Lenstar LS900 vs EchoScan US-800: comparison between optical and ultrasound biometry with and without contact lenses and its relationship with other biometric parameters

Noya Padín, Verónica; García Queiruga, Jacobo; Iacubitchii, Maria; Giráldez Fernández, María Jesús; Yebra-Pimentel Vilar, Eva; Pena Verdeal, Hugo

Abstract

Background: Due to the increasing use of contact lenses (CL) and the interest in ocular and body size relationships, this study aimed to compare measurements from two biometers (contact ultrasonic EchoScan US-800 and non-contact optical Lenstar LS900) with and without CL and to explore the relationship between ocular and body biometric parameters. Design and Methods: This cross-sectional study measured ocular biometry using two biometers along with their body height and right foot length in 50 participants. Differences between biometry data from the two devices were compared and correlations between ocular and body biometric values were analyzed. Results: All parameters showed interbiometric differences (p≤0.030), except crystalline lens thickness during CL wear (p=0.159). Comparing measurements with and without CL, differences were observed in axial length (p<0.001), vitreous length measured by optical biometer (p=0.016), and anterior chamber depth by ultrasonic biometer (p<0.016). Lens thickness remained unaffected (p≥0.190). Body height and foot length were correlated with anterior chamber depth, vitreous length, and axial length (p≤0.019, r≥0.330). Most biometric parameters were correlated among them using both devices (p≤0.037, r≥0.296). Conclusions: These biometers are not interchangeable and CL affects measurements. Body height and foot length correlate with ocular dimensions, and most ocular biometric values correlate positively.

Full text

1 ABSTRACT 1 Background: Due to the increasing use of contact lenses (CL) and the interest in ocular and body size 2 relationships, this study aimed to compare measurements from two biometers (contact ultrasonic EchoScan 3 US-800 and non-contact optical Lenstar LS900) with and without CL and to explore the relationship 4 between ocular and body biometric parameters. 5 Design and Methods: This cross-sectional study measured ocular biometry using two biometers along with 6 their body height and right foot length in 50 participants. Differences between biometry data from the two 7 devices were compared and correlations between ocular and body biometric values were analyzed. 8 Results: All parameters showed interbiometric differences (p≤0.030), except crystalline lens thickness 9 during CL wear (p=0.159). Comparing measurements with and without CL, differences were observed in 10 axial length (p<0.001), vitreous length measured by optical biometer (p=0.016), and anterior chamber depth 11 by ultrasonic biometer (p<0.016). Lens thickness remained unaffected (p≥0.190). Body height and foot 12 length were correlated with anterior chamber depth, vitreous length, and axial length (p≤0.019, r≥0.330). 13 Most biometric parameters were correlated among them using both devices (p≤0.037, r≥0.296). 14 Conclusions: These biometers are not interchangeable and CL affects measurements. Body height and foot 15 length correlate with ocular dimensions, and most ocular biometric values correlate positively. 16 17 Key words: Ocular biometry, Lenstar LS900, EchoScan US800, Contact Lenses, Human morphology. 18 19 HIGHLIGHTS 20 1. The present study analyzes the effect of CL wear during ocular biometry measurements. 21 2. CL should not be worn when performing biometry, both with optical and ultrasound biometry. 22 3. Results obtained by optical and ultrasonic biometry are not interchangeable. 23 4. Biometric parameters such as body height or foot length correlate positively with ocular 24 dimensions. 25 2 1. INTRODUCTION 26 Ocular biometry enables clinicians to measure the dimensions of the components of the eye with specially 27 designed biometers [1]. Biometric measurements can be performed based on different physical principles, 28 with acoustic (ultrasound) and optical methods being two of the most commonly used [2,3]. Ultrasound 29 biometry uses a high-frequency sound wave to measure the length of the internal structures of the eye, as 30 well as its total length, by capturing the echoes as the sound bounces off the ocular structures [1,4]. There 31 are two types of ultrasound biometry depending on the location of the probe: the direct contact and the 32 immersion contact methods. In direct contact biometry, the probe is placed directly on the cornea after the 33 instillation of topical anesthesia. In immersion biometry, a capsule filled with saline solution is placed 34 between the patient’s cornea and the probe. To avoid discomfort, the use of an anesthetic is also necessary 35 [1,4]. Optical biometry allows non-contact biometric measurements, and there are also two techniques: 36 partial coherence interferometry and optical low-coherence reflectometry, both based on laser 37 interferometry. The partial coherence interferometry uses a multimode laser diode reflecting off the corneal 38 surface and the retinal pigment epithelium to measure axial length; it can also be combined with other 39 technologies such as a Scheimpflug camera device to obtain more ocular parameters [5]. The optical low- 40 coherence reflectometry uses a super luminescent diode, which has stronger penetrating power and is 41 capable of acquiring more parameters than partial coherence interferometry alone [1,6]. 42 Since its inception in the 1970s, the main uses of ocular biometry have been for the calculation of the 43 intraocular lens power in cataract or refractive surgery, or the assessment of pathological ocular disorders 44 [1,7]. Additionally, its use has become increasingly widespread in monitoring the progression of myopia in 45 the last decades due to the increase in availability of methods in the market for its control [8,9]. While non- 46 contact devices are of great interest in the adult population, those devices have a particular interest in 47 myopia control of the pediatric population as many children have difficulty with invasive testing; also 48 myopia may be better studied if biometry can be performed while wearing myopia control soft contact lens 49 (CL) in order to avoid as much as possible the manipulation of the eye during an examination [10]. 50 Moreover, the ongoing increase in the prevalence of myopia has encouraged studies that relate ocular 51 3 parameters, such as lens thickness (LT), anterior chamber depth (ACD), vitreous length (VL) or axial length 52 (AL), to other body parameters [11,12]; for example, a positive correlation has been observed between AL 53 and body height, which means the taller the subject, the greater the AL [13], and between the relative value 54 of foot length to body height and AL in men [14]. 55 Regarding the devices studied here, the EchoScan US-800, A-scan ultrasound has been the standard 56 technique for measuring ocular parameters for decades [15], although in recent years, optical devices have 57 begun to gain relevance, especially for use in children [9]. Among these devices, the Lenstar LS900 has 58 demonstrated excellent reproducibility and repeatability in previous studies [16,17]. The present study 59 aimed: 1) to compare the values obtained between two biometers, the contact ultrasound EchoScan US-800 60 (Nidek Co., Tokyo, Japan) and the low coherence optical reflectometry Lenstar LS900 (Haag Streit AG, 61 Köniz, Switzerland), with the naked eye and during CL wear, and 2) to assess the relationship between 62 biometric parameters and other body parameters (body height and foot length). 63 64 2. METHODS 65 2.1 Participants 66 A total of 51 participants (13 men and 38 women) with a mean age of 22.2 ± 2.25 years (range from 18 to 67 28 years) were randomly recruited among university students. Participants were included if they had an 68 auto-refracted sphere of - 10.00 to + 4.00D, astigmatism up to - 3.00D, anterior chamber angle ≥ 30º and a 69 compensated intraocular pressure (IOP) ≤ 20.85 mmHg [2,18,19]. Patients were excluded if they had ever 70 been diagnosed with an ocular infection, trauma disorders, diseases at the time of the study (glaucoma, 71 scleral or corneal anomalies, dry eye disease, meibomian gland dysfunction, etc.), undergone ocular 72 surgery, diagnosed with systemic disorders that could affect the measurements (diabetes mellitus, 73 rheumatoid arthritis, etc.), or had a history of hypersensitivity or allergy to anesthesia [2,10,20]. No 74 participant was under any type of topical, systemic treatment or used artificial tears at the time of the study. 75 All participants gave their written informed consent to be included in the study. The study protocol followed 76 4 the tenets of the Declaration of Helsinki and was approved by the Ethics Committee of the university 77 (Approval number: USC 04/2022). 78 79 2.2 Experimental procedure 80 In a single visit, participants underwent a battery of tests according to standard protocols always in the same 81 order based on the invasiveness of the test [10,21]: pachymetry, topography, anterior chamber angle, IOP, 82 objective refractive condition, biometry of the naked eye, body height, foot length, and ocular biometry 83 wearing CL. 84 85 2.2.1 Measurements of inclusion criteria parameters 86 Pachymetry, topography and anterior chamber angle were performed by the Visionix120 system (Visionix 87 Luneau Technologies, Chartres, France) [22]. The Visionix120 is a non-invasive optical diagnostic device 88 that uses a Scheimpflug image-based system and a Placido disk system projected on the cornea to assess 89 the anterior segment parameters of the eye. 90 IOP was measured by the Canon TX-10 non-contact tonometer (Canon Inc, Tokyo, Japan) in automatic 91 mode. The device automatically performs three measurements on each eye to avoid fluctuations due to the 92 cardiac pulse cycle [23]. The pachymetry values obtained by Visionix120 were used to calculate the 93 compensated IOP [24]. 94 The objective refractive condition of the eye was determined using the Shin-Nippon NVISION-K5001 95 open-field window autorefractometer (Rexxam Co., Kagawa, Japan), which allows a binocular view of a 96 fixation target at a distance of 4 meters, minimizing instrumental myopia [25]. This device estimates the 97 refractive error by averaging three measurements. 98 99 2.2.2 Measurements of studied parameters 100 The participant’s body height and the length of their right foot were both measured using measuring tapes 101 with 0.1 cm divisions. One tape was attached to the wall and another tape was attached to the floor [26]. 102 5 The ocular biometry was measured exclusively in the right eye of each participant to avoid both the effects 103 of overstating the precision of statistical estimates using two different devices, each twice: a non-contact 104 biometer based on low coherence optical reflectometry as its measurement principle [27], and a contact 105 ultrasound biometer [21]. The optical biometer was set to perform three automatically consecutive 106 measurements once the operator manually starts the process to calculate the final value as the average of 107 the values obtained. The ultrasound device was set at AUTO3 mode, which automatically performs a battery 108 of measurements when the probe touches the cornea and chooses the closest three measures to obtain the 109 average value. The first measurement was performed with the optical biometer and the second with the 110 ultrasonic device, following instillation of one drop of anesthesia. This order was decided upon in order to 111 follow a protocol from least to most invasive to avoid errors induced by corneal compression during contact 112 biometry [21]. 113 After performing ocular biometry measurements with the naked eye using both biometers, and allowing the 114 anesthetic effect to wear off, a CL was fitted in the right eye of each participant, and the biometry was 115 repeated with each device in the same order [10]. In all cases, a daily disposable silicone hydrogel CL of 116 Delefilcon A (Alcon Laboratories, Inc, Fort Worth, USA) at - 3.00D of power was used [28]. 117 118 2.3 Statistical analysis 119 Data analysis was performed using SPSS statistical software v.25.0 for Windows (SPSS Inc., Chicago, 120 USA). The chosen level of statistical significance was p ≤ 0.05. Prior to the analysis, the Shapiro-Wilk test 121 was performed for all variables to detect deviations from the normal distribution [29]. Since the variables 122 showed a non-normal distribution, non-parametric tests were performed in the data analysis. Differences 123 between intra- and inter-biometers data with the naked eye and during CL wear were analyzed using the 124 Wilcoxon paired-sample test. For graphical purposes, Bland-Altman procedures were used; this method 125 describes the correlation or similarity between two variables, representing averages versus differences 126 [29,30]. The 95% limits of agreement (95% LoAs) were calculated (Mean difference ± 1.96 x Standard 127 Deviation (SD)), as well as the exact 95% Confidence Intervals (95% CI) for Upper and Lower LoAs 128 6 considered as a pair (Mean difference ± ct0.025 x SD; Mean difference ± ct0.975 x SD) [29,31]. The relationship 129 between ocular biometry values and other body parameters (body height and foot length) was performed 130 with Spearman’s correlation test. Correlation between variables was described as weak (0.20 – 0.40), 131 moderate (0.41 – 0.60), good (0.61 – 0.80) or strong (0.81 – 1.00) [32]. 132 133 3. RESULTS 134 Based on the inclusion criteria, the final sample consisted of 50 participants (13 men and 37 women). One 135 participant was excluded for having an intraocular lens. The descriptive data of the final sample are shown 136 in Table 1. 137 138 3.1 Differences between biometers data with the naked eye 139 A statistically significant difference was found between all parameters when comparing the results obtained 140 by the optical vs. the ultrasound device with the naked eye (Wilcoxon paired-sample test, all p ≤ 0.030) 141 (Table 2A). The Bland and Altman plot of means versus differences between the measurements with the 142 naked eye by the devices studied is shown in Figure 1. In the ACD and AL plots (Figure 1A and 1D, 143 respectively), no discernible trend is observed, indicating an absence of systematic bias in the 144 measurements; however, the variability between devices is relatively high in both cases. On the other hand, 145 the LT and VL plots (Figure 1B and 1C, respectively) exhibit two separate data sets: whereas the variability 146 within each set is small, the between-set variability is large (around 4 mm in both plots), indicating the 147 presence of an external factor that is affecting one of the sets. 148 149 3.2 Differences between biometers data wearing CL 150 Statistically significant differences between biometers were observed during CL wear in the ACD, VL and 151 AL values (Wilcoxon paired-sample test, all p ≤ 0.001), but not in LT (Wilcoxon paired-sample test, p = 152 0.159) (Table 2B). The Bland and Altman plot of means versus differences between the devices studied is 153 shown in Figure 2. Similar to Figure 1, the plots for ACD and AL (Figure 2A and 2D) exhibit a no 154 7 discernible trend, and the variability is relatively smaller in these cases than in the previous analysis. 155 Meanwhile, the plots for LT and VL (Figure 2B and 2C, respectively) still display two distinct data sets, 156 suggesting that the factor affecting one of the sets in the naked eye measurements persists in the 157 measurements taken with the CL on eye. 158 159 3.3 Differences between naked eye data and CL wear data measured by the optical biometer 160 A statistically significant difference between VL and AL measured by the optical biometer with the naked 161 eye vs. during CL wear was found (Wilcoxon paired-sample test, all p ≤ 0.016), while ACD and LT with 162 the naked eye and during CL wear were not statistically different (Wilcoxon paired-sample test, all p ≥ 163 0.190) (Table 2C). The Bland and Altman plot of means versus differences between the devices studied is 164 shown in Figure 3. The plots for ACD, LT, and VL (Figure 3A, 3B, and 3C, respectively) show low 165 variability in measurements with and without CL, indicating that any observed differences are not 166 statistically significant or, if significant, are not clinically relevant. In contrast, the AL plot (Figure 3D) 167 exhibits higher and more pronounced variability in comparison with the other parameters studied. 168 169 3.4 Differences between naked eye data and CL wear data measured by the ultrasound biometer 170 When comparing biometric data with the naked eye vs. during CL wear measured by the ultrasound 171 biometer, statistically significant differences were observed in ACD and AL (Wilcoxon paired-sample test, 172 all p ≤ 0.005), while not in LT and VL (Wilcoxon paired-sample test, all p ≥ 0.395) (Table 2D). The Bland 173 and Altman plot of means versus differences between the devices studied is shown in Figure 4. The ACD 174 plot (Figure 4A) displays a scatter of points with an increasing trend, from which can be inferred that the 175 variability in the differences is greater at the extremes of the scale. For LT and VL plots (Figure 4B and 176 4C), most of the data points are close to zero, indicating small differences between values. However, the 177 wide 95% LoAs suggest high variability. Finally, the AL plot (Figure 4D) exhibits a scatter of points with 178 no discernible trend but high variability. 179 180 8 3.5 Correlation between body height, foot length and ocular biometry data measured by the optical 181 biometer 182 The relationship between ocular biometric data measured by the optical biometer (naked eye) and body 183 height was statistically significant with ACD, VL and AL values (Spearman’s correlation, r ≥ 0.338, all p 184 ≤ 0.016); similar results were obtained for the correlation between biometric data measured by this device 185 and foot length, where ACD, VL and AL showed a statistically significant correlation (Spearman’s 186 correlation, r ≥ 0.330, all p ≤ 0.019) (Table 3). In all cases the correlation was weak. 187 Regarding the correlations of the optical biometric data with each other measured by the optical biometer, 188 results showed a strong correlation between VL and AL (Spearman’s correlation, r = 0.986, p < 0.001), a 189 moderate correlation between ACD and LT (Spearman’s correlation, r = - 0.568, p < 0.001), and weak 190 correlations between ACD and AL, and LT with VL and AL (Spearman’s correlation, all r ≥ 0.344, all p ≤ 191 0.015) (Table 3). Nevertheless, no correlation was obtained among ACD with VL (Spearman’s correlation, 192 p = 0.054) (Table 3). 193 194 3.6 Correlation between body height, foot length and ocular biometry data measured by the 195 ultrasound biometer 196 With the biometry data obtained by the ultrasound biometer, statistically significant correlations were only 197 observed with AL. This was observed for both body height (Spearman’s correlation, r = 0.397, p = 0.004) 198 and foot length (Spearman’s correlation, r = 0.351, p = 0.012), although with weak correlations in both 199 cases (Table 3). 200 Regarding the correlations of the ultrasound biometric data with each other measured by that biometer, all 201 parameters showed weak correlations with AL (Spearman’s correlation, all r ≥ 0.344, all p ≤ 0.015), VL 202 showed weak correlation with ACD and moderate correlation with LT (Spearman’s correlation, both r ≥ - 203 0.331, both p ≤ 0.019), and no relationship was found between ACD and LT (Spearman’s correlation, p = 204 0.095) (Table 3). 205 206 9 4. DISCUSSION 207 Measuring ocular biometry is of high clinical relevance because of its application to some pathological 208 conditions, such as the increasing prevalence of high myopia [9], and the intraocular lens calculations 209 related to cataract or refractive surgery [1,7]. The technological development of biometers must ensure that 210 accuracy is maintained. From the first ultrasonic biometers to the new optical biometers, numerous 211 improvements have been achieved. Optical biometers have high resolution and, thanks to the possibility of 212 incorporating other technologies, can quickly measure not only the AL, central corneal thickness (CCT), 213 ACD, and LT, but also corneal curvature, white-to-white distance, and pupil diameter [6]. The main 214 advantage of optical biometers over ultrasonic is that those devices do not require contact, avoiding the use 215 of anesthetic and the possible transmission of infections [4]. In addition, a probe misalignment and corneal 216 compression with ultrasonic biometers may contribute to inaccurate readings [1,4]. However, ultrasound 217 biometry is still necessary, because optical methods cannot obtain readings in eyes with opacities (leukoma, 218 dense cataract, hemorrhages…) or in the presence of inadequate foveal fixation (due to poor visual acuity, 219 eccentric fixation, retinal pathologies or alterations…) [1]. 220 In the present study, the results of two biometers (non-contact optical Lenstar LS900 and contact ultrasound 221 EchoScan US-800) were compared both against each other and, as well as against their respective 222 measurements obtained with the naked eye and taken during CL wear. When comparing the biometrics 223 results between the optical and ultrasound biometers with the naked eye, significant differences were 224 observed, which means that these devices could not be interchangeable. Similar results were obtained by 225 Gursoy et al. [33] when comparing Lenstar vs. another ultrasound biometry device. Ultrasound biometry 226 values were higher in ACD and LT, whereas optical biometry values were higher in VL and AL. When 227 comparing both devices wearing CL, ultrasound biometry values were also higher in ACD and optical 228 biometry values were also higher in VL and AL, but no significant differences were observed for the LT 229 (Wilcoxon paired-sample test, p = 0.159). This variation in the LT may be due to accommodation. Figures 230 1B and 2B show that the LT results when biometers were compared are divided into two distinct groups. 231 During the ultrasound biometer measurements, one group of participants likely maintained relaxed 232 16 22. Piñero DP, López-Navarro A, Cabezos I, et al. Corneal topographic and aberrometric 367 measurements obtained with a multidiagnostic device in healthy eyes: intrasession repeatability. J 368 Ophthalmol. 2017;2149145. 369 23. García‐Resúa C, Pena‐Verdeal H, Miñones M, et al. Reliability of the non‐contact tono‐pachymeter 370 Tonopachy NT‐530P in healthy eyes. Clin Exp Optom. 2013;96(3):286-94. 371 24. Fayed MA, Chen TC. Pediatric intraocular pressure measurements: Tonometers, central corneal 372 thickness, and anesthesia. Surv Ophthalmol. 2019;64(6):810-25. 373 25. Kuo Y-C, Wang J-H, Chiu C-J. Comparison of open-field autorefraction, closed-field 374 autorefraction, and retinoscopy for refractive measurements of children and adolescents in Taiwan. 375 J Formos Med Assoc. 2020;119(8):1251-8. 376 26. Hadush MY, Berhe AH, Medhanyie AA. Foot length, chest and head circumference measurements 377 in detection of low birth weight neonates in Mekelle, Ethiopia: a hospital based cross sectional 378 study. BMC pediatrics. 2017;17(1):111. 379 27. O’Donnell C, Hartwig A, Radhakrishnan HJCL, et al. Correlations between refractive error and 380 biometric parameters in human eyes using the LenStar 900. Cont Lens Anterior Eye. 381 2011;34(1):26-31. 382 28. Del Aguila-Carrasco AJ, Dominguez-Vicent A, Perez-Vives C, et al. Assessment of modifications 383 in thickness, curvatures, and volume upon the cornea caused by disposable soft contact lens wear. 384 Eur J Ophthalmol. 2015;25(5):385-90. 385 29. Armstrong RA, Davies LN, Dunne MC, Gilmartin B. Statistical guidelines for clinical studies of 386 human vision. Ophthalmic Physiol Opt. 2011;31(2):123-36. 387 30. Bland JM, Altman D. Statistical methods for assessing agreement between two methods of clinical 388 measurement. The lancet. 1986;327(8476):307-10. 389 31. Carkeet A. Exact parametric confidence intervals for Bland-Altman limits of agreement. Optom 390 Vis Sci. 2015;92(3):e71-80. 391 17 32. Dunn G. Design and analysis of reliability studies: The statistical evaluation of measurement errors. 392 New York, London: Oxford University Press: Edward Arnold Publishers; 1989. 393 33. Gursoy H, Sahin A, Basmak H, et al. Lenstar versus ultrasound for ocular biometry in a pediatric 394 population. Optom Vis Sci. 2011;88(8):912-9. 395 ** Evaluated the differences between biometric measurements employing the Lenstar optical biometer 396 versus an ultrasound biometer. 397 34. She Z, Hung L-F, Beach KM, et al. Comparing low-coherence interferometry and A-scan 398 ultrasonography in measuring ocular axial dimensions in young rhesus monkeys. Exp Eye Res. 399 2022;217:108937. 400 35. Wang Q, Ji X, Lu D, et al. Comparison of A-Scan ultrasonography and the Lenstar optical biometer 401 in Guinea pig eyes. Exp Eye Res. 2021;207:108578. 402 36. Buckhurst PJ, Wolffsohn JS, Shah S, et al. A new optical low coherence reflectometry device for 403 ocular biometry in cataract patients. Br J Ophthalmol. 2009;93(7):949-53. 404 37. Lewis JR, Knellinger AE, Mahmoud AM, et al. Effect of soft contact lenses on optical 405 measurements of axial length and keratometry for biometry in eyes with corneal irregularities. 406 Invest Ophthalmol Vis Sci. 2008;49(8):3371-8. 407 * Evaluated CL wear during measurements with optical devices on irregular corneas. 408 38. Lee KE, Klein BE, Klein R, et al. Association of age, stature, and education with ocular dimensions 409 in an older white population. Arch Ophthalmol. 2009;127(1):88-93. 410 39. Roy A, Kar M, Mandal D, et al. Variation of axial ocular dimensions with age, sex, height, BMI- 411 and their relation to refractive status. J Clin Diagn Res. 2015;9(1):AC01. 412 413 18 FIGURE LEGENDS 414 Figure 1. Mean versus differences (Bland-Altman plot) between biometers data with the naked eye in n = 415 50 participants. The thick solid horizontal line indicates the mean difference while the closely dashed 416 horizontal lines indicate the 95% LoAs (Mean difference ± 1.96xSD). The widely dashed horizontal lines 417 indicate the 95% Confidence Interval of the LoAs. A) ACD by Lenstar vs. ACD by EchoScan, B) LT by 418 Lenstar vs. LT by EchoScan, C) VL by Lenstar vs. VL by EchoScan, D) AL by Lenstar vs. AL by EchoScan. 419 95% LoAs = 95% Limits of Agreement. 95% CI = 95% Confidence Interval. SD = Standard Deviation. 420 ACD = Anterior Chamber Depth. LT = Lens Thickness. VL = Vitreous Length. AL = Axial Length. 421 Figure 2. Mean versus differences (Bland-Altman plot) between biometers data during CL wear in n = 50 422 participants. The thick solid horizontal line indicates the mean difference while the closely dashed 423 horizontal lines indicate the 95% LoAs (Mean difference ± 1.96xSD). The widely dashed horizontal lines 424 indicate the 95% Confidence Interval of the LoAs. A) ACD during CL wear by Lenstar vs. ACD during 425 CL wear by EchoScan, B) LT during CL wear by Lenstar vs. LT during CL wear by EchoScan, C) VL 426 during CL wear by Lenstar vs. VL during CL wear by EchoScan, D) AL during CL wear by Lenstar vs. AL 427 during CL wear by EchoScan. 95% LoAs = 95% Limits of Agreement. 95% CI = 95% Confidence Interval. 428 SD = Standard Deviation. ACD = Anterior Chamber Depth. LT = Lens Thickness. VL = Vitreous Length. 429 AL = Axial Length. CL = Contact Lenses. 430 Figure 3. Mean versus differences (Bland-Altman plot) between ocular biometry with the naked eye and 431 during CL wear measured by Lenstar in n = 50 participants. The thick solid horizontal line indicates the 432 mean difference while the closely dashed horizontal lines indicate the 95% LoAs (Mean difference ± 433 1.96xSD). The widely dashed horizontal lines indicate the 95% Confidence Interval of the LoAs. A) ACD 434 by Lenstar vs. ACD during CL wear by Lenstar, B) LT by Lenstar vs. LT during CL wear by Lenstar, C) 435 VL by Lenstar vs. VL during CL wear by Lenstar, D) AL by Lenstar vs. AL during CL wear by Lenstar. 436 95% LoAs = 95% Limits of Agreement. 95% CI = 95% Confidence Interval. SD = Standard Deviation. 437 ACD = Anterior Chamber Depth. LT = Lens Thickness. VL = Vitreous Length. AL = Axial Length. CL = 438 Contact Lenses. 439 19 Figure 4. Mean versus differences (Bland-Altman plot) between ocular biometry with the naked eye and 440 during CL wear measured by EchoScan in n = 50 participants. The thick solid horizontal line indicates the 441 mean difference while the closely dashed horizontal lines indicate the 95% LoAs (Mean difference ± 442 1.96xSD). The widely dashed horizontal lines indicate the 95% Confidence Interval of the LoAs. A) ACD 443 by EchoScan vs. ACD during CL wear by EchoScan, B) LT by EchoScan vs. LT during CL wear by 444 EchoScan, C) VL by EchoScan vs. VL during CL wear by EchoScan, D) AL by EchoScan vs. AL during 445 CL wear by EchoScan. 95% LoAs = 95% Limits of Agreement. 95% CI = 95% Confidence Interval. SD = 446 Standard Deviation. ACD = Anterior Chamber Depth. LT = Lens Thickness. VL = Vitreous Length. AL = 447 Axial Length. CL = Contact Lenses. 448 449