane, for his encouragement and patience throughout the duration of this final degree thesis and I would like to specifically appreciate all the dedication that he has provided me. Fernando, I am deeply thankful for all the knowledge that you have shared with me and for all the good moments. I am also grateful to my colleagues and friends, David A., Ivan Pau, Javier F., Juan Carlos M., Lola R., Rubén B., Ruth G., Ruth P. for his helpful comments and ideas and for made more enjoyable to work at the BRC (Bioimpedance Research Center) and for the great moments together. I would like to thank to the Polytechnic University of Catalonia for giving me the opportunity to come to Borås and especially to Prof. Pere Joan Riu Costa, Dr. Ramon Bragós Bardia and Dr. Lluís Prat Viñas. And above all I want to thanks to my friends and family, my mother Maria del Carmen, my girlfriend Diana, my uncles Suso and Modesto, my aunts Elvira and Blanca, my grandparents Antonio and Efigenia and my cousins Elvi and Maria Jesus, for giving me always support and to encourage me whenever I needed it. I dedicate this work to all of you. Antonio Ansede v TABLE OF CONTENTS Abstract ................................................................................................................................... iii Acknowledgements .................................................................................................................. iv Table of Contents ..................................................................................................................... v List of Acronyms ................................................................................................................... viii CHAPTER 1 Introduction ................................................................................................... 9 1.1 Introduction........................................................................................................................ 9 1.2. Motivation ......................................................................................................................... 9 1.3. Goal ................................................................................................................................. 10 1.4. Work done ....................................................................................................................... 10 1.5. Structure of the Thesis Report ........................................................................................ 10 1.6. Out of Scope .................................................................................................................... 11 CHAPTER 2 Background .................................................................................................. 12 2.1. Introduction to Electrical Bioimpedance ....................................................................... 12 2.1.1. Historical Introduction .................................................................................................... 12 2.1.2. Electricals Properties of Biological Tissue ..................................................................... 12 2.1.3. The Dispersion Windows ................................................................................................. 14 2.2. Measurements of Electrical Bioimpedance ................................................................... 14 2.2.1. The Cell Membrane ............................................................................................................. 14 2.2.2. Cole Model and Cole Plot .................................................................................................... 16 2.2.3. Electrode configuration ....................................................................................................... 17 2.3. Applications of Bioimpedance ........................................................................................ 18 2.3.1. Body Composition ................................................................................................................ 18 2.3.1.1. BIA Analysis ................................................................................................................................ 19 2.3.1.2. Single Frequency BIA .................................................................................................................. 21 2.3.1.3. Multi frequency BIA ..................................................................................................................... 21 2.3.1.4. Spectroscopy BIA ......................................................................................................................... 22 2.3.1.5. Segmental BIA ............................................................................................................................. 22 2.3.1.6. BIVA Analysis ............................................................................................................................. 23 2.3.2. Respiration Rate (Impedance Pneumography) ................................................................. 24 2.3.3. Lungs Composition (Impedance Plethysmography) ......................................................... 25 CHAPTER 3 Instrumentation and Method ...................................................................... 27 3.1. Instrumentation .............................................................................................................. 27 3.1.1. SFB7 Body Composition Analyzer ..................................................................................... 27 vi 3.1.2. The BioImp Software ........................................................................................................... 28 3.1.3. Evaluation Board for the Impedance Converter Network Analyzer AD5933 ............... 29 3.1.4. Four Electrodes Analog Front End .................................................................................... 31 3.2. Method ............................................................................................................................. 32 3.2.1. Application Equivalent Modeling ................................................................................... 33 3.2.2 EBI Application Equivalent Load ....................................................................................... 34 3.2.3 EBI Application Equivalent Spectroscopy Measurements ............................................... 37 3.2.3.1. SFB7 measurements. .............................................................................................................. 37 3.2.3.2 AD5933+4-AFE measurements. ............................................................................................ 38 3.2.4 Performance Comparison .................................................................................................... 39 3.2.4.1 Model Parameter Estimation. ................................................................................................. 40 3.2.4.2 Statistics of the Model Parameter Estimation. ....................................................................... 41 3.2.4.3 Estimation Error Visualization. .............................................................................................. 41 CHAPTER 4 Results .......................................................................................................... 42 4.1 Overview ........................................................................................................................... 42 4.2 SFB7 Measurements ........................................................................................................ 42 4.2.1 Total Body Composition (TBC): ..................................................................................... 42 4.2.2 Respiration Rate (RR): .................................................................................................... 43 4.2.3 Lungs Composition (LC): ................................................................................................ 43 4.2.4 Segmental Body Composition (SBC): ............................................................................. 44 4.2.4.1 Arm-Arm (AA) ...................................................................................................................... 44 4.2.4.2 Trunk-Trunk (TT): ................................................................................................................. 44 4.2.4.3 Leg-Leg (LL): ........................................................................................................................ 45 4.2.1. Measurements Summary ..................................................................................................... 46 4.3 Modeling 2R1C ................................................................................................................ 46 4.4. Spectroscopy Measurements in 2R1C Models ............................................................... 48 4.4.1 TBC: .................................................................................................................................. 48 4.4.2 RR: ..................................................................................................................................... 49 4.4.3 LC: ..................................................................................................................................... 49 4.4.4 AA: ..................................................................................................................................... 50 4.4.5 LL: ..................................................................................................................................... 51 4.4.6 TT: ..................................................................................................................................... 51 4.4.7 Comparison among the Theoretical Model Values and the Estimated Model Values ... 52 4.5. SFB7 Vs. AD5933+4-AFE ............................................................................................. 53 CHAPTER 5 Discussion .................................................................................................... 61 5.1. Performance of 2R1C Components Estimation ............................................................ 61 5.2. Study of Errors: Relative Error and Standard Deviation ............................................. 61 vii 5.3. Spectral Results ............................................................................................................... 62 CHAPTER 6 Conclusions & Future Work ....................................................................... 64 6.1. Conclusions ..................................................................................................................... 64 6.2. Future work ..................................................................................................................... 64 REFERENCES ...................................................................................................................... 65 viii LIST OF ACRONYMS AA - Arm-Arm ADC - Analog Digital Converter AFE - Analog Front End BCM - Body Cell Mass BIA - Bioimpedance Analysis BIS - Bioimpedance Spectroscopy BIVA - Bioimpedance Vector Analysis DSP - Digital Signal Processor EBI - Electrical Bioimpedance ECW - Extra Cellular Water FC - Characteristic Frequency FFM - Fat Free Mass LC - Lungs Composition LL - Leg-Leg MF-BIA - Multi Frequency Bioimpedance Analysis RR - Respiration Rate SBC - Segmental Body Composition SF-BIA - Single Frequency Bioimpedance Analysis TBW - Total Body Water TT - Trunk- Trunk ICG - Impedance CardioGraphy ICF - IntraCellular Fluid 9 CHAPTER 1 INTRODUCTION 1.1 Introduction Nowadays there is many and several applications of Electrical Bioimpedance (EBI), these have been emerging as answers to clinical needs as methods for non-invasive monitoring and the detection of changes in the structure and composition of tissues produced by pathophysiological processes. A new Integrated Circuit (IC), the AD5933 is available to perform EBI measurements allowing for the development of small-integrated EBI measurement systems enabling portable and wearable systems. 1.2. Motivation The increase of ageing population demands that the model of hospital centered healthcare turns towards one model of healthcare more personalized and home care-based. In this way, the patient involvement on the health care process will increase potentially increasing prevention and early detection what will provide eventually in a better and more affordable treatment. Such changes create the need to develop a new efficient, sustainable and useful technology to equip the patient with a tool for a personal health monitoring and simultaneously allowing the healthcare staff to monitor remotely the conditions of the patients. The latter together with the technological development of the field of functional textiles point out to the emerging area of personalized healthcare monitoring as plausible area of operation. EBI technology combines both the potential to be used in personalized healthcare as well as being able to benefit from the technological developments within functional textiles. The only piece missing to enable the development of wearable EBI measurement systems for personalized healthcare monitoring is the electronic device that will allow the implementation of EBI measurement systems as wearable systems. Recently the manufacturer of integrated circuits, Analog Devices, has introduced in the market the first integrated system dedicated to measure EBI: the AD5933. The availability of this EBI spectrometer device opens up new horizons for the integration of the measurement systems of EBI into portable and even wearable systems. To adapt the AD5933 device for human measurements an Analog Front-End has been developed to realize measurements with four electrodes wires. There are many applications that could benefit from such a small device the AD5933 and a natural and mandatory step in any development process is to identify which medical and 16 Biological tissue contains various different types of cells in several different arrangements and following the same approach applied by Fricke on the frequency range of the Beta- Dispersion, explanatory and descriptive models can be obtained with circuit bridges. Figure 2.5 presents a series circuit bridge and Figure 2.6 presents a parallel circuit bridge. 2.2.2. Cole Model and Cole Plot Electrical Bioimpedance is a measurement of the ability of a material to create an opposition to the flow of electric current through the tissues and the impedance is denoted by Z. The impedance can be represented mathematically as a complex magnitude in cartesian coordinates, made up of two components, Resistance (R) and Reactance (X) (2.2), or in polar coordinates using magnitude (|Z|) and phase (θ) (2.3). (2.2) Figure 2.4 Current paths in a suspension of cells at low and high frequencies. Figure 2.5 Equivalent model used for skin tissues Figure 2.6 Equivalent model typically used for suspension of cells 17 (2.3) The electrical impedance of linear systems is ruled by Ohm‟s law, which relates voltage in volts and current in Amperes through an impedance in Ohms, as denoted in (2.4). (2.4) Based in Ohm‟s law, a deflection method to measure electrical bioimpedance consist on applying electrical current or voltage on the tissue and measure the response of the tissue to the electrical stimulus. Like the input value over the impedance load is know and the complementary value is measured, it is possible to obtain the value of the impedance load with Ohm‟s law. Most often the electrical stimulus is applied into the tissue through electrodes and the electrical response is also measured with electrodes. Therefore in order to do EBI measurements there are several electrode configurations available and some of them they will be explained in the following section. 2.2.3. Electrode configuration There are several electrode configurations, but in this section only two will be explained: The 2-Electrode and 4-Electrode configuration, which are the configuration relevant to this work. 2-Electrode configuration With this configuration, depicted in Figure 2.7, both electrodes are used to inject the electrical stimulus and to measure the response produced by the tissue. Therefore, in this configuration the electrode-electrolyte interface impedances are in series with the sample impedance providing a total measured voltage VMEAS containing the sum of the voltages caused by the three impedances, see (2.5). (2.5) Fig 2.7 Two electrodes configuration. 18 Therefore the impedance obtained by applying Ohm‟s law is the combination of the tissue impedance together with the electrode impedance. Unfortunately, these impedances can be difficult or almost impossible to separate and alternative measurement method must be used: The 4-Electrode configuration. 4-Electrode configuration In this method the electrical stimulus is applied with a pair of electrodes and the resulting response is measured with a different pair of electrodes, see Figure 2.8. Focusing on a current driven system, the voltage on the current injecting electrodes is not contained in the voltage response sensed by the system, therefore the influence of the electrodes impedance can be reduced (2.6). (2.6) 2.3. Applications of Bioimpedance To analyze EBI data for each of the applications that have been studied in this work, several approaches are available. In this section several methods used to obtain Body Composition parameters as well as the applications of Respiration Rate and Lungs Composition will be introduced. 2.3.1. Body Composition The measurement of body fluid volumes, see Figure 2.7, Extra Cellular Water (ECW), Intra Cellular Water (ICW) and their sum, total body water (TBW) is important in many pathologies. TBW is strongly related to Fat-Free-Mass (FFM) which contains, in healthy individuals, an average of 73.2% of water [12]. Figure 2.8 Four electrodes configuration for a current driven system 19 Similarly, Body Cell Mass (BCM), which is an important nutritional parameter, is also closely connected to ICW. Independent measurements of FFM and TBW permit to detect dehydration, which is frequent in elderly persons or athletes after heavy training. Conversely, an overhydratation may indicate the presence of edema in cardiac patients of lymphoedema after a mastectomy. The measurement of TBW is also useful for evaluation of diuretic therapy. Renal patients treated by hemodialysis accumulate fluid between treatments. It is important to evaluate their amount of excess fluid, in order to determine how much fluid they should lose by ultra filtration and also how this fluid loss is distributed between ECW and ICW. Measurements of BCM are also important for assessing the morbidity of patients infected by HIV [13]. For Body Composition analysis there are several EBI methods available and two of them; BIA Analysis and BIVA will be introduced in the following sections. 2.3.1.1. BIA Analysis Bioimpedance analysis (BIA) is a very well widespread method for assessment of Body Composition. It consists in the measurement of the impedance or opposition to the flow of an electric current through the body fluids contained mainly in the lean and fat tissue. Impedance is low in lean tissue, where intracellular fluid and electrolytes are primarily contained, but high in fat tissue. Impedance is thus proportional to TBW value. Due to the impedance in lean tissue, where intracellular fluid and electrolytes are primarily contained, is lower than in fat tissue and because of the measured body resistance differs depending on the amount of body fluids, it is Figure 2.7 Schematic of fat-free mass, total body water, intracellular water, extracellular water and body cell mass. 20 possible to estimate the body composition. The basic model of this method is depicted in Figure 2.8 and it is based on the assumption that the body is a cylindrical-shape ionic conductor. In such volume conductor, the impedance between faces of a cylinder of finite length L and a cross sectional area A, when a uniform current density is applied parallel to its axis, is given by (2.7) (2.7) The measured impedance is a function of both the tissue complex resistivity ρ and also the shape of the conductor. Although the body is not a uniform cylinder and its conductivity/resistivity is not constant, a relationship between the volume of body water (V) and the ratio length (L) to impedance can be established as in equations (2.8) and (2.9) [14]. (2.8) (2.9) It is easier to measure height than the conductive length, which is usually from wrist to ankle. Therefore, the empirical relationship denoted as in the equation 2.10. (2.10) Due to the inherent inhomogeneity of the human body, and that (2.9) holds true for a homogeneous cylinder, the coefficient K in (2.10) describes an equivalent cylinder, which must be adjusted to match the real geometry. The value of the coefficient depends on various factors. More information can be obtained reading [15]. Figure 2.8 Principles of BIA from physical characteristics to body composition. Cylinder model for the relationship between impedance and geometry. 21 With the aim to obtain the body composition parameters the follow BIA methods are currently used: Single Frequency-BIA Multi Frequency-BIA Spectroscopy-BIA Segmental-BIA 2.3.1.2. Single Frequency BIA Single Frequency Bioimpedance Analysis (SF-BIA) methods rely only on the whole-body impedance (Z50) or the resistance (R50) measured at 50 KHz between the wrist and the ankle, as shown in Figure 2.9. Impedance data are subsequently entered into predictive equations derived through statistical regression in order to determine TBW, from which FFM is calculated according to the assumption that FFM is hydrated at 73.2% [16]. Several studies showed that a frequency of 50 kHz is too low for full penetration of current through the cell [17]. Theoretically, these devices are unable to discriminate between ECW and ICW. Therefore, the use of these new devices is not recommended in cases of abnormal fluid distribution. i.e. pathological cases. 2.3.1.3. Multi frequency BIA As with SF-BIA, Multi Frequency Bioimpedance Analysis MF-BIA uses empirical linear regression models but includes measurement of impedance at multiple frequencies. MF-BIA uses impedance data measured at least 2 frequencies: 1 at a very low frequency, usually 5 kHz, and 1 at a high frequency, typically 50, 100, 200, or 500 kHz, to evaluate FFM, TBW, ICW and ECW [18]. At the lowest frequency, measurements of impedance can be used to determine ECW. At Figure 2.9 Standard placements of electrodes for wrist to ankle 4E-measurements for SF-BIA and MF-BIA applications. 22 higher frequencies, the current can pass through the cell membrane, and thus, the impedance measurements can be used to determine TBW. The impedance data are applied to regressionderived equations in order to predict TBW, ECW, and ICW. According to [19] MFBIA is more accurate and less biased than SF-BIA for the prediction of ECW, whereas SF-BIA, compared to MF-BIA, is more accurate and less biased for TBW in critically ill subjects. [20]. 2.3.1.4. Spectroscopy BIA Spectroscopy Bioimpedance Analysis operates on impedance data measured over a wide spectrum of frequencies, usually from 5 to 1000 kHz. To analyze the BIS-derived impedance data, the most practical approach for clinical applications is to perform biophysical modeling on the impedance data; the modeling procedure involves fitting the spectral data to the Cole model using nonlinear curve fitting [21] [13]. This procedure generates the Cole model parameters, as well as R0, R, and exponent α. Cole model terms are then applied to equations derived from the Hanai mixture theory [22], which is essentially based on the notion that the body is a conducting medium of water, electrolytes, and lean tissue, in addition to non-conducting material within it. ECW and ICW are thus calculated individually; TBW is calculated as the sum of ECW and ICW. Bioimpedance Spectroscopy Analysis provides a more direct, individualized measurement of ECW and ICW than other bioimpedance approaches. This is due, in large part, to the fact that it measures impedance over a broad spectrum of frequencies, rather than being limited to only 1 or 2 frequencies, like in the case of SF-BIA and MF-BIA, respectively. Of particular importance, BIS differs from SF-BIA and MF-BIA in that it does not rely on statistically derived regression equations to predict body water volumes. Furthermore, it relies far less on assumptions that may be violated in disease states. For example, 2 assumptions underlying the SF-BIA method that may not hold true in many clinical populations are that FFM is 73% hydrated and that ICW and ECW are normally distributed. 2.3.1.5. Segmental BIA Segmental Bioimpedance Analysis, the body is considered to comprise five interconnecting conducting cylinders, two legs, two arms, and the trunk, as shown in Figure 2.10. Although the trunk of the body represents as much as 50% of whole body mass, it only contributes around 10% to whole body impedance. This implies three aspects for body composition analysis when using the whole body-one cylinder- BIA approach: - Changes of the impedance are closely related to changes of the FFM of the limbs. - Changes of the FFM of the trunk are probably not adequately described by whole body impedance measurements - Even large changes in the fluid volume within the abdominal cavity have only minor influence on the measurement of FFM or BCM. 23 2.3.1.6. BIVA Analysis In Piccoli et al (1994) the RXc-graph method was proposed. Resistance, R and reactance, Xc are obtained with right-side configuration in SFBIA at 50 kHz after standardization by height H. In contrast to other bioimpedance methods this approach does not yield any absolute estimate of ECW, ICW or TBW, makes no assumptions about body geometry, hydration state, or the electrical model of cell membranes and is unaffected by regression adjustments. In BIVA, Z can be considered as a bivariate random vector: resistance R and reactance Xc. These two variables, standardized for height, are plotted as point vectors in the R–Xc plane. An individual vector can then be compared with the reference 50%, 75%, and 95% tolerance ellipses calculated in the healthy population of the same gender and race. That is called R–Xc graph method; see Figure 2.10. The ellipse varies with age and body size. For more information see [15] Figure 2.10 Five-Cylinder model of the body and the impedance for each segment: right arm, right leg, left arm, left leg, and trunk. 24 2.3.2. Respiration Rate (Impedance Pneumography) Impedance pneumography is a non-invasive technique for monitoring the respiration rate and involves measuring changes in the electrical impedance of the thoracic cavity. The resistivity of lung tissue increases due to the thinning of the alveolar walls as well as by the deformation of the epithelial cells and blood vessels through the expansion of the alveoli [23]. These changes in resistivity combined with diaphragm movements and changes in the chest circumference affect the overall impedance of the thoracic cavity during breathing [24]. Even though there exists a good correlation between the volume of inspired air and impedance of the thoracic cavity as pointed in [25] the volume of air in the lungs is not the responsible for the change in resistance. Figure 2.11 BIVA Pattern (RXc-graph) [2]. Figure 2.12 Electrode configurations for pneumography impedance measurements. In a) and in b) are respectively showed the bi-polar and the tetra-polar configuration. 25 These impedance changes are measured by injecting a low-amplitude, single high-frequency, constant current using a pair of „drive‟ electrodes and recording the resultant voltage changes either from the same pair of electrodes as in a 2-Electrode (bi-polar) configuration in the Figure 2.12.a or from a different pair of „receive‟ electrodes placed at an appropriate location as in a 4- Electrode (tetra-polar) configuration in the Figure 2.12.b [26]. Impedance pneumography is well tolerated by patients because it does not involve placing any device within the airway that may cause discomfort to the patient. Therefore, it is the most widely used technique for the purpose of long-term monitoring and home monitoring. 2.3.3. Lungs Composition (Impedance Plethysmography) The impedance plethysmography is a non-invasive technique used to detect the changes of the thoracic impedance. It is known that changes of the thoracic impedance correlate with changes of the clinical syndromes of a lung edema due to the accumulation of fluid in the lungs [27] [28], see Figure 2.13. Furthermore, the thoracic impedance decreases considerably before the first clinical syndromes appear. So lung edema can be well detected using impedance plethysmography during their appearance and disappearance. The electrode placement for measurements of lung edema is under investigation, but there are two methods commonly used. The first method is the tetra polar configuration, as the Figure 2.14.a shows and the other method uses 6 electrodes in the specific placement as shows the Figure 2.14.b. Figure 2.13 The figure illustrates in the left hand normal lungs and in the right hand lungs with Pulmonary Edema. The light blue color on the right side drawing represents water. 32 Since the AD5933 applies voltage and expects a current at its input RFB, the AFE interfacing with the AD5933 has a voltage input and a current output. Source output expressly generates a current resulting from the ratio from Vout and the impedance of the Tissue Under Study (TUS), which is the exactly the current expected by the AD5933 at the input RFB as indicated in Figure 3.7. At the TUS side, the AFE has a current source as output, exciting the TUS with 350 μA-rms, while the input is a differential voltage measurement channel. In essence the AFE operation can be describe as follows: after the removal of the DC bias component from the voltage output of the AD5933 with a low pass filter at the input of the first V2CC, the AC voltage from Vout drives the Voltage Controlled Current Source (VCCS) injecting an AC current Iout in the TUS directly proportional to Vout. The second V2CC senses at its differential input the voltage drop caused at the TUS by the AC current generated by the first V2CC flowing through the TUS. Finally an AC current proportional to the voltage drop in the TUS is generated at the output of the second V2CC with an added DC component equivalent to the DCbias removed originally from the Vout. The total gain introduced by the cascade combination of both V2CC in the AFE is selectable by the choosing of several resistors. 3.2. Method In order to evaluate the performance of the Impedance Spectrometer On-chip AD5933 +4- Electrode Analog Frond End (4E-AFE), Impedance spectroscopy measurements for several EBI applications have been taken. Experimental measurements on electrical equivalent models have been taken with the AD5933+4E-AFE system and the EBI spectrometer Impedimed SFB7. The obtained impedance spectral data have been used to estimate the values of the equivalent circuit under measurement and the estimated values have been mutually compared in terms of estimation accuracy. Figure 3.7 Functional block diagram of the Analog Front-End. 33 The measurements have been taken on 2R1C equivalent models [30], each of them representing an specific EBI application. The electrical equivalence has been established both in terms of frequency as well as Ohmic load dynamic range. The values of the passive electrical components models were obtained experimentally from experimental EBI measurements. Since the value of the working load depends on the EBI applications, the functioning of the AD5933+4E-AFE was adjusted specifically to the impedance values range of each model. Figure 3.8 shows the work flow that has been applied in this work. 3.2.1. Application Equivalent Modeling The equivalent circuit topology chosen to model the EBI applications has been the 2R1C [30]. To obtain the values of the passive components for each specific equivalent model, the following steps have taken: 1. Experimental Measurements. In this step EBI measurements have been taken in a healthy subject for the following EBI applications: - Total Body Water contents - Lungs Composition - Respiration Rate - Segmental measurements for Body Composition: Arm-Arm, Leg-Leg, Trunk-Trunk. The specific morphological data of the subject is shown in Table 3.1. The 4-Electrode EBI measurements were taken with the SFB7 spectrometer within the frequency range 3 kHz to 1 MHz. Figure 3.9 Modeling process overview Experimental Measurements with SFB7 •Lungs •RR •TBC •SBC BIOIMP Circuit Modelling •Averaging of 100 spectroscopy measurements for each application. TABLE 3.1 Sex Age (years) Length(cm) Weight(Kg) Male 24 173 79 34 2. BioImp 2R1C-paralell bridge Circuit Modeling. The obtained spectral impedance data from each of the EBI application specific measurements were processed with the BioImp software to obtain the values for the passive components of the equivalent circuit model. The obtained values and parameters obtained from each EBI applications are the following: - 2R1C model components: o Intracellular Resistance (Ri) o Extracellular Resistance (Re) o Membrane Capacitor (Cm) - Resistance at 0 Hz ( ) - Resistance at Hz( ) - Characteristic Frequency (Fc) The final value for each of the parameters was obtained from the averaging of 100 spectroscopy measurements using the batch processing tool available at BioImp. 3.2.2 EBI Application Equivalent Load Once typical values for the components of the 2R1C have been calculated by averaging the experimentally performed EBI measurements, a typical dynamic range for the resistance was obtained. The maximum value of the model is set by Re and the minimum is set by the parallel connection of Re//Ri. Since the circuit parameters have been obtained from a single subject and both Re and Ri have an strong dependency on the size and the form of the tissue under study i.e. the whole body. The values of the resistors of the 2R1C bridge have been adjusted considering for the variance in morphological parameters of the population i.e. size, complexion, high etc. The applied process is explained as follows in the next diagram, see Figure 3.10. 1. Increasing the Load Dynamic Range. The range of values of the load was increased not directly over the 2R1C model but on the Cole parameters. Since the EBI spectrometer measures bioimpedance down to 0 Ohms, the upper limit of the range was adjusted modifying the value of R0. The new R0‟ value was obtained by Figure 3.10 EBI Application Equivalent Load and Customization Process Overview Increasing Load Dynamic Range •RE=R0'=R0(1+α) AD5933+4E-AFE application customization •RG1, R4and RRFB 35 using a multiplicative factor (1+α) for R0, where is a constant set to 0.5. i.e. the value of R0 was increased 50%, see Equation 3.1. (3.1) Due to this modification the values of the equivalent 2R1C model for the resistances Ri and Re and Cm were recalculated aiming to keep the original frequency dependency. The value recalculation was done as follows. The 2R1C circuit was simplified for D.C: According to the previous figure Re‟ is equal to R0‟, that is the maximum value of the resistance that the EBI spectrometer will measure for a given EBI application and therefore the AD5933, will be adjusted for a RTUS max equal to the newly obtained Re‟ to make the best use of the ADC. See Equation 3.2. (3.2) In order to keep the value of the characteristic frequency of the 2R1C similar to the typical value of each corresponding EBI application, the value of Ri was recalculated keeping the obtained value of Re‟ and the original value of Cm as follows in Equation 3.3. (3.3) The calculated values were used to build the equivalent electrical model 2R-1C for each application. Each model was implemented with a resistor as Re, a capacitor as Cm and a potentiometer for Ri with the purpose to tune the value to keep original Fc of the EBI application. 2 AD5933+4E-AFE application customization: Once the values for each of the models have been obtained, the AD5933+4E-AFE was customized by adjusting the values of the following resistors: Rrfb, R4 and applying (3.4) and (3.5). Figure 3.11 2R1C parallel Bridge for f=0 36 From the AC Analysis: (3.4) From the DC Analysis: (3.5) In order to make the most of the dynamic range of the ADC of the sensing stage of the AD5933, a DC bias voltage was introduced at the input with the value. Solving (3.6) for it was found that the values for Rrfb and R4 and must be equal. (3.6) (3.7) (3.8) (3.9) By means of the AC analysis, an amplitude voltage at the input of the sensing stage of 1.5 Volts was applied and the resistance RG1, the gain resistance of INA111 on the 4E-AFE, was calculated with the Equation 3.10: (3.10) From the analysis, it was obtained that several parameters have common values for all the EBI applications. The Table 3.2 lists the parameters with the corresponding values common for the EBI applications: TABLE 3.2 Rcc1 Resistor of the voltage-current converter amplifier 7.5KΩ R4 Resistor current-voltage 10 KΩ Rrfb Feedback resistor 10 KΩ Vi Vac injection 1.98V Vdd Vdc supply 3V 37 This way RG1 is the only parameter left that must be adjusted specifically to each application. Table 3.3 lists each of the obtained values for RG1 for each EBI application. 3.2.3 EBI Application Equivalent Spectroscopy Measurements Hence, with the optimization of the system AD5933+4E-AFE already done, the impedance measurements for each model were made with both devices: the SFB7 and the AD5933+4E-AFE. 3.2.3.1. SFB7 measurements. 100 spectroscopy measurements have been taken in the frequency range from 3kHz to 1MHz for each of the equivalent circuits. Each obtained spectrum was saved in a text file, which contains the measurement frequency and both the measured resistance and reactance. The SFB7 distributes the measurement frequencies, exponentially distributing 256 measurement points between 3 kHz to 1 MHz. See Figure 3.12. This frequency distribution dedicates more measurement points to lower frequencies than for higher frequencies. TABLE 3.3 APPLICATION RG1(Ω) TBC 6619 Arm-Arm 2851 Leg-Leg 3858 Trunk-Trunk 801 LC 654 RR 481 38 3.2.3.2 AD5933+4-AFE measurements. 100 spectroscopy measurements have been taken in the frequency range from 3kHz to 100KHz. Each obtained spectrum was saved in a CSV file, which contains the measured frequencies, impedance, phase, resistance and reactance. AD5933 distributes the measurement frequencies lineally distribution 501 measurements points between 3kHz and 100KHz. See Figure 3.13. This frequency distribution dedicates the same number of point to high and low frequencies because the sample points are equidistant. Figure 3.12 Frequency distribution of SFB7 y = 3000,3e0,0227x R² = 1 0 200.000 400.000 600.000 800.000 1.000.000 1.200.000 1 11 21 31 41 51 61 71 81 91 101 111 121 131 141 151 161 171 181 191 201 211 221 231 241 251 Frequency(Hz) SFB7 SFB7 Exponential Regression (SFB7) Figure 3.13 Frequency distribution of AD4933+4-AFE y = 194x + 2806 R² = 1 0 20000 40000 60000 80000 100000 120000 1 21 41 61 81 101 121 141 161 181 201 221 241 261 281 301 321 341 361 381 401 421 441 461 481 501 Frequency(Hz) AD5933+4-AFE AD5933+4-AFE Linear Regression(AD5933+4-AFE) 39 All the obtained measurements, from both spectrometers, were processed with BioImp Software. In order to process the measurements taken by the AD5933+4E-AFE, a Matlab script was done in MATLAB to prepare the data files generated by the software controlling the AD5933+4E-AFE [31] to be imported by BioImp Software. In addition to the file conversion a specific file type definition was declared in BioImp to read the files as shown in Figure 3.14. 3.2.4 Performance Comparison The performance comparison was done based on the error obtained from the model parameters estimation, which was done from the impedance measurements from both systems, according to the work flow diagram showed in the Figure 3.15. Figure 3.14 File definition Impedimed Fig 3.15 Performance comparison Process Overview Model Parameter Estimation •BioImp batch processing Statistical Error Analysis •MS Excel •Average •Standarddeviation •Maximum and Minimum •Absolute and Relative Error Estimation Error Visualization •Matlab Plot 40 3.2.4.1 Model Parameter Estimation. Once the files containing the 2R1C-models impedance measurements were uploaded into BioImp, the data files were batch processed to estimate the 2R1C parameters, see Figure 3.16. The estimation results from each measurement were saved into MS Excel format, see Figure 3.17. This way for each EBI application model two files containing the estimated parameters were obtained from the measurements taken with the SFB7 and the AD5933+4-AFE spectrometer. Figure 3.16 Batch results of Ri, Re and Cm 41 3.2.4.2 Statistics of the Model Parameter Estimation. Each of the exported files from BioImp was processed with MS Excel to obtain the following statistical parameters from the values of Re, Ri, Cm and Fc: o Average o Standard deviation o Maximum and Minimum o Absolute and Relative Error Note that the errors have been calculated comparing the estimated values with the real value of the passive components used to build the 2R1C. 3.2.4.3 Estimation Error Visualization. To obtaine the plots with the comaprison results, the statistical values have been exported to a workspace of MatLab. Once in Matlab the average, maximum, minimum of the estimated values and the value of the component of each EBI model application have been plotted. Figure 3.17 Data output file export options of BioImp 48 4.4. Spectroscopy Measurements in 2R1C Models Once the equivalents electrical circuits and the optimization of AD5933+4E-AFE have been already done, the impedance measurements for each model were made by both, the device SFB7 and the AD5933+4E-AFE. The obtained measurements with each of the spectrometers for each application are shown in the follow figures. 4.4.1 TBC: Figure 4.7 contains the plots of the reactance and resistance spectrum of the TBC application obtained from the SFB7 and AD5993+4-AFE. In the resistance spectrum it is possible to observe that the maximum and minimum values are approximately 900 Ω and 390 Ω respectively, providing us with a dynamic range of approximately 510 Ω. As the graphs denote the curves, resistance and reactance for both devices are overlapped, which means that the measurements of SFB7 and AD5933+4-AFE are very similar. It is also possible to observe that at low frequencies, between 3kHz and 6KHz, the device AD5933+4-AFE produce certain dispersion on the measurement. Figure 4.7 the left part of the figure the spectra resistance and the right part the spectra reactance are plotted measured from, SFB7 and AD5993+4-AFE. In the reactance is possible to observe that there is only one dominant dispersion with a Fc value approximately of 30 kHz 49 4.4.2 RR: In Figure 4.8 the reactance and resistance spectra of the TBC application for SFB7 and AD5993+4-AFE are plotted. In the resistance curve is possible to observe that the maximum and minimum values are approximately 57 Ω and 16 Ω respectively, providing us with a dynamic range of approximately 41 Ω. As the graphs denote the curves, resistance and reactance for the both devices are overlapped, therefore it means that the measurements of SFB7 and AD5933+4- AFE are very similar. It is also possible to observe that at low frequencies, between 3kHz and 6KHz, the measurements performed with the device AD5933+4-AFE present a dispersion. 4.4.3 LC: In Figure 4.9 the reactance and resistance spectra of the TBC application for SFB7 and AD5993+4-AFE are plotted. In the resistance curve is possible to observe that the maximum and minimum values are approximately 80 Ω and 18 Ω respectively, providing us with a dynamic range of approximately 62 Ω. As the graphs denote the curves, resistance and reactance for the both devices are overlapped, therefore it means that the measurements of SFB7 and AD5933+4- AFE are very similar. It is also possible to observe that at low frequencies, between 3KHz and 6KHz, the measurements performed with the device AD5933+4-AFE present a dispersion. Figure 4.8 the left part of the figure the spectra resistance and the right part the spectra reactance are plotted measured from both, SFB7 and AD5993+4-AFE. In the reactance is possible to observe that there is only one dominant dispersion with a Fc value approximately of 27 kHz 50 4.4.4 AA: In the Figure 4.10 is plotted the reactance and resistance of the TBC application for SFB7 and AD5993+4-AFE. Figure 4.9 the left part of the figure the spectra resistance and the right part the spectra reactance are plotted measured from both, SFB7 and AD5993+4-AFE. In the reactance is possible to observe that there is only one dominant dispersion with a Fc value approximately of 35 kHz Figure 4.10 the left part of the figure the spectra resistance and the right part the spectra reactance are plotted measured from both, SFB7 and AD5993+4-AFE. In the reactance is possible to observe that there is only one dominant dispersion with a Fc value approximately of 35 kHz 51 In the resistance curve is possible to observe that the maximum and minimum values are approximately 365 Ω and 185 Ω respectively, providing us with a dynamic range of approximately 180 Ω. As the graphs denote the curves, resistance and reactance for the both devices are overlapped; therefore it means that the measurements of SFB7 and AD5933+4-AFE are very similar. It is also possible to observe that at low frequencies, between 3kHz and 6KHz, the measurements performed with the device AD5933+4-AFE present a dispersion. 4.4.5 LL: In the Figure 4.11 is plotted the reactance and resistance of the TBC application for SFB7 and AD5993+4-AFE. In the resistance curve is possible to observe that the maximum and minimum values are approximately 500 Ω and 240 Ω respectively, providing us with a dynamic range of approximately 260 Ω. As the graphs denote the curves, resistance and reactance for the both devices are overlapped; therefore it means that the measurements of SFB7 and AD5933+4- AFE are very similar. It is also possible to observe that at low frequencies, between 3KHz and 6KHz, the measurements performed with the device AD5933+4-AFE present a dispersion. 4.4.6 TT: In the Figure 4.12 is plotted the reactance and resistance of the TBC application for SFB7 and AD5993+4-AFE. Figure 4.11 the left part of the figure the spectra resistance and the right part the spectra reactance are plotted measured from both, SFB7 and AD5993+4-AFE. In the reactance is possible to observe that there is only one dominant dispersion with a Fc value approximately of 25 kHz 52 In the resistance curve is possible to observe that the maximum and minimum values are approximately 98 Ω and 31 Ω respectively, providing us with a dynamic range of approximately 67 Ω. As the graphs denote the curves, resistance and reactance for the both devices are overlapped, therefore it means that the measurements of SFB7 and AD5933+4-AFE are very similar. It is also possible to observe that at low frequencies, between 3kHz and 6kHz, the measurements performed with the device AD5933+4-AFE present a dispersion. 4.4.7 Comparison among the Theoretical Model Values and the Estimated Model Values With one of the measurements taken by each device, SFB7 and AD5933 +4- AFE, and once the data were processed with BioImp software to obtain the value of Re, Ri, Cm and Fc then Figure 4.13 has been produced. It shows the comparison between the estimated model values Re, Ri, Cm and the characteristic frequency Fc obtained for each application and device with the theoretical measured values, which are enlisted in Table 4.5. TABLE 4.5 THEORETICAL VALUES 2R1C COMPONENTS Application Re(Ω) Ri(Ω) Cm(nF) Fc(kHz) TBC 933 835 2.99 30.1 RR 59.4 21.5 77.1 25.5 LC 81 23.4 46.2 33 AA 369.75 354 6.37 34.5 TT 99.6 34 46.5 25.6 LL 511.2 440 6.61 25.3 53 In Figure 4.13 it is possible to observe the calculated value and the obtained value from the measurements of both devices are quite similar, the maximum difference is in the value of Ri, between the calculated value and obtained value from the devices in the TBC application. 4.5. SFB7 Vs. AD5933+4-AFE In order to show the performance comparison between the SFB7 and the AD5933+4-AFE the following statistical parameters were chosen: - Absolute and Relative Error - Average - Standard deviation - Maximum and Minimum Figure 4.13 This graph shows in a) the membrane capacitance, in b) the intracellular resistance, in c) the extracellular Resistance and in d) the characteristic frequency per each application and measured with, both devices the SFB7 and AD5933+4-AFE, and also the theoretical value calculated. 54 Therefore, with all the measurements already processed with BioImp software, all the estimation results were exported to a MS Excel file. Once all the data were exported to MS Excel, the statistical values were calculated with the statistical tools available within MS Excel. Firstly, the absolute and relative error comparisons between SFB7 and AD5933+4-AFE were done. Figure 4.14 and 4.15 show both errors. Figure 4.14 Relative error of Ri, Re, Cm and Fc for each application and device: SFB7 and AD5933+4-AFE. 55 As it is possible to see in Figure 4.14 in all the calculated values for each application the relative error produced by the AD5933+4-AFE is slightly bigger than the produced by the SFB7. The maximum relative error with regarding to the real value of each parameter is about of 8% and it occurs in the TT application, also the average value of the relative error is approximately around 2%. The absolute error per application and circuital value of the used model is shown in Figure 4.15 and it was shown in the case of the relative error but in the respective units that the error obtained with the AD5933+4-AFE is larger than the error obtained with the SFB7. The maximum absolute error in the case of the calculated value from the impedance measurements is for the resistors about 10 Ω, for the capacitor about 4pF and for Fc is 2kHz Secondly, the standard deviation was calculated also with MS Excel. Figure 4.16 shows the comparison of the standard deviation obtained with each of the devices. Fig 4.15 Absolute error of Ri, Re, Cm and Fc per each application and device (SFB7 and AD5933+4-AFE) 56 Although the relative error obtained with the AD5933+4-AFE is most of the time larger than the error obtained with the SFB7, as shown in Figure 4.15, the standard deviation of the SFB7 is larger than the standard deviation produced by the AD5933+4-AFE in all the circuital values calculated in each of the applications. Finally in Matlab, the average, maximum, minimum of the estimated values and the real value of the component of each EBI model application have been plotted on Figures 4.17, 4.18, 4.19, 4.20, 4.21 and 4.22. Figure 4.16 Standard deviation of Ri, Re, Cm for each application and impedance spectrometer device: SFB7 and AD5933+4-AFE. 57 Figure 4.17 the maximum, minimum, average and real values of the 2R1C model (Re, Ri and Cm) and the characteristic frequency (Fc) for the segmental body composition Arm-Arm application. Figure 4.18 the maximum, minimum, average and real values of the 2R1C model (Re, Ri and Cm) and the characteristic frequency (Fc) for the Respiration Rate application. 64 CHAPTER 6 CONCLUSIONS & FUTURE WORK 6.1. Conclusions After all the results shown in the chapter 4 and discussion in chapter 5, it can be concluded that the combination AD5933+4-AFE is suitable to realize spectroscopy measurements of the EBI applications studied in this master thesis. This result has significant repercussions as in economic terms due to the high integration of AD5933 (I2C interface, DDS, Amplifiers, PGA, Low pass Filter, ADC, DSP and Temperature sensor) as in the electronic instrumentation because it opens up new horizons for the integration of the measurement systems of EBI into portable and even wearable systems. 6.2. Future work Once the main target of this work has been done, with the aim to solve some problems or to improve the applied method and the system AD5933+4-AFE the next points has brought up:  The α parameter was chosen arbitrarily and due to it all the models have been affected. Therefore, in order to improve this step in the estimation a study in several subjects should be realized to reach better approximation to α parameter by empiric way.  The frequency distribution of measurements performed by the SFB7 and AD5933+4- AFE is exponential and linear, respectively. With the goal of improving the comparison between both devices, it will be better to change by software the frequency distribution of the AD5933.  To use resistors and capacitors with a low tolerance, in order to obtain a lower variability between the models measurements  To add a larger delay at low frequencies or adding a filter in the AFE to reduce de dispersion at low frequencies.  Due to the spectral variability of the impedance under study in each application and to make suitable the combination AD5933+4-AFE for many applications, it will be a good improvement to add a differential PGA (Programmable Gain Amplifier) or scalable resistors array. 65 REFERENCES [1] S. Grimnes and Ø. G. Martinsen, Bioimpedance & Bioelectricity Basics: Academic Press, 2000. [2] A. Piccoli, B. Rossi, L. Pillon, and G. Bucciante, "A new method for monitoring body fluid variation by bioimpedance analysis: the RXc graph," 1994. [3] F. Seoane, "Electrical Bioimpedance Cerebral Monitoring: Fundamental Steps towards Clinical Application," in Signal & Systems. vol. PhD Göteborg: Chalmers University of Technology, 2007, p. 154. [4] M. Phillippson, "Les Lois De La Resistance Electrique Des Tissus Vivants," Bull. Acad. roy. Belgique, 1921. [5] H. Fricke, "A Mathematical Treatment of the Electrical conductivity of Colloids and Cell Suspensions," Journal of General Physiology, pp. 375-383, Jan 4 1924. [6] K. S. Cole, "Electric impedance of suspensions of spheres," Journal of General Physiology, vol. 12, pp. 29-36, 1928. [7] J. Nyboer, S. Bango, A. Barnett, and R. H. Halsey, "Radiocardiograms: Electrical impedance changes of the heart in relation to electrocardiograms and heart sounds," J. Clin. Invest., vol. 19, p. 773, 1940. [8] T. Olsson and L. Victorin, "Transthoracic impedance, with special reference to newborn infants and the ratio air-to-fluid in the lungs," Acta Paediatr Scand Suppl, vol. 207, p. Suppl 207:1ff, 1970. [9] R. F. Kushner and D. A. Schoeller, "Estimation of total body water by bioelectrical impedance analysis," Am J Clin Nutr, vol. 44, pp. 417-24, Sep 1986. [10] P. Aberg, I. Nicander, J. Hansson, P. Geladi, U. Holmgren, and S. Ollmar, "Skin cancer identification using multifrequency electrical impedance – A potential screening tool," IEEE Trans. Bio. Med. Eng., vol. 51, pp. 2097-2102, 2004. [11] H. P. Schwan, "Electrical properties of tissue and cell suspensions," Adv Biol Med Phys, vol. 5, pp. 147-209, 1957. [12] H. Morel and M. Y. Jaffrin, "Total body water measurement: using the multifrequency BIS-Hanai approach with 50 kHz single frequency," in 13th International Conference on Electrical Bioimpedance and the 8th Conference on Electrical Impedance Tomography, 2007, pp. 799-802. [13] A. De Lorenzo, A. Andreoli, J. Matthie, and P. Withers, "Predicting body cell mass with bioimpedance by using theoretical methods: a technological review," J Appl Physiol, vol. 82, pp. 1542-58, May 1997. [14] U. Kyle, I. Bosaeus, A. D. De Lorenzo, P. Deurenberg, M. Elia, J. M. Gomez, B. L. Heitmann, L. Kent-Smith, J. Melchior, M. Pirlich, H. Scharfetter, A. M. W. J. Schols, and C. Pichard, "Bioelectrical impedance analysis--part I: review of principles and methods," 2004. [15] M. Y. Jaffrin and H. Morel, "Body fluid volumes measurements by impedance: A review of bioimpedance spectroscopy (BIS) and bioimpedance analysis (BIA) methods," Medical Engineering & Physics, vol. 30, pp. 1257-1269, 2008. [16] G. B. Forbes, Human body composition : growth, aging, nutrition, and activity / Gilbert B. Forbes: Springer-Verlag, 1987. 66 [17] F. J. Deurenberg P Fau - Schouten and F. J. Schouten, "Loss of total body water and extracellular water assessed by multifrequency impedance." [18] U. G. Kyle, I. Bosaeus, A. D. De Lorenzo, P. Deurenberg, M. Elia, J. M. Gomez, B. L. Heitmann, L. Kent-Smith, J. C. Melchior, M. Pirlich, H. Scharfetter, A. M. W. J. Schols, and C. Pichard, "Bioelectrical impedance analysis--part I: review of principles and methods," Clin Nutr, vol. 23, pp. 1226-43, Oct 2004. [19] R. Patel, E. L. Peterson, N. Silverman, and B. J. Zarowitz, "Estimation of total body and extracellular water in post-coronary artery bypass graft surgical patients using single and multiple frequency bioimpedance," 1996. [20] W. Hannan, S. J. Cowen, C. E. Plester, K. C. Fearon, and A. Beau, "Comparison of bioimpedance spectroscopy and multi-frequency bio-impedance analysis for the assessment of extracellular and total body water in surgical patients," 1995. [21] K. S. Cole, "Permeability and impermeability of cell membranes for ions.," Quant. Biol., vol. 8, pp. 110–122, 1940. [22] T. Hanai, N. Koizumi, and R. Gotoh, "Dielectric properties of emulsions," Colloid & Polymer Science, vol. 184, pp. 143-148, 1962. [23] P. Nopp, N. Harris, T. Zhao, and B. Brown, "Model for the dielectric properties of human lung tissue against frequency and air content," Medical and Biological Engineering and Computing, vol. 35, pp. 695-702, 1997. [24] L. A. Baker Le Fau - Geddes, H. E. Geddes La Fau - Hoff, C. J. Hoff He Fau - Chaput, and C. J. Chaput, "Physiological factors underlying transthoracic impedance variations in respiration." [25] L. E. Baker, L. A. Geddes, and H. E. Hoff, "quantitative evaluation of impedance spirometry in man," The American journal of medical electronics, 1965. [26] A. F. Pacela, "Impedance pneumography--a survey of instrumentation techniques." [27] A. Fein, R. F. Grossman, J. G. Jones, P. C. Goodman, and J. F. Murray, "Evaluation of transthoracic electrical impedance in the diagnosis of pulmonary edema," Circulation, 1979. [28] F. Larsen, L. Mogensen, and B. Tedner, "Transthoracic electrical impedance at 1 and 100 kHz--a means for separating thoracic fluid compartments?," Clinical Physiology, 1987. [29] F. Seoane, J. Ferreira, J. J. Sanchéz, and R. Bragós, "Analog Front-End Enables Electrical Impedance Spectroscopy System On-Chip for Biomedical Applications," Physiol. Meas. , vol. 29 pp. S267-S278, 2008. [30] H. Fricke and S. Morse, "The Electric Resistance and Capacity of Blood for Frequencies Between 800 and 4.5 Million Cycles," Journal of General Physiology, vol. 9, pp. 153-67, 1925. [31] J. Ferreira and J. J. Sanchez, "Electrical Bioimpedance Measurement System for Limb Edema Monitoring," in School of Engineering. vol. MSc Borås: University College of Borås, 2007, p. 104. [32] H. Scharfetter, P. Hartinger, H. Hinghofer-Szalkay, and H. Hutten, "A model of artefacts produced by stray capacitance during whole body or segmental bioimpedance spectroscopy," Physiological Measurement, vol. 19, pp. 247-261, 1998. [33] J. Taylor, An Introduction to Error Analysis: University Science Books, 1997. [34] M. Balleza, N. Calaf, T. Feixas, M. González, D. Antón, P. J. Riu, and P. Casan, "Medición del patrón ventilatorio mediante tomografía por impedancia eléctrica en pacientes con EPOC " Archivos de bronconeumología, vol. 45 núm. 07, pp. 320-324, 2009. 67 [35] M. Balleza, J. Fornos, N. Calaf, T. Feixas, M. González, D. Antón, P. J. Riu, and P. Casan, "Seguimiento del patrón ventilatorio en reposo mediante tomografía por impedancia eléctrica " Archivos de Bronconeumología, vol. 43, núm 6, pp. 300-303, 2007. [36] L. Nescolarde, J. Rosell, and T. Doñate, "Relationship between segmental and wholebody phase angle in peritoneal dialysis patients," Physiological Measurement, vol. 29, p. N49, 2008.