Comparison between ocular biometric parameters and intraocular pressure with and without contact lenses

Noya Padín, Verónica; Nores Palmas, Noelia; Giráldez Fernández, María Jesús; Yebra-Pimentel Vilar, Eva; Pena Verdeal, Hugo

Abstract

Objectives: To compare the values of central corneal thickness, anterior chamber depth (ACD), lens thickness (LT), vitreous length (VL), axial length (AL), and corneal-compensated intraocular pressure performed with and without contact lenses (CL). Methods: Forty volunteer participants (16 men and 24 women, 24.262.9 years) were recruited. In a single visit, participants underwent autorefraction, keratometry, biometry, topography, pachymetry, and tonometry with the naked eye (without CL). Then, biometry, pachymetry, and tonometry were repeated twice wearing two CL (somofilcon A and nesofilcon A) fitted in a random order. Results: Many of the ocular biometric values were affected by the use of CL during measurements (paired t test; all P#0.003), except for LT and VL (Wilcoxon test; both P$0.272). Corneal-compensated intraocular pressure was also affected by contact lens wear during measurements, obtaining lower values when wearing them (Wilcoxon test; all P#0.001). Conclusions: Central corneal thickness, ACD, AL, and cornealcompensated intraocular pressure measurements cannot be performed while wearing CL. However, LT and VL measurements were not affected by any contact lens use. In addition, it was observed that ACD results from both devices are not interchangeable either when measured with the naked eye or using any contact lens.

Full text

1 TITLE: Comparison between ocular biometric parameters and intraocular pressure with 1 and without contact lenses 2 Running title: Ocular measurements with contact lenses 3 4 AUTHORS: 5 Veronica Noya-Padin, MSc1* - ORCID: 0000-0002-2815-3653 6 Noelia Nores-Palmas, MSc1 - ORCID: 0000-0001-8317-8039 7 Maria Jesus Giraldez, PhD1,2 - ORCID: 0000-0001-9595-7486 8 Eva Yebra-Pimentel, PhD1,2 - ORCID: 0000-0002-4012-4636 9 Hugo Pena-Verdeal, PhD1,2 - ORCID: 0000-0001-5103-1374 10 1Applied Physics Department (Optometry Area), Facultade de Óptica e Optometría, 11 Universidade de Santiago de Compostela, Santiago de Compostela, Spain 12 2Optometry Group, Instituto de Investigación Sanitaria Santiago de Compostela (IDIS), 13 Santiago de Compostela, Spain 14 *Corresponding author: Veronica Noya-Padin. Facultade de Óptica e Optometría, 15 Campus Vida s/n, Universidade de Santiago de Compostela, 15782 Santiago de 16 Compostela, Galicia, Spain. Tel.: +34 881 813610; e-mail: veronicanoya[email protected] 17 18 Number of tables: 4 / Number of figures: 1 19 Date of submission: March 29th, 2023 20 21 Conflict of interest statement: None. 22 Funding: This research did not receive any specific grant from funding agencies in the 23 public, commercial, or not-for-profit sectors. 24 2 Objectives: To compare the values of central corneal thickness, anterior chamber depth, 25 lens thickness, vitreous length, axial length and corneal-compensated intraocular 26 pressure performed with and without contact lenses. Methods: 40 volunteer participants 27 (16 men and 24 women, 24.2 ± 2.9 years) were recruited. In a single visit, participants 28 underwent autorefraction, keratometry, biometry, topography, pachymetry, and 29 tonometry with the naked eye (without contact lenses). Then, biometry, pachymetry and 30 tonometry were repeated twice wearing two contact lenses (Somofilcon A and 31 Nesofilcon A) fitted in random order. Results: Many of the ocular biometric values 32 were affected by the use of contact lenses during measurements (paired t-test; all p ≤ 33 0.003), except for lens thickness and vitreous length (Wilcoxon test; both p ≥ 0.272). 34 Corneal-compensated intraocular pressure was also affected by contact lens wear during 35 measurements, obtaining lower values when wearing them (Wilcoxon test; all p ≤ 36 0.001). Conclusions: Central corneal thickness, anterior chamber depth, axial length 37 and corneal-compensated intraocular pressure measurements cannot be performed while 38 wearing contact lenses. However, lens thickness and vitreous length measurements were 39 not affected by any contact lens use. In addition, it was observed that anterior chamber 40 depth results from both devices are not interchangeable either when measured with the 41 naked eye or using any contact lens. 42 43 Keywords: contact lenses; EchoScan US-800; intraocular pressure; ocular biometry; 44 Visionix120 45 3 On daily routine visual assessment, numerous instruments are used for diverse ocular 46 biometric parameters measurement of the eye, such as the central corneal thickness 47 (CCT), the anterior chamber depth (ACD), the lens thickness (LT), the vitreous length 48 (VL) or the total axial length (AL).1 The values of these parameters are relevant due to 49 their relationship with some pathological processes: the CCT measured by pachymetry 50 is an important parameter during a follow-up of conditions such as keratoconus, corneal 51 edema or glaucoma 2; a shallow ACD and the closure of the drainage angle have been 52 identified as a risk factor for glaucoma as well 3; LT and its position are necessary data 53 to apply in the intraocular lens calculation formulas, allowing to predict the position in 54 which the intraocular lens will tend to locate itself 4; VL measurement has been 55 recommended in keratoconus patients who are going to undergo corneal transplantation 56 because of its better accuracy than with AL in deciding the graft size 5; or finally, the 57 AL measured by ocular biometry, has become essential in myopia control,6 as well as in 58 conjunction with the keratometry values, being also fundamental to calculate the 59 intraocular lens power in cataract surgery.7 Those ocular biometric values could be 60 acquired using either classical or newly developed technologies. While the Ultrasound 61 A principle has been classically proposed as the gold standard for routine assessment, 62 its use in ophthalmic practice is limited by several drawbacks, such as direct contact 63 with the eye or possible ocular indentation.8 New technologies such as Scheimpflug 64 cameras have been introduced in the assessment of the anterior eye structures to 65 overcome those disadvantages.9 These technologies are clinically employed due to their 66 accuracy, the non-contact protocols, and the less experience-dependent factor. However, 67 in some cases, such as media opacity, optical technologies are limited, and the use of 68 ultrasonic devices is still necessary.10 Another parameter of great relevance in daily 69 4 routine visual assessment is the Corneal Compensated Intraocular Pressure (IOPcc), due 70 to its relationship with pathologies such as glaucoma.11 71 In the last decades, there has been a continuous increase in Contact Lenses (CL) 72 prescriptions as an option for refractive compensation.12 This increase could be due to 73 factors such as the quality of vision with the CL, the inherent comfort for physical and 74 leisure activities, or the full field of vision without interference from the frame,13 as well 75 as the development of CL for myopia control, which has boosted the use of CL in 76 children.14 Moreover, CLs are essential in some visual conditions, such as 77 anisometropia or keratoconus.15, 16 For these reasons, many people wear their CL 78 continuously and even come to their visual exams with them on; although the 79 recommendation is not to wear them for a period before the visual examination (there is 80 no consensus on the recommended period of time due to the lack of evidence in this 81 regard)17 since the measurements could be altered due to the transient alteration of the 82 corneal curvature generated by the CL.17 Thus, knowing how they affect the usual 83 clinical measurements earns interest. Far from these reasons, there is an added interest 84 for clinical practice in the use of CL in some measurement techniques which usually 85 require the use of anesthesia because of their invasiveness, for example, ultrasonic 86 ocular biometry where the CLs have the potential to replace the use of the drug. There 87 were already small studies evaluating the use of CL during the performance of certain 88 tests, particularly on the OCT technique, although this area still needs more research.18, 89 19 90 91 METHODS 92 Sample size calculation, inclusion and exclusion criteria and participants 93 5 For the sample size calculation, the software PS Power and Sample Size Calculations 94 Version 3.1.2 (Copyright© by William D. Dupont and Walton D. Plummer) was used. 95 Since the corneal and CL thickness are usually recorded in the same magnitudes (µm), 96 the sample size calculation was based on this parameter. Previous reports indicate that 97 the mean standard deviations (SD) of CCT are usually distributed with a value of 38 98 µm,20 while the lowest value of the CL thickness employed in the present study was 80 99 µm (Table 1). Based on those data, to have 90% power (Type II error associated) for a 100 significance level of 𝛼 = 0.05 (Type I error associated) with a confidence level of 95% 101 to detect a clinical difference between values with and without CL, the minimum 102 number of subjects required was 6 per group; since two different CL were employed in 103 a paired analysis, a minimum of 12 participants are needed in the final study group to 104 avoid statistical interference. The calculation was adjusted for a 30% loss to follow-up 105 between both CL fittings regarding possible drop-out during the measurement session 106 due to discomfort with the CL or impossibility to insert the CL, or also for possible 107 exclusion from the final analysis based on inclusion and exclusion criteria. Based on the 108 above, the calculation yielded a total sample size of 16 participants to accomplish the 109 study´s aims. 110 A total of 42 qualified volunteer participants (17 men and 25 women) with a mean age 111 of 24.3 ± 3.0 (from 19 to 33 years) from subjects attending the Optometry Clinic of the 112 center were recruited. The inclusion criteria were as follows: an auto-refracted sphere of 113 -8.00 to +4.00 D, astigmatism of ≤ 3.00 D,19, 21 keratometric readings in both flatter and 114 steeper meridian equal or lower to 47.20 D,22 and IOPcc equal or lower than 21.0 115 mmHg,11 and an anterior chamber angle both nasal and/or temporal equal or higher than 116 30º.23 The exclusion criteria were as follows: history of conjunctival, scleral, or corneal 117 6 disease, previous diagnosis of dry eye disease or glaucoma, a history of ocular surgery 118 and/or, diabetes mellitus, thyroid disorders, hypersensitivity or allergy to anesthesia 119 and/or systemic disorders that could affect the measurements.21, 24 No participant was 120 under any type of topical and systemic treatment or used artificial tears at the time of 121 the study. Current CL wearers, all of them users of spherical and/or low-astigmatism 122 soft CL, were asked not to wear their CL on the previous 24 hours of the examination. 123 Informed consent was obtained from all participants prior to their inclusion in the study. 124 The study protocol adhered to the principles outlined in the Declaration of Helsinki and 125 received approval from the Ethics Committee of the University. 126 Study Design 127 In a single-one-day visit, participants underwent a battery of tests always in the same 128 order with the naked eye (without CL): Autorefraction, keratometry, biometry, 129 topography, pachymetry and tonometry. Then, biometric parameters, pachymetry and 130 tonometry were repeated twice wearing two different CLs. The CL fitting order for each 131 participant was randomly assigned in arrival order (Somofilcon-Nesofilcon to odd- 132 numbered subjects and Nesofilcon-Somofilcon to even-numbered subjects, ensuring a 133 1:1 ratio). The fit of each CL was checked with an SL-D4 slit-lamp (Topcon 134 Corporation, Tokyo, Japan). The examination lasted about 30 minutes per participant; 135 approximately 10 minutes for inclusion criteria measurements and study parameter 136 measurements with the naked eye, and 10 minutes to fit each CL and repeat the study 137 parameter measurements. After the fit of the CL, 5 minutes had waited to perform the 138 measurements.25 139 Autorefraction, keratometry, tonometry and anterior chamber angle were used as 140 inclusion criteria. Topography was used for the CL fitting. Biometry and pachymetry 141 7 reading, with and without CL, were used as studied parameters. IOPcc measurements 142 were used both as inclusion criteria and as study parameters. In all tests, measurements 143 both with and without CL were taken by the same researcher only in the right eye of 144 each participant to avoid inter-observer variations or overstating the precision of 145 statistical estimates.8 146 Contact Lenses 147 During the protocol, two aspheric daily disposable CL were fitted: Somofilcon A 148 (Clariti 1 Day, CooperVision Inc., Pleasanton United States) and Nesofilcon A (Biotrue 149 ONEday, Bausch & Lomb Inc., New York, United States).26, 27 The parameters of the 150 CL are detailed in Table 1. In all participants, the lens power used was -3.00 D to ensure 151 the same thickness in all cases based on the manufacturer’s data provided. 152 Autorefraction and keratometry 153 The NVISION-K 5001 autorefractor (Rexxam Co., Kagawa, Japan) was used to obtain 154 non-cycloplegic refractive error measurements and keratometry values.28 This device is 155 a wide-view window autorefractor which allows a binocular view of a far fixation target 156 to minimize instrument myopia. Precision and vertex distance settings on the device 157 were set at 0.12 D and 12 mm respectively. Autorefraction and keratometry were 158 performed only once, and the results were used to determine if subjects were compliant 159 with the inclusion criteria. 160 Tonometry 161 Tonometry was performed with the Canon TX-10 non-contact tonometer (Canon Inc., 162 Tokyo, Japan),29 which blows an air pulse onto the eye and measures the patient’s 163 cornea deformation to obtain the IOP value. This device automatically performs a 164 8 battery of three measurements and the average is displayed on the device screen as the 165 IOP value in mmHg. 166 Topography, pachymetry and anterior chamber angle 167 Topography, pachymetry and anterior chamber angle measurement procedures were 168 performed by the multi-diagnostic Visionix120 (Visionix Luneau Technologies, 169 Chartres, France).30 This device is based on a Placido disk system projected on the 170 corneal surface to provide all corneal topographic information and a Scheimpflug 171 imaging-based system to perform the pachymetry and anterior chamber angle 172 measurement. The latter uses a monochromatic blue light (455 nm) to obtain 173 pachymetry (CCT, and was also used to calculate the IOPcc), anterior chamber angle 174 and ACD measurements. Topography was performed only once to evaluate the fit of the 175 CL prior to placement. 176 Biometry 177 The NIDEK EchoScan Ultrasound Model US-800 biometry (Nidek Co., Tokyo, Japan) 178 was used to obtain ACD, LT, VL and AL.31 The EchoScan US-800 is an ultrasonic 179 device which applies the ultrasonic principle based on the reflection method. When the 180 cornea is touched by the probe, a 10 MHz ultrasonic pulse is transmitted which is 181 reflected from each intraocular tissue. The echo generated is received by the probe, and 182 the time it delays coming to the probe is converted into distance. In all measurements, 183 the instrument was set to automatically perform measurements when the probe touches 184 the cornea. By an automatic correction process, the device averages the better three 185 consecutive measurements to avoid the data being affected by the movement of the eye 186 and/or probe. To enable the measurements, preemptive anesthesia was administered to 187 the naked eye beforehand, owing to the contact between the probe and the cornea. 188 9 Statistical analyses 189 Data analysis was performed with the SPSS statistical software v. 25.0 for Windows 190 (SPSS Inc., Chicago, United States). Significance was set at a p ≤ 0.05 for all the tests. 191 Previous to the analysis, the normal distribution of the data was tested using the 192 Shapiro-Wilk test 32; CCT, ACD and AL data showed a normal distribution (Shapiro- 193 Wilk test: all p ≥ 0.055) while LT, VL and IOPcc data showed a non-normal 194 distribution (Shapiro-Wilk test: all p ≤ 0.029). 195 Differences between parametric variables (CCT, ACD and AL) with the naked eye and 196 wearing CL measured by the Visionix120 or EchoScan US-800, as well as between 197 ACD measured by Visionix120 or EchoScan US-800, were assessed using a t-test. To 198 avoid type I errors arising from multiple comparisons in the variables difference 199 analysis, statistical significance (p < 0.050) was divided by the number of comparisons 200 performed to give a p < 0.025 for CCT, LT, VL, AL and IOPcc and p ≤ 0.0125 for 201 ACD.32, 33 Differences between LT, VL and IOPcc with the naked eye and wearing CL 202 were assessed using the Wilcoxon test. The 95% Confidence Intervals (95% CI) of the 203 differences between the values of the measurements were also provided (Mean 204 difference ± 1.96 x SD/√n).32, 34 205 206 RESULTS 207 Based on the inclusion criteria, the final sample consisted of 40 right eyes of 40 208 participants (16 men – 24 women) with mean ± SD age of 24.2 ± 2.9 years in a range 209 from 19 to 33 years (Descriptive statistics of the sample are shown in Table 2). One 210 participant was excluded for having a cylindrical refractive error greater than 3.00 D 211 while another was excluded for having keratometry values greater than 47.20 D. 212 16 CL. Differences in ACD and AL measured by EchoScan appear to depend on the type 357 of CL used, and LT and VL also measured by EchoScan seem to be not dependent on 358 the use of CL. Thus, the CL wear to perform measurements with the EchoScan US-800 359 is not a good option to replace the use of anesthesia when measuring the biometry of the 360 eye, except for inner parameters (LT and VL). IOPcc values measured by a non-contact 361 tonometer are also affected by CL-wear during measurements. The ACD values 362 measured by Visionix120 and EchoScan US-800 are not interchangeable. 363 17 REFERENCES 364 365 1. Rio-Cristobal A, Martin R. Corneal assessment technologies: current status. 366 Surv Ophthalmol 2014;59(6):599-614. 367 2. Szczotka-Flynn LB, Efron N. Aftercare. In: Contact Lens Practice: Elsevier; 368 2018. p. 364-384. e2. 369 3. Devereux JG, Foster PJ, Baasanhu J, et al. Anterior chamber depth measurement 370 as a screening tool for primary angle-closure glaucoma in an East Asian 371 population. Arch Ophthalmol 2000;118(2):257-263. 372 4. Meng J, Wei L, He W, et al. Lens thickness and associated ocular biometric 373 factors among cataract patients in Shanghai. Eye Vis (Lond) 2021;8(1):22. 374 5. Messina M, Umapathy T, Avadhanam V, et al. Vitreous cavity length in 375 keratoconus: implications for keratoplasty. Eye (Lond) 2018;32(2):359-363. 376 6. Chen Y, Wang D, Chen L, et al. Association of refraction and ocular biometry in 377 highly myopic eyes. Clin Exp Optom 2021;104(5):589-594. 378 7. Sheard R. Optimising biometry for best outcomes in cataract surgery. Eye 379 (Lond) 2014;28(2):118-125. 380 8. Findl O, Kriechbaum K, Sacu S, et al. Influence of operator experience on the 381 performance of ultrasound biometry compared to optical biometry before 382 cataract surgery. J Cataract Refract Surg 2003;29(10):1950-1955. 383 9. Kanclerz P, Khoramnia R, Wang X. Current developments in corneal 384 topography and tomography. Diagnostics (Basel) 2021;11(8):1466. 385 18 10. Li Y, Li HX, Liu YC, et al. Comparison of immersion ultrasound and low 386 coherence reflectometry for ocular biometry in cataract patients. Int J 387 Ophthalmol 2018;11(6):966-969. 388 11. Munteanu GZ, Munteanu ZVI, Roiu G, et al. Detection of Intraocular 389 Hypertension during Opportunity Screening (Check-Up Medical Inspections). J 390 Pers Med 2022;12(5):777. 391 12. Ibrahim NK, Seraj H, Khan R, et al. Prevalence, habits and outcomes of using 392 contact lenses among medical students. Pak J Med Sci 2018;34(6):1429. 393 13. Rah MJ, Walline JJ, Jones-Jordan LA, et al. Vision specific quality of life of 394 pediatric contact lens wearers. Optom Vis Sci 2010;87(8):560-566. 395 14. Morgan PB, Efron N. Quarter of a century of contact lens prescribing trends in 396 the United Kingdom (1996 - 2020). Cont Lens Anterior Eye 2022;45(3):101446. 397 15. Evans BJ. Orthoptic indications for contact lens wear. Cont Lens Anterior Eye 398 2006;29(4):175-181. 399 16. Downie LE, Lindsay RG. Contact lens management of keratoconus. Clin Exp 400 Optom 2015;98(4):299-311. 401 17. Goudie C, Tatham A, Davies R, et al. The effect of the timing of the cessation of 402 contact lens use on the results of biometry. Eye (Lond) 2018;32(6):1048-1054. 403 18. Lewis JR, Knellinger AE, Mahmoud AM, et al. Effect of soft contact lenses on 404 optical measurements of axial length and keratometry for biometry in eyes with 405 corneal irregularities. Invest Ophthalmol Vis Sci 2008;49(8):3371-3378. 406 19. Ferrer-Blasco T, Esteve-Taboada JJ, Dominguez-Vicent A, et al. Effect of 407 contact lenses on ocular biometric measurements based on swept-source optical 408 coherence tomography. Arq Bras Oftalmol 2019;82(2):129-135. 409 19 20. Ganesh D, Lin SR. Global metrics on ocular biometry: representative averages 410 and standard deviations across ten countries from four continents. Eye (Lond) 411 2022;37(3):511-515. 412 21. Pinero DP, Cabezos I, Lopez-Navarro A, et al. Intrasession repeatability of 413 ocular anatomical measurements obtained with a multidiagnostic device in 414 healthy eyes. BMC Ophthalmol 2017;17(1):193. 415 22. Beiko GH, Haigis W, Steinmueller A. Distribution of corneal spherical 416 aberration in a comprehensive ophthalmology practice and whether keratometry 417 can predict aberration values. J Cataract Refract Surg 2007;33(5):848-858. 418 23. Sihota R, Kamble N, Sharma AK, et al. 'Van Herick Plus': a modified grading 419 scheme for the assessment of peripheral anterior chamber depth and angle. Br J 420 Ophthalmol 2019;103(7):960-965. 421 24. Pena-Verdeal H, Garcia-Resua C, Vazquez-Sanchez C, et al. Inter-eye 422 osmolarity differences in patients with symptomatic and non-symptomatic dry 423 eyes. Arq Bras Oftalmol 2020;83(2):103-108. 424 25. Boychev N, Laughton DS, Bharwani G, et al. How should initial fit inform soft 425 contact lens prescribing. Cont Lens Anterior Eye 2016;39(3):227-233. 426 26. Kapfelsberger A, Eckstein J, von Ahrentschildt A, et al. Ultraviolet and Visible 427 Transmittance of Soft Contact Lenses with and without Ultraviolet Blockers. 428 Optom Vis Sci 2021;98(11):1270-1278. 429 27. Pena-Verdeal H, Noya-Padin V, Losada-Oubina M, et al. Changes of 430 symptomatology, tear film and ocular surface integrity one week during 431 Somofilcon-A and Omafilcon-A lens wear. Eur J Ophthalmol 432 2022:11206721221131131. 433 20 28. Kuo Y-C, Wang J-H, Chiu C-J. Comparison of open-field autorefraction, 434 closed-field autorefraction, and retinoscopy for refractive measurements of 435 children and adolescents in Taiwan. J Formos Med Assoc 2020;119(8):1251- 436 1258. 437 29. Garcia-Resua C, Pena-Verdeal H, Minones M, et al. Reliability of the non- 438 contact tono-pachymeter Tonopachy NT-530P in healthy eyes. Clin Exp Optom 439 2013;96(3):286-294. 440 30. Gordon-Shaag A, Piñero DP, Kahloun C, et al. Validation of refraction and 441 anterior segment parameters by a new multi-diagnostic platform (VX120). J 442 Optom 2018;11(4):242-251. 443 31. Yong K-L, Gong T, Nongpiur ME, et al. Myopia in Asian subjects with primary 444 angle closure: implications for glaucoma trends in East Asia. J Ophthalmol 445 2014;121(8):1566-1571. 446 32. Armstrong RA, Davies LN, Dunne MC, et al. Statistical guidelines for clinical 447 studies of human vision. Ophthalmic Physiol Opt 2011;31(2):123-136. 448 33. Pena-Verdeal H, Garcia-Resua C, Ramos L, et al. Diurnal variations in tear film 449 break-up time determined in healthy subjects by software-assisted interpretation 450 of tear film video recordings. Clin Exp Optom 2016;99(2):142-148. 451 34. Bland JM, Altman DG. Measuring agreement in method comparison studies. 452 Stat Methods Med Res 1999;8(2):135-160. 453 35. Schafer J, Steffen R, Reindel W, et al. Evaluation of surface water 454 characteristics of novel daily disposable contact lens materials, using refractive 455 index shifts after wear. Clin Ophthalmol 2015;9:1973-1979. 456 21 36. Firat PG, Cankaya C, Doganay S, et al. The influence of soft contact lenses on 457 the intraocular pressure measurement. Eye (Lond) 2012;26(2):278-282. 458 37. Patel S, Illahi W. Non-contact tonometry over soft contact lenses: effect of 459 contact lens power on the measurement of intra-ocular pressure. Cont Lens 460 Anterior Eye 2004;27(1):33-37. 461 38. Liu YC, Huang JY, Wang IJ, et al. Intraocular pressure measurement with the 462 noncontact tonometer through soft contact lenses. J Glaucoma 2011;20(3):179- 463 82. 464 39. Sapkota K, Franco S, Lira M. Intraocular pressure measurement with ocular 465 response analyzer over soft contact lens. Cont Lens Anterior Eye 466 2014;37(6):415-419. 467 40. Roggla V, Leydolt C, Schartmuller D, et al. Influence of Artificial Tears on 468 Keratometric Measurements in Cataract Patients. Am J Ophthalmol 2021;221:1- 469 8. 470 471 22 FIGURE LEGENDS 472 Figure 1. Mean (parametric parameters) or Median (non-parametric parameters) 473 values when measuring with the naked eye or wearing CL. n = 40. A) Mean of central 474 corneal thickness, B) Mean of anterior chamber depth measured by Visionix and 475 EchoScan, C) Median of lens thickness, D) Median of vitreous length, E) Mean of axial 476 length, F) Median of IOPcc. IOPcc: Corneal Compensated Intraocular Pressure; CL: 477 Contact lenses. Error bars represent the 95% confidence interval. Significant 478 differences (p ≤ 0.05, paired t-test) between naked eye measurements and CL wearing 479 measurements are indicated by a (*). 480