HPLC with Fluorescence Detection for Determination of Bisphenol A in Canned Vegetables: Optimization, Validation and Application to Samples from Portuguese and Spanish Markets

Vilarinho, Fernanda; Lestido Cardama, Antía; Sendón García, Raquel; Rodríguez Bernaldo de Quirós, Ana Isabel; Vaz, María de Fátima; Sanches Silva, Ana Teresa

Abstract

Bisphenol A (BPA) is one of the chemicals used to produce both polycarbonate plastics and epoxy resin coatings. Research has shown that small amounts of BPA can migrate into the foods and beverages enclosed in these types of containers. In this research, an analytical method based on high-performance liquid chromatography with fluorescence detection (HPLC-FLD) was developed and validated for the determination of BPA in canned vegetables. The results were confirmed by liquid chromatography coupled to tandem mass spectrometry (LC-MS/MS). Attenuated total reflectance Fourier transform infrared spectroscopy (ATR-FTIR) was performed, to identify the coating material of each tin can. Nineteen cans of vegetables were taken as study samples (eleven samples from the Spanish market, and eight samples from the Portuguese market). Excellent linear correlation (r2 = 0.9999) was observed over the range of 0.01 to 0.25 mg/L. Limit of detection (LOD) and limit of quantification (LOQ) values were calculated to be 0.005 mg/kg and 0.01 mg/kg, respectively. Good recoveries, between 72% and 90% were obtained at three different levels of concentration (RSD% = 4.6). BPA was not detected in the samples. The proposed HPLC-FLD was found to be suitable for the determination of BPA in canned vegetables

Full text

coatings Article HPLC with Fluorescence Detection for Determination of Bisphenol A in Canned Vegetables: Optimization, Validation and Application to Samples from Portuguese and Spanish Markets Fernanda Vilarinho 1, Antia Lestido-Cardama 2, Raquel Sendón2, Ana Rodríguez Bernaldo de Quirós2, Maria de Fátima Vaz 3and Ana Sanches-Silva 4,5,* 1Department of Food and Nutrition, National Institute of Health Dr. Ricardo Jorge, I.P., 1600-560 Lisbon, Portugal; [email protected] 2Department of Analytical Chemistry, Nutrition and Food Science, Pharmacy Faculty, University of Santiago de Compostela, 15705 Santiago de Compostela, Spain; [email protected] (A.L.-C.); [email protected] (R.S.); [email protected] (A.R.B.d.Q.) 3IDMEC, Instituto Superior Técnico, Departamento de Engenharia Mecânica, University of Lisbon, 1649-004 Lisbon, Portugal; [email protected] 4National Institute for Agricultural and Veterinary Research (INIAV), 2780-157 Vairão, Portugal 5Center for Study in Animal Science (CECA), ICETA, University of Porto, 4099-002 Porto, Portugal *Correspondence: [email protected] Received: 9 June 2020; Accepted: 29 June 2020; Published: 30 June 2020   Abstract: Bisphenol A (BPA) is one of the chemicals used to produce both polycarbonate plastics and epoxy resin coatings. Research has shown that small amounts of BPA can migrate into the foods and beverages enclosed in these types of containers. In this research, an analytical method based on high-performance liquid chromatography with fluorescence detection (HPLC-FLD) was developed and validated for the determination of BPA in canned vegetables. The results were confirmed by liquid chromatography coupled to tandem mass spectrometry (LC-MS/MS). Attenuated total reflectance Fourier transform infrared spectroscopy (ATR-FTIR) was performed, to identify the coating material of each tin can. Nineteen cans of vegetables were taken as study samples (eleven samples from the Spanish market, and eight samples from the Portuguese market). Excellent linear correlation (r 2 =0.9999) was observed over the range of 0.01 to 0.25 mg/L. Limit of detection (LOD) and limit of quantification (LOQ) values were calculated to be 0.005 mg/kg and 0.01 mg/kg, respectively. Good recoveries, between 72% and 90% were obtained at three different levels of concentration (RSD% =4.6). BPA was not detected in the samples. The proposed HPLC-FLD was found to be suitable for the determination of BPA in canned vegetables. Keywords: bisphenol A; epoxy resins; canned vegetables; HPLC-FLD; LC-MS/MS 1. Introduction People are exposed to contaminants through several sources. Packaging can be a potential source of contamination through the migration of substances from the packaging material into the food [ 1 ]. This fact is especially relevant, as most of the food found at the supermarkets is packaged in order to prevent its deterioration and to extend its shelf life [ 1 – 3 ] In order to ensure food safety, the monitoring of migration from food contact materials has become a priority issue. For researchers, the study of the contaminant bisphenol A (BPA) and its impact on the endocrine system are of major concern, because BPA exhibits estrogenic properties arising from its structural resemblance to the human 17β-estradiol [4] and disturbs the endocrine and reproductive systems [5–7]. Coatings 2020,10, 624; doi:10.3390/coatings10070624 www.mdpi.com/journal/coatings Coatings 2020,10, 624 2 of 12 Bisphenol A [2,2-bis(4 hydroxyphenyl)propane] has the chemical formula C15H16O2and results from the condensation of 2 mol of phenol with 1 mol of acetone in the presence of an acid catalyst [ 4 ]. BPA is mainly (approximately 95%) used in the production of synthetic polymers including epoxy resins (as the internal coating of food and beverage cans) and polycarbonates (storage containers, reusable plastic bottles, plates, among others) [ 2 , 4 ]. BPA is also used in the production of plasticized PVC as inhibitor of the polymerization of vinyl chloride and, in some plasticizers as antioxidant [8]. With several limitations, BPA is authorized to be used as a monomer in the manufacture of plastic materials intended to come into contact with food in the European Union. According to Regulation (EU) no. 10/2011 [ 9 ] and its amendments, BPA has a specific migration limit of 0.05 mg/kg [ 10 ]. Moreover, BPA is not authorized to be used for the manufacture of polycarbonate infant feeding bottles [ 11 ] and polycarbonate drinking cups or bottles which, due to their spill proof characteristics, are intended for infants and young children (Regulation (EU) no. 2018/213) [10]. In recent years many methods for measuring the BPA in food have been developed. In 2019 F. Vilarinho et al. [ 4 ] carried out an exhaustive compilation and discussion of the analytical methodologies used for the determination of BPA as result of its migration from food contact materials. The determination of BPA in food is mainly performed by liquid chromatography (LC) [ 4 , 12 , 13 ] with different detectors: ultraviolet (LC/UV) [ 14 ], fluorescence (LC/FLD), electrochemical detection (LC/ECD), mass spectrometry (LC/MS), and tandem mass spectrometry (LC/MS-MS) [ 15 ] or by gas chromatography (GC) coupled with mass spectrometry (GC/MS) and tandem mass spectrometry (GC/MS-MS) [ 4 , 13 , 16 ]. According to Vilarinho et al. [ 4 ] the MS detector is the most usually used for both LC and GC, due to its unquestionable advantages which comprehend high selectivity, high sensitivity, and universality. It is important to notice that the choice of the detector, greatly influences the detection limit (LOD) and the quantification limit (LOQ) [4]. In the present study, the development and validation of an analytical method by high-performance liquid chromatography with fluorescence detection (HPLC-FLD) to quantify BPA in canned vegetables are reported. The liquid chromatography-tandem mass spectrometry (LC-MS/MS) was used to confirm the results obtained. Attenuated total reflectance Fourier transform infrared spectroscopy (ATR-FTIR) was performed to identify the coating material of each of cans of samples selected for the study. Nineteen canned vegetables (eleven samples from the Spanish market, and eight samples from the Portuguese market) were evaluated in order to check their compliance with the legislation in force in the European Union. 2. Materials and Methods 2.1. Chemicals and Standard Solutions All reagents were analytical grade. Acetonitrile LC Grade (Lichrosolv, Merck KGaA, Darmstadt, Germany). Methanol LC-MS Grade was provided from Merck (Darmstadt, Germany). Ultrapure water (type I) was obtained from an Automatic Plus purification system (Wasserlab, Navarra, Spain). Standard Bisphenol A (BPA; 2,2-bis(4-hydroxyphenyl)propane; 4,4’-isopropylidene-diphenol), with a purity >99% was acquired from Aldrich-Chemie (Steinheim, Germany). BPA stock solution was prepared by dissolving the BPA in acetonitrile (ACN) in order to obtain 1000 mg L −1 . Standard solutions of concentration between 0.005 and 0.25 mg L −1 were prepared by diluting the stock solution with the ACN/Water (10:90). 2.2. Samples Preparation Nineteen commercial canned vegetables were purchased in three different local supermarkets of Lisbon (Portugal) and Santiago de Compostela (Spain). The canned samples have different food products: namely, three samples of canned fine peas, two of artichoke hearts, two of sweet corn, one of chilli peppers, one of asparagus (thick gauge), one of red peppers—all natural, one of rolled mushrooms, one of Almagro eggplant, one of whole peeled tomato, one of white cabbage, one of cooked Coatings 2020,10, 624 3 of 12 lentils, one of vegetables of Macedonia, one of mung bean sprouts and two of green beans (Table 1). All canned foodstuffs were sampled in duplicate, therefore the result is the mean of two independent measurements Samples were stored at room temperature until analysis. Table 1. Samples selected for the present study (P—Portugal; E—Spain). Origin (Country) Samples Used for Trial 1E Alegrias Riojanas 2E Asparagus buds 3E Red peppers—whole in brine 4E Artichoke hearts 5E Sweet corn 6E Peas with extra-fine carrots 7E Laminated mushrooms 8E Eggplant Almagro 9E Fine peas 10E Whole peeled tomato 11E White cabbage 1P Very fine peas 2P Cooked lentils 3P Vegetables of Macedonia 4P Sweet corn 5P Mungo bean sprouts 6P White cooked quinoa 7P Green beans 8P Artichoke Hearts Following the preparation of foods described by Ner í n et al. (2002) [ 17 ] and because the objective of the paper was also to first screen BPA in canned vegetables, the liquid phase in contact with canned foods was analyzed. To determine the amount of BPA present in each sample, the net content of each can was leaked into a container. Then 900 µ L were removed from this container and 100 µ L of can was added. This mixture was filtered and then injected into HPLC-FLD and LC-MS/MS. 2.3. HPLC-FLD The high-performance liquid chromatography (HPLC) system used in this study was an Agilent Technologies 1200 Series (Waldbronn, Germany) constituted by a quaternary pump, a degassing device, an autosampler, a column thermostat system, a diode array detector, a fluorescence detector and an Agilent ChemStation for LC and ChemStation for LC 3D Systems software (version B.04.0x). 2.4. LC-MS/MS The liquid chromatography-tandem mass spectrometry (LC-MS/MS) system used in this study consisted of an Accela autosampler, a column thermostat system and Accela 1250 pump fitted with a degasser, coupled to a TSQ Quantum Access Max Triple Quadrupole controlled by Xcalibur 2.1.0 (Thermo Fisher Scientific, San Jos é , CA, USA). MS data were acquired in the negative Atmospheric Pressure Chemical Ionization (APCI) mode. The operating conditions were: spray voltage 2500 V; nebulizer gas (N 2 ), 35 psi; vaporizer temperature 400 ◦ C; capillary temperature 350 ◦ C. As the collision gas was used Argon (1.5 mTorr). The selected precursor ion for BPA was m /z 227.2 and the two SRM transitions monitored were 227.2–212.1 and 227.2–133.2 m/z with a collision energy of 21 and 29 V respectively (Table 3). 2.5. Chromatographic Conditions The chromatographic conditions for the HPLC-FLD and the LC-MS/MS analyses are similar. Table 2shows the chromatographic conditions for BPA separation/quantification and Table 3indicates Coatings 2020,10, 624 4 of 12 the gradient elution conditions [ 18 ]. For LC-MS/MS analysis the organic solvent used was methanol instead of acetonitrile. Table 2. Chromatographic conditions for the determination of bisphenol A (BPA) as results of its migration to canned vegetables. Chromatographic Conditions of HPLC-FLD and LC-MS/MS Methods Analytical column Luna®C18 (2) 100 Å (150 mm ×3 mm, 5 µm particle size) Temperature 25 ◦C Injection volume 20 µl FLD detector λem 305 nm; λex 225 nm MS spectrometer conditions Acquired in Negative atmospheric pressure chemical ionization (APCI) mode Spray voltage 2500 V Nebulizer gas (N2), 35 psi Vaporizer temperature 400 ◦C Capillary temperature 350 ◦C Collision gas Argon (1.5 mTorr) Selected precursor ion for BPA m/z 227.2 SRM transitions monitored 227.2 >212.1, collision energy: 21V SRM transitions monitored 227.2. >133.2, collision energy: 29V Table 3. Gradient elution conditions of HPLC-FLD and LC-MS/MS methods. Time (min) %A %B* 0.0 70.0 30.0 2.00 70.0 30.0 23.00 0.0 100.0 30.00 0.0 100.0 32.00 70.0 30.0 35.00 70.0 30.0 *B corresponds to acetonitrile in HPLC-FLD or methanol in LC-MS/MS. 2.6. ATR-FTIR Infrared spectra were acquired using an ATR-FTIR spectrometer (ATR-PRO ONE FT-IR4700, Jasco, Japan) equipped with a Diamond optical element (in the range from 4000–650 cm −1 ). ATR-FTIR spectrometer was controlled by the software Spectra Manager ™ Suite (version 2). The identification was carried out by using the IR Spectral libraries of Polymers & Related Compounds (Bio-Rad Laboratories, Inc., Hercules, CA, USA) using the software KnowItAll®(version 17.4.135.B). 2.7. Recovery To determine recovery of BPA, a spiking experiment using a canned vegetable BPA free (blank) was performed. The spiking levels chosen were 0.01, 0.05 and 0.25 mg kg −1 of BPA, respectively. HPLC results were plotted into the BPA calibration curve. Spiking assays were performed by different operators, in three different days and the recovery was calculated by comparing the theoretical spiking and practical BPA value obtained from HPLC. Coatings 2020,10, 624 5 of 12 2.8. Precision Repeatability and precision were determined by running spiking experiments on a representative sample at 3 different spiking levels (0.01, 0.05 and 0.25 mg kg −1 ). Repeatability was carried out in a single day and precision (internal reproducibility) was performed during three different days and with different operators. Data are expressed in terms of % of relative standard deviation (%RSD). 3. Results and Discussion 3.1. Method Validation Validation parameters including linearity, precision (internal reproducibility), limit of detection (LOD) and limit of quantification (LOQ) were determined according to the International Conference on Harmonisation (ICH) guidelines Q2 (R1) [19,20]. The use of a series of standards solutions of known concentration allowed to test the linearity of the method. The calibrations curves were constructed using six concentration levels (0.005, 0.010, 0.025, 0.050, 0.100 and 0.250 mg L −1 ) and they were fitted to a linear equation. The linearity was tested in acetonitrile/water (10:90, v/v) as solvent. Three calibration curves were performed in three different days. All of them have shown good linearity in the studied concentration range with coefficients of determination (r 2 ) equal to 0.9999, indicating suitability for BPA quantification in the selected calibration range. Table 4shows the parameters of linearity. Table 4. Recovery, repeatability and intermediate precision of the HPLC-FLD method. Compound Recovery (%) (n=6) Repeatability (RSD%) (n=6) Intermediate Precision (RSD%) (n=18) BPA 0.01a) mg kg−1 0.05 a) mg kg−1 0.25 a) mg kg−1 0.01 a) mg kg−1 0.05 a) mg kg−1 0.25 a) mg kg−1 0.01 a) mg kg−1 0.05 a) mg kg−1 0.25 a) mg kg−1 90 77 72 9.94 3.48 3.26 4.53 4.70 0.57 a) Fortification level. The limits of detection and quantification, calculated according to ACS guidelines [ 21 ] (defined as three and ten times the height signal of the noise level, respectively) were presented in Table 5. ThemethodpresentsenoughsensitivitytodetecttheBPAattheregulatorylevel,oncetheexperimentalLOD (0.005 mg kg −1 ) is considerably lower than the specific migration limit (SML) of 0.05 mg kg −1 established by the Regulation (EU) no. 10/2011 and its amendments, which are in force [4,10]. Table 5. Linearity, limit of detection and quantification of BPA according to the International Conference on Harmonisation (ICH) guidelines Q2 (R1), and using high-performance liquid chromatography (HPLC) with fluorescence detection (FLD). Day Equation r2LOD (mg kg−1) LOQ (mg kg−1) Range (mg kg−1) 1y=224.71x−0.1121 0.9999 0.005 0.01 0.01–0.25 2y=223.92x−0.2539 3y=219.54x−0.113 Figure 1shows the chromatogram obtained by HPLC-FLD with 0.005 mg kg −1 standard solution, that is the limit of detection of the method. As an important step in the validation process, is to demonstrate that the proposed methodology is fit to its purpose. This was evaluated through recovery assays, spiking a BPA free-sample. A working session with six replicates for each spiking level was performed in three different days. According to the results report in Table 4, it is confirmed that the method internal reproducibility is appropiate even at low concentrations, with average recoveries (n=18) ranging from 72 % to 90%. Repeatability Coatings 2020,10, 624 6 of 12 was always lower than 3.48%, except at 0.01 mg kg −1 , which was 9.94%, and intermediate precision was always lower than 4.70%, indicating good precision. Poustka et al. (2007) [ 22 ] in their research, reported the results of bisphenols monitoring in various categories of canned food obtained from the Czech retail market through the validation of analytical procedure HPLC coupled with FLD. They obtained recoveries, not below 75%, and LOD and LOQ of 0.002 and 0.01 mg kg−1, respectively. In its turn, Errico et al. (2014) [ 23 ] in its work performed the determination of bisphenol A (BPA) in canned tomatoes by a method based on solid-phase extraction followed by liquid chromatography coupled to fluorescence spectrophotometry obtaining a LOQ of 0.26 µg BPA kg−1tomato. Coatings 2019, 9, x FOR PEER REVIEW 6 of 11 Figure 1. Chromatograms obtained by HPLC-FLD with a 0.005 mg·k g−1standard solution: luminescence units (LU) vs. retention time (min). As an important step in the validation process, is to demonstrate that the proposed methodology is fit to its purpose. This was evaluated through recovery assays, spiking a BPA free-sample. A working session with six replicates for each spiking level was performed in three different days. According to the results report in Table 4, it is confirmed that the method internal reproducibility is appropiate even at low concentrations, with average recoveries (n = 18) ranging from 72 % to 90%. Repeatability was always lower than 3.48%, except at 0.01 mg kg−1, which was 9.94%, and intermediate precision was always lower than 4.70%, indicating good precision. Poustka et al. (2007) [22] in their research, reported the results of bisphenols monitoring in various categories of canned food obtained from the Czech retail market through the validation of analytical procedure HPLC coupled with FLD. They obtained recoveries, not below 75%, and LOD and LOQ of 0.002 and 0.01 mg kg−1, respectively. In its turn, Errico et al. (2014) [23] in its work performed the determination of bisphenol A (BPA) in canned tomatoes by a method based on solid-phase extraction followed by liquid chromatography coupled to fluorescence spectrophotometry obtaining a LOQ of 0.26 μg BPA kg−1 tomato. Figures 2 and 3 show the chromatogram obtained by HPLC-FLD with a fortification level of 0.01 mg kg−1and a chromatogram obtained by HPLC-FLD of a sample of “sweet corn” where BPA was not detected, respectively. Figure 2. HPLC-FLD chromatogram of a blank sample (“sweet corn”) with a fortification level of 0.01 mg kg−1. Luminescence unit (LU) vs Retention Time (min). Figure 1. ChromatogramsobtainedbyHPLC-FLDwitha0.005mg · kg −1 standardsolution: luminescence units (LU) vs. retention time (min). Figures 2and 3show the chromatogram obtained by HPLC-FLD with a fortification level of 0.01 mg kg −1 and a chromatogram obtained by HPLC-FLD of a sample of “sweet corn” where BPA was not detected, respectively. Coatings 2019, 9, x FOR PEER REVIEW 6 of 11 Figure 1. Chromatograms obtained by HPLC-FLD with a 0.005 mg·k g−1standard solution: luminescence units (LU) vs. retention time (min). As an important step in the validation process, is to demonstrate that the proposed methodology is fit to its purpose. This was evaluated through recovery assays, spiking a BPA free-sample. A working session with six replicates for each spiking level was performed in three different days. According to the results report in Table 4, it is confirmed that the method internal reproducibility is appropiate even at low concentrations, with average recoveries (n = 18) ranging from 72 % to 90%. Repeatability was always lower than 3.48%, except at 0.01 mg kg−1, which was 9.94%, and intermediate precision was always lower than 4.70%, indicating good precision. Poustka et al. (2007) [22] in their research, reported the results of bisphenols monitoring in various categories of canned food obtained from the Czech retail market through the validation of analytical procedure HPLC coupled with FLD. They obtained recoveries, not below 75%, and LOD and LOQ of 0.002 and 0.01 mg kg−1, respectively. In its turn, Errico et al. (2014) [23] in its work performed the determination of bisphenol A (BPA) in canned tomatoes by a method based on solid-phase extraction followed by liquid chromatography coupled to fluorescence spectrophotometry obtaining a LOQ of 0.26 μg BPA kg−1 tomato. Figures 2 and 3 show the chromatogram obtained by HPLC-FLD with a fortification level of 0.01 mg kg−1and a chromatogram obtained by HPLC-FLD of a sample of “sweet corn” where BPA was not detected, respectively. Figure 2. HPLC-FLD chromatogram of a blank sample (“sweet corn”) with a fortification level of 0.01 mg kg−1. Luminescence unit (LU) vs Retention Time (min). Figure 2. HPLC-FLD chromatogram of a blank sample (“sweet corn”) with a fortification level of 0.01 mg kg−1. Luminescence unit (LU) vs Retention Time (min). 3.2. LC-MS/MS In order to optimize and obtain the appropriate signal response in LC-MS/MS, acetonitrile and methanol were introduced as mobile phases. The best signal was achieved with methanol-water as a mobile phase. Coatings 2020,10, 624 7 of 12 MS data were acquired in selected reaction monitoring (SRM) mode once the optimization of the MS/MS parameters was performed using the built-in perfusion system. The ion m/z 227.2 has been assigned as the deprotonated molecule [M–H] − , the transition m/z 227.2 >212.1 was related with the additional loss of oxygen [M–H–O] − and the transition m/z 227.2 >133.2 with the loss of phenol group [4]. Coatings 2019, 9, x FOR PEER REVIEW 7 of 11 Figure 3. Chromatogram obtained by HPLC-FLD with a blank sample of “sweet corn”: luminescence unit (LU) vs. retention time (min). 3.2. LC-MS/MS In order to optimize and obtain the appropriate signal response in LC-MS/MS, acetonitrile and methanol were introduced as mobile phases. The best signal was achieved with methanol-water as a mobile phase. MS data were acquired in selected reaction monitoring (SRM) mode once the optimization of the MS/MS parameters was performed using the built-in perfusion system. The ion m/z 227.2 has been assigned as the deprotonated molecule [M–H]−, the transition m/z 227.2 > 212.1 was related with the additional loss of oxygen [M–H–O] − and the transition m/z 227.2 > 133.2 with the loss of phenol group [4]. LC-MS/MS is valuable to confirm the results of HPLC-FLD. However, this technique has a higher LOD (0.01 mg kg−1) than HPLC-FLD (0.005 mg kg−1). The Figure 4 shows the chromatogram obtained by LC-MS/MS for the 0.01 mg kg-1standard solution of the BPA. In our study, any of the samples presented detectable BPA according to HPLC-FLD, therefore it was not necessary to analyse them by LC-MS/MS. However, we decided to analyse the samples by LC-MS/MS and also none of them showed detectable amounts of BPA by this technique. Patrón 10ppb BPA en LC-MS/MS Muestra 5E en LC-MS/MS RT:3.91 - 31.95 SM:7B 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 Time (min) 0 20 40 60 80 100 120 140 160 0 20 40 60 80 100 120 140 160 Relative Abundance 10.03 8.49 12.80 9.89 13.38 6.85 11.31 14.23 12.51 14.46 8.18 8.06 9.44 11.11 14.86 15.05 17.47 16.28 6.62 15.53 5.80 16.83 17.79 5.65 18.45 19.90 21.77 4.42 5.55 22.60 18.92 20.05 21.22 22.91 20.47 24.95 23.19 31.77 29.97 29.30 28.36 25.49 26.10 27.56 30.24 5.54 5.86 4.91 6.39 7.01 8.69 7.86 9.61 9.68 10.88 11.62 11.95 12.38 12.93 14.29 15.12 15.32 16.19 16.54 17.32 18.67 19.42 20.19 20.68 21.80 22.38 23.02 24.07 25.01 29.45 28.0526.66 29.90 31.42 NL: 1.31E2 m/z= 132.70-133.70 F: - c APCI SRM ms2 227.200 [132.850-133.550, 211.750-212.450] MS 5E NL: 1.81E3 m/z= 211.60-212.60 F: - c APCI SRM ms2 227.200 [132.850-133.550, 211.750-212.450] MS 5E A B Figure 3. Chromatogram obtained by HPLC-FLD with a blank sample of “sweet corn”: luminescence unit (LU) vs. retention time (min). LC-MS/MS is valuable to confirm the results of HPLC-FLD. However, this technique has a higher LOD (0.01 mg kg −1 ) than HPLC-FLD (0.005 mg kg −1 ). The Figure 4shows the chromatogram obtained by LC-MS/MS for the 0.01 mg kg−1standard solution of the BPA. Coatings 2019, 9, x FOR PEER REVIEW 7 of 11 Figure 3. Chromatogram obtained by HPLC-FLD with a blank sample of “sweet corn”: luminescence unit (LU) vs. retention time (min). 3.2. LC-MS/MS In order to optimize and obtain the appropriate signal response in LC-MS/MS, acetonitrile and methanol were introduced as mobile phases. The best signal was achieved with methanol-water as a mobile phase. MS data were acquired in selected reaction monitoring (SRM) mode once the optimization of the MS/MS parameters was performed using the built-in perfusion system. The ion m/z 227.2 has been assigned as the deprotonated molecule [M–H]−, the transition m/z 227.2 > 212.1 was related with the additional loss of oxygen [M–H–O] − and the transition m/z 227.2 > 133.2 with the loss of phenol group [4]. LC-MS/MS is valuable to confirm the results of HPLC-FLD. However, this technique has a higher LOD (0.01 mg kg−1) than HPLC-FLD (0.005 mg kg−1). The Figure 4 shows the chromatogram obtained by LC-MS/MS for the 0.01 mg kg-1standard solution of the BPA. In our study, any of the samples presented detectable BPA according to HPLC-FLD, therefore it was not necessary to analyse them by LC-MS/MS. However, we decided to analyse the samples by LC-MS/MS and also none of them showed detectable amounts of BPA by this technique. Patrón 10ppb BPA en LC-MS/MS Muestra 5E en LC-MS/MS RT:3.91 - 31.95 SM:7B 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 Time (min) 0 20 40 60 80 100 120 140 160 0 20 40 60 80 100 120 140 160 Relative Abundance 10.03 8.49 12.80 9.89 13.38 6.85 11.31 14.23 12.51 14.46 8.18 8.06 9.44 11.11 14.86 15.05 17.47 16.28 6.62 15.53 5.80 16.83 17.79 5.65 18.45 19.90 21.77 4.42 5.55 22.60 18.92 20.05 21.22 22.91 20.47 24.95 23.19 31.77 29.97 29.30 28.36 25.49 26.10 27.56 30.24 5.54 5.86 4.91 6.39 7.01 8.69 7.86 9.61 9.68 10.88 11.62 11.95 12.38 12.93 14.29 15.12 15.32 16.19 16.54 17.32 18.67 19.42 20.19 20.68 21.80 22.38 23.02 24.07 25.01 29.45 28.0526.66 29.90 31.42 NL: 1.31E2 m/z= 132.70-133.70 F: - c APCI SRM ms2 227.200 [132.850-133.550, 211.750-212.450] MS 5E NL: 1.81E3 m/z= 211.60-212.60 F: - c APCI SRM ms2 227.200 [132.850-133.550, 211.750-212.450] MS 5E A B Figure 4. Chromatograms obtained by LC-MS/MS for the 0.01 mg kg −1 standard solution of the BPA: (A) the transition 227.2 >133.1 m/z; (B) the transition 227.2 >212.1 m/z. Coatings 2020,10, 624 8 of 12 In our study, any of the samples presented detectable BPA according to HPLC-FLD, therefore it was not necessary to analyse them by LC-MS/MS. However, we decided to analyse the samples by LC-MS/MS and also none of them showed detectable amounts of BPA by this technique. 3.3. ATR-FTIR For the analysis of the cans by ATR-FTIR, small pieces of each sample (can) were cut with an area suitable to be used in the spectrophotometer. Pieces of the cover, bottom and side of each tin can were analyzed. Subsequently, each piece of the sample was analyzed by ATR-FTIR regarding their inner part. The identification was carried out by comparing obtained spectra with those of the IR Spectral libraries of Polymers and Related Compounds (Bio-Rad Laboratories, Inc.) database. Analyzing the results listed in Table 6, it can be seen that polyester, phenoxy and epoxy coatings are those that are used in the samples. However, the polyester is the most used, as 78% of the samples have a polyester coating. Figure 5shows, as an example, two spectra obtained for the samples 2P (phenoxy coating) and 3S (polyester coating), respectively. Table 6. Main compound identified for each of the samples individually. Samples/Country Part of the Tin Can Evaluated Main Compound Identified 1S Inner Body Epoxy coating Inner Cover Polyester 1P Inner Bottom Polyester Inner Side Phenoxy coating Inner Cover Polyester 2S Inner Bottom Modified phenolic coating Inner Side Epoxy coating Inner Cover Epoxy coating 2P Inner Bottom Polyester Inner Side Phenoxy coating Inner Cover Polyester 3S Inner Bottom Epoxy coating Inner Side Polyester Inner Cover Polyester 3P Inner Bottom Polyester Inner Side Polyester Inner Cover Polyester 4S Inner Bottom Polyester Inner Side Polyester Inner Cover Polyester 4P Inner Bottom Polyester Inner Side Phenoxy coating Inner Cover Phenoxy coating 5S Inner Bottom Polyester Inner Side Polyester Inner Cover Polyester Coatings 2020,10, 624 9 of 12 Table 6. Cont. Samples/Country Part of the Tin Can Evaluated Main Compound Identified 5P Inner Bottom Polyester Inner Side Polyester Inner Cover Polyester 6S Inner Bottom Phenoxy coating Inner Side Polyester Inner Cover Polyester 6P Inner Bottom Phenoxy coating Inner Side Polyester Inner Cover Polyester 7S Inner Bottom Phenoxy coating Inner Side Phenoxy coating Inner Cover Polyester 7P Inner Bottom Polyester Inner Side Polyester Inner Cover Phenoxy coating 8S Inner Bottom Epoxy coating Inner Side Phenoxy coating Inner Cover Polyester 8P Inner Bottom Epoxy coating Inner Side Epoxy coating Inner Cover Phenoxy coating 9S Inner Bottom Polyester Inner Side Phenoxy coating Inner Cover Phenoxy coating 10S Inner Bottom Epoxy coating Inner Side Phenoxy coating Inner Cover Phenoxy coating 11S Inner Bottom Epoxy coating Inner Side Epoxy coating Inner Cover Phenoxy coating S: Spain; P: Portugal. Phenoxy resin has high molecular weight, a thermoplastic polymer having pendant hydroxyl groups and no oxirane ring [ 24 ]. Therefore, a broad peak of the hydroxyl group was observed at 3450.0 cm −1 in the FTIR spectrum, shown in Figure 5a [ 24 ]. In Figure 5b the band at 3454 cm −1 is related to stretching vibrations of OH groups. The bands in the range between 2800–3000 cm −1 correspond to stretching vibrations of CH groups such as CH 2 and CH 3 [ 25 ]. Weak bands at 1410 and 1462 cm −1 observed in the spectrum of polyester can be assigned to aromatic ring. The strong band at 1269 cm –1 that appears in the spectrum of polyester is due to the twisting vibration of CH 2 groups [ 25 ]. The similarity between the two spectra is because Phenoxy coating is a thermoplastic polymer derived from BPA ((CH3)2C(C6H4OH)2) and epoxy-epichlorohydrin (Cl–CH2–(C2H3O)) [26].