Accuracy of refractive index spectroscopy by broadband interferometry

Arosa Lobato, Yago; Rodríguez-Fernández, Carlos Damián; Doval Casas, Alejandro; López Lago, María Elena; Fuente Carballo, Raúl de la

Abstract

Uncertainties of refractive and group index in dispersion measurement by spectrally resolved white light interferometry are deeply analyzed. First, the contribution to uncertainty of the different parameters affecting both indices is identified. Afterwards, results are presented for a 1.5 mm thick fused silica sample over a broad spectral range, from 400 to 1000 nm, and the effects that mostly deteriorate the measurement accuracy are established. Finally, the different contributions are quadratically combined to determine the total uncertainty of the two indices

Full text

Measurement 187 (2022) 110225 Available online 6 October 2021 0263-2241/© 2021 The Author(s). Published by Elsevier Ltd. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/). Accuracy of refractive index spectroscopy by broadband interferometry Yago Arosa , Carlos Dami´ an Rodríguez-Fern´ andez , Alejandro Doval , Elena L´ opez Lago, Raúl de la Fuente * Nanomateriais, Fot´ onica e Materia Branda, Departamento de Física Aplicada, Universidade de Santiago de Compostela, Spain ARTICLE INFO Keywords: Dispersion Spectroscopy Spectral interferometry Uncertainty ABSTRACT Uncertainties of refractive and group index in dispersion measurement by spectrally resolved white light interferometry are deeply analyzed. First, the contribution to uncertainty of the different parameters affecting both indices is identified. Afterwards, results are presented for a 1.5 mm thick fused silica sample over a broad spectral range, from 400 to 1000 nm, and the effects that mostly deteriorate the measurement accuracy are established. Finally, the different contributions are quadratically combined to determine the total uncertainty of the two indices. 1. Introduction The possibility of using spectrally resolved white light interferometry (SRWLI) as a refractometric technique was already suggested in the early 1990s [1]. The first studies [2,3], focused on the measurement of the refractive index and its dispersion curve, n(λ), in a relatively wide spectral range (tens of nanometers). Soon after, these works were extended to incorporate the measurement of other dispersion parameters, mainly chromatic dispersion, dn/dλ, and group index, d( σ n)/d σ , where σ =1/λ is the wavenumber, although some authors evaluated higher order dispersion parameters [4,5]. All these studies aim to measure the refractive index or some kind of spectral derivative by combining spectroscopy with interferometry, and we will refer to these techniques as Refractive Index Spectroscopy by Broadband Interferometry (RISBI). Like other techniques based on SRWLI, to apply RISBI we basically need three elements: a broadband source, an interferometer, and a spectrometer. RISBI was used to measure the dispersion of reference materials, both isotropic [6–10] and anisotropic [11–13], to measure dispersion in fibers [14–17], to determine the group delay of dispersive mirrors [18,19], to measure ocular dispersion [20], to calculate thermo-optical coefficients [21], visualization of thermal lens effect [22] or to model the dispersion of families of fluids [23,24]. Furthermore, while the first works applied RISBI in the visible range, dispersion results in the near infrared [25,26] or even the UV range [27] can be found in the literature. Of course, the validity of the obtained results was analyzed in many works, but the analysis was often based on comparison with the values obtained using other methods. However, in some works, [1,26,28], the precision or the uncertainty of the measurements was alluded to, but always in a simplified way, and only regarding to refractive index. In this work, we aim to evaluate the accuracy of RISBI techniques to measure the refractive and group index dispersion of a generic sample in a broad and rigorous way. To perform the study, we derived the theoretical expressions describing the contribution to uncertainty of each parameter involved in a generic RISBI measurement. The previous expressions are used, without any loss of generality, to evaluate the uncertainty of a real measurement taken with our experimental RISBI system on a fused silica sample. Note that the parameters governing the uncertainty in our device are similar to those governing uncertainty in any other RISBI device and, in consequence, our uncertainty evaluation can be easily extrapolated to other experimental setups and to the measurement of other kind of samples. The rest of this contribution is organized as follows: in the next section, the different parameters that affect the measurements are identified and the fundamental formulas to calculate the two indices and their corresponding uncertainties are presented. In Section 3, the uncertainties associated to each of these parameters are modeled and evaluated separately, and, in Section 4, the results obtained with a thin sample of fused silica are presented and discussed. The discussion is first devoted to each parameter, and, afterwards, to their combination, which yields the final uncertainties of the indices. Finally, in Section 5, the conclusions are presented. * Corresponding author. E-mail address: [email protected] (R. de la Fuente). Contents lists available at ScienceDirect Measurement journal homepage: www.elsevier.com/locate/measurement https://doi.org/10.1016/j.measurement.2021.110225 Received 13 July 2021; Received in revised form 7 September 2021; Accepted 22 September 2021 Measurement 187 (2022) 110225 2 2. Foundations In a SRWLI experiment (see Fig. 1), two broadband beams are superimposed, and the interference pattern is resolved in the spectral domain. At an arbitrary wavelength, λ, the irradiance can be written as: I(λ) = I0(λ)[1+V(λ)cosφ(λ)],(1) with I 0 the background irradiance, V the fringe visibility, and φ, the phase difference between the light beams. In the case in which one of the beams passes through a transparent plate of thickness d, the phase difference is given by: φ(λ) = 2 π λ[d(n−nA)− nAl],(2) where n and n A are the plate and air refractive index, respectively, at wavelength λ, and l is the path difference travelled by one of the beams with respect to the other, in air. In this expression, it is assumed that no path difference has been introduced other than that generated in the thin plate and in air propagation. It is important to stress that the computed phase differs from that given by the equation by a multiple of 2 π . This is because the arc cosine is a multivalued function or, in other words, different phases differing by a multiple of 2 π result in the same value of the cosine. Therefore, when calculating the refractive index, we have an ambiguity that results from this lack of definition of the phase. Specifically, the refractive index at a wavelength λ is calculated as n=nA+1 d(λφ 2 π +nAl+kλ),(3) where k is an unknown integer. Using this equation and the well-known formula for the propagation of uncertainties, we obtain the uncertainty of the refractive index, Δn as 1 : In addition, in the third and fifth summands we have introduced n to abbreviate the formula and n geq is the group index at the so-called equalization wavelength, that at which the irradiance has a stationary point [4,7,39], that is, the wavelength at which the phase in Eq. (2) presents a minimum and, therefore, its derivative is zero Applying this last condition it is obtained: l=d[ngeq ngeqair −1](5) As discussed in the introduction, in addition to refractive index, another parameter that can be calculated with RISBI is the group index, n g : ng=d d σ (n σ ) = n+ σ dn d σ ,(6) With σ =1/λ, the wavenumber. Like refractive index dispersion, the group dispersion, that is, the wavelength dependence of the group refractive index, belongs to one of the fundamental dispersion characteristics of transparent optical materials. Nonlinear and ultrafast optics, optical fibers and waveguide photonics are specific areas in which group dispersion is of wide interest. From Eqs. (2) and (3), the following expression is obtained: ng=ngA +1 d(ngAl+1 2 π dφ d σ ),(7) being n gA the group index of air. One way to compute the phase derivative is to determine the fringe periodicity, Λ. It is [29]: Λ( σ ) = 2 π |dφ/d σ |,(8) and therefore: ng=ngA +1 d(ngAl±1 Λ),(9) where the modulus and the ±sign come from the fact that the phase derivative is zero at the equalization wavelength and changes its sign at each side: negative for shorter wavelengths, and positive for longer wavelengths. The corresponding uncertainty is, Δn g : Δng=1 d (ngAΔl)2+[(ng−ngA)Δd]2+(1 Λ2ΔΛ)2 √.(10) Eqs. (4) and (10) are the key equations of this study. To determine the value of the refractive and group index uncertainty, each term will be analyzed separately. Note that in the case of double pass interferometers, as the Michelson interferometer, one of the beams crosses the sample twice, so in the previous equations d =2e, where e is the sample thickness. 3. Uncertainties In this section, the uncertainties of the different parameters contributing to the refractive and group indices are discussed in detail. With respect to the uncertainty in the refractive index of air, an excellent review can be found in ref. [30]. 3.1. Spectrometer calibration The most common way to compute spectral irradiance is by means of a spectrometer. Here, no particular configuration will be considered, it is simply assumed that the resolution of the device is sufficient to resolve the interference fringes and that a linear camera is the spectrum sensing system. Calibration of the spectrometer consists in determining which wavelength falls in any pixel of the spectrometer camera. In a typical calibration procedure, various sources of known wavelength illuminate the camera, to establish the correspondence between pixels and wavelengths, λ(N), being N the pixel number, for the source wavelengths. Then, interpolation, polynomial fitting, or physical modeling [31–36] are used to determine a relation between wavelength and any pixel in the camera. In this work, we do this last step by means of a WLSI-based approach [37]. The process of calibration is described in detail in that reference; here, only a brief discussion is given. Once the available sources of known wavelength allow us to identify the λ arriving at some of the pixels of the spectrometer camera, the Δn=1 d (λ 2 π Δφ)2 +(λΔk)2+(n−ngeq λdΔλ)2+(nAΔl)2+[(n−nA)Δd]2+[(d+l)ΔnA]2 √.(4) 1 In the third term on the root, the following approximation was performed: (ng/ngeq)air ≃1 Y. Arosa et al. Measurement 187 (2022) 110225 3 proper RISBI device without sample is used so that a broadband source illuminates the camera. By applying RISBI procedure, an array of differential phase values, φ, corresponding to every pixel is also used in order to get a set of (N,φ) points. That ϕ measurement is made for a suitable path difference, chosen to have many spectral fringes. Pairs of values of wavelength and phase difference are interesting for our purpose because if there is no sample in the interferometer, the spectral phase will be linear with the wavenumber σ =1/λ That means: φ(N) = 2 π lnA σ (N)− 2k π ,(11) where l is, as before, the path difference in air travelled by the beams. This equation can be inverted to express the wavenumber as: nA σ (N) = aφ(N)+b,(12) with a =1/(2 π l) and b =k/l. Now, the known values of the calibration wavelengths are used to perform a linear fit and determine the constants a and b. They are employed to transform the measured phase difference into wavenumber for every pixel, providing the desired relationship σ (N) or λ(N). Note that this fitting can be highly enhanced by choosing a path difference l showing many resolved spectral fringes, but, obviously, the relationship λ(N) is independent of the value of l. since different path differences corresponds to different sets of φ values. Finally, Eq. (12) furnishes the uncertainty equation: Δ σ = (aΔφ)2+(φΔa)2+Δ2b √/nA,(13) together with: Δλ=Δ σ / σ 2=λ2Δ σ .(14) To know how to evaluate Δφ in Eq. (13), see Section 3.4. 3.2. The path difference in air The uncertainty of the path difference in air can be calculated by taking advantage of the linear relationship between phase and wavenumber in the absence of sample. Once the previous calibration was carried out, both magnitudes are related by the slope a as: l=1/(2 π a)⇒Δl=lΔa/|a|,(15) where we apply the absolute value to compute the uncertainty. Although in this section we apply the same methodology as in the previous one, it is common for the path difference to be measured with a specific spectrometer of smaller spectral range than the one used to measure dispersion. This will be the case when the latter spectrometer cannot resolve the fringes associated with the measurement of l (smaller than the sample thickness, but of the same order of magnitude). 3.3. The sample thickness If the sample is solid, its thickness can be measured directly with a high-resolution gauge. We must take into account the instrument resolution, δd r , the flatness, δd f , and parallelism, δd p , of the sample surfaces. The corresponding uncertainties are Δdr=δdr/ 3 √, Δdf=δdf/ 3 √, and Δdp=δdp/ 3 √, respectively. In addition to these uncertainties, we must include the statistical uncertainty that results from carrying out the same measurement several times. The uncertainty estimation of liquid samples is a little more complicated since they must be introduced into a transparent cell to be measured. You can proceed in two ways: (a) measuring the total thickness of the cell, the thickness of the walls and subtracting both magnitudes to obtain the internal thickness, d; (b) directly measuring the internal thickness by applying RISBI to a liquid of known refractive index. In this case, the phase difference added by the walls must be subtracted. Typically, in both methods, the estimated uncertainty is twice that of solid samples. 3.4. Phase computation The calculus of the phase uncertainty is undoubtedly the most complicated step. It is determined by errors in the measurement of irradiance, but the latter also affects the background irradiance and visibility. To understand the effect of the irradiance variation on the phase, pay attention to Fig. 2. It plots the actual (I) and measured irradiance (I +ΔI) as a function of phase in one cycle and a bit. The oscillations are assumed to be very fast with respect to the background and visibility variation, so that we can keep the latter constant during a cycle. The irradiance variation affects both the upper and lower envelopes of the measurement as well as the phase. On the one hand, there is (or it can be) a change in the maximum and minimum values, which is associated to changes in the envelopes. On the other hand, there is also (or at least it could be) a lateral shift at any point of the interferogram, which is associated with a change in phase. Consequently, the uncertainty of the phase depends on the uncertainty of the irradiance and its envelopes. Each contribution can be determined separately. Let us begin with the contribution of irradiance. The assumption that Light Source Interferometer Spectrometer Fig. 1. Outline of a SRWLI device. Fig. 2. Graph example of irradiance variation. Y. Arosa et al. Measurement 187 (2022) 110225 4 the envelopes are constant implies that the background irradiance and visibility are also constant. Therefore, from Eq. (1) it is obtained that the maximum phase variation due to irradiance is: I+ΔI=I0[1+Vcos(φ+Δφ)],(16) or reusing Eq. (1): ΔI=I0V[cosφ(cosΔφ−1)− sinφsinΔφ](17) In addition, we can approximate cosΔφand sinΔφ, so that to order Δφ 2 we have: ΔI= − I0V[cosφ(Δφ)2/2+sinφΔφ],(18) or, by reordering: cosφ(Δφ)2/2+sinφΔφ+ΔI/(I0V) = 0.(19) The solutions to this equation are written as: Δφ=−sinφ ± sin2φ−2cosφΔI/(I0V) √cosφ ,(20) where for consistency, ΔI and cosφ have opposite sign and, therefore, the solutions are real. Although both solutions are correct, only certain branches of these remain bounded and are physically valid (see Fig. 3). On the other hand, the uncertainty is taken as positive. Therefore, the physical solution for the uncertainty can be summarized in a single equation as: Δφ0= |tanφ|[ 1+2|cosφ|ΔI/(I0Vsin2φ) √−1],(21) where ΔI is assumed to be positive. This solution has two branches that meet at φ =0 (see Fig. 3), at which: Δφ0= 2ΔI/(I0V) √,(22) value which corresponds to an extreme of the solution. On the other hand, at φ =± π /2, the solutions tend asymptotically to: Δφ0=ΔI/(I0V).(23) Under normal conditions ΔI/(I0V)<1, the maximum value of uncertainty corresponds to the one given in Eq. (22), while the minimum value is the one corresponding to Eq. (23). In a second step, we must compute the contribution of the envelopes to the phase uncertainty. The envelopes, I±, can be put as a function of background irradiance and visibility according to the following expressions: I±=I0(1±V),(24) and therefore, the irradiance can be recast as: I=I+ 2(1+cosφ)+I− 2(1−cosφ).(25) If we consider separately the variation of each envelope, we obtain: I=I±+ΔI± 2[1±cos(φ+Δφ)]+I∓ 2[1∓cos(φ+Δφ)] (26) Proceeding similarly to the previous case, and after some simple calculations, we arrive at: cosφ(Δφ)2/2+sinφΔφ−ΔI±(1±cosφ)/(2I0V) = 0,(27) which is similar to Eq. (19). Therefore, solving the equation we obtain the contribution of the envelopes to the phase uncertainty: Δφ±= |tanφ|[ 1+|cosφ|(1±cosφ)ΔI±/(I0Vsin2φ) √−1],(28) with ΔI±positive. In particular, there is no contribution of the upper envelope to the phase uncertainty when φ= (2m+1) π ,m∈Z,and there is not contribution of the lower envelope when φ=2m π . On the other hand, maximum uncertainty is Δφ±= 2ΔI±/(I0V) √. Putting all together and applying the law of propagation of uncertainties we obtain: Δφ= Δ2φ0+Δ2φ++Δ2φ− √.(29) Examples of each relative contribution to the total phase uncertainty are shown in Fig. 4. As uncertainty decreases, the contribution of the irradiance approaches the sum of the contributions of the two envelopes. 3.5. Phase ambiguity According to Eq. (3), the measured refractive index, here n k , is a function of an arbitrary integer k for a given wavelength and sample: nk=λ dk+a,(30) with a independent of k. Because of k being an arbitrary integer, the raw refractive index measured with RISBI is also arbitrary in terms of k. The variation of the refractive index produced by two consecutive values of k is: Δn=λ d.(31) That means that the error in the refractive index changes in steps of Δn. Therefore, in the case that the reference refractive index is known within a band smaller than Δn, the uncertainty associated to k can be considered null. To overcome the ambiguity, the refractive index of the sample at a fixed wavelength, λ 0 , is usually measured by a second method, which provides a way to determine k. In general terms, the result of measuring the reference refractive index by whatever procedure is n 0 ±δn, where n 0 is the measured refractive index at the wavelength λ 0 and δn its uncertainty. Hence, it can be said that the real value of the refractive index at λ 0 is contained in the interval [n 0 - δn, n 0 +δn] with a high degree of confidence. This implies that we can get the correct value of k provided that: δn⩽Δn0 2=λ0 2d,(32) because in this case there is only one possible value of n k in the above interval [n 0 - δn, n 0 +δn]. In consequence, n k is the value that must be taken as correct instead of the measured reference refractive index n 0 . On the other hand, if Eq. (31) is not fulfilled, the value of k cannot be reliably assured. In view of this discussion, it can be said that the uncertainty in the value of k is: Δk=floor(2d λ0 δn)(33) Fig. 3. Mathematical variation of the phase (dashed line) and the associated uncertainty (continuous line). Y. Arosa et al. Measurement 187 (2022) 110225 5 Here we assume that near λ 0 , δn ≈δn 0 . So, to reduce Δk as much as possible, it is better to measure the reference refractive index at a longer wavelength. The k uncertainty contributes to refractive index uncertainty as: Δn λ=1 dfloor(2d λ0 δn)(34) In Fig. 5 this contribution is plotted for λ =λ 0 as a function of the sample thickness, d. For a fixed λ, Δn makes a jump just when the argument of the floor function is an integer and, then decreases with d until another jump arises. In this figure, the upper envelope is just 2δn/ λ 0 , and the lower envelope is 2δn/ λ 0 – 1/d. So, independently of the value of d, the maximum contribution to the refractive index uncertainty at wavelength λ is: 2δn λ/ λ 0 . 3.6. Fringe spacing The local fringe spacing, Λ( σ ), corresponds to the distance in wavenumber between two consecutive maxima or minima. It is also twice the distance between consecutive extremes (i.e., the absolute difference between a maximum and its adjacent minimum). So, to determine Λ( σ ), the extremes of the irradiance must be located. Calling σ ±= σ ±Λ/4the wavenumber for two consecutive minima and Fig. 4. Phase uncertainty related to irradiance Δφ 0 (blue), upper envelope Δφ + (green) and lower envelope Δφ - (red). In each case, the irradiance uncertainty is constant and equal to 5 ×10 −1 (top), 1 ×10 −1 (middle) and 5 ×10 −2 (bottom). (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.) Fig. 5. Refractive index uncertainty normalized by the reference wavelength and caused by phase ambiguity against sample thickness. In this example, λ 0 = 633 nm and δn =1 ×10 −4 . Y. Arosa et al. Measurement 187 (2022) 110225 6 maxima, Λ( σ ) takes the form: Λ( σ ) = 2( σ +− σ −), σ = ( σ ++ σ −)/2,(34) and for the uncertainty: ΔΛ( σ ) = 2 (Δ σ +)2+(Δ σ −)2 √.(35) By definition, irradiance maxima belong to the previously defined upper envelope, I + , and irradiance minima belong to the lower envelope, I - . As discussed above, a change in irradiance leads to a change in phase which, in turn, varies the wavenumber of the irradiance extrema. However, a change in the value of envelopes, does not affect the position of extrema but only affects to their value. So, the uncertainty in wavenumber can be written as (see Eq. 7): Δ σ =Δφ 2 π [dng−ngA(d+l)],(36) where Δφ is related to ΔI by Eq. (20). Note that once the period is determined using Eq. (34) for two extrema, it can be interpolated for any value of σ , and, for this reason, Eq. (36) applies to every σ It can also be calculated directly from the irradiance pattern by noticing that: and taking care that σ ±= σ ±Λ/4when the uncertainty at σ is computed using Eq. (35). 4. Results and discussion In order to evaluate the impact of the different uncertainty sources on the retrieval of refractive and group indices over a broad spectral range, we analyze in this section measurements of standard experiments carried out in our laboratory. Although results are specific to our experimental setup, the analysis is general enough to be applicable to measurements with typical RISBI devices. The main features of our RISBI system have been described elsewhere [10]. It can perform realtime dispersion measurements in a broad spectral range, from 260 to 1650 nm. With the purpose of keeping the discussion as general as possible, the results shown in this work are restricted to the interval from 400 to 1000 nm, where only one single spectrometer is used. Despite RISBI can operate with just two spectral acquisitions, one of the interferogram and another of the background irradiance, we have repeated each one of them 10 times to have statistically meaningful measurements. Indeed, it could be possible to extract the background irradiance from the interferogram, but we prefer to acquire it separately, since it allows for a double-check of the validity of the data. In Fig. 6 we show how a standard measurement with our device looks like. As it is usual in our experiments, the interferogram contains the stationary phase point, which assures good visibility in the whole spectral range, even for high dispersive samples. To obtain information about the sample dispersion, the phase must be extracted from Eq. (1). There are a variety of methods to extract the phase from a rapid varying interferogram such as the one shown in Fig. 6. For example, applying the Hilbert, Fourier, and wavelet transforms, phase shifting methods, or phase calculation from minimum and maximum detection [39–41]. Unfortunately, many of these methods do not work well near the stationary phase point. Here, we use the procedure detailed in [10,39]. First, we separate the measured background irradiance in Eq. (1) (blue line in Fig. 6); second, we determine the upper and lower envelopes by interpolation of their maxima and minima, respectively. In this step, the stationary phase point must be omitted. The visibility function is obtained by subtraction of the two envelopes, it is removed from the equation and the cosine is extracted. Finally, the phase principal value is obtained by applying the arc cosine function. In this section, we present results obtained with a fused silica optical window, 2 mm thick. Characteristics and manufacturer’s specifications are detailed in Table 1. As in the preceding section, we analyze the impact on the measurement of each parameter separately. Fig. 6. Example of interferogram and background irradiance against pixel number. Table 1 Sample parameters. Diameter 20.0 mm Material UV Grade Fused Silica Antireflection Coating Uncoated Surface Flatness λ/10 @ 632.8 nm Surface Quality 20–10 scratch-dig Thickness 2.0 mm Thickness Tolerance ±0.1 mm Wavefront Distortion λ/10 @ 632.8 nm Parallelism <5 arc seconds Clear Aperture Central, 80% of diameter Diameter Tolerance +0.0/-0.2 mm [I( σ +)− I0( σ +)]/V( σ +) = cos[φ( σ +Λ/4)] = − cos[φ( σ −Λ/4)] = [I0( σ −)− I( σ −)]/V( σ −),(37) Y. Arosa et al. Measurement 187 (2022) 110225 7 4.1. Spectrometer calibration A low-pressure mercury discharge lamp was used to carry out the spectrometer spectral calibration, which is performed without sample in the interferometer. The spectra lines of this lamp in the range of measurement are shown in Fig. 7. Only the marked lines were used for calibration purposes. In order to remove any dispersion in the interferometer and ensure that induced phase difference is a linear function of the wavenumber in air, two interferograms corresponding to two different optical paths in air were taken and their phases subtracted. The computation of phase and its uncertainty was done using the method described in Section 3.4. Among the three sources of uncertainty (slope and intercept of the linear regression, and phase), the phase is the most important one. Hence, the calibration can be considered correct, but there is a need of improving the phase measurement (this will be discussed below, in Section 4.4). Fig. 8 shows that this phase uncertainty produces fast oscillations in the uncertainty of the (a) wavenumber and (b) wavelength within a period. In a real experiment, such a rapid change in uncertainty will be unrealistic and the phase is averaged over one cycle, which corresponds to the red line in Fig. 8(b). Fig. 7. Spectrum of a low-pressure Hg discharge lamp. Red points correspond to the lines used to perform the fit. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.) Fig. 8. (a) Wavenumber and (b) wavelength uncertainty. The red line is the local mean. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.) Fig. 9. Ten different measurements of the sample thickness, d, and their mean (green line). (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.) Table 2 Sample thickness uncertainty and mean. Uncertainty (nm) Mean (mm) Resolution 58 – Flatness 9 – parallelism 69 – Statistical 135 2.05713 Total 160 2.05713 Y. Arosa et al. Measurement 187 (2022) 110225 8 4.2. Path difference in air The slope of the linear regression of the calibration process can be used to extract path difference l. The uncertainty of this magnitude is straightforward to obtain, and it is Δl =141 nm for a path difference of l =1.9860 mm. Notice that these magnitudes correspond to a Michelson interferometer where light passes through the sample twice and that it will be the half in a Mach-Zehnder type interferometer. 4.3. Sample thickness The fused silica plate thickness was measured with a high accuracy micrometer with resolution of 100 nm. The measure was repeated 10 times to get its statistical mean and its standard deviation (see Fig. 9). To calculate the uncertainty due to the sample flatness and parallelism, a beam diameter of 5 mm and centered in the sample was considered. Results are shown in Table 2. Please, note that in the case of using a Michelson interferometer, light passes through the sample twice, so both, means and uncertainties of the table, have to be multiplied by a factor of two. 4.4. Phase There are several sources of errors that contribute to phase uncertainty. On the one hand, mechanical and thermal vibrations vary the phase in the interferometer and so the interference pattern; on the other hand, lack of visibility and noise in the spectrometer deteriorate the image interferogram. In order to take into account the different contributions, we first took up to 10 interferograms and calculated their mean plus the standard deviation, as well as the mean and standard deviation of their envelopes (in blue, in Fig. 10). Furthermore, we applied a filter in the Fourier space and quantified the difference between the filtered and unfiltered interferograms (in green, in Fig. 10) as well as the difference of the corresponding envelopes. This gives an idea of the influence of noise. As a limit, absolute values of the spectrum smaller than one thousandth of the maximum value were filtered. Both statistical and noise contributions to the uncertainty of the irradiance are noisy, however the statistical contribution is greater in magnitude. Regarding the uncertainty of the envelopes, it is mainly constant with large modulations. While we believe that those modulations at the border of the spectrum are caused by lack of visibility, we do not recognize any possible cause of those in the center of the spectrum. However, note that they are smaller than 1% of the values of the envelope. In Fig. 11 the three contributions to the phase uncertainty are separately shown. In agreement with the analysis in Section 3.4, all contributions present oscillations from period to period following Eqs. (21) and (28). The maxima and minima (zeros) of the phase contributions related to the upper and lower envelopes alternate at phase values multiple of π , coinciding with the maxima of the phase contribution related to the irradiance, while the minima of the later (greater than zero) occur when the phase is an odd multiple of π /2. Except at the spectrum borders, the greater contribution is related to the irradiance uncertainty, followed by the one related to the upper envelope. In Fig. 12, which is taken as the final result of this section, the total phase uncertainty and its local average between 0.1 and 0.3 rad are shown. 4.5. Phase ambiguity The uncertainty due to the phase ambiguity, Δk, depends on the sample thickness, the wavelength at which the reference refractive index is measured, λ 0 , and the uncertainty of this measure, δn. In Fig. 13 we plot Δk against δn for our sample in both Michelson and Mach-Zehnder type interferometers, taking as reference the emission wavelength of a He-Ne laser, 632.8 nm. In the first case, there is no contribution of the phase ambiguity for δn less than 7.7 ×10 −5 (the double, in the case of a Mach-Zehnder interferometer). This accuracy can be easily achieved by widely used refractometric techniques as those based on many interferometric approaches or minimum deviation methods. Indeed, commercial refractometers are available with an accuracy in the order of 5 ×10 −5 or better. We note that the impact of the phase ambiguity in the refractive index uncertainty increases with sample thickness, just the opposite to the contributions of phase, thickness, and path difference in air. In Fig. 14(a) it is shown the largest sample thickness for which the phase ambiguity contribution to the refractive index uncertainty is zero as a function of δn. In Fig. 14(b) it is plotted the contribution to the refractive index uncertainty of the minimum value of the sample thickness that produces Δk =1, at a wavelength of 1 μ m, for a varying δn too. As before, the reference wavelength is 632.8 nm. 4.6. Fringe spacing As discussed in Section 3.6 the local periodicity is first calculated as Λ( σ ) = 2( σ +− σ −)with σ ±= σ ±Λ/4, corresponding to maxima and minima of the interferogram. That means that σ is close to a point where the irradiance equals the background (φ = π /2). Then, the periodicity is interpolated to cover all the measured points. The result is plotted in Fig. 15. The periodicity goes from approximately 10 nm at the spectrum borders to infinity at the equalization wavelength. Since the periodicity is proportional to the phase uncertainty, it shows oscillations as depicted in Fig. 16. 4.7. Spectrometer calibration and material dispersion The wavenumber and wavelength uncertainties associated to the spectrometer calibration were modeled in section 3.1 (Eqs. 13 and 14) and, in section 4.1, they were quantified for the Vis-NIR spectrometer of our RISBI device (Fig. 8). The next step consists in introducing the wavelength uncertainty in Eq. (4) to compute its contribution to the total refractive index uncertainty (third term in the square root). Since the expression of the group index in Eq. (9) is independent of wavelength, it means that there is not a similar contribution to the group index uncertainty. Therefore, it could seem that the measured group index dispersion is independent of spectrometer calibration and of any miscalibration. This is meaningless. The explanation is that Eq. (10) refers only to the group index uncertainty that comes from the effect of the parameters that appear in the expression used to compute the group Fig. 10. Statistical (blue) and noise (green) contribution to irradiance (top), upper (medium) and lower (bottom) envelope uncertainty. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.) Y. Arosa et al. Measurement 187 (2022) 110225 9 index given in Eq. (9). However, the dependence on Δλ, and thus on any calibration error, appears as soon as the dispersion curve, n g (λ), is considered. That is, if for a particular wavelength λ there is a miscalibration Δλ, it means that the value of the group index at this wavelength will be proportional to Δλ, and the same can be say about the refractive index uncertainty. Hence, apart from the contribution of wavelength miscalibration to the refractive index uncertainty taken into Fig. 11. Contributions to the phase uncertainty related to the irradiance (blue), upper envelope (green) and lower envelope (red). (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.) Fig. 12. Total phase uncertainty and its local average. Fig. 13. Error in k as a function of uncertainty of the reference refractive index for a Mach-Zehnder (blue line) and Michelson (red line) type interferometer. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.) Fig 14. Largest d for which Δk is zero as a function of δn (top) and the contribution to refractive index uncertainty at a wavelength of 1 μ m if Δk = 1 (bottom). Y. Arosa et al.