A Polynomial-Exponent Model for Calibrating the Frequency Response of Photoluminescence-Based Sensors

Torre Vega, Ángel De La,Medina Rodríguez, Santiago,Segura Luna, José Carlos,Fernández Sánchez, Jorge Fernando

Abstract

This work has been partly supported by the Spanish Ministry of Economy, Industry and Competitiveness (CTQ2014-53442-P, Grant BES-2009-026919 and Torres Quevedo Grants PTQ-15-07922 and PTQ-15-07912), and the CEI BioTic Granada Campus (Project CEIbioTIC14-2015).

Full text

Article A Polynomial-Exponent Model for Calibrating the Frequency Response of Photoluminescence-Based Sensors Angel de la Torre 1, Santiago Medina-Rodríguez 2, Jose C. Segura 1and Jorge F. Fernández-Sánchez 3,* 1 Department Signal Theory, Networking and Communications, University of Granada, 18071 Granada, Spain; [email protected] (A.d.l.T.); [email protected] (J.C.S.) 2By Techdesign S.L., 28500 Arganda del Rey (Madrid), Spain; [email protected] 3Department of Analytical Chemistry, University of Granada, 18071 Granada, Spain *Correspondence: [email protected]; Tel.: +34-958-240451 Received: 8 July 2020; Accepted: 13 August 2020; Published: 18 August 2020   Abstract: In this work, we propose a new model describing the relationship between the analyte concentration and the instrument response in photoluminescence sensors excited with modulated light sources. The concentration is modeled as a polynomial function of the analytical signal corrected with an exponent, and therefore the model is referred to as a polynomial-exponent (PE) model. The proposed approach is motivated by the limitations of the classical models for describing the frequency response of the luminescence sensors excited with a modulated light source, and can be considered as an extension of the Stern–Volmer model. We compare the calibration provided by the proposed PE-model with that provided by the classical Stern–Volmer, Lehrer, and Demas models. Compared with the classical models, for a similar complexity (i.e., with the same number of parameters to be fitted), the PE-model improves the trade-off between the accuracy and the complexity. The utility of the proposed model is supported with experiments involving two oxygen-sensitive photoluminescence sensors in instruments based on sinusoidally modulated light sources, using four different analytical signals (phase-shift, amplitude, and the corresponding lifetimes estimated from them). Keywords: calibration; chemical sensor; photoluminescence; oxygen sensing; frequency response; Stern–Volmer model; Lehrer model; Demas model; polynomial-exponent model 1. Introduction An accurate operation of photoluminescence sensors requires the appropriate calibration of the models describing the relationship between the analytical signal and the concentration of the analyte to be determined [ 1 , 2 ]. The calibration consists in the estimation of the model parameters that fit some calibration data according to some optimization criterion. Therefore, given a model (with M parameters) and some calibration data, the calibration procedure can be mathematically formulated as the multidimensional search of the M parameters minimizing the error between the calibration data and the model [3]. The classical Stern–Volmer model [ 1 ] provides a basic description of the photoluminescence phenomenon. This simple model (with two parameters) describes the luminescent response with a first order differential equation. Even though the Stern–Volmer model provides a qualitative description valid for a wide range of photoluminescence sensors [ 1 , 4 ], this calibration model is insufficient for an accurate description of most of the luminescence instruments because of the heterogeneous distribution of the analyte, illumination and luminescence in the sensing phase [3,5–8]. Sensors 2020,20, 4635; doi:10.3390/s20164635 www.mdpi.com/journal/sensors Sensors 2020,20, 4635 2 of 17 In order to provide a more accurate description of the relationship between the analyte concentration and the instrument response, when the 1-site Stern–Volmer model is insufficient, the 2-site Lehrer [ 9 ] and Demas [ 6 , 7 , 10 , 11 ] models (which are extensions of the former) are commonly used for calibration. Additionally, the Demas model provides scalability, and can be extended to more than 2 sites [ 11 , 12 ]. More complex models have also been proposed in the literature, in order to accurately describe the interaction between the quencher and the luminophore in the sensing phase, for example, by modeling the diffusion of the quencher in the sensor volume [ 8 , 13 , 14 ], the spatial distribution of the luminescence lifetime [ 12 , 15 , 16 ], or the dynamics of the luminescence [ 17 ]. Such models involve complex mathematical description (including integrals for the volume of the luminescent material) and a large number of parameters to be fitted. Even though these models are useful to understand specific aspects of the quenched photoluminescence, their applicability for calibrating instruments is limited due to the number of parameters to be fitted [18]. The model complexity is a relevant aspect to be considered for calibrating an instrument. A model with more parameters to be fitted provides a more accurate fitting of the calibration data. However, the calibration data are always affected by noise, and if the number of parameters is too large or the calibration data are insufficient (in terms of statistical robustness), the calibration curve will be affected by over-fitting (low calibration error but large instrumental error because the calibration curve is over-adapted to the errors in the calibration data) [ 19 , 20 ]. Additionally, fitting a complex model implies a multidimensional search of the M parameters of the calibration curve, and the search and convergence becomes more difficult as M increases [ 21 ]. In that sense, a good calibration model is one providing a good accuracy with a low number of parameters [2]. In this work, we propose a new model to describe the relationship between the analytical response and the analyte concentration in chemical sensors based on photoluminescence. In its basic form, the proposed model can be considered an extension of the 2-parameter Stern–Volmer model, by including an exponent to the analytical signal as an additional parameter to be fitted (thus yielding to a 3-parameter model). The proposed model can be scaled and admits a simple extension to M parameters (with M arbitrarily large), which provides flexibility of the model depending on the availability of calibration data. For M parameters, the model is defined with a M− 2 degree polynomial and an exponent (its calibration requires M− 1 coefficients for the polynomial and an additional parameter for the exponent), and therefore the proposed model is referred to as the polynomial-exponent model. The proposed model is motivated by the limitations of the classical models, particularly for the description of photoluminescence sensors using modulated light sources [ 22 , 23 ]. The polynomial-exponent model is compared in this paper with the classical models commonly used for calibrating photoluminescence sensors: the Stern–Volmer, the Lehrer, and the 2-site Demas models (requiring 2, 3, and 4 parameters, respectively). The proposed and the classical models have been evaluated in experiments involving two oxygen sensing phases in two different photoluminescence measuring systems based on sinusoidally modulated light sources. The experimental results show that the proposed polynomial-exponent model provides a good trade-off between accuracy and number of parameters. It is applicable to the amplitude, phase-shift, or lifetime measurements of luminescence, and it is thought to be of potential utility for enzymatic biosensors coupled with the optical oxygen transducer, and related chemical sensors employing quenching/enhancement of luminescence. 2. Limitations of the Classical Calibration Models The polynomial-exponent model that we propose in this manuscript is motivated by the limitations of the classical models conventionally considered for calibrating photoluminescence sensors when they are applied to measurements in the frequency domain. In this section, we analyze in detail the classical calibration models and their limitations when applied in phase-resolved photoluminescence. Sensors 2020,20, 4635 3 of 17 2.1. Photoluminescence Response with Modulated Excitation 2.1.1. The Stern–Volmer Equation Luminescence is the result of a radiative deactivation of the molecules of a luminescent material from an excited state to the fundamental state. The simplest first order linear photoluminescence model describes the population of the excited state N(t) linearly increasing with the illumination xexc(t) and linearly decreasing according to the radiative and non-radiative deactivation constants Γr and Γnr [1]: d N(t) d t =a0xexc(t)−(Γr+Γnr)N(t), (1) where t is the time variable and a0 is a constant that relates the illumination used as excitation with the increase of the excited state population in the luminescent material. This is a non-homogeneous first order linear differential equation characterized by a lifetime τq inversely proportional to the sum of the deactivation constants. In the presence of a quencher, the simplest quenching model suggests a linear increase of the non-radiative deactivation constant with the quencher concentration C , thus yielding to the Stern–Volmer equation: τq=1 Γr+Γnr =1/Γr 1+Γnr/Γr =τ0 1+kC (2) where τ0 is the inverse of Γr (i.e., the lifetime at null quencher concentration) and k is the Stern–Volmer constant [ 1 , 4 , 24 ]. The photoluminescent emission xem(t) is proportional to the radiative decay, i.e., to the product Γr·N(t) , and therefore the photoluminescence is described by a similar differential equation: d xem(t) d t =a1xexc(t)−xem(t) τq(3) where the coefficient a1provides the relation between excitation and emission. 2.1.2. Frequency Response of the Stern–Volmer Model In the case of an illumination source modulated with a sinusoidal or periodic signal, the response of the photoluminescent sensor is better described through its frequency response. The differential equation is transformed to the frequency domain [ 25 – 27 ] by substitution of the derivative operator by jω (where j is the imaginary unity, ω= 2 πf is the angular frequency, in radians per second, and f is the frequency in Hz): jωXem(jω) = a1Xexc(jω)−1 τq Xem(jω)(4) and the frequency response can be obtained from this equation: H(jω) = Xem(jω) Xexc(jω)=τqa1 1+jωτq(5) which is usually rewritten as: H(jω) = M0 τq τ0 1 1+jωτq(6) where M0 is the modulation-factor at null quencher concentration and low frequency (i.e., the ratio between the emission and excitation at these conditions). In phase-resolved luminescence spectroscopy, the frequency response describes the relationship between the excitation and the emission for each frequency component in the excitation signal. At each frequency, H(jω) is a complex number, Sensors 2020,20, 4635 4 of 17 with modulus and argument, the modulus representing the ratio between the amplitudes of the emission and excitation sinusoidal components of frequency ω , and the argument representing the phase-shift between both sinusoidal components [27,28]. According to this linear first order model, the modulation-factor predicted for a given concentration Cand modulation frequency ωis [1,22]: m(C,jω) = q|H(jω)|2=M0 τq(C) τ0 1 q1+ (ωτq(C))2(7) and the phase-shift is [1,22]: φ(C,jω) = −arctan(ωτq(C)) (8) 2.1.3. Analyte Determination with Modulated Illumination Sources From the previous equations, the lifetime τq can be estimated either from the modulation-factor or from the phase-shift [22,23]: τφ(C) = −tan(φ(C,jω)) ωτm(C) = τ0 m(C,jω)/M0 p1−(ωτ0)2(m(C,jω)/M0)2(9) However, in phase-resolved spectroscopy, the modulation-factor at null concentration is difficult to be measured at null frequency, while it is easier to be estimated at the modulation frequency. The modulation-factor at the modulation frequency ωfor null concentration is given by: m0(jω) = m(0, jω) = M0 p1+ (ωτ0)2(10) and therefore the following expression is more practical for the estimation of the lifetime based on the modulation-factor: τm(C) = τ0 m(C,jω)/m0(jω) p1+ (ωτ0)2[1−(m(C,jω)/m0(jω))2](11) (this expression requires the parameter τ0, usually estimated as τφat null concentration). In summary, a phase-resolved photoluminescence sensor (i.e., using a modulated light source as excitation), described by a first order linear model, can easily be calibrated with Equations (9), (11) and (2). At a given frequency ω and concentration C , the modulation-factor m(C , jω) and the phase shift φ(C , jω) can be measured from an analysis of the excitation and emission signals xexc(t) and xem(t) . From them, the modulation-factor and the phase-shift based lifetimes τm(C) and τφ(C) are estimated with Equations (9) and (11). By collecting measurements at different concentrations, the Stern–Volmer model can be calibrated using Equation (2), i.e., the parameters of the Stern–Volmer model ( τ0 and k ) providing the best fitting of the calibration data set can be estimated. After the calibration, from a new measurement (with an excitation signal modulated at frequency ω ) obtained at an unknown concentration, the modulation-factor and phase-shift can be measured, the corresponding lifetimes can be derived (either τm or τφ ), and using the calibrated Stern–Volmer model, the concentration can be estimated, providing the analyte determination C(τq). As can be noted from the equations, for the first order model, the luminescence lifetime only depends on the concentration, and the lifetime estimations obtained either from phase-shift or modulation-factor and at different frequencies should be identical. Due to noise (mainly that affecting the emission signal) and the error propagation, the accuracy of a phase-resolved photoluminescence sensor strongly depends on the modulation frequency, and multifrequency sensors with an appropriate combination of τm and τφ estimated at several frequencies provide the best results for the complete range of concentrations [22,23,29,30]. Sensors 2020,20, 4635 5 of 17 2.2. Limitations of the Stern–Volmer Model The Stern–Volmer model provides a very good qualitative description of a wide range of photoluminescent sensors. The accuracy provided by this model is acceptable for some sensing phases, but it is insufficient for most of them, due to deviations of the photoluminescence from the first order linear model [ 1 , 3 , 4 , 11 ]. These deviations are associated with the presence of different excitation and/or deactivation mechanisms, the distribution of the quencher in the volume of the luminescent material, biased measurements of the modulation-factor or phase-shift due to instrumental configuration, etc. Taking into account the mathematical formulation of the model, two different situations should be considered: (a) the relationship between the quencher concentration and the non-radiative deactivation constant is not linear (i.e., Equation (2) is insufficient); or (b) the order of the differential equation is higher than one. From the mathematical perspective, both situations are completely different, since in the first case, the photoluminescence sensor is a first order system (i.e., with a monoexponential evolution of the luminescent response, with a well defined lifetime, and providing identical lifetime estimations from modulation-factor or from phase-shift and at whatever modulation frequency), while in the second case there is not a well-defined luminescence lifetime, the estimated values depend on the estimation procedure (so they are referred to as “apparent lifetimes”) and therefore the calibration parameters depend on the sensor configuration (modulation frequency, whether the apparent lifetime is estimated either from the modulation-factor or from the phase-shift, etc.). For the first situation (first order differential equation but nonlinear relationship between quencher concentration and deactivation constant), the Lehrer model is an extension of the Stern–Volmer model [1,9]: τq=τ0x 1+kC + (1−x)(12) where a proportion x of the luminescent material is assumed to be affected by the quencher while the rest, 1 −x , is not affected. Compared with the Stern–Volmer model, the 3-parameter Lehrer model increases the flexibility in the description of the τq(C) curves, providing a better calibration for many photoluminescence sensors. The Demas model can also be considered an extension of the Stern–Volmer model [1,7]: τq=τ0x 1+k1C+(1−x) 1+k2C(13) and represents two independent quenching contributions, each one with a different Stern–Volmer constant. The 4-parameter Demas model increases the flexibility with respect to the Stern–Volmer and Lehrer models and provides more accurate calibrations. Additionally, the Demas model is scalable (and can be extended to more than two sites by including more terms): τq=∑ n τ0,n 1+knC(14) where τ0,nis the contribution of the n-th site to the lifetime at null quencher concentration. For the second situation (higher order differential equation), a multi-exponential photoluminescence model is proposed [1,22,23], providing the following frequency response: H(C,jω) = ∑ n M0,n τq,n τ0,n 1 1+jωτq,n(15) where the index n represents each mono-exponential process contributing to the multi-exponential response; for each process n , M0,n is the modulation factor at low frequency and null concentration, τ0,n is the lifetime at null concentration, and the relation between the concentration and the lifetime for each mono-exponential process is described by a Stern–Volmer equation [1,24]: Sensors 2020,20, 4635 6 of 17 τq,n(C) = τ0,n 1+knC(16) with its corresponding Stern–Volmer constant kn . In this model, each mono-exponential process requires three parameters to be calibrated ( M0,n , τ0,n , and kn ). By transforming the frequency response into the corresponding differential equation, it can be demonstrated that an N -process model is equivalent to a N-order differential equation. Interestingly, the Demas multi-site first order model and the multi-exponential model provide an equivalent equation for low frequencies. Effectively, if ωτq,n 1 for all the processes, the dependence with frequency disappears, and Equation (15) can be rewritten as: H(C) = ∑ n M0,n 1+knC(17) and, since the modulation-factor is proportional to the lifetime at low frequency (according to Equation (11)), both models are identical at low frequency. However, the differences between both models become significant for phase-resolved photoluminescence sensors using higher modulation frequencies (or for multifrequency photoluminescence). Experimental results suggest that deviation from the simple Stern–Volmer model is usually better explained with a multiexponential model than with a multi-site quenching of the lifetime, since the apparent lifetimes estimated at different frequencies, either from modulation-factor or phase-shift, provide inconsistent estimations [ 23 , 29 – 32 ], while a multiexponential model provides a globally consistent description of the photoluminescence response at different modulation frequencies [23]. Even though the multiexponential model is preferable for a global description of the photoluminescence multifrequency response, the sensor calibration is usually performed with the Demas model and using apparent lifetimes (either from modulation-factor or phase-shift based and estimated at a single frequency) because the combination of the different measured parameters in a single analyte determination is not very common and it is a relatively recent proposal [ 23 , 29 , 30 ]. In addition, the Demas model is expected to provide a reasonable calibration when it is applied to the apparent lifetimes because at moderate frequencies the relation between m(C , jω) and τm(C) , or between φ(C , jω) and τφ(C) , is close to linear, and usually very high modulation frequencies are not applied in phase-resolve photoluminescence (because the emission amplitude is low at high modulation frequencies, and the response would be too noisy for an accurate analyte determination) [23]. 2.3. Limitations of the Demas Model According to the previous discussion, even though the Demas multi-site description is not appropriate for a multiexponential photoluminescence sensor, at very low frequencies, it is perfectly consistent (since it is mathematically equivalent to the multiexponential model). On the other hand, if the deviation from the Stern–Volmer model is moderate, or at moderate frequencies (relative to the involved lifetimes), the Demas model provides a reasonable calibration, usually with accuracy significantly improved with respect to the Stern–Volmer model (because the inconsistency can be compensated by the model flexibility). However, the inconsistency of the Demas model limits its accuracy in multiexponential photoluminescence sensors excited with modulated light sources. Additionally, the instrumental bias (affecting the modulation-factor and the phase-shift measurements) and its propagation to the apparent lifetimes also contribute to limiting the accuracy of the Demas-based calibration. An accurate estimation of the modulation-factor and phase-shift requires the acquisition of the excitation and emission signals xexc(t) and xem(t) and the subsequent digital signal processing [ 1 , 27 , 28 ]. The illumination is usually provided by an electrical signal which feeds a driver circuit connected to the illumination source, and the excitation signal is acquired by Sensors 2020,20, 4635 7 of 17 sampling the electrical signal used for the illumination system. The acquisition of the excitation signal is not difficult since this signal is usually provided by a signal generator, with large amplitude and high signal-to-noise ratio. However, the acquisition of the emission signal in a photoluminescence sensor is usually affected by several problems. The photoluminescence response is converted into an electrical signal by a photodetector (for example, a photodiode, a phototransistor, or a photomultiplier tube), and the electrical signal has to be amplified and sampled. Depending on the intensity of the photoluminescence response and the required sensitivity, large gains are usually required in the transduction, and usually the transductors and the amplification circuitry cause linear and nonlinear distortions (group delay, frequency response associated with the instrument including gain and phase-shift distortions, nonlinear distortion due to the transducer response, etc.). A significant bias in the estimation of the phase-shift is very common in photoluminescence sensors (associated with group delay in the photodetector and frequency response of the photodetector and preamplifier). The main source of bias in the amplitude measurements would be associated with the polarization of the electronic and optoelectronic components, but in a linear system it does not affect frequency based measurements (only at null frequency). Therefore, the bias in the modulation-factor is expected to be moderate and would be associated with nonlinear effects in the photodetector. These effects can be estimated, calibrated, and compensated in order to remove them from the frequency response of the photoluminescence sensor and therefore to obtain a more consistent calibration of the sensor using a Demas or a multiexponential model. However, an accurate calibration and compensation of the instrumentation effects for the operation conditions (or a join calibration of both the instrumentation and the photoluminescent subsystems) are not easy [1,24,27,28]. 3. The Polynomial-Exponent Model The calibration based on the Demas model presents two advantages: firstly, the model is based on the physics of the sensing phase and therefore it provides a physical interpretation of the calibration parameters. On the other hand, it is a scalable model. However, it also presents some disadvantages: firstly, the consistency of the model is limited for sensors that deviate from the Stern–Volmer model operating at moderate frequency. Secondly, an instrumental bias would increase the inconsistency of the model. Finally, the calibration of a N -sites model (requiring the optimization of 2 N parameters) can be simplified to a N -dimensional search [ 2 ], but it becomes unpractical for large N , thus limiting the scalability of the model. Taking into account the advantages of the Demas-based calibration, and also its limitations (associated with inconsistencies when the luminescent sensing phases requires a differential equation of order higher than one, or because of instrumental nonlinearity or bias), we propose to describe the relationship between the analytical signal φ (where φ would be either the phase-shift, the modulation-factor or the corresponding lifetimes) and the concentration C(φ) with a polynomial function applied to φα, i.e., applied to the analytical signal corrected with an exponent: C(φ) = p ∑ n=0 anφnα=a0+a1φα+a2φ2α+. . . +apφpα(18) The calibration of this polynomial-exponent model (which will be referred to as PE model) requires the estimation of p+ 2 parameters, where p is the polynomial degree ( p+ 1 parameters for the polynomial coefficients and an additional one for the exponent α). Even though the polynomial-exponent model is not intended for a physical description of the photoluminescence (but for the calibration of photoluminescence sensors), it provides a very simple estimation of the sensor response at null concentration φ0 and the sensitivity at a given conentration, KC. These parameters are defined as: φ0≡lim C→0φKC≡1 φ(C) ∂φ(C) ∂C K0≡lim C→0 1 φ(C) ∂φ(C) ∂C (19) Sensors 2020,20, 4635 8 of 17 and, for the PE model, they are obtained, respectively, from the polynomial roots and from the first derivative of the polynomial. For example, in the case of the first degree PE model (PE1), the parameters providing the physical description of the model are: φ0=−a0 a11/α K0= 1 a0α KC= 1 αa1φα C (20) and those for the PE2 model (second degree) are: φ0=  −a1+qa2 1−4a0a2 2a2  1/α K0= 1 αa1φα 0+2αa2φ2α 0 KC= 1 αa1φα C+2αa2φ2α C (21) (φCbeing the analytical signal at the concentration C). In fact, if the analytical signal is the lifetime, the first degree PE model can be considered as an extension of the Stern–Volmer model: C(τ) = a0+a1τατ(C) = −a1/a0 1−C/a0−1/α τSV (C) = τ0 1+kC (22) and the comparison of both expressions of the lifetime reveals that the Stern–Volmer model is a particular case of the PE1 model for α=− 1, where the lifetime at null concentration is τ0=−a1/a0 and the sensitivity (or Stern–Volmer constant) is K=−1/a0. The scalability of the PE model is easier than in the Demas model because the estimation of the polynomial coefficients for a given exponent α does not require any search (but a one-step matrix inversion), and the optimization of the p -degree PE-model requires a one-dimensional search (for α ) independently of the polynomial degree (details about the PE-model calibration are provided in Section 1 of the Supplementary Materials). In this work, we evaluate 1st and 2nd degree PE models (PE1 and PE2) requiring the estimation of three and four parameters, respectively. Additionally, in order to analyze the role of the exponent included in the model, we also evaluate the PE2 model with the exponent constrained to α=− 1, i.e., a 3 parameter model which will be referred to as P2 model (note that P1 model corresponds to the Stern–Volmer model, and, therefore, the P2 model could be considered another three-parameter extension of the Stern–Volmer model). Octave/MatLab functions for the calibration of the P2, PE1, and PE2 models are provided in Section 2 of the Supplementary Materials. We also evaluate the Stern–Volmer, Lehrer, and Demas models as reference. These reference models, require the estimation of 2, 3, and 4 parameters, respectively. In these models, the response at null concentration, φ0 , is a parameter of the model, and the sensitivity at C =0 is K0=k , K0=x·k and K0=x·k1+ ( 1 −x)·k2 for the Stern–Volmer, Lehrer, and Demas models, respectively. 4. Experimental Design The accuracy of the proposed PE model for calibrating photoluminescence-based sensors has been evaluated with experiments involving two different sensing phases for oxygen detection. In these experiments, the phase-shift or the modulation-factor were measured by phase-resolved luminescence spectroscopy in the frequency domain. The first sensing phase, Platinum(II) 5,10,15,20-meso-tetrakis-(2,3,4,5,6-pentafluorophenyl)- porphyrin immobilized in polystyrene (PtTFPP/PS), is an oxygen-sensing film coated at the end of an optical fibre. The photoluminescence sensor was designed for measuring oxygen in gas phase ( pO2 ) in the range 0.5–20 kPa. The sensing film was illuminated with an ultraviolet LED excited with a sinusoidal signal at a frequency of 5145 Hz. The photoluminescent emission was transduced and amplified with a photomultiplier tube and a low-noise preamplifier. Both the excitation and emission electrical signals were digitized with an oscilloscope, and digital signal processing implementing the Sensors 2020,20, 4635 9 of 17 I/Q method was applied to obtain the modulation-factor and the phase-shift for each measurement. Details about this sensing film, the instruments involved in the acquisition of the excitation and emission signals, and the I/Q method for estimating the modulation-factor and phase-shift from the recorded signals can be found in our previous works [23,31]. The second sensing phase (N1008-AP200/19) is an iridium-based oxygen-sensing dye, Ir(2-(2,4-difluorophenyl) pyridine) 2(4,7-diphenyl-1, 10-phenanthroline), immobilized into AP200/19, a nanostructured aluminium oxide-hydroxide solid support. The sensor using this sensing-phase was designed for measuring oxygen in gas phase at low partial pressures (in the range 0.25–10 kPa). The sensing film was illuminated with an ultraviolet LED excited with a sinusoidal signal at a frequency of 30,100 Hz. The photoluminescent emission was transduced and amplified with a photomultiplier tube and a low-noise preamplifier. Both the excitation and emission electrical signals were processed with a dual-phase lock-in amplifier (LIA) (SR830, Stanford Research Systems, Sunnyvale, CA, USA). The LIA instrument provided the phase-shift between the excitation and emission signals. Details about the N1008-AP200/19 sensing film, the involved instruments, and the procedure for estimating the phase-shift can be found in our previous works [33,34]. Six different calibration experiments were defined with the measurements obtained with these sensors: experiments 1, 2, 3, and 4 involve the PtTFPP/PS sensor, and provide the concentration ( O2 partial pressure, kPa) using the phase-shift φ , the phase-shift based lifetime τφ , the modulation-factor m , or the modulation-factor based lifetime τm as analytical signal, respectively. Experiments 5 and 6 involve the N1008-AP200/19 sensor and provide the O2 partial pressure using the phase-shift φ and the corresponding lifetime τφ , respectively. Obviously, the pair of calibration experiments involving a measurement (either φ or m ) and the corresponding lifetime are correlated, since they are based in the same primary measurements. For the experiments 1, 2, 3, and 4, the analytical signal was obtained at different oxygen concentrations, with 250 measurements registered at each concentration. For the experiments 5 and 6, 100 measurements are available at each concentration. The data acquisition was performed with increasing oxygen concentration and leaving a sufficient time at each new concentration in order to allow the stabilization of the gas station and the sensing phase before the data recording. For each experiment, the measurements were randomly assigned to disjoint calibration and evaluation subsets (125 measurements at each concentration for the experiments 1, 2, 3, and 4; fifty measurements at each concentration for the experiments 5 and 6). The data in the calibration datasets were averaged for the calibration procedures (in order to accurately analyze the residual of the fitting curves). The data in the evaluation datasets were used individually in order to evaluate the impact of the calibration procedures over the accuracy in the oxygen determination. For each experiment, tables with the mean and standard deviations of the analytical signals at each concentration (for the calibration and evaluation datasets) are provided in Section 3 of the Supplementary Materials. 5. Results and Discussion The calibration dataset for each experiment was used for calibrating the Stern–Volmer (SV), Lehrer (L), Demas (D), 2nd degree polynomial (P2), and 1st and 2nd degree polynomial-exponent (PE1 and PE2) models. The optimization criterion applied for calibration was the minimization of the relative error in the concentration (in this case pO2 ) [ 2 ]. The SV and P2 optimizations are obtained directly from the calibration dataset (these models do not require any iterative procedure for the calibration). Since the L, PE1, and PE2 optimizations require a one-dimensional search, the calibration was configured for 100 iterations for these models. The D optimization was performed with 500 iterations (since it involves a two-dimensional search). The SV, L, and D models have been calibrated with the Octave/MatLab functions provided in our previous work [ 2 ]. The P2, PE1, and PE2 models have been calibrated with the Octave/MatLab functions provided in Section 2 of Supplementary Materials. Sensors 2020,20, 4635 16 of 17 10. Eftink, M.R.; Ghiron, C.A. Exposure of tryptophanyl residues in proteins. Quantitative determination by fluorescence quenching studies. Biochemistry 1976,15, 672–680. [CrossRef] 11. Carraway, E.R.; Demas, J.N.; DeGraff, B.A.; Bacon, J.R. Photophysics and photochemistry of oxygen sensors based on luminescent transition-metal complexes. Anal. Chem. 1991,63, 337–342. [CrossRef] 12. Mills, A. Optical sensors for oxygen: a log-gaussian multisite-quenching model. Sens. Actuators Chem. 1998,51, 69–76. [CrossRef] 13. Bossi, M.L.; Daraio, M.E.; Aramendía, P.F. Luminescence quenching of Ru(II) complexes in polydimethylsiloxane sensors for oxygen. J. Photochem. Photobiol. Chem. 1999,120, 15–21. [CrossRef] 14. Ogurtsov, V.I.; Papkovsky, D.B. Modeling of luminescence-based oxygen sensors with non-uniform distribution of excitation and quenching characteristics inside active medium. Sens. Actuators Chem. 2003,88, 89–100. [CrossRef] 15. Draxler, S.; Lippitsch, M.E.; Klimant, I.; Kraus, H.; Wolfbeis, O.S. Effects of Polymer Matrixes on the Time-Resolved Luminescence of a Ruthenium Complex Quenched by Oxygen. J. Phys. Chem. 1995,99, 3162–3167. [CrossRef] 16. Mills, A. Response characteristics of optical sensors for oxygen: models based on a distribution in tau0 or kq. Analyst 1999,124, 1301–1307. [CrossRef] 17. Badocco, D.; Mondin, A.; Fusar, A.; Pastore, P. Calibration Models under Dynamic Conditions for Determining Molecular Oxygen with Optical Sensors on the Basis of Luminescence Quenching of Transition-Metal Complexes Embedded in Polymeric Matrixes. J. Phys. Chem. 2009 ,113, 20467–20475. [CrossRef] 18. Trettnak, W.; Gruber, W.; Reininger, F.; Klimant, I. Recent progress in optical oxygen sensor instrumentation. Sens. Actuators Chem. 1995,29, 219–225. [CrossRef] 19. Hawkins, D.M. The Problem of Overfitting. J. Chem. Inf. Comput. Sci. 2004,44, 1–12. [CrossRef] 20. Faber, N.M.; Rajkó, R. How to avoid over-fitting in multivariate calibration—The conventional validation approach and an alternative. Anal. Chim. Acta 2007,595, 98–106. [CrossRef] 21. Fletcher, R. Practical Methods of Optimization, 2nd ed.; John Wiley & Sons Ltd.: Chichester, UK, 2008. 22. Torre-Vega, A.; Medina-Rodríguez, S., Medina-Rodríguez, C., Fernández-Sánchez, J.F. Chapter 6: Progress in Phosphorescence Lifetime Measurement Instrumentation for Oxygen Sensing. In Quenched-Phosphorescence Detection of Molecular Oxygen: Applications in Life Sciences; Papkovsky, D.B., Dmitriev, R.I., Eds.; RSC Detection Science Series No. 11; The Royal Society of Chemistry: London, UK, 2018; pp. 117–144. 23. Medina-Rodríguez, S.; de la Torre-Vega, A.; Sainz-Gonzalo, F.J.; Marín-Suárez, M.; Elosúa, C.; Arregui, F.J.; Matias, I.R.; Fernández-Sánchez, J.F.; Fernández-Gutiérrez, A. Improved Multifrequency Phase-Modulation Method That Uses Rectangular-Wave Signals to Increase Accuracy in Luminescence Spectroscopy. Anal. Chem. 2014,86, 5245–5256. [CrossRef] 24. Schäferling, M. The Art of Fluorescence Imaging with Chemical Sensors. Angew. Chem. Int. Ed. Engl. 2012,51, 3532–3554. [CrossRef] 25. Proakis, J.G; Manolakis, D.G. Digital Signal Processing—Principles, Algorithms and Applications, 4th ed.; Pearson Prentice Hall: Upper Saddle River, NJ, USA, 2007. 26. Sundararajan, D. The Discrete Fourier Transform: Theory, Algorithms and Applications; World Scientific Publishing Co. Pte. Ltd.: Singapore, 2001. 27. Cajlakovic, M.; Bizzarri, A.; Konrad, C.; Voraberger, H. Optochemical Sensors Based on Luminescence. In Encyclopedia of Sensors; Grimes, C.A., Dickey, E.C., Pishko, M.V., Eds.; American Scientific Publishers: Valencia, CA, USA, 2006; Volume 7, pp. 291–313. 28. Langer, P.; Müller, R.; Drost, S.; Werner, T. A new method for filter-free fluorescence measurements. Sens. Actuators Chem. 2002,82, 1–6. [CrossRef] 29. Medina-Rodríguez, S.; Medina-Rodríguez, C.; de la Torre-Vega, A.; Segura-Luna, J.C.; Mota-Fernández, S.; Fernández-Sánchez, J.F. Real-time optimal combination of multifrequency information in phase-resolved luminescence spectroscopy based on rectangular-wave signals. Sens. Actuators Chem. 2017 ,238, 221–225. [CrossRef] 30. Medina-Rodríguez, C.; Medina-Rodríguez, S.; de la Torre-Vega, A.; Fernández-Gutiérrez, A; Fernández-Sánchez, J.F. Direct estimation of the standard error in phase-resolved luminescence measurements: Application to an oxygen measuring system. Sens. Actuators Chem. 2016,224, 521–528. Sensors 2020,20, 4635 17 of 17 31. Medina-Rodríguez, S.; de la Torre-Vega, A.; Fernández-Sánchez, J.F.; Fernández-Gutiérrez, A. An open and low-cost optical-fiber measurement system for the optical detection of oxygen using a multifrequency phase-resolved method. Sens. Actuators Chem. 2013,176, 1110–1120. [CrossRef] 32. Medina-Rodríguez, S.; de la Torre-Vega, A.; Fernández-Sánchez, J.F.; Fernández-Gutiérrez, A. Evaluation of a simple PC-based quadrature detection method at very low SNR for luminescence spectroscopy. Sens. Actuators Chem. 2014,192, 334–340. [CrossRef] 33. Marín-Suarez del Toro, M.; Fernández-Sánchez, J.F.; Baranoff, E.; Nazeeruddin, M.K.; Graetzel, M.; Fernández-Gutiérrez, A. Novel luminescent Ir(III) dyes for developing highly sensitive oxygen sensing films. Talanta 2010,82, 620–626. [CrossRef] 34. Medina-Rodríguez, S.; Marín-Suárez, M.; Fernández-Sánchez, J.F.; de la Torre-Vega, A.; Baranoff, E.; Fernández-Gutiérrez, A. High performance optical sensing nanocomposites for low and ultra-low oxygen concentrations using phase-shift measurements. Analyst 2013,138, 4607–4617. [CrossRef] c 2020 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).