Full text
a Corresponding author: Eeva Boman, Department of Oncology, Tampere University Hospital, PO BOX 2000 (Teiskontie 35), FI-33521 Tampere, Finland; phone: (+3853) 311 63201; email: [email protected] Treatment accuracy without rotational setup corrections in intracranial SRT Eeva Boman,1,2a Mika Kapanen,1,2 Marko Laaksomaa,1 Hanna Mäenpää,1 Simo Hyödynmaa,2 and Pirkko-Liisa Kellokumpu-Lehtinen1,3 Department of Oncology,1 Tampere University Hospital, Tampere, Finland; Department of Medical Physics,2 Tampere University Hospital, Tampere, Finland; School of Medicine,3 University of Tampere, Tampere, Finland [email protected] Received 20 October, 2015; accepted 22 February, 2016 The aim of this study was to evaluate the impact of actual rotational setup errors on dose distributions in intracranial stereotactic radiotherapy (SRT) with different alternatives for treatment position selection. A total of 38 SRT fractions from 18 patients were retrospectively evaluated with rotational setup errors obtained from actual treatments. The planning computed tomography (CT) images were rotated according to online cone-beam CT (CBCT) images and the dose distribution was recalculated to the rotated CT images using three different patient positionings derived from: 1) an automatic 6D match neglecting rotation correction (Auto6D); 2) an automatic 3D match (Auto3D); and 3) a manual 3D match from actual treatment (Treat3D). The mean conformity index (CI) was 0.92 for the original plans and 0.91 for the Auto6D plans. The mean CI decreased significantly (p < 0.01) to 0.78 and 0.80 for the Auto3D and the Treat3D plans, respectively. The mean minimum dose of the planning target volume (PTVmin) was 91.9% of the prescribed dose for the original plans and 92.1% for the Auto6D plans, while for the Auto3D and the Treat3D plans PTVmin decreased significantly (p < 0.01) to 78.9% and 80.2%, respectively. No significant differences were seen between the Auto6D and the original treatment plans in terms of the dose parameters. However, the Auto3D and the Treat3D plans were statistically significantly inferior (p < 0.01) to the Auto6D and the original plans. In addition, a significant negative correlation (p < 0.01, |r| > 0.38) was found in the Auto3D and the Treat3D cases between the rotation error and CI, PTVmin or minimum dose of gross tumour volume. In SRT, a treatment plan of comparable quality to 6D rotation correction can be achieved by using 6D registration without a rotational correction in the selection of patient positioning. This was demonstrated for typical rotation errors seen in clinical practice. PACS number(s): 87.55, 87.57 Key words: SRT, rotational setup correction, patient positioning I. INTRODUCTION Brain metastases occur in 10%–30% of all cancer patients.(1) Treatment options are whole brain radiotherapy, local therapies (SRT or surgery), and steroids. Several trials have shown a benefit in local control, quality of life, and even improved survival for singleor oligometastatic patients treated with SRT.(2,3,4) In SRT, small brain lesions are treated with small GTV-to-PTV margins in a single or few fractions. Because of the relatively high dose per fraction (typically 6-21 Gy per fraction), steep dose gradients are required around the planning target volume (PTV) to avoid damage JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 4, 2016 86 86
87 Boman et al.: SRT accuracy without 6DOF correction 87 Journal of Applied Clinical Medical Physics, Vol. 17, No. 4, 2016 to surrounding organs, mainly the central nervous system.(5) Target localization uncertainty should be kept very small (typically within 1 mm) in SRT to not compromise the local control of the treatment(6) and to minimize the risk of injury to the surrounding brain parenchyma.(7) SRT presumes adequate patient immobilization. Thermoplastic stereotactic masks are nowadays commonly used in SRT treatments. These have been validated in many studies as fulfilling the high accuracy standard of SRT.(8,9) Image-guidance with orthogonal kV-images,(10) oblique images,(11) cone-beam computed tomography (CBCT) images,(12) reflecting markers with bite blocks,(13) and, more recently, patient surface matching systems(14) have shown to be sufficent in patient positioning in SRT keeping the residual setup errors under the desired limits. Intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT) have increased the dose conformity to PTVs and decreased the dose to normal tissues when compared to traditional static beams in linear accelerator-based treatments.(15) The use of VMAT also decreases the overall treatment time, reducing the possibility of intrafraction movements by the patient. The use of flattering filter-free (FFF) beams decreases the treatment time even further in high dose per fraction treatments.(16) To further increase accuracy in cranial SRT, patient rotational setup errors can be corrected using a six degrees of freedom (6DOF) couch.(17,18,19) In all of these studies, the correction of the rotational setup error increased plan quality and organ-at-risk (OAR) sparing. It is evident that in the patient setup, the use of the 6DOF couch improves treatment quality. However, the 6DOF couch may not be available in many cancer centers due to the high cost. Gevaert et al.(18) compared the calculated dose with and without rotational corrections and concluded a loss of 5% dose coverage when rotational correction was not used, pointing to a 0.5° rotation error as a threshold angle for correction. The aim of this study was to evaluate the added value of the use of 6DOF registration in treatment without the presence of the 6DOF couch system. For this purpose, only translational corrections were taken from the 6DOF registration and the dose was recalculated using this 6D patient positioning on rotated CT images to illustrate the effect of uncorrected actual rotations on dose distribution. The resulting dose distribution was compared to that of the original plan and to those obtained by accounting for residual position errors from the 3DOF registration and the actual treatment localized by a physician. II. MATERIALS AND METHODS Eighteen singleor oligometastatic SRT patients with a total of 38 fractions were retrospectively evaluated in this study. The number of treated fractions ranged from a single fraction to 5 fractions, depending on the tumor size and patient-specific details. Patients were immobilized using the BrainLab mask system (Brainlab AG, Feldkirchen , Germany). A slice thickness of 1 mm was used in the initial treatment planning CT (Philips Big Bore, Philips Medical Systems, Fitchburg, WI). 6D registration was used between the recent MRI and planning CT in Eclipse registration software (Aria 11, Varian Medical Systems, Palo Alto, CA). Treatment contouring and planning were performed with the Eclipse v.11 treatment planning system (Varian Medical Systems). GTV-to-PTV margins of 1–2 mm were used. The VMAT technique (Rapid Arc, Varian Medical Systems) with noncoplanar or coplanar subarcs was used in the treatment planning with 6 MV or 6 MV flattening filter-free (6MV-FFF) beams. High definition multileaf collimation with leaf width of 2.5 mm was used in the planning. The dose calculation was performed with a grid size of 1 mm using anisotropic analytical algorithm v.11 (Varian Medical Systems). The treatments were delivered on a TrueBeam STx 1.6 with HD-MLC (Varian Medical Systems). Patient positioning was based on online CBCT images with a slice thickness of 2 mm. The CBCT images were retrospectively analyzed. A schematic illustration of the workflow and the coordinate system are presented in Fig. 1. An automatic 6D registration based on skull volume was performed between the CBCT and original planning CT (CTplan) using Eclipse
88 Boman et al.: SRT accuracy without 6DOF correction 88 Journal of Applied Clinical Medical Physics, Vol. 17, No. 4, 2016 registration software (Varian Medical Systems). The resulting rotational deviations in all 3D directions (pitch, roll, yaw) were used in the rotation of the original CT images so that the resulting rotated CT (CTrot) presented the actual treatment orientation. The rotation isocenter was the center of the CT images. The CT rotation was done using homemade software on Matlab (The MathWorks, Inc., Natick, MA). After rotation, the new CTrot images were matched by automatic 6D registration to the CBCT image (CTrot-CBCT 6D). No rotational deviations should be seen in this registration. The actual treatment position (Treat3D) obtained from the oncologist’s 3D match was marked on the CBCT image and copied onto CTrot by using an automatic 6D registration between the CBCT and CTrot. Automatic 3D/6D registrations were used to copy the treatment position points for the Auto3D/Auto6D plans from CTplan to CTrot, respectively. All the registrations were done using the skull as the volume of interest. The dose distributions with actual rotations were calculated in CTrot for three different positionings of the patient — Treat3D, Auto3D, and Auto6D — and compared to the original treatment plan (planned). To evaluate the target coverage and plan quality, the Paddick conformity index (CI) and gradient index (GI) were used.(20,21) CI is defined as (1) CI = V100 PTV VPTV V100 PTV V100 in which V100PTV is the prescribed dose volume in PTV, VPTV is the PTV volume, and V100 is the volume which receives the prescribed dose. CI =1 means the best target coverage and lower values indicate a worse plan quality. GI is defined by (2) GI = V50 V100 in which V50 is the volume which receives higher than half of the prescribed dose. GI value increases as the dose gradient outside the target volume becomes shallower, indicating worse plan quality. Fig. 1. Workflow of the different image registrations and definition of rotations to construct the needed CT image set for localizations and dose calculations. The coordinate system with rotation axes is also presented.
89 Boman et al.: SRT accuracy without 6DOF correction 89 Journal of Applied Clinical Medical Physics, Vol. 17, No. 4, 2016 The Wilcoxon matched-pairs, signed-rank test (SPSS, v22, IBM Corp., Armonk, NY) for nonparametrically distributed data was performed to test the differences between the groups. Statistical significance was considered at p < 0.01 (two-tailed, |Z| > 2.576) including the Bonferroni correction between the groups planned–Auto6D, planned–Auto3D, planned–Treat3D, Auto6D–Auto3D, and Auto6D–Treat3D, which was made by dividing 0.05 by number of comparisons (i.e., by five). The dose distributions in four different cases (planned, Treat3D, Auto3D, and Auto6D) were studied for 38 different patient setups. The mean PTVvol was 8.0 cm3 (± 7.6 cm3), with a maximum volume of 26.6 cm3 and a minimum volume of 0.9 cm3. The locations and shapes of the PTVs are presented in Fig. 2. From 38 patient setups, there were two patients with more than one lesion and treatment isocenter was located outside of the PTV. One patient had 1 fraction with two lesions and the other had 3 fractions with three lesions resulting in four different patient setups. To demonstrate the use of 6D registration without rotational corrections for complex shaped targets or two separate spherical target volumes with reasonably large rotational errors, two patient examples are given. In the first example, the PTV is quite large (14.6 cm3) and complex in shape. The treatment plan consisted of two slightly noncoplanar full arcs. In the second example, two spherical PTVs were treated simultaneously using four noncoplanar half arcs. The volumes of the PTVs were 20.1 cm3 and 5.2 cm3. The distance between the PTVs was 3 cm. The GTV-to-PTV margin was 2 mm in both examples. III. RESULTS The mean rotational setup errors were -0.29° ± 0.88° (range from -2.1° to 1.6°), -0.26° ± 0.95° (from -2.0° to 2.4°) and -0.47° ± 0.80° (from -2.7° to 0.8°) for the pitch, roll and yaw rotations, respectively. The mean distance between actual treatment positioning and the position obtained from Auto6D registration was 1.2 mm (± 0.6 mm), ranging from 2.6 mm to 0.1 mm, while the corresponding distance between treatment position and that obtained from the Auto3D was 1.7 mm (± 0.8 mm), altering from 3.8 mm to 0.2 mm. CI, GI, maximum, and minimum doses of PTV and GTV (PTVmax, PTVmin, GTVmin) were investigated for each plan, and the mean, standard deviation (SD), minimum (Min), and maximum (Max) values are presented in Table 1 for each investigated group. The Wilcoxon test results for the groups planned–Auto6D, planned–Auto3D, planned–Treat3D, Auto6D–Auto3D, and Auto6D–Treat3D are shown in Table 2. These results show that the plans from the Auto6D group have almost the same mean CI, PTVmin, and GTVmin values as the planned group and no statistically significant difference was found between these two groups. The mean CI, PTVmin, and GTVmin values were reduced Fig. 2. Schematic illustration of the investigated PTVs for 18 patients.
90 Boman et al.: SRT accuracy without 6DOF correction 90 Journal of Applied Clinical Medical Physics, Vol. 17, No. 4, 2016 significantly (p < 0.01) for the Auto3D and Treat3D plans when compared to the original and Auto6D plans. Only a minor change was seen in the PTVmax among all the investigated groups. No significant difference was seen in GI values between the groups. Significant correlation (p < 0.01) was found between the square root of the quadratic sum of the rotations (quadratic sum) and CI in the Auto3D group, rauto3D = -0.46. In the Treat3D group, the correlation was nearly significant, rtreat3D = -0.38 (p = 0.02). Significant correlation (p < 0.01) was found between the rotation and PTVmin, rauto3D = -0.54 and rtreat3D = -0.52, and between rotation and GTVmin, rauto3D = -0.54 and rtreat3D = -0.42. The correlation was calculated for the percentage difference from the original plan values for the CI, PTVmin, and GTVmin values by dividing those with the original plan values to minimize the effect of variation in the original plans. No significant correlation was found in the Auto6D case between the rotation and dose parameters (rCI = -0.24 (p = 0.15), rPTVmin = -0.33 (p = 0.04), rGTVmin = -0.26 (p= 0.11)). CI, PTVmin, and GTVmin scaled to original plan values with respect to quadratic Table 1. The mean, SD, Min, and Max results for CI, GI, PTVmax, PTVmin, and GTVmin for the four different groups. CI Mean SD Min Max planned 0.92 0.02 0.82 0.94 Auto6D 0.91 0.03 0.82 0.94 Auto3D 0.78 0.12 0.43 0.94 Treat3D 0.80 0.10 0.62 0.93 GI Mean SD Min Max planned 4.23 0.83 3.14 5.93 Auto6D 4.23 0.82 3.14 5.91 Auto3D 4.24 0.83 3.14 5.94 Treat3D 4.22 0.83 3.12 5.92 PTVmax Mean SD Min Max planned 119.8 6.5 110.6 136.5 Auto6D 119.8 6.4 110.6 136.4 Auto3D 119.5 6.4 110.5 136.6 Treat3D 119.9 6.6 110.6 136.8 PTVmin Mean SD Min Max planned 91.9 3.1 84.9 96.4 Auto6D 92.1 4.0 81.8 97.5 Auto3D 78.9 11.7 53.7 97.0 Treat3D 80.2 10.7 55.1 96.1 GTVmin Mean SD Min Max planned 105.5 3.6 98.2 115.6 Auto6D 105.7 3.4 98.4 115.6 Auto3D 99.9 5.4 82.7 112.8 Treat3D 101.1 5.7 90.2 115.1 Table 2. The Zand p-values from the Wilcoxon signed-rank test for the related samples between the groups (|Z|/p). CI GI PTVmax PTVmin GTVmin planned – Auto6D 1.62/0.11 1.17/0.24 0.30/0.76 2.05/0.04 1.36/0.17 planned – Auto3D 5.29/<0.01a 1.33/0.19 3.10/<0.01a 5.14/<0.01a 5.09/<0.01a planned – Treat3D 5.34/<0.01a 1.95/0.05 0.76/0.44 5.30/<0.01a 4.30/<0.01a Auto6D – Auto3D 5.29/<0.01a 2.42/0.02 3.09/<0.01a 5.30/<0.01a 5.27/<0.01a Auto6D – Treat3D 5.30/<0.01a 1.02/0.31 0.23/0.82 5.29/<0.01a 4.67/<0.01a a Groups with |Z| > 2.567, p < 0.01 were considered to have a significant difference.
91 Boman et al.: SRT accuracy without 6DOF correction 91 Journal of Applied Clinical Medical Physics, Vol. 17, No. 4, 2016 sum of the rotations are presented in Fig. 3. The multiple lesion cases are shown separately in Fig. 3. These images show a threshold of quadratic sum of the rotations of around 0.5° when all scaled parameters are under the limits of CI < 90%, PTVmin < 90%, and GTVmin < 95% for the Auto3D group. For the quadratic sum higher than 0.5°, most of the scaled parameters are under those limits. In the first example, the PTV is nonspherical in shape, as seen in Fig. 4(a). The rotational setup errors were 0.9°, -1.6°, and 2.7° for the pitch, roll, and yaw rotations, respectively. The distance from the treatment positioning of the patient to the positioning given by Auto6D registration was 2.2 mm, and the distance from the treatment positioning to that given by Auto3D registration was 0.9 mm. In the second example, where there are two separate targets (Fig. 4(b)), the rotational setup errors were -1.8°, -0.3°, and 0.2°, the pitch, roll, and yaw rotations, respectively. The distance from the treatment positioning to the Auto6D positioning was 1.0 mm, and Fig. 3. The percentage difference from the original plan to the CI, PTVmin, and GTVmin values with respect to the quadratic sum of the rotations. Auto6D plans are indicated by a square (□), Auto3D plans by triangle (∆), and Treat3D plans by a circle (○). S indicates single lesion and M is for multiple (2–3) lesions. The multiple lesions are shown by filled symbols for Auto6D/3D plans and by a line (-) for Treat3D plans.
92 Boman et al.: SRT accuracy without 6DOF correction 92 Journal of Applied Clinical Medical Physics, Vol. 17, No. 4, 2016 the distance from the treatment positioning to the Auto3D positioning was 0.8 mm. The isodose lines for the 100% prescription isodose and the dose-volume histograms (DVHs) for PTV and GTV are shown in Fig. 4 for the different cases (planned, Auto6D, Auto3D, Treat3D). It is evident from the DVH figures that the coverages for PTV and GTV are decreased significantly for the Auto3D and Treat3D plans, while for Auto6D the coverages are almost identical to the original plan in both examples. IV. DISCUSSION 6DOF couches have become a common tool in stereotactic brain treatments. Previously, the 6DOF couch has been shown to improve the treatment quality by taking into account daily rotational errors.(18,19) Common to these studies is that they compare the 6D registration with rotational correction results only to those of the 3D registration, which is also shown in this study to significantly impair the dose distribution. We have shown that by appropriate selection of the patient positioning (Auto6D) in brain SRT treatments, a comparable treatment quality is achieved without the correction of rotational errors. The investigated PTVs were very different in shape, size, and location. Thus, the results give the impression that the Auto6D method works best, not only for spherical PTVs, but also for more complex PTV shapes and even for several simultaneous targets (the amount of targets varied from one to three). This is demonstrated in Fig. 4. The examples of two patients (A and B) are shown with PTV outlines (red), isodoses of the prescription dose of 100% for the original plans (orange), Auto6D plans (cyan), Auto3D plans (brown), and Treat3D plans (green). The locations of the patient positionings are also shown for Auto6D (6D), Auto3D (3D), and Treat3D (green +). On the left, the DVHs are shown for each case’s PTVs and GTVs.
93 Boman et al.: SRT accuracy without 6DOF correction 93 Journal of Applied Clinical Medical Physics, Vol. 17, No. 4, 2016 the examples given of nonspherical PTVs. Although there were only two cases and 4 fractions with multiple targets, their results give also an impression that the Auto6D method is the best method for patient positioning amongst the studied methods also, for multiple targets, and that at least it is better than the Auto3D. However, more studies might be needed to make any conclusions for the multiple target cases. Gevaert et al.(18) proposed 0.5° as a threshold angle for the correction of rotation errors to keep the dose coverage of PTV over 95% with the 3D registration. Our data suggest that a 3D match is suboptimal in terms of CI, PTVmin, and GTVmin when the quadratic sum of the rotations exceeds 0.5° (scaled parameters CI < 90%, PTVmin < 90%, and GTVmin < 95% in Fig. 3). For 6D registration, our results suggest that the threshold of rotations could be even larger. The quadratic sum of rotation values of 2°–3° still give adequate plan quality values for scaled CI < 95%, PTVmin < 95%, and GTVmin < 95%. Although the quadratic sum does not describe physically the actual rotation angle, it still seems to provide reasonable one-number measure of overall rotation. These findings are demonstrated to be valid for brain metastases treated with the SRT. The used margins were 1–2 mm. For stereotactic radiosurgery (SRS), the clinical tolerances and practices are different and these findings may not be directly valid for SRS treatments. Besides the setup error, the total treatment accuracy depends also on the patient intrafraction movement(18) and machine isocenter accuracy. These are not considered in this work, but should be kept in mind in estimation of sufficient margins for the SRT. With mask fixation the intrafraction movement is very small, but not negligible. Verbakel et al.(8) found that, for BrainLab mask system, the intrafraction motion was randomly distributed around the zero in every direction and that the maximum deviation was 1 mm and 1° when the average treatment time was 16 min. For shorter treatment times the deviation might be less. In this paper, we have shown the 6D registration to be superior over 3D registration. However, the 6D registration may not be possible in all online registration software, and implementation in daily practice might be difficult. At least on the newest versions of Varian (Varian Medical Systems) and Elekta (Elekta Ltd., Crawley, UK) accelerators and in the ExacTrac (Brainlab) system this option is available. These results apply to Varian 6D registration and need to be confirmed before adaptation to the 6D registration systems of other vendors. V. CONCLUSION We conclude that the outcome of SRT treatments without the option of 6DOF couch could be improved by using 6D registration in the treatment setup. This was demonstrated for typical actual rotation errors below 3° seen in treatment situations. ACKNOWLEDGMENTS This study was financially supported by Pirkanmaa Hospital District research fund (grant number 15013). COPYRIGHT This work is licensed under a Creative Commons Attribution 3.0 Unported License.
94 Boman et al.: SRT accuracy without 6DOF correction 94 Journal of Applied Clinical Medical Physics, Vol. 17, No. 4, 2016 REFERENCES 1. DeVita VT Jr., Lawrence TS, Rosenberg SA. DeVita, Hellman, and Rosenberg’s cancer: principles and practice of oncology, tenth edition. Alphen aan den Rijn, Netherlands: Wolters Kluwer; 2014. 2. Andrews DW, Scott CB, Sperduto PW, et al. Whole brain radiation therapy with or without stereotactic radiosurgery boost for patients with one to three brain metastases: phase III results of the RTOG 9508 randomised trial. Lancet. 2004;363(9422):1665–72. 3. Aoyama H, Shirato H, Tago M, et al. Stereotactic radiosurgery plus whole-brain radiation therapy vs stereotactic radiosurgery alone for treatment of brain metastases: a randomized controlled trial. JAMA. 2006;295(21):2483–91. 4. Rades D, Kueter KD, Pluemer A, Veninga T, Schild SE. A matched-pair analysis comparing whole-brain radiotherapy plus stereotactic radiosurgery versus surgery plus whole-brain radiotherapy and a boost to the metastatic site for one or two brain metastases. Int J Radiat Oncol Biol Phys. 2009;73(4):1077–81. 5. Milano MT, Usuki KY, Walter KA, Clark D, Schell MC. Stereotactic radiosurgery and hypofractionated stereotactic radiotherapy: normal tissue dose constraints of the central nervous system. Cancer Treat Rev. 2011;37(7):567–78. 6. Noël G, Simon JM, Valery CA, et al. Radiosurgery for brain metastasis: impact of CTV on local control. Radiother Oncol. 2003;68(1):15–21. 7. Wiggenraad R, Verbeek-de Kanter A, Kal HB, Taphoorn M, Vissers T, Struikmans H. Dose-effect relation in stereotactic radiotherapy for brain metastases. A systematic review. Radiother Oncol. 2011;98(3):292–97. 8. Verbakel WF, Lagerwaard FJ, Verduin AJ, Heukelom S, Slotman BJ, Cuijpers JP. The accuracy of frameless stereotactic intracranial radiosurgery. Radiother Oncol. 2010;97(3):390–94. 9. Minniti G, Valeriani M, Clarke E, et al. Fractionated stereotactic radiotherapy for skull base tumors: analysis of treatment accuracy using a stereotactic mask fixation system. Radiat Oncol. 2010;5(1). 10. Li G, Yang TJ, Furtado H, et al. Clinical assessment of 2D/3D registration accuracy in 4 major anatomic sites using on-board 2D kilovoltage images for 6D patient setup. Technol Cancer Res Treat. 2015;14(3):305–14. 11. Wurm RE, Erbel S, Schwenkert I, et al. Novalis frameless image-guided noninvasive radiosurgery: initial experience. Neurosurgery. 2008;62(5 Suppl):A11–17. 12. Haertl PM, Loeschel R, Repp N, Pohl F, Koelbl O, Dobler B. Frameless fractionated stereotactic radiation therapy of intracranial lesions: impact of cone beam CT based setup correction on dose distribution. Radiat Oncol. 2013;8:153. 13. Ryken TC, Meeks SL, Pennington EC, et al. Initial clinical experience with frameless stereotactic radiosurgery: analysis of accuracy and feasibility. Int J Radiat Oncol Biol Phys. 2001;51(4):1152–58. 14. Peng JL, Kahler D, Li JG, et al. Characterization of a real-time surface image-guided stereotactic positioning system. Med Phys. 2010;37(10):5421–33. 15. Andrevska A, Knight KA, Sale CA. The feasibility and benefits of using volumetric arc therapy in patients with brain metastases: a systematic review. J Med Radiat Sci. 2014;61(4):267–76. 16. Shi LW, Lai YQ, Lin Q, Ha HM, Fu LR. The effect of flattening filter free on three-dimensional conformal radiation therapy (3D-CRT), intensity-modulated radiation therapy (IMRT), and Volumetric Modulated Arc Therapy (VMAT) plans for metastatic brain tumors from non-small cell lung cancer. Health Phys. 2015;109(1):1–9. 17. Takakura T, Mizowaki T, Nakata M, et al. The geometric accuracy of frameless stereotactic radiosurgery using a 6D robotic couch system. Phys Med Biol. 2010;55(1):1–10. 18. Gevaert T, Verellen D, Engels B, et al. Clinical evaluation of a robotic 6-degree of freedom treatment couch for frameless radiosurgery. Int J Radiat Oncol Biol Phys. 2012;83(1):467–74. 19. Dhabaan A, Schreibmann E, Siddiqi A, et al. Six degrees of freedom CBCT-based positioning for intracranial targets treated with frameless stereotactic radiosurgery. J Appl Clin Med Phys. 2012;13(6):215–25. 20. Paddick I. A simple scoring ratio to index the conformity of radiosurgical treatment plans. Technical note. J Neurosurg. 2000;93(Suppl 3):219–22. 21. Paddick I, Lippitz B. A simple dose gradient measurement tool to complement the comformity index. J Neurosurg. 2006;105(Suppl.):194–201.